GE Healthcare
Revolution™ ACT
Product Data Sheet
Released
Revolution ACT has not been submitted to, approved or cleared by the US FDA. Not for sale in the US.
Rendered PDF File Page 3 of 29 DOC1774956, Rev:2
Index
Index
ACT introduction
Revolution ACTTechnologies:
1. Smart introduction
Core Technologies:
a. NEW Clarity panel detector
New
b. Clarity panel detector
Smart Dose:
Advanced Acquisition
Organ Dose Modulation (ODM)
Ultra Kernel
Volumetric Image Space
with sub-mm Reconstruction (ViSR)
imaging
c. Smart
Smart Dose:Flow:
a. ASiR
Pitch booster with IQ enhance
b. Organ
NEW Dose
user interface
Modulation (ODM)
c. Volumetric
Volume helical
Imagedigital
Spacetilt Reconstruction (ViSR)
Smart Flow:
Spectral calibration
2. ACT Hardware:
a. Pitch booster with IQ enhance
a. Gantry Design
b. New user interface
b. Tube
c. Volume helical digital tilt
c. Generator
d. Spectral calibration
d. Patient Table
Revolution ACT Hardware:
e. Host Console
a. Gantry
3. Smart Design
Flow technologies
b. Tube
4. Helical Scanning
c. Generator
5. Axial Scanning
d. Patient
6. Optidose Table
technologies
e. Host Console
7. Software Clinical Applications – Essential
Smart Flow Technologies
8. Software Clinical Applications - Advanced
Helical Scanning
9. Image Quality Specifications
Axial Scanning
10. Installation and service information
OptiDose Technologies
Software Clinical Applications
Software Clinical Applications - Advanced
Image Quality Specifications
Installation and Service Information
Released
Millions of patients around the world are still waiting to receive the standard of care they need
and in an affordable way. Healthcare providers are seeking better technology to help more
patients and drive better outcomes.
Simple and fast, CT is arguably the most valuable diagnostic imaging tool in healthcare. Yet its
capacity to improve patient care is far from tapped.
All revolutions start somewhere. Our revolution began with the Revolution CT system —
designed from the ground up for pioneering the future of CT. Now we introduce
Revolution ACT.
Revolution ACT is designed not only as a product, but as a solution to help you provide better
patient care — and provide it to more patients. It’s the centerpiece of our commitment to help
you achieve your CT services aims.
You’ll find many of the technologies inherited from our most advanced Revolution CT system –
like the new Clarity panel detector, new user interface, or Organ Dose Modulation.
Helping you feel more confident about moving to the next level, clinically and economically —
that’s intelligent innovation.
Revolution ACT: It’s exactly what’s needed to help you achieve success. And exactly what
you’d expect from GE.
Released
CORE TECHNOLOGIES
The new Clarity panel detector is yet another innovation in GE’s long history of advancements in CT
going back to 1976. To clinicians and patients, this technology is the center piece of our commitment
to continuously provide higher image quality. Revolution ACT innovative detector technology is
designed to provide long term protection of your investment while delivering clinical capability and
economic returns today.
Clarity panel detector inherits many elements our premium Revolution CT product including the
integrated detector design and data processing chipsets. The new detector design leverages the
latest advancement in the semi-conductor and electronics industries making it compact, better
integrated and more efficient.
Highly innovative segmented panel designs built with advanced packaging and miniaturization
technology for lower power consumption and improved thermal performance. Since thermal control
is localized on the integrated panel, the detector achieves better precision in performance and
reduces the detector startup time.
The higher density of interconnects and modern A/D chips deliver better* signal linearity with fast
response time. An integrated DAS on Detector or DoD design, the analog signal transmission length
from Detector to DAS is cut to a virtual zero reducing noise by up to 20% and improving image
quality*.
CORE TECHNOLOGIES
Advanced Acquisition*
A higher sampling density provides better representation of the original signal and can potentially
lead to improved image quality and fidelity. The term “conjugate samples” refers to two projection
samples that are 180º apart in their respective projection angle, but at an equal distance to the iso-
center. Taking both sets of projection measurements into consideration, we obtain a significantly
increased sampling density and, therefore, potentially improve the overall acquisition. Using GE’s
novel conjugate cone beam reconstruction algorithm, the conjugated rows of projections are
considered jointly in the back projection step and delivers improved Z-axis visualization performance.
Conjugate cone-beam acquisition enables 32-slice imaging on Revolution ACT.
*Option
Ultra Kernel
CORE TECHNOLOGIES
Noise and artifacts stood in the way of producing high-quality CT images in low-dose exams that
reduce risk to patients. Advanced image reconstruction techniques provide the breakthrough,
enabling high image quality in multi-slice CT exams at significantly less dose than before. ASiR uses
sophisticated statistical modeling to remove noise in images while preserving anatomical detail. This
enables a same- or better-quality image to be produced with lower tube current or tube voltage,
thus lower dose, for all CT applications in which it is used. ASiR is one of the key dose reduction
technologies flowing down from our premium product line. ASiR is the most-used dose-lowering
software in the industry. It allows healthcare providers to lower dose by up to 40% to their patients
as compared to standard image reconstruction without reducing imaging quality. ASiR also improves
low-contrast detectability and can have equivalent IQ to an acquisition with 1.67 times the mA.
Operators can manage longer coverage scans easier. The maximum tube power and current of ASiR
allows achieving the same image quality at a lower mA with less tube heat output, which enables the
tube for longer duration under helical scan. Revolution ACT performance will be equivalent to 40KW
and 333mA (@120kV, 3.3MHu tube heat capacity equivalence with ASiR). This improved efficiency
helps lower power consumption and operational costs to the administrators.
*Option
Released
CORE TECHNOLOGIES
ANRT Is a reconstruction technology used to reduce the noise of diagnostic images and retain
detailed information of the image structure. ANRT noise-reduction technology can, while maintaining
image noise level, lessen the dose of some X-ray testing, or improve image noise level and image
quality. ANRT is only applicable on digital oblique images.
Released
CORE TECHNOLOGIES
An intelligent filter that assesses homogeneity and connectedness of objects within an image is
employed to achieve noise reduction. In anatomical regions such as the brain, this 3D filter reduces
noise without compromising resolution, for clear visualization of brain, tumor, and pediatric cases.
With ViSR the scanner delivers up to 20% image quality improvement at the same dose, or the same
image quality with up to 36% dose reduction1.
1. In clinical practice, the use of VISR may reduce CT patient dose depending on the clinical task, patient size,
anatomical location and clinical practice. A consultation with a radiologist and a physicist should be made to
determine the appropriate dose to obtain diagnostic image quality for the particular clinical task
2. When ASiR is installed, 3D Neuro Filter would be disabled
*Option
Released
CORE TECHNOLOGIES
CORE TECHNOLOGIES
Digital tilt exams are more efficient since operators can setup
the entire workflow from their console without the need for
multiple trips between console and gantry controller.
Spectral calibration calculates the beam hardening vectors based on a theoretical model of
x-ray attenuation through beam-path and the data from scanning a 20 cm water phantom.
Applying the result from spectral calibration in reconstruction, the uniformity of the CT
numbers in image, especially the off-center image, is improved compared with the
conventional beam hardening correction method.
Released
CORE TECHNOLOGIES
Gantry specifications
Aperture 65 cm
Maximum SFOV 43 cm
Focal spot to detector distance 94.9 + 0.1 cm
Tilt + 30 degrees (digital)
Rotation Speed 0.6*, 0.98, 1.0, 1.2, 1.5, 2.0, 3.0, 4.0 seconds
Laser Alignment Internal & external with + 1 mm accuracy
* Partial Scan, 0.6 second equivalent rotation speed in Segmented CINE mode.
The "short geometry design" improves geometry efficiency compared to conventional long
geometry system. For example, Revolution ACT distance from focal spot to detector is
94.9 cm. The geometric efficiency of Revolution ACT is approximately 26% higher than that
of a long geometry scanner. This means that the effective generator power output of
Revolution ACT is equivalent to a 32 kW generator.
Released
CORE TECHNOLOGIES
Tube
Anode Heat Storage 2.0 MHU and 3.3MHU equivalent with ASiR
Anode Heat Dissipation (max) 6200W (500 kHU/min)
0.8 mm (W) x 0.6 mm (L)
Focal Spot
Nominal focal spot size (IEC 60336:2005)
Tube Current (max) 200 mA and 333mA equivalent with ASiR
Power generation
Generator NEW Compact Generator
Scan Watch streamlines patient scan workflow to reduce the patient waiting time. It displays
the tube usage percentage on host UI so that the operator can know the current tube usage
status before setting up a new exam.
Released
CORE TECHNOLOGIES
Patient table
Vertical Scannable Range 700mm- 900mm
Maximum Horizontal Range 1520 mm
Maximum Vertical Range 441mm-900mm
Maximum Scan Range 1350 mm (with extender)
Movement Speed 0.5 mm/sec to 100 mm/sec
Loading Capacity 180 kg (400 lb)
Positioning Accuracy 0.25 mm
Convenient controls on the gantry for cradle movement and patient positioning. Optimized
table dimensions designed to accommodate large variety of patients. Latch free control is
provided and readily accessible on the table for emergency release. Vertical movement of
the table enables optimal patient position and convenient loading and unloading.
Released
CORE TECHNOLOGIES
Imaging system
New Clarity panel detector
Detector
HiLight scintillator with DAS on detector (DoD)
Smart Flow
SmartPrep
SmartPrep allows intermittent monitoring of IV
contrast enhancement in an area of interest. The
contrast flow is monitored by low-dose scans until In the enhanced filming tool, besides the fixed
the contrast enhancement reaches the preferred layout 1:1, 2:1, 4:1, 6:1, 8:1, 9:1, 12:1, 15:1, 8:1, 20:1,
point and then auto trigger function will 24:1, 25:1, 30:1, 35:1 and 42:1, clinicians could also
automatically initiate the scan prescription choose to create your own customized layout to
match your specific needs. And it also allows
editing of the annotations on the image.
Real-time scout
Scout image can be displayed simultaneously with
the acquisition. With the real-time scout image
display, you can decide to stop scout acquisition
once the necessary anatomy is covered.
Connect Pro
Connect Pro assists easy retrieval of patient critical
information from HIS/RIS using DICOM connection
(and then sends this information to patient
schedule). HIS/RIS & PACS requires ConnectPro
Dynamic Transition
With SmartPrep procedure, Dynamic Transition
allows the scan phase to start automatically when
10 PMR
the Hounsfield unit (HU) of the transition region-of- Prospective multiple reconstruction (PMR): Up to 10
interest (ROI) reaches the desired enhancement sets of reconstructions can be pre-programmed as
threshold. part of the scan protocol prior to acquisition. The
operator can select different start/end location,
slice thickness, interval, reconstruction algorithms
Enhanced filming workflow and display fields of view for each reconstruction.
The new filming tool provides you a friendly and
powerful user interface. It allows you to open Up to 10 IPS transfer speed of images real-time
multiple sessions at the same time for one or more during acquisition to up to 4 different destinations
patients; you could set up multiple films in parallel.
Released
Smart Flow
Smart Flow
Full simultaneity allows scan and recon to work All essential dose information is now conveniently
concurrently with image display, processing and and clearly displayed in a dedicated section of the
analysis (including computationally intensive screen.
features such as MPR, MPVR and 3D/MIP) while still
running image archival, filming and networking
processes.
SmartPlan
ImageWorks
ImageWorks software is designed to take
advantage of the Revolution ACT CT scanner’s
After acquiring the scout image, entering to next computer and image processor. This desktop
series (Axial/helical/cine) which SmartPlan is environment includes image management and
enabled in protocol management, SmartPlan is networking.
triggered and modifying the initial localizer for the
Axial series. Because some of the image analysis and display
features of ImageWorks replicate those in Exam
Rx, the next section describes only features that
are incremental or significantly different.
Image access
Point and click interface along with a pictorial
directory (browser) allows for easy selection by
exam, series or image
Smart Flow
Smart Flow
Industry standards
Revolution ACT complies with a wide variety of
industry standards to facilitate more rapid
adoption of features and performance
improvements as the computing and medical
imaging industry evolves.
HIPPA
Password protected user login and authentication
Image anatomization tool
Filming protocol
3M-952 Standard
Table
Helical Scanning
*Option
Released
Axial Scanning
Dose reporting
Neuro DSA
Workflow allows subtraction and visualization of
two image sets such as non-contrast and contrast
enhanced CT.
DentaScan
DentaScan is a fast simple non-invasive software
package, which provides a panoramic dental view
enabling accurate measurements to be taken.
DentaScan facilitates dental prosthetic implants
and other dental surgical procedures by providing
highly detailed information of the teeth and the
surrounding bone structure.
*Option
Released
System LCD is measured with an 8 inch (20 cm) CTDI is measured using a 16 and 32 cm CTDI
CATPHAN phantom. The phantom is widely phantom
accepted as a standard for measuring image
quality performance on multi-slice CT scanners.
Helical scans are tested with 120 kV and 0.938:1
The statistical LCD shall meet the following
minimum specifications: pitch using a standard reconstruction kernel and
35 cm FOV. The system CTDIvol expressed in
5 mm @ 0.30% at 13.3 mGy mGy/100 mAs shall meet the following
5 mm @ 0.30% at 10.0 mGy w/ 40% ASiR* specifications (+15%):
*when ASiR option is enabled 9.44mGy (CTDIvol) for body scan
X/Y – lp/cm
50% > 8.5
10% > 13.0
0% > 18.0
Uniformity
Just as you provide the best care possible to your Relative humidity rate of change:
patients, we’re redesigned Revolution ACT with 5% RH/hour max.
your demanding clinical needs and outstanding
patient care in mind. Power requirements
The only facility input to the system is
200/220/240/380/400/420/440/460/480 V
System Component Dimensions nominal, 3 phase Delta or Wye, 50/60 Hz, 40kVA
service, 15 KVA average power; main disconnect to
Component Size (cm) Weight (kg) be located within 5 feet (1.5 m) of the PDU. The
W H D facility must also provide a protective disconnect
Gantry 178.3 174.1 92.1 870 device with low voltage, low energy local and
Computer 17.8 44.8 44.5 15 multi-point remote capability, in the line feeders to
console the PDU.
Table & cradle 56.8 91.4 213.8 340
PDU 55.0 106.2 70.0 300
Complete, detailed specifications of all power
Total weight 1525
requirements are available upon request. For most
installations, Revolution ACT does not require any
The minimum room size for Revolution ACT would power conditioning equipment to be used in
be as shown in the below figure. conjunction with the PDU. Regulators are not
recommended for use with this system. For those
sites with known large power line transients, a
suppresser filter for the system computer and
peripherals may be useful. In general, suppresser
filters are not recommended.
Cooling requirements
The cooling requirements do not include cooling
for the room lighting, personnel or non-CT
equipment present. Cooling requirements are listed
by subsystem to allow planning for each room of
the CT suite.
Equipment room: If a separate equipment room is GE Healthcare offers you continuous care through
used to house the PDU, the allowable temperature various innovative technologies and education
range is 60°-84°F (15°-29°C) at 30%-60% relative opportunities to ensure that your Revolution ACT
humidity (non-condensing). meets your demanding clinical needs.
Warranty
The published Company warranty in effect on the date
of shipment shall apply. The Company reserves the
right to make changes.
All specifications are subject to change.
Released
About GE Healthcare
Our “healthymagination” vision for the future invites the world to join
us on our journey as we continuously develop innovations focused on
reducing costs, increasing access, and improving quality around the world.
Headquartered in the United Kingdom, GE Healthcare is a unit of General
Electric Company (NYSE: GE). Worldwide, GE Healthcare employees are
committed to serving healthcare professionals and their patients in more
than 100 countries. For more information about GE Healthcare, visit our
website at www.gehealthcare.com.
GE Healthcare
9900 Innovation Drive
Wauwatosa, WI 53226
U.S.A.
www.gehealthcare.com
Released
Revolution ACT has not been submitted to, approved or cleared by the US FDA. Not for sale in the US.