Vegabar54 4-20ma Hart 36726-En 120324
Vegabar54 4-20ma Hart 36726-En 120324
VEGABAR 54
4 … 20 mA/HART
Document ID:
36726
Process pressure
Contents
Contents
1 About this document
1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Symbolism used. . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3 Product description
3.1 Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 7
3.2 Principle of operation . . . . . . . . . . . . . . . . . . . . . . .. 8
3.3 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 9
3.4 Packaging, transport and storage . . . . . . . . . . . . . .. 9
3.5 Accessories and replacement parts . . . . . . . . . . . . .. 10
4 Mounting
4.1 General instructions . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.2 Mounting steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.3 Mounting steps, external housing. . . . . . . . . . . . . . . . 14
6.5 . . . . . . . . ..
6.6 Saving the parameter adjustment data . . . . . . . . . . .. 42
2 VEGABAR 54 • 4 … 20 mA/HART
Contents
9 Dismounting
9.1 Dismounting steps . . . . . . . . . . . . . . . . . . . . . . . . . . 51
9.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
10 Supplement
10.1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
10.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Supplementary documentation
Information:
Supplementary documents appropriate to the ordered version come
36726-EN-120324
with the delivery. You can find them listed in chapter "Product
description".
Editing status: 2012-03-09
VEGABAR 54 • 4 … 20 mA/HART 3
1 About this document
Ex applications
This symbol indicates special instructions for Ex applications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
This arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
36726-EN-120324
4 VEGABAR 54 • 4 … 20 mA/HART
2 For your safety
regulations.
VEGABAR 54 • 4 … 20 mA/HART 5
2 For your safety
2.6 CE conformity
This device fulfills the legal requirements of the applicable EC
guidelines. By attaching the CE mark, VEGA provides a confirmation
of successful testing. You can find the CE conformity declaration in the
download area of "www.vega.com".
6 VEGABAR 54 • 4 … 20 mA/HART
3 Product description
3 Product description
3.1 Structure
Scope of delivery The scope of delivery encompasses:
VEGABAR 54 • 4 … 20 mA/HART 7
3 Product description
Type label The type label contains the most important data for identification and
use of the instrument:
1
2
11
3
4
5
6
7
8 10
Scope of the operating This operating instructions manual applies to the following instrument
instructions manual versions:
l Software from 3.82
8 VEGABAR 54 • 4 … 20 mA/HART
3 Product description
Functional principle Sensor element is the CERTEC® measuring cell with robust,
dependent on the process fitting also front-flush, abrasion-resistant
ceramic diaphragm. The process pressure causes a capacitance
change in the measuring cell via the ceramic diaphragm. This change
is converted into an appropriate output signal and outputted as
measured value.
The CERTEC® measuring cell is also equipped with a temperature
sensor. The temperature value can be displayed via the indicating and
adjustment module as well as processed via the signal output (with
digital versions).
Seal concept As a standard feature, the CERTEC® measuring cell is equipped with
a lateral, recessed seal.
Instruments with double seal have an additional front seal.
Instruments with hygienic fitting are equipped with a gap-free form
seal.
3.3 Operation
The instrument can be adjusted with the following adjustment media:
l With indicating and adjustment module
l with the suitable VEGA DTM in conjunction with an adjustment
software according to the FDT/DTM standard, e.g. PACTware and
PC
l with manufacturer-specific adjustment programs AMS™ or PDM
l With a HART handheld
VEGABAR 54 • 4 … 20 mA/HART 9
3 Product description
Transport Transport must be carried out under consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
Transport inspection The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or
concealed defects must be appropriately dealt with.
Storage Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
l Not in the open
l Dry and dust free
l Not exposed to corrosive media
l Protected against solar radiation
l Avoiding mechanical shock and vibration
Storage and transport l Storage and transport temperature see chapter "Supplement -
temperature Technical data - Ambient conditions"
l Relative humidity 20 … 85 %
External indicating and VEGADIS 61 is suitable for external measured value indication and
adjustment of plics® sensors. It is connected to the sensor with an up
36726-EN-120324
adjustment unit
to 25 m long, four-wire, screened standard cable.
You can find further information in the operating instructions
"VEGADIS 61" (Document-ID 27720).
10 VEGABAR 54 • 4 … 20 mA/HART
3 Product description
External indicating and VEGADIS 62 is suitable for measured value indication and adjustment
adjustment unit VEGA- of sensors with HART protocol. It is looped into the 4 … 20 mA/HART
DIS 62 signal cable.
VEGADIS 62 is suitable for measured value indication with sensors
without HART protocol.
You can find further information in the operating instructions
"VEGADIS 62" (Document-ID 36469).
Measuring instrument The measuring instrument holder is used for wall/tube mounting of
holder VEGABAR series 50 pressure transmitters and VEGAWELL 52
suspension pressure transmitters. Supplied reducers enable the
adaptation to different instrument diameters. The material used is
316L.
Protective cover The protective cover protects the sensor housing against soiling and
intense heat from solar radiation.
You will find additional information in the supplementary instructions
manual "Protective cover" (Document-ID 34296).
Electronics module The electronics module is a replacement part for pressure transmitter
VEGABAR. One version is available for each type of signal output.
You find further information in the operating instructions "Electronics
module VEGABAR series 50 and 60 " (Document-ID 30175).
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 11
4 Mounting
4 Mounting
4.1 General instructions
Suitability for the pro- Make sure that all parts of the instrument exposed to the process, in
cess conditions particular the sensor element, process seal and process fitting, are
suitable for the existing process conditions. These include above all
the process pressure, process temperature as well as the chemical
properties of the medium.
You can find the specifications in chapter "Technical data" and on the
type label.
Mounting position Select an installation position you can easily reach for mounting and
connecting as well as later retrofitting of an indicating and adjustment
module. The housing can be rotated by 330° without the use of any
tools. You can also install the indicating and adjustment module in four
different positions (each displaced by 90°).
Ventilation and pressure The ventilation of the electronics housing as well as the atmospheric
compensation pressure compensation for the measuring cell are realised via a filter
element in the area of the cable gland.
36726-EN-120324
12 VEGABAR 54 • 4 … 20 mA/HART
4 Mounting
1 1 1
2 2 2
Caution:
Due to the filter effect, the pressure compensation is time delayed.
When opening/closing the housing cover quickly, the measured value
can change for a period of approx. 5 s by up to 15 mbar.
Information:
Make sure that the filter element is always free of buildup during
operation. A high-pressure cleaner may not be used for cleaning.
With instrument versions in protection IP 66/IP 68, 1 bar, the ventilation
is realised via the capillaries in the permanently connected cable. The
filter element is replaced by a blind stopper.
Temperature limits Higher process temperatures often mean also higher ambient
temperatures. Make sure that the upper temperature limits stated in
chapter "Technical data" for the environment of the electronics housing
and connection cable are not exceeded.
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 13
4 Mounting
Sealing/Screwing in Seal the thread with teflon, hemp or a similar resistant seal material on
threaded versions the process fitting thread 1½ NPT.
à Screw VEGABAR 54 into the welded socket. Tighten the hexagon
on the process fitting with a suitable wrench. Wrench size, see
chapter "Dimensions".
Warning:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.
Wall mounting 1 Mark the holes according to the following drilling template
2 Depending on the mounting surface, fasten the wall mounting
plate with 4 screws
14 VEGABAR 54 • 4 … 20 mA/HART
4 Mounting
90 mm (3.54")
70 mm (2.76") 8 mm
(0.32")
3mm
(0.12")
mm
3,5 4")
R .1
(0
110 mm (4.33")
93 mm (3.66")
Tip:
Mount the wall mounting plate so that the cable entry of the socket
housing points downward. The socket housing can be displaced by
180° to the wall mounting plate.
Warning:
The four screws of the socket housing must only be hand screwed. A
torque > 5 Nm (3.688 lbf ft) can damage the wall mounting plate.
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 15
5 Connecting to power supply
Tip:
We recommend using VEGA overvoltage arresters B63-48 and
ÜSB 62-36G.X.
Voltage supply Power supply and current signal are carried on the same two-wire
cable. The voltage supply range can differ depending on the
instrument version.
The data for power supply are specified in chapter "Technical data".
Provide a reliable separation between the supply circuit and the mains
circuits according to DIN VDE 0106 part 101. The VEGA power supply
units VEGATRENN 149A Ex, VEGASTAB 690 as well as all
VEGAMETs and VEGASCANs meet this requirement.
Keep in mind the following additional factors that influence the
operating voltage:
l Output voltage of the power supply unit can be lower under
nominal load (with a sensor current of 20.5 mA or 22 mA in case of
fault message)
l Influence of additional instruments in the circuit (see load values in
chapter "Technical data")
Connection cable The instrument is connected with standard two-wire cable without
screen. If electromagnetic interference is expected which is above the
test values of EN 61326 for industrial areas, screened cable should be
used.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm
(0.2 … 0.35 in) ensures the seal effect of the cable gland. If you are
using cable with a different diameter or cross-section, exchange the
seal or use a suitable cable gland.
We generally recommend the use of screened cable for HART
multidrop mode.
36726-EN-120324
Cable gland ½ NPT On the instrument with cable entry ½ NPT and plastic housing there is
a metallic ½" threaded insert moulded into the plastic housing.
16 VEGABAR 54 • 4 … 20 mA/HART
5 Connecting to power supply
Caution:
No grease should be used when screwing the NPT cable gland or
steel tube into the threaded insert. Standard grease can contain
additives that corrode the connection between threaded insert and
housing. This would influence the stability of the connection and the
tightness of the housing.
Cable screening and If screened cable is necessary, connect the cable screen on both ends
grounding to ground potential. In the sensor, the screen must be connected
directly to the internal ground terminal. The ground terminal on the
outside of the housing must be connected to the potential equalisation
(low impedance).
If potential equalisation currents are expected, the connection on the
processing side must be made via a ceramic capacitor (e. g. 1 nF,
1500 V). The low frequency potential equalisation currents are thus
suppressed, but the protective effect against high frequency interfer-
ence signals remains.
Warning:
Within galvanic plants as well as vessels with cathodic corrosion
protection there are considerable potential differences. Considerably
equalisation currents can be caused via the cable scrren when the
screen is earthed on both ends. To avoid this, the cable screen must
only connected to ground potential on one side of the switching
cabinet in such applications. The cable screen must not be connected
to the internal ground terminal in the sensor and the outer ground
terminal on the housing not to the potential equalisation!
Information:
The metallic parts of the instrument (transmitter, process fitting, etc.)
are conductively connected with the inner and outer ground terminal
on the housing. This connection exists either as a direct metallic
contact or via the shielding of the special connection cable on
instruments with external electronics. You can find specifications on
the potential connections within the instrument in chapter "Technical
data".
VEGABAR 54 • 4 … 20 mA/HART 17
5 Connecting to power supply
1 Loosen the four screws on the housing base with an Allen key size
4
18 VEGABAR 54 • 4 … 20 mA/HART
5 Connecting to power supply
3
1
3 Loop the connection cable through the cable entry on the housing
base1)
Information:
The cable gland can be mounted in three positions each displaced by
90°. Simply exchange the cable gland against the blind plug in the
suitable thread opening.
4 Connect the wire ends as described under "Single/Double
chamber housing" according to the numbering
5 Connect the screen to the internal ground terminal, connect the
outer ground terminal above on the housing to potential equal-
isation
6 Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
7 Attach the mounting plate again and tighten the screws
The electrical connection of the sensor to the external housing is
finished.
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 19
5 Connecting to power supply
Display
I²C 1
1 2 5 6 7 8
4 2
Wiring plan
Display
I2C
1 2 5 6 7 8
20 VEGABAR 54 • 4 … 20 mA/HART
5 Connecting to power supply
Electronics compart-
ment
Display
1
I2C
1 2 5 6 7 8
3 2
VEGABAR 54 • 4 … 20 mA/HART 21
5 Connecting to power supply
Connection compart-
ment
Display
2
1 2 I2C
Wiring plan
I2C
1 2
22 VEGABAR 54 • 4 … 20 mA/HART
5 Connecting to power supply
Fig. 15: VEGABAR 54 in IP 68 version 25 bar and axial cable outlet, external
housing
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 23
5 Connecting to power supply
Display
5
I²C
1 1 2 5 6 7 8
36726-EN-120324
24 VEGABAR 54 • 4 … 20 mA/HART
5 Connecting to power supply
Terminal compartment,
housing socket
1 2 3 4
2 4
1 3
5 6
Display
I2C
1 2 5 6 7 8
VEGABAR 54 • 4 … 20 mA/HART 25
5 Connecting to power supply
36726-EN-120324
26 VEGABAR 54 • 4 … 20 mA/HART
6 Set up with the indicating and adjustment module PLICSCOM
Note:
You can find detailed information on the adjustment in the operating
instructions manual "Indicating and adjustment module".
VEGABAR 54 • 4 … 20 mA/HART 27
6 Set up with the indicating and adjustment module PLICSCOM
Note:
If you intend to retrofit the instrument with an indicating and adjustment
module for continuous measured value indication, a higher cover with
an inspection glass is required.
36726-EN-120324
28 VEGABAR 54 • 4 … 20 mA/HART
6 Set up with the indicating and adjustment module PLICSCOM
1 2
1.1
Adjustment system The sensor is adjusted via the four keys of the indicating and
adjustment module. The LC display indicates the individual menu
items. The functions of the individual keys are shown in the above
illustration. Approx. 10 minutes after the last pressing of a key, an
automatic reset to measured value indication is triggered. Any values
not confirmed with [OK] will not be saved.
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 29
6 Set up with the indicating and adjustment module PLICSCOM
HART mode
Standard
Address 0
Level or process press- VEGABAR 54 can be used for level as well as for process pressure
ure measurement measurement. Default setting is level measurement. The mode can be
changed in the adjustment menu.
Depending on the application only the respective subchapter "Level or
process pressure measurement" is of importance. There, you find the
individual adjustment steps.
Level measurement
Parameter ad- Set up VEGABAR 54 in the following sequence:
justment "Level
measurement" 1 Selecting adjustment unit/density unit
2 Carry out a position correction
3 Carry out min. adjustment
4 Carry out max. adjustment
In the menu item "Adjustment unit" you select the physical unit in which
the adjustment should be carried out, e.g. mbar, bar, psi…
The position correction compensates the influence of the mounting
position or static pressure on the measurement. It does not influence
the adjustment values.
Information:
The steps 1, 3 and 4 are not necessary for instruments which are
already preset according to customer specifications!
You can find the data on the type label on the instrument or in the
menu items of the min./max. adjustment.
The indicating and adjustment module enables the adjustment without
filling or pressure. Thanks to this, you can carry out your settings
already in the workshop without the instrument having to be installed.
The actual measured value is also displayed in the menu items for
min./max. adjustment.
36726-EN-120324
Select unit In this menu item you select the adjustment unit as well as the unit for
the temperature indication in the display.
30 VEGABAR 54 • 4 … 20 mA/HART
6 Set up with the indicating and adjustment module PLICSCOM
To select the adjustment unit (in the example switching over from bar
to mbar), proceed as follows:2)
1 Push the [OK] button in the measured value display, the menu
overview is displayed.
▶ Basic adjustment
Display
Diagnostics
Service
Info
2 Confirm the menu "Basic adjustment" with [OK], the menu item
"Unit" will be displayed.
Unit
Unit of measurement
bar▼
Temperature unit
°C▼
Information:
When switching over to adjustment in a height unit (in the example
from bar to m), the density also has to be entered.
Proceed as follows:
1 Push the [OK] button in the measured value display, the menu
overview is displayed.
2 Confirm the menu "Basic adjustment" with [OK], the menu item
"Units of measurement" will be displayed.
3 Activate the selection with [OK] and select the requested unit with
[->] (in the example m).
4 Confirm with [OK], the submenu "Density unit" appears.
Unit of measurement
Density unit
▶ kg/dm³
pcf
5 Select the requested unit, e.g. kg/dm³ with [->] and confirm with
[OK], the submenu "Density" appears.
36726-EN-120324
2) Selection options: mbar, bar, psi, Pa, kPa, MPa, inHg, mmHg, inH2O,
mmH2O.
VEGABAR 54 • 4 … 20 mA/HART 31
6 Set up with the indicating and adjustment module PLICSCOM
Unit of measurement
Density
0001000
kg/dm³
6 Enter the requested density value with [->] and [+], confirm with
[OK] and move to position correction with [->].
The adjustment unit is thus switched over from bar to m.
Proceed as follows to select the temperature unit:3)
à Activate the selection with [OK] and select "Temperature unit
with [->].
à Activate the selection with [OK] and select the requested unit with
[->] (e.g. °F).
à Confirm with [OK].
The temperature unit is hence switched over from °C to °F.
+0000 mbar
53 mbar
32 VEGABAR 54 • 4 … 20 mA/HART
6 Set up with the indicating and adjustment module PLICSCOM
Information:
For an adjustment with filling, simply enter the actual measured value
indicated at the bottom of the display.
If the adjustment ranges are exceeded, the message "Outside
parameter limits" appears. The editing procedure can be aborted with
[ESC] or the displayed limit value can be accepted with [OK].
Information:
The displayed pressure for 100 % corresponds to the nominal
measuring range of the sensor (in the above example 1 bar =
1000 mbar).
2 Set the requested percentage value with [->] and [OK].
3 Confirm with [OK] and edit the requested mbar value.
4 Set the requested mbar value with [+] and [->].
5 Confirm with [OK] and move to the menu overview with [ESC].
The max. adjustment is finished.
Information:
For an adjustment with filling, simply enter the actual measured value
indicated at the bottom of the display.
If the adjustment ranges are exceeded, the message "Outside
parameter limits" appears. The editing procedure can be aborted with
[ESC] or the displayed limit value can be accepted with [OK].
VEGABAR 54 • 4 … 20 mA/HART 33
6 Set up with the indicating and adjustment module PLICSCOM
Information:
The steps 1, 3 and 4 are not necessary for instruments which are
already preset according to customer specifications!
You can find the data on the type label on the instrument or in the
menu items of the zero/span adjustment.
The indicating and adjustment module enables the adjustment without
filling or pressure. Thanks to this, you can carry out your settings
already in the workshop without the instrument having to be installed.
The actual measured value is displayed in addition to the menu items
for zero/span adjustment.
3 Select the menu item "Application" with [->] and edit with [OK].
Warning:
Note the warning: "Output can change".
Select unit In this menu item you select the adjustment unit as well as the unit for
the temperature indication in the display.
To select the adjustment unit (in the example switching over from bar
to mbar), proceed as follows:4)
1 Push the [OK] button in the measured value display, the menu
36726-EN-120324
overview is displayed.
4) Selection options: mbar, bar, psi, Pa, kPa, MPa, inHg, mmHg, inH2O,
mmH2O.
34 VEGABAR 54 • 4 … 20 mA/HART
6 Set up with the indicating and adjustment module PLICSCOM
▶ Basic adjustment
Display
Diagnostics
Service
Info
2 Confirm the menu "Basic adjustment" with [OK], the menu item
"Unit" will be displayed.
Unit
Unit of measurement
bar▼
Temperature unit
°C▼
+0000 mbar
53 mbar
VEGABAR 54 • 4 … 20 mA/HART 35
6 Set up with the indicating and adjustment module PLICSCOM
+0000.0 mbar
0000.0 mbar
Information:
The zero adjustment shifts the value of the span adjustment. The span,
i.e. the difference between these values, however, remains un-
changed.
Information:
For an adjustment with pressure, simply enter the actual measured
value indicated at the bottom of the display.
If the adjustment ranges are exceeded, the message "Outside
parameter limits" appears. The editing procedure can be aborted with
[ESC] or the displayed limit value can be accepted with [OK].
+1000.0 mbar
0000.0 mbar
Information:
The displayed pressure for 100 % corresponds to the nominal
measuring range of the sensor (in the above example 1 bar =
1000 mbar).
2 Set the requested mbar value with [->] and [OK].
3 Confirm with [OK] and move to the menu overview with [ESC].
The span adjustment is finished.
Information:
For an adjustment with pressure, simply enter the actual measured
value indicated at the bottom of the display.
36726-EN-120324
36 VEGABAR 54 • 4 … 20 mA/HART
6 Set up with the indicating and adjustment module PLICSCOM
Copy sensor data This function enables reading out parameter adjustment data as well
as writing parameter adjustment data into the sensor via the indicating
and adjustment module. A description of the function is available in the
operating instructions manual "Indicating and adjustment module".
The following data are read out or written with this function:
l Measured value presentation
l Adjustment
l Damping
l Linearisation curve
l Sensor-TAG
l Displayed value
l Display unit
l Scaling
l Current output
l Unit of measurement
l Language
The following safety-relevant data are not read out or written:
l SIL
l HART mode6)
l PIN
l Application
Copy sensor data
Linearisation curve A linearization is necessary for all vessels in which the vessel volume
does not increase linearly with the level - e.g. in a horizontal cylindrical
or spherical tank - and the indication or output of the volume is
required. Corresponding linearization curves are preprogrammed for
these vessels. They represent the correlation between the level
percentage and vessel volume. By activating the appropriate curve,
the volume percentage of the vessel is displayed correctly. If the
volume should not be displayed in percent but e.g. in l or kg, a scaling
can be also set in the menu item "Display".
Linearisation curve
Linear
Enter the requested parameters via the appropriate keys, save your
settings and jump to the next menu item with the [->] key.
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 37
6 Set up with the indicating and adjustment module PLICSCOM
The values of the following menu items are not reset with "Reset:
Factory setting
Like basic adjustment, in addition, special parameters are reset to
default values.8)
Peak value
The min. and max. temperature or pressure values are each reset to
the actual value.
36726-EN-120324
38 VEGABAR 54 • 4 … 20 mA/HART
6 Set up with the indicating and adjustment module PLICSCOM
Optional settings Additional adjustment and diagnosis options such as e.g. scaling,
simulation or trend curve presentation are shown in the following menu
schematic. You will find a detailed description of these menu items in
the operating instructions manual "Indicating and adjustment module".
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 39
6 Set up with the indicating and adjustment module PLICSCOM
Information:
Depending on the version and application, the highlighted menu
windows may not always be available.
Basic adjustment
▶ Basic adjustment 1
Display
Diagnostics
Service
Info
Unit 1.1 Position correction 1.2 Min. adjustment 1.3 Max. adjustment 1.4
Unit of measurement Offset 000.0 % 100.00 %
bar ▼ = P
= =
Temperature unit 0.2 mbar 0.0 mbar 100.00 mbar
°C ▼ 0000 mbar 0.0 mbar 0.0 mbar
1s Linear ▼ Sensor
Display
Basic adjustment 2
▶ Display
Diagnostics
Service
Info
Displayed value 2.1 Displayed value ▼ 2.1 Display unit 2.2 Scaling 2.3
0 % = 0.0
Pressure ▼ Scaled Volume ▼ l
l▼ 100 % = 100.0
l
Lighting 2.4
Switched off ▼
36726-EN-120324
40 VEGABAR 54 • 4 … 20 mA/HART
6 Set up with the indicating and adjustment module PLICSCOM
Diagnostics
Basic adjustment 3
Display
▶ Diagnostics
Service
Info
Service
Basic adjustment 4
Display
Diagnostics
▶ Service
Info
SIL 4.5 HART mode 4.6 Copy sensor data 4.7 PIN 4.8
Application 4.9
Level ▼
Info
Basic adjustment 5
Display
Diagnostics
Service
▶ Info
Instrument type 5.1 Calibration date 5.2 Last change using PC 5.3 Sensor characteristics 5.4
Serial number
VEGABAR 54 • 4 … 20 mA/HART 41
6 Set up with the indicating and adjustment module PLICSCOM
36726-EN-120324
42 VEGABAR 54 • 4 … 20 mA/HART
7 Set up with PACTware and other adjustment programs
VEGACONNECT exter-
nally
1 2
TWIST
3
LO
EN
CK
OP
USB
4
VEGABAR 54 • 4 … 20 mA/HART 43
7 Set up with PACTware and other adjustment programs
Necessary components:
l VEGABAR 54
l PC with PACTware and suitable VEGA DTM
l VEGACONNECT
l Power supply unit or processing system
VEGACONNECT via
HART
2 4
CK
LO
TWIST
USB
3
N
OPE
Fig. 23: Connecting the PC via HART to the signal cable
1 VEGABAR 54
2 HART resistance 250 Ω (optional depending on processing)
3 Connection cable with 2 mm pins and terminals
4 Processing system/PLC/Voltage supply
Necessary components:
l VEGABAR 54
l PC with PACTware and suitable VEGA DTM
l VEGACONNECT
l HART resistance approx. 250 Ω
l Power supply unit or processing system
Note:
With power supply units with integrated HART resistance (internal
resistance approx. 250 Ω), an additional external resistance is not
necessary. This applies, e. g. to the VEGA instruments VEGATRENN
149A, VEGADIS 371, VEGAMET 381. Common Ex separators are
also usually equipped with a sufficient current limitation resistance. In
such cases, VEGACONNECT 4 can be connected parallel to the
4 … 20 mA cable.
Note:
Keep in mind that for setup of VEGABAR 54, DTM-Collection in the
actual version must be used.
44 VEGABAR 54 • 4 … 20 mA/HART
7 Set up with PACTware and other adjustment programs
VEGABAR 54 • 4 … 20 mA/HART 45
8 Maintenance and fault rectification
Cleaning If necessary, clean the diaphragm. Make sure that the materials are
resistant to the cleaning process, see resistance list under "Services"
on "www.vega.com". The wide variety of applications of isolating
diaphragms makes special cleaning instructions necessary for each
application. Please ask the VEGA agency serving you.
Fault rectification The first measures to be taken are to check the output signals as well
as to evaluate the error messages via the indicating and adjustment
module. The procedure is described below. Further comprehensive
diagnostics can be carried out on a PC with the software PACTware
and the suitable DTM. In many cases, the causes can be determined
and the faults rectified this way.
24 hour service hotline Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since
we offer this service world-wide, the support is only available in the
English language. The service is free of charge, only the standard
telephone costs will be charged.
46 VEGABAR 54 • 4 … 20 mA/HART
8 Maintenance and fault rectification
? 4 … 20 mA signal missing
l Connection to voltage supply wrong
à Check connection according to chapter "Connection steps"
and if necessary, correct according to chapter "Wiring plan"
l No power supply
à Check cables for breaks; repair if necessary
l Operating voltage too low or load resistance too high
à Check, adapt if necessary
? E017
l Adjustment span too small
à repeat with modified values
? E036
l no operable sensor software
à Carry out a software update or send instrument for repair
? E041
l Hardware error
à Exchange the instrument or send it in for repair
Reaction after fault rec- Depending on the reason for the fault and the measures taken, the
36726-EN-120324
tification steps described in chapter "Set up" may have to be carried out again.
9) Fault message can also appear if the pressure is higher than the nominal
range.
VEGABAR 54 • 4 … 20 mA/HART 47
8 Maintenance and fault rectification
48 VEGABAR 54 • 4 … 20 mA/HART
8 Maintenance and fault rectification
l Sensor
l Voltage supply
l VEGACONNECT
l PC with PACTware
VEGABAR 54 • 4 … 20 mA/HART 49
8 Maintenance and fault rectification
Prepare update Connect the signal conditioning instrument to power supply and
provide the connection from the PC to the instrument via the interface
adapter. Start PACTware and go via the menu "Project" to the VEGA
project assistant. Select "USB" and "Set instruments online". Activate
the project assistant with "Start". The assistant establishes the
connection automatically and opens the parameter adjustment window
"Sensor # online parameter adjustment". Connect this parameter
adjustment window before you carry out further steps.
Load software into sen- Select with the right mouse key the sensor in the project and go to
sor "Additional function". Then click to "Software update". The window
"Sensor # software update" opens. PACTware checks now the sensor
data and displays the actual hardware and software version of the
sensor. This takes approximately 60 s.
Push the button "Update software" and select the previously extracted
hex file. Then the software update can be started. The additional files
are installed automatically. Depending on the sensor, this procedure
lasts up to 1 h. Then the message appears ""Software update
successfully executed".
You can download a return form (23 KB) from our homepage at www.
vega.com under: "Downloads - Forms and certificates - Repair form".
By doing this you help us carry out the repair quickly and without
having to call back for needed information.
50 VEGABAR 54 • 4 … 20 mA/HART
9 Dismounting
9 Dismounting
9.1 Dismounting steps
Warning:
Before dismounting, be aware of dangerous process conditions such
as e.g. pressure in the vessel, high temperatures, corrosive or toxic
products etc.
9.2 Disposal
The instrument consists of materials which can be recycled by
specialised recycling companies. We use recyclable materials and
have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and
the respective national laws. Pass the instrument directly on to a
specialised recycling company and do not use the municipal collecting
points. These may be used only for privately used products according
to the WEEE directive.
Correct disposal avoids negative effects on humans and the environ-
ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 51
10 Supplement
10 Supplement
10.1 Technical data
General data
Parameter, pressure Gauge pressure, absolute pressure, vacuum
Measuring principle Ceramic-capacitive, dry measuring cell
Communication interface I²C bus
process fitting
Max. torques
- Screws for PMC 1", PMC 1¼" 2 Nm
52 VEGABAR 54 • 4 … 20 mA/HART
10 Supplement
Output variable
Output signal 4 … 20 mA/HART
HART output values according to HART standard 5.0
- Primary Value Pressure
- Secondary Value Temperature10)
Signal resolution 1.6 µA
Failure signal current output (adjustable) mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA
Max. output current 22 mA
Load see load diagram under Power supply
Fulfilled NAMUR recommendations NE 43
100 %
90 %
1 2
10 %
tT tA t
tS
Fig. 24: Sudden change of the process variable. tT: dead time; tA: rise time; tS: jump response time
1 Process variable
2 Output signal
10) Depending on the respective instrument and the process fitting, the tem-
perature can deviate from the real process temperature.
VEGABAR 54 • 4 … 20 mA/HART 53
10 Supplement
Input variable
Adjustment
Adjustment range of the min./max. adjustment relating to the nominal measuring range:
- Percentage value -10 … 110 %
- Pressure value -20 … 120 %
Adjustment range of the zero/span adjustment relating to the nominal measuring range:
- zero -20 … +95 %
- Span -120 … +120 % 11)
- Difference between zero and span max. 120 % of the nominal range
Recommended max. turn down 10 : 1 (no limitation)
Nominal measuring ranges and overload capability in bar/kPa
The specifications are only an overview and refer to the measuring cell. Limitations due to the
material and process fitting version are possible. The specifications on the type label apply.
54 VEGABAR 54 • 4 … 20 mA/HART
10 Supplement
VEGABAR 54 • 4 … 20 mA/HART 55
10 Supplement
Deviation determined according to the limit point method according to IEC 6077012)
Applies to the digital signal output (HART, Profibus PA, Foundation Fieldbus) as well as to
analogue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
Deviation with version 0.1 %
- Turn down 1 : 1 up to 5 : 1 < 0.1 %
- Turn down > 5 : 1 < 0.02 % x TD
Deviation with version 0.2 %
- Turn down 1 : 1 up to 5 : 1 < 0.2 %
- Turn down > 5 : 1 < 0.04 % x TD
analogue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
56 VEGABAR 54 • 4 … 20 mA/HART
10 Supplement
In the compensated temperature range 0 … +100 °C (+32 … +212 °F), reference temperature
20 °C (68 °F).
Average temperature coefficient zero signal
- Turn down 1 : 1 < 0.05 %/10 K
- Turn down 1 : 1 up to 5 : 1 < 0.1 %/10 K
- Turn down up to 10 : 1 < 0.15 %/10 K
Outside the compensated temperature range:
Average temperature coefficient zero signal
- Turn down 1 : 1 typ. < 0.05 %/10 K
Thermal change of the current output
Applies also to the analogue 4 … 20 mA current output and refers to the set span.
Thermal change, current output < 0.05 %/10 K, max. < 0.15 %, each with
-40 … +80 °C (-40 … +176 °F)
0,15 %
-0,15 %
Long-term stability (according to DIN 16086, DINV 19259-1 and IEC 60770-1)
Applies to digital HART interface as well as to analogue current output 4 … 20 mA under
reference conditions. Specifications refer to the set span. Turn down (TD) is the relation nominal
measuring range/set span.
Long-term drift of the zero signal:
- For one year < 0.05 % x TD
- For five years < 0.1 % x TD
- For ten years < 0.2 % x TD
Ambient conditions
Ambient, storage and transport temperature
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 57
10 Supplement
Process conditions
The specifications of the pressure stage and product temperature are used as an overview. The
specifications on the type label are applicable.
Pressure stage, process fitting
- Thread 316L PN 60
- Thread 316L 1" for ball valve fitting PN 40
- PMC 316L PN 10
- Flange 316L PN 40
58 VEGABAR 54 • 4 … 20 mA/HART
10 Supplement
VEGABAR 54 • 4 … 20 mA/HART 59
10 Supplement
Voltage supply
Operating voltage
- Non-Ex instrument 12 … 36 V DC
- EEx-ia instrument 12 … 30 V DC
Operating voltage with illuminated indicating and adjustment module
- Non-Ex instrument 20 … 36 V DC
- EEx-ia instrument 20 … 30 V DC
Permissible residual ripple
- < 100 Hz Uss < 1 V
- 100 Hz … 10 kHz Uss < 10 mV
16) Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.
60 VEGABAR 54 • 4 … 20 mA/HART
10 Supplement
Ω
1000
750
3
500
2
1
250
4
12 14 16 18 20 22 24 26 28 30 32 34 36 V
17) Instruments with gauge pressure measuring ranges cannot detect the am-
bient pressure when submerged, e.g. in water. This can lead to falsification
of the measured value.
18) Only with instruments with absolute pressure ranges.
VEGABAR 54 • 4 … 20 mA/HART 61
10 Supplement
Approvals
Instruments with approvals can have different technical data depending on the version.
That's why the associated approval documents have to be noted with these instruments. They are
part of the delivery or can be downloaded under www.vega.com via "VEGA Tools" and "serial
number search" as well as via "Downloads" and "Approvals".
36726-EN-120324
62 VEGABAR 54 • 4 … 20 mA/HART
10 Supplement
10.2 Dimensions
The following dimensional drawings represent only an extract of the
possible versions. Detailed dimensional drawings can be downloaded
on www.vega.com under "Downloads" and "Drawings".
The two chamber housings are not available with instruments with
4 … 20 mA signal output
Plastic housing
~ 69 mm ~ 84 mm (3.31")
(2.72")
ø 79 mm ø 79 mm
(3.03") (3.31")
M16x1,5
112 mm (4.41")
112 mm (4.41")
M20x1,5/
½ NPT
M20x1,5/
1 ½ NPT 2
Fig. 27: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated indicating and adjustment module the
housing is 9 mm/0.35 in higher
1 Single chamber version
2 Double chamber version
Aluminium housing
M16x1,5
116 mm (4.57")
120 mm (4.72")
M20x1,5/ M20x1,5
½ NPT
M20x1,5/
1 ½ NPT 2
36726-EN-120324
Fig. 28: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated indicating and adjustment module the
housing is 9 mm/0.35 in higher
1 Single chamber version
2 Double chamber version
VEGABAR 54 • 4 … 20 mA/HART 63
10 Supplement
M16x1,5
116 mm (4.57")
120 mm (4.72")
M20x1,5 M20x1,5
M20x1,5/
1 ½ NPT 2
Fig. 29: Housing versions in protection IP 66/IP 68 (1 bar) - with integrated indicating and adjustment module the
housing is 9 mm/0.35 in higher
1 Single chamber version
2 Double chamber version
~ 59 mm ~ 69 mm ~ 87 mm (3.43")
(2.32") (2.72")
ø 80 mm ø 86 mm
ø 79 mm
(3.15") (3.39")
(3.11")
M16x1,5
117 mm (4.61")
120 mm (4.72")
112 mm (4.41")
M20x1,5/
M20x1,5/ ½ NPT
½ NPT M20x1,5/
1 2 ½ NPT 3
Fig. 30: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated indicating and adjustment module the
housing is 9 mm/0.35 in higher
1 Single chamber version, electropolished
2 Single chamber version, precision casting
2 Double chamber version, precision casting
36726-EN-120324
64 VEGABAR 54 • 4 … 20 mA/HART
10 Supplement
M16x1,5
117 mm (4.61")
120 mm (4.72")
112 mm (4.41")
M20x1,5/ M20x1,5
½ NPT
M20x1,5/
1 ½ NPT 2 3
Fig. 31: Housing versions in protection IP 66/IP 68 (1 bar) - with integrated indicating and adjustment module the
housing is 9 mm/0.35 in higher
1 Single chamber version, electropolished
2 Single chamber version, precision casting
2 Double chamber version, precision casting
65 mm
(2 9/16")
(2 43/64")
68 mm
42mm
(1 21/32")
1
59 mm (2 21/64")
(3 5/8")
92 mm
40mm
(1 37/64")
90 mm (3 35/64")
110 mm (4 21/64")
2
~ 66 mm (2 19/32")
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 65
10 Supplement
SW 36 mm
(1.42") SW 27 mm
(1.06")
45 mm
(1.77")
44 mm
(1.73")
G¾A G½A
22 mm
(0.87")
21 mm
(0.83")
ø 24 mm ø 18 mm
(0.95") (0.71")
ø 27 mm
(1.06")
GA GB
SW 41 mm SW 32 mm
(1.61") (1.26")
44 mm
(1.73")
59 mm
(2.32")
G1A G1A
21 mm
(0.83")
ø 30 mm
(1.18")
40 mm
(1.58")
ø 40 mm ø 40 mm
(1.58") (1.58")
GC GD
Fig. 33: VEGABAR 54, threaded connection: GA = G¾ A acc. to DIN 3852-E, GB = G½ A front-flush, GC = G1 A
absolutely front-flush, GD = G1 A (DIN 3852-E) x 40 mm absolutely front-flush
36726-EN-120324
66 VEGABAR 54 • 4 … 20 mA/HART
10 Supplement
44,5 mm
(1.75")
45 mm
(1.77")
ø 36 mm ø 50,5 mm
(1.42") (1.99")
CI CD
SW41
44,5 mm
(1.75")
67,5 mm
(2.66")
G1 A
ø 52 mm
(2.05")
GK RI
Fig. 34: VEGABAR 54, hygienic fitting: CI = Clamp DN15-20 according to DIN 32676, CD = Clamp 1" according to
DIN 32676, ISO 2852, GK = G1 A conus 40° front-flush, RI = Ingold connection
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 67
10 Supplement
45 mm
(1.77")
ø 62,5 mm
(2.46")
RE
Fig. 35: VEGABAR 54, hygienic fitting: RE = bolting DN 25 PN 40 acc. to DIN 11851
36726-EN-120324
68 VEGABAR 54 • 4 … 20 mA/HART
10 Supplement
d2
50 mm
(1.97")
b
f
d4
k
D
mm DN PN D b k d2 d4 f
1 DB 15 40 95 16 65 4 x ø14 45 2
FZ 20 40 105 18 75 4 x ø14 58 2
DA 25 40 115 18 85 4 x ø14 68 2
2 F1 1“ 150 lbs 108 14,2 79,2 4 x ø15,7 50,8 1,6
mm DN PN D b k d2 d4 f
1 DB 15 40 3.74" 0.63" 2.56" 4 x ø0,55" 1.77" 0.08"
FZ 20 40 4.13" 0.71" 2.95" 4 x ø0,55" 2.28" 0.08"
DA 25 40 4.53" 0.71" 3.35" 4 x ø0,55" 2.68" 0.08"
2 F1 1“ 150 lbs 4.25" 0.56" 3.12" 4 x ø0,62" 2" 0.06"
VEGABAR 54 • 4 … 20 mA/HART 69
10 Supplement
47,8 mm d2
(1 7/8")
b
f
d4
k
D
FA
mm DN PN D b k d2 d4 f
FA 25 40 115 18 85 4xø14 68 2
inch DN PN D b k d2 d4 f
FA 25 40 4 17/64" 45/64" 3 11/32" 4xø 35/64" 2 43/64" 5/64"
36726-EN-120324
70 VEGABAR 54 • 4 … 20 mA/HART
10 Supplement
SW32 ø 36 mm
46
45,5 mm
(1 51/64")
(2 3/64")
(1 57/64")
(1 27/64")
52 mm
48 mm
M44x1,25 M30x1,5 M8
ø 25 mm
21 mm
(1 9/64")
(53/64")
29 mm
ø 26,3 mm
(63/64")
(1 1/32")
ø 27,4 mm
(1 5/64")
BA/BB BF/BS PC
43,7 mm
43,7 mm
(1 23/32")
(1 23/32")
M8 M8
ø 34,6 mm ø 40,4 mm
(1 23/64") (1 19/32")
PD PE
Fig. 38: VEGABAR 54, fitting for the paper industry: BA/BB = M44 x 1.25; BF = M30 x 1.5; BS = M30 x 1.5 for
headbox, PC = PMC 1", PD = PMC 1¼", PE = PMC 1½"
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 71
10 Supplement
SW36 SW36
104,8 mm (4 1/8")
104,8 mm (4 1/8")
G1" G1"
(1 3/16")
(1 3/16")
30 mm
30 mm
ø 24,7 mm ø 24,7 mm
(31/32") (31/32")
ø 29,3 mm ø 29,3 mm
(1 5/32") (1 5/32")
VP VR
Fig. 39: VEGABAR 54, connection for paper industry: VP = G1" for PASVE, VR = G1" for PASVE with O-ring seal
36726-EN-120324
72 VEGABAR 54 • 4 … 20 mA/HART
10 Supplement
VEGA系列产品在全球享有知识产权保护。
进一步信息请参见网站<http://www.vega.com>。
10.4 Trademark
All the brands as well as trade and company names used are property
of their lawful proprietor/originator.
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 73
Index
INDEX
A Mounting, socket versions 14
Accessory
- External indicating and adjustment unit 10 P
- External indicating and adjustment unit Position correction 32, 35
VEGADIS 62 11 Pressure compensation 12
- Flanges 11 Process conditions 12
- Indicating and adjustment module 10
- Interface adapter 10
- Measuring instrument holder 11
R
Recycling 51
- Protective cover 11
Repair form 50
Application area 8
Replacement parts
- Electronics module 11
C Reset 38
Cable entry 16 Return form 50
Cable screening 17
Check output signal 46
Connection cable 16
S
Safety data sheet 50
Connection compartment
Seal concept 9
- Double chamber 22
Service hotline 46
Connection VEGACONNECT
Span adjustment 36
- directly on the sensor 43
- externally 43
- to the signal cable 44 T
Copy sensor data 37 Temperature limits 13
Total deviation 48
Type label 8
D
Disposal 51
U
Unit of measurement 30, 34
E
Electronics and connection compartment 20
Electronics compartment V
- Double chamber 21 Voltage supply 9, 16
Error messages 47
W
F WEEE directive 51
Fault rectification 46 Wiring plan
- Double chamber housing 22
- External electronics 25
H
- Single chamber housing 20
HART multidrop 30
Hotline 46
Z
Zero adjustment 36
M
Maintenance 46
Max. adjustment 33
36726-EN-120324
Min. adjustment 32
Moisture 12
Mounting external housing 14
Mounting position 12
74 VEGABAR 54 • 4 … 20 mA/HART
Index
36726-EN-120324
VEGABAR 54 • 4 … 20 mA/HART 75
Printing date:
VEGA Grieshaber KG
Am Hohenstein 113
77761 Schiltach
Germany
Phone +49 7836 50-0
Fax +49 7836 50-201
E-mail: [email protected]
www.vega.com
ISO 9001