Aspirator Template
Aspirator Template
Calibration Certificate
ATEC certifies that the above equipment has been calibrated in accordance with our calibration procedures using
standard equipment calibrated and accredited directly or indirectly to the US national institute of standards and
technology (NIST).
This certificate must not be reproduced without written approval of ATEC.
Any repair, miss use, etc…without recalibration, then ATEC will not be liable to any consequences that may cause
serious problems to the patients.
NOTE
Safety testing checks the electrical safety of the equipment and is important to prevent a shock to the patient or
the staff. By following the IEC 60601 and its limits, we make sure that the medical electrical equipment is safe.
Revised by Approved by
ENG/ Tamer Farid ENG/ Tamer Tawfik
1
ATEC Biomedical Calibration Lab
Calibration Certificate
2
ATEC Biomedical Calibration Lab
Calibration Certificate
PHYSICAL CONDITION
Status Status
Parameter Parameter
Pass Fail Pass Fail
Chassis / Housing PASS Displays / indicators PASS
Carts / casters PASS Audible signal PASS
Line cord PASS Controls PASS
AC plug PASS Accessories PASS
SAFETY TESTS
CLASS TYPE
CURRENT CONSUMPTION (A) MAINS VOLTAGE
L1-L2 (V)
PROTECTIVE EARTH RESISTANCE L1-EARTH (V)
TEST CURRENT (A) L2-EARTH (V)
RESISTANCE (˂0.5Ω) ISOLATION RESISTANCE(Ω) OVER
ENCL. LEAKAGE CURRENT (µA) EARTH LEAKAGE CURRENT (µA)
NORM. POL LIMIT (100) PASS/FAIL NORM. POL LIMIT (500) PASS/FAIL
PASS PASS
NO L2 LIMIT (500) PASS/FAIL NO L2 LIMIT (1000) PASS/FAIL
PASS PASS
NO EARTH LIMIT (500) PASS/FAIL
PASS
REV. POL LIMIT (100) PASS/FAIL REV. POL LIMIT (500) PASS/FAIL
PASS PASS
NO L2 LIMIT (500) PASS/FAIL NO L2 LIMIT (1000) PASS/FAIL
3
ATEC Biomedical Calibration Lab
Calibration Certificate
PASS PASS
NO EARTH LIMIT (500) PASS/FAIL
PASS
4
ATEC Biomedical Calibration Lab
Calibration Certificate
PERFORMANCE TEST
CHANNEL: 1
PRESSURE ACCURACY
No. UNIT SET VALUE READING REFERENCE PASS/FAIL
1 mmHg ± 10% PASS
2 mmHg ± 10% PASS
3 mmHg ± 10% PASS
4 mmHg ± 10% PASS
5 mmHg ± 10% PASS
CHANNEL: 2
PRESSURE ACCURACY
No. UNIT SET VALUE READING REFERENCE PASS/FAIL
1 mmHg 30 30 ± 10% PASS
2 mmHg ± 10% PASS
3 mmHg ± 10% PASS
4 mmHg 50 50 ± 10% PASS
5 mmHg ± 10% PASS
REMARKS