Module-5-Sensors in Analytical Techniques
Module-5-Sensors in Analytical Techniques
Gangadhar B Bagihalli
Syllabus:
Electrode System: Introduction, types of electrodes. Ion selective electrode – definition, construction,
working and applications of glass electrode. Determination of pH using glass electrode. Reference
electrode - Introduction, calomel electrode – construction, working and applications of calomel
electrode. Concentration cell – Definition, construction and Numerical problems.
Introduction:
Note:1
1
Dr. Gangadhar B Bagihalli
Note:2
Electrochemical cells: “An electrochemical cell is a device which converts chemical energy into
electrical energy or electrical energy into chemical energy”
Note:3
1. Galvanic cells-are devices which convert chemical energy into electrical energy. Example-
Dry cell, Lead-acid cell, Ni-Cd cell etc.
2. Electrolytic cells-are devices which convert electrical energy into chemical energy.
Galvanic cells: The Daniel cell is an example of a galvanic cell. It consists of a zinc rod dipped
in a 1M ZnSO4.This forms one half cell. A copper rod dipped in a 1M CuSO4 constitutes the
other half cell. These two electrodes are connected internally by salt bridge and externally
connected through a voltmeter. The Daniel cell is based on the red ox reaction.
-
e
Salt bridge
+
- Copper
Zinc
_
Anode Cathode
ZnSO4 CuSO4
Solution Solutio
n
At the zinc electrode zinc goes into the solution as Zn2+ ions liberating electrons at zinc
electrode oxidation takes place
Zn Zn2+ + 2e-
At copper electrode reduction of Cu2+ ions takes place with the deposition of metallic copper on
copper rod by gaining electrons, ic at copper electrode reduction takes place
2
Dr. Gangadhar B Bagihalli
Cu2+ + 2e- Cu
By IUPAC convention the electrode at which oxidation takes place is anode and the electrode
where the reduction occurs is cathode.
It can be defined as “the potential developed at the interface between the metal and
solution, when it is in contact with a solution of its own ions”
Standard electrode potential is the electrode potential when the electrode is in contact with
a solution of unit concentration; at 298K. If the electrode involves a gas then the gas at one
atmosphere pressure
EMF of a cell:
EMF of a cell: “Emf is the potential difference between the two electrodes of a galvanic cell
which causes the flow of current from one electrode to other”
The EMF of a cell represents the driving forces for cell reactions. It can be represented by
thermodynamic relation.
∆G= -nFE
Where, ∆G- Free energy change accompanying during cell reaction, n- number of electrons
transferred during cell reaction and F- Faraday constant. The cell reactions are spontaneous when
∆G is –ve.
0.0591 Species.at.Cathode
Ecell E 0 cell . log
n Species.at. Anode
3
Dr. Gangadhar B Bagihalli
Types of electrodes:
Depending on the electrochemical applications the electrodes have been classified into six types:
1. Metal-metal ion electrode: It consists of a metal (M) dipping into a solution containing
its own ions (Mn+). The electrode potential depends on the nature of the electrode and
activity of metal ions.
e.g., Zn / Zn2+
2. Metal-metal salt ion electrode: An electrode of this type consists of a metal in contact
with a sparingly soluble salt of the same metal, immersed in a solution containing anion
of the same metal salt. The electrode potential depends upon the activity of the reversible
ions.
e.g., Hg / Hg2Cl2/ Cl-, Ag /AgCl /Cl-
3. Gas electrode: It consists of a gas bubbling about inert metal electrode in a solution
containing ions to which the gas is reversible. The electrode potential depends upon the
pressure of the gas and the activity of the reversible ions.
e.g., Pt / H2 (g) / H+, Pt /Cl2 / Cl-
4. Amalgam electrode: This electrode is same as that of metal-metal ion electrode except
the fact that here metal-amalgam is in contact with its own metal ions. e.g., Pb-Hg/Pb2+
5. Redox selective electrode: An electrode of this type consists of an inert metal (e.g., Pt)
in contact with a solution containing ions of a substance in two different oxidation states.
The electrode potential depends upon the ratio of the activities of the two ions. The role
of inert metal electrode is to pick up electrons and to provide electrical contact.
e.g., Pt / Fe3+, Fe2+
6. Ion selective electrode: Ion-selective electrode (or membrane electrode) is used for
determining the activities of various ions in aqueous solutions. The potential developed
by this electrode depends only on the activity of ionic species of interest.
e.g., Glass electrode, which is specific to H+ ions.
Reference electrodes:
“Reference electrode are the electrode with reference to those, the electrode potential of
any electrode can be measured”
4
Dr. Gangadhar B Bagihalli
Calomel electrode:
Calomel electrode is extensively used as secondary reference electrode. This is a metal-
metal salt type electrode. Hg2Cl2 is called calomel. Calomel electrode consists of Hg, Hg2Cl2 and
solution of KCl. Calomel electrode is shown in Fig. Calomel electrode is made up of glass tube.
Hg is placed at the bottom of the glass tube which has side tubes at both sides. Platinum wire is
sealed through the glass tube and dipped into the Hg, which provides the external electrical
contact. Hg is covered with paste of Hg+ Hg2Cl2 and KCl solution. Depending on the
concentration of KCl the calomel electrode is called decinormal, normal and saturated.
Figure
Calomel electrode may act as anode or cathode with respect to other electrode.
5
Dr. Gangadhar B Bagihalli
It is simple to construct.
The electrode potential is reproducible and stable.
It is used as a reference electrode.
Limitations: The calomel electrode contains mercury, which poses health hazards. It potential
alters at higher temperatures (60C).
6
Dr. Gangadhar B Bagihalli
Therefore, the electrode potential of single electrode can be measured by using some
reference electrode whose valve is known.
In order to measure the electrode potential of given electrode, it is coupled with calomel
electrode through salt bridge. Then the valve of emf of cell, Ecell is measured using electronic
voltmeter.
For example, to measure the potential of the zinc electrode is coupled with calomel
electrode through a salt bridge and voltmeter, as in diagram.
7
Dr. Gangadhar B Bagihalli
8
Dr. Gangadhar B Bagihalli
Ion selective electrode consists of ion selective membrane in contact with an analyte
solution on one side and an internal reference solution on the other side. An internal reference
electrode is constituted with reference solution. The electrode can be represented as
The observed potential of an ion selective electrode is a kind of junction potential that develops
across a membrane that separates the analyte solution from internal std. solution.
Where, C1 & C2 are the concentration of the analyte and reference solution respectively. If the
concentration of the inner std. solution is kept constant, then
The overall potential of the electrode also includes that of the reference electrode and thus,
EM = Ej + Eref
The electrode is combined with an external reference electrode and EMF of the so formed cell is
determined by potentiometric method. Knowing the electrode potential of the reference
electrode, the electrode potential of the ion selective electrode is calculated which in turn gives
the ion concentrations.
9
Dr. Gangadhar B Bagihalli
Advantages:
ISEs are one of the few techniques which can measure both positive and negative ions
(H+, Na+, Ca2+, NH+, Cl-, F-, NO3-, S2-, etc).
Glass electrode can be used in aqueous solutions over a wide temperature range and is
unaffected by sample colour or turbidity.
It is not affected by oxidizing or reducing species present in the analyte.
It is particularly useful in biological/medical applications because it measures the activity
of the ion directly, rather than the concentration.
It is simple to operate, requires little sample and gives reproducible pH value.
Glass electrode:
Figure
10
Dr. Gangadhar B Bagihalli
H+ + MG1 M+ + HG1
The exchange of ions by the inner and outer membrane gives rise to a boundary potential.
This boundary potential consists of two potentials E1 & E2, which are associated with the outer
and inner membranes respectively.
Eb = E1- E2
From thermodynamic consideration it has been shown that E1 & E2 are related to the
hydrogen ion activities at each face. This is represented by Nernst equation type relation
Where j1 and j2 are constants, a1 and a2 are activities of H+ ions in the outer and inner
solution, respectively. a11 & a12 are the activities of H+ ions at the external and internal surface of
the glass membrane.
Since the glass at the inner & outer membranes are identical J1 = J2 and a11 = a12
Then,
Eb = E1- E2
Eb = L’ +0.0591 log C1
EG = Eb + Eref + Easy
11
Dr. Gangadhar B Bagihalli
Note: 1. In Glass electrode the glass used is made up of Na2O (22%), CaO (6%) and
SiO2 (72%). It can sense hydrogen ions up to a pH of about 9.
2. Asymmetric potential: In EG = Eb + Eref
Eb = 0, when C1 = C2. However, it has been observed that, when C1 = C2 a small
potential is developed, that potential is called as asymmetric potential (Easy).
Therefore, EG = Eb + Eref + Easy
To determine the pH of a given solution, the glass electrode dipped in a solution whose
pH needs to be determined. Figure shows the arrangement of electrodes in determining the pH.
It is combined with saturated calomel electrode.
pH Meter
Glass
Calomel
electrode
electrode
Solution of
unknown pH
Figure
Cell representation
Saturated calomel elctrode\\soln of unknown pH\glass\HCl (0.1M)\AgCl\Ag
Ecell = EG - ECal
Ecell = E0G - 0.0591 pH - Ecal
pH = E0G – Ecal-Ecell/0.0591
pH = K - Ecell/0.0591, where K- Glass electrode assembly constant.
E0G is calculated by dipping the glass electrode in solution of known pH.
12
Dr. Gangadhar B Bagihalli
pH Meter:
The advantage of pH meter is that there is no need to evaluate K which get cancelled by
external potential.
A reference electrode and a SCE are connected to pH meter. The electrode assembly is
dipped in a standard buffer solution and meter is calibrated to pH of the buffer.
The pH of buffer pHb is related to Ecell(b) by equation
pHb = K – Ecell(b)
__________ ---------------------- (1)
0.0591
Next electrode assembly is dipped in test solution and pH of test solution is given by
pH = K – Ecell
__________ ---------------------- (2)
0.0591
Substracting 1 from 2 we get
Concentration Cells:
A concentration cell is a Galvanic cell that has two identical half-cells of the same
material differing only in concentrations. Here, potential arises due to a decrease in free energy
due to difference of concentration (activity) between the two electrodes. Cell potential can be
calculated using the Nernst Equation. A concentration cell produces a voltage as it attempts to
reach equilibrium, which will occur when the concentration in both cells are equal.
13
Dr. Gangadhar B Bagihalli
Concentration cells consist of two electrodes of the same substance in contact with solutions
of different concentrations and potential difference arises as long as concentration difference
exists. These cells are with or without transference of material from one electrode to the other.
The above example is for electrolyte concentration cell without transference. Its
electrodes and cell reactions are:
For spontaneous cell reaction, c2 c1, the more concentrated solution should be in the
right half-cell and less concentrated solution should be in the left half-cell.
14
Dr. Gangadhar B Bagihalli
REVIEW QUESTIONS
Sensors:
Introduction
Development in Science and Technology during last century has resulted accelerated
growth in emission of anthropogenic pollution in to environment contributing to increasing
incidences of chronic diseases. Toxic gases, metals, pesticides and other dangerous chemicals
have spread throughout our environment. Therefore, monitoring of composition of air, water and
soil has become an area of intense interest. In health sector, for diagnosis and health monitoring,
it is required to determine clinically important chemicals present in body fluids. For various
industrial and agricultural applications qualitative detection and quantitative analysis of chemical
species is the major requirement. These demands have led to the development of many important
analytical approaches. These techniques are not suitable for routine and online analysis.
Therefore there is a growing demand to develop new instruments, techniques and procedure so
as to make analysis convenient, fast, reliable and selective. To meet these demands, Chemical
sensors of various types have been intensively developed. These sensors have number of
advantages over other classical methods of analysis like low cost, less time of analysis,
portability of the device, sample non-destruction and accuracy of result. Chemical sensors find
numerous applications like in home safety, food freshness, medical diagnosis and environmental
protection and are therefore likely to have a positive impact on our lives in the nea future.
15
Dr. Gangadhar B Bagihalli
A chemical sensor is a device that is capable of giving real time analytical information
about a test sample. It interacts with a specific chemical or biological analyte, detects it and
produces the signal proportional to its quantity. Analyte is the target species which is being
detected and measured using a sensor.
a) Receptor:
Receptor is a chemical or biological recognition element which is capable of interacting with
analyte specifically and selectively. It produces a signal corresponding to interaction in the form
of changes in potential, conductivity, current, mass, heat, pH, colour etc. If the receptor is of
biological origin (e. g., DNA, antibodies and enzymes), the device is referred to as a biosensor. If
the receptor is of chemical origin, the device is referred to as a chemical sensor.
b) Transducer:
During the interaction of sensing element with analyte, certain physical or chemical properties of
the sensing element changes proportionately to the analyte concentration. Transducer is used to
convert the signal created by the receptor-analyte interaction into a readable value or measurable
from physical quantity. In most chemical sensors, the sensing element and the transducer are
packaged together, in direct spatial contact, in the same unit.
16
Dr. Gangadhar B Bagihalli
Interaction between the receptor and anlyte can be monitored by several methods.Different
parameters are measured in each method like change in potential, conductivity, current, mass,
heat, pH, colour etc. For measurement of each parameter, different types of transducer are used.
There are several types of sensors available depending on type of transducer is used like
Electrochemical sensors, Conductometric sensors, Thermometric sensors and Optical sensors etc.
Electrochemical Sensors:
1. Diffusion of the analyte to the electrode/electrolyte interface I(in the liquid phase)
2. Adsorption onto the electrode surface
3. Electrochemical reaction with electron transfer
4. Desorption of the products
5. Diffusion of the products away from the reaction zone to the bulk of electrolyte or gas
phase
1. The oxygen sensor is used for detection of dissolved oxygen in water boiler and to
monitor dissolved oxygen concentration in metal, glasses and in hydrogen fuel.
2. They are used in security and defense applications like detection of toxic gass,
warfare agents etc.
17
Dr. Gangadhar B Bagihalli
3. They are used in water analysis and environmental monitoring, like measurement of
toxic metal concentration in water, detection of oxides of nitrogen and sulphur, CO,
pH of water etc.
4. They are used in diagnostic and health-care applications, like in situ monitoring of
glucose serum uric acid, blood Ca2+, Fe+2, Acetylcholine etc.
5. They are used in soil parameter analysis, evaluation and in agriculture applications.
1. Potentiometric Sensor:
In this sensor, change in potential during the chemical interaction between receptor and
analyte is measured using combination of an indicator electrode and reference electrode.
Commonly, Platinum is used an inert electrode whereas saturated calomel electrode is used as a
reference electrode. Indicator electrode is used to measure change in potential due to redox
reaction occurring on the surface of the electrode. Using an ion-selective electrode as an
indicator electrode, concentration of particular ion can be measured with good selectivity. pH
glass electrode is an example for this class of electrode. In potentiometric sensor, measurement is
taken at zero current.
Potentiometric Sensor
2. Amperometric Sensor
An amperometric sensor is similar to a potentiometric sensor consisting of an electrolyte solution
in which the electrode are immersed. But, it is operated under an externally applied fixed
voltage. This potential causes the analyte to react and a current to pass. At fixed applied
potential, the magnitude of the current generated is directly proportional to the concentration of
analyte. Here, current passing through the cell is measured.
18
Dr. Gangadhar B Bagihalli
In a sample amperometric sensor, two electrode namely counter electrode and working
electrode are used. Theoretically, in amperometric, potential between two electrodes must be
maintained constant. But, in reality, redox reactions at each electrode surface results in potential
change. Therefore to keep potential constant, a three-electrode system is commonly used. In the
three-electrode system, the current at the sensing electrode can be measured at constant potential.
This helps in maintaining a constant potential and measure the current at constant potential using
counter and working electrode.
19
Dr. Gangadhar B Bagihalli
3. Conductometric Sensor
20
Dr. Gangadhar B Bagihalli
Specific conductance (k) is given by, k = 1/R X 1/a where 1/a is known as cell constant and R is
the resistance of the solution.
The conductitivity cell is dipped in the electrolytic solution taken in a beaker and it is connected
to a conductance measuring device called as conductivity meter.
1. Conductometric sensor can be used to monitor any chemical which can change the
electrolytic conductionce of solution on chemical reaction.
2. It is used to estimate acids, bases and their mixtures in a sample.
3. It is used to check the amount of ionic impurities in water samples.
4. It is used in measuring acidity or alkalinity of sea water and fresh water.
5. Conductometric biosensors are used in biomedicine, environment monitoring,
biotechnology and agriculture related applications.
Conductmetric Sensors
Optical Sensors:
Interaction of electromagnetic radiation with matter forms the basis of broad range
analytical methods commonly known as spectrochemical methods of analysis. Commonly,
electromagnetic radiation in the ultraviolet-visible-infrared domains is used for analytical
purposes. A broad range of chemical sensors have been developed on the ground of interaction
of the sensing element with electromagnetic radiation. Sensors based on the transduction of
interaction of electromagnetic radiation with the chemical species are called as optical sensors.
21
Dr. Gangadhar B Bagihalli
In a simple optical sensor used to measure absorption of light, main components used are
a light source, a wavelength selector, a photodetector and a display of the output.
In modern optical sensors, optical components such as a lens, optical couplers and
connectors are used for coupling light into optical fibers and solid-state optoelectronic
components. This has enabled the development of commercial portable optical sensor system.
Simple optical sensors are used to determine the concentration of colored chemical
species in solution. They are based on measurement of absorbance of transmittance of light of
particular wavelength by colored chemical species in the solution. They are governed by Beer-
Lamberts law.
1. Optical Sensors can be used in the determination of any chemical species which can
interact with electromagnetic radiation.
2. Optical Sensors have been developed for number of different types of chemical and
biochemical molecules and ions. For example, ion in solution (e.g., pH, metal ions
and anions), gases (e.g., CO2, O2, NH3, SO2, NO2, NO, etc.,), Vapors (e.g.,
moisture, volatile organic compounds, etc.,) and molecules (e.g., glucose, pesticides,
DNA, bacteria, etc.,) can be determined using optical sensors.
3. Optical sensors find important and varied uses in environmental, biotechnological,
food, pharmaceutical, medical and related applications.
4. Optical fiber based (bio)sensors are used in screening of drugs, detection of food-
borne pathogens, detection of explosives and environmental monitoring.
22
Dr. Gangadhar B Bagihalli
Thermometric Sensors:
Main component of a thermometric sensor is a small tubular catalytic reactor fitted with a
tempetature transducer. Analyte (reactant) is fed in to the reactor. The wall of the reactor is
coated with catalyst or enzyme capable of catalyzing the reaction, liberating heat energy. Heat
liberated is quantified by means of a temperature transducer. The change in temperature is
converted to the output voltage by transducers which are amplified and fed to the data storage
and processing unit.
In order to convert chage in temperature into an electric signal, two main kinds of
transducers which exhibit thermoelectric effect are used.
1. Thermal bio sensors are based on the temperature change induced by a simple
enzymatic reaction. They are used in the determination of metaboloties, bioprocess
monitoring, and environmental control. An example is the determination of glucose
using the glucose oxide-catalyzed reaction.
2. Thermometric chemical sensors are used for determination of combustible gases that
react with oxygen at the surface of a suitable catalyst.
23
Dr. Gangadhar B Bagihalli
Review Questions:
Analytical Techniques:
INTRODUCTION:
Analytical Chemistry is branch of chemistry that deals with chemical analysis. Chemical
analysis involves identifying and determining the percentage composition in a sample of a
substance. The identification is carried out by qualitative analysis. Quantitative analysis gives the
percentage composition of the constituents. The sample under analysis is an analyte. Some
important techniques of quantitative analysis are titrimetry or volumetry, gravimetry and
instrumental methods.
In titrimetry, a known amount of the analyte present in the solution is allowed to react
with a standard solution of an appropriate reagent. The volume of the standard reagent required
to react with known amount of the analyte is determined by titration.
24
Dr. Gangadhar B Bagihalli
The detector identifies the analytical signal from the analyte and converts it into an a
transduced signal (electrical signal). The latter is then processed by the signal processor which
transforms the signal in such a form that it can be recorded on a meter, x-y recorder or
digitalizer. The signal processor can amplify, attenuate, differentiate, integrate or compare the
analytical signal. Modern analytical instruments make use of microprocessor based signal
processors.
(i) They require small quantities (less than a milligram) of the analyte.
(ii) Methods adopted are easy when a large number of analytes have to be analyzed
CONDUCTOMETRY:
25
Dr. Gangadhar B Bagihalli
by other ions of different mobility. Therefore, the conductance of a solution depends on the
number and mobility of ions. The equivalence point is determined graphically by plotting
conductance (ordinate) against titer values (abscissa). From the equivalence point the
concentration of analyte can be determined. Conductometer consists of a conductivity cell& a
conductance measuring device.
K = 1/ ρ
L/A = x is constant and called as cell constant. It is measured by the with the solution of
known specific conductance, Ex. KCl.
26
Dr. Gangadhar B Bagihalli
Instrument:
The cell has 2 platinum electrodes placed at unit distance apart. The assembly responds
rapidly to the changes in the concentration of the analyte. The simple arrangement of
conductometric titration is represented as follows. Advantages of conductometric titrations
Applications:
27
Dr. Gangadhar B Bagihalli
Consider the titration of HCI against NH4 OH. The conductance first falls, due to the
replacement of highly mobile H+ ion by the ions during neutralization. After the equivalence
point, the conductance does not change appreciably since NH4OH is a weak base. A plot of
conduct against the volume of NH4OH added is shown in Fig. 1b The point of intersection of the
two curves gives the neutralization point.
28
Dr. Gangadhar B Bagihalli
Consider the titration of acetic acid against NaOH. The conductance of the acid will be
initially low since acetic acid is a weak electrolyte. When NaOH is added to the acid, the salt
formed is highly ionized and the conductance increases. On complete neutralization of the acid,
further addition of base leads to an increase in the number of mobile OH ions. Hence the
conductance increases sharply. Thus a plot of conductance against volume of base added gives
two straight lines as shown in the Fig. 1c. The point of intersection of the two lines gives the
neutralization point.
d) Acid mixture with strong bse (HCl & CH3COOH v/s NaOH)
Consider the titration of acid mixture (HCl & CH3COOH) with strong base NaOH On the
addition of NaOH, the conductance of the acid mixture solution decreases steeply, because the
highly mobile H+ ions in the strong acid is gradually replaced by less mobile Na+ ions.
Addition of base beyond the first neutralization point leads to moderate increase in conductance.
The conductance increases moderately, because the poorly conducting weak acid (acetic acid) is
gradually converted to highly conducting salt (sodium acetate) until the second neutralization.
29
Dr. Gangadhar B Bagihalli
Further addition of NaOH, increases number of highly mobile OH ions. Hence conductance
increases steeply.
Where, € is the molar extinction coefficient, t is the path length and is a constant for a given
substance at a given wavelength. If t, the path length is kept constant, then, A Vs C. Hence, a plot of
absorbance against concentration gives a straight line.
30
Dr. Gangadhar B Bagihalli
Instrumentation:
(ii) A filter which provides the desired wavelength range wherein the solution gives the
maximum absorbance.
Light from a tungsten lamp is allowed to fall on the solution taken in the sample cell after
passing through the filter. First, a blank solution is taken in the sample cell and placed in the path
of the light beam. Its absorbance is adjusted to zero (or transmittance to 100) on the mater. Next
the analyte solution is placed in the path of the light beam and the quantity of light absorbed is
measured as its absorbance (or transmittance)
31
Dr. Gangadhar B Bagihalli
APPICATIONS OF COLORIMETRY:
Principle:
When a solution containing cupric ions is treated with ammonia solution, a soluble
complex with a characteristic deep blue colour is formed.
The absorbance of this solution is measured at 620 nm (λ max) since the complex shows
maximum absorbance at this wavelength.
A series of standard solutions of Cu2+ are prepared and their absorbances are measured at
620 nm after treating with ammonia solution. A plot of absorbance against concentration gives a
calibration curve. The absorbance of the test solution is measured in a similar way. Its
concentration is determined for the graph.
Procedure:
Transfer a known volume of steel solution into a 100 ml flask so as to give an absorbance
between 0.1 and 0.5. Add 5 ml of ammonia and make it up to the mark with distilled water.
Stopper the flask and mix the solution well.
32
Dr. Gangadhar B Bagihalli
Measure the absorbance of the solutions against blank ( 5 ml ammonia and water in a
standard flask) at 620 nm using a photoelectric colorimeter. By plotting calibration curve
concentration of copper (in mg/ml) against absorbance as shown in the fig. using this curve we
can determine the concentration of copper in the given sample solution.
Potentiometry:
Theory
A Potentiometric titration may be defined as a titration in which the end point is detected by
measuring the change in potential of a suitable electrode during the titration. The electrode which
responds to the change in concentration of the ion in the solution is called the indicator electrode.
The indicator electrode is combined with the reference electrode to form the cell and the e.m.f
of the so formed cell is measured during the titration The e.m.f of the cell changes gradually till
the end point and changes rapidly at very close to the end point and again the change is gradual
after the end point When e.m.f., is plotted as ordinate and the volume of titrant added as abscissa,
the point of inflection of the curve corresponds to the equivalence point or end point of the
titration
33
Dr. Gangadhar B Bagihalli
Reference electrode eg- calomel for potentiometer, glass electrode for pH meter
1. The solution to be titrated is taken in the beaker and titrant is taken in the micro burette.
2. Immerse the electrodes in the beaker along with stirrer
3. The electrodes are connected to the potentiometer which gives the emf values
1. The FAS solution is taken in the beaker and K2Cr2O7 is taken in the micro burette.
1. The electrodes are connected to the potentiometer which gives the emf values
2. Go on adding the titrating reagent with an in travel of 0.5 ml
3. There will be sudden jump in the emf value after adding some amount of reagent
4. carry out the experiment for five more readings
5. plot first derivative curve ΔE/ ΔV Vs volume of reagent added
6. The volume correspond to the maximum peak in the graph is called equivalence point or
end point of the reaction
9. Using formula normality and amount can be calculated.
34
Dr. Gangadhar B Bagihalli
∆E/∆v
Volume of K2Cr2O7
Applications:
1. Colored solution can also be titrated.
2. Acid-base titration can also be done in this method.
3. In this method Oxidation-reduction titrations can also be carried out.
4. Precipitation reactions can also be carried out potentiometrically.
REVIEW QUESTIONS:
35