INSTRUCTION MANUAL

ultrasonic level meters ULM – 70

Firmware: v.2.0

Read carefully the instructions published in this manual before the first use of the level meter. Keep the manual
at a safe place. The manufacturer reserves the right to implement changes without prior notice.
Table of contents

1 . Basic description.......................................................................................................................... 4
2 . Range of applications.................................................................................................................. 4
3 . Variants of sensors...................................................................................................................... 5
4 . Dimensional drawings................................................................................................................ 5
5. Installation and putting into operation................................................................................. 6
6. Mechanical mounting ................................................................................................................ 6
7. Electrical connection................................................................................................................... 10
8. Examples of ULM-70 connection............................................................................................... 12
8.1 . Wiring diagram of the level meter with current output and PDU unit.......................... 12
8.2 . Wiring diagram of the level meter with current output and PDU unit
in versions for explosive areas............................................................................................ 13
8 .3 . Wiring diagram of the level meter with current output and MGU unit......................... 15
8 .4 . Wiring diagram of the level meter with Modbus / RS485 and MGU unit...................... 15
8 .5 . Wiring diagram of the level meter with Modbus / RS485 and PDU unit....................... 15
9 . Setting elements.......................................................................................................................... 16
10 . Operation and Settings............................................................................................................ 17
10.1 . Basic configuration............................................................................................................ 17
10.2 . Service settings................................................................................................................... 20
10 .3 . Additional functions.......................................................................................................... 22
11. Protocol HART®........................................................................................................................... 25
12. Protocol Modbus........................................................................................................................ 25
13 . Function and status indication.............................................................................................. 25
14 . Order code................................................................................................................................... 26
15 . Accessories.................................................................................................................................. 27
16 . Safety, protection, compatibility and explosion proof..................................................... 27
17 . Use, manipulation and maintenance................................................................................... 28
18 . Marking of labels....................................................................................................................... 28
19 . Technical specifications........................................................................................................... 30
20 . Packing, shipping and storage................................................................................................ 32
21 . Menu structure.......................................................................................................................... 33
Used symbols

To ensure maximum safety of control processes, we have defined the following safety instructions
and information. Each instruction is labelled with the appropriate pictogram.
Alert, warning, danger
This symbol informs you about particularly important instructions for installation and operation of
equipment or dangerous situations that may occur during the installation and operation. Not observing
these instructions may cause disturbance, damage or destruction of equipment or may cause injury.

Information
This symbol indicates particularly important characteristics of the device.

Note
This symbol indicates helpful additional information.

Safety

All operations described in this instruction manual have to be carried out only by trained
personnel or an accredited person. Warranty and post warranty service must be exclusively
carried out by the manufacturer.
Improper use, installation or set-up of the level meter can result in crashes in the applica-
tion (overfilling of the tank or damage of system components).
The manufacturer is not responsible for improper use, losses of work caused by either
direct or indirect damage, and for expenses incurred during installation or use of the
level meter.

1. Basic description

Ultrasonic level meter ULM® is a compact measuring device consisting of two parts - main level
meter (the body with measuring electronics) and display module. Using the electroacoustic
converter, the level meters transmit the sequence of ultrasonic pulses that spread towards the
surface level. The converter recuperates reflected acoustic waves that are subsequently pro-
cessed in the electronic module. The intelligent evaluation block filters out interfering signals,
compares the cleaned received signal with the false reflection map (e.g. from mixers, ladders,
reinforcement etc.) and selects a suitable reflection (echo). Based on the period during which
the individual pulses spread towards the surface level and back and based on the measured
temperature in the tank, the instant distance to the surface level is calculated. According to the
level height, the level meter output is set and the measured value is displayed on the display.

2. Range of applications

Thanks to their non-contact sensing principle, level meter are suitable for continuous level
measurement of liquids, wastewater, sludge, bulk materials, suspensions, adhesives, resins in
various open and closed containers, sumps, channels or troughs. The applicability for measur-
ing the level of bulk materials is limited and the measuring range is shortened.

4 ULM–70 © Dinel, s.r.o.


3. Variants of sensors

ULM­­–70_­–02 Measuring range from 0.15m to 2m, plastic PVDF transmitter,


mechanical connection with thread G 1".

ULM­–70_–06 Measuring range from 0.25m to 6m, plastic PVDF transmitter,


mechanical connection with thread G 1 ½".

ULM­–70_–10 Measuring range from 0.4m to 10m, plastic PVDF transmitter,


mechanical connection with thread G 2 ¼".

ULM–70_–20 Measuring range from 0.5m to 20m, plastic PVDF transmitter,


mechanical connection with aluminium alloy flange.

4. Dimensional drawings
ULM–70_–02 ULM–70_–06 ULM–70_–10
Pg11 cable gland
M16 x 1,5

Ground
terminal

ULM–70_–20

variant ULM-70
with protective conductor

po 45° po 45°

Flange (for type 20) according to


standard: DIN 2527, PN10, DN100
© Dinel, s.r.o. ULM–70 5
5. Installation and putting into operation

Please follow next 3 steps:


• Mechanical mounting - see chapter 6
• Electrical connection - see chapter 7
• Settings - see chapter 9

6. Mechanical mounting

• Install the level meter in the vertical position into


the upper lid of the tank or reservoir using a weld-
ing flange, a fastening nut or a flange so that the
level meter axis can be perpendicular to the sur-
face level of the measured liquid (Fig. 1).
• The min. dimensional parameters to install the
level meter into a lid or a ceiling of a tank are given
in Fig. 3.
• When installing in an open channel (reservoir,
drain etc.), install the level meter onto a bracket as Fig. 1: Recommended installation in the tank
close as possible to the expected max. level.
• The reference plane for the measurement is the d > 1/12 c
ULM–70–02 ; 10
(min. 200 mm)
lower edge of the transducer (Fig.2). In connec-
tion with the measurement principle, no signals d > 1/8 c
ULM–70–06
reflected in the area immediately under the level (min. 200 mm)
meter can be evaluated. The zone (Fig. 2) deter- ULM–70–20
d > 1/10 c
mines the min. distance possible between the (min. 200 mm)
level meter and the highest surface level. The min.
distances to the medium are given in the chapter
"Specifications".
• It is necessary to install the level meter so that
the bin level cannot interfere with the dead zone
when filled up to the maximum. If the measured
level interferes with the dead zone, the level meter
will not work properly.

d – Distance from
m – Dead zone the tank wall
(blind zone, blocking reference plane for
the measurement c – Maximum reach
distance)
of the level meter

Fig. 2: Level meter dead zone Fig. 3: Installation distance from the tank wall

6 ULM–70 © Dinel, s.r.o.


• If the maximum surface level in the tank interferes with the dead zone, the level meter has
to be mounted into a higher installation neck. In this way, the tank can be filled nearly up
to the maximum volume. The inner neck surface has to be even and smooth (without edg-
es and welded joints); the inner edge should be rounded where the ultrasonic wave leaves
the pipe. The neck diameter should be as large as possible but the neck height should be
as low as possible. Recommended dimensions of the input neck are given in Fig. 4.

a – Neck height
b – Neck width
m –Dead zone (blind zone,
blocking dist.)
ULM–70– a<3b
02 ; 06 b > 100 mm
a < 1,5 b
ULM–70–10
b > 100 mm
a < 1,5 b
ULM–70–20
b > 150 mm

Fig. 4: Possible installation of the installation neck

• If the emitted acoustic signal of the level meter is affected by near objects (roughness on
walls of the tank, various partitions, mixers etc.), it is necessary to map false reflections
by activating the mode "TEACHING". In case of installed mixers, it is necessary to put the
mixers to position under the level meter (direct the mixer paddle to the ultrasonic signal
beam) (Fig. 5 and 6).

Fig. 5: False echo from obstacles in the tank Fig. 6: False echo from the mixer paddle

• Do not install the level meter in or above the filling point (Fig. 7).

Fig. 7: Level meter installation outside the influence of filling

© Dinel, s.r.o. ULM–70 7


• In case the level of bulk solids is measured,
the measurement range is reduced Due to horn
adapter
absorption of acoustic waves by a bulk me-
dium, shortening of the measuring range
occurs by up to 50% depending on the grain
size. We therefore recommend selecting a
level meter with greater range than the max-
imum range of measuring the medium. It is
also appropriate to use a directional horn
(see image 8), which reduces the shorten-
ing of the measuring range, because it bet-
ter concentrates acoustic energy while pre-
serving the same beam angle, and improves
the sensitivity when receiving the reflected
echo. We recommend to consult the use with
the manufacturer.

Fig. 8: Level meter installation in silo or hopper

• During filling, mixing and other processes, foam


can arise on the surface level of the measured liquid.
The foam considerably absorbs the ultrasonic sig-
nal which might cause malfunction of the level meter
(Fig. 9). For such cases, it is necessary to set up "SENSITIV-
ITY" mode to "high" or contact the manufacturer if need.
In case of a thin layer of foam, it is also possible to use
the directional horn for improving receipt of the reflec-
ted echo.

• Scattering or attenuation of Fig. 9: Foam on the surface


the ultrasonic signal can result
if the surface level has been
moderately stirred or rippled
(by a mixer, coming liquid etc.).
It can result in reduction of the
measurement range or unreli-
able function of the level meter
(Fig. 10).
• Rotating mixer blades can cause
that the surface is stirred, which
results in false reflections of the
ultrasonic signal from the surface
level and unreliable operation of
the level meter (Fig. 13).(obr. 11).
For a rippled or swirling level,
you can use the directional Fig. 10: Moderately stirred Fig. 11: Intensely stirred surface
horn to eliminate scattering of surface
the ultrasonic signal.

8 ULM–70 © Dinel, s.r.o.


• If the level sensor is mounted to ventilation
bottlenecks and places with ba- holes
rriers, or near uneven walls or
the filling area, where the tran-
sleeves for
smission signal could be dis-
mounting
torted, we recommend using a
guide tube (acoustic horn). The
tube must be made from a sin-
gle material with a smooth inner
surface (see image 12a, 12b ).
The minimum diameter of the
pipe shall be dimension "h"
as shown in the table below.
After installing, you must perform
the procedure "TEACHING". We
recommend consulting with the
manufacture on the construction
of the guide tube.
ULM–70–02 h ≥ 70 mm
Fig. 12a: Short guide
ULM–70–06 h ≥ 100 mm Fig. 12b: Total guide tube
tube installation
ULM–70–10 h ≥ 150 mm installation

ULM–70– 20 h ≥ 200 mm

• The level meter must not be installed in places with direct solar radiation and must be
protected against weather effects. If the installation in places with direct solar radiation is
inevitable, it is necessary to mount a shielding cover above the level meter(Fig. 13).
• It is suitable to run the cable under a cable bushing (obliquely down in slack) according to
Fig. 14 to prevent penetration of humidity. Then the rain and condensing water can flow
off freely.
• The cable bushing and connector have to be sufficiently tightened to prevent penetra-
tion of humidity.

Fig. 13: Solar radiation Fig. 14: Prevention to avoid


shielding cover intrusion of humidity

© Dinel, s.r.o. ULM–70 9


7. Electrical connection

The level meter is connected to consequential (evalu-


ating) device with a suitable cable with the outer
diameter of 6 to 8 mm using screw terminals located
under the display module. The recommended cross
section of cores for the current version 2 x 0,5 ÷ 0,75
mm2 and for the version with Modbus communication
2 x 2 x 0,25 mm2 (twisted pair, shielded). Plus pole
(+U) is connected to the terminal (+), minus pole (0 V)
to the terminal (-) and the shielding (only for shielded Fig. 16: Wiring diagram of the level meter
cables) to the terminal ( ). Communication wires A with current output ULM-70 _-_-_- I
and B of the line RS-485 (for version "M" - Modbus)
are connected to the terminals A and B.
metal display unit connector
clip
Procedure to connect the cable to the level meter:
1. Unscrew the nut of the upper transparent lid.
2. Take the upper edge of the display module and
take it out carefully by mild swinging up. Terminal block

3. If you cannot grasp the module, you can use


a small screwdriver. Insert it as far as the seam
and use from several sides to slightly lift the
module. Fig. 17: Inside view of screw terminals of the level
meter with current output ULM-70_-_-_-I
4. Release the cable outlet and thread the stripped
supply cable in.
5. Connect the cable to the screw terminals ac-
cording to the diagram in Fig. 17 or 18. Firmly
tighten the terminals and the cable outlet.
6. If the level meter with Modbus is involved as a
terminal for RS-485, we recommend (to avoid
reflections on the line) to connect 120Ω termi-
nation resistor. This is done by moving a small
Fig. 18: Wiring diagram of the level meter
lever of the switch marked 120Ω to the ON posi-
with Modbus ULM-70_-_-_-M
tion. On the level meters connected to the line
RS-485 as an intermediate device, the termina-
tion resistors are not connected (switch remains metal clip display unit connector
off).
7. Insert the display module back into the head so ON
120

that the connector is properly connected. 120Ω switch

8. Slide silicone seal on the thread of the level me- + IS


A B GND
ter body, then tighten the nut of the upper lid. Terminal block
Connect the cable to consequential device.

Fig. 19: Inside view of screw terminals of the


level meter with Modbus ULM-70_-_-_-M

10 ULM–70 © Dinel, s.r.o.


Data connection RS-485 / MODBUS:
Maximum distance of the module from the continuous line (length of T segment) is 3 m (fig. 35).
At the end units, it is necessary to connect terminating resistors Rz 120 Ω, whilst the terminating
resistors must not be connected on the continuous units(fig. 35).
The cable must be a shielded twisted pair with a core cross-section of 0.35 .. 0.8 mm2 with an imped-
ance of approximately 120 Ω.
The cable shielding is connected to a shielding clamp on the connector of line RS485 and is con-
nected with the clamp of the PE switchboard (direct grounding) only at a single point.
In the event that the RS-485 line is lead outside of one lighting arrester, it is necessary to protect it
via suitable overvoltage protection.
In the event of communication problems caused by strong interference, it is appropriate to install
the system inside a metal switchboard and to install the strong interference sources (e.g. frequency
converters) away from this switchboard.

GRLM-70 MODBUS GRLM-70 MODBUS GRLM-70 MODBUS FCU-400


Switch ON (120 Ω) Switch OFF Switch OFF Switch SW. 1 ON
(120 Ω)
Fig. 20: Connection of multiple units via RS485 to a network

Electrical connection can only be made when de-energized!


The source of the power voltage must comprise of a stabilised safe low power source with
galvanic separation. In the event that a switch-mode power supply is used, it is essential that
its construction effectively suppresses common mode interference on the secondary side. In
the event that the switch-mode power supply is equipped with a PE safety terminal, it must
be unconditionally grounded! Spark-safe devices type ULM–70Xi must be powered from
a spark-safe power source meeting the above-mentioned requirements.
Due to the possible occurrence of an electrostatic charge on the non-conductive parts of the
sensor, it is necessary to ground all sensors intended for use in environments with an explosion
hazard ULM-70Xi. This is done using the screw located on the head of the level gauge under
the cable gland. Connect the screw directly at the point of installation of the level gauge to
the conductive tank or to a conductive grounded structure.
In the event that the sensor is installed in an outdoor environment at a distance greater than 20
m from the outdoor switchboard, or from an enclosed building, it is necessary to supplement
the electrical cable leading to the sensor with suitable overvoltage protection.
© Dinel, s.r.o. ULM–70 11
User manual - METER SRP-N118

Power +24V
a) b)
optional ACTIVE supply
In the event of strong ambient electromagnetic interference, paralleling(depending
of conductors with power
on version) current output
distribution, or for distribution to distances over 30 m, we recommend grounding the level meter - +
(see 6above)
mm and using a shielded cable. Ground the shielding of the cable 1on the2 side of the

max. 1.6 mm
3 4
max. 2 mm

5 mm
power supply, or the shielding is possible to connect only on inside pin of the level meter marked
see fig. 17 and 19 (the shielding of the cable is always connected in a single
10 11location).
12 13 14 15
n.c. n.c.
Figure 4.2. Recommended dimensions of cable stripping
Measures
a) for must also
big connectors (1 tobe
6),designed and
b) for small implemented
connectors in the
(7 to 13, 15 electrical
to 17) installation to reduce the
effects of static electricity to a safe level. R1 R2
(optional) (optional)
Installation in hazardous areas must be carried out in accordance with EN 60079-14 (Electrical
equipment for explosive gaseous atmospheres - Part 14: Electrical installations
Figure in hazardous
4.7. Terminals description (relay ou
areas other than mining areas) and, where applicable, in accordance with other standards
that apply to the area. Power +24V
supply (optional)
(depending on version) OC4
- +
1 2 3 4
8. Examples of ULM-70 connection 10 11 12 13 14 15
- + + - + -
OC1 OC2 OC3
8 .1 . WIRING DIAGRAM OF THE LEVEL METER WITH CURRENT OUTPUT AND PDU UNIT
Figure 4.3. Method of connecting cables to the clamping connectors (optional) (optional) (optional)

OC1 ÷ OC4: Umax = 30V DC,


! All connections must be made while power supply is disconnected !
Imax = 30mA, Pmax = 100mW

PDU-420-W PDU-4xx-P Figure 4.8. Terminals descript


(input 4 ... 20 mA ) (input 4 ... 20 mA )
0/1 - 5V
0/2 - 10V R1 R2
(option) (option)
GND + connected
internally internally
connected
RS - 485 connected internally
+ -
5 6
17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 16 17 18 19 20
GND

PE

n.c.

n.c.
DATA+
DATA-

31 32 33 34 35
+ -
Power supply

Uo
(depending
on version)

GND + 24V (+5%, -10%)


0/4 - 20mA
(0V) - (+U)Imax
+ = 100 mA (+U) + + -
(0V) -

Figure 4.4. Terminals description (relay outputs)


ULM-70N-_ _-_-I ULM-70N-_ _-_-I
Figure 4.9. Connection of 2-w

Connection of PDU-420-W is valid for firmware version 6.00 or higher. The older ver-
sions (up to version 5.99), the level meter output +U is connected to the terminal 7 and
the output 0V to the terminal 10.

12 ULM–70 © Dinel, s.r.o.


Figure 4.3. Method of connecting cables to the clamping connectors
8 .2 . WIRING DIAGRAM OF THE LEVEL METER WITH CURRENT OUTPUT
! All connections must be made while power supply is disconnected !
AND PDU UNIT IN VERSIONS FOR EXPLOSIVE AREAS

Figure 4.3. Method of connecting cables to the clamping connectors

! All connections must be made while power supply is disconnected !


0/1 - 5V
0/2 - 10V R1 R2
(option) (option)
GND PDU-420-W-230V
+
RS - 485 0/1 - 5V
(input 4 ... 20 mA )
0/2 - 10V R1 R2
(option) (option)
GND +
RS - 485
17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
GND

PE

n.c.

n.c.
DATA+
DATA-

GND

PE

n.c.

n.c.
DATA+
DATA- + +Uo -

Power supply
Uo -

Power supply

(depending
on version)
+

(depending
GND

on version)
GND + 24V24V
(+5%,
(+5%, -10%)
-10%)
0/4 - 20mA Imax = 100 mA
0/4 - 20mA Imax = 100 mA

Figure 4.4. Terminals description (relay outputs)


Figure 4.4. Terminals description (relay outputs)

ULM-70Xi-_ _-_-I

(+U) +

(0 V) –

IRU-420-I

N L
explosive non-explosive
area area 230 V AC

Connection of PDU-420-W is valid for firmware version 6.00 or higher. The older ver-
sions (up to version 5.99), the terminal 9 of the IRU unit is connected to the terminal
10 of the PDU unit and the terminal 10 of the IRU unit is connected to the terminal 11
of the PDU unit.

© Dinel, s.r.o. ULM–70 13


Power RS - 485
(optional) Imax = 100mA
supply GND DATA
(depending on version) OC4 + Uo - DATA-
- +
0/1 - 5V 1 2 3 4 5 6 7 8 9
0/2 - 10V R1 R2
16 17 18 19 20
10 11 12 13 14 (option)
(option) 15 31 32 33 34 35
GND + - + + - + - n.c.
RS - 485 + + n.c.
OC1 OC2 OC3
(optional)
User manual - METER SRP-94 (optional) (optional) 0/4-20mA
User manual - METER SRP-94 GND

17 16 1615 Double
14 13 12PDU-4xx-P-230V
11 10 9 8 7 6 P5max = 100mW
Imax = 30mA,
OC1 ÷ OC4: Umax = 30V DC,
4 3 2 10/1 - 5V, 0/2 - 10V
(input 4 ... 20 mA )
numeration means, that depending on device version, particular
31Double

GND

PE

n.c.

n.c.
16 numeration means, that depending
to the topon device number.
version, particular

DATA+
DATA-
terminal can be marked according or bottom
31
terminal can be marked according to the top Figure
+ Uoor bottom4.8.number. -
Terminals description (OC-type outputs)

Power supply
(depending
on version)
GND
Power
R4
+
R4
(optional) 24V
+24V (+5%, -10% RS - 485
+5%, -10%)
Power
supply
0/4 - 20mA
(optional) Imax
+24V = -10%
Imax
+5%, 100
= 100mARS
mA GND
- 485 DATA+
supply + UoGND
Imax = 100mA - internally
DATA-connected
DATA+
(depending on version)
(depending on version) + Uo+- - DATA-

1 2
3 4 5
5 6 6 7 8 9
1 2 3 4 5 166 17 7 18 8 19 9 20 16 17 18 19
Figure104.4. Terminals
11 12 13 14 15 description
31
31 32
(relay
32 3333 34
outputs)
34 35 35 31 32 33
10 11 12 13 14 15 31 n.c.
32 33 34 35n.c.
n.c. + + n.c.
GND
0/4-20mA
+ +
- + GND
0/4-20mA
R1 R2 R3
(optional) (optional) (optional)
R1 R2 R3 0/1 - 5V, 0/2 - 10V
(optional) (optional) (optional) 0/1 - 5V, 0/2 - 10V
Figure 4.6. Terminals description (relay outputs)
Figure 4.6. Terminals description (relay outputs) 9
Power +24V +5%, -10% RS
Figure - 485
4.9. Connection of 2-wire current converters
Powersupply Imax = 100mA RS - 485
(optional)+24V +5%, -10% GND DATA+
supply
(depending on version) OC4
(optional)
+ UoGND
Imax = 100mA - DATA- DATA+
+
(depending on version) OC4
- + Uo - DATA-
1 2 + 3- 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
10 11 12 13 14 15 31 32 33 34 35
10 11
- 12
OC1
+ 13
+ 14
- 15
+ - 31 n.c.
32 33 34 35n.c.
OC2 OC3 + + GND
+ (optional)
- (optional)
+ - + (optional)
- n.c. 0/4-20mAn.c.
ULM-70Xi-_ _-_-I OC1 OC2
OC1 ÷(optional)
(optional) OC4: Umax
OC3
= 30V DC,
(optional)
+ +
0/4-20mA
GND

Imax = 30mA, Pmax = 100mW 0/1 - 5V, 0/2 - 10V


OC1 ÷ OC4: Umax = 30V DC,
Imax = 30mA, Pmax4.7.
(+U) + Figure = 100mW
Terminals description
0/1(OC-type
- 5V, 0/2 - 10Voutputs)

Figure 4.7. Terminals description (OC-type outputs)

(0 V) –

IRU-420-I

11

11

N L
230 V AC
explosive non-explosive
area area

14 ULM–70 © Dinel, s.r.o.


8 .3 . WIRING DIAGRAM OF THE LEVEL METER WITH CURRENT OUTPUT AND MGU UNIT

ULM-70N-_ _-_-I
(+U) +

(0 V) - MGU-800
(with module IUI4)

8 .4 . WIRING DIAGRAM OF THE LEVEL METER WITH MODBUS / RS485 AND MGU UNIT

MGU-800
ULM-70N-_ _-_-M (without additional modules)
(0V) WH
(+U) BN

Cab.Sh.
(B) GN
(A) YE

8 .5 . WIRING DIAGRAM OF THE LEVEL METER WITH MODBUS / RS485 AND PDU UNIT

+24V DC
RS 485 - Master Imax=120 mA
_
S A B +
12 13 14 18 19
(0V) WH
Cab.Sh.

(+U) BN
(B) GN
(A) YE

ULM-70N-_ _-_-M

© Dinel, s.r.o. ULM–70 15


9. Setting elements

Settings are performed using 3 buttons located Display


of measured values
on the display module DM-70. All the settings
are available in the menu of the level meter.

button
• Set-up mode access
• Confirmation of selected item in the menu Set-up elements
• Move the cursor in the line
• Saving of set-up data Teaching
mode activation
button
Units
• Move in the menu Echo receiving* warning and infor-
• Change of values mation inscriptions
Lock of level
button meter
• Cancelling of carried out changes * Slow flashing while the reflected signal (echo)
• Shift one level up is received from the measured level.

• Status indication (left lower corner of the display):


symbol „E“ -a  lternating flashing - correct reception of the reflected wave indicating the level
symbol „T“ - permanent shine - "TEACHING" mode is activated
- inverse shine - activation of the "TEACHING" mode
symbol - permanent shine - the level meter is locked against unauthorized setting using
a password, enter password to unlock (see MENU - PASSWORD)
• Warning inscriptions:
NO ECHO - when empty tank, after you perform the procedure TEACHING
- level meter is not able to measure (check the media or change the sensitivity)
FIXED OUTPUT - the output stream is fixed to a constant value (see DIAGNOSTIC - CUR-
RENT)
LOW POWER - low  supply voltage (must be in the range - see Technical specifications)
NO PASSWORD - the level meter is protected using a password against unauthorized setting
NO DATA AVAILABLE - display module doesn't communicate with the electronics of the lev-
el meter (e.g. incorrectly inserted display module into connector or
measuring module is not functional).
• Information inscriptions:
DISTANCE TO LEVEL - the display shows the current distance
(see DIAGNOSTIC - DISTANCE)
CURRENT - the display shows the current proud (see DIAGNOSTIC - CURRENT)

Level meter GRLM-70_-__-_-_-_-L is supplied without the display module (display) DM-70. To
setup the level meter, it is necessary to connect a display module to it (or it can be configured via
HART or MODBUS). When the settings are completed, the display module may be disconnected
and the level meter then measures without it.

16 ULM–70 © Dinel, s.r.o.


10. Operation and setting

Set the level meter using 3 buttons placed on the display module U
UNNIITTSS
(see Chapter Set-up elements). After 5 min. of inactivity, the level LEVEL: mm
meter automatically returns back to the measurement mode. If the DISPLAY: %
password is active, the level meter will be also locked. The val-
ues that have not been confirmed using the button will not be SAVE

saved! After the meter is locked, you cannot change the setting!
When you attempt to edit, the words "NO PASSWORD" will appear
on the display. How to unlock the level meter is given on page 16. Dinel®
After connection of the supply voltage to the level meter the display
shows the logo "Dinel" and the text "Starting" (approx. 15 s). Then, STARTING

the level meter goes to the measuring mode and the display shows
the current measured value.

10 .1 . BASIC CONFIGURATION
BASIC SETTINGS
After the first start of the level meter it is necessary to perform the SERVICE
DIAGNOSTIC
basic configuration (setting of the measuring range, choice of units CLONE SETTINGS
PASSWORD
and possibly damping). The settings are accessible in the basic LANGUAGE
INFO
menu by pressing the "BASIC SETTINGS".

MIN LEVEL and MAX LEVEL


You can freely define the minimum / maximum distance from the front surface of the
level meter (item "LEVEL"for currents 4 / 20 mA). The "DISPLAY" is intended to set the value
displayed on the display. Setting the units is done in the "UNITS".

MIN A CTU A L L EVEL


MIN LEVEL 0 2 8 7 0 mm
MAX LEVEL
UNITS OUTPUT: 0 4 . 0 0 mA
DAMPING
LEVEL: 0 6 0 0 0 mm
SENSITIVITY
TEACHING DISPLAY: 00000%
ACTUAL LEVEL: Actual distance to
level
OUTPUT: current 4 mA / 20 mA
MIN LEVEL
MAX A CTU A L L EVEL
0 2 8 7 0 mm
LEVEL: Definition of the min / max
MAX LEVEL
UNITS OUTPUT: 2 0 . 0 0 mA level
DAMPING
LEVEL: 0 0 2 5 0 mm
SENSITIVITY
TEACHING
DISPLAY: The value showed on the
DISPLAY: 00100%
display

If in the bottom of the display appears (when entering the values) the ​​inscription "OUT OF LIM-
ITS", the value specified for the item "LEVEL" is outside the measuring range of the level meter.
If the inscription "SPAN TOO SMALL" is shown, it must be specified a larger span between Min
and Max values. For more information, see chapter "Specifications".
The decimal point position of the item 'LEVEL' is firmly set (according to the selected units), in
the item "DISPLAY" it is freely adjustable

© Dinel, s.r.o. ULM–70 17


1. To enter to the menu press the same button to select "BASIC SETTINGS". Then, using
and select "MIN LEVEL" or "MAX LEVEL".

2. Now it is shown the item "MIN LEVEL" ("MAX LEVEL"). By pressing and set the out-
put current "OUTPUT", the distance for the defined current "LEVEL" the value ​​on the dis-
play "DISPLAY".

3. By pressing button save the data. By next presses of the button leave the menu.
The level meter returns to measurement mode.

UNITS
Level meter can process and convert a large number of different physical values. The setting
is done in the item "UNITS".

LEVEL: Unit selection (mm, cm, m, in, ft)


UNITS
MIN LEVEL
MAX LEVEL DISPLAY: The unit showed on the display
UNITS LEVEL: mm
DAMPING (%, mm, cm, m, in, ft, l, hl, m3, gal, bbl,
DISPLAY: %
SENSITIVITY
TEACHING
mA)
TEMPERATURE °C
TEMPERATURE: Temperature unit (°C, °F)

1. To enter to the menu press the same button to select "BASIC SETTINGS". Then, using
and select "UNITS".

2. Now the menu item "UNITS" is shown. By pressing the and button make the set-
tings of individual items.

3. By pressing button save the data. By next presses of the button leave the menu.
The level meter returns to measurement mode.

DAMPING
Setting the response time of the measurements. The function is useful for suppressing
level fluctuations, waves and rapid changes of the level. The reaction time will depend on the
exponential function. Damping with a defined delay in seconds represents the time when
exponential reaches 2/3 of its maximum value.

DAMPING
MIN LEVEL
MAX LEVEL
UNITS
DAMPING 05 The damping time can be set in the
SENSITIVITY
TEACHING
interval from 0 to 99 s.

1. To enter to the menu press the same button to select "BASIC SETTINGS". Then, using
and select "DAMPING".

2. Now the menu item "DAMPING" is shown. By pressing the and button make the set-
tings of individual items.

3. By pressing button save the data. By next presses of the button leave the menu.
The level meter returns to measurement mode.

18 ULM–70 © Dinel, s.r.o.


SENSITIVITY
The setting is defined in three steps of the level meter sensitivity.
"LOW" – Low sensitivity in case of surrounding interferences affecting the measurement.
"MEDIUM" – Medium sensitivity (suitable for most applications).
"HIGH" – Enhanced sensitivity for measured media partly absorbing the ultrasonic signal
MIN LEVEL SENSITIVITY
MAX LEVEL
UNITS
You can set the sensitivity in three
DAMPING LOW
degrees:
SENSITIVITY
TEACHING
LOW – MEDIUM – HIGH.

TEACHING
The mode serves for suppressing false reflections resulting from reflection of the ultra-
sonic signal from roughnesses on walls of the tank, various partitions, mixers or other obsta-
cles. The sensor starting this mode detects false reflections and save them in the memory.
Then these false reflections will not affect the subsequent measurement (they are masked).
Before starting the mode it is necessary to empty the tank as much as possible (pref-
erably completely).
TEACHING
MIN LEVEL
MAX LEVEL
UNITS
SET LEVEL DISTANCE:
If there are no above obstacles
DAMPING 06000 mm
SENSITIVITY in the tank, it is not necessary to
TEACHING START
start this mode.

1. To enter to the menu press the same button to select "SERVICE". Then, using and
select "TEACHING".
2. Now it is shown the item "TEACHING". By pressing set the value "LEVEL DISTANCE" (dis-
tance to the level) – supposed distance from the face of the sensor to the medium level. If the
distance to the level is not precisely known, enter a value rather lower (in the tolerance field
as shown in Fig. 19).
3. After entering the "SET LEVEL DISTANCE" by pressing button the system starts "teaching"
(false reflection mapping). During the mapping, the display shows flashing sign "RUNNING".
4. The mapping of false echoes can be completed when you see the inscription "Press OK to
stop" and you press .
5. The procedure is completely finished when you can see the inscription "DONE". It is then
possible to exit the menu by repeated pressing the button .

TEACHING TEACHING
SET LEVEL DISTANCE: SET LEVEL DISTANCE: The mode "TEACHING" will stop
06000 mm 06000 mm automatically after ca. 1000 meas-
RUNNING DONE urements.
PRESS ESC TO EXIT

If during the scanning of the tank in the bottom of the display


appears the dialog "press OK to stop" (see figure) the level meter TEACHING
already found no further obstacles and "TEACHING" mode may be SET LEVEL DISTANCE:
terminated. If it is not terminated, the level meter is still ready for 06000 mm
the possible presence of obstacles (e.g. paddles of the agitator). RUNNING
Once it registers a further obstacle, the dialogue disappears and PRESS OK TO STOP

the obstacle is erased. This process may be repeated up to 1000


cycles. After this the "TEACHING" mode is automatically stopped.

© Dinel, s.r.o. ULM–70 19


In case of installed mixers, it is necessary to
position the mixers under the level meter (di-
rect the mixer blade to the ultrasonic signal Obstacles

beam).

Level Distance
Note: If there are significant obstacles in the
upper half of the tank, multiple false reflec- 1/2 of Tank
tions can occur especially in closed tanks. In
such cases it is necessary to reduce the level in
the tank as much se possible to correctly mask Tolerance zone for
"LEVEL DISTANCE" value
these possible multiple false reflections.
Fig. 21: Level distance zone
10 .2 . SERVICES SETTINGS
In the supplemented configuration, you can set parameters BASIC SETTINGS
of sensitivity, mapping of false reflections, temperature differ- SERVICE
DIAGNOSTIC
ence compensation, behaviour in case of fault conditions or CLONE SETTINGS
PASSWORD
HART® communication. Here, you can set the sensor into the INFO

initial state or reset it as well. The settings are accessible in the


basic menu under the item "SERVICE".

MEDIUM TEMPERATURE
The level meter is equipped with automatic temperature compensation. If for instance in
the tank there is a difference of 10°C between the temperature of the measured material
(medium) and the temperature at the mounting site of the level meter (see the mode "DIA-
GNOSTICS, page 20), the measuring accuracy will be reduced by around 1% of the set range. If
this function is activated, this temperature difference can be compensated. If in the tank (open
channel) is a big difference between the temperature of the measured medium (liquid) and
temperature in the place of installation of the ULM (see mode, "DIAGNOSTICS" page 20), it is
advised to improve the precision of the measurement by the zone temperature compensation.
Otherwise, this mode is not necessary to run.

MEDIUM TEMPERATURE MEDIUM TEMPERATURE Inactive compensation (initial state), the


FAILURE MODE
HART YES word "NO" appears on the display.
FACTORY DEFAULT
RESET
+20°C See the "UNITS" menu for tempera-
ture unit selection (°C or °F).

After start of the zone temperature compensation mode it is necessary to set the tempera-
ture of the surface of the medium. The level meter then calculates the average value from the
medium temperature and the temperature at the installation place of the level meter. With
such an average temperature it counts in calculating of the velocity of acoustic waves propaga-
tion and for determination of the level position.

FAILURE MODE
It defines the output current of the level meter when the measured medium level is in the dead
zone ("DEAD ZONE") or outside the measurement range in case of echo loss ("NO ECHO").

NO ECHO: Current in case of echo loss


MEDIUM TEMPERATURE FAILURE MODE
FAILURE MODE
HART
DEAD ZONE: Dead zone current
NO ECHO: 3.75 mA
FACTORY DEFAULT
RESET DEAD ZONE: 22 mA The values can be set in three steps: 3.75 mA,
22 mA and LAST (last measured data).

20 ULM–70 © Dinel, s.r.o.


HART
HART® mode (point to point, multidrop) and multidrop mode address setting. Up to 15 units
can be connected to one two-wired cable in the multidrop mode.

MEDIUM TEMPERATURE
FAILURE MODE
HART In case of the address "00", the point to
POLLING ADDRESS
HART
FACTORY DEFAULT
point mode is enabled. The range from
00
RESET "01" to "15" is reserved for addresses in
the multidrop mode.

MODBUS
This item is part of a menu with Modbus output level meter ULM-70 _-_-_- M. Modbus mode is
intended for the settings of the level Modbus addresses, baud rate and parity settings.

MEDIUM TEMPERATURE MODBUS


ADDRESS: 1 to 247 (default 1)
FAILURE MODE
ADDRESS: 001
BAUD RATE: 4800, 9600, 19200
MODBUS
FACTORY DEFAULT
BAUD RATE: 9600
(default 9600)
RESET
FORMAT: 8N1 FORMAT : 8N1, 8O1, 8E1, 8N2 (default 8N1)
FORMAT:
number of stop bits: 1
parita:N – non parity 2
O – odd parity
E – even parity
data: 8 – number of bits

FACTORY DEFAULT
To reset the initial values of the level meter set by the manufacturer, press the button
(see the Factory default table, p. 28).

MEDIUM TEMPERATURE FACTORY DEFAULT


After you press the button
FAILURE MODE "RUNNING" will be displayed for
MODBUS
FACTORY DEFAULT ARE YOU SURE? about 3 sec. After the initial values
RESET
are set, "DONE" will be appear on
the display.

FACTORY DEFAULT FACTORY DEFAULT

RUNNING
DONE
PRESS ESC TO EXIT

RESET
Complete restart of the level meter. The same effect has also a short-time interruption of the
supply voltage. To enable the resetting, press the button .

MEDIUM TEMPERATURE
FAILURE MODE
RESET
During the restart process, "RUN-
MODBUS
FACTORY DEFAULT ARE YOU SURE?
NING" will be displayed. Then the
RESET level meter will be automatically
turned off and on.

© Dinel, s.r.o. ULM–70 21


10 .3 . A dditional functions
Additional functions include modes to display temperature in the tank or to find out the actual
flowing current in the loop. Besides, to lock modifications using a password and information
about the level meter version. All of the functions are accessible from the main menu.

DIAGNOSTICS
It contains information about the actual temperature inside the BASIC SETTINGS
SERVICE
tank (or about the compensated temperature) "TEMPERATURE" DIAGNOSTIC
CLONE SETTINGS
and current flowing through the loop "CURRENT". If the tem- PASSWORD
LANGUAGE
perature compensation ("MEDIUM TEMPERATURE") is activated, INFO

the corrected temperature is displayed.

DISTANCE TO LEVEL DISTANCE TO LEVEL


TEMPERATURE
CURRENT
The temperature is measured inside
01220 mm
the tank where the level meter is
installed.
If the temperature of the measured
TEMPERATURE
DISTANCE TO LEVEL
TEMPERATURE
medium is different, we recommend
CURRENT you to carry out the temperature
+23°C
compensation "MEDIUM TEMPERA-
TURE" because of accuracy (see
p. 15). Then the displayed tempera-
ture is an average value from the
DISTANCE TO LEVEL CURRENT temperature set in the "MEDIUM
TEMPERATURE
CURRENT TEMPERATURE" and the actual tem-
17.30 mA perature measured by the sensor.

CLONE SETTINGS
This mode is intended for copying of the level meter (ULM–70 BASIC SETTINGS
body) configuration into the display module (DM–70) and SERVICE
DIAGNOSTIC
CLONE SETTINGS
back. The display module can then be removed from the level PASSWORD
INFO
meter body and put into another level meter and make there
the settings transfer (cloning).
The "CLONE SETTINGS" mode transfers all data, excluding setting of the "Teaching" and HART®.
CLONE SETTINGS CLONE SETTINGS CLONE SETTINGS
SENSOR DISPLAY SENSOR DISPLAY
MODULE MODULE
NOW CLONING
DISPLAY SENSOR DISPLAY SENSOR
MODULE MODULE
ARE YOU SURE?

1. Press to enter the menu and select the item "CLONE SETTINGS". Copying of the settings
from the body of the level meter to display module is done by selecting "SENSOR - DISPLAY
MODULE". To transfer the settings from the display module to another level meter select
the item DISPLAY MODULE - SENSOR.
2. The selected mode starts by pressing button During transmission the display shows
"NOW CLONING".

3. After completing the process in the middle of the screen displays "DONE". It is then pos-
sible to leave the menu and the mode by pressing the button .

22 ULM–70 © Dinel, s.r.o.


CLONE SETTINGS

WARNING–CLONING IS NOT
Incompatible type of level meter. Transfer of the settings can be
POSSIBLE realized only with the same type of level meter (e.g. ULM–70–02
WRONG SENSOR TYPE - ULM–70-02, ULM–70–10 - ULM–70–10) and with the firmware
PRESS ESC TO EXIT version 2.0 and later.

CLONE SETTINGS

WARNING–CLONING IS NOT The data set is not stored into the display module (DM–70). The
POSSIBLE transfer can not be done. It is necessary to repeat the procedure
NO SETTINGS SAVED
of the copying the settings in the mode "CLONE SETTINGS".
PRESS ESC TO EXIT

PASSWORD
You can lock the level meter data against unauthorized editing.
TEACHING
After activating the password the data may be read, but can not be
SET LEVEL DISTANCE:
edited. If you try to edit the settings (without true password) the 06000 mm
display shows "NO PASSWORD". START
The password can be any 5-digit numeric combination. The com- NO PASSWORD

bination of numbers 00000 is reserved for disabling the password.

1. Use the buttons and in the menu "PASSWORD" to select the mode "ENTER" for en-
tering the password or the mode "CHANGE" for changing the password (when activated,
the words are displayed inversely). Press the button once again to confirm the selec-
tion. You can change the password only when the level meter is unlocked. Otherwise, the
words "NO PASSWORD" will be displayed.
2. Now you can edit the password. The actual edited item is displayed inversely. Press the but-
ton to move to the next position (clockwise direction), button serves to change the
values (0 ... 9).
3. After the operation is completed, confirm the edited data by pressing the button .

Display of status information to con-


BASIC SETTINGS
SERVICE
PASSWORD firm data:
DIAGNOSTIC
CLONE SETTINGS ENTER "YES" – correctly edited password
PASSWORD
00000
LANGUAGE
INFO "NO" – incorrectly edited password
"OK" – the password saved (only in
case of "CHANGE")
PASSWORD
The password is automatically hidden
CHANGE after it is edited or changed ("00000"
00000 will appear).
To deactivate the password, edit the
numerical combination "00000" in the
mode "CHANGE".

The level meter with activated password will be automatically locked after 5 minutes
of inactivity or after 5 min. from switching to measuring mode. Locking of level meter
is indicated in the lower left corner of the screen by the letter "L".

If the password is lost, contact the manufacturer.

© Dinel, s.r.o. ULM–70 23


LANGUAGE
Setting the language of display menu.
LANGUAGE
BASIC SETTINGS
SERVICE You can set three kinds of language:
DIAGNOSTIC
CLONE SETTINGS
PASSWORD
ENGLISH
ČESKY – ENGLISH – по русски
LANGUAGE
INFO

INFO
Information about the type, serial number and production date of the level meter (type, serial
number – SN and firmware version – SW).
BASIC SETTINGS SENSOR SENSOR
SERVICE DISPLAY MODULE
DIAGNOSTIC ULM-70N-06-I HART
CLONE SETTINGS SN: 140025
PASSWORD
LANGUAGE SW: 2.0
INFO

SENSOR DISPLAY MODULE


DISPLAY MODULE
DM-70
SN: 140041
SW: 4.1

11. Protocol HART®

Universal communication interface for data communication of peripheral devices with the
level meter. Data transmission runs through the same line as the 4 ÷ 20mA current loop
without impact on analog communication. For setting the level meter and collection of
measured data, it is necessary to have available a HART communicator, by which it is possible
to communicate directly with the level meter, or using it, to mediate communication with
a peripheral device, see image 22.

ULM–70_-_ _-_-I

≈ 250 R power supply unit

HART®
convertor
UHC-01

RS232 / USB

Obr. 22: Typical hardware configuration with HART

24 ULM–70 © Dinel, s.r.o.


HART® Specifications
The implemented HART® protocol is revision 5.
It contains universal commands 0, 1, 2, 3, 6, 11, 12, 13, 14, 15, 16, 17, 18, 19 and standard
(practical) commands 34, 35, 40, 42, 44, 49.

Meaning of variables
PV - distance to the surface
SV - value shown on the display
TV - temperature at the measurement point
QV - level height

12. Protocol Modbus

Data communication takes place along a series line of a standard RS-485 with protocol Modbus
RTU. A list of relevant variables is provided in a separate annex. To set up the level meter and
collect measured data, you can use the software application "Basic SCADA level", which is freely
available at the Website www.dinel.cz. Connecting the level meter to a peripheral device can be
performed using a converter URC-485, see image 23.

ULM–70_-_ _-_-M

power supply unit

convertor
URC-485 RS485 / USB

Obr. 23: Typical hardware configuration with Modbus

13 . Function and status indication

Function and status indicationare signalled by:


• display module (see chapter 9)
• setting the fault current to the value selected in MENU - SERVICE - FAULT MODE (applies to
the current version with HART® communication - I)
• status messages in HART® communication (valid for current version with HART® communication - I)
• Status registers: STATUS 1 and STATUS 2 in MODBUS communication (valid for version with
MODBUS communication - M)

© Dinel, s.r.o. ULM–70 25


14. Order code

PRODUCT
ULM-70
PERFORMANCE
N Normal – usable in non-explosive areas only
Xi Ex. proof – suitable for explosive areas

MAXIMUM REACH
02 0,15 ... 2 m
06 0,25 ... 6 m
10 0,4 ... 10 m
20 0,5 ... 20 m

PROCESS CONNECTION
G thread G1"
F flange

OUTPUT TYPE
I current output (HART®)
M RS-485 (Modbus RTU)

ELECTRICAL CONNECTION
B1 plastic cable gland M16,
B2 plastic cable gland M20
B3 plastic cable gland M20 for 2 cables
H1 plastic cable gland for protective hose

SET-UP ELEMENTS
D basic version with display
C basic version with LCD display
L without display, metal lid

ULM-70 N - 06 - G - I - B1 - D EXAMPLE OF CODING

26 ULM–70 © Dinel, s.r.o.


15. Accessories

Standard – incl. in the price of the level sensor Optional – for extra charge
• 1 pc of Seal (for ULM–70_–02, 06, 10) • Fixing nuts G1" and G1 ½" and G2 ¼
• free-to-download programme Basic Scada • Horn adapter ST–G1,
Level (for the Modbus version) STG1,5 and ST–G2,25
• for version Modbus convertor URC-485

16. Safety, protection, compatibility and explosion proof

The level meter ULM–70 is equipped with protection against reverse polarity and output cur-
rent overload.
Protection against dangerous contact is secured by low safety voltage that complies with
EN 33 2000-4-41.
Electromagnetic compatibility according to EN 55022/B, EN 61326/Z1 and EN 61000-4-2 to 6.
Explosion proof of ULM–70Xi type complies with the following standards: EN 60079-0 : 2007;
EN 60079-11 : 2007 ; EN 60079-26 : 2007 and examined by FTZÚ-AO 210 Ostrava - Radvanice
certificate No.: FTZÚ 09 ATEX 0277X.

Special conditions for safe use ULM–70Xi:


The device is designed for connection to the isolating repeater IRU-420. When the other ap-
proved supply unit is used, whose output parameters satisfy above mentioned output param-
eters, it is necessary to have a galvanic separation or, if supply unit without galvanic separation
is used (Zener barriers), it is necessary provide potential equalization between sensor and point
of barrier earthing.
For application in zone 0 the present explosive atmospheres - mixture of air with flammable
gases, vapour or mists must comply: 0,8 bar < p < 1,1 bar. The device must be installed in such
a way, to prevent mechanical damage of sensor face. It is necessary carried out earthing by
screw which is placed on head of level meter.

The device must be installed in such a way, to prevent mechanical damage


of sensor face.

© Dinel, s.r.o. ULM–70 27


17. Use, manipulation and maintenance

The level meter does not require any personnel for its operation. Follow-up displaying device is
used to inform the technological entity operating personnel on the measured substance level
height during the operation.
Maintenance of this equipment consists in verification of integrity of the level meter and of
the supply cable. Depending on the character of the substance measured, we recommend to
verify at least once per year the clarity of the ultrasound transducer emitting field and to clean
it, respectively. In case any visible defects are discovered, the manufacturer or reseller of this
equipment must be contacted immediately.

It is forbidden to perform any modifications or interventions into the ULM–70 level meter
without manufacturer's approval. Potential repairs must be carried out by the manufac-
turer or by a manufacturer authorized service organization only.

Installation, commissioning, operation and maintenance of the ULM–70 level meter has
to be carried out in accordance with this instruction manual; the provisions of regula-
tions in force regarding the installation of electrical equipment have to be adhered to.

Installation in areas with potentially explosive atmospheres must be carried out in ac-
cordance with standard EN 60079-14 (Electrical apparatus for explosive gas atmospheres
- Part 14: Electrical installations in hazardous areas other than mines) and respectively in
accordance with other standards that apply to a given area.

The device must be installed to prevent tensile overload rope electrode level meter.

18. Marking of labels

délka for
Labels 70type
mmof- ULM-70N-_
výška 20 _-_-I-_-_:
mm
Temperature range
ULM-70N-02-G-I-B-D according to type:
Dinel Ser. No.: ______

www.dinel.cz
Made in Czech Republic U = 18 ... 36 V
02 ta = -30 ... +70°C
I = 4 ... 20 mA
IP67
06 ta = -30 ... +70°C
t a = -30 ... +70 °C
10 ta = -30 ... +60°C
Example 20 ta = -30 ... +60°C
Štítek of label
pro for type ofULM-70N-02-G-I-B-D
varianty ULM-70N-02-G-B-D
Symbol of producer: logo Dinel ®

Internet address:ULM-70N-06-G-I-B-D
Dinel www.dinel.cz Ser. No.: ______

Level meter type:UULM–70N–_


www.dinel.cz
Made in Czech Republic = 18 ... 36 V
_–_–I–_–_
Serial number: Ser. No.:
I = 4 ... 20xxxxx
mA – (from the left: production year, serial prodution number)
IP67
Supply voltage: Uit=a =18 -30......36
+70V°C
=
Output current range: I = 4 ÷ 20 mA
Štítek
Ambient pro varianty
temperature range: ULM-70N-06-G-I-B-D
ta = -30 ... +_ _ °C (see. Temperature range according to type)
Protection class: IP67
Dinel
Compliance mark:
ULM-70N-10-G-I-B-D Ser. No.: ______

Electro-waste
Made in Czech Republic take-back
www.dinel.cz
U = 18 ... system
36 V mark:
I = 4 ... 20 mA
28 t a = -30 ... +60 °C
IP67 ULM–70 © Dinel, s.r.o.

Štítek pro varianty ULM-70N-10-G-I-B-D


Labels
délka for
70 type
mmof- výška
ULM-70Xi-_
20 mm_-_-I-_-_:
Clasification non-explo-
sive performance:
Dinel ULM-70Xi-02-G-I-B-D
I = 4 ... 20 mA
Ser. No.: ______
II 1/2 G Ex ia IIB T5 Ga/Gb 02 II 1/2G Ex ia IIB T5 Ga/Gb
www.dinel.cz U i = 30 V FTZÚ 09 ATEX 0277X
Made in Czech Republic P i = 0,99 W, I i = 132 mA 06 II 1/2G Ex ia IIB T5 Ga/Gb
C i = 370 nF, L i = 0,9 mH IP67
t a = -30 ... + 70 °C 10 II 1/2G Ex ia IIA T5 Ga/Gb
1026
20 II 2G Ex ia IIA T5 Gb
Example of label for type of ULM-70Xi-02-G-I-B-D
Štítek pro varianty ULM-70Xi-02-G-I-B-D
Symbol of producer: logo Dinel® Temperature range
Dinel
Internet
délka address:
70 mm
ULM-70Xi-06-G-I-B-D
Iwww.dinel.cz
- výška
= 4 ... 20 mA
20 mm
Ser. No.: ______
II 1/2 G Ex ia IIB T5 Ga/Gb according to type:
www.dinel.cz U = 30 V FTZÚ 09 ATEX 0277X
Level meter type: PULM–70Xi–_ = 0,99 W, I = 132_–_–I–_–_
i

02 ta = -30 ... +70°C


Made in Czech Republic mA
i i

C = 370 nF, L = 0,9 mH IP67


Serial number: Ser. t =No.:
-30 ... xxxxx
+ 70 °C –
i i

a
06 ta = -30 ... +70°C
Dinel ULM-70N-02-G-M-B-D Ser.
(from the left: production year, serial prodution number)
1026No.: ______

10 ta = -30 ... +60°C


Output
Štítek current
www.dinel.cz
pro range:
U = 18I ...
Republic varianty = 436...ULM-70Xi-06-G-I-B-D
V 20 mA
20 ta = -30 ... +60°C
Made in Czech

Max. internal values:RS-485 Ui = 30 V =, Ii =RTU)


(Modbus 132 mA; Pi = 0,99 IP67 W; Ci = 370 nF; Li = 0,9 mH
t a = -30 ... +70 °C
Ambient temperature
Dinel range: ta = -30 ... +_ _Ser.
ULM-70Xi-10-G-I-B-D °CNo.:(viz.______
Teplotní rozsah dle typu)
I = 4 ... 20 mA II 1/2 G Ex ia II A T5 Ga/Gb
Label of non-explosive U = 30 V device: , Performance: II _G Ex ia II_ T5 _/_
Štítek
Made in Czechpro
www.dinel.cz
Republic varianty
Number of certificate
P = 0,99 W, I = ULM-70N-02-G-M-B-D
i FTZÚ 09 ATEX 0277X

C = 370of nF, intrinsically safety: FTZÚ IP67 09 ATEX 0277X


i i 132 mA
i L = 0,9 mH
i

Protection class: IP67 at = -30 ... + 60 °C

Dinel
Compliance mark:
ULM-70N-06-G-M-B-D Ser. 1026No.: ______

, No. of authorized person examining control of system quality: 1026


Štítek
www.dinel.cz
Electro-waste protake-back
Made in Czech Republic
varianty
U = 18 ... 36 ULM-70Xi-10-G-I-B-D
system
V
mark:
RS-485 (Modbus RTU)
t = -30 ... +70 °C
IP67
ULM-70Xi-20-F-I-B-D
Dinel
Labels for typeI =of
a

4 ...ULM-70N-_
20 mA II_-_-M-_-_:
2 G Ex ia II A T5 Gb
Ser. No.: ______

Štítek pro varianty ULM-70N-06-G-M-B-D


www.dinel.cz
Made in Czech Republic
U i = 30 V
P i = 0,99 W, I i = 132 mA
FTZÚ 09 ATEX 0277X

C i = 370 nF, L i = 0,9 mH IP67 Temperature range


t a = -30 ... + 60 °C

Dinel ULM-70N-10-G-M-B-D Ser.


1026No.: ______
according to type:
www.dinel.cz 02 ta = -30 ... +70°C
Štítek pro varianty
U = 18
ULM-70Xi-20-F-I-B-D
36 V
Made in Czech Republic ...
RS-485 (Modbus RTU)
IP67 06 ta = -30 ... +70°C
t a = -30 ... +60 °C
10 ta = -30 ... +60°C
Štítek pro varianty ULM-70N-10-G-M-B-D 20 ta = -30 ... +60°C

Varianty značení:
Dinel ULM-70N-20-F-M-B-D
Example of label for type of ULM-70N-10-G-M-B-D
Ser. No.: ______

ULM-70Xi-__-_-I-_-_
www.dinel.cz
U = 18 36 V
Made in Czech Republic ...

Symbol of producer:RS-485logo DinelRTU)


(Modbus ® D
L IP67
t a = -30 ... +60 °C
Internet address: www.dinel.cz
B
Level meter type: ULM–70N–_ _–_–I–_–_ H
Štítek pro varianty ULM-70N-20-F-M-B-D
Serial number: Ser. No.: xxxxx – (from the left: production year, serial prodution number)
Supply voltage: U= 18 ... 36 G V=
F
Data output: RS-485 (Modbus
01
02
RTU)
Ambient temperature06range: ta = -30 ... +_ _ °C (see. Temperature range according to type)
Varianty
Protection značení:
class: IP6720
10

Compliance mark:
ULM-70N-__-_-M-_-_
Electro-waste take-back system mark:
D
L
Real label size is 70x20mm.
B
H

© Dinel, s.r.o. ULM–70 29


G
F
01
02
19. Technical specifications

Technical specifications – Level meter


ULM–70_–02 0.15 ... 2 m
ULM–70_–06 0.25 ... 6 m
Measuring range 1)
ULM–70_–10 0.4 ... 10 m
ULM–70_–20 0.5 ... 20 m
Adjustable measuring range (SPAN) Min. 200 mm
ULM–70N–_ _ 18 ... 36 V DC
Supply voltage
ULM–70Xi–_ _ 18 ... 30 V DC

ULM–70_–_ _-_-I 4 ... 20 mA (Limit values 3.9 ... 20.5 mA), HART®
Output
ULM–70_–_ _-_-M RS-485 with protocol Modbus RTU

ULM–70_–_ _-_-I 4 ... 20 mA / Max. 22 mA


Current consumption
ULM–70_–_ _-_-M Max. 20 mA
ULM–70_–02 ; 10 < 1 mm
Resolution ULM–70_–06 < 2 mm
ULM–70_–20 < 2.5 mm
Accuracy (within the total range) 0.15 %
Temperature error Max. 0.04% / K
ULM–70_–02 120 kHz
ULM–70_–06 75 kHz
Operating frequency
ULM–70_–10 50 kHz
ULM–70_–20 30 kHz
ULM–70_–02 ; 10 10°
Beamwidth (-3 dB) ULM–70_–06 14°
ULM–70_–20 12°
ULM–70_–02 ; 06 -30 ... +70 °C
Ambient temperature range
ULM–70_–10 ; 20 -30 ... +60 °C
Short-time temperature stress resistance +90 °C / 1 hour
Max. operation overpressure (on transmission surface) 0.1 MPa
Sensitivity 3 steps (low – medium – high)
Damping 0 ... 99 s
Measuring period 1 ... 4 s
Rise time ca. 30 s
Additional technical data for Ex proof 2) – Max. internal values Ui=30 V DC; Ii=132 mA; Pi=0.99 W; Ci=370 nF; Li=0.9 mH
Failure indication Adjustable in modes:
(echo loss, level in dead zone 3), internal failure) 3.75 mA ; 22 mA ; Last measured value
Maximal resistance of current output load (U = 24 V DC) Rmax = 270 Ω 4)

ULM–70_–02 Screwing with thread G 1''


ULM–70_–06 Screwing with thread G 1½''
Mechanical connection
ULM–70_–10 Screwing with thread G 2¼''
ULM–70_–20 Aluminium alloy flange
1)
In case the level of bulk-solid materials is measured, the measurement range is reduced.
2)
Allowed pressure range in the zone 0: 80 ... 110 kPa.
3)
Dead zone = Blind zone = Blocking distance
4)
Including 250Ω resistor in case of HART® connection.

30 ULM–70 © Dinel, s.r.o.


Technical specifications – Level meter
Protection class IP67
ULM–70_–02 0.3 kg
ULM–70_–06 0.4 kg
Weight
ULM–70_–10 0.6 kg
ULM–70_–20 3.1 kg

Technical specifications – Display module


Display type Matrix OLED, LCD1
Resolution 128 x 64 pixel
Character height / Number of digits measured value 9 mm / 5 Digits
OLED Yellow
Display colour
LCD black with white background light
Buttons Membrane switch panel

Ambient temperature OLED -30 ... +70 °C


range LCD -20 ... +70 °C
Weight 46 g
1) OLED- suitable for indoor and low-light applications. LCD – suitable for outdoor applications particularly with direct sunlight.

Used materials
Sensor part Variants Standard material
Lid All types aluminium alloy with powder coating
Glass All types polycarbonate
Body All types aluminium alloy with powder coating
Housing with thread All types plastic PP

Electroacoustic converter All types plastic PVDF

Display module ULM-70_...-D (with display) plastic POM


Cable gland All types plastic PA
Flange ULM-70-20-F aluminium alloy with powder coating

Factory default
ULM–70_–02 ULM–70_–06 ULM–70_–10 ULM–70_–20
MIN LEVEL 1)
2 000 6 000 10 000 20 000
MAX LEVEL 2) 150 250 400 500
UNITS mm; %; °C mm; %; °C mm; %; °C mm; %; °C
DAMPING 2 5 10 10
SENSITIVITY MEDIUM MEDIUM MEDIUM MEDIUM
MEDIUM TEMPERATURE NO NO NO NO
FAILURE MODE – NO ECHO 3.75 mA 3.75 mA 3.75 mA 3.75 mA
FAILURE MODE – DEAD ZONE 3) 22 mA 22 mA 22 mA 22 mA
POOLING ADDRESS (HART®) 00 00 00 00
PASSWORD No password No password No password No password
1)
Distance to min. level
2)
Distance to max. level
3)
Dead zone = Blind zone = Blocking distance
© Dinel, s.r.o. ULM–70 31
A rea classification (according to EN 60079-10 and EN 60079-14)
ULM–70N–_ _ Performance for non-explosive areas
Explosive proof – suitable for explosive areas (combustible gases or vapours)
ULM–70Xi–02
II 1/2G Ex ia IIB T5 Ga/Gb with JBIP1), the whole level meter
ULM–70Xi–06
– zone 1, front head part – zone 0
Explosive proof – suitable for explosive areas (combustible gases or vapours)
ULM–70Xi–10 II 1/2G Ex ia IIA T5 Ga/Gb with Isolating repeater (IRU–420), the whole level meter
– zone 1, front head part – zone 0
Explosive proof – suitable for explosive areas (combustible gases or vapours)
ULM–70Xi–20
II 2G Ex ia IIA T5 Gb with Isolating repeater (IRU–420), the whole level meter – zone 1
1) Intrinsically safe isolation converter (e.g. Dinel IRU-420).

20. Packaging, shipping and storage

The device DLM–35 is packaged in a polyethylene bag, and the entire consignment is placed into a
cardboard box. A suitable filler material is used in the cardboard box to prevent mechanical damage
during transport. Remove the device from the packaging only just before using, thereby protecting it
from potential damage.
A forwarding company will be used to ship goods to the customer. Upon prior agreement, ordered goods
can be picked up in person at company headquarters. When receiving, please check to see that the
consignment is complete and matches the order, or to see if any damage has occurred to the packaging
and device during transport. Do not use a device clearly damaged during transport, but rather contact
the manufacturer in order to resolve the situation.
If the device is to be further shipped, it must be wrapped in its original packaging and protected against
impact and weather conditions.
Store the device in its original packaging in dry areas covered from weather conditions, with humidity
of up to 85 % without effects of chemically active substances. The storage temperature range is -20°C
to +60°C.

Level gauges (sensors) of ULM-70_- 02, 06, 10, 20 type variants are equipped with protective
caps to prevent damage to the ultrasonic sensor. The cover must be removed during operation!
If possible, remove the cover after installation. If the technology does not allow it, remove the
cap just before installation. Store the cover after removal! If the level gauge (sensor) has to
be handled again, the cap must be placed back on. If handling without the protective cap is
detected, the level gauge/sensor will be regarded as mechanically damaged.

32 ULM–70 © Dinel, s.r.o.


19 . Menu structure
BASIC SETTINGS MIN AC TUAL MAX A C TUA L UNITS DAMPING
MIN LEVEL 0 2 8 7 0 mm 0 2 8 7 0 mm
SERVICE MAX LEVEL
DIAGNOSTIC UNITS OUTPUT: 0 4 . 0 0 mA OUTPUT: 2 0 . 0 0 mA LEVEL: mm
CLONE SETTINGS 05
DAMPING LEVEL: 0 6 0 0 0 mm LEVEL: 0 0 2 5 0 mm
PASSWORD SENSITIVITY DISPLAY: %
LANGUAGE TEACHING DISPLAY: 00000% DISPLAY: 00100%
INFO

BASIC SETTINGS MEDIUM TEMPERATURE MEDIUM TEMPERATURE FAILURE MODE HART SENSITIVITY
SERVICE FAILURE MODE

© Dinel, s.r.o. ULM–70


DIAGNOSTIC HART YES POLLING ADDRESS SENSITIVITY: MEDIUM
NO ECHO: 4.00 mA
CLONE SETTINGS FACTORY DEFAULT 00 1 2 3 4 5 6! 7! 8!
PASSWORD RESET
LANGUAGE +20°C MATERIAL: LIQUID
INFO

BASIC SETTINGS DISTANCE TO LEVEL DISTANCE TO LEVEL TEMPERATURE FACTORY DEFAULT TEACHING
SERVICE CURRENT
DIAGNOSTIC TANK MUST BE EMPTY
CLONE SETTINGS +23°C
01220 mm
PASSWORD
LANGUAGE ARE YOU SURE? START
INFO

BASIC SETTINGS CLONE SETTING CURRENT RESET


SERVICE
DIAGNOSTIC SENSOR DISPLAY
CLONE SETTINGS MODULE
PASSWORD 17.30 mA
DISPLAY SENSOR
LANGUAGE ARE YOU SURE?
MODULE
INFO

BASIC SETTINGS PASSWORD BASIC SETTINGS LANGUAGE


SERVICE SERVICE
DIAGNOSTIC DIAGNOSTIC
CLONE SETTINGS ENTER CLONE SETTINGS ENGLISH
PASSWORD PASSWORD
00000 LANGUAGE
LANGUAGE
INFO INFO

BASIC SETTINGS SENSOR SENSOR DISPLAY MODULE


SERVICE DISPLAY MODULE GRLM-70N-10-I HART DM-70
DIAGNOSTIC
CLONE SETTINGS ELECTRODE: 1000 mm SN: 140085
PASSWORD
SN: 140015 SW: 2.1 SW: 4.1
LANGUAGE
INFO

33
Notes
industrial electronics

Dinel, s. r. o.
U Tescomy 249
760 01 Zlín
Czech Republic

phone: +420 577 002 002


e-mail: [email protected]

www.dinel.cz