007dw+rapid Eng
007dw+rapid Eng
Rigaku Corporation
Contents
1. Overview ....................................................................................................................................................... 4
2. Configuration ............................................................................................................................................... 5
3.3. Detector.................................................................................................................................................. 9
5. Software ...................................................................................................................................................... 10
7. Heat Loads...................................................................................................................................................11
8.4. Environment......................................................................................................................................... 13
9. Warranty..................................................................................................................................................... 16
i
12. Others.......................................................................................................................................................... 17
ii
Version Approved Checked Created
number Data of revision Revision history by by by
iii
1. Overview
Rigaku proposes to provide a complete single crystal x-ray diffraction system for small molecule data collection and structural
determination. This device can generate high-intensity X-rays (31kW/mm2) by the combination with the confocal mirror., a
Rigaku partial- , 3-axis goniometer, a curved image plate area detector, and all the accessories and peripherals required to fully
integrate the system.
Features
(1) High brightness x-ray generator
The RA-Micro 7HFM tabletop rotating anode X-ray generator achieving 31 kW/mm2 brilliance.
(2) Oscillation photograph method
The blind area can be greatly decreased by measuring the oscillation photograph in the vicinity of =50°.
(3) Large measurement range
460 mm x 256 mm curved imaging plate: In a one image capture the angular range of 2-theta
is -60 to 144 degrees
(tangential direction) providing coverage of 204 degrees.
(4) High speed data reading
Installing IP is one piece. To shorten duty time, the high-speed reading method is adopted. ( 1.5 times compared with
conventional ratio)
(5) It corresponds to three readout modes
It corresponds to three modes of 100×100 μm, 100×150 μm and 200×200 μm. As for 200×200 μm mode,
high-speed readout is possible (26sec/ 1frame). A small sample of the lattice can greatly shorten the measurement
time by using 200×200 μm mode.
(6) Large dynamic range and low level noise.
Device is compliant with a linear dynamic range of 1 x 106. A dual PMT system gives the Rigaku R-AXIS
RAPID II the largest dynamic range available of any detector. A noise level of <1 X-ray photon/ADU with a
sensitivity of 1 X-ray photon/pixel (CuK ) ensures that reflections will be accurately measured and weak reflections
will have the best possible S/N.
(7) Automatic measurement and data analysis
All processes from the measurement to data processing were automated.
(8) Two absorption correction methods
The absorption correction by empirical method or numerical method is possible.
(9) The target corresponding dual wavelength
The target material can be easily changed without leak of the vacuum.
(10) The mirror corresponding dual wavelength
The mirror can be easily switched by using the rotational mechanism.
4
2. Configuration
2.3. Software
2.3.1. Instrument control and data analysis software CrystalClear 1
2.3.2. Structure solution software CrystalStructure(Optional) 1
5
2.4. Standard accessories
2.4.1. Grease 1
2.4.2. Bomb regulator 1
2.4.3. Gas regulator (He gas) 1
6
2.7. Option
2.7.1. Cooling water system 1
2.7.2. Low temperature system using liquid nitrogen 1
3. Hardware specification
3.1. -ray generating unit
3.1.1. -ray generator
(1) Maximum output rating 1.2 kW
(2) Tube voltage setting range 20 60 kV in 1 kV steps
(3) Tube current setting range 10 30 mA in 1 mA steps
It does not become 10mA or more from 20 to 39kV the tube current
(4) Stability 0.01% against 10% mains fluctuation
(5) X-ray take-off direction Point focus X-ray take-off in two horizontal directions (right port only)
(6) X-ray shutter Rotary shutter
(7) Safety features Abnormal cooling water flow and pressure detection
Abnormal generator overload detection
Abnormal tube voltage detection
Broken filament and overload detection
TMP error detection
Vacuum error detection
Target error detection
High voltage unit error detection
Emergency stop switch
Leak current breaker
Fail-safe mechanism Optional
(8) Alarm display
3.1.2. Target
Mo/Cu, Cu/Cr (Selectable)
7
3.2. Goniometer
3.2.1. Confocal Max Flux mirror VariMax DW(Mo/Cu), VariMax DW(Cu/Cr)
System Dual Wavelength Optic incorporates two VMHF optics with
different working wavelengths.
Correspond to Dual wavelength It is possible to switch easily.
Adjustment The position of each optic is defined by four adjustable screws in
regard to the rotatable shell.
Safety mechanism Four screws have the release function.
Substitution of gas He gas
Mirror size More then 150mm long.
Divergence angle Variable slit
0.2deg (High resolution measurement) 0.6deg (High intensity
measurement)
3.2.2. He gas More than 99.995 purity, 7m3bomb It is not attached.
3.2.3. Goniometer (Including goniometer head and collimator)
(Chi)axis -16 deg ~ + 55 deg, 0.0001deg/ step
(Phi)axis 0 deg ~ +360 deg, 0.0001deg/ step
(Omega)axis -85 deg ~ + 265 deg, 0.0001deg/ step
Sphere of confusion Less than 20 m
Sample to detector distance Fixed at 127.4 mm
Diffraction angle coverage 2 = - 60 deg ~+144deg (equator line direction), 45 deg (meridional direction)
Collimator 1.0 mm single pinhole collimators
0.1 mm single pinhole collimators (Optional)
Goniometer head IUCr 49 mm type (with manual XYZ mechanism)
(range of movement X and Y = ±1 mm, Z = ±2 mm
Eucentric goniometer head (Optional)
Focusing mechanism (Optional)
8
3.3. Detector
3.3.1. Image plate and IP reader
Detector : Imaging Plate
Readout method : Laser scanning readout
Readout source : Semiconductor laser (Maximum rated output 30mW
Image plate
1) Detector area : 470 mm x 256 mm curved image plate
2) Pixel size : 100×100 μm, 100×150 μm, 200×200 μm,
3) Pixel count : 4600×2560, 4600×1700, 2300×1280,
Sensitivity : 1 X-ray photon/pixel
Dynamic range : 1.05 x 106
Readout time : 3000 lines / minute
51 seconds / image at 100 x 100 μm
26 seconds / image at 200 x 200 μm
Erasing time : 20 seconds
Duty time : 90 seconds
3.6.2. Interface to PC
Control RS-232C
Data transfer PCI bus
9
4. Low temperature system (Optional)
Air- or water-cooled type : -180 ˚C ~ RT (Low temperature system using liquid nitrogen)
* Please refer to the specifications for the low temperature device for details.
5. Software
5.1. Data collection software: Rapid Auto
Display features
10
TwinSolve features
Twin indexing
Twin relationships
Twin integration
6. Standard accessories
Grease 1
7. Heat Loads
11
8. Installation requirements
8.1. Dimensions and weight
Main unit
Dimension (mm) Weight
Component (kg)
Width Depth Height
8.2.2. Grounding
Grounding D 100
Connecting more than two instruments to ground together can induce noise and malfunctioning of the
instruments. They should be grounded separately
12
8.3. Cooling water
Please note that there is a possibility that the tube and CCD may corrode when any additives are used.
The use of tap water may result in algae growth and occlusion within the X-ray tube.
To avoid the need to change the cooling water frequently, it is recommended to use distilled water instead of tap water.
8.4. Environment
13
8.4.2. Magnetic field
(1) Do not place any magnetic objects which may affect the CRT image of the computer close to the system.
(2) Make sure there is no leakage of magnetic flux (force) near the X-ray tube.
8.4.3. Atmosphere
(1) The following contaminants in the air may damage the instrument.
a) Flammable and corrosive gases
b) Dust
c) Salt
(2) Avoid direct sunlight
(3) No other heat source next to the system
8.4.5. Humidity
8.4.6. Vibration
No physical vibration sensed by human body
8.4.7. Others
Other environmental requirements such as dust and corrosive gas are pursuant to those for conventional
electronic devices.
14
8.5.3. Layout
A 1.0-m distance for maintenance purposes is required around the system for the RA-Micro7HF DW + VariMax DW
with RAPID.
4 8
No. Item Port Cable length Leading length Power line Recital
3-phase
1 Power cable (Main body) Round crimp contact R5.5-5 6m 4m Included
200V 20 A
Power cable 3-phase
2 Round crimp contact R22-8 6m 4m Included
(XG) 200V 15A
Grounding cable
3 Round crimp contact R22-6 6m 4m --- Included
(XG)
Cooling hose
4 --- --- --- --- Included
(Main body)
Power cable
5 Depends on specifications of cooling water system.
(Cooling water system)
Signal cable
6 Depends on specifications of cooling water system
(Cooling water system)
Power cable 1-phase
7 3P plug 7m 4m Included
(CPU) 100V 15A
Cooling hose
8 --- 5m 4m Included
(XG to target)
1-phase
9 Target controller 3P plug 5m 2m Included
100V 15A
15
Dimensions
1070
600
1800
850
The dimensions do not include outstretching items such as the main power connector and the X-ray on lamp.
The controller for the goniometer and X-ray generator are located in the main body.
9. Warranty
The warranty period will be for one year after the acceptance date.
However, the warranty does not cover accidents that are caused by the following factors:
(1) Failures caused by fires, natural disasters, or acts of God.
(2) Use other than defined in the agreement (use for unintended purposes).
(3) Modifications made without the consent of Rigaku.
(4) Operating conditions that are significantly different from the specifications.
(5) Serious mistakes committed by the user or third party (including failures caused by an operating error) no matter
whether they are accidental or intentional.
(6) Consumable items or parts that have reached the end of their useful life.
16
11. Acceptance
The system shall be accepted when the contents of the inspection report, which is in accordance with the supplier’s
inspection standard, are reproduced after installation of the system is completed.
12. Others
The specification and external appearance are subject to change without notice for the purpose of improvement.
17