Document Information for: 5409509SCOMS

Type Drawing Print


Name 5409509SCOMS
Revision 2
State Release
ECO 2135369
Description Pharos Bearing Washer Cleaning OMS
Originator 212006294_james_m_rossmiller

File List
1. 5409509SCOMS_s1_r2.pdf
2. 5409509SCOMS_s1_r2.pptx
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.

Approval Information
Person Action Date and Time
212008849_john_a_holma Approved 11/06/2012 02:23:00 pm GMT
214006619_michael_j_utschig Approved 10/29/2012 07:22:43 pm GMT

This page is generated automatically by the GEHC MyWorkshop System.


Printed documents are for Reference Only and may be out-of-date.
Check the database to ensure you have the correct revision.

Approved Document - 5409509SCOMS_r2.pdf Page 1 of 7


DOCUMENT SECURITY
This document is considered GE Healthcare proprietary.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.

Note: Revision of this document is controlled by a PDM system (e.g., eDocs, GPCP, ePDM, myWorkshop). Before using this
document, make sure it is the latest revision. Access the PDM system to verify the current revision. If you do not have access
to, or are unfamiliar with, the PDM system, consult your quality representative.
See the GEHC Myworkshop System to determine the status of this document.

Revision History

Pages Description of Change Change Control Num ber Author Release Date
Rev.
Changed

1 All Initial Release of OMS for new part ECO 2123893 John Holma See ECO

Added use of "minnow basket" to


2 All ECO 2135369 G. Dzioba See ECO
process.

CRITICAL TO QUALITY
INSTRUCTION VERIFY REFERENCE
(CTQ) OR SAFETY
Title: Pharos Washer Plate Solvent Cleaning OMS AREA: SOLVENT CLEANER Rev
First Made For: Global Tube Engineering RDO = “TUB” PRODUCT: Pharos 2

GEHC Proprietary – Do Not Distribute


Copyright © 2004, GE
Approved Healthcare
Document - 5409509SCOMS_r2.pdf
GE Healthcare
Milwaukee,
Page 2 of Wisconsin,
7 USA
5409509SCOMS
Sheet 1 OF 6
Pharos Washer Plate
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.

Solvent Cleaning
5409509
See the GEHC Myworkshop System to determine the status of this document.

Always use nitrile gloves when


handling parts.

CRITICAL TO QUALITY
INSTRUCTION VERIFY REFERENCE
(CTQ) OR SAFETY
Title: Pharos Washer Plate Solvent Cleaning OMS AREA: SOLVENT CLEANER Rev
First Made For: Global Tube Engineering RDO = “TUB” PRODUCT: Pharos 2

GEHC Proprietary – Do Not Distribute


Copyright © 2004, GE
Approved Healthcare
Document - 5409509SCOMS_r2.pdf
GE Healthcare
Milwaukee,
Page 3 of Wisconsin,
7 USA
5409509SCOMS
Sheet 2 OF 6
Pharos Washer Plate Solvent Cleaning
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.

Before using the cleaner, confirm that the Green Tag has not expired. Refer to
2
2345469SCMDO for additional equipment operation details.
See the GEHC Myworkshop System to determine the status of this document.

Verify basket fixture is resting


3
properly on robot arm.
TRANSPORT BASKET

MINNOW
BASKET

4 With nitrile gloves on, place up to


WASHER
24 washer plates into a minnow
basket and place minnow basket
into Transport Arm Basket.

CRITICAL TO QUALITY
INSTRUCTION VERIFY REFERENCE
(CTQ) OR SAFETY
Title: Pharos Washer Plate Solvent Cleaning OMS AREA: SOLVENT CLEANER Rev
First Made For: Global Tube Engineering RDO = “TUB” PRODUCT: Pharos 2

GEHC Proprietary – Do Not Distribute


Copyright © 2004, GE
Approved Healthcare
Document - 5409509SCOMS_r2.pdf
GE Healthcare
Milwaukee,
Page 4 of Wisconsin,
7 USA
5409509SCOMS
Sheet 3 OF 6
5 Close front door (cycle will not run unless door is
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.

fully closed).

Verify recipe #2 is selected as the


See the GEHC Myworkshop System to determine the status of this document.

6
“ACTIVE RECIPE”.

Start wash cycle by pressing and holding


7 “PRESS TO BEGIN BASKET CYCLE”
touch button for 3 seconds.

If alarm is triggered during operation, contact PTL and/or MPE.


8
CRITICAL TO QUALITY
INSTRUCTION VERIFY REFERENCE
(CTQ) OR SAFETY
Title: Pharos Washer Plate Solvent Cleaning OMS AREA: SOLVENT CLEANER Rev
First Made For: Global Tube Engineering RDO = “TUB” PRODUCT: Pharos 2

GEHC Proprietary – Do Not Distribute


Copyright © 2004, GE
Approved Healthcare
Document - 5409509SCOMS_r2.pdf
GE Healthcare
Milwaukee,
Page 5 of Wisconsin,
7 USA
5409509SCOMS
Sheet 4 OF 6
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.

During cycle, upper left corner of main screen will flash between “BASKET IS
PROCESSING” and “HEATING MODE IS ENABLED”
9
Also, robot arm will be in the lowered position
See the GEHC Myworkshop System to determine the status of this document.

Once cycle is done, upper right section of main screen


10 will read “READY TO BEGIN BASKET PROCESSING”

CRITICAL TO QUALITY
INSTRUCTION VERIFY REFERENCE
(CTQ) OR SAFETY
Title: Pharos Washer Plate Solvent Cleaning OMS AREA: SOLVENT CLEANER Rev
First Made For: Global Tube Engineering RDO = “TUB” PRODUCT: Pharos 2

GEHC Proprietary – Do Not Distribute


Copyright © 2004, GE
Approved Healthcare
Document - 5409509SCOMS_r2.pdf
GE Healthcare
Milwaukee,
Page 6 of Wisconsin,
7 USA
5409509SCOMS
Sheet 5 OF 6
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.

AFTER the cleaning cycle is complete.

IMPORTANT: CLEAN NITRILE GLOVES MUST BE USED WHEN


See the GEHC Myworkshop System to determine the status of this document.

11 REMOVING PARTS FROM MACHINE AFTER CYCLE COMPLETES.

12 With nitrile gloves on, remove washer plates from fixture once cycle is complete

13 Verify there are no marks on the parts introduced by the fixture.

14 Verify the parts are free of stains or other contamination


(machining chips, particles, etc)

15 Place completed washer plates into container designated by the traveler.

CRITICAL TO QUALITY
INSTRUCTION VERIFY REFERENCE
(CTQ) OR SAFETY
Title: Pharos Washer Plate Solvent Cleaning OMS AREA: SOLVENT CLEANER Rev
First Made For: Global Tube Engineering RDO = “TUB” PRODUCT: Pharos 2

GEHC Proprietary – Do Not Distribute


Copyright © 2004, GE
Approved Healthcare
Document - 5409509SCOMS_r2.pdf
GE Healthcare
Milwaukee,
Page 7 of Wisconsin,
7 USA
5409509SCOMS
Sheet 6 OF 6