Carbondioxide Sensor
Carbondioxide Sensor
Instruction Sheet
for the PASCO
Model CI-6561
stopper
5. Use the switch on the top of 2. Insert the sensor assembly and probe into the
the sensor to select a upright sampling bottle. Be sure the stopper fits
measurement range: snugly into the neck of the bottle (See Figure 2).
Figure 1: Range
10K: 0 to 10,000 ppm or Switch
100K: 0 to 100,000 ppm. 3. Bring the sensor/bottle assembly inside and connect
it to a ScienceWorkshop interface. Ensure the
6. In the Setup window, double click on the CO2
interface is turned on and connected to a computer.
Sensor icon, click on the Measurement tab, and
select the concentration range (10K or 100K). The 4. On the top of the Carbon Dioxide Sensor, move the
range is 10K by default. Click OK. range switch button to the 10K range position.
7. (Optional): To calibrate the sensor, follow the (Note: This is the default position.)
calibration instructions on this page.
5. In the DataStudio Setup window, double click on
Note: To change the sampling rate (rate of data the CO2 Sensor icon (to open the Sensor Properties
collection), use the Plus and Minus buttons in the dialog); then click on the Measurement tab, and
General tab of the Sensor Properties dialog. The select the 10K concentration range. Click OK.
sampling rate depends on the interface used. The
default rate is 1 sample/second. One sampling cycle 6. In DataStudio, open a display and click the Start
occurs every time the green light flashes. Although button to observe the carbon dioxide reading. Wait
the sample rate of the interface can be set much at least 90 seconds for the reading to stabilize.
higher, the sensor only updates once per second.
7. Press the Calibrate button on the sensor and hold
8. To collect data, open a display and click the Start for 3 seconds. The green light will stay on to
button on the main toolbar. indicate calibration has begun.
8. Wait approximately one minute. When the green
light flashes again, calibration is finished.
2 ®
012-08402A Carbon Dioxide Sensor
9. To verify the calibration, observe the sensor’s Suggested Experiment: Cellular Respiration
reading in DataStudio. The reading should be stable of Peas
at approximately 400 ppm (±50 ppm).
Equipment required: 1/4 cup of pea seeds, piece of
Note: If the sensor’s reading drifts or seems too high dark cloth, CI-6561 CO2 Gas Sensor with bottle and
or low, recalibrate. stopper, ScienceWorkshop interface, DataStudio,
Temperature Sensor (optional)
Collecting Carbon Dioxide Samples
Use the supplied sampling bottle to collect air or other 1. Connect the Carbon Dioxide Gas Sensor to a
samples, such as an atmospheric sample or the CO2 ScienceWorkshop interface.
gas produced from germinating seeds. Small amounts 2. Put the dry pea seeds into the bottle supplied with
of solid samples, such as plants or other organisms, the sensor. Insert the probe with stopper and
can be placed in the bottom of the bottle. sensor into the bottle.
3. In your data collection software, click the Start
CAUTION: Do not try to remove the probe button to take a reading of the concentration of
from the sensor. The probe is not removable.
carbon dioxide gas in the bottle. Record data for
To collect an atmospheric sample: When collecting about 10 minutes; this is the baseline reading
air, keep the bottle upright and the neck open (without before germination.
the stopper). Wait long enough to allow atmospheric 4. Disconnect the sensor from the sensor/bottle
air to fill the bottle. Insert assembly.
the probe with stopper
5. Moisten the peas with a few teaspoons of water.
into the top neck of the
Place a dark cloth over the bottle, and allow the
bottle. Ensure the stopper
peas to germinate in the bottle overnight.
fits snugly into the neck
of the bottle before taking 6. Following germination, reconnect the sensor to the
a reading. interface and click the Start button. Monitor the
CO2 gas reading for about 10 minutes.
To collect expired air or
CAUTION: Do not allow the probe or sensor
other gas samples:
assembly to get wet at any time!
Collect the sample in a
plastic bag and clamp the 7. Compare the rate of carbon dioxide generation
Figure 2: Sensor
bag. To take a reading, placement in bottle before and after germination. Explain your results.
unclamp the bag and What is causing the change in carbon dioxide
immediately place the concentration?
plastic bag over the probe and stopper. In some cases, 8. (Optional): Repeat step 6 with the bottle placed in
a valve, tubing with stopcock, or chamber may be water baths of different temperatures.
necessary to prevent atmospheric gas from entering
Note: If you have a Temperature Sensor or
the bag and mixing with the collected sample.
thermometer available, you can monitor both the CO2
and temperature every 5 minutes for 20 minutes.
® 3
Carbon Dioxide Sensor 012-08402A