PHAR229 Final Notes
PHAR229 Final Notes
Introduction to Optics
Spectroscopy: Study of the interaction of electromagnetic radiation (based on wave or particles) with matter this interaction produce
,
The energy associated with a given segment of the spectrum is proportional to its frequency.
Spectroscopy: Study of the interaction of electromagnetic radiation (wave or particles) with matter
Certain electromagnetic radiation phenomena:
I
Transmission: Propagation of electromagnetic waves through a medium without significant absorption
or scattering, maintaining its original direction and frequency
Reflection: Bouncing back of electromagnetic waves when they encounter a boundary between two
different media
Refraction: Occurs when waves pass from one medium to another, causing a change in direction and
Based on wave
a
speed due to the change in the medium's refractive index
description Diffraction: Bending and spreading of electromagnetic waves as they encounter an obstacle or pass
through a narrow opening, resulting in interference patterns
Interference: Occurs when two or more electromagnetic waves superpose and interact with each other
Constructive interference, where the waves reinforce each other
Destructive interference, where the waves cancel each other out
Scattering: Redirection of electromagnetic waves in different directions when they interact with
particles or irregularities in a medium, causing waves to scatter in various angles, changes in intensity,
frequency, and polarization
Polarization: Orientation of the electric field vector of an electromagnetic wave (can be linearly
polarized, circular or elliptical)
Based on quantum mechanics (photon):
ΔE = h x f = (h x c) / Λ
1) Atoms, ions and molecules exist in discrete energy states only
E0 (grounded), E1-E2-E3… (excited) states
2) When an atom, ion or molecule changes energy state, it absorbs or emits energy equal to the
energy difference (ΔE = E1 - E0)
3) Wavelength or frequency of radiation absorbed or emitted during a transition proportional ΔE
Sub-electronic energy levels (Vibrational and Rotational energy levels)
Emission: Energy of photon is released (sending out gas, heat, light etc.) by the atoms whose electrons make a
transition between two energy levels
Chemiluminescence: Emissions of species excited by chemical reactions
Photoluminescence: Emissions of photons in excited states produced by radiational activation such as
light source (fluorescence, phosphorescence)
Luminescence: emission from cool bodies or emission from hot bodies that is not due to thermal
excitation (e.g. electricity, laser etc.)
Spectroscopic Measurements
A= Absorbance Each chemical has different absorption
𝜀= Molar absorptivity coefficient (M-1cm-1)
b= cell width (cm)
c= concentration (M)
T= fraction of light transmitted
I= Transmitted intensity of light beam
Io= Initial intensity of light beam
Po > P1
Absorbing
of
solution
concentration C A
1) Line spectrum: Radiating species are individual atoms/ions that are well separated
2) Band spectrum: Produced in spectral sources due to presence of gaseous radicals or small molecules,
arising from numerous vibrational levels superimposed on the ground state electronic energy level
3) Continuum spectrum: Produced when solids such as carbon or tungsten heated to incandescence
Firsis
Blackbody radiation: Produced by innumerable atomic and molecular oscillations excited in
the condensed solid due to thermal energy
ΔE = ΔEelectronic + ΔEvibrational + ΔErotational
For each electronic state , vibrational state
. For each vibrational state rotational state
many ,
many
Atomic Spectroscopy
Principle: Interaction of electromagnetic radiation with “free” atoms (isolated atoms in gas phase) to observe
specific “spectral lines” (discrete lines of light) produced after an atom undergoes excitation/de-excitation
Aim: Measuring wavelength and intensities of radiation absorbed or emitted by atoms to identify chemical
elements in a sample and their concentrations
Atomic excitation: Absorbed energy from the light causes atom’s electrons to move to higher energy levels
Atomic de-excitation: Atom releases its energy and electrons move to lower energy states
Types of Atomic Spectroscopy:
1) Atomic Absorption Spectroscopy (AAS)
2) Atomic Emission Spectroscopy (AES)
3) Atomic Fluorescence Spectroscopy (AFS)
logTc
C
pripendecular
Atomic Spectroscopy Experiments: each step influences accuracy and precision of result.
Sample intro Forming gas phase atoms Excitation/Emission Detecting photons Readout
Light source: Hollow Cathode Lamp
Advantage sharp lines specific for element of interest Ionizes inert gas to high potential
:
1) Types of flame: Selection of right region in flame is important for optimal performance
Primary combustion zone: Not in thermal equilibrium, not used
Interconal region: Highest temperature; often used in spectroscopy be , can narrower
Outer cone: Cooler region rich in oxygen due to surrounding air; gives metal oxide formation
2) Electrothermal atomization: For really small
samples
Sample is contained in a heated, graphite furnace which is heated by passing an electrical current
To prevent oxidation of the furnace, it is sheathed in gas (Argon usually)
No nebulization, sample is introduced as a drop
3) Plasma atomization:
Higher temperature values can be reached, which allows efficient atomization
Samples can be introduced via nebulization, electrothermal methods or ablation (removal)
Then, samples are transported into plasma with argon
Molecular Spectroscopy
1) UV-Vis spectroscopy
Analytical signal originates due excitation of an atom/molecule by UV or visible radiation (190-700 nm)
Absorption spectrums obtained, showing which particular wavelengths (Λ) of light the chemical species
can absorb
Atoms and molecules can absorb electromagnetic radiation, but only at certain energies and thus Λ
Electrons in bonding/non-bonding (lower energy) molecular orbitals undergo a transition to anti-
V bonding (higher energy) molecular orbitals
σ σ* transitions (alkanes):
Energy required is large
Absorption max. around 125 nm which are never observed in ordinarily accessible UV region
n σ* transitions (O, N, S, halogens):
Saturated compounds containing atoms with unshared electrons
Require less energy than σ σ* type
Radiation between 150 – 250 nm (most absorption peaks appear below 200 nm)
Molar absorptivities (𝜀) for peaks in n σ* are low to intermediate (100 – 3,000 M-1cm-1)
n π* & π π* transitions (carbonyls & unsaturated FGs):
Mostly preferred, because energies required for these processes bring absorption peaks into
experimentally convenient spectral region (200 – 700 nm)
Molar absorptivities (𝜀) for peaks in n π* are generally low (10 – 100 M-1cm-1)
Molar absorptivities (𝜀) for peaks in π π* fall in between 1,000 – 10,000 M-1cm-1
Absorption with d and f electrons:
Most of the transition metal ions absorb in UV or visible region of the spectrum (200-700 nm)
This type of transitions create very bright colors that can be observed under UV-VIS
Transition metal ions such as Ni2+, Cr2O72-, Cu2+, Co2+
Charge-Transfer absorption:
Molar absorptivities of these species are very large (𝜀max > 10,000)
Complexes that exhibit charge-transfer absorption are called “charge-transfer complexes”
Fe(SCN)2+, Fe(phenyl)32+, Starch-I5-
Solvent effect: selection of appropriate solvent is very important
Intensely absorbing compounds must be examined in dilute solutions, so that meaningful light
energy can be received by the detector, and this requires the use of completely transparent
(non-absorbing) solvents
Most commonly used solvents are water, ethanol, hexane and cyclohexane
Min. wavelength limits are especially important in UV-Vis, HPLC, CE etc. (used in detectors)
Qualitative/Quantitative analysis by UV-Vis Spectro:
Λmax is used where the UV absorbance is greatest
Different standards with different C values are prepared which provide different A values
y=mx+b equation is obtained from A vs. C graph and used to find concentration of the
unknown sample
2) IR spectroscopy the measurement of infrared interaction with matter by absorption
is ,
reflection used to
emission , or .
study
It
identify chemical substances
is and or functional group in solid , goe,
Every molecular compound has a unique infrared absorption spectrum, thus exact match between the liquid Form.
spectrum of known compound with the spectrum of unknown analyte directly identifies the analyte
No two molecules give exactly the same IR spectrum (except enantiomers)
Less satisfactory in quantitative analysis than UV-Vis due to lower sensitivity and deviations from Beer’s
Law, and absorbance measurements are less precise
Looking for presence/absence of functional groups:
Polar bond is usually IR-active
Non-polar bond in a symmetrical molecule will absorb weakly or not at all
IR does not cause electronic transitions, but induce transitions in vibrational and rotational states -M
within the ground electronic state of the molecule (stretching & bending vibrations)
UV-Vis related to wavelength, whereas IR related to wave number
Each IR absorption peak defines different type of stretching or bending vibrations
Instruments for IR spectroscopy: Dispersive instruments, FT-IR (mostly used), Filter Photometers; all
of which are different devices with different designs
Haga "A
ory
* 3) Fluorescence Spectroscopy
Used for identification of species having fluorescence property *
Following of the analyte is easier in FS Aug> Hong
Molecular fluorescence is measured by exciting the sample at “absorption/excitation wavelength”, and
by measuring the emission at a longer wavelength called “emission/fluorescence wavelength”
Fluorescence emission is usually measured at right angles to the incident beam
The short lived emission that occurs is called “fluorescence” whereas luminescence that is much longer
lasting is called “phosphorescence” (continuous process, not preferred)
Analysis: similar approach
Emitted electromagnetic wavelength is proportional to concentration
Calibration curve is constructed by the analysis of standard solutions
Finally, sample solution property is measured and unknown material is determined
Ion exchange resins are used to eliminate ions that would otherwise interfere with analysis
Ion exchange resins are particularly useful for chromatographic separation of ionic species
from solution
concentrating ions
very
dilute .
Chromatography
Definition: Components of a mixture are carried through the stationary phase by the flow of the mobile phase,
and separations occur based on differences among migration rates of the mobile phase components
Stationary Phase: Fixed in place either in a column or on a planar surface
shouldn't interact with each
Mobile Phase: Moves over the stationary phase, carrying with it the analyte mixture They
other
Elution: Process in which solutes are washed through the stationary phase by the movement of mobile phase
Eluent: Solvent used (carrier) that moves chemicals of the mixture through the stationary phase
Eluate: Mobile phase that exists the column (material emerging from the column)
Classification of Chromatography
According to shape: Planar vs. Column
According to mobile phase: Liquid vs. Gas vs. Supercritical fluid
According to separation mechanism:
1) Adsorption Chromatography: Type of liquid chromatography in which chemicals are
retained based on their adsorption at the surface of the support (stationary phase)
D
2) Ion Exchange Chromatography: ions held in a porous insoluble solid are exchanged for ions
in a solution that is brought into contact with the solid
3) Size Exclusion Chromatography: separates molecules based on their size by filtration
through a gel
Large particles do not enter the cell and eluted earlier; need less volume for elution
Small particles enter the cell and eluted later; need more volume for elution
4) Partition Chromatography: Separation of components between two liquid phases which
are original solvent and another film of solvent used in the column liquid-liquid chromatography
Plate Theory:
-15903 .2 , &ni Chromatographic column contains large number of separate layers called theoretical plates
·is so i sti Separate equilibrations of the sample between stationary and mobile phase occur in these plates
,
The thinner the plate, the higher the number of plate, the more effective the separation
Variables that influence the Plate Height (H in mm)
Particle size: Less sized column packing material less plate height
Film coat thickness: Less film coat thickness less plate height
control it
by adding organic solvent Viscosity: Mobile phase viscosity can be decreased to change interactions of the substances
Temperature: Increased temperature Reduced band broadening effect interaction of the substance with mobile and
solid phase
Flow rate: H increases with increased linear flow rate (cm/s) .
Distribution Constants:
All chromatographic separations are based on differences
in extent to which solutes are distributed between mobile
and stationary phase ( A-mobile A-stationary)
Distribution constant value is used to understand the
analysis system, but determining the concentarion of
each matter in both phases is very hard
Retention Time (tR): Time between injection of a sample and appearance of a solute peak at the detector of the
chromatographic column
Dead Time (tM): Time it takes for non-retained species to pass through the column It doesn't affect the separation but should keep
·
, we it as minimum
is solute
500 jojet, as possible. Determine it
by specific marker .
lon columnly
Retention factor (k): Compares the migration rates of the solutes on columns ( k=[ tR - tM ] / tM ) should be higher than 2
issziojesda isazioja ;a
Selectivity factor (α): α = KB / KA where B is the more strongly retained species, whereas A is the more rapidly
eluted species, thus α is always greater than unity
Number of theoretical plates (N): N= L / H where L is the length of the column packing and H is plate height;
efficiency increases as N increases and It decreases
,
Resolution (RS): Quantitative measure of the ability to separate two analytes; better when retention time
differences are higher and peak widths are smaller
Tailing factor (S): Coefficient that shows the degree of peak symmetry at 10% peak height (AS = B / A)
*
Relative Standard Deviation (RSD): Used to investigate reproducibility of the results
Gas Chromatography
Principle: Partition of molecules between mobile phase and stationary phase
Applications:
Characterization of unformulated drugs, with regards to detection of process impurities
Limit tests for impurities in drug substances
Quantification of drugs in formulations, particularly if the drug lacks chromophore
Characterization of raw materials used in synthesis of drug molecules
Characterization of volatile oils which may be used as excipients
Measurement of drugs and their metabolites in biological fluids
Strengths:
Quantitative accuracy and precision like HPLC
Much greater separation power than HPLC when used with capillary columns
Readily automated
Determines the compounds that lack chromophores
Mobile phase does not vary and not require disposal and cheap compared to organic solvents in HPLC
Limitations:
Only thermally stable and volatile compounds can be analyzed
Sample introduction is more difficult because of the small volumes of sample injected
Aqueous solutions and salts cannot be inject into the instrument
Column Oven: controls the temperature change
Isothermal: Keeps oven at one temperature, only useful for series of very similar compounds differing
in boiling points such as alcohols (not very useful in general)
Gradient: Changing temperature profile which provides better resolution and analysis time
Column types:
Packed: Filled with a coated inert solid that are chemically bonded to the support such as fire brick,
alumina and graphite
Longer column life and less bleeding (source of background noise)
Higher capacity/concentration
Low resolution and low S/N
Capillary: Coated on inside diameter of the capillary wall with film thickness of 0.1 – 5μ, where ticker
film provides better resolution but also more bleeding (typically fused silica coated with polyamide)
High resolution and high S/N
Low capacity and higher cost
Detectors: (1) Flame ionization, (2) Thermal conductivity, (3) Electron capture
Derivatization: When the analyte is not a volatile material, chemical reaction with special reagents changes the
non-volatile substance into a volatile substance
Loss of sample
Side reaction
Time consumption
&
&
V
Direct
En Dropping mercury we
it polacography
mmmanent miga
art
~
X
F Presidual current
important parameters
=
peakcurrenttial
:
&
>+