AsteionVP_Outline, 1 September 2005

AsteionVP
- Value Package -
Outline

© Toshiba Medical Systems Corporation 2005 1


Contents

Purpose of development
Scope of new development
Comparison with AsteionVF
Options

© Toshiba Medical Systems Corporation 2005 2


Purpose of Development
Purpose
Keep single slice system available - PC based instead of O2
Offer up-to-date specifications

Target markets
Main markets: Japan and China
Other areas as well - except USA and Canada

Features
Optimized operability
Low dose technologies
Optimized workflow with fast reconstruction and 3D/MPR generation
Real time technology

Available as from July 2005

© Toshiba Medical Systems Corporation 2005 3


Scope of new development

LCD monitor

Operation panels

Table drive motor

Console
© Toshiba Medical Systems Corporation 2005 4
Scope of new development
System New colours
Improved EMC

Gantry operation panels


Style of Aquilion32/64

Couch Classic design


Table slide motor/servo as Super4 and other systems

Console Reconstruction in PC - costs reduction


PC as with Super4 - XEON processors - Windows 2000
SSCONT and ADI boards
Data acquisition firmware and drivers
LCD monitor

© Toshiba Medical Systems Corporation 2005 5


Software
Based on Aquilion 16 PC and Asteion Super4

Recon Scan Display subsystem 3D/MPR


(DSS)
New
Modified 0% 80% 100% 100%
Same as As/S4
Driver Common middleware IAP Engine
(FSYS)
0% 100% 100%

Scan Pre- Exceed 100%


control processing
MKS(Nutcracker) 100%
firmware firmware
Windows 100%
95% 98% PC(Intel system CPU) 100%

© Toshiba Medical Systems Corporation 2005 6


Compared
AsteionVP AsteionVF
Power 50 kVA 50 kVA
Full rotation scan 1s (0.75s) 1s (0.75s)
Reconstruction time 1.2s 3s (1s)
Slice width (0.8) 1,2,3,5,7,10 (0.8) 1,2,3,5,7,10
Field of view 480mm 480mm
X-ray tube 2 MHU (3.5 MHU) 2 MHU (3.5 MHU)
Generator 24 kW (36 kW) 24 kW (36 kW)
3D Standard option
Real time technology Real time helical 1fps Real time helical 1fps
Flash start 1fps Flash start 1fps
Real EC Real EC
(Sure start)
(Fluoro)

© Toshiba Medical Systems Corporation 2005 7


Compared
AsteionVP AsteionVF
Monitor LCD 48cm CRT
Others Quantum Denoising filter
T mode filming
Guided mode in 9 languages
Console
Components 1 navibox - 1 computer 1 navibox - 1 computer
Size 340x620x665 480x750x700
Approximate weight 80kg 120kg
Power 200V 1.9A 200V 3.9A

© Toshiba Medical Systems Corporation 2005 8


Compared
AsteionVP AsteionVF
CPU XEONx2 3.06GHz O2 R5K 300MHz
Memory 3GB 256MB
System disk 80GB 9GB
Image disk 36GB 4GB
Image data 32000 images 8000 images
Raw data disk 36GB 4GB
Raw data 3000 rotations 1500 rotations
1s helical scan
MOD Option Standard
CD-RW Standard DICOM

Arm down holder Option Standard


Triangular mat Standard Option
Arm up holder Standard Standard

© Toshiba Medical Systems Corporation 2005 9


Options
Sub-second scan kit - 0.75s full rotation scan - requires 3.5 MHU tube
key required - load at Tube & Generator - disables 2MHU selection

3.5 MHU tube - suitable for 0.75s scan unlike CXB-200B

Flash start - SureStart with one frame per second


key required

36 kW - generator is 42 kW Spellman
key required - load at Tube & Generator selection

0.8 mm slice - standard 1mm - S and SS - key required

MOD - for archiving - also for site back-up - key to be typed

© Toshiba Medical Systems Corporation 2005 10


Manuals
New manuals
Site planning
Operation card
Disposal
Maintenance - Asteion Super4 concept
Service manuals for hardware - two volumes

© Toshiba Medical Systems Corporation 2005 11


Site Planning - Installation
• Small navibox
• Some extra space required behind gantry - for service and for headrest
At least 1150 between gantry and wall
• Epoxy is STRONGLY recommended

• Load options at much as possible during system configuration


0.75s - 0.8mm, etc. have implications - e.g. for calibration

• Allow three hours between power ON and calibration


Calibrate at normal OLP
Calibrate NRA - all FOVs one by one
Calibrate Air - all FOV together

© Toshiba Medical Systems Corporation 2005 12