Machine Translated by Google

BC-5390
Auto Hematology Analyzer
Machine Translated by Google

BC-5390
Auto Hematology Analyzer

The new BC-5390 Auto Hematology Analyzer provides rapid and reliable test from just 33uL of blood. Using three

Mainstream technologies: laser scatter, low cytometry and chemical dye, BC-5390 can provide accurate differential readout.

In order to save time and increase walk-away automation, an autoloader is equipped to hold 40 tubes one time and achieve

60 samples/hr throughput. With the popular windows based software, you can easily perform the routine tests, manage

patient results, set up auto-cleaning and connect with LIS server. BC-5390 is your ideal choice to streamline daily work.

Semi-conductor laser combined with chemical dye Up to 60 samples processed per hour
advanced method of cytometry Supports whole blood mode for capillary sample
21 parameters with complete 5-part Large storage capacity: up to 100,000 samples
dierentiation(CBC+DIFF) on white blood cells Support bi-directional LIS connection
40-tube autoloader with random access Customization on reference range, auto-cleaning
Closed tube for STAT samples schedule and report format
Capability to ag abnormal samples

Only 33 ÿl sample volume for CBC + DIFF results


Machine Translated by Google

White blood cell differentiation M-5 reagents and BC-5D controls

The semi-conductor laser system collects Only 4 routine M-5 reagents are used in
forward scatter and side scatter counting cycle featuring economical
information of white blood cells in a ow consumption and 2 years shelf life. To help

cell and differentiates lymphocytes, precision monitoring, three levels of BC-5D

monocytes, neutrophils and eosinophils controls are provided in a ready-to-use kit


according to the cell size and granule and the assay value table can be automati-

complexity. Specific chemical dye cally imported through USB memory.


Eosinophilia sample
treatment to eosinophils can separate

WBC/BASO from neutrophils in DIFF scattergram Windows based software

even on Eosinophilia sample. In addition, The analyzer's windows-based software is


basophils are measured in an impedance simple to use and plays a powerful informa-

channel upon lysing action on the RBC tion hub to store 100000 patient results. You
0 100 200 fL
and other WBC groups. can set-up password access, reference range,
auto-cleaning schedule and so on. Also, the

Advanced impedance measurement for built-in report format tool can help to
RBC
RBC and PLT customize the final report type to include

Mindray patented digital sheath ow and microscopic counting, ESR, blood type and

weak signal detection techniques are diagnostic remarks.


0 100 200 fL
applied to ensure accurate counting and

sizing of RBC and PLT in the impedance QC monitoring and patient archive
PLT channel. Dynamic discriminator 60 QC les are designed to store LJ QC

adjustment can improve the separation of results. 300 data points can be recorded in

RBC and PLT groups when large single le. 2 common QC programs are

0 10 20 30 fL platelet and micro RBC are present. enabled for full quality assurance purposes.
Benefiting from these improvements, Patient data is archived and can be

MCV, RDW and PLT performance are searched and presented in trend curve for

greatly enhanced. case follow-up.

Dierent holder adapters supports three types of sample including

Venous whole blood sample, Predilute sample and Capillary whole blood

sample. Capillary whole blood sample can be well tested directly which is

more convenient for users and makes the analyzers an ideal choice to
work with pediatric and geriatric samples.

Venous whole blood sample Pre-diluted sample Capillary whole blood sample

Anti-aging samples with more than 12 hours transportation to central labs


which is a quite frequent and normal situation for small to medium labs in

countryside. Therefore, diagnosing anti-aging samples accurately becomes


a significant issue for developing countries or cities. With enhanced

1 hour 4 hours 8 hours 12 hours 24 hours reagent system and upgraded reagent formula, BC-5390 is an
ideal solution that can better differentiate white blood cells on

anti-aging samples under room temperature.


Machine Translated by Google

BC-5390
Auto Hematology Analyzer

Technical Specifications:

Principles Data Storage Capacity


Impedance method for WBC/BAS, RBC and PLT counting; Up to 100,000 results including numeric and
Cyanide free reagent for hemoglobin test; graphical information
Flow Cytometry (FCM) + Laser scatter + Chemical dye method for
Communication
WBC differential analysis
LAN Port supports HL7 protocol
Parameters

21: WBC, LYM(#,%), NEU(#,%), MON (#,%), EOS (#,%), BAS (#,%), RBC, HGB, Operating Environment

HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV Temperature: 15 °C~30°C

3 histograms and 1 scattergram Humidity: 30~85%


Air Pressure: 70~106 kPa

Analysis Mode
Autoloader Whole Blood Mode Power Requirement

Closed Whole Blood Mode AC100-240V ÿ 300VA 50/60Hz

Closed Predilute Mode


Dimension and Weight
Dimension (WxDxH, inches): 22.4 x 23.2 x 20.6
Throughput
Up to 60 samples per hour with autoloader Weight: ÿ 143 pounds

Up to 51 samples per hour with closed mode


Up to 53 samples per hour with predilute mode

Sample Volume
Whole Blood CBC+DIFF 33 ÿL
Whole Blood CBC 24 ÿL
Prediluted: 20 ÿL

Test Panel
CBC CBC+DIFF

Performance

Carryover Precision WBC Linearity


(x 103 /ÿL) ÿ 1.0% RBC (x 106 / ÿ 0.15 (SD) or 3.0% (CV%) 0.3-200

ÿL) ÿ 1.0% ÿ 1.5% 0.2-8.0

HGB (g/dL) ÿ 1.0% ÿ 1.5% 0.5-25

HCT (%) ÿ 1.0% ÿ 2.0% 2 -75

PLT (x 103 /ÿL) ÿ 1.0% ÿ 7.5 (SD) or 5.0% (CV%) 5-2000

is a trademark of Shenzhen Mindray Bio-Medical Electronics Co., Ltd.


Mindray Building, Keji 12th Road South,
2016 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. Specifications subject to changes without prior notice.
High-tech Industrial Park, Nanshan, Shenzhen 518057, PR China
P/N:ENG-BC5390FDA-210285x4-20170210
Tel: +86 755 8188 8998 Fax: +86 755 26582680
Email: [email protected] www.mindray.com