Chapter 2 - Gas Chromatography (Student's Note)
Chapter 2 - Gas Chromatography (Student's Note)
CHAPTER 2
GAS CHROMATOGRAPHY
1
Course Learning Outcomes
2
General Chromatography
Chromatography
Gas Liquid
GC : Volatile solutes
(gas has lowest density, can only elute volatile comps with small molecular
weight, Mw)
(Sample MUST be volatile at temperatures BELOW 350oC)
3
GC separation is based on :
4
Gas-solid chromatography (GSC)
5
Gas-solid chromatography (GSC)
6
Gas-solid chromatography (GSC)
➢ Limited application :
• Semi-permanent retention of polar molecules.
• Tailing of elution peaks.
7
Capillary Column
8
Gas-liquid chromatography (GLC)
9
The chromatographic process
- Partitioning
Separation is achieved by partitioning the sample
component between the phases
(gas)
MOBILE PHASE
Sample
Sample
out
in
STATIONARY PHASE
(liquid coated onto a solid support)
10
Schematic diagram of
gas chromatography
Sample is injected & vaporized
onto the head of the column
As each
component
emerges from the
column, it will be
detected
11
Performing GC separations
➢ From the time the mixtures are injected until they reach the detector,
they are being retained by the liquid stationary phase in the column.
12
GC Instrumental components
2. Columns
4. Detectors
13
1. Carrier gas
(mobile phase)
14
Carrier gas (mobile phase)
15
➢ Pressure regulators, gauges & flow meters are required to
control the flow rate of the gas.
Inlet
Outlet Gauge
Gauge
Gas
Shutoff Inlet
Pressure Control
16
➢ Type of gases for carrier gas (mobile phase)
✓ Helium (common)
✓ Nitrogen
✓ Hydrogen
✓ Argon
17
➢ Carrier gas :
18
Advantages & disadvantages of carrier gas
19
20
Typical carrier gases
21
➢ The carrier gas flow can be quantified by:
Average linear velocity (ū) is the distance passed (cm)/sec by mobile phase
in the column (cm/sec).
22
Recommended linear velocities & flow rates
23
Van Deemter Plot for GC
➢ Theories of band broadening in GC can be explained by
the Van Deemter equation.
24
Van Deemter Plot
➢ A plot of plate height vs average linear velocity of mobile
phase.
25
➢ With N2, the optimum flow rate is lower compared to H2.
Gas density : N2 > He > H2 DM : H2 > He > N2
➢ B α diffusion coefficient of analyte in mobile phase, DM
➢ H2 is lighter than N2 (heavier; larger Mw), therefore DM in H2 is higher than
in N2.
26
➢ The slope of the curve at higher flow rate is shallower when using H2 due
to smaller C term (the equilibrium of analyte between the mobile phase
(H2) & stationary phases is more efficient).
➢ The more rapidly a solute diffuses between phases, the smaller is the C
term.
27
The mobile phase mass transfer (CMv)
CMv = (fM(k’)dp2/DM)v
dp = particle diameter of packing
DM = mobile phase diffusion coefficient
28
Sample Problem 2.1
GC, HPLC & SFC are 3 types of separation methods with different mobile
phases. Compare & discuss how each mobile phase affects the B term
of van Deemter in each of the above separation method.
29
Sample Problem 2.2
Which term (A, B or C) is most significant in GC? Explain why & suggest
one approach to minimum the effect.
30
Effect of mobile phase flow rate on H for GC and LC
31
2. Column
32
Type of Column
Types of column
33
Column Dimension
34
Packed column
35
Tube made of glass, metal (stainless steel, copper,
aluminium) or teflon with 2-6 mm I.D. & 1-5 m in
length.
Solid support :
✓ Small & uniform particles size
✓ Mechanical stability
✓ Porous
Solid support ✓ High surface area
✓ Serve to hold the ✓ Inert
liquid stationary
phase
36
Column Wall
Carrier
Gas
Solid Support
Liquid Phase
37
➢ Small diameter particle (dp), decrease the time required
for solute equilibration, thus improve column efficiency.
CMv = (fM(k’)dp2/DM)v
38
➢ Flow of the mobile phase through a capillary column is
relatively unimpeded when in comparison with a packed
column.
B/v = 2ɣDM/v
39
➢ Due to the difficulty of packing the tubing
uniformly, these types of columns have a larger diameter
than capillary columns & have a limited range of length.
40
Capillary column
41
Wall-coated open tubular column, WCOT - The inner
wall are directly coated with liquid stationary phase
(no support material).
42
Tubes / capillary (capillary made of metal, glass, stainless steel or
thin fused silica (SiO2))
Stationary phase
SCOT: the inner wall of the capillary is lined with a thin layer of solid
support onto which the stationary phase has been adsorbed.
43
Fused silica capillary column
➢ Much thinner (0.1 mm) than the glass capillary columns & are given
strength by the polyimide coating.
44
(imparts a brownish color to the columns)
45
Compared with packed column, capillary column offers :
✓ Higher Rs with narrow peak
✓ Shorter analysis times due to the higher flow rates applied
✓ Greater sensitivity
Disadvantage :
X Lower sample capacity
100 ng for a 0.25 mm I.D. column with 0.25 µm thick film
5 µg for a 0.53 mm I.D. column with 5 µm thick film
46
Properties of GC columns
47
➢ Capillary columns tend to have lower H (greater N per given column
length) since there is no A term (no packing).
➢ Narrow ID column (eg. 250 µm) has lower C term compared to wide ID
column (eg. 500 µm)
48
Column temperature
49
GC columns
50
➢ Film thickness primarily affect the retentive character & the
capacity of a column.
51
2. Thick films and wide bore columns are required with thermal
conductivity and IR detectors.
✓ Have high sample capacity.
✓ Low Rs & long tR.
3. Thin films and narrow bore columns are useful for separating
species of low volatility (high boiling point) in a reasonable
time & more efficient.
Disadvantages :
▪ Low sample capacity
▪ Require high sensitive detector
52
GC column comparisons
Description Thin film narrow bore Thick film narrow bore Thick film wide bore
53
Sample Problem 2.3
If a GC column is changed from packed to capillary column, explain
the effect of the changes on the H.
54
Stationary Phase
➢ Ideal properties :
▪ Low volatility
▪ Thermal stability
▪ Chemical inertness
▪ Low viscosity
55
➢ Choice :
✓ Depend on the analyte of interest using the principle “like-dissolve-
like”.
✓ polar liquid for polar analytes
✓ nonpolar liquid for nonpolar analytes
➢ When the match is good, the order of elution is determined by the boiling
point of the analytes.
56
➢ Order of elution:
➢ Less retained analytes elute earlier.
➢ Given similar molecular characteristics, more volatile analytes (low
boiling point) elutes earlier. Low molecular masses elute earlier.
57
➢ Polar stationary phases contain functional groups e.g., –CN, -CO- and –
OH.
➢ Analytes:
58
Examples of liquid stationary phases :
1. Polydimethyl siloxanes (PDMS):
➢ Non to strong polarity depending on R groups.
➢ Most stable, robust & versatile.
R R
O Si O Si
R R n
59
➢ The most basic polysiloxane is the 100% methyl substituted.
➢ The % give the amount of substitution of the named group for methyl
groups on the polysiloxane backbone.
➢ E.g., 5% phenyl PDMS (5% phenyl group and 95% methyl group).
60
Examples of liquid stationary phases :
HO-CH2-CH2-(O-CH2-CH2)n-OH
61
Stationary Phase Common Maximum Common Applications
Trade Name Temperature (C)
General-purpose nonpolar
Polydimethyl siloxane phase; hydrocarbons;
(NON POLAR) HP-1, SE-30 350 polynuclear aromatics; drugs;
steriods; PCBs
5% Phenyl-polydimethyl Fatty acid methyl esters;
siloxane HP-5, SE-52 alkaloids; drugs; halogenated
350
compounds
62
Typical chromatograms for open tubular columns coated with
(a) PDMS; (b) 5% phenyl methyldimethyl siloxane; (c) 50% phenyl methyldimethyl siloxane; (d) 50%
poly(trifluoropropyl-dimethyl) siloxane; (e) PEG; (f) 50% poly(cyanopropyl-dimethyl) siloxane. (Courtesy of
J&W Scientific)
63
Nonpolar versus polar stationary phase
64
Bonded & cross-linked stationary phases
➢ Purpose :
65
Column bleeding
➢ Untreated column slowly lose their stationary phase owing to ‘bleeding’ in
which a small amount of immobilized liquid is carried out of the column
during elution.
66
➢ Process :
▪ Cross-linking:
67
Solid support (packing)
68
➢ A problem in GC is the physical adsorption of polar analytes (eg.
alcohols/aromatic hydrocarbons) on the silicate surfaces of column
packings & capillary walls.
69
➢ The SiOH groups on the support surface have a strong affinity for polar
molecules & tend to retain them by adsorption.
➢ NOTE: Operating pH for silica stationary phase = pH 2-8, siloxane bond (Si-O-
Si) to the stationary phase hydrolysed below pH 2 while above pH 8 it will
dissolved
70
Example of peak tailing
71
Asymmetric peak shapes
72
Deactivation of support materials
➢ Process:
73
1. Silanization using
DMCS
2. Washing with
methanol
74
Kovats Retention Index (RI)
➢ Retention index relates the retention time of a solute to the
retention time of linear alkanes
75
76
3. Sample injection
system
77
Injection port / inlets
(sample introduction)
➢ The inlet is a piece of hardware attached to the column head.
➢ To flash/ instantly evaporate the sample & introduce it into the column.
➢ Tinj > 50 oC higher than the boiling point of least volatile component of
the sample (250°-275° C is adequate) to ensure fast & complete
vaporization.
78
➢ Injection using a microsyringe (gases are injected by a gas-tight syringe)
through a rubber septum.
79
➢ Column efficiency requires :
▪ Sample volume should not be too large due to it will cause broad peak
due to slow vaporization of sample.
80
Packed column
➢ Volume injected :
▪ Liquid samples – 0.1-10 µL
▪ Gas samples – 1-10 mL
➢ All the vaporized sample from the injector enters onto the
column.
81
Capillary columns
➢ Volume ~10-3 µL.
➢ The size of the capillary column limits the amount of analyte that can be
injected, otherwise, overloading occurs.
82
Recommended procedure for sample loading into syringe
83
Injector types
1) Split injector
2) Splitless injector
3) On-column injector
84
85
Septum Nut Septum
Viton O-Ring
Liners
86
Splitless Packing Split Packing
Liner
Ground
4 inches long. Ground
Portion
Portion
6 mm O.D. / 4 mm I.D.
6 mm O.D. / 2 mm I.D. O-Rings O-Rings
87
1) Split Injection (sample splitter)
88
➢ Large part of sample vented out via the split valve will:
▪ Prevents overloading of the column.
▪ Produces narrow & sharp peaks.
➢ Split ratio: the proportion of sample vented out from the column (typically
write 20:1 to 100:1).
▪ E.g. 50:1, that is, for every 50 units of gaseous sample that are vented
out to waste, 1 unit goes on the column.
▪ Higher ratio – Too much sample vented out, small amount of sample
enter the column, hence small peak & not reproducible.
89
(a) Split ratio : 1/25 (b) Split ratio: 1/75
90
2) Splitless Injection (split vent is closed)
➢ Split vent opened 30-90 seconds after injection, removes bulk of the
solvent (large solvent peak may overlap with the analyte peak), but
leaves most of the sample condensed at the top of the column.
91
➢ Purpose :
▪ Thus, reduce solvent peak and separate the analyte from the
solvent.
▪ Low conc. (e.g. sample extracted using SPME) : ensure all the
samples will be transferred into the column.
92
➢ Requires a low initial column temperature to dissolve & refocus the solutes
at the start of the column & then temperature programming.
➢ Transfer times are slower and peaks are broader when compared to split
injection.
➢ Two techniques can be used to get a narrow band (refocus the solutes) at
the head of the column: solvent focusing and analyte focusing.
➢ The difference between the two methods is the initial temperature of the
column oven.
93
1) Solvent focusing, the initial oven temperature is low enough (~20 °C
below the boiling point of solvent) to allow the solvent to recondense at
the head of the column. This forms a zone of liquid solvent that
traps/focus all of the analytes at the head of the column.
94
The Splitless Process
➢ Column temperature is kept at relatively low temp, lower (by at least 10 oC)
than the boiling point of the solvent.
➢ Volatile solvents will condense at the top of the column & solute will
dissolve & refocus in the solvent (solvent trapping).
➢ Less volatile solute will re-condense on top of the colder column & refocus
too (cold trapping).
➢ Then, split vent will be opened to remove bulk of the solvent. (Note: large
solvent peak may tailing and will overlap with the analyte peak).
95
➢ Sample that is concentrated at the top of the column and then will
introduce into the column giving a sharp peak.
96
toluene
toluene
➢ Advantages :
✓ Good for trace analysis (less than 100 ppm).
✓ Good for high boiling point compounds.
✓ Easily automated.
97
98
3) On column injection
➢ Used for samples that easily decompose above their boiling point
or thermally labile compounds.
99
➢ Sample is directly injected into the column as a liquid without
vaporization with the aid of a syringe with a long, narrow needle
whereby the injector is maintained at low temperature.
➢ Advantages:
100
101
4. Detectors
102
Detectors
103
Flame Ionization Detector
(FID)
➢ Analytes containing carbon burn in hydrogen-oxygen flame and produce
ions.
➢ CH + O → CHO+ + e-
104
➢ Useful :
✓ Responds to the number of carbon atoms entering the detector per
unit time, it is mass-sensitive, rather than a concentration-sensitive.
✓ General detector for the analysis of organic compounds or
hydrocarbons.
✓ Insensitive to nonhydrocarbons (H2, He, N2, O2, CO, CO2, H2O, NH3,
NO, H2S).
✓ High sensitivity.
✓ Large linear response range.
✓ Low noise.
✓ Robust.
✓ Easy to use.
105
Electron-Capture Detector
(ECD)
➢ The sample eluate from a column is passed over a
radioactive β-emitter like nickel-63.
➢ An electron from the emitter causes ionization of
the carrier gas (often N2) and will produce a burst
of electrons.
➢ In the absence of organic species, a background
current between a pair of electrodes results from
this ionization process.
➢ When analytes that can capture electrons e.g with
halogens, peroxides & nitro groups enter the
detector, electrons from the background current
are captured, therefore the current is suddenly
decreased.
106
➢ Advantages:
✓ Non-destructive.
✓ Sensitive.
ECD cannot be used with inert gas like He due to no interaction of electrons
with the carrier gas molecules to form a background current.
107
Thermal Conductivity Detector
(TCD)
108
➢ The temperature of the source depends upon the thermal conductivities
of the surrounding gases.
➢ Carrier gases such as helium and hydrogen has very high thermal
conductivities so the addition of even a small amount of sample is
readily detected.
109
➢ Advantages:
✓ Non-destructive.
110
GC detectors & their detection limits
Symbol Detector Type Selectivity Detection limit
Most organic cmpds that ionize in
FID Flame Ionization 100 pg
flame
Thermal
TCD Universal 1 ng
Conductivity
Nitrogen
NPD Nitrogen, phosphorus 10 pg
Phosphorus
Sulphur, phosphorus, tin, boron,
Flame
FPD arsenic, germanium, selenium, 100 pg
Photometric
chromium
111
Detection techniques & their applications
Detector Used for
Mass spectrometer (MS) Identification of unknown compounds
Flame Ionization Detector Compounds containing Carbon
(FID)
112
Interfacing GC with
spectrometric detectors
113
Mass Spectrometry Detector
➢ The MS does this by breaking each molecule into ionized fragments and
detecting these fragments using their mass to charge ratio.
114
➢ Advantages of MS:
115
116
Method Development for GC
117
Areas to optimize
➢ Injector
➢ Carrier gas
➢ Column Temperature
✓ Minimal temperatures give good Rs, but increase the elution times.
118
Factors which affect GC separations
119
b) Polarity of compounds : Polar compounds will move slowly
if the column is polar. Differences in polarity of the
compounds is only important if you are separating a mixture
of compounds which have widely different polarities.
120
d) Column polarity : Usually, all compounds will move slower
on polar columns, but polar compounds will show a larger
effect.
121
➢ Factors to be considered in selecting the column :
122
Column Length vs. resolution
123
Column Diameter vs. resolution
124
The heights of the later eluting peaks are
ISOTHERMAL reduced & the peak widths increased
because they are more affected by the lower
Constant column temperature temperature program used.
throughout the analysis.
TEMPERATURE PROGRAMMING
Oven temperature:
Column temperature is 60oC for 1 min
increased either continuously or 60-180oC at 20o/min
in steps as the analysis
proceeds.
Increase in temperature
✓ Decrease retention time
✓ Sharpens peak
125
Temperature Program Method
126
➢ When column temperature increases, the early eluting
compounds (low boiling point) will overlapped & late eluting
compounds (high boiling point) separated.
127
➢ In the analysis of a broad boiling point range, isothermal
analysis may not produce an efficient separation. The temp
programming able to solve general separation problem such
as long elution times that cause band broadening which lead
to poor Rs.
128
➢ A temperature program allows comps that elute early in the
analysis to separate adequately, while shortening the time
it takes for late eluting comps to pass through the column.
129
Solution to general elution problem
130
Increase column length
α = (tR)B – tM/(tR)A – tM
131
➢ The most effective way to alter Rs is to change the
selectivity or the capacity factor of the column.
132
Overloaded column
133
➢ The front of an overloaded peak has
gradually increasing concentration.
Asymmetric peak
➢ The solute is so soluble in the
overloaded zone that little solute trail
behind the peak.
134
Sample Preparation for GC
a) Derivatization
b) Pyrolysis
c) Headspace
135
a) Chemical Derivatization
136
➢ Chemical derivatization can : (Advantages)
✓ Increase the volatility & decrease the polarity of
compounds.
✓ Reduce thermal degradation of samples by increasing
their thermal stability.
✓ Reduce tailing.
✓ Increase specific detector response by incorporating
functional groups which lead to higher detector signals,
e.g. CF3 groups (addition of halogen atoms) for ECD.
137
➢ Common derivatization methods can be classified into 4
groups depending on the type of reaction applied :
▪ Silylation
▪ Acylation
▪ Alkylation
▪ Esterification
138
b) Pyrolysis
139
➢ Large molecules cleave at their weakest points and
produce smaller, more volatile fragments. These fragments
can be separated by gas chromatography.
140
c) Headspace
141
▪ G = Gas phase (headspace); The gas phase is
commonly referred to as the headspace and lies
above the condensed sample phase.
142
THANK YOU FOR YOUR ATTENTION
THANK YOU
Q & A SESSION
CHM 510
ANALYTICAL SEPARATION METHODS
143