18.robotic Dental Implant Placement Workflow For Edentulous Jaws
18.robotic Dental Implant Placement Workflow For Edentulous Jaws
Supported by the Shandong Municipal Health Commission [project no. 202208050566]; the Qingdao Natural Science Fund [project no. 23–2-1–133-zyyd-jch]; and by
the Qingdao Municipal Health Commission [project no. 2021-WJZD183], PR China.
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
a
Postgraduate student, Department of Oral Implantology, The Affiliated Hospital of Qingdao University, School of Stomatology of Qingdao University, Qingdao, PR China.
b
Attending, Department of Oral Implantology, The Affiliated Hospital of Qingdao University; School of Stomatology of Qingdao University, Qingdao, PR China.
c
Resident, Department of Stomatology, Qingdao Women and Children's Hospital, Qingdao University, Qingdao, PR China.
d
Professor, Department of Oral Implantology, The Affiliated Hospital of Qingdao University, School of Stomatology of Qingdao University, Qingdao, PR China.
1
W.W. and X.L. contributed equally to this article.
A B
Figure 2. Data collection. A, Remaining teeth meet data registration
requirements and merge. B, Template with radiopaque markers.
A B
C D
Figure 3. Project design. A, Data registration and merge. B, Added crowns and adjusted occlusion. C, D, Designed implant position and implantation
steps for robot (red arrows: vertical maxillary first molars).
The patient and robotic device were positioned to prosthesis was attached to the implants using a pick-up
ensure that the robot's eye was approximately 1 m above technique13 and immediately loaded (Fig. 6).
the surgical area and that the extraoral, robot, and probe The STL data of the preoperative design and the
markers were located at the same level and could be DICOM data of the postsurgical CBCT were imported
detected by the stereo visual sensor (green indicated a into the software program (DentalNavi; YakeRobot
good connection). The “calibration probe” option was Technology Ltd) to conduct the deviation analysis.
selected on the software program, the positioning plate Preoperative planning and postoperative CBCT were
registration was completed, and implant calibration was manually matched based on the anatomic structure, and
performed using the calibration probe marker, the entry the deviations between the actual and planned 3D im
and exit paths of the robot handpiece were recorded, plant positions were measured and recorded. The results
and the spatial mapping relationship between the vision of the accuracy analysis in the patient were platform
and the CBCT image was established (Fig. 4C, D). Fi (4.10 ±0.90 mm), apex (0.60 ±0.19 mm), depth (0.49
nally, the entry and exit paths were reconfirmed. ±0.15 mm), and angle (1.43 ±0.78 degrees) (Fig. 7,
The surgeon (B.Z.) replaced and installed the drills Table 1).
according to preset implant procedures and controlled
the robot to complete the osteotomy preparation and
implant placement. The process was displayed in real-
DISCUSSION
time on the DentalNavi screen. The axis, angle, and
depth of the robotic implantation could have been ad Precise placement of implants is essential for long-term
justed at any time according to the surgical situation. On esthetic and restorative outcomes.14 The static CAIS
the day of surgery, owing to the patient's concern that facilitated implantation in edentulous patients, with
sinus elevation might lead to maxillary sinus perforation, greater precision and predictability, than freehand. A
bleeding, or infection, the plan was changed to tilt the review of the accuracy analysis of edentulous implants
implants at the maxillary first molar sites (Fig. 5). All 6 using static CAIS, including 2037 implants, showed that
implants achieved sufficient primary stability above 35 the mean deviation of implant platforms and apex was
Ncm. After surgery, a prefabricated computer-aided 1.35 mm and 1.34 mm, respectively and the mean de
design and computer-aided manufactured interim viation of angle was 3.59 degrees.15 The accuracy of
D
Figure 4. Preoperative preparation. A, Fixing registration of intraoral attachment. B, Cone beam computed tomography scan data matched with
preoperative plan. C, D, Preoperative registration to detect extraoral, robot, and probe markers.
C E
Figure 5. Surgical procedure. A, Robotic surgery overview. B, Osteotomy preparation. C, Implant placement. D, E, Real-time monitoring on
DentalNavi (changed plan during surgery, red arrows: tilted maxillary first molar).
A B C
D E
Figure 6. Immediate loading. A-C, Digital fabrication (red arrow: palatal extension for guide positioning). D, Composite resin abutments and
protector caps placed. E, Immediate loading completed.
A B C D
Figure 7. Deviation analysis on DentalNavi. A, Three-dimensional model. B, Cross-sectional view. C, Sagittal view. D, Coronal view.
Table 1. Deviations of implant position (mean ±standard deviation) This technology has a complete intelligent decision-
Deviation making system that could make reasonable decisions
Platform (mm) 0.49 ±0.21 based on the intraoperative force and positional feedback.
Apex (mm) 0.60 ±0.19
Depth (mm) 0.49 ±0.15
The robotic arm completes the autonomous operations
Angular (degree) 1.43 ±0.78 under system control, which should reduce the operating
difficulty and the learning cycle for the surgery. Surgeons
are not required to keep their heads down all the time to
implant placement with static CAIS is affected by factors operate; they only needed to observe and make timely
that include image acquisition, data processing, software adjustments, which was beneficial to their health, espe
design, guide template printing, template fixation, and cially for the cervical spine. The stability of the robotic arm
template types.16–18 However, an implant template is was higher than that of the human hand. If needed,
not needed for robotic surgery, which avoids associated custom suction tubes could be designed using a software
errors. The results of accuracy analysis in the present program for automatic intraoperative suction.
patient were platform (0.49 ±0.21 mm), apex (0.60 The surgical plan, including the type, position, and
±0.19 mm), depth (0.49 ±0.15 mm), and angle (1.43 direction of the implant, could be flexibly adjusted
±0.78 degrees), more accurate than CAIS, and consistent during surgery without any attachment replacement. For
with a previous study on the accuracy analysis of the presented patient, 6 implants were planned with 2
edentulous jaws.19 treatment options involving the maxillary first molar:
whether to perform sinus floor elevation or tilt the im 3. Siqueira R, Chen Z, Galli M, et al. Does a fully digital workflow improve the
accuracy of computer-assisted implant surgery in partially edentulous
plant. Because access for the robot to place a distally patients? A systematic review of clinical trials. Clin Implant Dent Relat Res.
tilted implant in the posterior area may have been dif 2020;22:660–671.
4. Tahmaseb A, Wismeijer D, Coucke W, Derksen W. Computer technology
ficult, sinus floor elevation was chosen. The patient had applications in surgical implant dentistry: A systematic review. Int J Oral
provided informed consent and was aware that the plan Maxillofac Implants. 2014;29:25–42.
5. Ozan O, Turkyilmaz I, Ersoy AE, et al. Clinical accuracy of 3 different types
could have been changed. On the day of surgery, owing of computed tomography-derived stereolithographic surgical guides in
to the patient’s concern about sinus floor elevation, the implant placement. J Oral Maxillofac Surg. 2009;67:394–401.
6. Vinci R, Manacorda M, Abundo R, et al. Accuracy of edentulous computer-
plan was changed to place tilted implants. Because of aided implant surgery as compared to virtual planning: A retrospective
the patient's large mouth opening and cooperation, the multicenter study. J Clin Med. 2020;9:774.
7. Barrak I, Joób-Fancsaly Á, Braunitzer G, et al. Intraosseous heat generation
full process of robot implantation was completed, and during osteotomy performed freehand and through template with an
the interim restoration was designed only to the second integrated metal guide sleeve: An in vitro study. Implant Dent. 2018;27:342–350.
8. Monaco C, Arena A, Corsaletti L, et al. 2D/3D accuracies of implant position
premolar. With other digital procedures, a change of after guided surgery using different surgical protocols: A retrospective study.
plans during the operation could only have been J Prosthodont Res. 2020;64:424–430.
9. Liu YF, Wu JL, Zhang JX, et al. Numerical and experimental analyses on the
achieved by freehand placement. Nevertheless, extraoral temperature distribution in the dental implant preparation area when using
marker fixation should be used to ensure robust stability a surgical guide. J Prosthodont. 2018;27:42–51.
10. Alevizakos V, Mitov G, von See C. Guided implant placement using an
during the procedure. internally cooling surgical template: A technical note. J Oral Implantol.
Limitations for this system include that adequate 2020;46:533–535.
11. Li Z, Xie R, Bai S, Zhao Y. Implant placement with an autonomous dental
patient opening is needed and robotic surgery in the implant robot: A clinical report. J Prosthet Dent 2023.
posterior area, especially in the second molar region, 12. Lindhe J, Nyman S. The role of occlusion in periodontal disease and the
biological rationale for splinting in treatment of periodontitis. Oral Sci Rev.
may be difficult. The preoperative preparation is com 1977;10:11–43.
plex and time-consuming. During surgery, the robot 13. Gallucci GO, Bernard JP, Bertosa M, Belser UC. Immediate loading with
fixed screw-retained provisional restorations in edentulous jaws: The pickup
system could automatically adjust and follow the pa technique. Int J Oral Maxillofac Implants. 2004;19:524–533.
tient's small range of movement. However, excessive 14. Taheri Otaghsara SS, Joda T, Thieringer FM. Accuracy of dental implant
placement using static versus dynamic computer-assisted implant surgery:
movement would lead to significant errors. Additionally, An in vitro study. J Dent. 2023;132:104487.
the surgeon has no surgical tactile perception. 15. Eftekhar Ashtiani R, Ghasemi Z, Nami M, et al. Accuracy of static digital
surgical guides for dental implants based on the guide system: A systematic
review. J Stomatol Oral Maxillofac Surg. 2021;122:600–607.
16. Vieira DM, Sotto-Maior BS, Barros CA, et al. Clinical accuracy of flapless
computer-guided surgery for implant placement in edentulous arches. Int J
SUMMARY Oral Maxillofac Implants. 2013;28:1347–1351.
17. Nomiyama LM, Matumoto EK, Corrêa MG, et al. Comparison between
The robotic implant system demonstrated the standar flapless-guided and conventional surgery for implant placement: A 12-
dization, repeatability, and predictability of the implant month randomized clinical trial. Clin Oral Investig. 2023;27:1665–1679.
18. Vercruyssen M, Laleman I, Jacobs R, Quirynen M. Computer-supported
process and was minimally invasive, precise, efficient, implant planning and guided surgery: A narrative review. Clin Oral Implants
safe, and worthy of clinical promotion. Although the Res. 2015;26:69–76.
19. Wang W, Xu H, Mei D, et al. Accuracy of the Yakebot dental implant robotic
preoperative preparation of the robotic implant system system versus fully guided static computer-assisted implant surgery
was tedious, the robotic system will become more so template in edentulous jaw implantation: A preliminary clinical study. Clin
Implant Dent Relat Res 2023.
phisticated, mature, and smaller with improvements in
technology. Corresponding author:
Dr Baodong Zhao
Department of Oral Implantology
The Affiliated Hospital of Qingdao University
PATIENT CONSENT School of Stomatology of Qingdao University
No. 59 Haier Road
Written informed consent was obtained from the patient. Qingdao, Shandong 26600
PR CHINA
Email: [email protected]