IR Spectroscopy and FTIR Spectroscopy - How An FTIR Spectrometer Works and FTIR Analysis
IR Spectroscopy and FTIR Spectroscopy - How An FTIR Spectrometer Works and FTIR Analysis
Credit: iStock
Contents
What is IR spectroscopy?
What is IR spectroscopy?
The human eye can only see a small part of the much broader spectrum of
electromagnetic radiation (Figure 1). On the high-energy side of the visible
spectrum lies the ultraviolet (UV) region, while on the lower energy side is the
infrared (IR). IR regions most useful for the analysis of organic compounds tend to
be 2,500 to 16,000 nm in wavelength. Far-, mid- and near-IR (NIR) are included
under the umbrella of “molecular spectroscopy”.
IR spectroscopy is the study of the interaction of IR light with matter, where IR light
is characterized by the wavenumber range spanning 12,800 to 10 cm-1. Historically,
by convention, IR tends to be described in “wavenumber”, where any wavenumber
is inversely proportional to its wavelength. Thus, a shorter wavelength will have a
larger wavenumber, referring to the fact that more waves would be able to fit into
a given distance. Far-IR is typically defined as radiation between 500 and 20 cm-1,
mid-IR between 4,000 and 500 cm-1 and NIR as typically between ~ 10,000 and
4,000 cm-1.
Figure 2: Animation showing the 3-dimensional movements that can occur for
molecular atomic bonds when excited by IR light. These movements cause the IR
spectral absorbance bands we observe. Credit: Taken from YouTube
https://www.youtube.com/watch?v=0S_bt3JI150?utm_source=363938&utm_medium=pd
f&utm_campaign=pdf_lead_conversion
The difference between IR and FTIR is that the latter is constructed from an
interferogram as the raw signal. This represents the light intensity as a function of
the position of a mirror inside the interferometer, not as a function of wavelength
(as occurs in dispersive instruments). This is the “FT“. The signal must first be
Fourier-transformed (FT) to produce the intensity as a function of wavenumber.
Today, most research and development grade mid-IR instruments are of the FT
type. Their development can be traced back to the 1890s and the work of Albert
Michelson, who, while studying the speed of light, invented the “interferometer”,
for which he received the Nobel Prize. An FTIR instrument uses an interferometer
(Figure 4), which consists of a source, beam splitter, two mirrors, a laser and a
detector. The energy goes from the source to the beam splitter which splits the
beam into two parts. One part is transmitted to a moving mirror, the other is
reflected to a fixed mirror. The moving mirror moves back and forth at a constant
velocity, controlled by the calibrating laser’s response. The two beams are reflected
from the mirrors and recombined back at the beam splitter, generating an
interference pattern, which is transmitted though the sample compartment (and if
present the sample where absorbance occurs) to the detector. This signal is then
subjected to the FT function to generate a spectrum.
1. The source: IR energy is emitted from a glowing black-body source and the
beam passes through an aperture, which controls the amount of energy.
4. The detector: The beam finally passes to the detector for final measurement.
5. The computer: The signal is digitized, the FFT calculation takes place and final
infrared spectrum is presented to the user.
There are now a whole range of FTIR instruments and versatile interchangeable
accessories that allow gaseous, liquid and solid samples of different sizes and
forms to be analyzed by the same basic instrument. It is worth noting that normal
glass is not mid-IR transmissible, so all instrument optics and sampling accessories
have to be constructed from other suitable IR optical materials. Early techniques
developed for solid samples required the analyte to be ground and mixed with IR-
transmissible substrates, often potassium bromide (KBr), under high pressure into
a small solid clear disk. These were then mounted in a holder for transmission
measurements. Liquid (non-water containing) samples were often formed as thin
films between two such IR-transmissible discs with a small spacer. Time and
reproducibility were both issues with this method.
Over the last 30 years, there has been increasing adoption of alternatives, notably
the now ubiquitous techniques of “ATR” (attenuated total reflectance). This device
can accommodate small amounts of liquid or solid sample placed onto a crystal
window with no real sample preparation required, allowing a spectrum to be
gathered in a few seconds. When solids are analyzed, they are firmly pressed onto
the crystal window by a top fixing clamp (Figure 6). Most published applications for
solid samples now use this form of device. Other types of devices, such as a
reflective hemisphere for diffuse reflectance or gas sample sealed cells are also
available for specific applications. There are even 96-position microtiter format
plates made from gold and other IR-compatible materials, allowing for high-
throughput screening using specially adapted FTIR accessory units.9
Figure 6: Use of an ATR sampling accessory.
FTIR spectra are information rich, but this very fact can make using and
interpreting them challenging. A useful primer for interpreting FTIR can be found
here. Even a simple, pure, single compound sample, like vanillin (Figure 8), has a
multi-peak spectrum. In such cases, library matching approaches to an
authenticated standard may identify it in a single component mixture, but this
would be impossible in complex mixtures of other compounds. The problem is
that most organics contain combinations of carbon (C), hydrogen (H), nitrogen (N)
or oxygen (O) atoms on single or double bonds, and so the same multiple
absorption peaks overlap from compound to compound. Simply trying to “eyeball”
lots of spectra to try to judge if or how the samples they come from are different in
some way then rapidly becomes an overwhelming challenge, even for an
experienced analyst. To address this problem, FTIR data is frequently used in
combination with statistical modeling approaches, such as multivariate
analysis (MVA), which in a chemistry context is commonly termed “chemometrics”.
FTIR spectral data is very amenable to MVA techniques, which at their heart simply
need multiple spectra to be collected from each sample and constructed into a
single data matrix. Here, each table row is an individual sample’s complete
spectrum, and each column is the aligned absorbance for particular consecutive
wavenumbers across all the samples. In this form, techniques like principal
components analysis (PCA) can be applied to explore and visualize possible class-
based relationships between different sample group’s spectral responses
efficiently via their scores plots. This gives an immediate “interpretability” of
sample differences, which is very difficult to gauge from simply over-laying spectra
from different sample classes. An example of a PCA scores plot is shown in Figure
9 for raw vs processed bio-oil sample differentiations.
Figure 9: Bio oil samples showing raw (red) compared to processed (green)
material. The FTIR spectrum (left) and the derived PCA scores plot (right) are
shown. Credit: Author.
When quantitation is the goal (e.g., concentration values), MVA approaches, such
as partial least-squares regression, can be used to build quantitative value-based
calibration predictions for properties such as chemical concentrations, using data
previously gathered on each sample’s composition from other assay techniques.
These known values are then fed into the algorithmic calculation to identify
spectral features that best correlate with the external value of interest. This latter
approach is very popular as, when properly planned and validated, it can allow
FTIR to effectively replace wet chemistry assays for new unknown samples (of the
same types), saving time and money. One useful consequence of constructing MVA
models is that the statistical output table produced, in addition to the sample
plots, allows the most important wavelength used in building the model to be
identified. From this information it is often possible to interpret some direct
chemical insights.
IR spectrum chart
The chart below (Figure 10) shows the bands produced by the major functional
groups (1500 cm-1 and above). The region 500-1500 cm-1, which is in the mid-IR
region, is called the fingerprint region and provides molecular fingerprints unique
to specific compounds that cannot be faked.
Figure 10: Chart showing the IR bands produced by the major functional groups
and the fingerprint region specific to each compound.
Advantages, disadvantages and uses of mid vs near-IR/FTIR
spectroscopy
The shape and structure of mid-IR vs NIR spectra are quite different, with mid-IR
spectra containing sharper and more defined spectral absorption bands (Figure
11) for organic species, which thus lend the technique to structural elucidation and
compound identification. Also, detailed data tables of characteristic molecular
function group wavenumber regions have been collated and published over the
years, many within specific application areas. Organic molecules strongly absorb
mid-IR radiation, so good spectra can be obtained from relatively little sample
material (e.g., a few powder grains). Disadvantages include its intolerance of water
(which quenches the IR signal even when present at just a few percent). In
addition, the very fact that organics generally absorb mid-IR so well means that the
resulting spectrum will only be from a few microns of sample penetration and
represent limited homogeny. Therefore, considerably more careful sample
preparation or analytical replication is needed.
FTIR sits in quite a unique “sweet spot” in terms of equipment costs, ease of use
and the information it can produce. Its flexibility has seen it applied to more areas
than there is room to discuss in this article. A range of pharmaceutical and medical
applications are becoming increasingly common.11
Article
Microplastic Detection
READ MORE
References (Click to expand)
-
2. Freitag S, Thain SC, Squier AH, Hogan EJ, Crittenden PD. Assessing
metabolic changes of the reindeer lichen C. portentosa to increasing
environmental N inputs using metabolomic fingerprinting and profiling
techniques. Comp. Biochem. Physiol. 2009;153(2, Supplement):S57.
doi:10.1016/j.cbpa.2009.04.518
7. Lewis W, Welch PD. Historical notes on the fast Fourier transform. Proc.
IEEE. 1967;55(10):1675-1677.doi:10.1109/TAU.1967.1161903
13. Kümmel T, van Marwick B, Rittel M, et al. Rapid brain structure and
tumour margin detection on whole frozen tissue sections by fast
multiphotometric mid-infrared scanning. Sci Rep. 2021;11(1):11307.
doi:10.1038/s41598-021-90777-4
14. Finlayson D, Rinaldi C, Baker MJ. Is infrared spectroscopy ready for the
Clinic? Anal Chem. 2019;91(19):12117-12128.
doi:10.1021/acs.analchem.9b02280
©2024 Technology Networks, all rights reserved, Part of the LabX Media Group