CL-900i (EJK024) Series Maintenance Guidance - V1.0 - EN
CL-900i (EJK024) Series Maintenance Guidance - V1.0 - EN
Maintenance Guidance
IVD Global Technical Support Dept
Table of Contents
1 Active Maintenance Objects and Cycle...................................................................................... 3
1.1 Active Maintenance Objects............................................................................3
1.2 Active Maintenance Cycle............................................................................... 3
2 Active Maintenance Materials and Tools List............................................................................. 3
3 Maintenance Items List..................................................................................................................5
4 Active Maintenance Items Guide................................................................................................. 7
4.1 Checking Consumables Status.......................................................................7
4.2 Checking System Status..................................................................................7
4.3 Checking the Probe.......................................................................................... 8
4.4 Checking the Syringe....................................................................................... 9
4.5 Checking Overflowing of the Dispersion Carousel.................................... 10
4.6 Cleaning the Cap of the Wash Buffer Tank................................................. 11
4.7 Cleaning the Dust Screen..............................................................................12
4.8 Cleaning the Radiating Fan...........................................................................13
4.9 Cleaning Vortexer Holes................................................................................ 14
4.10 Cleaning the Incubation, Maintenance, and Waste Drainage Holes...... 15
4.11 Cleaning the Gripper...................................................................................... 16
4.12 Replacing the Spring of the Gripper............................................................ 17
4.13 Cleaning Waste Drainage Probes................................................................ 18
4.14 Probe Special Wash....................................................................................... 19
4.15 Prime and Drain.............................................................................................. 19
4.16 Cleaning the Probe Swab..............................................................................21
4.17 Replacing the Dispersion Aspirate Peristaltic Pump Tubes..................... 22
4.18 Cleaning the dispersion aspirating probes..................................................26
4.19 Cleaning the Dispersion Dispense Probes................................................. 28
4.20 Checking the Substrate Leakage Box......................................................... 30
4.21 Checking the Isolation Chamber.................................................................. 31
5 System Performance Test...........................................................................................................32
5.1 Daily Clean.......................................................................................................32
5.2 DCF Diagnosis................................................................................................ 32
5.3 Substrate Background Test........................................................................... 33
5.4 System Repeatability......................................................................................33
6 Maintenance Software Upgrade................................................................................................ 34
7 Maintenance FAQs.......................................................................................................................34
8 Maintenance Report and Records.............................................................................................35
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
Maintenance Guide
1 Active Maintenance Objects and Cycle
Table 3-2 Tools and materials that need to be prepared by the customer.
SN Supplies and Tools Applicable Maintenance Source
1 Clean gauze Checking the syringe, wiping the Provided by the
rotor, sample probe, exterior of customer
dispersion aspirate / dispense probe
and reagent compartment
2 Dust-free cotton Cleaning the wash well and vortexer Provided by the
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
swabs customer
3 Suction cleaner Cleaning fans and the dust screen Provided by the
customer
4 Hair brush Cleaning fans and the dust screen Provided by the
customer
5 Beaker Washing probe and swab Provided by the
customer
6 Alcohol Washing sample probe and Provided by the
dispersion probe customer
7 Fiber-free gloves Wiping sample probe and the Provided by the
exterior of dispersion probe customer
8 Keyboard wash Cleaning the keyboard Provided by the
solution customer
9 Dispersion Replacing and cleaning the Accessory kit
unclogging device dispersion aspirate probe
10 Dispersion aspirate Replacing and cleaning the Accessory kit
probe assembly dispersion aspirate probe
Note: Please contact the person in charge of the instrument department before the
maintenance and inform the person of the maintenance content and the required time.
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
When performing annual maintenance, the engineer should perform all maintenance items
listed in the following table for about 4.5 hours.
SN Type Performe Items Maintenance System Maintenance
r Cycle Status Time
1 Checking Engineer Checking the power line Annual Standby 2min
and ensuring that the
ground voltage should
be ≤ 5V
2 Checking Engineer Checking instrument Annual Standby 1min
status
3 Checking Engineer Checking sample probe Annual Standby 3min
4 Cleaning Engineer Cleaning the Cap of the Annual Standby 3min
Wash Buffer Tank
5 Cleaning Engineer Cleaning the analyzer Annual Standby 8min
panel
6 Cleaning Engineer Cleaning the dust Annual Standby 10min
screen
7 Checking Engineer Checking overflowing of Annual Maintenance 10min
the dispersion carousel
8 Cleaning Engineer Cleaning the radiating Annual Maintenance 15min
fan
9 Cleaning Engineer Cleaning the vortexer Annual Maintenance 5min
holes
10 Cleaning Engineer Cleaning the incubation, Annual Maintenance 8min
maintenance,
photometry, and waste
drainage holes
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
11 Cleaning Engineer Cleaning Waste Annual Shutdown 5min
Drainage Probes
12 Executing Engineer Probe special wash Annual Maintenance 2min
13 Checking Engineer Checking the syringe Annual Maintenance 15min
14 Checking Engineer Checking the isolation Annual Maintenance 10min
chamber
15 Checking Engineer Checking the substrate Annual Maintenance 15min
leakage box
16 Executing Engineer Priming and Cleaning Annual Maintenance 12min
17 Cleaning Engineer Cleaning the sampling Annual Shutdown 5min
probe swab
18 Cleaning Engineer Replacing the Annual Shutdown 20min
Dispersion Aspirate
Peristaltic Pump Tubes
19 Cleaning Engineer Cleaning the dispersion Annual Maintenance 20min
aspirate probes
20 Cleaning Engineer Cleaning the dispersion Annual Shutdown 8min
dispense probes
21 Replacing Engineer Replacing the gripper Annual Shutdown 15min
spring
22 Executing Engineer Executing DCF Annual Standby 2min
diagnostics
23 Executing Engineer Executing routine Annual Standby 20min
cleaning
24 Executing Engineer Executing substrate Annual Standby 24min
background detection
25 Executing Engineer Executing System Annual Standby 25min
Repeatability
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
Select Utility - Status - Fluidics
Check if negative pressure of the vacuum pump is ≤ -30Kpa and
other negative pressure is abnormal
Check Feel whether the reagent carousel can refrigerate. Check the
current of the two radiators range from 2.0 A to 6.5 A.
Maintenance Takes 1min
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
Methods of cleaning the dispersion carousel:
Wash the dispersion carousel assembly with running water, dry it
with dust-free gauze and then air-dry it.
Use dust-free gauze moistened with alcohol to clean the interior of
the dispersion chamber before drying it with dry dust-free gauze.
Use dust-free gauze moistened with alcohol to clean the gripping
holes on the lower surface of the dispersion carousel cover and
wipe off the wash buffer crystal and substrate coagulation. Then dry
the holes with a dry dust-free gauze.
5. If overflow occurs, clean the substrate tubes:
Use gauze moistened with alcohol to gently wipe the exterior of the
substrate tube, and then use gauze moistened with deionized water
to wipe out the alcohol. Note: Do not damage the substrate tube
during cleaning.
6. After overflow occurs, clean the gripper according to the gripper
finger cleaning guide.
7. After overflow occurs, clean the incubation / waste drainage /
photometer hole according to the actual conditions.
8. Install the dispersion assembly according to the reverse order of
the disassembling steps.
Check After finishing the step 4, make sure that the dispersion carousel
cover, the dispersion chamber and the dispersion carousel are free
of wash buffer crystals or solidified substrate and show metallic
luster. and show metallic luster.
Maintenance takes 10min (No overflow occurs.)
60min (If overflow occurs.)
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
2. Use gauze moistened with deionized water to clean the tank cap.
3. Finally, wipe it with a clean napkin and reload it for use.
Check Confirm whether the wash buffer tank cap is clean.
Maintenance takes 3min
Precautions 1. Do not reinstall the dust screens until they are dry completely.
2. Install the dust screens correctly to avoid gaps.
System status Shut Down
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
3. Open the backplate of the analyzer and remove the other two
dust screens.
4. Use the suction cleaner, hair brush or fresh water to clean the
dust screens, and then dry them in air.
5. Reinstall the dust screens when they are dry.
Check Visual inspection. Observe the dust situation on the dust screens
under the light, and confirm there is no obvious blockage.
Maintenance takes About 10min
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
Materials required Suction cleaner, hair brush and gauze
Precautions 1. Pay attention to the direction of the fan during installation.
2. Make sure that the fan can rotate smoothly.
System status Shutdown
How to do 1. Confirm the instrument is powered off.
2. Use a hexagon wrench to loosen the four hexagon screws on
each fan, disconnect the cables, and remove the fan.
3. Use the suction cleaner, hairbrush or gauze to clean the fan
assembly.
4. Reinstall the cleaned fan assembly.
Check The fan is cleaned without obvious dust and rotates smoothly..
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
Check Visually check the vortexer hole is clean.
4. After cleaning, check if there is cotton fiber left in each hole and
reinstall the shield cover.
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
Check None
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
6. Hang a new finger clamping spring on the spring guide post, and
note that the spring hook must be accurately hung in the groove of
the spring column.
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
Maintenance Item Replacing the Spring of the Gripper
7. Install the spring and tighten the screws in reverse order of step 3
to step 4, then reinstall the shield cover
8. When the maintenance instruction is executed successfully, click
Exit to complete the maintenance.
Check Start up the instrument, and confirm that the gripper works properly.
Maintenance takes 15min
4. Gently wipe the probe outer wall from top to bottom with a piece
of dustless gauze dipped with alcohol to ensure that the outer wall is
wiped clean.
4. Reinstall the transparent shield cover (after finishing other
maintenance (if any)).
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
Check No stains on the probe surface, exposing metallic luster.
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
content unit
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
5. After draining and priming, click OK to complete the
maintenance.
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
2. Place tissue under the pump head to prevent the liquid from
spilling onto other parts.
3. As shown in the following figure, from top to bottom, there are 5
tubes on the pump head to be installed.
Tube press block
Roller
Pump tube
clamp
Knurled nut
Pump tube
slot
4. Unscrew the two knurled nuts and open the tube press block to
remove the five tubes from top to bottom. Take caution to protect the
pump tube from being scratched.
5. Install the new tubes from bottom to top. After installing each
pump tube, rotate the peristaltic pump roller at least 10 times to
ensure that the pump tubes are in their respective channels to avoid
pump tube intertwining.
Note: Make sure the direction of each tube is correct. Otherwise,
dispersion wash may fail. If the direction cannot be confirmed, refer
to other tubes that have not been removed.
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
6. After installing the new pump tubes, first tighten the inner knurled
nut and then the outer one to fix the tube press block.
Note: Do not tighten the two nuts simultaneously.
7. Power on the analyzer for final confirmation.
Note: Before replacing the pump tubes, pull the roller to check
whether it can rotate smoothly. If no, replace the dispersion aspirate
peristaltic pump assembly.
Check After replacing the tubes, test the flow to check if the dispersion
aspiration function is normal:
1. Prepare 4 empty cuvettes, marked with A1, A2, A3, A4
respectively. Use the calibrated pipette to add 1 ml pure water into
the 4 empty cuvettes, and place the 4 cuvettes on an empty cuvette
tray.
2. remove four aspirate probes A1~A4 and put them into four
cuvettes respectively. Make sure the probe tip reaches the bottom of
the cuvette and coincides with the axis of the cuvette without tilting.
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
5. After the peristaltic pump stops, take out aspirate probes A1~A4,
use the 200ul pipette to measure the liquid volume in the marked
cuvettes A1~A4 and record the volume.
6. Calculate the aspirate flow of the A1~A4 aspirate probes
respectively.
Aspirate flow = (1000ul- remaining liquid volume in the cuvette after
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
aspiration (ul))/5 s
Note: If the pipette cannot measure the volume accurately, a ruler
can be used.
7. After the flow test, install back the aspirate probes and other
components.
Maintenance takes 20 min
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
positions on the dispersion carousel.
4. The engineer starts cleaning. Take A4 probe as example.
Unscrew the round nut of A4 aspirating needle on the aspirate board
counter-clockwise and lift it upward, and pull the aspirating needle
from the dispersion carousel. Do not bend or bump the probes.
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
needle, and use the unclogging device to clean its interior. After
exterior cleaning, connect the syringe tool (included in the
accessory kit) with the quick joint of the aspirating needle. Prepare a
beaker full of probe cleanser, immerge the other end of the probe
into the beaker, and push/pull the syringe plunger for over 10 times.
Lift the probe tip above the liquid level and discharge the liquid in the
tube and prepare another beaker filled with deionized water or
distilled water, immerge the probe tip into the liquid, pull the syringe
plunger to aspirate liquid, put the probe into another empty beaker
and push the syringe plunger to discharge the liquid. Repeat this
action for at least 10 times to clean the probe interior. After interior
cleaning, disconnect the probe assembly from the syringe
connector, swing the probe and tube to remove the liquid leftover,
and then use clean gauze (without scraps) to dry the probe exterior.
Please note that the tube is connected tight during the cleaning
process
7. Put the probe back: Take the clean aspirating needles and
put the probe back into the aspirate board. Take care during the
process to prevent the probe from colliding or scraping with other
accessories. Turn the round nut clockwise to lock the aspirating
needle. Gently lift the aspirating needle to check whether the
aspirating needle is smoothly rebounded automatically. Connect the
quick joint of the aspirating needle guide tube with the
corresponding joint on the dispersion aspirate adapter board. When
connecting them, press the joints gently and screw them clockwise
until the buckles of the joints are fastened and a click is heard.
Check whether the joint is fastened by shaking it gently. Fix the tube
onto the press plate and the black ring symbol on the tube should
align with the edge of the press plate.
8. Click [Finish], the gripper will throw the cuvettes, click the
[Exit] button to complete the maintenance.
Check The surface of the aspirate probes is free of dirt and crystals, and
shows metallic luster. Result of routine cleaning shows Pass,
Maintenance takes 20 min
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
them.
System Standby
status
How to do 1. Confirm the system is in the Standby status and the remaining reaction
cuvettes are ≥ 4
2. Select Utility – Maintenance – Maintenance.
3. Click Dispersion Probes/Tubes Maintenance – Execute.
4. The engineer starts the maintenance, unscrewing counterclockwise the
round nut on the dispense probe and removing the dispense probe from the
dispersion carousel.
Note: Keep the hose connection tight without loosening or falling off.
5. Use the clean gauze to wipe the outer wall of the dispense probes till the
surface of the probes shows the metallic luster.
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
6. After cleaning, reinstall the dispense probes on the dispense carousel and
tighten the round nut.
7. Check if the hose connection is loose. If yes, tighten it.
8. Repeat Step 3 ~ 5, and clean other 3 dispense probes.
9. Click OK to finish the cleaning.
10. Close the shield cover of the analyzer.
11. Fill out the maintenance log.
Check The surface of the dispense probes is free of dirt and crystals, and shows
metallic luster.
Maintenance 8 min
takes
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
3. Loosen the two screws fixing the substrate leakage box and take
the box out carefully.
The substrate
leakage box is under
the substrate bottle
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
operation to prevent biological risk.
2. Cut off the parts connecting the original tubes and the connectors
to make sure the sealing stability of during reinstallation. The tubes
must be inserted into the root of the connector.
System status Powered off
How to do 1. Confirm the system is powered off.
2. Remove the left side plate of the analyzer.
Find the substrate leakage box under the substrate bottle.
3. Visually check the isolation chamber for foreign particles.
4. If there are foreign particles in the isolation chamber, replace the
isolation chamber.
5. Reinstall the front panel of the analyzer and the shielding cover.
Check Visually check the isolation chamber is clean.
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
Item Index
Item Index
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
Sample CV CV ≤ 1.5%
7 Maintenance FAQs
Common Maintenance Errors Possible Reasons
Dispersion wash buffer enters the tube slowly. The cap of the wash buffer tank is not
clean during maintenance, causing the
air vent to be clogged.
Inaccurate dispensing of sample or reagent affects The syringe is not checked for leak or
results. air bubbles during maintenance.
The system temperature or reagent refrigeration The dust screen and the radiating fan
temperature is high. are not cleaned during maintenance.
The gripper fails to grip cuvettes occasionally. The gripper is not cleaned or not
cleaned thoroughly during maintenance.
The gripper fails to grip cuvettes or reset frequently. The spring of the gripper is not replaced
during maintenance.
The result is inaccurate, causing The inner wall washing of the probe is
the carried contamination between sample reagent. not checked or the outer wall is not
cleaned during maintenance.
The wash well is clogged. The wash well is not cleaned during
NO:ZSH-22002-CL-900i Version:1.0
IVD Global Technical Support Dept
maintenance.
The device service life exceeds the limit. The Relevant parts are replaced during
prompt is red. maintenance, but the counter is not
reset.
NO:ZSH-22002-CL-900i Version:1.0