Lecture 13 Temperature Measurement
Lecture 13 Temperature Measurement
TEMPERATURE MEASUREMENT
12/24/2020 1
Outline
• Thermocouples
– overview, reference junction, proper connections, types, special
limits of error wire, time constants, sheathing, potential problems,
DAQ setup
• RTDs
– overview, bridges, calibration, accuracy, response time, potentail
problems
• Thermistors
• Infrared Thermometry
– fundamentals, emissivity determination, field of view
• Other
– Non-electronic measurement, thin-film heat flux gauge
• Temperature Controllers
• How to Choose
– Standards, cost, accuracy, stability, sensitivity, size, contact/non-
contact, temperature range, fluid type
12/24/2020 2
Thermocouples
• Seebeck effect
– If two wires of dissimilar metals are joined at both ends and one
end is heated, current will flow.
– If the circuit is broken, there will be an open circuit voltage across
the wires.
– Voltage is a function of temperature and metal types.
– For small DT’s, the relationship with temperature is linear
DV = DT
– For larger DT’s, non-linearities may occur.
12/24/2020 3
Measuring the Thermocouple Voltage
• If you attach the thermocouple directly to a voltmeter, you will
have problems.
12/24/2020 5
Other types of thermocouples
• Many thermocouples don’t have one copper wire. Shown
below is a “Type J” thermocouple.
12/24/2020 6
Isothermal Block
• The block is an electrical insulator but good heat
conductor. This way the voltages for J3 and J4 cancel out.
Thermocouple data acquisition set-ups include these
isothermal blocks.
V = (T1 − Tblock )
12/24/2020 7
Software Compensation
• How can we find the temperature of the block?
Use a thermister or RTD.
• Once the temperature is known, the voltage
associated with that temperature can be
subtracted off.
• Then why use thermocouples at all?
– Thermocouples are cheaper, smaller, more flexible and
rugged, and operate over a wider temperature range.
• Most data acquisition systems have software
compensation built in. To use Labview,you’ll need
to know if you have a thermister or RTD.
12/24/2020 8
Hardware Compensation
• With hardware compensation, the temperature of the
isothermal block again is measured, and then a battery
is used to cancel out the voltage of the reference
junction.
• This is also called an “electronic ice point reference”.
With this reference, you can use a normal voltmeter
instead of a thermocouple reader. You need a separate
ice-point reference for every type of thermocouple.
12/24/2020 9
Making Thermocouple Beads
12/24/2020 10
Time Constant vs. Wire Diameter
12/24/2020 11
Time Constant vs. Wire Diameter, cont.
12/24/2020 12
Thermocouple Types
12/24/2020 14
Sheathing and SLE
12/24/2020 15
Sheathing, cont.
12/24/2020 17
Potential Problems, cont.
• Shunt impedence
– As temperature goes up, the resistance of many
insulation types goes down. At high enough
temperatures, this creates a “virtual junction”. This is
especially problematic for small diameter wires.
• Galvanic Action
– The dyes in some insulations form an electrolyte in the
water. This creates a galvanic action with a resulting
emf potentially many times that of the thermocouple.
Use an appropriate shield for a wet environment. “T
Type” thermocouples have less of a problem with this.
12/24/2020 18
Potential Problems, cont.
• Thermal shunting
– It takes energy to heat the thermocouple, which results in a small decrease
in the surroundings’ temperature. For tiny spaces, this may be a problem.
– Use small wire (with a small thermal mass) to help alleviate this problem.
Small-diameter wire is more susceptible to decalibration and shunt
impedence problems. Extension wire helps alleviate this problem. Have
short leads on the thermocouple, and connect them to the same type of
extension wire which is larger. Extension wire has a smaller temperature
range than normal wire.
• Noise
– Several types of circuit set-ups help reduce line-related noise. You can set
your data acquisition system up with a filter, too.
– Small-diameter wires have more of a problem with noise.
12/24/2020 19
Potential Problems
12/24/2020 20
Data Acquisition Systems for Thermocouples
12/24/2020 21
Things to Note During System Assembly
12/24/2020 22
RTDs (Resistance Temperature Detectors)
12/24/2020 23
RTD geometry
12/24/2020 25
Resistance/Temperature Conversion
12/24/2020 26
Accuracy and Response Time
12/24/2020 28
Thermistors
12/24/2020 29
Thermistor Non-Linearity
12/24/2020 30
Resistance/Temperature Conversion
12/24/2020 31
Infrared Thermometry
12/24/2020 32
Benefits of Infrared Thermometry
12/24/2020 34
Emissivity
12/24/2020 35
Ways to Determine Emissivity
12/24/2020 36
Spectral Effects
12/24/2020 37
Price and Accuracy
12/24/2020 38
Non-Electronic Temperature Gages
12/24/2020 39
Non-Electronic Temperature Gages, cont.
12/24/2020 40
Thin-Film Heat Flux Gauge
12/24/2020 42
Thin-Film Heat Flux Gauge, cont.
• Difficulties with these gauges
– The distance between the two sides is very
small, so the temperature difference is small.
The uncertainty in the temperature difference
measurement can be large.
– Watch where you place them. If the effective
conductivity of the gauges is different than the
conductivity of the material surrounding it, it will
be either easier or harder for heat to pass
through it. Heat will take the path of least
resistance, so if you don’t position the gauge
carefully, you may not be measuring the actual
12/24/2020
heat flux. 43
Temperature Controllers
12/24/2020 44
Proportional controllers
• Proportional controllers
– Power can be varied. For example, in a heating
unit the average power supplied will decrease
the closer one gets to the set point.
– Power is often varied by turning the controller
on and off very quickly rather than using a VFD
– Some proportional controllers use proportional
analog outputs where the output level is varied
rather than turning the controller on and off.
12/24/2020 45
PID
• Combines proportional with integral and derivative control.
• With proportional control, the temperature usually stabilizes a
certain amount above or below the setpoint. This difference is
called offset.
• North America
– NEMA (National Electrical Manufacturers
Association), UL (Underwriters Laboratories),
12/24/2020
CSA (Canadian Standards Association
48
Enclosure Ratings
12/24/2020 49
Choice Between RTDs, Thermocouples,
Thermisters
• Cost – thermocouples are cheapest by far, followed by RTDs
• Accuracy – RTDs or thermisters
• Sensitivity – thermisters
• Speed - thermisters
• Stability at high temperatures – not thermisters
• Size – thermocouples and thermisters can be made quite small
• Temperature range – thermocouples have the highest range, followed
by RTDs
• Ruggedness – thermocouples are best if your system will be taking a
lot of abuse
12/24/2020 50
Simplified Uncertainty Analysis for Lab 1
12/24/2020 51
Random Error
• 95% confidence interval – 95% of temperature readings
will fall in this range
– =+/- 2 standard deviations
– For your lab, during calibration, take at least 35 data points
(N=35) at one temperature. Then calculate the average and
standard deviation using the equations below.
– Excel can also be used.
1 N
T = Ti
N i =1
1
1 N 2
(Ti − T )
2
ST =
N − 1 i =1
12/24/2020 52
Bias Error
• Conduction and radiation errors should be
negligible.
• For our lab, we will do a simplified analysis.
• Once you have a calibration curve fit, find the
deviation between the curve fit and each data
point. Use the magnitude of the maximum
deviation as your bias error.
• In ME 120 you’ll learn a lot more about calculating
uncertainties!
12/24/2020 53
Thank you for your attention!
12/24/2020 54