LIS Serial Installation
LIS Serial Installation
You are here: Panther System > Panther Procedures > Service Procedures > LIS Serial
Installation
Purpose
This procedure provides instructions for installing the LIS network interface from the Panther
System to a customer's LIS (Laboratory Information System).
Time Required
30 minutes
References
Panther System LIS Host Interface Guide
Pre-Installation Planning
Customer Planning
2. The customer needs to provide an RS232 LIS connection. This can be in one of
the following forms
RS232 Direct
RS232 Converter
Configuration Options
Prior to performing the configuration, determine the following options from the
customer or use the defaults:
Does the customer use uni-directional LIS (only sends results to the LIS system
and does not receive orders for tests from the LIS system)? If so, check Enable
Unidirectional Mode.
Does the customer want the system to automatically generate test order queries
to the LIS? If so, check Enable Test Order Query.
Does the customer want to send results automatically? If so, check Automatic
Result Sending.
Does the customer require result verification? If so, ensure Enable Result
Verification is checked in Admin > System Configuration > Results
Management.
Does the customer only want to verify certain result types? If so, complete the
following steps:
d. Click Add and select the result types the customer would like to hold.
Does the customer want result error flags to be compressed so that multiple
flags are sent on one result comment record? If so, check Enable Result Flag
Sending In a Compressed Format.
Does the customer want to activate LIS logging? If so, turn on both LIS logging
options (LIS logging is turned off by default).
For RS232 Serial connections, determine if the LIS server uses custom or
default values for Baud Rate, Data Bits, Stop Bits, and Parity.
Does the customer want to query the Panther System for results? If so, check
Enable Host Query Responding.
Does the customer want results for calibrators or controls? If so, check Enable
Calibration Result Sending or Enable Control Result Sending.
1. At the back of the PC, connect the RS232 cable to COM Port 1 or Com Port
2.
Do not use COM Port 3—that port is used to connect the system to
the CPU.
2. Connect the other end of the RS232 cable to the matching Com Port on the
customer LIS system.
Serial Settings
for host
transmissions.
17 Baud Rate Select Baud Rate 19200, Specifies the Baud
9600, 4800, 2400, or Rate used for
1200. Default is 9600. transmissions
between the Panther
System and the host.
18 Data Bits 7 or 8. Default is 8. Overwrites the
ASTM1381
requirements.
19 Stop Bits 0 or 1. Default is 1. Overwrites the
ASTM1381
requirements.
5. Launch the Panther System main assay software and log in as an Admin or
Super Operator User.
6. Select Admin.
8. For Serial Configurations, select Serial for the Interface Layer and check the
Enable LIS connection.
9. Select Enable Automatic Test Order Query and Enable Automatic Result
Sending, based upon customer preferences.
10. Check the proper Baud Rate and Data Bits values and enter the proper
setting in the Com Port field.
12. Check the Messages screen to confirm that no LIS Error Messages are
displayed.
15. If the customer requests Result Verification, click the Enable Result
Verification selection.
2. Select Enable ASTM 1381 Logging and Enable ASTM 1394 Logging, and
click Save.
3. If the Panther System is running SW Ver. 5.0 or higher, complete the following
steps:
d. After the verification is complete, disable ASTM 1381 and ASTN 1394
logging.
4. If the Panther System is running SW version lower than 5.0, complete the
following steps:
a. Navigate to Tasks > Perform Maintenance > Service, and ope the
Dashboard.