Module 2: Electrode System: Introduction, types of electrodes. Ion selective electrode – definition, construction, working and
applications of glass electrode. Determination of pH using glass electrode. Reference electrode - Introduction, calomel
electrode – construction, working and applications of calomel electrode. Concentration cell – Definition, construction and
Numerical problems.
Sensors: Introduction, working principle and applications of Conductometric sensors, Electrochemical sensors, Thermometric
sensors, and Optical sensors.
Analytical Techniques: Introduction, principle and instrumentation of Colorimetric sensors; its application in the estimation
of copper, Potentiometric sensors; its application in the estimation of iron, Conductometric sensors; its application in the
estimation of weak acid.
Electrochemical Systems
Working of glass electrode: When the glass electrode is dipped into any solution containing H+
ions, the Na+ ions of the glass membrane are exchanged for H+ ions of the test solution.
1
H+ + Na+ GI- Na+ + H+GI-
If a thin walled bulb containing an acid is immersed in another solution containing H+ ions (fig),
a potential is developed across the glass membrane. This is called the boundary potential E b. It is
a potential developed across the glass membrane when concentration of the solution inside and
outside the glass membrane are different. The Eb is due to the difference in potential (E1-E2)
developed across the gel layer of the glass membrane between the two liquid.
Mathematically it is represented as,
Eb = E1-E2
Where, E1= Potential due to H+ present in outside solution (Unknown solution)
E2 = Potential due to H+ present in inside solution (known solution)
According to Nernst equation
Eb = 2.303RT log C2
nF
C1
Eb = 0.0591 log C1 –0.0591 log C2...................(1)
n n
Where, C2 is the concentration of H+ ions of the solution into which glass membrane is dipped.
The concentration of H+ ion inside the bulb (C1) is constant i.e. C1 = 0.1 M.
Thus, Eb = 0.0591 log C2 + K or
n
= K + 0.0591 log C2
n
2
Glass electrode selects only H+ ions ignoring other ions.
Hence C2 = H+
Eb= K + 0.0591 log [H+]
Where, log [H+] = -pH
Thus, Eb= K - 0.0591pH ---------------------(2)
The combined glass electrode is dipped into acidic solution, then the potential of the glass electrode
is given by….
EG = Eb +EAg-AgCl………………………(3)
From equation 1, theoretically if C1= C2, Eb Should be 0, however it has been observed practically
that even when C1 = C2, a small potential is developed which is called as asymmetric potential
(Easym). Hence equation 3 can be rewritten as
EG = Eb +EAg-AgCl + Easym ………………………(4)
Substituting the value of Eb from equation (2) in equation (4)
EG = K - 0.0591pH + E Ag/AgCl + E assy
3
Determination of pH using glass electrode
E cell
pH E E G cal
0.0591
pH = K - Ecell where, K is known as glass electrode assembly constant.
0.0591
E G value or K is evaluated by dipping the glass electrode in a solution of known pH and
measuring emf of the cell formed when coupled with a calomel electrode. Next the same assembly
is dipped into test solution and the pH of test solution can be determined.
4
They are of two types;
I. Primary Reference Electrodes: Eg. Standard hydrogen electrode (SHE). Its potential is arbitrarily
considered as zero, and it helps to determine the potential of other electrodes.
Limitations of primary reference electrode (SHE)
(a) Construction and working is difficult on account of difficulties involved in maintaining
concentration at unity and pressure in the gas (1 atm).
(b) Platinum is highly susceptible for poisoning by impurities in the gas.
(c) It cannot be used in the presence of oxidizing agents.
To over come all these difficulties, secondary reference electrodes are used.
Calomel electrode
It is a metal – metal salt ion electrodes and also a secondary reference electrode.
Construction: It consists of mercury, mercurous chloride and a solution of KCl. It is made up of
glass tube, Hg is placed at the bottom of the glass tube. It is covered with a paste of calomel (HgCl2)
and KCl solution. A solution of KCl is introduced above the paste through the side tube. Electrical
contact is made through Pt wire (dipped in Hg). This glass tube is taken into another glass tube,
which has a side tube and a porous plug at the bottom. This porous plug act as salt bridge. The
conc of KCl solution used is either decinormal, normal or saturated. Correspondingly, the electrode
is known as decinormal, normal or saturated calomel electrode. This constitutes one half cell.
Calomel electrode can act as either anode or cathode. It is connected to the other half cell through
salt bridge.
5
Half cell representation
The calomel electrode is represented as, Hg /Hg2Cl2 (s)/ Cl- (anode)
Or Cl-/Hg2Cl2 (s)/ Hg (Cathode)
Working
Depending on the potential of electrode with which the calomel electrode is connected, calomel
electrode acts as anode or cathode.
Applications:
1. It is used as secondary reference electrode in all potentiometric determinations.
2. Used in glass or combined electrode to determine the pH of the unknown solution.
Advantages:
1. Construction is simple and electrode is portable.
2. The cell potential is stable and reproducible.
3. Cell potential does not vary with temperature.
Limitations:
1. Calomel electrode cannot be used beyond a temperature of 60 °C.
2. Hg employed in construction poses pollution problem.
7
To determine the potential of unknown electrode, test electrode eg. Zn/Zn2+ is coupled with a
saturated calomel electrode to form a cell.
If test electrode is anode:
The cell representation is as follows:
Zn (s)/Zn2+// Cl-/ HgCl2(s) / Hg (l)
Zn Zn2+ + 2e-
At cathode:
Hg2Cl2 + 2e- 2Hg + 2Cl-
Over all reaction:
8
= Ecal - EZn
= 0.242 - EZn
= EMn+/M - 0.242
9
Concentration Cells
In the case of galvanic cells discussed earlier, the electrical energy arises from the chemical
reactions which take place in the cells. There is however another category of cells in which the
EMF arises not due to any chemical reaction but due to transfer of matter from one half to the other
because of a difference in concentration of the species involved.
Thus the concentration cell is an electrochemical cell in which both anode and cathode are made
up of same elements (identical electrodes) and immersed in same ionic solution (electrolyte) but
of different concentration.
Concentration cells are of two types:
1) Electrolyte concentration cell
2) Electrode concentration cell
1. Electrolyte concentration cell: In these cells, the two electrodes of same element is in contact
with solution of same metal ion (electrolyte) but of different concentration.
Eg. Two copper electrodes immersed in copper sulphate solutions of different concentrations C 1
and C2. Cell is represented as
Cu / Cu2+( C1) // Cu2+(C2) / Cu
Construction:
10
At anode Cu → Cu 2+ (C1) + 2e-
At cathode Cu2+ (C2) + 2e- → Cu
EMF of cell is given by
E E E anode
cell = cathode
2.303RT 2.303RT
E nF logC 2 E nF logC1
=
log C 2
2.303RT
E cell
nF C1
2. For above cell to be operative, E cell should be positive which is possible only if C2 > C1.
Thus the direction of spontaneous reaction is from the more concentrated one to the less
concentrated one.
3. In a concentration cell, electrode with lower electrolyte concentration acts as anode and the
C 2
one with higher electrolyte concentration acts as cathode. Higher the ratio
C 1 , higher
is the emf.
11
Sensors
Introduction
A sensor is a detection device that can sense the measured data and convert it into electrical
signals or other required forms of information output according to particular rules, in order to meet
the requirements of data transmission, processing, storage, and display, as well as recording and
control.
Sensors enable objects to have senses such as touch, taste, and smell, and they gradually
become alive as a result of their evolution. Thermal elements, photosensitive elements, gas-
sensitive elements, force-sensitive elements, magnetic-sensitive elements, humidity-sensitive
elements, sound-sensitive elements, radiation-sensitive elements, color-sensitive elements, and
taste-sensitive elements are the ten categories according to their basic sensing functions.
12
Sensors are frequently categorized based on their functioning principle, input data, and
application scope. They may be loosely split into three types based on their diverse functioning
principles:
1) Physical sensor – Sensors built with the physical qualities of specific transforming
components and the special physical properties of certain functional materials are known as
physical sensors.
2) Chemical sensor – Chemical sensors convert the composition and concentration of
inorganic and organic chemical compounds into electrical signals using the electrochemical
reaction concept.
3) Biological sensor - It's a sensor that detects and measures biochemical compounds using
biologically active chemicals.
Conductometric sensor:
Working Principle: The relationship between conductance and receptor techniques is the basis
for developing conductometric detection. The conducting material acts as an
electrochemical transducer to convert the sensing signals into electrical signals.
The basic principle of conductometric detection involves a reaction that can change the ionic
species concentration. This reaction leads to changes in electrical conductivity or current flow. In
this method, two inert metal electrodes are used. The two electrodes are separated at a certain fixed
distance before applying AC voltage, which later causes current flow. The change in conductance
between the electrodes due to ionic composition changes during recognition is measured. The
speed, practicality, specificity, sensitivity, near real-time detection, and need for low sample
volume are some advantages of conductometric sensors.
Applications of conductometric sensors:
(i) Detection of foodborne pathogens.
(ii) Detection of alkali metal ions, organic acids,
(iii) Detection of circulating tumor cells.
(iv) Toxic heavy metal ion detection
13
(v) Conductometric sensors for gas sensing.
Additional Material: Conductometric detection is based on measurement of specific conductance
of an analyte and is preferable because it can be applied for detection of both electroactive and
electroinactive species. The conductivity electrodes can either be in contact with the solution or
insulated using a thin layer. Conductivity detection is mostly associated with capillary
electrophoresis (CE).
Contact-mode conductivity measurement, the electrodes are in direct contact with the sample.
Although better contact with sample results in good sensitivity and lesser response time, it is
accompanied with the increased risk of degradation of electrodes and contamination of sample
Contactless conductivity detection, on the other hand, does not suffer from electrode instability.
In this case, the electrodes are separated from the buffer or sample by a thin layer of insulation.
For achieving good capacitive coupling, the insulation layer needs to be as thin as possible.
Conducting polymers are also used for conductance-based detection. They are organic
conjugated compounds with an extended π-orbital system allowing movement of electrons from
one end to other. Conducting polymers have unique electrical and optical properties that can be
tuned by reaction with redox active agents. Electrical conductivity of these types of polymers
depends on their ability to transport charge carriers along the polymer backbone. Interaction with
redox active species results in the change in conductivity of the polymers which can be measured
either in DC or AC conditions. Commonly used conducting polymers are polyaniline,
polyacetylene, polypyrrole, polyfluorine, etc.. The advantages of conducting polymers are their
environment stability, easy processibility, compatibility with biomolecules, ability to be coated on
the surface of desired dimensions, and modification to bind to biomolecules. These properties
taken together have paved the way for developing conducting polymer-based sensors for detection
of bacteria, proteins, trace metal ions, and many more. Such a conductometric biosensor was
developed by Dixit et al. [80] in which polyaniline thin film doped with Fe-Al was deposited on
glass substrate. On top of this film, an interdigitated Au electrode array was designed for detection
of E. coli in liquid samples. The current–voltage characteristics of the film upon exposure to
microorganism were measured using source measuring unit. Change in conductivity of the film is
due to the interaction of charged microorganism with the polymer film resulting in redox reactions
at the polymer film interface.
14
Electrochemical sensors
Electrochemical sensors are devices that give information about the composition of a
system in real time by coupling a chemically selective layer (the recognition element) to an
electrochemical transducer. Electrochemical sensors are made on the basis of ion conduction.
According to the formation of their electrical characteristics, electrochemical sensors can be
divided into potential sensors, conductivity sensors, electricity sensors, polarographic sensors, and
electrolytic sensors.
Electrochemical sensors are mainly used to analyze gas, liquid, or solid components
dissolved in liquids, the measurement of liquid pH, conductivity, and oxidation-reduction
potential.
Working Principle
The electrochemical sensor works by reacting with the measured gas and generating an
electric signal proportional to the gas concentration. A typical electrochemical sensor consists of
a sensing electrode and a counter electrode and is separated by a thin electrolytic layer.
The gas first reacts with the sensor through the tiny capillary-shaped opening, then the
hydrophobic barrier layer, and finally reaches the electrode surface. Using this method can allow
an appropriate amount of gas to react with the sensing electrode to form a sufficient electrical
signal while preventing electrolytes from leaking out of the sensor. The electrochemical sensor
contains the following main components:
a) Breathable membrane (also called hydrophobic membrane): The breathable membrane
is used to cover the sensing (catalytic) electrode, and in some cases, it is used to control
the molecular weight of the gas reaching the electrode surface.
b) Electrode: The electrode material should be a catalytic material that can perform semi-
electrolytic reactions over a long period of time. Generally, electrodes are made of precious
metals, such as platinum or gold, which react effectively with gas molecules after catalysis.
c) Electrolyte: The electrolyte must be able to carry out the electrolysis reaction and
effectively transfer the ionic charge to the electrode.
d) Filter: Sometimes a scrubber filter is installed in front of the sensor to filter out unwanted
gas. The most commonly used filter material is activated carbon.
15
Advantages and disadvantages of electrochemical sensors
Advantages of electrochemical sensors
The use of electrochemical sensors can accurately measure some odors, analyze proteins,
and can also be used for new drug development and disease treatment.
Detection speed is relatively fast, very simple, convenient.
The cost is relatively low.
Electrochemical sensors are useful in many aspects, such as medicine and industrial
analysis.
Disadvantages of electrochemical sensors
Short lifespan. They usually need to be replaced every one to three years.
Electrolytic solution, need to replenish it regularly, which is a bit troublesome.
Application of electrochemical sensors
i. Humidity sensor - The sensor used to measure relative humidity is coated with
piezoelectric quartz crystal, which is made of small quartz crystal by
photolithography and chemical etching technology.
ii. Detection of toxic gases like nitrogen oxide, hydrogen sulfide and sulfur dioxide with high
selectivity and sensitivity.
Optical Sensors
The method of sensing light rays is known as optical sensing. The sensor type used for
optical sensing is known as optical sensor. Optical Sensor converts light rays into electrical signal.
The purpose of an optical sensor is to measure a physical quantity of light and, depending on the
type of sensor, then translates it into a form that is readable by an integrated measuring device.
Working principle:
The operating principle is the transmitting and receiving of light in an optical sensor, the
object to be detected reflects or interrupts a light beam sent out by an emitting diode. Depending
on the type of device, the interruption or reflection of the light beam is evaluated. This makes it
possible to detect objects independently of the material they are constructed from (wood, metal,
plastic or other). Special devices even allow for a detection of transparent objects or those with
different colors or variations in contrast.
Optical sensors use visible or ultraviolet light to interrogate sensors for analysis. Optical
sensors can be represented in general terms as a wavelength-selectable light source, the sensor
material itself interacting with analytes, and a light detector (Fig.). What the detector monitors
varies by technique (e.g., refractive index, scattering, diffraction, absorbance, reflectance,
photoluminescence, chemiluminescence, etc.), can cover different regions of the electromagnetic
spectrum, and can allow measurement of multiple properties (e.g., intensity of light, lifetime,
polarization, etc.).
16
Figure: Working operation of optical sensing. General arrangement of spectroscopic
measurements: (A) light reflection; (B) light refraction; (C) light absorption; (D) fluorescent
emission.
Applications:
17
of resistance thermometers are commonly made of metal wires as platinum and copper, and at
low temperatures, carbon, germanium, and
rhodium iron are often used for the element.
Because they are almost made of platinum,
we often call them platinum resistance
thermometers.
As the temperature changes, the
resistance value of the metal also changes.
And for different metals, the resistance value changes differently with the temperature, which
can be directly used as the output signal.
Resistance changes in 2 different ways:
Positive temperature coefficient
temperature increases & resistance
increases
Temperature decreases & resistance
decreases
Negative temperature coefficient
temperature increases &resistance
decreases
Temperature decreases &resistance
increases
Application:
1. Sensing Application - The thermal conversion method of temperature sensors is often
used to measure physical quantities, such as flow rate, radiation, gas pressure and type,
humidity, thermochemical reaction, etc.
2. Biomedical Domain - Special temperature sensors are often applied for biomedical
applications. These temperature sensors have low power consumption, long-term
stability, and high reliability, with an accuracy of less than 0.1℃ between 32℃ C and
44℃
3. Industrial Application - Integrated temperature sensors can be applied in automation
and microbe thermal detection.
4. Consumer Products - Many low-cost integrated temperature sensors and transmitters
have been used in consumer products such as washing machines, refrigerators, and air
conditioners.
18
Analytical Techniques:
Application: (i) Colorimetry is versatile method of determining the concentration of metals and
nonmetals present in small quantities in ores, soil, samples and alloys.
(ii) For the determination of the course of the reaction by measuring the rate of formation and
disappearance of the light-absorbing compound
19
(iii) A compound can be identified by determining the absorption spectrum in the visible region of
the light spectrum
(iv) For determination of the concentration of a colored compound
(v) Colorimetric estimation of Cu in CuSO4. Draw out 2, 4, 6, 8, and 10 ml cm3 of the Copper
sulphate solution into 50cm3 volumetric flask. Add 5cm3 of ammonia solution to each of them and
dilute upto the mark with distilled water and mix well. Prepare a blank solution by diluting 5cm 3
of ammonia solution in 50cm3 volumetric flasks. For test solution add 5ml of NH3 and make up to
the mark. Measure the absorbance of each of these against blank solution at 620 nm. Plot a graph
of absorbance (OD) against volume of copper sulphate solution and determine the volume of
copper sulphate solution in the test sample as shown in the figure and find the amount of copper
present in it.
Thus, the concentration can be calculated, provided Eo of the electrode is known. The principle
involved in potentiometric titration is the measurement of emf between two electrodes, an indicator
electrode, (the potential of which is function of the concentration of the ion to be determined) and
a reference electrode of constant potential. In this method, the measurement of emf is made while
the titration is in progress. The equivalence point of the reaction is revealed by a sudden change in
potential in the plot of emf readings against the volume of titrant.
Instrumentation: A potentiometer consists of: (i) Calomel electrode as a reference electrode, (ii)
Platinum electrode as an indicator electrode, (iii) a device for measuring the potential and (iv)
Magnetic stirrer.
20
Application: Potentiometric estimation of FAS using standard K2Cr2O7 solution: Pipette out 25ml
of FAS into a beaker. Add 1 t.t dil H2SO4, immerse calomel electrode + platinum electrode into it.
Connect the assembly to a potentiometer and measure the potential by adding K2Cr2O7 in the
increments of 0.5ml.
Plot graph E/ V against volume of K2Cr2O7, and determine the equivalence point. From the
normality and volume K2Cr2O7, solutions calculate the normality and the weight of FAS in the
given solution.
21
The conductance of solution depends on the number and mobility of ions. The substitution of ions
with different ionic mobility affects the electrolytic conductivity. Therefore, the equivalence point
can be determined by means of conductivity measurement for a neutralization reaction between an
acid and a base. Equivalence point is determined graphically by plotting conductance against titer
values.
Instrumentation: Conductometer consists of: (1) conductivity cell having two platinum
electrodes; and a (ii) conductometer. A simple arrangement of conductometric titration is depicted
in figure. The solution to be titrated is taken in the beaker.
Application in the estimation of weak: Pipette out 50ml of sample into a beaker. Immerse the
conductivity cell into it. Connect the conductivity cell to a conductivity meter and measure the
conductance by adding NaOH from the burette by increment of 1 ml. Plot a graph of conductance
against volume of NaOH. Determine the neutralization point from the graph as shown below.
Weak acid with a strong base: (CH3COOH Vs NaOH): In the conductometric titration of a
weak acid with a strong base, the conductance of the acid will be initially low due to poor
dissociation of acetic acid. On complete neutralization of the acid, further addition of base leads
to an increase in the number of more mobile OH- ions. Hence conductance increases sharply.
22