Acetabular Positionning
Acetabular Positionning
Keywords: Acetabular cup position, angular position, anteversion, depth, height, inclination,
mediolateral, pelvic tilt, superoinferior
MeSH terms: Acetabulum, replacement, arthroplasty, hip, biomechanics
best outcomes are associated with reconstruction of the anteversion. To achieve a target radiographic anteversion,
anatomic COR. This results in a reduction in JRF and the operative angle should be higher.
abductor muscle force.15,16,19
During surgery in lateral position, operative inclination is
During THA, the acetabular reaming vector can cause assessed by looking from a point perpendicular to the plane
minor displacements of COR superiorly and posteriorly. of anteversion of the cup inserter. Radiographic inclination
It is believed that superior displacements of up to 13 mm can be assessed by standing directly behind the patient and
(and medial of 7.5 mm)24 have no clinical consequence. looking (from a point perpendicular to the coronal plane) at
However, more recent work suggests that when COR is the insertion angle with floor [Figure 2]. The radiographic
raised >3 mm (and medialized >5 mm) restoration of offset projection will be greater than this due to anteversion.
within 5 mm becomes difficult.26,27 The operative anteversion is assessed by looking down,
Angular position of cup to project the insertion angle onto sagittal plane, and
measuring against the longitudinal axis. This angle will be
Angular position includes anteversion and inclination
greater than the projected radiographic angle [Figure 2].
(abduction angle) of cup.
An illustration of these differences is the interpretation
A variety of values has been recommended to help reduce
wear, bearing‑related noise, impingement, and risk of of Lewinnek’s safe zone by Murray.31 Lewinnek et al.2
dislocation. Inclination past 45° has been shown to increase
wear rate,28,29 metal‑on‑metal bearings with inclinations past
55° have shown raised blood levels of metal ions,30 CoC
bearings outside a zone of 35°–45° inclination, and 15°–35°
anteversion are 29 times more likely to squeak.11 However,
the vast majority of studies relate to dislocation, with a
wide range of recommended values. The most commonly
quoted study is by Lewinnek et al.2 He found an increased
dislocation rate in cups placed outside anteversion angles
of 5°–25° and 30°–50° of inclination. To a certain extent,
the variation in recommended angles can be attributed to
different ways of measuring these angles. An understanding
of this is vital in achieving optimum cup placement.
Quantifying angular position
Angular position is the angle subtended by acetabular
axis, which passes through the center of socket
and is perpendicular to plane of the socket face
(Calandruccio, 1987). In practical terms, this is represented
by the inserter handle of an uncemented socket. The
angle of this axis can be measured in three different
ways, depending on the plane or the axis from which
it is measured. Inclination and anteversion can thus be
operative, radiologic, or anatomic [Table 1].31
Operative and radiographic angles are the most relevant Figure 2: Schematic diagram showing surgeon position to assess
surgically in lateral position of the patient. Due to radiographic and operative inclination. Position 1 assesses the projection
of the operative inclination on the coronal plane, and therefore, the
anteversion, the radiographic inclination is higher than radiographic angle. Position 2 perpendicular to the vertical plane passing
operative inclination and increases with increasing through inserter handle assesses operative inclination
reported 25° as the upper limit of anteversion beyond defined safe zones.41‑45 As previously discussed, there is
which 15% hips dislocated. This is a radiographic angle large variation in recommended values46‑53 partly due to the
and, on conversion, equivalent to 38° operative anteversion, use of different reference frames and angular definitions.
above which 21% hips dislocated. Therefore, Murray
Yoon et al.32 have attempted to resolve these
recommends that Lewinnek’s statement should have been
interpreted as ‑ hips should be inserted at <35° of operative inconsistencies by amalgamating data from nine different
anteversion to avoid dislocations. papers recommending safe zones and computed angles to
a unified system by transferring to a pelvic reference plane
Pelvic tilt and functional acetabular positions (based on a pelvic tilt of −8° standing and −4° lying34).
The angle between the coronal plane of the patient and the Using their common reference, the averaged target
anterior pelvic plane (APP) (plane passing through both orientation is 41° inclination and 16° anteversion for
anterior superior iliac spines [ASIS] and pubic tubercles) is radiographic angles. This equates to operative angles of
called pelvic tilt [Figure 3].32 It can be judged on radiographs 39° inclination and 21° anteversion. Based on the evidence
by looking at symphysis pubis to sacrococcygeal distance so far, this seems to be a good generalization that can be
with normal values of 32 mm (range 8–50 mm) in offered to minimize the incidence of dislocation.
women and 47 mm (range 15‑72 mm) in men.33 The
distance increases with forward tilt and decreases when Section II ‑ Planning and Execution in Cup
tilted backward. The average pelvis has a posterior tilt, Positioning
−12° and −8° in standing and supine, respectively,34 that Callanan et al.54 studied the accuracy of cup positioning and
tilts further back when sitting [Figure 3]. concluded that only 50% of hips were within the targeted
Dynamic changes to pelvic tilt affects radiographic safe zone for both anteversion and inclination (63%
anteversion,35 with 0.7°–0.8° increase for each degree of for inclination and 79% for version). Surgeon volume
posterior tilt.36,37 There is a mean increase in anteversion and high body mass index (BMI) were independent risk
of 7° (range −2° to 18°) in standing compared to supine, factors, with obesity, low volume surgeons, and minimally
with 52% of patients having more than 5° increase.36 There invasive surgery conferring a 1.3‑fold, 2‑fold, and 6‑fold
is a larger change moving from standing to sitting with an higher risk of malposition, respectively. Techniques that
average increase of 15.6°.38 can be used to improve the accuracy of positioning are
Forward or anteriorpelvic tilt causes a decrease in discussed below.
inclination and vice versa.37 Although the mean change Templating
is not significant (mean 2°),39 as the change in inclination
happens in a nonlinear manner, patients with larger posterior Templating is an essential step in the THA surgical process.
tilts have a larger change in inclination (mean change 0.29° Two‑dimensional templating is the most common method,
per degree of posterior pelvic tilt as opposed to 0.47° per and while, this does have limitations, it provides enough
degree when patient has 15° posterior tilt).40 information to carry out uncomplicated primary THAs
[Figure 4].
Safe zone considering differing angular measurements
and reference frames
There is controversy whether a generic safe zone exists
for cup position as some researchers have found that a
significant proportion of cups that dislocate lie within
Figure 4: X-ray pelvis with hip joints and proximal 1/3 rd of femur
Figure 3: Schematic representation on saw bones demonstrating pelvic anteroposterior view showing templating for an uncemented total hip
tilt - the difference between the anterior pelvic plane and the coronal plane arthroplasty
Predicting implant size from templating is imprecise, number were placed within safe zone (80.75% vs. 62.34%)
with correct size prediction ranging from 16%–62%,55,56 and had lower dislocation rate (1.03% vs. 2.49%) using
improving to 52%–98% if a range of sizes one above navigation. A meta‑analysis of cup position in navigated
and below is accepted. Some of this inaccuracy has been versus nonnavigated groups61 showed that a significantly
attributed to differences in magnification, poor placement lower number of cups were outliers with regard to the safe
of calibrating ball, and distortion due to projection.57 position in the navigated group (10.7% vs. 41.8%).
Work from our institution58 suggest that a major factor is
There is active, semi‑active, and passive navigation. Active
incorrect placement of cup template on radiographs and
navigation uses robots to implant cups. Semi‑active systems
that an understanding of relationship between teardrop and
allow the surgeons to move the robotic arms but do not
the most inferior extent of posterior acetabular rim (IPAR)
allow the arm to move beyond a milling boundary that has
may help to improve sizing.
been determined by preoperative three‑dimensional imaging.
The “teardrop” is a radiographic landmark created by The data from robotic THA is only starting to come through
superimposition of the most distal part of the medial wall and is not presented here. Passive navigation only guides
of the acetabulum and the tips of the anterior and posterior the surgeon to implant in the right position and consists of
horns. A line drawn between both teardrops can be used three types of navigation ‑ imageless navigation, computed
as a reference from which to measure inclination and leg tomography‑based navigation, and fluoroscopic navigation.
length. However, rather than teardrop, it is the posterior
Most passive navigation systems use APP as the reference
and inferior part of the semilunar weight bearing bone of
plane. Although this aids accurate placement of cup in
the acetabulum that is the most inferior and we call this
relation to the pelvis, it does not consider pelvic tilt.
IPAR [Figure 5]. The IPAR and teardrop were at different
Babisch et al.62 produced a nomogram for pelvic tilt and
levels in 86% cases, and the inferior extent of most cups
used it to adjust for pelvic tilt during navigation resulting
tends to sit between the radiographic projection of IPAR
in 98% accuracy of cup placement. Although current
and teardrop on postoperative X‑rays. Placing template at
navigation systems consider mean pelvic tilt, the future
the level of teardrop would tend to undersize and placing at
lies in adjusting cup position according to functional pelvic
IPAR would oversize the component [Figure 5].
position of each individual.
However, the aim of templating is not to predict implant
Despite these advantages, navigation is not commonly used
sizes (other than in extreme situations), as this can be
due to cost and operative time implications.
done easily during surgery, but to predict cup position in
terms of restoring COR and height, depth, and angular Patient positioning and its significance
position [Figure 4].
Supine positioning, with the use of anterior, anterolateral,
Computerized navigation – passing fad or lasting or direct lateral approaches, has the advantage that it is
benefits? easier to assess whether the pelvis is square and to assess
limb lengths. In addition, surgery takes place in the coronal
Navigation can lead to more reliable cup placement within
plane which is the functional plane.
safe zones.59 An evidence‑based analysis60 showed that
though there was no significant difference in mean angles However, the vast majority of THA uses lateral decubitus
between navigated and nonnavigated cups (though with position using posterior, anterolateral, anterior, or the
less variation in the navigated group), a significantly higher newer direct superior approach. Here, pelvic position tends
Figure 5: X-ray pelvis with both hip joints and proximal 1/3rd of femur anteroposterior view showing the difference in position between teardrop and inferior
extent of posterior acetabular rim. To size the acetabulum accurately in templating as well as to understand cup positioning, it’s important to recognize
the distance between the two
to have wide variation at setup. Pelvic tilt can range from lead to lateral uncoverage, and therefore, a medialized
25° posterior to 20° anterior (55% hips having posterior position should be chosen [Figures 6 and 7].
tilt, 38% anterior) with 16.1% patients having a tilt of
Ifan anatomic position is chosen, the transverse acetabular
10° or greater.39 Therefore, while we tend to think that the
ligament (TAL) and labrum can be used to define the cup
pelvis is aligned with the coronal plane of the patient when
not only in version but also its depth of placement.64,66
positioned on operating table, the truth is that we do not
Beverland et al.65 has described his technique of reaming
know its real orientation.
the acetabulum conservatively to no more than 4 mm
While maximum variation is for pelvic tilt, 8° bigger than femoral head size until the final reamer is
(2 standard deviation [SD] ± 32°), it is also adducted, cradled by labrum and TAL. The depth, version, and height
−4° (2SD ± 12°), and externally rotated, −8° (2SD ± 14°).63 (not inclination) of a hemispherical cup is guided by this.
It is recommended that while positioning, both ASIS and
If medialization is chosen, reamer/osteotomes are used
pubic tubercles are considered, and position of posterior
to expose the true floor. Care must be taken to restore
support in craniocaudal direction is used to control
global offset, as conventional reaming displaces COR
pelvic tilt.64 To control pelvic obliquity, a helpful method
medially by a mean of 5 mm and elevates it by 3.7 mm,
described by Beverland et al.65 is to draw parallel lines
compared to the anatomic position which displaces it by
at lumbosacral region using a spirit level with the patient
a mean of <1 mm in each direction.27 In planning, stem
seated preoperatively. In the lateral position, these lines
offset should be measured from intended final position of
then demonstrate the degree of pelvic adduction before
hip COR (based on final cup position) rather than COR
application of posterior support. This may not be as
of femoral head. The profile of the lesser trochanter (LT)
helpful in high BMI patients.
should also be noted, as an externally rotated hip, as
How to Achieve Target Depth evidenced by a prominent LT can cause a 20% reduction in
measurement of femoral offset19 on templating.
Individualized depth of cup placement is based on
achieving optimum cup fixation as the priority, with the To check global offset intraoperatively, there are several
main variable being adequate cup coverage. Where there techniques described. Jigs are available to measure from
is an adequate acetabulum an anatomic position, with its a fixed point above acetabulum before hip dislocation,
associated advantages, can be chosen [Figure 6]. Care measuring length, and offset. The senior author has
should be exercised in choosing a medialized position described his technique of using a suture from a Judd
in a person with large native offset, as medialization can pin inserted into the illium just superior to acetabulum to
result in medial movement of COR up to 14 mm27 making measure intraoperative leg length [Figure 8].67,68 The angle
it difficult to find a stem that would restore global offset. of this suture, with respect to the floor, can also be used to
Similarly, in a person with a deep acetabulum further determine the restoration of global offset. For this technique,
medialization can lead to impingement and loss of ROM. it is necessary to accurately place the limb in same position
On the other hand, in an acetabulum that is shallow or each time, as any change in angular position of the limb can
tending toward dysplasia, the anatomical position would affect measurement significantly. Another technique is direct
measurement of the distance from COR of the femoral head
to greater trochanter before neck resection and restoring it
along with the length of any medialization. The use of a
caliper to measure diameter of the resected head and using
this measurement to estimate the distance to the prosthetic
head from fixed points on greater trochanter and resected
Figure 6: X-ray pelvis with both hip joints and proximal half of femur,
anteroposterior view in this patient, an anatomical position has been chosen Figure 7: Preoperative and postoperative radiographs of pelvis with both hips
for the cup. Reducing the acetabular offset would mean that a femoral stem and proximal femur anteroposterior view in a patient with shallow acetabulum
with greater offset would have been required. Risk of impingement and where the medialized position has been chosen. Choosing the anatomic
reduced range of movement could result from medialization position in this patient would have resulted in lateral uncoverage of the cup
neck to accurately reproduce femoral offset and height when patients, care must be taken to ensure that inclination is
an anatomic COR is chosen has been described [Figure 8].69 not changed by abutment of the cup inserter against soft
tissues (angled/offset inserter is useful in such situations).
How to Achieve Target Height However, orthopedic surgeons are not very good at visually
The TAL can also be used to determine height.64 Ideally, estimating angles subtended by Jigs.66 A digital protractor
the reamer should be cradled by TAL and labrum66 which placed on the inserter handle can help improve the accuracy
ensures that the cup will not migrate proximally. The effect of insertion angles.70
of medialization must be considered as it can elevate COR Achieving the correct angle between inserter handle
by up to 18.9 mm.24 The inferior extent of the cotyloid and the floor or the operating table does not ensure
fossa is projected as teardrop on radiographs, and at the satisfactory cup inclination. Hill et al.71 reported a 12.7°
correct inclination, the inferior extent of the cup should variation between implanted inclination angle and final
not be above this. However, in dysplasia, sometimes position on radiographs despite careful implantation.
medialization and elevation of COR ID required to achieve They indicated that although this is in large part related
good lateral coverage. to the difference between operative and radiographic
inclination, it is also partly due to adduction of the
How to Achieve Target Angular Position
pelvis on operating table. They, therefore, recommend
Due to the disparity between operative and radiographic an operative inclination target of 35° to prevent outliers
angles, Grammatopoulos et al.63 recommended implanting at above what is seen as a critical inclination of 50° of
the cup in 5° less inclination and 8° more anteversion to radiographic inclination.
achieve the target radiographic position. The most common
technique used to guide angular position remains the use a We have discussed variability of pelvic position at setup.
jig during cup insertion. In smaller incisions and high BMI There is also significant intraoperative movement of the
pelvis during surgery. Factors influencing movement
include surgeon, pelvic supports (two ASIS supports better),
approach (posterior more movement than anterolateral),
and procedure type (resurfacing more than THA).62 Such
pelvic movement during surgery can affect cup position.
A solution is to use internal landmarks that are independent
of patient position. The relationship between the
superolateral most point of the bony acetabulum and the
lateral extent of the templated acetabular cup can be used
for this purpose. The distance between these points can
be measured during templating and can then be recreated
intraoperatively to guide inclination [Figure 9].
The use of TAL to guide anteversion has been reported to
reduce dislocation rates to 0.6%.64 Another intrapelvic guide
to cup version is the acetabular notch or psoas groove.
Especially in anatomical position of cup, care should be
Figure 8: Peroperative photograph showing a Judd pin is inserted at the exercised that there should be adequate anteversion so that
supraacetabular level, and the suture with a knot is used to mark the
distance to a fixed point on the greater trochanter before dislocation. This the anterior margin of cup is deep to the notch to prevent
suture can be used to assess leg length and offset intraoperatively psoas irritation.
Figure 9: X-ray (L) hip joint anteroposterior view and peroperative photograph showing measuring the lateral overhang of template and reproducing it
intraoperatively to achieve correct cup inclination
Where it is suspected that the femoral version is abnormal, Financial support and sponsorship
the version of the implanted stem should be considered,
Nil.
before the definitive cup is implanted. In these cases, the
use of combined anteversion angle is recommended.24,25 It Conflicts of interest
is beyond the scope of this article to discuss this technique There are no conflicts of interest.
in detail, however, it is recommended that the surgeon is
familiar with this concept where abnormal femoral anatomy References
is encountered. 1. Jenkins PJ, Clement ND, Hamilton DF, Gaston P, Patton JT,
Howie CR. Predicting the cost‑effectiveness of total hip and
Spinopelvic Interaction knee replacement: A health economic analysis. Bone Joint J
2013;95‑B: 115‑21.
It has long been recognized that severe hip arthritis and 2. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR.
its treatment with arthroplasty can change the alignment Dislocations after total hip‑replacement arthroplasties. J Bone
of the spine. Recently, there has been more focus on Joint Surg Am 1978;60:217‑20.
the relationship between functional cup position and 3. Little NJ, Busch CA, Gallagher JA, Rorabeck CH, Bourne RB.
spinopelvic kinematics. Due to changes in lumbar Acetabular polyethylene wear and acetabular inclination and
femoral offset. Clin Orthop Relat Res 2009;467:2895‑900.
lordosis and resultant pelvic tilt, a cup positioned within
4. Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH.
conventionally defined safe zones may become unstable Effect of femoral component offset on polyethylene wear in total
with activities such as sitting and standing. hip arthroplasty. Clin Orthop Relat Res 2001;388:125‑34.
In people with normal spine/pelvis mobility, there is a 5. Schmalzried TP, Shepherd EF, Dorey FJ, Jackson WO,
dela Rosa M, Fa’vae F, et al. The John Charnley Award. Wear is a
posterior tilt of 20°–35°, moving from standing to sitting, function of use, not time. Clin Orthop Relat Res 2000;381:36‑46.
with a mean of 4° less in stiffer spines, and 13° less than 6. Asayama I, Chamnongkich S, Simpson KJ, Kinsey TL,
those with hypermobile spines.72 Mahoney OM. Reconstructed hip joint position and abductor
muscle strength after total hip arthroplasty. J Arthroplasty
This has implications in patients undergoing THA with 2005;20:414‑20.
stiffer spines or spinal fusions. Such patients are not able 7. Kurtz WB, Ecker TM, Reichmann WM, Murphy SB. Factors
to tilt when sitting and therefore tend toward anterior affecting bony impingement in hip arthroplasty. J Arthroplasty
impingement in sitting, increasing the risk of posterior 2010;25:624‑34.e1‑2.
instability while patients with unbalanced spines tend to 8. Bonnin MP, Archbold PH, Basiglini L, Fessy MH,
extend their spine and hip to retain balance and tend to get Beverland DE. Do we medialise the hip centre of rotation in
total hip arthroplasty? Influence of acetabular offset and surgical
posterior impingement in standing. technique. Hip Int 2012;22:371‑8.
Phan et al.73 divide these patients into four groups and have 9. Bonnin MP, Archbold PH, Basiglini L, Selmi TA, Beverland DE.
suggested cup anteversion options: for flexible balanced Should the acetabular cup be medialised in total hip arthroplasty.
Hip Int 2011;21:428‑35.
spines, the cup is placed within a standard safe zone; a rigid
10. Sexton SA, Yeung E, Jackson MP, Rajaratnam S, Martell JM,
balanced spine (anterior impingement in sitting) should be Walter WL, et al. The role of patient factors and implant position
compensated by placing the cup in greater anteversion; a in squeaking of ceramic‑on‑ceramic total hip replacements.
flexible unbalanced spine (posterior impingement) should J Bone Joint Surg Br 2011;93:439‑42.
have spinal surgery first or a cup in less anteversion; finally, 11. Walter WL, O’toole GC, Walter WK, Ellis A, Zicat BA. Squeaking
the rigid unbalanced spine should have spinal surgery first in ceramic‑on‑ceramic hips: The importance of acetabular
component orientation. J Arthroplasty 2007;22:496‑503.
or a cup placed in a standard safe zone.
12. Cassidy KA, Noticewala MS, Macaulay W, Lee JH, Geller JA.
Effect of femoral offset on pain and function after total hip
Conclusion arthroplasty. J Arthroplasty 2012;27:1863‑9.
We have discussed the range of factors that play a role 13. Frueh WW, Hozack WJ. Management of limb length
in achieving planned cup position. The anatomy of discrepancy after total hip arthroplasty. Semin Arthroplasty
2005;16:127‑31.
the pelvis, stiffness of spinopelvic junction, functional
14. Parvizi J, Sharkey PF, Bissett GA, Rothman RH, Hozack WJ.
positions of pelvis, its position at setup and movement Surgical treatment of limb‑length discrepancy following total hip
during surgery, reference frames used, the way angles arthroplasty. J Bone Joint Surg Am 2003;85‑A: 2310‑7.
are measured and surgical technique, all have an impact 15. Abolghasemian M, Samiezadeh S, Jafari D, Bougherara H,
on cup position [Table 2]. An improved understanding Gross AE, Ghazavi MT. Displacement of the hip center of
of these factors may necessitate a move away from the rotation after arthroplasty of Crowe III and IV dysplasia:
A radiological and biomechanical study. J Arthroplasty
traditional technique of trying to place cup within the
2013;28:1031‑5.
same defined safe zone for every patient. The future may 16. Bicanic G, Delimar D, Delimar M, Pecina M. Influence of the
lie in individualized target positions, taking into account acetabular cup position on hip load during arthroplasty in hip
functional pelvic movement, to reduce dislocation risk, and dysplasia. Int Orthop 2009;33:397‑402.
improve outcomes. 17. Traina F, De Fine M, Biondi F, Tassinari E, Galvani A, Toni A.
The influence of the centre of rotation on implant survival using 38. Lazennec JY, Boyer P, Gorin M, Catonné Y, Rousseau MA.
a modular stem hip prosthesis. Int Orthop 2009;33:1513‑8. Acetabular anteversion with CT in supine, simulated standing,
18. Robinson M, Bornstein L, Mennear B, Bostrom M, Nestor B, and sitting positions in a THA patient population. Clin Orthop
Padgett D, et al. Effect of restoration of combined offset on Relat Res 2011;469:1103‑9.
stability of large head THA. Hip Int 2012;22:248‑53. 39. Zhu J, Wan Z, Dorr LD. Quantification of pelvic tilt in total hip
19. Sariali E, Klouche S, Mouttet A, Pascal‑Moussellard H. The arthroplasty. Clin Orthop Relat Res 2010;468:571‑5.
effect of femoral offset modification on gait after total hip 40. Maratt JD, Esposito CI, McLawhorn AS, Jerabek SA, Padgett DE,
arthroplasty. Acta Orthop 2014;85:123‑7. Mayman DJ. Pelvic tilt in patients undergoing total hip
20. Charnley J. Low Friction Arthroplasty of the Hip – Theory and arthroplasty: When does it matter? J Arthroplasty 2015;30:387‑91.
Practice. Berlin, Heidelberg: Springer‑Verlag; 1979. 41. Abdel MP, von Roth P, Jennings MT, Hanssen AD, Pagnano MW.
21. Nevelos J, Ingham E, Doyle C, Streicher R, Nevelos A, What safe zone? The vast majority of dislocated THAs are within
Walter W, et al. Microseparation of the centers of the Lewinnek safe zone for acetabular component position. Clin
alumina‑alumina artificial hip joints during simulator Orthop Relat Res 2016;474:386‑91.
testing produces clinically relevant wear rates and patterns. 42. Esposito CI, Gladnick BP, Lee YY, Lyman S, Wright TM,
J Arthroplasty 2000;15:793‑5. Mayman DJ, et al. Cup position alone does not predict risk of
22. Wroblewski BM, Siney PD, Fleming PA. Microseparation, fluid dislocation after hip arthroplasty. J Arthroplasty 2015;30:109‑13.
pressure and flow in failures of metal‑on‑metal hip resurfacing 43. Grammatopoulos G, Thomas GE, Pandit H, Beard DJ, Gill HS,
arthroplasties. Bone Joint Res 2012;1:25‑30. Murray DW. The effect of orientation of the acetabular component
23. Widmer KH. Containment versus impingement: Finding a on outcome following total hip arthroplasty with small diameter
compromise for cup placement in total hip arthroplasty. Int hard‑on‑soft bearings. Bone Joint J 2015;97‑B:164‑72.
Orthop 2007;31 Suppl 1:S29‑33. 44. Paterno SA, Lachiewicz PF, Kelley SS. The influence of
24. Malik A, Maheshwari A, Dorr LD. Impingement with total hip patient‑related factors and the position of the acetabular
replacement. J Bone Joint Surg Am 2007;89:1832‑42. component on the rate of dislocation after total hip replacement.
25. Ranawat CS, Maynard MJ. Modern techniques of cemented total J Bone Joint Surg Am 1997;79:1202‑10.
hip arthroplasty. Tech Orthop 1991;6:17‑25. 45. Pierchon F, Pasquier G, Cotten A, Fontaine C, Clarisse J,
26. Dastane M, Dorr LD, Tarwala R, Wan Z. Hip offset in total hip Duquennoy A. Causes of dislocation of total hip arthroplasty.
arthroplasty: Quantitative measurement with navigation. Clin CT study of component alignment. J Bone Joint Surg Br
Orthop Relat Res 2011;469:429‑36. 1994;76:45‑8.
27. Meermans G, Doorn JV, Kats JJ. Restoration of the centre 46. Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G,
of rotation in primary total hip arthroplasty: The influence Stöckl B. Reducing the risk of dislocation after total hip
of acetabular floor depth and reaming technique. Bone Joint J arthroplasty: The effect of orientation of the acetabular
2016;98‑B: 1597‑603. component. J Bone Joint Surg Br 2005;87:762‑9.
28. Hirakawa K, Mitsugi N, Koshino T, Saito T, Hirasawa Y, Kubo T. 47. Dorr LD, Wan Z. Causes of and treatment protocol for instability
Effect of acetabular cup position and orientation in cemented of total hip replacement. Clin Orthop Relat Res 1998;355:144‑51.
total hip arthroplasty. Clin Orthop Relat Res 2001;388:135‑42. 48. Jolles BM, Zangger P, Leyvraz PF. Factors predisposing to
29. Patil S, Bergula A, Chen PC, Colwell CW Jr., D’Lima DD. dislocation after primary total hip arthroplasty: A multivariate
Polyethylene wear and acetabular component orientation. J Bone analysis. J Arthroplasty 2002;17:282‑8.
Joint Surg Am 2003;85‑A Suppl 4:56‑63. 49. Masaoka T, Yamamoto K, Shishido T, Katori Y, Mizoue T,
30. De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, Shirasu H, et al. Study of hip joint dislocation after total hip
De Smet K. Correlation between inclination of the acetabular arthroplasty. Int Orthop 2006;30:26‑30.
component and metal ion levels in metal‑on‑metal hip resurfacing 50. McCollum DE, Gray WJ. Dislocation after total hip arthroplasty.
replacement. J Bone Joint Surg Br 2008;90:1291‑7. Causes and prevention. Clin Orthop Relat Res 1990;261:159‑70.
31. Murray DW. The definition and measurement of acetabular 51. Seki M, Yuasa N, Ohkuni K. Analysis of optimal range of socket
orientation. J Bone Joint Surg Br 1993;75:228‑32. orientations in total hip arthroplasty with use of computer‑aided
32. Yoon YS, Hodgson AJ, Tonetti J, Masri BA, Duncan CP. design simulation. J Orthop Res 1998;16:513‑7.
Resolving inconsistencies in defining the target orientation 52. Widmer KH, Zurfluh B. Compliant positioning of total hip
for the acetabular cup angles in total hip arthroplasty. Clin components for optimal range of motion. J Orthop Res
Biomech (Bristol, Avon) 2008;23:253‑9. 2004;22:815‑21.
33. Siebenrock KA, Kalbermatten DF, Ganz R. Effect of pelvic tilt 53. Yoshimine F. The influence of the oscillation angle and the
on acetabular retroversion: A study of pelves from cadavers. Clin neck anteversion of the prosthesis on the cup safe‑zone that
Orthop Relat Res 2003;407:241‑8. fulfills the criteria for range of motion in total hip replacements.
34. Lembeck B, Mueller O, Reize P, Wuelker N. Pelvic tilt makes The required oscillation angle for an acceptable cup safe‑zone.
acetabular cup navigation inaccurate. Acta Orthop 2005;76:517‑23. J Biomech 2005;38:125‑32.
35. Tannast M, Fritsch S, Zheng G, Siebenrock KA, Steppacher SD. 54. Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE,
Which radiographic hip parameters do not have to be Freiberg AA, et al. The John Charnley Award: Risk factors for
corrected for pelvic rotation and tilt? Clin Orthop Relat Res cup malpositioning: Quality improvement through a joint registry
2015;473:1255‑66. at a tertiary hospital. Clin Orthop Relat Res 2011;469:319‑29.
36. Polkowski GG, Nunley RM, Ruh EL, Williams BM, Barrack RL. 55. González Della Valle A, Comba F, Taveras N, Salvati EA.
Does standing affect acetabular component inclination and The utility and precision of analogue and digital preoperative
version after THA? Clin Orthop Relat Res 2012;470:2988‑94. planning for total hip arthroplasty. Int Orthop 2008;32:289‑94.
37. Mellano CR, Spitzer AI. How does pelvic rotation or tilt affect 56. Knight JL, Atwater RD. Preoperative planning for total hip
radiographic measurement of acetabular component inclination arthroplasty. Quantitating its utility and precision. J Arthroplasty
angle during THA? J Orthop 2015;12:222‑7. 1992;7:403‑9.
57. Scheerlinck T. Primary hip arthroplasty templating on Placement of the acetabular component. Bone Joint J 2016;98‑B
standard radiographs. A stepwise approach. Acta Orthop Belg 1 Suppl A: 37‑43.
2010;76:432‑42. 66. Beverland D. The transverse acetabular ligament: Optimizing
58. Edakalathur J, Bhaskar D, Thomas S, Board TN, Rajpura A. version. Orthopedics 2010;33:631.
IPAR (Inferior edge of the posterior acetabular rim) and Its 67. Desai A, Barkatali B, Dramis A, Board TN. A simple
Relevance in Positioning of the Cup. Oral Presentation At: intraoperative technique to avoid limb length discrepancy in total
European Hip Society, Munich; 07 September, 2016. hip arthroplasty. Surgeon 2010;8:119‑21.
59. Dorr LD, Malik A, Wan Z, Long WT, Harris M. Precision and 68. Desai AS, Dramis A, Board TN. Leg length discrepancy
bias of imageless computer navigation and surgeon estimates after total hip arthroplasty: A review of literature. Curr Rev
for acetabular component position. Clin Orthop Relat Res Musculoskelet Med 2013;6:336‑41.
2007;465:92‑9. 69. Hill JC, Archbold HA, Diamond OJ, Orr JF, Jaramaz B,
60. Moskal JT, Capps SG. Acetabular component positioning in Beverland DE. Using a calliper to restore the centre of the
total hip arthroplasty: An evidence‑based analysis. J Arthroplasty femoral head during total hip replacement. J Bone Joint Surg Br
2011;26:1432‑7. 2012;94:1468‑74.
61. Gandhi R, Marchie A, Farrokhyar F, Mahomed N. Computer 70. Meermans G, Goetheer‑Smits I, Lim RF, Van Doorn WJ, Kats J.
navigation in total hip replacement: A meta‑analysis. Int Orthop The difference between the radiographic and the operative
2009;33:593‑7. angle of inclination of the acetabular component in total hip
62. Babisch JW, Layher F, Amiot LP. The rationale for tilt‑adjusted arthroplasty: Use of a digital protractor and the circumference of
acetabular cup navigation. J Bone Joint Surg Am 2008;90:357‑65. the hip to improve orientation. Bone Joint J 2015;97‑B:603‑10.
63. Grammatopoulos G, Pandit HG, da Assunção R, Taylor A, 71. Hill JC, Gibson DP, Pagoti R, Beverland DE. Photographic
McLardy‑Smith P, De Smet KA, et al. Pelvic position measurement of the inclination of the acetabular component in
and movement during hip replacement. Bone Joint J total hip replacement using the posterior approach. J Bone Joint
2014;96‑B:876‑83. Surg Br 2010;92:1209‑14.
64. Archbold HA, Mockford B, Molloy D, McConway J, Ogonda L, 72. Kanawade V, Dorr LD, Wan Z. Predictability of acetabular
Beverland D. The transverse acetabular ligament: An aid to component angular change with postural shift from standing to
orientation of the acetabular component during primary total hip sitting position. J Bone Joint Surg Am 2014;96:978‑86.
replacement: A preliminary study of 1000 cases investigating 73. Phan D, Bederman SS, Schwarzkopf R. The influence of sagittal
postoperative stability. J Bone Joint Surg Br 2006;88:883‑6. spinal deformity on anteversion of the acetabular component in
65. Beverland DE, O’Neill CK, Rutherford M, Molloy D, Hill JC. total hip arthroplasty. Bone Joint J 2015;97‑B:1017‑23.