the detector was equal to the pitch of the screw for one
complete revolution of the stepper motor in 200 steps.
The microcontroller based data acquisition
and control system includes a 4 line 20 character LCD
display to display the current position of the detector
on the slotted line and the detector current in
microamperes as shown in Figure 2. Necessary
interfacing electronics, firmware and algorithm for
conversion of the ADC data and tracking of position
of detector is provided.
3000 ACKNOWLEDGMENT
One of the authors is thankful to UGC, Delhi for
2500
y = 32.826x provision of grants for carrying out this research work
2
2000
R = 0.9996 under the MANF scheme.
Output (mV)
References
1500
Data
1000
Fit to data 1. Laboratory Manual, Te0320 - Microwave Communication
Lab, Department Of Telecommunication Engineering, Srm
University S.R.M. Nagar, Kattan Kulathur – 603 203, July
500
2013.
2. David M. Pozar, Microwave Engineering Fourth Edition,
0 University of Massachusetts at Amherst, John Wiley & Sons,
0 20 40 60 80 100 Inc. (2005).
Input (mV)
3. D.V. Ahire* and P., R. Chaudhari, „Variation Of Dielectric
And Attenuation Properties For Tree Leaves With Moisture
Fig. 4 Current to voltage converter amplifier used Contents At Microwave Frequency‟ nternational Journal of
Physics and Mathematical Sciences I, Vol. 3 (1) January -
with the slotted line scanning and data acquisition March, pp.1-6 (2013).
system. 4. B.C. Glancy and A „Krall Automated Microwave Dielectric
It is seen that all the points lie well along a straight Constant Measurement System‟, Report, Naval Surface
line as is seen form the best fitting line joining the Weapon Centre, Virgenia 22448-5000, March – 1987.
5. Shaikh Yusuf H., Khan A. R and Behere S. H., „AVR
data points and the value of R2. A slight departure of Microcontroller Based Data Acquisition System for
the data points from the straight line is attributed to Laboratory Experiments‟, Advances in Applied Science
cumulative effect of all uncertainties including the Research, 2012, 3 (1) : 208-215
ADC error. It was found that it is convenient to 6. Neha chauhan and Ravi Kumar, „Data Acquisition System‟
Internatonal Journal Of Innovative Research In Technology,
calibrate the amplifier directly to reproduce the IJIRT | Volume 1 Issue 6 (2014).
current values rather than voltage obtained from ADC, 7. Badri Ram, „Fundamentals of Microprocessors and
this step was accomplished in the controlling program Microcomputers‟ Fifth edition (1991) : 311.
at the computer by providing a single conversion 8. Vijaya Bhaskar M. and Padma Suvarna R., „Embedded
System Design For Measurement of Dielectric Constant Of
factor to convert the ADC reading into current in Conducting Polymers At Microwave Frequencies‟,
microamperes. International Journal of Development Research, Vol. 5, Issue,
The current to voltage conversion and the gain of the 03, pp. 3778 – 3781, March, 2015.
amplifier were in agreement with the theoretical 9. Yusuf H Shaikh, Munde S. V., Microcontroller Based
Iontophoresis Power Supply with Data Acquisition System „
design values and the performance of the system was International Journal of Advances in Management ,
satisfactory, however during the development phase, Technology & Engineering Sciences‟ March 2013 vol.2 issue
there were issues like spikes and variations due to 6 127-129
movement of the detector. The kinks were more 10. Y. H. Shaikh, A. R. Khan, Patange, K. B., J. M Pathan, S. H.
Behere “Resistance of cell in fractal growth in
prominent near the peak, indicating that it is related to electrodeposition”International Journal of Artifical Life
the voltage, at low voltages, this effect is present but is research” Vol II No1 (2011).
getting merged in the curve. The kinks were found to 11. Ziad Salem, Ismail Al Kamal, Alaa Al Bashar “A Novel
be due to mechanical errors in the movement of the Design of an Industrial Data Acquisition System”, Proc. of
Int. Conf. on Inf. And Comm. Tech, ICTTA 2006, : 2589.
screw, the screw and bolt fitting had some design 12. V.G. Sangam , B.M. Patre, „Microcontroller based data
problems of fitting. The mechanical design has to be acquisition system for an amperometric biosensor and the
attended to for smoother movement of the detector analysis of glucose concentration‟ Jl. of Instrum. Soc. of India
assembly. Also noise pickup has to be suppressed and 2009 39(1):30.
13. P. Asimakopoulos, G. Kaltsas and A. G. Nassiopoulou, „A
necessary filtering was implemented. After removal of microcontroller based interface circuit for data acquisition and
the noise in the electronics and mechanical parts the control of a micromechanical thermal flow sensor‟ Institute of
standing waveform at the slotted line was reasonably Physics Publishing Journal of Physics: Conference Series 10
good as is seen from the selected screenshot in Figure (2005) 301 (http://iopscience.iop.org/1742-6596/10/1/074).
3, however some nonlinear readings are still seen
occasionally, particularly near the peaks that are
contributed by the shaking of the moving platform of
the detector.
IV. CONCLUSION
The microwave bench slotted line scanning and data
acquisition system along with the necessary
electronics, hardware and controlling software is
successfully designed, constructed and tested and
details of design, construction and performance are
presented.