0% found this document useful (0 votes)
28 views

MenAMI Codes-II CC 03

Uploaded by

ali.yildiirm
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
28 views

MenAMI Codes-II CC 03

Uploaded by

ali.yildiirm
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 116

AMI Codes-II CC

Version 6.00 and higher

Operator’s Manual

A-96.250.581 / 050517
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
E-mail: [email protected]

Document Status

Title: Monitor AMI Codes-II CC Operator’s Manual

ID: A-96.250.581

Revision Issue

00 April 2010

01 June 2010

02 Aug. 2013 Update to Rev. 5.40, main board V2.4

03 June 2016 Update to Rev. 6.00, main board V2.5

© 2016, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved

subject to change without notice


AMI Codes-II CC

Table of Contents

1. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1. Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2. General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3. Restrictions for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2. Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1. Instrument Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.2. Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.1. Installation Check List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2. Mounting of Instrument Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.3. Connecting Sample and Waste . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.3.1 FEP Tube at Sample Inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.3.2 FEP Tube at Sample Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.4. Installation of Flow Cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.5. Install the Option pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.5.1 pH as Option ex works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.5.2 pH Option as Retrofit Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6. Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.6.1 Connection Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6.2 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.7. Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.8. Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.8.1 Alarm Relay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.8.2 Relay 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.9. Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.9.1 Signal Output 1 and 2 (current outputs) . . . . . . . . . . . . . . . . . . 34
3.10 Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.10.1 Signal Output 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.10.2 Profibus, Modbus Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.10.3 USB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.1. Prepare Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.2. Peristaltic Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.3. Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.4. Fill or Flush Reagent System . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.5. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.6. Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

A-96.250.581 / 050517 1
AMI Codes-II CC

5. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.1. Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.2. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.3. Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.4. Changing Parameters and values. . . . . . . . . . . . . . . . . . . . . . . . . 44
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.1. Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.2. Stop of Operation for Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 46
6.3. Refill or replace Reagents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.3.1 Reagents for measuring Free Chlorine and Total Chlorine . . . . 49
6.4. Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.5. Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
6.6. Cleaning the protective Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.7. Cleaning the Photometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.8. Cleaning the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.8.1 Disassemble the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
6.8.2 Assemble the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
6.9. Maintenance of pH sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
6.10. Tube Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6.10.1 Replace the Pump Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6.10.2 Replace the Reagent Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
6.11. Cleaning the solenoid valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.12. Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
6.13. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
7. Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
7.1. General Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
7.2. Calibration Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.2.1 Process calibration tc or fc . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.2.2 Process pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.2.3 Standard pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.3. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
8.1. Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
8.2. Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
8.3. Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
8.4. Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
8.5. Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
9. Program List and Explanations . . . . . . . . . . . . . . . . . . . . . . . . . 81
1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81

2 A-96.250.581 / 050517
AMI Codes-II CC

2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
4 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
10. Material Safety Data sheets . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
10.1. Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
11. Default Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
12. Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
13. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113

A-96.250.581 / 050517 3
AMI Codes-II CC
Safety Instructions

AMI Codes-II CC - Operator’s Manual


This document describes the main steps for instrument setup, oper-
ation and maintenance.

1. Safety Instructions
General The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work environ-
ment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Target Operator: Qualified person who uses the equipment
audience for its intended purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
OM Location The AMI Operator’s Manual shall be kept in proximity of the instru-
ment.
Qualification, To be qualified for instrument installation and operation, you must:
Training  read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
 know the relevant safety rules and regulations.

4 A-96.250.581 / 050517
AMI Codes-II CC
Safety Instructions

1.1. Warning Notices


The symbols used for safety-related notices have the following sig-
nificance:

DANGER

Your life or physical wellbeing are in serious danger if such


warnings are ignored.
 Follow the prevention instructions carefully.

WARNING

Severe injuries or damage to the equipment can occur if such


warnings are ignored.
 Follow the prevention instructions carefully.

CAUTION

Damage to the equipment, minor injury, malfunctions or incor-


rect process can be the consequence if such warnings are ig-
nored.
 Follow the prevention instructions carefully.

Mandatory The importance of the mandatory signs in this manual.


Signs
Safety goggles

Safety gloves

A-96.250.581 / 050517 5
AMI Codes-II CC
Safety Instructions

Warning Signs The importance of the warning signs in this manual.

Electrical shock hazard

Corrosive

Harmful to health

Flammable

Warning general

Attention general

6 A-96.250.581 / 050517
AMI Codes-II CC
Safety Instructions

1.2. General Safety Regulations


Legal The user is responsible for proper system operation.
Requirements All precautions must be followed to ensure safe operation
of the instrument.
Spare Parts Use only official SWAN spare parts and disposables. If other parts
and are used during the normal warranty period, the manufacturer’s
Disposables warranty is voided.
Modifications Modifications and instrument upgrades shall only be carried out by
an authorized Service Technician. SWAN will not accept responsi-
bility for any claim resulting from unauthorized modification or alter-
ation.
WARNING

Risk of Electrical Shock


If proper operation is no longer possible, the instrument must be
disconnected from all power lines, and measures must be taken
to prevent inadvertent operation.
 To prevent from electrical shock, always make sure that the
ground wire is connected.
 Service shall be performed by authorized personnel only.
 Whenever electronic service is required, disconnect instru-
ment power and power of devices connected to.
– relay 1,
– relay 2,
– alarm relay

WARNING

For safe instrument installation and operation you must read


and understand the instructions in this manual.

WARNING

Only SWAN trained and authorized personnel shall perform the


tasks described in this document.

A-96.250.581 / 050517 7
AMI Codes-II CC
Safety Instructions

1.3. Restrictions for use


The sample must not contain any particles, which may block the
flow cell. Sufficient sample flow is coercive for the correct function
of the instrument.
If the sample contains only little disinfectant concentrations, or
there is the danger of biological growth, we recommend to use the
optional Cleaning module from Swan.

WARNING

Health hazard
Some reagents are etching and can cause severe burns or eye
damage.
 For safe handling of the reagents you must read and under-
stand the instructions in this manual, as well as the Material
Safety Data Sheets (MSDS)

Download The current Material Safety Data Sheets (MSDS) for the below list-
MSDS ed Reagents are available for downloading at www.swan.ch.
 OXYCON ON-LINE DPD
 OXYCON ON-LINE Buffer
 OXYCON ON-LINE KI
 Buffer solution pH 4
 Buffer solution pH 7
 Buffer solution pH 9

8 A-96.250.581 / 050517
AMI Codes-II CC
Product Description

2. Product Description
Application The AMI Codes-II CC is a complete monitoring system for the auto-
Range matic, continuous measurement and dosing control of chlorine
based on the DPD colorimetric method APHA 4500 Cl-G and on
EN ISO 7393-2.
It can be used for measuring disinfectants in:
 Pools and sanitary water
 Cooling water
 Waste water effluent
It is also applicable for water containing additives like corrosion in-
hibitors, cyanuric acid and antiscaleants.
Determines free available chlorine, total chlorine and calculates
monochloramine, dichloramine and combined chlorine.
Necessary reagents:
 Oxycon on-line DPD
 Oxycon on-line Buffer
 Oxycon on-line KI.

Disinfectant
Measurement Disinfectant Measuring range Accuracy

Free available chlorine, 0.00–1.00 ppm ± 0.01 ppm


Total chlorine 1.00–3.00 ppm ± 0.06 ppm
3.00–5.00 ppm ± 0.20 ppm

Signal Two signal outputs programmable for measured values (freely


Outputs scaleable, linear or bilinear) or as continuous control output (control
parameters programmable).
Current loop: 0/4–20 mA
Maximal burden: 510 
Third signal output available as an option. The third signal output
can be operated as a current source or as a current sink (selectable
via switch).
Relay Two potential-free contacts programmable as limit switches for
measuring values, controllers or timer for system cleaning with au-
tomatic hold function. Both contacts can be used as normally open
or normally closed.
Maximum load: 1 A/250 VAC

A-96.250.581 / 050517 9
AMI Codes-II CC
Product Description

Alarm Relay One potential free contact.


Alternatively:
 Open during normal operation, closed on error and loss of
power.
 Closed during normal operation, open on error and loss of
power.
Summary alarm indication for programmable alarm values and in-
strument faults.
Input For potential-free contact to freeze the measuring value or to inter-
rupt control in automated installations (hold function or remote-off)
Special Possibility to interrupt the measurement by activating the input. See
Feature Program List and Explanations, 5.3.4, p. 103.
Safety No data loss after power failure. All data is saved in non-volatile
Features memory. Over voltage protection of in- and outputs.Galvanic sepa-
ration of measuring inputs and signal outputs.The analyzer is facto-
ry tested and ready for installation and operation.
Communica-  Third signal output (can be used in parallel to the USB interface)
tion Interface  USB Interface for logger download
(optional)  RS485 with Fieldbus protocol Modbus or Profibus DP
pH Optional pH measurement is possible (pH correction or calibration).
Relay Box The AMI Relay Box is designed for the direct power supply and ac-
tivation of dosing devices which are controlled with an AMI trans-
mitter, e. g. to connect two solenoid valves or one motor valve for
disinfectant additions.
Cleaning An optional Cleaning module is available.
Module

10 A-96.250.581 / 050517
AMI Codes-II CC
Product Description

Glossary Used abbreviations to the chlorine forms measured:

Abbr. Term Comment

fac Free available chlorine Immediate reaction with DPD


(includes cyanuric acid fractions)

tc1 Total chlorine 1 Immediate reaction of DPD + KJ


(mainly monochloramine)

tc2 Total chlorine 2 Reaction with DPD + KJ after 2


min. reaction

cmc Calculated monochloramine cmc = tc1 - fac

cdc Calculated dichloramine cdc = tc2 - tc1

ccc Calculated combined chlorine ccc = tc2 - fac

pH/T pH and Temperature (Optional)

B/s Bubbles per second Sample flow


[ppm]
Value

@ 2 min
tc2
cdc

ccc

tc1 +KI
cmc

fac DPD+Buffer

Time

0 sec 3-5 sec 2 min

A-96.250.581 / 050517 11
AMI Codes-II CC
Product Description

Fluidics The sample flows through the sample inlet [R] and the inlet filter [H]
into the constant head [A]. Adjust the flow regulating valve [F] so
that always a small part of the sample flows through the overflow
tube [B] into the constant head drain [Q].
Free available A part of the sample flows through the photometer inlet [C] into the
chlorine mixing chamber [E], where the reagents [M] and [N] are added by
the peristaltic pump [D] and mixed with the sample. The mixed
sample flows through the photometer [K] where the free available
chlorine is measured.

Level

C
D E F

I J G

K
L
M N O

R
P Q

A Constant head J Solenoid valve V 1


B Overflow tube K Photometer
C Photometer inlet L Air bubble detector
D Peristaltic pump M Reagent Oxycon on-line DPD
E Mixing chamber N Reagent Oxycon on-line Buffer
F Flow regulating valve O Reagent Oxycon on-line KI
G Solenoid valve V2 P Photometer drain
H Inlet Filter Q Constant head drain
I Photometer air inlet R Sample inlet

12 A-96.250.581 / 050517
AMI Codes-II CC
Product Description

Total chlorine 1 After the measurement of free available chlorine is finished the so-
lenoid valve [G] is energized for a short time and reagent [O] is
added to determine total chlorine 1.
Total chlorine 2 To ensure the necessary reaction time of 2 min. for the determina-
tion of total chlorine 2, the sample flow is stopped by the solenoid
valve [J]. After the time has elapsed, total chlorine is measured and
solenoid valve [J] opens again.

Level

C
D E F

I J G

K
L
M N O

R
P Q

After the measurement the sample flows through the outlet of the
photometer where it will be aerated through air inlet [I] to generate
bubbles. Then the sample flows through the bubble detector [L] into
the photometer drain [P].

A-96.250.581 / 050517 13
AMI Codes-II CC
Product Description

General It is distinguished between two different kinds of measurement:


 Free available chlorine (fac) measurement
 Overall measurement including free available chlorine (fac),
total chlorine 1 (tc1) and total chlorine 2 (tc2).
The Interval fac and the Interval tc2 can be set individually:
 Interval fac between 0 and 12 min.
 Interval tc2 between 0 and 60 min (overall measurement).

NOTICE: If the fac interval is set to zero, it is nevertheless


measured in the overall measurement.
The default setting for fac measurement interval is 5 min and for
overall measurement interval 20 min.
Time The blue bar represents the sample flow through the photometer.
sequence of an At the beginning the solenoid valve [J] is de-energized and the
overall sample flows through the photometer. A short time before the mea-
measurement surement starts, a zero point measurement is performed. Then the
peristaltic pump starts and a small portion of the reagents “Oxycon
on-line DPD” and “OXYCON ON-LINE Buffer” is pumped into the
mixing chamber. During this time, the solenoid valve [G] is de-ener-
gized and the “Oxycon On-line KI” circulates between the peristaltic
pump and the valve block.
A short time after, if the mixture is in the photometer, the free avail-
able chlorine (fac) of the sample is measured.
After about 25 sec the peristaltic pump starts again. Now the sole-
noid valve [G] is energized and a small portion of the reagents
“Oxycon on-line DPD”, “OXYCON ON-LINE Buffer” and “Oxycon
On-line KI” is pumped into the mixing chamber.
A short time after, the solenoid valve [J] is energized and the sam-
ple flow through the potometer is stopped. Immediately after that,
the tc 1 is measured.
The “calculated monochloramine” (cmc) is calculated and dis-
played.
After about 120 sec the tc 2 is measured and then the “calculated
dichloramine” (cdc) and the “calculated combined chlorine” (ccc)
are calculated and displayed.
After the measurement is finished, the solenoid valve [J] is de-ener-
gized and sample flows through the photometer and a flush time of
2 minutes follows.

14 A-96.250.581 / 050517
AMI Codes-II CC
Product Description

G G

fac J tc 1 tc 2 J fac
D

min
1 2 3 4 5 6 7 8
A B E

A Reaction time for measuring tc 2 tc1 Total chlorine 1


B Flush time tc2 Total chlorine 2
C Overall measurement time OXYCON ON-LINE DPD
D Flush time end OXYCON ON-LINE Buffer
E fac measurement OXYCON ON-LINE KI
G Valve 1 for adding OXYCON ON-LINE KI Zero point measurement
J Valve 2 for closing photometer inlet Sample measurement
fac Free available chlorine Peristaltic pump active for a short
time

The programmed measuring intervals of tc2 [F] and fac [G] may
overlap each other.
The tc2 measurement has priority and therefore always takes place
in the programed measuring interval [F]. The fac measurement is
filled in the gaps [H] of the overall measurement and may be shifted
(orange arrow) in respect to programmed interval. If 2 fac measure-
ments are programmed during a overall measurement one of them
is cancelled (red arrow).

fac fac fac


tc2

min
0 10 20 30 40 50 60
H
G
F

A-96.250.581 / 050517 15
AMI Codes-II CC
Product Description

2.1. Instrument Specification

Power Supply Voltage: 100–240 VAC (± 10%)


50/60 Hz (± 5%)
or 24 VDC (± 10%)
Power consumption: max. 30 VA
Electronics Aluminium with a protection degree of IP 66/NEMA 4X
housing Ambient temperature: -10 to +50 °C
Limit range of operation: -25 to +65 °C
Storage and transport: -30 to +85 °C
Humidity: 10–90% rel., non condensing
Display: backlit LCD, 75 x 45 mm
Sample Flow rate: min. 10 l/h
requirements Temperature: 5–50 °C (41–122 °F)
Inlet pressure: 0.15–2 bar (2–28 PSI)
Outlet pressure: pressure free
NOTICE: No oil, no grease, no sand.

On-site The analyzer site must permit connections to:


requirements Sample inlet: Tube 6 x 8 mm
Sample outlet: 1/2” hose nozzle for flexible tube
diam. 20x15 mm
Measuring Free chlorine, monochloramine, total residual chlorine
Range 0.00–1.00 ppm ± 0.01 ppm
1.00–3.00 ppm ± 0.06 ppm
3.00–5.00 ppm ± 0.2 ppm
pH measurement (optional)
2–12 pH 0.01 pH

16 A-96.250.581 / 050517
AMI Codes-II CC
Product Description

Dimensions Panel: 400x850x200 mm


Mounting hole distance 374x824
Screws: 5 mm or 6 mm diameter
Weight: 12.0 kg / 26.5 lbs without reagents and
sample water
17.0 kg / 37.5 lbs with reagents and sam-
ple water

400 mm / 15¾”
13 mm / ½”
374 mm / 14¾”

13 mm / ½”
6 x dia. 6.5 mm / ¼”

AMI Codes II CC

824 mm / 32 7/16”
412 mm / 16 ½ ”

850 mm / 33½”

30 mm / 1 3 /16”

A-96.250.581 / 050517 17
AMI Codes-II CC
Product Description

2.2. Instrument Overview

A
G
B H

J
K
C
L
M
N
O

D
Q
E
R
F

A Panel J Flow regulating valve


B Transmitter K Sample inlet
C Peristaltic pump L Valve 1 (Photometer)
D Reagent Oxycon on-line DPD M Valve 2 (KI dosing)
E Reagent Oxycon on-line Buffer N Inlet filter
F Reagent Oxycon on-line KI O Photometer
G Temperature sensor P Air bubble detector
H pH sensor Q Constant head drain
I Constant head R Photometer drain

18 A-96.250.581 / 050517
AMI Codes-II CC
Installation

3. Installation

3.1. Installation Check List

Check Instrument’s specification must conform to the National Electrical


Code, all state and local codes, and all plant codes and standards
for electrical equipment.
On-site require- 100–240 VAC ( 10%), 50/60 Hz ( 5%) or 24 VDC (±10%),
ments isolated power outlet with ground connection and 30 VA.
Sample line with min. 10 l/h and 0.15–2 bar.
Waste line with pressure free drain.
Installation Mount the instrument in vertical position.
Display should be at eye-level.
Mount the filter, filter vessel, and constant head cover. Connect the
sample and waste line. See Connecting Sample and Waste, p. 21
Electrical Wiring Do not switch on the Instrument until all electrical connections are
made.
Connect all external devices like limit switches, current loops and
pumps.
Connect power cord.
See Electrical Connections, p. 27
If ordered: See Install the Option pH, p. 23
Option pH
Reagents Prepare reagents. See Refill or replace Reagents, p. 47.
Insert the suction lances.
Power-up Perform exactly in this order:
 Lock pump tubes.
 Turn on the sample flow and wait until the flow cell is completely
filled.
 Switch on power.
Start <Fill system>. See Fill or Flush Reagent System, p. 39
Instrument Program all parameters for external devices (interface, recorders,
Setup etc.). Program all parameters for instrument operation (limits,
alarms, measuring interval).
Run-in period Let the instrument run continuously for 1 h.
pH sensor If ordered: Calibrate pH sensor. See Calibration, p. 52
calibration

A-96.250.581 / 050517 19
AMI Codes-II CC
Installation

Process Make 3 manual measurements. Use a high quality photometer,


calibration e.g. Chematest from Swan. Calculate average value and compare
this value to the value, indicated by the AMI. If necessary, correct
the value. The zero point is done automatically before each mea-
surement.

3.2. Mounting of Instrument Panel


The first part of this chapter describes the preparing and placing of
the system for use.
 The instrument must only be installed by trained personnel.
 Mount the instrument in vertical position.
 For ease of operation mount it so that the display is at eye
level.
 For the installation a kit containing the following installation
material is available:
– 6 Screws 6x60 mm
– 6 Dowels
– 6 Washers 6.4/12 mm
Mounting The instrument is only intended for indoor installation.
requirements For dimensions see Dimensions, p. 17.

20 A-96.250.581 / 050517
AMI Codes-II CC
Installation

3.3. Connecting Sample and Waste

3.3.1 FEP Tube at Sample Inlet


Use plastic tube (FEP, PA, or PE 6 x 8 mm) to connect the sample
line.

Mounting of A Screw connection


A
SERTO fitting B Compression ferrule
B C Knurled nut
D Flexible tube
C

3.3.2 FEP Tube at Sample Outlet

WARNING

Risk of water pollution


The drain of the photometer outlet contains DPD.
 At no means recirculate it into the water system.

A Tube from photometer


B Drain Photometer
A C C Tube from constant head
D Drain constant head
B D E Hose nozzles
F 1/2” tubes
E
F

Connect the 1/2” tubes [F] to the hose nozzles [E] and place it into
a pressure free drain with sufficient capacity.

A-96.250.581 / 050517 21
AMI Codes-II CC
Installation

3.4. Installation of Flow Cell

CAUTION

Fragile Part
Handle the constant head tube with care.

To avoid damage during the transport, the constant head tube [C] of
the AMI Codes-II CC is not installed.
A Constant head cover
A
B Overflow tube
C Constant head tube
D Gasket
B E Flow cell block
Level

To install the constant head tube proceed as follows:

1 Unpack the constant head tube [C].


2 Push the constant head tube into the flow cell block [E].
3 Put the constant head cover [A] onto the constant head tube.
4 Check if the overflow tube [B] is aligned with the upper Level
mark.

22 A-96.250.581 / 050517
AMI Codes-II CC
Installation

3.5. Install the Option pH

3.5.1 pH as Option ex works


If the pH option was ordered with the AMI Codes-II CC, the pH sen-
sor cable as well as the temperature sensor are already connected
to the AMI transmitter.
A A Connector cap
B Connector
B C pH Sensor
D Temperature sensor
E Constant head cover
F Protective cap
C G Constant head

D
E
F

1 Carefully pull off the protective cap [F] from the pH sensor [C]
by turning it clockwise.
2 Store the protective cap in safe place.
3 Rinse the pH sensor tip with clean water.
4 Insert the pH sensor into one of the holes in the constant head
cover [E].
5 Insert temperature sensor [D] into the small hole.
6 Remove the connector cap from the connector of the pH sen-
sor. Store it in safe place.
7 Screw the connector [B] onto the pH sensor.

A-96.250.581 / 050517 23
AMI Codes-II CC
Installation

3.5.2 pH Option as Retrofit Kit

A B C

F E D

A 2 Clamps with screws D Front end PCB


B pH sensor E Temperature sensor
C Sensor cable F Short overflow tube

24 A-96.250.581 / 050517
AMI Codes-II CC
Installation

WARNING

Risk of electrical shock.


Do not perform any work on electrical components if the trans-
mitter is switched on. Failure to follow safety instructions could
result in serious injury or death.
 Always turn off AC power before manipulating electric parts.

1 Screw the clamps for calibration solution onto the panel. Use
the already drilled holes [A].

AMI Codes-II

2 Stop sample flow. Wait until flow cell is empty.


3 Switch off the AMI transmitter (disconnect power).
A A Constant head cover
B Constant head tube
C Short overflow tube
B D Flow cell block
Level

4 Remove the constant head cover [A] and the constant head
tube [B] from the flow cell block [D].
5 Insert the short overflow tube [C] into the outlet hole leading to
the photometer.
6 Install the constant head tube and the constant head cover.

A-96.250.581 / 050517 25
AMI Codes-II CC
Installation

A Front end PCB


B pH sensor plug
C Temperature sensor plug

A B C

7 Open the cover of the AMI transmitter housing.


8 Install the front end PCB.
9 Feed the cable of the pH sensor through one of the cable
glands (see Cable thicknesses, p. 27) into the AMI transmitter
housing.
10 Connect it to the BNC socket.
11 Feed the cable of the Temperature sensor through one of the
cable glands into the AMI transmitter housing.
12 Connect the temperature sensor cable to the plug as follows:
 Terminal 19: line
 Terminal 20: shield.
13 Close the cover of the AMI transmitter housing.
14 Install the pH sensor, see pH as Option ex works, p. 23.
15 Turn on sample flow and wait until flow cell has been filled com-
pletely.
16 Switch power ON. The instrument automatically detects the
front end PCB during start-up.

26 A-96.250.581 / 050517
AMI Codes-II CC
Installation

3.6. Electrical Connections

WARNING

Risk of electrical shock.


Do not perform any work on electrical components if the trans-
mitter is switched on. Failure to follow safety instructions could
result in serious injury or death.
 Always turn off AC power before manipulating electric parts.
 Grounding requirements: Only operate the instrument from
an power outlet which has a ground connection.
 Make sure the power specification of the instrument corre-
sponds to the power on site.
Cable In order to comply with IP66, use the following cable thicknesses
thicknesses

A B C

A PG 11 cable gland: cable Øouter 5–10 mm


B PG 7 cable gland: cable Øouter 3–6.5 mm
C PG 9 cable gland: cable Øouter 4–8 mm

NOTICE: Protect unused cable glands


Wire  For Power and Relays: Use max. 1.5 mm2 / AWG 14
stranded wire with end sleeves.
 For Signal Outputs and Input: Use 0.25 mm2 / AWG 23
stranded wire with end sleeves.

A-96.250.581 / 050517 27
AMI Codes-II CC
Installation

WARNING

External Voltage.
External supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks
 Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in-
stallation.
– relay 1
– relay 2
– alarm relay

WARNING

To prevent from electrical shock, do not connect the instrument


to the power unless the ground wire (PE) is connected.
 Do not connect unless specifically instructed to do so.

WARNING

The mains of the AMI Transmitter must be secured by a main


switch and appropriate fuse or circuit breaker.

28 A-96.250.581 / 050517
AMI Codes-II CC
Installation

3.6.1 Connection Diagram

CAUTION

Use only the terminals shown in this diagram, and only for the
mentioned purpose. Use of any other terminals will cause short
circuits with possible corresponding consequences to material
and personnel.

A-96.250.581 / 050517 29
AMI Codes-II CC
Installation

3.6.2 Power Supply

WARNING
Risk of electrical shock
Do not perform any work on electrical components if the trans-
mitter is switched on. Failure to follow safety instructions could
result in serious injury or death.
 Always turn off AC power before manipulating electric parts.
 Installation and maintenance of electrical parts must be per-
formed by professionals.

A
B
C

A Power supply connector


B Neutral conductor, Terminal 2
C Phase conductor, Terminal 1
D Protective earth PE

NOTICE: The protective earth wire (Ground) has to be


connected to the grounding terminal.
Installation The installation must meet the following requirements.
requirements  Fuse 1.6 AT
 Mains cable to comply with standards IEC 60227 or IEC
60245; flammable rating FV1
 Mains equipped with an external switch or circuit-breaker
– near the instrument
– easily accessible to the operator
– marked as interrupter for AMI Codes-II CC

30 A-96.250.581 / 050517
AMI Codes-II CC
Installation

3.7. Input
NOTICE: Use only potential-free (dry) contacts.
The total resistance (sum of cable resistance and resistance of
the relay contact) must be less than 50 Ω.
Terminals 30 and 31
For programming see chap. 9, menu Installation, 5.3.4, p. 103.

3.8. Relay Contacts


3.8.1 Alarm Relay
NOTICE: Max. load1 A / 250 VAC
Alarm output for system errors.
Error codes see Troubleshooting, p. 68.

NOTICE: With certain alarms and certain settings of the AMI


transmitter the alarm relay does not switch. The error, however,
is shown on the display.

Terminals Description Relay connection


NC1) 10/11 Active (opened) during normal 1) 11
Normally operation.
0V
Closed Inactive (closed) on error and 10
loss of power.
12

NO 12/11 Active (closed) during normal 11


Normally operation. 0V
Open Inactive (opened) on error and 10
loss of power.
12

1) usual use

A-96.250.581 / 050517 31
AMI Codes-II CC
Installation

3.8.2 Relay 1 and 2


NOTICE: Max. load 1 A/250 VAC
Relay 1 and 2 can be configured as normally open or as normally
closed. Standard for both relays is normally open. To configure a
Relay as normally closed, set the jumper in the upper position.

NOTICE: Some error codes and the instrument status may


influence the status of the relays described below.

Relay Jumper
config. Terminals pos. Description Relay configuration
Normally 6/7: Relay 1 Inactive (opened) during
Open 8/9: Relay 2 normal operation and 6
loss of power.
Active (closed) when a 0V
programmed function is 7
executed.
Normally 6/7: Relay 1 Inactive (closed) during
Closed 8/9: Relay 2 normal operation and 6
loss of power.
Active (opened) when a 0V
programmed function is 7
executed.

A
B

A Jumper set as normally open (standard setting)


B Jumper set as normally closed
For programming see menu Installation 5.3.2 and 5.3.3, p. 98.

32 A-96.250.581 / 050517
AMI Codes-II CC
Installation

CAUTION

Risk of damage of the relays in the AMI Transmitter due to


heavy inductive load.
Heavy inductive or directly controlled loads (solenoid valves,
dosing pumps) may destroy the relay contacts.
 To switch inductive loads > 0.1 A use an AMI relay box avail-
able as an option or suitable external power relays.

Inductive load Small inductive loads (max 0.1A) as for example the coil of a power
relay can be switched directly. To avoid noise voltage in the
AMI Transmitter it is mandatory to connect a snubber circuit in par-
allel to the load.
A snubber circuit is not necessary if an AMI relaybox is used.
A A AC or DC power supply
B C B AMI Transmitter
C External power relay
D E D Snubber
E Power relay coil

Resistive load Resistive loads (max. 1A) and control signals for PLC, impulse
pumps and so on can be connected without further measures
A B A AMI Transmitter
B PLC or controlled pulse pump
C C Logic

Actuators Actuators, like motor valves, are using both relays: One relay con-
tact is used for opening, the other for closing the valve, i.e. with the
2 relay contacts available, only one motor valve can be controlled.
Motors with loads bigger than 0.1A must be controlled via external
power relays or an AMI relay box.
A A AC or DC power supply
B C B AMI Transmitter
C Actuator

A-96.250.581 / 050517 33
AMI Codes-II CC
Installation

3.9. Signal Outputs

3.9.1 Signal Output 1 and 2 (current outputs)


NOTICE: Max. burden 510 Ω
If signals are sent to two different receivers, use signal isolator
(loop isolator).
Signal output 1: Terminals 14 (+) and 13 (-)
Signal output 2: Terminals 15 (+) and 13 (-)
For programming see Program Overview, p. 74, Menu Installation

3.10 Interface Options


A AMI Transmitter
A B Slot for interfaces
C Frontend PCB
D Screw terminals
B

The slot for interfaces can be used to expand the functionality of


the AMI instrument with either:
 Third signal output
 a Profibus or Modbus connection
 an USB Interface

34 A-96.250.581 / 050517
AMI Codes-II CC
Installation

3.10.1 Signal Output 3


Terminals 38 (+) and 37 (-).
Requires the additional board for the third signal output 0/4–20 mA.
The third signal output can be operated as a current source or as a
current sink (switchable via switch [A]). For detailed information see
the corresponding installation instruction.
NOTICE: Max. burden 510 Ω.

A
Third signal output 0/4 - 20 mA PCB
A Operating mode selector switch

3.10.2 Profibus, Modbus Interface


Terminal 37 PB, Terminal 38 PA
To connect several instruments by means of a network or to config-
ure a PROFIBUS DP connection, consult the PROFIBUS manual.
Use appropriate network cable.
NOTICE: The switch must be ON, if only one instrument is
installed, or on the last instrument in the bus.
OFF

ON

Profibus, Modbus Interface PCB (RS 485)


A On - OFF switch

A-96.250.581 / 050517 35
AMI Codes-II CC
Installation

3.10.3 USB Interface


The USB Interface is used to store Logger data and for Firmware
upload. For detailed information see the corresponding installation
instruction.
The optional third signal output 0/4 – 20 mA PCB [B] can be
plugged onto the USB interface and used in parallel.

USB Interface
A USB interface PCB
B Third signal output 0/4 - 20 mA PCB

36 A-96.250.581 / 050517
AMI Codes-II CC
Instrument Setup

4. Instrument Setup
After installation according to checklist proceed as following:

4.1. Prepare Reagents


1 Prepare reagents. See Refill or replace Reagents, p. 47.
2 Insert the suction lances into the containers.

4.2. Peristaltic Pump


The instrument is delivered with opened occlusion frames.

1 Activate the peristaltic pump tubes by closing the occlusion


frame [B].

A Turn to lock
B Occlusion frame
A
C Rotor
D Pump tube
B

C
D

A-96.250.581 / 050517 37
AMI Codes-II CC
Instrument Setup

4.3. Establish Sample Flow

WARNING

Water pollution
The drain of the photometer outlet contains DPD.
 At no means recirculate it into the water system.

A A Cover
B Constant head tube
B C Flow cell block
D Flow regulating valve
Level
E Filter
F Filter vessel

C
D

1 Open the flow regulating valve (D) and wait until the flow cell is
completely filled.
2 Switch on power.
3 Adjust the sample flow so that always a small part of the sample
drains off through the overflow tube.
4 Start <Fill system>, see Fill or Flush Reagent System, p. 39.

38 A-96.250.581 / 050517
AMI Codes-II CC
Instrument Setup

4.4. Fill or Flush Reagent System


Fill or flush the reagent tubing:
 upon the initial instrument setup,
 after refilling the reagent containers,
 before a system shut-down to flush the system with deminer-
alized water until no more reagent is left in the system.

Service 3.2.2
Navigate to menu <Maintenance /
Verification Service/Fill system>.
Fill System Press [Enter].
Cleaning

Fill System 3.2.2.5


The peristaltic pump is activated for
1.5 minutes.
Progress

<Enter> to stop

Fill System 3.2.2.5


Press [Exit] 4 x to go back to the oper-
ating mode.
Progress

Done

1 Check tubing and flow cell for leaks and repair if necessary.
2 Let the instrument run continuously for 1 hour.

4.5. Programming
Programming Program all parameters for external devices (interface, recorders,
etc.)
Program all parameters for instrument operation (disinfectant, lim-
its, alarms).
Program the DPD value of the Verikit in menu <Installation/
Sensors/Ref. Verification>.
If pH option is installed, program the two buffers you want to use for
calibration in menu <Installation/Sensors/Standards>.
See Program List and Explanations, p. 81.

A-96.250.581 / 050517 39
AMI Codes-II CC
Instrument Setup

4.6. Calibration
1 Calibrate pH sensor (if option pH is installed).
See Standard pH, p. 54.
Perform process calibration.
See Process Calibration of tc & fc, p. 52
If ordered: The instrument should be operating for 1h before performing a pH
Calibration of calibration.
pH sensor Program the two buffers you want to use for calibration in menu
<Installation/Sensors/Standards>. Calibrate the pH sensor with
two buffers, e.g. pH 7.00 and pH 9.00. See chapter Calibration, p.
52 for details.

40 A-96.250.581 / 050517
AMI Codes-II CC
Operation

5. Operation

5.1. Keys

A B C D

Exit Enter

A to exit a menu or command (rejecting any changes)


to move back to the previous menu level
B to move DOWN in a menu list and to decrease digits
C to move UP in a menu list and to increase digits
to switch between display 1 and 2
D to open a selected sub-menu
to accept an entry

Program
RUN 14:10:45 Main Menu 1
Access, Exit Enter
R1
0.22 ppm Messages
fac
Diagnostics
tc2 0.26 ppm
R2
Maintenance
ccc 0.04 ppm Operation
Exit
35 B/s 25.4°C Installation

A-96.250.581 / 050517 41
AMI Codes-II CC
Operation

5.2. Display
A B C D E

RUN 12:56:02

R1 fac 0.22 ppm


12:55:35
F
12:52:40

R2
tc2 0.26 ppm G
12:52:40
ccc 0.04 ppm
L 47 B/s 26.8 °C H
K I

A RUN normal operation


HOLD input closed or cal delay: Instrument on hold (shows
status of signal outputs).
OFF input closed: control/limit is interrupted (shows status
of signal outputs).
B ERROR Error Fatal Error

C Reagent low, indicates remaining reagents in % (17% = 340 ml)


D Keys locked, transmitter control via Profibus
E Time
F Time stamp (time of the last valid measuring value)
G Process values
H Sample temperature (only with option pH)
I Cleaning solution low, indicates remaining cleaning solution in %
K Sample flow
L Relay status
Relay status, symbols
upper/lower limit not yet reached
upper/lower limit reached
control upw./downw. no action
control upw./downw. active, dark bar indicates control intensity

motor valve closed


motor valve: open, dark bar indicates approx. position
timer
timer: timing active (hand rotating)

42 A-96.250.581 / 050517
AMI Codes-II CC
Operation

5.3. Software Structure

Main Menu 1
Messages
Diagnostics
Maintenance
Operation
Installation

Menu Messages 1
Messages 1.1
Reagent Status Reveals pending errors as well as an event history
Pending Errors (time and state of events that have occurred at an
Maintenance List earlier point of time).
Message List It contains user relevant data.

Menu Diagnostics 2
Diagnostics 2.1
Identification Provides user relevant instrument and sample data.
Sensors
Sample
I/O State
Interface

Menu Maintenance 3
Maintenance 3.1
Calibration For instrument calibration, relay and signal output
Service simulation, and to set the instrument time.
Simulation It is used by the service personnel.
Set Time 23.09.06 16:30:00

Menu Operation 4
Operation 4.1
Sensors User relevant parameters that might need to be
Relay Contacts modified during daily routine. Normally password
Logger protected and used by the process-operator.
Subset of menu 5 - Installation, but process-related.

Menu Installation 5
Installation 5.1
Sensors For initial instrument set up by SWAN authorized
Signal Outputs person, to set all instrument parameters. Can be
Relay Contacts protected by means of password.
Miscellaneous
Interface

A-96.250.581 / 050517 43
AMI Codes-II CC
Operation

5.4. Changing Parameters and values


Changing The following example shows how to change the logger interval:
parameters 1 Select the parameter you want to
Logger
Sensors 5.1.2
4.4.1
change.
Sensor
Log type
interval FOME
30 min
Disinf.logger Free chlorine 2 Press [Enter]
Clear no
Temperature NT5K
Standards
3 Press [ ] or [ ] key to
Logger 4.1.3
highlight the required parameter.
Log interval Interval. 30min
5 min
Clear logger no 4 Press [Enter] to confirm the selec-
10 min tion or [Exit] to keep the previous
30 min parameter).
1 Hour

 The selected parameter is


Logger 4.1.3
highlighted but not saved yet.
Log interval 10 min
Clear logger no 5 Press [Exit].

 Yes is highlighted.
Logger 4.1.3
Log interval
Save ?
6 Press [Enter] to save the new pa-
Clear logger no rameter.
Yes
No  The system reboots, the new
parameter is set.

Changing 1 Select the value you want to


values Alarm Conductivity 5.3.1.1.1
change.
Alarm High 5.00 ppm
Alarm Low 0.00 ppm 2 Press [Enter].
Hysteresis 1.00 ppm 3 Set required value with [ ] or
Delay 5 Sec
[ ] key.
4 Press [Enter] to confirm the new
Alarm Conductivity 5.3.1.1.1
value.
Alarm High 0.20 ppm
Alarm Low 0.00 ppm 5 Press [Exit].
Hysteresis 1.00 ppm Yes is highlighted.
Delay 5 Sec
6 Press [Enter] to save the new val-
ue.

44 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

6. Maintenance

6.1. Maintenance Schedule

WARNING

Stop operation before maintenance.


 Stop sample flow.
 Shut off power of the instrument.

Daily (dirty water) Check sample supply for dirt.


up to every Clean all filters and strainers, if necessary.
2 weeks (clean Clean AMI Codes protection filter, if necessary.
water) Check sample flow (see also Troubleshooting, p. 68)

Every 4–6 weeks Clean reagent containers and prepare new reagents.
Let instrument run for 1 h. Make 3 manual measure-
ments. Compare average value to displayed value. If nec-
essary, perform process calibration.

Monthly Recommendation: Check photometer with verification kit


Verification, p. 51

Yearly Exchange reagent pump tubes.

By occurrence: E020, FOME dirty: Cleaning the Photometer, p. 56


E022, Reagent empty: Refill or replace Reagents, p. 47
E065, Reagents low: Refill or replace Reagents, p. 47

If option pH is installed

Weekly Perform a process calibration, see Process pH, p. 53.

Every 2 months Perform a standard calibration, see Standard pH, p. 54.

A-96.250.581 / 050517 45
AMI Codes-II CC
Maintenance

6.2. Stop of Operation for Maintenance


1 Put the suction lances into a bucket with clean water.
2 Start <Fill system>.
 The reagent tubes are flushed with water.
3 Remove the suction lances from the water.
4 Start <Fill system> again.
 The water will be pumped out of the reagent tubes.
5 Stop sample flow.
6 Wait until the flow cell is empty.
7 Shut off power of the instrument.

If pH option is installed:
8 Remove the pH sensor [A] from the flow cell.
9 Fill the protective cap [B] with water.
10 Put the protective cap onto the sensor tip.

A pH sensor
B Protective cap

46 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

6.3. Refill or replace Reagents


The liquid level in the containers 2 and 3 is monitored.
The following messages are displayed:
Container almost Maintenance E065 - Reagents low and the
empty remaining reagent volume in %
(starting at 17 % = 340 ml).
Container empty Error E022 - Reagent empty

NOTICE: Before refilling the reagents, rinse the containers with


demineralized water.
Reagent The 2 liter reagent container for DPD & Buffer will last for 19 days of
consumption operation (with measurement interval of 4 minutes). The 2 liter re-
agent container for KI will last for 66 days of operation (with mea-
surement interval of 20 minutes). The provided reagent set (for 8
containers) therefore lasts for 5 respectively 17 months of opera-
tion.
As the reagent consumption is not linear find further examples be-
low:

Measuring interval Duration of DPD/Buffer Duration of KI

one canister reagent set one canister reagent set


fc tc filling filling

2 minutes 10 minutes ~ 14 days 3 months ~ 33 days 8 months

4 minutes 20 minutes ~ 19 days 5 months ~ 66 days 17 months

WARNING

Health hazard
Some reagents can cause severe burns or eye damage.
 For safe handling of the reagents you must read and under-
stand the Material Safety Data Sheets (MSDS), see Material
Safety Data sheets, p. 106

A-96.250.581 / 050517 47
AMI Codes-II CC
Maintenance

Oxycon On-line DPD

WARNING

Severe eye irritation and severe skin irritation.


Concentrated Oxycon On-line DPD contains more than 10%
mineral acids.
 Do not swallow.
 Avoid any contact with eyes and skin.
 Wear protective goggles,
 Wear protective gloves.
 IF IN EYES: Rinse cautiously with water for several minutes.
Remove contact lenses, if present and easy to do. Continue
rinsing.
 Immediately call a POISON CENTER or doctor/ physician.

Oxycon On-line Buffer


Contains citric acid potassium salt, do not swallow.

Oxycon On-line KI

CAUTION

Severe eye irritation and skin irritation.


This product is corrosive and causes severe burns.
 Harmful if swallowed.
 Avoid any contact with eyes and skin
 IF IN EYES: Rinse cautiously with water for several minutes.
Remove contact lenses, if present and easy to do. Continue
rinsing.

48 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

Canister set up
A A Suction lance without
level detector
B (canister 1)
B Suction lance with
level detector
(canisters 2 and 3)
C Level detector
D 2 L mark
C E Canister 1: Oxycon
on-line DPD
F Canister 2: Oxycon
D on-line Buffer
2L 2L 2L
G Canister 3: Oxycon
E on-line KI
OXYCON ON-LINE
DPD-Reagenz
DPD Réactif
OXYCON ON-LINE
Pufferlösung
Tampon
OXYCON ON-LINE
KI-Lösung
KI-Solution
H Holder
DPD Reagent Buffer Solution

F
G

6.3.1 Reagents for measuring Free Chlorine and Total Chlorine


To prepare 2 L reagents the following chemicals are necessary:
 1 x 50 ml of Oxycon On-line DPD
 1 x 240 gr of Oxycon On-line Buffer
 1 x 60 gr of Reagent Oxycon On-line KI
 6 liters demineralized water

Prepare 1 Rinse the canister [E] labelled “OXYCON ON LINE DPD-Re-


Oxycon agent” with demineralized water.
On-line DPD 2 Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3 Slowly pour the content of one bottle of concentrate Oxycon
On-line DPD (50 ml) into the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.

A-96.250.581 / 050517 49
AMI Codes-II CC
Maintenance

6 Put the canister [E] into the holder [H].


7 Remove the screw cover and insert the suction lance [A] and
tighten the screw cover.

Prepare 1 Rinse the canister [F] labelled “OXYCON ON LINE Buffer” solu-
Oxycon tion with demineralized water.
On-line Buffer 2 Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3 Slowly pour the content of one bag of buffer Oxycon On-line
Buffer into the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
6 Put canister [F] into holder [H].
7 Remove the screw cover, insert the suction lance [B] and tight-
en the screw cover.

Prepare 1 Rinse the canister [G] labelled “OXYCON ON LINE KI” solution
Oxycon with demineralized water.
On-line KI 2 Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3 Slowly pour the content of one bag of Oxycon On-line KI into
the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
6 Put canister [G] into holder [H].
7 Remove the screw cover, insert suction lance [B] and tighten
the screw cover.
8 Fill reagent system. See Fill or Flush Reagent System, p. 39

50 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

6.4. Verification
The “Verification kit for AMI Photometer” is available as
an accessory. An optical window with a precisely deter-
mined absorbance value is placed into the light beam
of the photometer. The actual measured absorbance
will be compared to the reference value labeled on
each kit.

Set reference Prior to performing the verification the DPD reference value, e.g.
value: 0.255, needs to be set in menu 5.1.4
<Installation>\<Sensors>\<Ref. Verification>.
Verification Basically follow the dialog in menu 3.2.1
procedure: <Maintenance>\<Service>\<Verification>.

NOTICE: Start any time, if a measuring cycle is in progress wait


for next prompt.

1 Stop sample flow by closing regulating valve. Wait for next


prompt: Constant head will be drained and an automatic zero
will be defined.
2 Open cuvette of the photometer and insert the verification filter.
[Enter] to continue.
3 Align the triangle shape either to the front or back side and ad-
just for minimal absorbance (see AMI Display).
4 Press [Enter] to save the verification measurement. The verifi-
cation is successful if the difference is within the limits. [Enter] to
continue.
5 Remove filter, close cuvette and open regulating valve. [Enter]
to finish and [Exit] to the main display.
Verification Can be reviewed in menu 2.2.1.5
history: <Diagnostics>\<Sensors>\<Photometer>\<Ver. History>

A-96.250.581 / 050517 51
AMI Codes-II CC
Maintenance

6.5. Calibration
Process Let the instrument run for 1 h.
Calibration
of tc & fc NOTICE: Perform process calibration for free chlorine or total
residual chlorine only if:
• the sample concentration is close to the desired process
value (stable value).
• you are sure that the reagents are mixed completely and
correctly.
• if the difference to the manual measurement is significant.
• Keep in mind the accuracy of your manual measurement.
Use CHEMATEST 25 (or equivalent photometer) to determine the
sample disinfectant concentration. Determine the sample disinfec-
tant value by 3 manual DPD measurements. Take the sample direct
form the constant head. Calculate the average value. Compare this
value to the value, indicated by the AMI transmitter. Keep in mind
the accuracy of your manual measurement. Only correct the instru-
ment if the difference is significant.

Calibration 3.1.1 Free av. chlorine 3.1.1.4


Free av. chlorine Current Value x.xx ppm
Total chlorine 2 Enter Slope x.xxx
Process pH
Standard pH Process Value x.xx ppm
Save <Enter>

Free av. chlorine 3.1.1.4


Current Value x.xx ppm
Enter the correct value Slope x.xxx
with the [ ] or [ ]
key. Process Value x.xx ppm
Save <Enter>

Free av. chlorine 3.1.1.5


Current Value x.xx ppm
Slope x.xxx

Calibration Successful
Press 3 x [Exit]

Possible error messages see Calibration Errors, p. 69.

52 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

Zero A zero is automatically done before each measurement.

Process pH Use a Chematest 25 photometer (or equivalent) to determine the


sample pH value. Insert the electrode through a hole in the con-
stant head cover into the constant head.

NOTICE: Make sure your reference instrument is calibrated


correctly!

Maintenance 3.1 Calibration 3.1.2


Calibration Free av. chlorine
Service Enter Total chlorine 2
Simulation Process pH
Set Time 01.01.05 16:30:00 Standard pH
Cleaning

Process pH 3.1.2.4

Current Value 7.78 pH


Offset x mV

Process Value 7.78 pH


Save <Enter>

Enter the correct value Process pH 3.1.2.4


with the [ ] or [ ] Current Value 7.78 pH
key. Offset x mV

Process Value 7.70 pH


Save <Enter>

Process pH 3.1.2.5 Process pH 3.1.2.5


Current Value 7.70 pH Current Value 7.70 pH
Offset y mV Enter Offset y mV
Slope x.xx mV
Process Value 7.70 pH
Calibration successful Save <Enter>

 Possible error messages see Calibration Errors, p. 69.

A-96.250.581 / 050517 53
AMI Codes-II CC
Maintenance

Standard pH 1 Navigate to menu <Maintenance>/


Maintenance 3.1
<Calibration>.
Calibration Enter
Simulation 2 Press [Enter].
Set Time 01.06.04 16:30:00 3 Remove the pH sensor from the
Enter
Fill System flow cell.
Cleaning
4 Follow the instructions on the dis-
Calibration 3.1.3 play.
Free av. chlorine
Total chlorine 2
Process pH
Standard pH

Display 1 Rinse and dry the pH sensor and put it into standard 1
instructions 2 Standard 1, current value (Progress is shown).
3 Rinse and dry the pH sensor and put it in standard 2
4 Standard 2, current value (Progress is shown)
5 Rinse and dry the pH sensor and put it into the flow cell

 Possible error messages see Calibration Errors, p. 69.

54 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

6.6. Cleaning the protective Filter


Switch off the instrument according to instructions in Stop of Opera-
tion for Maintenance, p. 46.

A A Flow cell block


B Flow regulating valve
B
C Filter shaft
D Filter
E Filter vessel
C

Normally the filter in your sample supply line will retain most debris.
If the filter shows deposits, proceed as follows:

1 Close the main tap of the sample inlet.


2 Close flow regulating valve [B].
3 Unscrew and remove the filter vessel [E] from the
flow cell block [A].
4 Hold the filter [D] on the shaft [C] and unscrew and remove it.
5 Backwash the filter under pressure of tap water.
6 Clean the outside of the filter.
7 Install the filter and the filter vessel again.
8 Establish the sample flow.
9 Adjust sample flow with the regulating valve.

A-96.250.581 / 050517 55
AMI Codes-II CC
Maintenance

6.7. Cleaning the Photometer


Clean the photometer after indication by alarm (E020, FOME dirty).
Switch off the instrument according to instructions in Stop of Opera-
tion for Maintenance, p. 46.
Material Small brush.
Procedure A Flow regulating valve
B Photometer cover
A
B C Photometer
D Brush

1 Close the flow regulating valve [A].


2 Wait until the sample flow through the photometer has stopped.
3 Unscrew the cover [B] from the photometer [C].

4 Clean the Photometer with a small brush [D].


5 Screw the cover to the photometer.
6 Open the flow regulating valve.

Clean the photometer after indication by alarm (E020, FOME dirty).

56 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

6.8. Cleaning the Flow Cell

CAUTION

Acrylic glass parts are fragile and scratch-sensitive.


Possible damage of acrylic glass parts due to scrubbing materi-
als.
 Never use organic solvents or scrubbing materials to clean
acrylic glass parts.
 Use soft detergent and rinse well. Eliminate lime deposits with
a common household deliming agent in standard concentra-
tion.
 Do not drop the constant head tube.

A-96.250.581 / 050517 57
AMI Codes-II CC
Maintenance

6.8.1 Disassemble the Flow Cell


The flow cell can be disassembled easily. Before disassembling the
flow cell, switch off the instrument according to instructions in Stop
of Operation for Maintenance, p. 46.

A A Constant head cover


B Overflow tube
C Constant head tube
B D Flow cell block
E Flow regulating valve
Level

D
E

Cleaning 1 Switch off the instrument according to instructions in Stop of


Operation for Maintenance, p. 46.
2 Remove the constant head cover [A].
3 Remove the constant head tube [C] from the flow cell block.
4 Pull the overflow tube [B] out of the flow cell block [D].
5 Clean all acrylic parts with a soft brush (bottle cleaner) and
soapy water.
6 Remove lime deposits with a common household deliming
agent with standard concentrations.
7 Rinse the parts carefully with clean water.

58 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

6.8.2 Assemble the Flow Cell


A A Constant head cover
B Overflow tube
C Constant head tube
B D Gasket
E Flow cell block
Level

1 Replace the gasket [D] before reassembling the flow cell.

NOTICE: A film of teflon paste (e.g. Fomblin from Solvay


Solexis) on the gaskets improves tightness and life time.
2 Push the overflow tube [B] through the flow cell block as far as it
reaches the drain.
3 Install the constant head tube [C] onto the flow cell block.
4 Put the cover onto the constant head.
5 Align the overflow tube with the upper level mark.

A-96.250.581 / 050517 59
AMI Codes-II CC
Maintenance

6.9. Maintenance of pH sensor


A Connector
B pH sensor shaft
C Flow cell cover
A D Flow cell

Clean 1 Remove the pH sensor [B] from the flow cell.


pH sensor 2 Unscrew and remove the connector [A] from the pH sensor.
Prevent the connectors from getting wet
3 If necessary wipe the pH sensor shaft and the green tip cau-
tiously with a soft, clean, and damp paper tissue.
4 Remove grease with a tissue moistened with alcohol.
CAUTION: Alcohol is inflammable.
5 If the sensor is very dirty put it into 1% diluted hydrochloric acid
for 1 min.
CAUTION! hydrochloric acid is corrosive!
6 Rinse the pH sensor with clean water.

60 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

6.10. Tube Replacement

6.10.1 Replace the Pump Tubes


The pump tube [D] of the peristaltic pump is exposed to a minimal
wear.
It is therefore recommended to exchange the pump tube annually.

CAUTION

Pollution of reagents possible.


If the occlusion frames are opened during operation, already
mixed reagents will flow back into the reagent canisters and pol-
lute the reagents.
 Never open the occlusion frames if the instrument is in opera-
tion.
 Proceed according to Stop of Operation for Maintenance, p.
46 before opening the occlusion frames.

Overview A Pump housing


B Occlusion frame
A
closed
B C Rotor
D Pump tube
C
E Pump inlet
D F Pump outlet

A-96.250.581 / 050517 61
AMI Codes-II CC
Maintenance

Dismount The pump tube can easily be dismounted and mounted.


pump tubes Proceed as follows:
A Pump housing
B Occlusion frame open
A
C Rotor
B D Pump tube
C E Pump inlet
D F Pump outlet
E
F

1 Switch off the instrument according to instructions in Stop of


Operation for Maintenance, p. 46
2 Open the occlusion frame (B) by turning it counter-clockwise.
3 Remove the occlusion frame (B) from the rotor (C) by pulling
the complete occlusion frame out of the holder.
4 Dismount all reagent tubes from the occlusion frame.
5 Exchange the occlusion frame.
6 Install all reagent tubes.
7 Lock the occlusion frames.
8 Start fill system.

62 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

6.10.2 Replace the Reagent Tubes


Tube
numbering
Q Level
1 2

3 1

4 2

6 7
6 6
7 7
5 5

3 4 5

M N O

Nr. from to
1 Pump outlet rear frame Flow cell block, connection 1
see Flow cell block side view Q
2 Pump outlet middle frame Flow cell block, connection 2
see Flow cell block side view Q
3 Reagent container (M) Pump inlet rear frame
Oxycon on-line DPD
4 Reagent container (N) Pump inlet middle frame
Oxycon on-line Buffer
5 Reagent container (O) Valve block connection 5
Oxycon on-line KI
6 Valve block connection 6 Pump inlet front frame 6
7 Pump outlet front frame Valve block connection 7

A-96.250.581 / 050517 63
AMI Codes-II CC
Maintenance

6.11. Cleaning the solenoid valve


Disassemble The solenoid valves are mounted at the bottom of the flow cell
the solenoid block. The solenoid valve should be disassembled if it does not
valve switch anymore or if it is clogged.
A Valve 1 switches the sample flow
through the photometer on and off.
B Valve 2 is used for adding the
Reagent Oxycon on-line KI.

A
B

1 Switch off the instrument according to instructions in Stop of


Operation for Maintenance, p. 46
2 Loosen the nut (A).

3 Remove the solenoid coil (B) from


the valve body (C).
B

64 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

4 Loosen the fixing screws of the


valve body with a 2.5 mm Allen key
(D).

NOTICE: The O-rings inside the


valve body may stick on the flow
cell and fall down if the valve body
is removed.
5 Remove the valve body from the
flow cell.
E

6 Remove the base plate (G) with a


screw driver size 0 (F).

The membrane (H) is now


visible.
7 Clean base plate (G) and mem-
H brane (H) with clean water.
8 If necessary replace the mem-
brane.

Assemble Assemble the solenoid valve in reverse order.

A-96.250.581 / 050517 65
AMI Codes-II CC
Maintenance

6.12. Replacing Fuses

WARNING

External Voltage.
External supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks.
 Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in-
stallation.
– relay 1
– relay 2
– alarm relay

When a fuse has blown, find out the cause and fix it before
replacing it with a new one.
Use tweezers or needle-nosed pliers to remove the defective fuse.
Use original fuses provided by SWAN only.
A B C D E F G

A 1.6 AT/250V Instrument power supply


B 1.0 AT/250V Relay 1
C 1.0 AT/250V Relay 2
D 1.0 AT/250V Alarm relay
E 1.0 AF/125V Signal output 2
F 1.0 AF/125V Signal output 1
G 1.0 AF/125V Signal output 3

66 A-96.250.581 / 050517
AMI Codes-II CC
Maintenance

6.13. Longer Stop of Operation


1 Put the suction lances into a bucket with clean water.
2 Start <Fill system>.
 The reagent tubes are flushed with water.
3 Remove the suction lances from the water.
4 Start <Fill system> again.
 The water will be pumped out of the reagent tubes.
5 Stop sample flow.
6 Wait until level in flow cell has fallen to the shorter tube inside
the cell.
7 Shut off power of the instrument.
8 Empty the flow cell completely.
9 Open the occlusion frames of the peristaltic pump, see Replace
the Pump Tubes, p. 61.

If option pH is installed
10 Unscrew and remove the connector from the pH sensor.
11 Put the connector cap onto the sensor connector.
12 Fill 3.5 molar KCl (if not available: water) into the rubber cap.
13 Remove the pH sensor from the flow cell and place the rubber
cap on the tip of the sensor

CAUTION

Damage of pH sensor
Wrong storage will damage the pH sensor.
 Never store the pH sensor dry.
 Store the pH sensor with tip pointing downwards in a frost-
protected room.

A-96.250.581 / 050517 67
AMI Codes-II CC
Troubleshooting

7. Troubleshooting
This chapter provides some hints to make troubleshooting easier.
For any detailed information how to handle or clean parts please
see Maintenance, p. 45. For any detailed information how to pro-
gram the instrument please see Program List and Explanations, p.
81.

7.1. General Instructions


NOTICE: The sample for the manual measurement (with DPD)
must be taken directly from the flow cell.
If you need further help please contact your dealer. Note serial
number of instrument and all diagnostic values before doing so.
Diagnostic Zero photometry: 10’000–16’000 Hz (mostly near 16 000 Hz)
values Slope photometry: 0.8–1.2
pH offset: New pH sensor: near 0, old pH sensor >  50 mV
pH slope: typically: 55–62 mV/pH unit.
Frequently
Problem Possible Reasons
asked
questions Unstable values  Sample taken too close to feeding line
 Sample flow too irregular or too low
Codes display  Wrong manual measurement or old
higher or lower than chemicals have been used. Repeat the
manual measure- verification.
ment  Reagents of AMI Codes mixed wrongly or
not completely
Sample flow alarm,  Check sample flow at photometer outlet.
but there is sample It must be at least 100 ml/min. For that
place the photometer outlet tube into a
measuring cup for 1 minute.
 Check sample line for pressure fluctua-
tion.
 Check for regular air bubble pattern.
 Check flow alarm values in menu 5.3.1.3,
p. 96)

68 A-96.250.581 / 050517
AMI Codes-II CC
Troubleshooting

7.2. Calibration Errors

7.2.1 Process calibration tc or fc


Possible error Slope error:
message
Possible cause Corrective Action
Wrong manual measure- Repeat the manual measurement.
ment. Use fresh reagents.
 Wrong reagent mixture  Make a correct mixture.
 Reagents not complete-  Mix long and intensively.
ly solved in water.

7.2.2 Process pH
Possible error Offset error:
message
Possible cause Corrective Action
Manual measurement Repeat the manual measurement.
wrong.
Slope of last calibration Set default calibration values, see
wrong. 5.4.2, p. 104
Repeat the calibration
pH sensor dirty, old or Clean or replace pH sensor, see
defect. Maintenance of pH sensor, p. 60.
Cable connector corroded. Replace cable and sensor.

7.2.3 Standard pH
Possible Offset error or Slope error:
error message
Possible cause Corrective Action
Old, dirty or wrong buffer Check buffers expiration date if nec-
solutions. essary order new buffer.
Verify programmed buffer Change programmed buffer values
values with the values of or use correct buffer solution.
the buffer solution used.
pH sensor dirty, old or Clean or replace pH sensor, see
defect. Maintenance of pH sensor, p. 60.
Cable connector corroded. Replace cable and sensor.

A-96.250.581 / 050517 69
AMI Codes-II CC
Troubleshooting

7.3. Error List


Error
Non-fatal Error. Indicates an alarm if a programmed value is ex-
ceeded.
Such Errors are marked E0xx (bold and black).
Fatal Error (blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
 Errors which disappear if correct measuring conditions are re-
covered (i.e. Sample Flow low).
Such Errors are marked E0xx (bold and orange)
 Errors which indicate a hardware failure of the instrument.
Such Errors are marked E0xx (bold and red)

HOLD 14:10:45 Error or fatal Error


R1 Error not yet acknowledged.
0.22 ppm
0.26 ppm Check Pending Errors 1.1.5 * and
R2
0.04 ppm take corrective action.
28 B/s 25.4°C Press [ENTER].

Main Menu 1
Reagent level low
Messages Indicates the remaining reagent in per-
Diagnostics cent
Maintenance Navigate to menu Messages.
Operation Press [ENTER].
Installation

Messages 1.1 Navigate to menu Pending Errors.


Pending Errors Press [ENTER].
Maintenance List
Message List

Pending Errors 1.1.5 Press [ENTER] to acknowledge the


Error Code E002 Pending Errors. The Error is reset and
saved in the Message List.
Alarm low
* Menu numbers see
<Enter> to Acknowledge
Program Overview, p. 74

70 A-96.250.581 / 050517
AMI Codes-II CC
Troubleshooting

Error Description Corrective action

E001 Alarm high fac – check process


(Free available chlorine) – check programmed value 5.3.1.1.1.1, p.
94

E002 Alarm low fac – check process


(Free available chlorine) – check programmed value 5.3.1.1.1.26,
p. 94

E003 Alarm high tc1 – check process


(Total chlorine 1) – check programmed value 5.3.1.1.2, p.
95

E004 Alarm low tc1 – check process


(Total chlorine 1) – check programmed value, 5.3.1.1.2, p.
95

E005 Alarm high tc2 – check process


(Total chlorine 2) – check programmed value 5.3.1.1.3, p.
95

E006 Alarm low tc2 – check process


(Total chlorine 2) – check programmed value 5.3.1.1.3, p.
95

E007 Sample Temp.high – check sample temperature


– check programmed value 5.3.1.5, p. 96

E008 Sample Temp. low – check sample temperature


– check programmed value 5.3.1.5, p. 96

E009 Sample Flow high – check Inlet pressure


– readjust sample flow
– check program value 5.3.1.4.2, p. 96

E010 Sample Flow low – check Inlet pressure


– readjust sample flow
– clean instrument, see Cleaning the
protective Filter, p. 55
– check program value 5.3.1.4.36, p. 96

A-96.250.581 / 050517 71
AMI Codes-II CC
Troubleshooting

Error Description Corrective action

E011 Temp. shorted – check wiring of temperature sensor, see


Connection Diagram, p. 29
– check temperature sensor

E012 Temp. disconnected – check wiring of temperature sensor, see


Connection Diagram, p. 29
– check temperature sensor

E013 Case Temp. high – check case/environment temperature


– check program value 5.3.1.2, p. 96

E014 Case Temp. low – check case/environment temperature


– check program value 5.3.1.3, p. 96

E015 Valve 1 defective – Check Valve 1, see Cleaning the


solenoid valve, p. 64

E016 DIS invalid – This error appears after start-up and will
disappear after the first valid
measurement is finished.

E017 Control Timeout – check control device or programming in


Installation, Relay contact, Relay 1/2
5.3.2 and 5.3.3, p. 98

E018 Reagent Pump – shut off power


– check wiring, see Connection Diagram,
p. 29

E019 Photometer not con- – shut off power


nected – check wiring, see Connection Diagram,
p. 29

E020 Photometer dirty – check process


– clean the photometer, see Cleaning the
Photometer, p. 56

E021 Absorbance too high – Appears if disinf. value is too high


– Check dosing system or process

E022 Reagent empty – refill reagents,


see Refill or replace Reagents, p. 47

72 A-96.250.581 / 050517
AMI Codes-II CC
Troubleshooting

Error Description Corrective action

E023 Cleaning Solution – refill cleaning solution, see 6.3.,  47

E024 Input active – See If Fault Yes is programmed in Menu


5.3.4, p. 103

E026 IC LM75 – call service

E028 Signal output open – check wiring on signal outputs 1 and 2

E030 EEprom Frontend – call service

E031 Calibration Recout – call service

E032 Wrong Frontend – call service

E039 Alarm high pH – check process


– check programmed value 5.3.1.6, p. 97

E040 Alarm low pH – check process


– check programmed value 5.3.1.6, p. 97

E049 Power-on – none, normal status

E050 Power-down – none, normal status

E065 DPD / Buffer – Operating display, upper status line: The


number next to the triangle, indicates
the remaining reagents in%. Refill
reagents on time. See Refill or replace
Reagents, p. 47

E067 Cleaning Solution – Operating display, lower status line: The


number next to the triangle, indicates
the remaining cleaning solution in%.
Refill cleaning solution on time.

A-96.250.581 / 050517 73
AMI Codes-II CC
Program Overview

8. Program Overview
For explanations about each parameter of the menus see Program
List and Explanations, p. 81.

 Menu 1 Messages informs about pending errors and mainte-


nance tasks and shows the error history. Password protection
possible. No settings can be modified.
 Menu 2 Diagnostics is always accessible for everybody. No
password protection. No settings can be modified.
 Menu 3 Maintenance is for service: Calibration, simulation of
outputs and set time/date. Please protect with password.
 Menu 4 Operation is for the user, allowing to set limits, alarm
values, etc. The presetting is done in the menu Installation
(only for the System engineer). Please protect with password.
 Menu 5 Installation: Defining assignment of all inputs and
outputs, measuring parameters, interface, passwords, etc.
Menu for the system engineer. Password strongly recom-
mended.

8.1. Messages (Main Menu 1)


Reagent Status DPD/Buffer 1.1.1* * Menu numbers
1.1* Potassium Iodide
Cleaning Solution
Pending Errors Pending Errors 1.2.5*
1.2*
Maintenance List Maintenance List 1.3.5*
1.3*
Message List Number 1.4.1*
1.4* Date, Time

74 A-96.250.581 / 050517
AMI Codes-II CC
Program Overview

8.2. Diagnostics (Main Menu 2)


Identification Designation AMI Codes-II CC * Menu numbers
2.1* Version V6.00 - 02/16
Peripherals PeriClip 1 / 1.03 2.1.3.1*
2.1.3* PeriClip 2 only with cleaning module
Factory Test Instrument 2.1.4.1*
2.1.4* Motherboard
Front End
Operating Time Years / Days / Hours / Minutes / Seconds 2.1.5.1*
2.1.5*
Sensors Photometer Absorbance
2.2* 2.2.1* (Raw value) Hz
Cal. History Number 2.2.1.4.1*
2.2.1.4* Date, Time
Factor fc
Factor tc
Ver. History Number 2.2.1.5.1*
2.2.1.5* Date, Time
Absorbance
Reference value
pH Electrode Current Value / (Raw value)
2.2.2* Cal. History Number 2.2.2.5.1*
2.2.2.5* Date, Time
Offset / Slope
Miscellaneous Case Temp. 2.2.3.1*
2.2.3*
Sample Sample ID 2.3.1*
2.3* Sample Flow B/s
(Raw value) Hz
Temperature / (Nt5k)
I/O State Alarm Relay 2.4.1*
2.4* Relay 1 & 2 2.4.2*
Input
Signal Output 1 & 2
Interface Protocol 2.5.1* (only with RS485
2.5* Baud rate interface)

A-96.250.581 / 050517 75
AMI Codes-II CC
Program Overview

8.3. Maintenance (Main Menu 3)


Calibration Free av. chlorine Current Value * Menu numbers
3.1* 3.1.1* Factor
Process Value 3.1.1.4*
Total chlorine 2 Current Value
3.1.2* Factor
Process Value 3.1.2.4*
Process pH Current Value
3.1.3* Offset
Process Value 3.1.3.4*
Standard pH (Progress) 3.1.4.5*
3.1.4*
Service Verification (Progress) 3.2.1.1*
3.2* 3.2.1*
Fill System (Progress) 3.2.2.5*
3.2.2*
Simulation Alarm Relay 3.3.1*
3.3* Relay 1 3.3.2*
Relay 2 3.3.3*
Signal Output 1 3.3.4*
Signal Output 2 3.3.5*
Valve 1 3.3.6*
Valve 2 3.3.7*
Set Time (Date), (Time)
3.4*
Cleaning Parameter Mode 3.5.1.1*
3.5* 3.5.1* Start time / Calendar 3.5.1.xx*
Delay 3.5.1.3*
Signal Outputs 3.5.1.4*
Output/Control 3.5.1.5*
Fill Channel 11 (Progress) 3.5.2.5*
3.5.2*
Fill Channel 12 (Progress) 3.5.3.5*
3.5.3*

76 A-96.250.581 / 050517
AMI Codes-II CC
Program Overview

8.4. Operation (Main Menu 4)


Sensors Filter Time Const. 4.1.1*
4.1* Hold after Cal. 4.1.2*
Interval fac 4.1.3*
Interval tc2 4.1.4*
Default pH 4.1.5*
Relay Contacts Alarm Relay Free av. chlorine Alarm High 4.2.1.1.1*
4.2* 4.2.1* 4.2.1.1* Alarm Low 4.2.1.1.25*
Hysteresis 4.2.1.1.35*
Delay 4.2.1.1.45*
Total chlorine 1 Alarm High 4.2.1.2.1*
4.2.1.2* Alarm Low 4.2.1.2.25*
Hysteresis 4.2.1.2.35*
Delay 4.2.1.2.45*
Total chlorine 2 Alarm High 4.2.1.3.1*
4.2.1.3* Alarm Low 4.2.1.3.25*
Hysteresis 4.2.1.3.35*
Delay 4.2.1.3.45*
Calc. Monochl. Alarm High 4.2.1.4.1*
4.2.1.4* Alarm Low 4.2.1.4.25*
Hysteresis 4.2.1.4.35*
Delay 4.2.1.4.45*
Alarm pH Alarm High 4.2.1.5.1*
4.2.1.5* Alarm Low 4.2.1.5.25*
Hysteresis 4.2.1.5.35*
Delay 4.2.1.5.45*
Relay 1 & 2 Parameter
4.2.2* & 4.2.3* Setpoint 4.2.x.200*
Hysteresis 4.2.x.300*
Delay 4.2.x.40*
Input Active 4.2.4.1*
4.2.4* Signal Outputs 4.2.4.2*
Output / Control 4.2.4.3*
Fault 4.2.4.4*
Delay 4.2.4.5*

A-96.250.581 / 050517 77
AMI Codes-II CC
Program Overview

Logger Log Interval 4.3.1*


4.3* Clear Logger 4.3.2*
Eject USB Stick 4.3.3* (If USB interface is installed)
Display Screen 1 Row 1 4.4.1.1*
4.4* 4.4.1* Row 2 4.4.1.2*
Row 3 4.4.1.3*
Screen 2 Row 1 4.4.2.1*
4.4.2* Row 2 4.4.2.2*
Row 3 4.4.2.3*

* Menu numbers

78 A-96.250.581 / 050517
AMI Codes-II CC
Program Overview

8.5. Installation (Main Menu 5)


Sensors Dimension 5.1.1* * Menu numbers
5.1* Interpolation 5.1.2*
Ref. Verification 5.1.3*
Standards Standard 1 5.1.4.1* only with pH option
5.1.4* Standard 2 5.1.4.2*
Cleaning 5.1.5 only with Cleaning Module
Signal Outputs Signal Output 1 & 2 Parameter 5.2.1.1 & 5.2.2.1*
5.2* 5.2.1* & 5.2.2* Current Loop 5.2.1.2 & 5.2.2.2*
Function 5.2.1.3 & 5.2.2.3*
Scaling Range Low 5.2.x.40.10*
5.2.x.40 Range High 5.2.x.40.20*
Relay Contacts Alarm Relay Disinfection Free av. chlorine Alarm High 5.3.1.1.1.1*
5.3* 5.3.1* 5.3.1.1* 5.3.1.1.1* Alarm Low 5.3.1.1.1.25*
Hysteresis 5.3.1.1.1.35*
Delay 5.3.1.1.1.45*
Total chlorine 1 Alarm High 5.3.1.1.2.1*
5.3.1.1.2* Alarm Low 5.3.1.1.2.25*
Hysteresis 5.3.1.1.2.35*
Delay 5.3.1.1.2.45*
Total chlorine 2 Alarm High 5.3.1.1.3.1*
5.3.1.1.3* Alarm Low 5.3.1.1.3.25*
Hysteresis 5.3.1.1.3.35*
Delay 5.3.1.1.3.45*
Case Temp. high 5.3.1.2*
Case Temp. low 5.3.1.3*
Sample Flow Flow Alarm 5.3.1.4.1*
5.3.1.4* Alarm High 5.3.1.4.2*
Alarm Low 5.3.1.4.35*
Temperature Alarm High 5.3.1.5.1*
5.3.1.5* Alarm Low 5.3.1.5.25*
pH Alarm High 5.3.1.6.1*
5.3.1.6* Alarm Low 5.3.1.6.25*
Hysteresis 5.3.1.6.35*
Delay 5.3.1.6.45*

A-96.250.581 / 050517 79
AMI Codes-II CC
Program Overview

Relay 1 & 2 Function 5.3.2.1 & 5.3.3.1*


5.3.2* & 5.3.3* Parameter 5.3.2.x & 5.3.3.x*
Setpoint 5.3.2.x & 5.3.3.x*
Hysteresis 5.3.2.x & 5.3.3.x*
Delay 5.3.2.x & 5.3.3.x*
Input Active 5.3.4.1*
5.3.4* Signal Outputs 5.3.4.2*
Output/Control 5.3.4.3*
Fault 5.3.4.4*
Delay 5.3.4.5*
Miscellaneous Language 5.4.1*
5.4* Set defaults 5.4.2*
Load Firmware 5.4.3*
Password Messages 5.4.4.1*
5.4.4* Maintenance 5.4.4.2*
Operation 5.4.4.3*
Installation 5.4.4.4*
Sample ID 5.4.5*
Line break detection 5.4.6*
Interface Protocol 5.5.1* (only with RS485
5.5* Device Address 5.5.21* interface)
Baud Rate 5.5.31*
Parity 5.5.41* *Menu numbers

80 A-96.250.581 / 050517
AMI Codes-II CC
Program List and Explanations

9. Program List and Explanations

1 Messages
1.1 Reagent Status
1.1.1 DPD/Buffer: Shows the fill level of the DPD/Buffer.
Potassium Iodide: Shows the fill level of the Potassium Iodide.
Cleaning solution: Shows the fill level of the Cleaning solution.
A behind the reagent means that the filling level is ok. If the fill-
ing level falls below 17%, it is shown in%.
1.2 Pending Errors
1.2.5 Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
1.3 Maintenance List
1.3.5 Demands necessary maintenance, e.g. preparing new reagents.
1.4 Message List
1.4.1 Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Designation: View the Designation of instrument.
Version: Firmware of instrument (e.g. V6.00-02/16)
2.1.3 Peripherals: PeriClip 1: Firmware of peristaltic pump (e.g. 1.06)
2.1.4 Factory Test: Test date of the Instrument, Motherboard and Fron-
tend. QC factory test.
2.1.5 Operating Time: Years/Days/Hours/Minutes/Seconds

A-96.250.581 / 050517 81
AMI Codes-II CC
Program List and Explanations

2.2 Sensors
2.2.1 Photometer:
Absorbance: Process value, depends on sample.
Raw value: Shows the actual photometer signal in Hz.
2.2.1.4 Cal. History: Shows the diagnostic values of the last calibrations.
Number: Counter for the calibrations
Date, Time: Date and time assigned to a number.
Factor fc: Factor fc is the multiplier applied to the slope of the free
chlorine calibration line. The ideal factor is 1.00.
Factor tc: Factor tc is the multiplier applied to the slope of the total
chlorine calibration line. The ideal factor is 1.00
2.2.1.5 Ver. History: Shows the verification values of the last verifications:
Number: Calibration counter.
Date, Time: Date and time of the calibration.
Absorbance: Measured absorbance of the reference kit.
Reference value: True value of the reference kit according to
label.
2.2.2 pH Electrode: Only if option pH is installed.
Current Value: Shows the actual measured pH value.
Raw value: Shows the actual electrode voltage in mV.
2.2.2.5 Cal. History: Shows the calibration values of the last pH sensor
calibrations. Offset in mV and slope in mV/pH.
Typical offset of pH electrode: <  30 mV.
Max. tolerated offset: <  60 mV
Typical slope of pH electrode: 55–65 mV/pH unit.
Max. limits: 40–65 mV/pH
2.2.3 Miscellaneous:
2.2.3.1 Case Temp: Shows the current temperature in °C inside the trans-
mitter.
2.3 Sample
2.3.1 Sample ID: Shows the identification assigned to a sample. This
identification is defined by the user to identify the location of the
sample.
Sample Flow: Shows the actual sample flow in B/s (bubbles per
second). Sample flow must be above 5 B/s.
Raw value: Shows the raw value of the sample flow in Hz.
Temperature: Only if option pH is installed.
Actual sample temperature in °C and in Ohm (NT5K)

82 A-96.250.581 / 050517
AMI Codes-II CC
Program List and Explanations

2.4 I/O State


Shows current status of all in- and outputs.
2.4.1 & 2.4.2
Alarm Relay: Active or inactive
Relay 1 & 2: Active or inactive
Input: Open or closed
Signal Output 1 & 2: Actual current in mA
Signal Output 3: Actual current in mA (if option is installed)

2.5 Interface
Only available if optional interface is installed.
Shows the programmed communication settings.

3 Maintenance
3.1 Calibration
In this menu, you can correct measuring values (disinfectant and
pH) or calibrate offset and slope of pH electrode.
3.1.1 Free av. chlorine: Possibility to correct the free available chlorine
value. See Process Calibration of tc & fc, p. 52, for more details.
3.1.2 Total chlorine 2: Possibility to correct the total chlorine value. See
Process Calibration of tc & fc, p. 52, for more details.
3.1.3 Process pH: Only available, if option pH has been installed. Cor-
rection of pH electrode. See Process Calibration of tc & fc, p. 52,
for details.
3.1.4 Standard pH: Only available, if option pH has been installed. Cali-
bration of pH electrode with the two standard solutions pro-
grammed in Installation 5.1.3. See Standard pH, p. 54, for details.
3.2 Service
3.2.1 Verification: Performs a verification using the reference kit. Follow
dialog. See Verification, p. 51
3.2.2 Fill System: Activates the reagent pump. The function Fill System
is used to fill or flush the reagent tubes.

A-96.250.581 / 050517 83
AMI Codes-II CC
Program List and Explanations

3.3 Simulation
To simulate a value or a relay state, select the
 alarm relay,
 relay 1 or 2
 signal output 1 or 2
 valve 1 or 2
with the [ ] or [ ] key.
Press the [Enter] key.
Change the value or state of the selected item with the [ ] or
[ ] key.
Press the [Enter] key.
 The value is simulated by the relay/signal output.
Alarm Relay: Active or inactive
Relay 1 & 2: Active or inactive
Signal Output 1 & 2: Actual current in mA
Signal Output 3: Actual current in mA (if option is installed)
Valve 1 (photometer) Active or inactive
Valve 2 (KI supply) Active or inactive

At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val-
ues will be reset.
3.4 Set Time
Adjust date and time.
3.5 Cleaning
Automatic cleaning process using the Cleaning module-II con-
trolled by transmitter of AMI Codes-II CC.
 Measurement is interrupted.
 Flow error interrupts cleaning.
 Cleaning does not start: E022 Reagent empty, E023 Cleaning
solution, no flow.
3.5.1 Parameter:
3.5.1.1 Mode: Select daily, weekly or off.
3.5.1.xx Start time/Calendar: Time off day respectively weekday(s) for the
automatic start of the cleaning process.

84 A-96.250.581 / 050517
AMI Codes-II CC
Program List and Explanations

3.5.1.3 Delay: During cleaning plus the delay time, the status of the signal-
and control outputs is as get below.
Range: 60–6’000 sec
3.5.1.4 Signal Outputs: Select the operation mode of the signal outputs
during cleaning:
Cont.: Signal outputs continue to issue the measured value.
Hold: Signal outputs hold the last valid measured value.
Measurement is interrupted. Errors, except fatal errors,
are not issued.
Off: Signal outputs are switched off (set to 0 or 4 mA).
Errors, except fatal errors, are not issued.

3.5.1.5 Output/Control: Relay or signal output:


Cont.: Controller continues normally.
Hold: Controller continues based on the last valid value.
Off: Controller is switched off.

3.5.2 Fill Channel 11: Activates the cleaning pump and switches the
valve to the cleaning solution 1 (right container)
3.5.3 Fill Channel 12: Activates the cleaning pump and switches the
valve to the cleaning solution 2 (left container)

A-96.250.581 / 050517 85
AMI Codes-II CC
Program List and Explanations

4 Operation
4.1 Sensors
4.1.1 Filter Time Constant: Used to damp noisy signals. The higher the
filter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 Sec
4.1.2 Hold after Cal: Delay permitting the instrument to stabilize again af-
ter calibration. During calibration- plus hold-time, the signal outputs
are frozen (held on last valid value), alarm values, limits are not ac-
tive.
Range: 0–6‘000 Sec
4.1.3 Interval fac: The measuring interval of “Free available chlorine” can
be set according to the process requirements.
Range: 0–12 min
4.1.4 Interval tc2: The measuring interval of “Total chlorine 2” can be set
according to the process requirements. If the interval is set to 0 min
only fac is measured.
Range: 0–60 min
4.1.5 Default pH: If the pH value of the sample is known it can be entered
in this menu. This menu only appears if the pH option is not in-
stalled.
Range: 0–14 pH
4.2 Relay Contacts
See 5.3 Relay Contacts, p. 94
4.3 Logger
The instrument is equipped with an internal logger. The data can be
 copied to a PC with an USB stick if option USB interface is in-
stalled.
The logger can save approx. 1500 data records. Records consists
of: Date, time, alarms, measuring values, temperature, flow.
Range: 1 Second to 1 hour
4.3.1 Log Interval: Select a convenient log interval. Consult the table be-
low to estimate the max logging time. When the login buffer is full,
the oldest data record is erased to make room for the newest one.
(circular buffer)

Interval 1s 5s 1 min 5 min 10 min 30 min 1h Event


Driven
Time 25 min 2h 25 h 5d 10 d 31 d 62 d

86 A-96.250.581 / 050517
AMI Codes-II CC
Program List and Explanations

4.3.2 Clear Logger: If confirmed with yes, the complete logger data is de-
leted. A new data series is started.
4.4 Display
Process values are displayed on two screens. Toggle screens with
the [ ] key. Each screen displays max. 3 process values.
4.4.1 Screen 1
4.4.1.1 Row 1
4.4.1.2 Row 2
4.4.1.3 Row 3
Possible settings for all rows are:
 None
 Free av. chlorine (Free available chlorine)
 Total chlorine 1
 Total chlorine 2
 Calc. monochl. (Calculated monochloramine)
 Calc. comb. cl. (Calculated combined chlorine)
 Calc. dichloramine (Calculated dichloramine)
 pH
4.4.2 Screen 2
Same as screen 1.

A-96.250.581 / 050517 87
AMI Codes-II CC
Program List and Explanations

5 Installation
5.1 Sensors
5.1.1 Dimension: The measuring value can be displayed as ppm or mg/l
5.1.2 Interpolation:
Yes: Calculates the average of the 2 last measuring values of free
chlorine measurement. Use this mode to avoid high spikes in the
control loop.
No: The true measuring value is displayed and available at the
outputs.
5.1.3 Ref. Verification: Set absorbance value of verification kit according
to label.
Range: 0.200–0.600
5.1.40 Standards: Program the two standard solutions for the calibration
of the pH electrode. If pH electrode is not connected the pro-
grammed standards are not active.
Range: 1.00 pH - 13.00 pH
5.1.5 Cleaning: Program if one or two solutions are used on Cleaning
module. See 3.5 Cleaning, p. 84
Range: 1 Solution, 2 Solutions
5.2 Signal Outputs
5.2.1&5.2.2 Signal Output 1 and 2: Assign process value, the current loop
range and a function to each signal output.
NOTICE: The navigation in the menu <Signal Output 1> and
<Signal Output 2> is equal. For reason of simplicity only the
menu numbers of Signal Output 1 are used in the following.
5.2.1.1 Parameter: Assign one of the process values to the signal output.
Available values:
 Free av. chlorine (Free available chlorine)
 Total chlorine 1
 Total chlorine 2
 Calc. monochl. (Calculated monochloramine)
 Calc. comb. cl. (Calculated combined chlorine)
 Calc. dichloramine (Calculated dichloramine)
 Temperature
 Sample flow
 pH

88 A-96.250.581 / 050517
AMI Codes-II CC
Program List and Explanations

5.2.1.2 Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current
range.
Available ranges: 0–20 mA or 4–20 mA
5.2.1.3 Function: Define if the signal output is used to transmit a process
value or to drive a control unit. Available functions are:
 Linear, bilinear or logarithmic for process values.
See As process values, p. 89
 Control upwards or control downwards for controllers.
See As control output, p. 91
As process The process value can be represented in 3 ways: linear, bilinear or
values logarithmic. See graphs below.
[mA]

20
(0 - 20 [mA])

(4 - 20 [mA])

B A
10 12

0/4 X
0.0 0.1 0.2 0.3 0.4 0.5

A linear X Measured value


B bilinear

[mA]

20
(0 - 20 [mA])

(4 - 20 [mA])

10 12

0/4 X
2 4 6
1 10 100 1’000 10’000
0 1 2 3 4

X Measured value (logarithmic)

A-96.250.581 / 050517 89
AMI Codes-II CC
Program List and Explanations

5.2.1.40 Scaling: Enter beginning and end point (Range low & high) of the
linear or logarithmic scale. In addition, the midpoint for the bilinear
scale.
Parameter Free av. chlorine
5.2.1.40.10 Range low: 0–10 ppm or 0–10 mg/l
5.2.1.40.20 Range high: 0–10 ppm or 0–10 mg/l
Parameter Total chlorine 1
5.2.1.40.11 Range low: 0–10 ppm or 0–10 mg/l
5.2.1.40.21 Range high: 0–10 ppm or 0–10 mg/l
Parameter Total chlorine 2
5.2.1.40.12 Range low: 0–10 ppm or 0–10 mg/l
5.2.1.40.22 Range high: 0–10 ppm or 0–10 mg/l
Parameter Calc. monochl.
5.2.1.40.13 Range low: 0–10 ppm or 0–10 mg/l
5.2.1.40.23 Range high: 0–10 ppm or 0–10 mg/l
Parameter Calc. comb. cl.
5.2.1.40.14 Range low: 0–10 ppm or 0–10 mg/l
5.2.1.40.24 Range high: 0–10 ppm or 0–10 mg/l
Parameter Calc. dichloramine
5.2.1.40.15 Range low: 0–10 ppm or 0–10 mg/l
5.2.1.40.25 Range high: 0–10 ppm or 0–10 mg/l
Parameter Temperature
5.2.1.40.16 Range low: -30 to +120 °C
5.2.1.40.26 Range high: -30 to +120 °C
Parameter Sample flow
5.2.1.40.17 Range low: 0 –600 B/s
5.2.1.40.27 Range high: 0 –600 B/s
Parameter pH
5.2.1.40.18 Range low: 0 –14 pH
5.2.1.40.28 Range high: 0 –14 pH

90 A-96.250.581 / 050517
AMI Codes-II CC
Program List and Explanations

As control out- Signal outputs can be used for driving control units. We distinguish
put different kinds of controls:
 P-controller: The controller action is proportional to the devia-
tion from the setpoint. The controller is characterized by the
P-Band. In the steady-state, the setpoint will never be
reached. The deviation is called steady-state error.
Parameters: setpoint, P-Band
 PI-controller: The combination of a P-controller with an
I-controller will minimize the steady-state error. If the reset
time is set to zero, the I-controller is switched off.
Parameters: setpoint, P-Band, reset time.
 PD-controller: The combination of a P-controller with a
D-controller will minimize the response time to a fast change
of the process value. If the derivative time is set to zero, the
D-controller is switched off.
Parameters: setpoint, P-Band, derivative time.
 PID-controller: The combination of a P-, an I - and a D-con-
troller allows a proper control of the process.
Parameters: setpoint, P-Band, reset time, derivative time.
Ziegler-Nichols method for the optimization of a PID controller:
Parameters: Setpoint, P-Band, Reset time, Derivative time
Y

B A

X
a

A Response to maximum control output Xp = 1.2/a


B Tangent on the inflection point Tn = 2L
X Time Tv = L/2

The point of intersection of the tangent with the respective axis will
result in the parameters a and L.
Consult the manual of the control unit for connecting and program-
ming details. Choose control upwards or downwards.

A-96.250.581 / 050517 91
AMI Codes-II CC
Program List and Explanations

Control upwards/downwards
Setpoint: User-defined process value (Measured value or flow)
P-Band: Range below (upwards control) or above (downwards
control) the set-point, within the dosing intensity is reduced from
100% to 0% to reach the set-point without overshooting.
5.2.1.43 Control Parameters: if Parameters = Free av. chlorine
5.2.1.43.10 Setpoint: 0–10 ppm or 0–10 mg/l
5.2.1.43.20 P-Band: 0–10 ppm 0–10 mg/l
5.2.1.43 Control Parameters: if Parameters = Total chlorine 1
5.2.1.43.11 Setpoint: 0–10 ppm or 0–10 mg/l
5.2.1.43.21 P-Band: 0–10 ppm 0–10 mg/l
5.2.1.43 Control Parameters: if Parameters = Total chlorine 2
5.2.1.43.12 Setpoint: 0–10 ppm or 0–10 mg/l
5.2.1.43.22 P-Band: 0–10 ppm 0–10 mg/l
5.2.1.43 Control Parameters: if Parameters = Calc. monochl.
5.2.1.43.13 Setpoint: 0–10 ppm or 0–10 mg/l
5.2.1.43.23 P-Band: 0–10 ppm 0–10 mg/l
5.2.1.43 Control Parameters: if Parameters = Calc. comb. cl.
5.2.1.43.14 Setpoint: 0–10 ppm or 0–10 mg/l
5.2.1.43.24 P-Band: 0–10 ppm 0–10 mg/l
5.2.1.43 Control Parameters: if Parameters = Calc. dichloramine
5.2.1.43.15 Setpoint: 0–10 ppm or 0–10 mg/l
5.2.1.43.25 P-Band: 0–10 ppm 0–10 mg/l
5.2.1.43 Control Parameters: if Parameters = Temperature
5.2.1.43.16 Setpoint: -30 to +120 °C
5.2.1.43.26 P-Band: 0–100 °C
5.2.1.43 Control Parameters: if Parameters = Sample Flow
5.2.1.43.17 Setpoint: 0 –600 B/s
5.2.1.43.27 P-Band: 0 –200 B/s
5.2.1.43 Control Parameters: if Parameters = pH
5.2.1.43.18 Setpoint: 0 –14 pH
5.2.1.43.28 P-Band: 0 –14 pH

92 A-96.250.581 / 050517
AMI Codes-II CC
Program List and Explanations

5.3.2.32.31.3 Reset time: The reset time is the time till the step response of a sin-
gle I-controller will reach the same value as it will be suddenly
reached by a P-controller.
Range: 0–9’000 Sec
5.3.2.32.31.4 Derivative time: The derivative time is the time till the ramp re-
sponse of a single P-controller will reach the same value as it will
be suddenly reached by a D-controller.
Range: 0–9’000 Sec
5.3.2.32.31.5 Control timeout: If a controller action (dosing intensity) is constantly
over 90% during a defined period of time and the process value
does not come closer to the setpoint, the dosing process will be
stopped for safety reasons.
Range: 0–720 min

A-96.250.581 / 050517 93
AMI Codes-II CC
Program List and Explanations

5.3 Relay Contacts


5.3.1 Alarm Relay: The alarm relay is used as cumulative error indicator.
Under normal operating conditions the contact is active.
The contact is inactive at:
 Power loss
 Detection of system faults like defective sensors or electronic
parts
 High case temperature
 Lack of reagents
 Process values out of programmed ranges.
Program alarm levels, hysteresis values and delay times for the fol-
lowing parameters:
 Free chlorine
 Total chlorine 1
 Total chlorine 2
 Case Temperature high and low
 Sample flow
 Temperature
 pH

5.3.1.1 Disinfection
5.3.1.1.1 Free av. chlorine
5.3.1.1.1.1 Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E001 is displayed in the mes-
sage list. Range: 0.00–10.00 ppm
5.3.1.1.1.26 Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E002 is displayed in the message
list.
Range: 0.00–10.00 ppm
5.3.1.1.1.36 Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range. 0.00–10.00 ppm
5.3.1.1.1.46 Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0.00–28‘800 Sec

94 A-96.250.581 / 050517
AMI Codes-II CC
Program List and Explanations

5.3.1.1.2 Total chlorine 1


5.3.1.1.2.1 Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E003 is displayed in the mes-
sage list. Range: 0.00–10.00 ppm
5.3.1.1.2.26 Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E004 is displayed in the message
list.
Range: 0.00–10.00 ppm
5.3.1.1.2.36 Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range. 0.00–10.00 ppm
5.3.1.1.2.46 Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0.00–28‘800 Sec
5.3.1.1.3 Total chlorine 2
5.3.1.1.3.1 Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E005 is displayed in the mes-
sage list. Range: 0.00–10.00 ppm
5.3.1.1.3.26 Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E006 is displayed in the message
list.
Range: 0.00–10.00 ppm
5.3.1.1.3.36 Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range. 0.00–10.00 ppm
5.3.1.1.3.46 Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above or fallen below the pro-
grammed alarm.
Range: 0.00–28‘800 Sec

A-96.250.581 / 050517 95
AMI Codes-II CC
Program List and Explanations

5.3.1.2 Case Temp. high: Set the alarm high value for temperature of elec-
tronics housing. If the value rises above the programmed value
E013 is issued.
Range: 30–75 °C
5.3.1.3 Case Temp. low: Set the alarm low value for temperature of elec-
tronics housing. If the value falls below the programmed value
E014 is issued.
Range: -10 to +20 °C
5.3.1.4 Sample Flow: Define at which sample flow a flow alarm should be
issued.
5.3.1.4.1 Flow Alarm: Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will al-
ways be indicated in the display, pending error list, saved in the
message list and the logger.
Available values: Yes or no
NOTICE: Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.4.2 Alarm High: If the measuring values rises above the programmed
value E009 will be issued.
Range: 100–600 B/s
5.3.1.4.36 Alarm Low: If the measuring values falls below the programmed
value E010 will be issued.
Range: 5–80 B/s
5.3.1.5 Temperature: Only available if pH option is installed. Define the
measuring value, which should issue an alarm high respectively
low.
5.3.1.5.1 Alarm High: If the sample temperature rises above the pro-
grammed value E007 is issued.
Range: 30–70 °C
5.3.1.5.26 Alarm Low: If the sample temperature falls below the programmed
value E008 is issued.
Range: 0–20 °C

96 A-96.250.581 / 050517
AMI Codes-II CC
Program List and Explanations

5.3.1.6 Alarm pH: Only available if pH option is installed. Define the mea-
suring value, which should issue an alarm high respectively low.
5.3.1.6.1 Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E039 is displayed in the mes-
sage list.
Range: 0–14 pH
5.3.1.6.26 Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E040 is displayed in the message
list
Range: 0–14 pH
5.3.1.6.36 Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value
Range: 0–14 pH
5.3.1.6.46 Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec

A-96.250.581 / 050517 97
AMI Codes-II CC
Program List and Explanations

5.3.2 and 5.3.3 Relay 1 and 2: The contacts can be set as normally open or nor-
mally closed with a jumper. See Relay 1 and 2, p. 32.
The function of relay contacts 1 or 2 are defined by the user.
NOTICE: The navigation in the menu <Relay 1> and <Relay 2>
is equal. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1 First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
- End of Batch (relay 2 only)
2 Then enter the necessary data depending on the selected func-
tion.
5.3.2.1 Function = Limit upper/lower:
When the relays are used as upper or lower limit switches, program
the following:
5.3.2.20 Parameter: select a process value
5.3.2.300 Setpoint: If the measured value rises above respectively falls below
the set-point, the relay is activated.
Parameter Range
Free av. chlorine 0–10 ppm
Total chlorine 1 0–10 ppm
Total chlorine 2 0–10 ppm
Calc. monochl. 0–10 ppm
Calc. comb. cl. 0–10 ppm
Calc. dichloramine 0–10 ppm
Temperature -30 to +120 °C
Sample flow 0 –600 B/s
pH 0–14 pH

98 A-96.250.581 / 050517
AMI Codes-II CC
Program List and Explanations

5.3.2.400 Hysteresis: within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value
fluctuates around the alarm value.
Parameter Range
Free av. chlorine 0–10 ppm
Total chlorine 1 0–10 ppm
Total chlorine 2 0–10 ppm
Calc. monochl. 0–10 ppm
Calc. comb. cl. 0–10 ppm
Calc. dichloramine 0–10 ppm
Temperature 0 to +100 °C
Sample flow 0 –200 B/s
pH 0–14 pH
5.3.2.50 Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range. 0–600 Sec
5.3.2.1 Function = Control upwards/downwards:
The relays may be used to drive control units such as solenoid
valves, membrane dosing pumps or motor valves. When driving a
motor valve both relays are needed, relay 1 to open and relay 2 to
close the valve.
5.3.2.22 Parameter: Choose on of the following process values.
 Free av. chlorine (Free available chlorine)
 Total chlorine 1
 Total chlorine 2
 Calc. monochl. (Calculated monochloramine)
 Calc. comb. cl. (Calculated combined chlorine)
 Calc. dichloramine (Calculated dichloramine)
 Temperature
 Sample flow
 pH
5.3.2.32 Settings: Choose the respective actuator:
 Time proportional
 Frequency
 Motor valve

A-96.250.581 / 050517 99
AMI Codes-II CC
Program List and Explanations

5.3.2.32.1 Actuator = Time proportional


Examples of metering devices that are driven time proportional are
solenoid valves, peristaltic pumps.
Dosing is controlled by the operating time.
5.3.2.32.20 Cycle time: duration of one control cycle (on/off change).
Range: 0–600 Sec.
5.3.2.32.30 Response time: Minimal time the metering device needs to react.
Range: 0–240 Sec.
5.3.2.32.4 Control Parameters
Range for each Parameter same as 5.2.1.43, p. 92

5.3.2.32.1 Actuator = Frequency

Examples of metering devices that are pulse frequency driven are


the classic membrane pumps with a potential free triggering input.
Dosing is controlled by the repetition speed of dosing shots.
5.3.2.32.21 Pulse frequency: Max. pulses per minute the device is able to re-
spond to. Range: 20–300/min.
5.3.2.32.31 Control Parameters
Range for each Parameter same as 5.2.1.43, p. 92

5.3.2.32.1 Actuator = Motor valve


Dosing is controlled by the position of a motor driven mixing valve.
5.3.2.32.22 Run time: Time needed to open a completely closed valve
Range: 5–300 Sec.
5.3.2.32.32 Neutral zone: Minimal response time in % of the runtime. If the re-
quested dosing output is smaller than the response time, no
change will take place.
Range: 1–20 %

100 A-96.250.581 / 050517


AMI Codes-II CC
Program List and Explanations

5.3.2.32.4 Control Parameters


Range for each Parameter same as 5.2.1.43, p. 92
5.3.2.1 Function = Timer:
The relay will be active repetitively depending on the programmed
time scheme.

5.3.2.24 Mode: Operating mode (interval, daily, weekly)


5.3.2.24 Interval
5.3.2.340 Interval: The interval can be programmed within a range
of 1–1’440 min.
5.3.2.44 Run Time: Enter the time the relay stays active.
Range: 5–32’400 Sec.
5.3.2.54 Delay: during run time plus the delay time the signal and control
outputs are held in the operating mode programmed below.
Range: 0–6’000 Sec.
5.3.2.6 Signal Outputs: Select operating mode of the signal output:
Cont.: Signal outputs continue to issue the measured value.
Hold: Signal outputs hold the last valid measured value.
Measurement is interrupted. Errors, except fatal errors,
are not issued.
Off: Signal outputs are switched off (set to 0 or 4 mA).
Errors, except fatal errors, are not issued.

5.3.2.7 Output/Control: Select operating mode of the controller output:


Cont.: Controller continues normally.
Hold: Controller continues based on the last valid value.
Off: Controller is switched off.

A-96.250.581 / 050517 101


AMI Codes-II CC
Program List and Explanations

5.3.2.24 daily
The relay contact can be activated daily, at any time of a day.
5.3.2.341 Start time: to set the start time proceed as follows:
1 Press [Enter], to set the hours.
2 Set the hour with the [ ] or [ ] keys.
3 Press [Enter], to set the minutes.
4 Set the minutes with the [ ] or [ ] keys.
5 Press [Enter], to set the seconds.
6 Set the seconds with the [ ] or [ ] keys.
Range: 00:00:00–23:59:59
5.3.2.44 Run Time: see Interval
5.3.2.54 Delay: see Interval
5.3.2.6 Signal Outputs: see Interval
5.3.2.7 Output/Control: see Interval
5.3.2.24 weekly
The relay contact can be activated at one or several days, of a
week. The daily starting time is valid for all days.
5.3.2.342 Calendar:
5.3.2.342.1 Start time: The programmed start time is valid for each of the pro-
grammed days. To set the start time see 5.3.2.341, p. 102.
Range: 00:00:00–23:59:59
5.3.2.342.2 Monday: Possible settings, on or off
to
5.3.2.342.8 Sunday: Possible settings, on or off
5.3.2.44 Run Time: see Interval
5.3.2.54 Delay: see Interval
5.3.2.6 Signal Outputs: see Interval
5.3.2.7 Output/Control: see Interval
5.3.2.1 Function = Fieldbus:
The relay will be switched via the Profibus input. No further param-
eters are needed.

102 A-96.250.581 / 050517


AMI Codes-II CC
Program List and Explanations

5.3.3.1 Function = End of Batch


This function is only available on relay 2. It is used to communicate
with canal switching instruments from third-party suppliers. The re-
lay closes for 1 sec. after each valid measurement. If End of Batch
is selected, no further selection is possible.

5.3.4 Input: The functions of the relays and signal outputs can be de-
fined depending on the position of the input contact, i.e. no function,
closed or open.
5.3.4.1 Active: Define when the input should be active:
The measurement is interrupted during the time the input is active.
No: Input is never active.
When closed: Input is active if the input relay is closed
When open: Input is active if the input relay is open
5.3.4.2 Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
Cont.: Signal outputs continue to issue the measured
value.
Hold: Signal outputs issue the last valid measured value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off: Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
5.3.4.3 Output/Control: (relay or signal output):
Cont.: Controller continues normally.
Hold: Controller continues on the last valid value.
Off: Controller is switched off.
5.3.4.4 Fault:
No: No message is issued in pending error list and the
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Yes: Message E024 is issued and stored in the mes-
sage list. The Alarm relay closes when input is
active.

A-96.250.581 / 050517 103


AMI Codes-II CC
Program List and Explanations

5.3.4.5 Delay: Time which the instrument waits, after the input is deactivat-
ed, before returning to normal operation.
Range: 0–6‘000 Sec

5.4 Miscellaneous
5.4.1 Language: Set the desired language.
Available settings: German /English/French/Spanish
5.4.2 Set defaults: Reset the instrument to factory default values in three
different ways:
 Calibration: Sets calibration values back to default. All other
values are kept in memory.
 In parts: Communication parameters are kept in memory. All
other values are set back to default values.
 Completely: Sets back all values including communication
parameters.
5.4.3 Load Firmware: Firmware updates should be done by instructed
service personnel only.
5.4.4 Password: Select a password different from 0000 to prevent unau-
thorized access to the menus “Messages”, “Maintenance”, “Opera-
tion” and “Installation”.
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN representa-
tive.
5.4.5 Sample ID: Identify the process value with any meaning full text,
such as KKS number.
5.4.6 Line Break Detection: If activated, error message E028 is shown in
case of line break on signal outputs 1 and 2.

104 A-96.250.581 / 050517


AMI Codes-II CC
Program List and Explanations

5.5 Interface
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.

5.5.1 Protocol: Profibus


5.5.20 Device address: Range: 0–126
5.5.30 ID-Nr.: Range: Analyzer; Manufacturer; Multivariable
5.5.40 Local operation: Range: Enabled, Disabled
5.5.1 Protocol: Modbus RTU
5.5.21 Device address: Range: 0–126
5.5.31 Baud Rate: Range: 1200–115 200 Baud
5.5.41 Parity: Range: none, even, odd
5.5.1 Protocol: USB-Stick:
Only visible if an USB interface is installed. No further settings are
possible.

A-96.250.581 / 050517 105


AMI Codes-II CC
Material Safety Data sheets

10. Material Safety Data sheets

10.1. Reagents
Catalogue No.: A-85.410.120
Product name: OXYCON ON-LINE DPD

Catalogue No.: A-85.410.120


Product name: OXYCON ON-LINE Buffer

Catalogue No: A-85.419.200


Product name: OXYCON ON-LINE KI

Catalogue No.: A-85.112.300


Product name: Calibration Solution pH 4

Catalogue No.: A-85.113.300


Product name: Calibration Solution pH 7

Catalogue No: A-85.114.300


Product name: Calibration Solution pH 9

Download The current Material Safety Data Sheets (MSDS) for the above list-
MSDS ed Reagents are available for downloading at www.swan.ch.

106 A-96.250.581 / 050517


AMI Codes-II CC
Default Values

11. Default Values


NOTICE: The parameter Cleaning is only visible if an optional
Cleaning Module is connected to the AMI Codes II.
The parameters pH and Temperature are only visible if the
option pH is installed.
Operation:
Sensors: Filter Time Const.: ................................................................. 30 Sec
Hold after Cal.:..................................................................... 120 Sec
Interval fac: .............................................................................. 5 Min
Interval tc2 ............................................................................. 20 Min
default pH: (If pH option is not installed)................................... 7 pH
Alarm Relay ...................................................................... same as in Installation
Relay 1 and 2 ...................................................................... same as in Installation
Input ...................................................................... same as in Installation
Logger: Logger Interval:........................................................... Event Driven
Clear Logger:................................................................................ no
Display Display 1; Line 1 .....................................................Free av. chlorine
Display 1; Line 2 ....................................................... Total chlorine 2
Display 1; Line 3 ....................................................... Calc. comb. cl.
Display 2; Line 1 ....................................................... Total chlorine 1
Display 2; Line 2 ................................................. Calc. dichloramine
Display 2; Line 3 ....................................................... Total chlorine 2
Installation:
Sensor: Dimension:................................................................................. ppm
Interpolation:................................................................................. no
Ref. Verification: ...................................................................... 0.255
Standard: Standard 1: ......................................................... 7.00 pH
Standard: Standard 2: ......................................................... 9.00 pH
Cleaning ......................................................................... 2 Solutions
Signal Output Parameter:.............................................................Free av. chlorine.
1 and 2 Current loop:..................................................................... 4 - 20 mA
Function:.................................................................................. linear
Scaling: Range low:........................................................... 0.00 ppm
Scaling: Range high: ......................................................... 5.00 ppm
Scaling: Temperature: Range low:......................................... 0.0 °C
Scaling: Temperature: Range high: ...................................... 50.0 °C
Scaling: pH: Range low: ..................................................... 0.00 pH
Scaling: pH: Range high:................................................... 14.00 pH

A-96.250.581 / 050517 107


AMI Codes-II CC
Default Values

Scaling: Sample Flow: Range low: ...........................................0 B/s


Scaling: Sample Flow: Range high:......................................200 B/s
Alarm Relay Disinfection
Free av. chlorine, Alarm high: ..........................................10.00 ppm
Free av. chlorine, Alarm low:..............................................0.00 ppm
Free av. chlorine, Hysteresis:.............................................0.10 ppm
Free av. chlorine, Delay: ..........................................................5 Sec
Total chlorine 1, Alarm high:.............................................10.00 ppm
Total chlorine 1, Alarm low: ................................................0.00 ppm
Total chlorine 1, Hysteresis:...............................................0.10 ppm
Total chlorine 1, Delay: ............................................................5 Sec
Total chlorine 2, Alarm high:.............................................10.00 ppm
Total chlorine 2, Alarm low: ................................................0.00 ppm
Total chlorine 2, Hysteresis:...............................................0.10 ppm
Total chlorine 2, Delay: ............................................................5 Sec
pH: Alarm high: .................................................................. 14.00 pH
pH: Alarm low:...................................................................... 0.00 pH
pH: Hysteresis: .................................................................... 0.10 pH
pH: Delay: ................................................................................5 Sec
Sample Flow: Flow Alarm: .......................................................... yes
Sample Flow: Alarm High: ....................................................500 B/s
Sample Flow: Alarm Low: .........................................................5 B/s
Sample Temp.: Alarm High: .....................................................55 °C
Sample Temp.: Alarm Low: ........................................................5 °C
Case temp. high:......................................................................65 °C
Case temp. low: .........................................................................0 °C
Relay1 and 2 Function: ......................................................................... Limit upper
Parameter: ............................................................. Free av. chlorine
Setpoint:.............................................................................5.00 ppm
Hysteresis: .........................................................................0.10 ppm
Delay:.....................................................................................30 Sec
If Function = Control upw. or dnw:
Parameter: ............................................................Free av. chlorine
Settings: Actuator: ...........................................................Frequency
Settings: Pulse Frequency: ............................................ 120/min.
Settings: Control Parameters: Setpoint: .........................5.00 ppm
Settings: Control Parameters: P-band:...........................0.10 ppm
Parameter: ................................................... Total chlorine 1 and 2
Settings: Actuator: ...........................................................Frequency
Settings: Pulse Frequency: ............................................ 120/min.
Settings: Control Parameters: Setpoint: .........................5.00 ppm
Settings: Control Parameters: P-band:...........................0.10 ppm

108 A-96.250.581 / 050517


AMI Codes-II CC
Default Values

Parameter:.............................................................. Calc. monochl.


Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: .............................................120/min.
Settings: Control Parameters: Setpoint:......................... 5.00 ppm
Settings: Control Parameters: P-band: .......................... 0.10 ppm
Parameter:............................................................... Calc. comb. cl.
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: .............................................120/min.
Settings: Control Parameters: Setpoint:......................... 5.00 ppm
Settings: Control Parameters: P-band: .......................... 0.10 ppm
Parameter:........................................................ Calc. dichloramine
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: .............................................120/min.
Settings: Control Parameters: Setpoint:......................... 5.00 ppm
Settings: Control Parameters: P-band: .......................... 0.10 ppm
Parameter:...................................................................Temperature
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: .............................................120/min.
Settings: Control Parameters: Setpoint:............................... 30 °C
Settings: Control Parameters: P-band: .................................. 1 °C
Parameter:.................................................................. Sample Flow
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: .............................................120/min.
Settings: Control Parameters: Setpoint:........................... 200 B/s
Settings: Control Parameters: P-band: .............................. 20 B/s
Parameter:................................................................................... pH
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: .............................................120/min.
Settings: Control Parameters: Setpoint:................................ 7 pH
Settings: Control Parameters: P-band: .............................. 0.1 pH
Common settings:
Settings: Control Parameters: Reset time:........................... 0 Sec
Settings: Control Parameters: Derivative Time: ................... 0 Sec
Settings: Control Parameters: Control Timeout:................... 0 Min
Settings: Actuator: ................................................ Time proportional
Cycle time: ............................................................................ 60 s
Response time: ..................................................................... 10 s
Settings: Actuator ......................................................... Motor valve
Run time: ............................................................................... 60 s
Neutral zone: ........................................................................... 5%

A-96.250.581 / 050517 109


AMI Codes-II CC
Default Values

If Function = Timer:
Mode:.................................................................................... Interval
Interval:................................................................................. 1 min
Mode: ....................................................................................... daily
Start time: ........................................................................00.00.00
Mode:.....................................................................................weekly
Calendar; Start time: .......................................................00.00.00
Calendar; Monday to Sunday:.................................................. Off
Run time: ...............................................................................10 Sec
Delay:.......................................................................................5 Sec
Signal output:............................................................................. cont
Output/Control: .......................................................................... cont
Input: Active ............................................................................when closed
Signal Outputs ........................................................................... hold
Output/Control .............................................................................. off
Fault..............................................................................................no
Delay......................................................................................10 Sec
Miscellaneous Language:............................................................................. English
Set default:....................................................................................no
Load firmware: ..............................................................................no
Password: ........................................................... for all modes 0000
Sample ID: ....................................................................... - - - - - - - -
Line break detection .....................................................................no
Interface Protocol:....................................................... depending on interface

110 A-96.250.581 / 050517


AMI Codes-II CC
Index

12. Index
A Interface . . . . . . . . ... 34
Alarm Modbus . . . . . . ... 35
pH . . . . . . . . . . . . 97 Profibus . . . . . . ... 35
Sample flow . . . . . . . 96 USB . . . . . . . . ... 36
Alarm Relay . . . . 10, 31, 94 Interpolation . . . . . . 88, 107
Application range ....... 9 Interrupt measurement . . . 10

C L
Cable thicknesses . ..... 27 Logger ......... 86, 107
Cal. History . . . . . ..... 82
Calendar . . . . . . ..... 102 M
Calibration Measurement interval . . .. 47
pH . . . . . . . 40, 53, 83 Measuring Range . . . . .. 16
Changing values . . . . . . . 44 Message List . . . . . . . .. 81
Cleaning module . . 10, 84, 88 Modbus . . . . . . . . . . .. 35
Current outputs . . . . . . . 34 Mounting requirements . .. 20

D O
Disassemble the solenoid valve 64 On-site requirements .. 16, 19
Disinfectant . . . . . . . . . . . 9
P
E Pending Errors . . . . . . . . 81
Electrical wiring ....... 19 pH Electrode . . 19, 40, 60, 82
pH gain . . . . . . . . . . . . 68
F pH Measurement . . . . 10, 82
Fill System . . . . . . . . 39, 83 pH offset . . . . . . . . . . . 68
Flow Alarm . . . . . . . . 68, 96 Power Supply . . . . . . . . 16
Fluidics . . . . . . . . . . . . 12 Process calibration . . . . . 20
Process pH . . . . . . . . 53, 83
H Profibus . . . . . . . . . . . . 36
HART ............. 36
R
I Reagent consumption. . . . 47
Input . . . . . . . . . . . . 10, 31 Relay . . . . . . . . 9, 98, 108
Instrument Setup . . . . . . 19 Relay Box. . . . . . . . . . . 10

A-96.250.581 / 050517 111


AMI Codes-II CC
Index

S U
Sample Flow . . . . 38, 82, 96 USB Interface . ........ 36
Sample requirements . . . . 16
Signal Outputs . . . . 9, 34, 88 V
Simulation . . . . . . . . . . . 84 Ver. History . . . . . . . . . . 82
Slope photometry . . . . . . 68 Verification . . . . . 51, 83, 88
Software . . . . . . . . . . . . 43
Solenoid valve . . . . . . . . 64
W
Standard pH . . . . . . 54, 83
Wire .............. 27

T
Z
Terminals . . . 29, 31–32, 35
Tube numbering . . . . . . . 63
Zero photometry ....... 68

112 A-96.250.581 / 050517


AMI Codes-II CC
Notes

13. Notes

A-96.250.581 / 050517 113


AMI Codes-II CC

SWAN
is represented worldwide by subsidiary companies
and distributors.

cooperates with independent representatives


all over the world.

SWAN Products
Analytical Instruments for:

High Purity Water

Feedwater, Steam and Condensate

Potable Water

Pool and Sanitary Water

Cooling Water

Waste Water and Effluents

Made in Switzerland

114 A-96.250.581 / 050517

You might also like