GC Troubleshooting
GC Troubleshooting
• Signal connections • Temperature settings • Analytical conditions Sample loading capacity • Inject less sample (dilute, use split injection, • Verify signal settings and adjust if needed.
Poor Resolution exceeded reduce injection volume).
Electronics
• Repair or replace cables or boards.
Fast autosampler injection • Use wool or slow injection speed.
Identify the Cause: Causes Solutions
into open liner
Dirty or damaged
detector
• Perform detector maintenance or
replace parts.
• Define the problem clearly; for example, “Over the last 4 days, only the phenols in my sample have been tailing.” Non-selective • Choose appropriate stationary phase and
Flow/temperature • Verify steady flow rates and temperatures, then
stationary phase column dimensions.
settings wrong or variable adjust settings and/or replace parts if needed.
• Review sample and maintenance records to identify trends in the data or problem indicators, such as area counts • Optimize carrier gas linear velocity and
Poor efficiency • Remove contamination and use properly
decreasing over time or injector maintenance not being performed as scheduled. GC oven temperature program.
Carryover/Ghost Peaks Adsorption/reactivity
deactivated liner, seal, and column.
Causes Solutions
Unstable Baseline (Spiking, Noise, Drift) Low flow rates
• Verify injector and detector flow rates and
adjust if needed.
Incompatible stationary • Choose appropriate stationary phase. • Verify make-up gas flow and adjust if needed.
phase Causes Solutions
Slow GC oven program • Increase GC oven programming rate.
• Reduce amount injected, dilute sample. • Leak check connections and replace seals
Spiking Carrier gas leak or if needed.
Column overloading • Increase column inner diameter and/or contamination Poor analyte/solvent • Lower GC oven start temperature.
film thickness. • Replace carrier gas and/or detector gas filters. focusing
Injector or detector • Clean system and perform regular • Reduce retention of compounds by
Column film is too thick
contamination maintenance. decreasing film thickness and length.
Column contamination • Condition, trim, and rinse column. Sample carryover • See Carryover/Ghost Peaks solutions.
Tailing Peaks or stationary phase bleed
• Replace septum.
Causes Solutions Septum coring/bleed • Inspect inlet liner for septa particles and
Noise replace liner if needed.
• Use properly cleaned and deactivated liner,
Adsorption due to surface seal, and column. Loose cable or circuit • Clean and repair electrical connections.
activity or contamination • Trim inlet end of column. board connections
• Replace column if damaged.
• Verify flow rates are steady and reproducible;
Variable carrier gas or may need to replace or repair flow controller.
Adsorption due to chemical • Derivatize compound. detector gas flows
composition of compound Drift • Leak check system.
• Check for leaks at all connections, replace • Allow enough time for detector temperatures
Leak in system Detector not ready
For an in-depth discussion on how to choose the right column and critical seals if needed. and flows to equilibrate.
improve your chromatographic results, visit www.restek.com • Minimize dead volume.
Installation issues • Verify that the column is cut properly (square).
• Download our Guide to GC Column Selection and Optimizing Separations
(lit. cat.# GNBR1724- UNV) • Verify correct installation distances.
© 2013 Restek Corporation.
• Request our Simplifying Column Selection poster (GNWC1612-UNV) Lit. Cat.# GNWC1723-UNV All rights reserved.
Printed in the U.S.A.