TruePath Instructions
TruePath Instructions
Triaxial Testing
TABLE OF CONTENTS
1. Introduction ........................................................................................................................................1
2. TruePathTM System Hardware Setup..................................................................................................2
2.1. Unpacking:......................................................................................................................................2
2.2. Sigma-1 Load Frame Setup: ...........................................................................................................5
2.3. GeoJac Load Frame Setup: .............................................................................................................7
2.4. TruePath System Setup:..................................................................................................................8
2.4.1. Computer: ..............................................................................................................................10
2.4.2. Network Module:...................................................................................................................11
2.4.3. Precision Variable Power Supply:.........................................................................................13
2.4.4. Load Frame Modules:............................................................................................................14
2.4.5. DigiFlow Modules:................................................................................................................19
2.5. TruePath System Power Up..........................................................................................................22
3. TruePath System Software - Installation and Configuration............................................................22
3.1. Installation ....................................................................................................................................22
3.2. Starting TruePath ..........................................................................................................................24
3.3. TruePath Status Panel ...................................................................................................................27
3.4. TruePath Menu Items....................................................................................................................28
3.4.1. Setup Menu Items..................................................................................................................29
3.4.1.1. TestNet ...........................................................................................................................29
3.4.1.2. Sensors ...........................................................................................................................35
3.4.1.3. Actuators ........................................................................................................................41
3.4.1.4. Reading Schedules .........................................................................................................44
3.4.1.5. Preferences .....................................................................................................................47
3.4.1.6. Setup ‘Save’ ...................................................................................................................48
3.4.2. File Menu Items.....................................................................................................................48
3.4.3. Tools Menu Items..................................................................................................................49
3.4.3.1. Manual Mode - Load Frame..........................................................................................50
3.4.3.2. Manual Mode – Pumps ..................................................................................................52
GEOTAC is proud to present the TruePath automated stress path system, a system designed
specifically for geotechnical testing laboratories. The result of an extensive research and
development effort, the TruePath system incorporates recent advancements in electronics and
software to make automated testing more affordable than ever. More importantly, these
advances have simplified automated testing systems while making them more robust.
Advantages for geotechnical testing include the capability to perform deformation, load, or
pressure controlled ‘closed loop’ testing. Also the immediate feed back provided by real time
plots allows the user to make critical decisions and adjustments during the course of testing.
The ‘back bone’ of the TruePath system is GEOTAC’s TestNet system that provides
automation through data acquisition and control. A TestNet system is a series of interconnected
modules that in turn are connected to a computer via a serial (COM) port.
TestNet employs Distributed Data Acquisition and Control (DDAC), a new concept that offers
significant advantages over conventional methods used for data collection and test control. A
typical data acquisition system consists of a centrally located data logger/controller to which
sensors throughout the laboratory connect. This requires long bulky cables to carry analog
signals between the data logger and sensors. Long cable runs can result in signal degradation
and contamination from electrical noise. With DDAC, data conversion hardware is placed very
close to sensor locations where analog signals are converted to digital signals. Digital signals,
which easily transmit over long distances, are sent to and from a centrally located PC compatible
computer over a single cable, which can be connected to multiple test stations to form a network.
Compared to conventional systems, DDAC is flexible, robust, and offers improved signal quality
over a longer distance with less cable clutter.
The TruePath system ships complete. All sensors, cables, and software for performing triaxial
testing are included. A detailed discussion of the system components is presented below.
2.1. Unpacking:
Upon taking delivery of a TruePathTM System, inspect crates and cartons for signs of
visible damage, then unpack (see below) all components and inspect for any concealed
damage. Report any damage found immediately to the shipper. The sooner a claim is
filed, the better the chances of restitution being made.
The load frame shipped with the system is either a SIGMA-1TM model (5K, 10K, or 40K
pound capacity) or a GeoJacTM.
Unpack the remaining cartons and use the following list and figures below to check that
all items are present:
ITEM ACCESSORIES
Sigma-1 Load Frame Power cord
Load Cell Bolt
Load Cell Bottom Adapter
Deformation Sensor
Bracket
Strain Rods
Or
GeoJac Load Frame Power cord
Power Brick
AD-IO module (4 ch.)
Serial cables (2)
DC power cable
Load Cell piston adapter
Load Cell bottom Adapter
Deformation Sensor
Bracket and post
Refer to the figure above. Place load frame on a support bench or table that is sturdy,
leveled, and make sure that the area is dry, temperature controlled, and vibration free.
The area should also have minimal electrical noise, away from compressors, generators,
air conditioning units, etc. Complete the setup as follows:
A) Level Load Frame. Place a level on the platen and adjust the leveling feet to
achieve level. Rotate the level 90 degrees and adjust the leveling feet again to
achieve level. Repeat this process until level is achieved in both directions.
Refer to the figure above. Place the GeoJac on a support bench or table that is sturdy,
leveled, and make sure that the area is dry, temperature controlled, and vibration free.
The area should also have minimal electrical noise, away from compressors, generators,
air conditioning units, etc. Complete the setup as follows:
A. Level Load Frame. Place base on a level surface.
B. Attach Deformation Sensor Post and Bracket. Screw post into one of the tapped
holes on the GeoJac top plate. Slide bracket on to post and tighten clamping knob to
hold bracket in place.
TruePath System components can be set up quickly using the figures and the
instructions below. Be sure to read the instructions important information concerning
use and care of the system is given. Note, do not power up the system until the last step.
The minimal specifications for the TruePath system computer are as follows:
IBM PC compatible, with a Pentium CPU running at 120 MHz or faster,
Windows 95® operating system, minimum 16 MB RAM, 3.5 inch floppy drive,
1.2 GB hard drive, and a Super VGA card with a monitor offering 800x600 or
higher resolution. The computer should also have at least one free serial (COM)
Place computer near load frame. Attach the AC power cord, keyboard, and
mouse. Place monitor near computer and attach power cord to monitor and
connect cable from monitor to computer. Attach female end of one serial cable
to the 9 pin serial port on the rear of the computer (next to the 25 pin male
connector). Plug AC power cord from monitor and computer into an AC outlet.
The Network module is the first of seven TestNet modules incorporated in the
TruePath system. The Network module regulates the flow of commands and
data between the computer and the load frame. The front panel is shown below.
An ‘OFF/ON’ switch on the front regulates the power to the TestNet network
and can be used to reset the network. A red LED lights up when network power
is on. Two green on the front panel indicate communication status (TX =
transmit, RX = receive).
The rear panel is shown below. It has two serial ports and one power port.
One serial cable connects the module to the computer. The male end of this
serial cable connects to the female D-sub connector labeled ‘TestNet In’. The
second serial cable connects the Network module to load frame. The female end
of this cable connects to the male sub-D connector labeled ‘TestNet Out’. The
DC power cord from the Power Supply connects to the Network module at the
connector labeled “MODULE POWER”. The power cord is terminated with 3-
pin DIN connectors. NOTE: Do not connect this cable to the Network module
until the power supply is set up properly (see below). Also, switch the Network
module to ‘off’ before connecting or disconnecting this cable.
A linear regulated DC power supply, shown above, provides logic power for the
TestNet module circuitry and sensor excitation. An AC power cord terminated
with a standard 3-prong plug connects the power supply to an AC outlet (110
VAC, 60 Hz). Plug in the power supply, allow five minutes for warm up, and
then check the voltage reading on the front panel. The power supply was set to
14.9, +/- 0.1 current before shipping. If this setting has changed, insert a small
flat bladed screwdriver through the fine adjustment hole in the front panel and
into the slot of the adjustment post. Increase setting by turning the screwdriver
clockwise. NOTE: TestNet modules can be permanently damaged if the DC
power input exceeds the 15V limit. A safe practice is to turn the Network
module off when making adjustments to the power supply settings. The current
reading should be zero if nothing is connected to the front panel.
Two additional TestNet modules, an ADIO module and a Servo module, are
associated with the load frame. On the Sigma-1 load frame, the modules are
attached to the rear panel of the load frame as shown below.
The AD-IO module provides four channels of analog data acquisition. State-of-
the-art analog-to-digital (A/D) conversion technology is used to achieve 22-bit
resolution at 80 readings per second data rate on a single channel. Conversion
rates as high as 250 readings per second can be achieved without significant loss
in resolution. The four input signal ranges, +/- 10V, +/- 1V, +/- 0.1V, and +/-
0.01V, can be independently set for individual channels. This allows ADIO
modules to handle low level (mV) signals from strain gage type sensors along
side with high level signals without any signal conditioning.
Sensors are connected to ADIO modules at the four 6-pin miniature DIN
(MDIN) sockets labeled ‘CH 1’ through ‘CH 4’. The ADIO modules can
provide either 5 VDC or 10 VDC for sensor excitation. Excitation voltage is
determined by how a sensor connector is wired and can be different for different
channels. Sensors should not draw more than 500mA from each ADIO module
and the total power consumption by all sensors should not exceed 3A for the
entire network. Sensor excitation power is stepped down from 15 VDC network
The load frame rear panel with all the proper cables connected is shown below.
An AC power cable connects to the socket on the left. This socket also contains
an off/on switch for AC power. Four sensor cables are shown connected to the
ADIO module. The external load cell is connected to CH1, internal load cell (if
present) is connected to CH2, the deformation sensor is connected to CH3, and
the temperature sensor (if present) is connected to CH4. The male end of the
serial cable coming from the Network is shown connected to the 9 pin sub-D
connector on the Servo module labeled ‘TestNet In’. The female end of the
serial cable going to the Cell pump is shown connected to the 9 pin sub-D
connector on the AD-IO module labeled ‘TestNet Out’.
If a GeoJac is being used, the Servo module is attached to the GeoJac side panel.
The AD-IO module is external. The GeoJac Side panel, with and without
connections is shown below.
The both ends of the external AD-IO module are shown below.
Each DigiFlow pump has two TestNet modules, an ADIO module and a Servo
module . These modules are attached to the pump rear panel as shown below.
The cell pump ADIO module provides four channels of analog data.
Each DigiFlow pump requires 24 VDC power to operate the motor. The Sigma-
1 load frame at the two right most connectors labeled ‘POWER OUT’ supplies
this power. If a GeoJac is being used, 24 VDC is supplied by the Power Brick.
Two DC power cables have been supplied. Plug one end of each cable into the
load frame, then plug the other end into on of the DigiFlow pumps.
The DigiFlow cell pump rear panel with all the proper cables connected is
shown below.
All modules except the Network module contain termination DIP switches (red
block of tiny switches). The purpose of termination is to minimize signal
reflections in the cable and to designate the last module on the network. The
switches must be set to ‘OFF’ on all but the last module on the network. They
must be set to ‘ON’ for the last module, which in this case is the ADIO module
on the pore pump. Servo module DIP switches have been set to ‘OFF’
Once a successful power up has been achieved, proceed to the next section for
instruction on installing the system software.
TruePath Software allows for operation of the load frame and pumps and provides acquisition
and control for stress path triaxial testing. It is a Windows based true 32-bit program offering
benefits such as familiar graphical user interface and robust preemptive multitasking that allows
other applications to run in the background.
3.1. Installation
The program is distributed on CD. Please make sure to read the ‘Readme.txt’ file
included on CD for the latest information and instructions.
B) After you install ‘Windows Installer’ and reboot the system, double click
‘TruePath.msi’ and follow instructions provided by the installation program.
C) The installation program copies program files under directory ’C:\ Program Files
\ TruePath ’on the hard disk (unless another path is specified) and creates a
program folder named ‘Geotac’ under Start | Programs menu and places a
shortcut to TruePath. It will also place a shortcut to TruePath.exe on the desktop.
You can start TruePath by either double clicking the program icon in ‘Program Files \
TruePath’ folder or selecting TruePath the the Start | Programs | Geotac menu. The
more common way of starting the program though would be to double click the short cut
icon on the desktop. Upon startup, the ‘TruePath ’ main user interface window appears
as shown in below.
Click ‘OK’ and the first of a series of windows will pop up prompting the user to enter
configuration file information. Refer to the three-ring binder supplied with the system
for entry information. This information can be entered as the windows pop up, or the
windows can be dismissed (click either ‘Done’ or ‘OK’) and the information entered
later through ‘Setup’ on the menu bar. As the configuration windows pop up, refer to the
section below on the menu bar item ‘Setup’ for instructions on entering configuration
information.
The first window to pop up after clicking ‘OK’ requires the selection of a serial port.
Use the drop down box to select a serial port then click ‘Done’. If the program has not
been activated, the following message appears:
Enter the activation code then click ‘Activate’. If the correct COM port was selected, the
remaining set up widows will pop up. If the wrong COM port is chosen, the following
message will appear.
Click ‘OK’ and to allow the remaining setup windows to appear. When finished with
these windows, return to ‘Setup | TestNet’ menu item and choose another serial port.
Repeat this process until the correct serial port is chosen. Refer to the section below
concerning the menu item Setup for additional details on identifying the correct serial
port.
Once the program is launched, you can exit the program by clicking on the [X] button in
the far right corner of the title bar or by selecting Close (Alt. + F4) from the system
menu. TruePath starts with ‘C:\ Program Files \ TruePath ’ as the default path. That
means any data files created by the program will be placed in directory ‘C:\ Program
The user interface has three components: Status Panel, Menu Bar and Tab Panels. Each
of these is discussed in the sections below:
The Status Panel occupies the left half of the screen and is visible throughout the test. It
is a graphical representation of a stress path test setup that includes important status
information. A quick look at the status panel allows the operator to assess overall status
of the system any time during the test. Rectangular numeric display fields are used for
displaying sensor readings, platen displacement, and platen deformation rate.
The Speed and Position display fields and the load frame and Flow and Volume display
fields on the pumps contain values computed based on a built in encoder, not from a
deformation sensor. This is a safety measure in that encoder information is always
available for calculations while the deformation sensor information may be out of range
if not properly set by the user. Deformation sensor readings however are available for
data reduction if needed.
A standard Window style menu is located at the top of the main user interface window.
Using menu commands the operator can enter and edit test parameters, sensor calibration
information, reading schedules, and hardware settings and constants. This information is
stored in the ‘Configuration File’. Menu commands are also used and for testing the
A/D converter and servo (motor controller). The various menu options are shown and
discussed below.
3.4.1.1. TestNet
Serial Port, Baud Rate, and module settings are all entered through the
‘Setup |TestNet’ option. Select TestNet and the ‘TestNet Setup’ window
will appear.
Name order in the module list coincides with the module location on the
network; first module (top of list) is closest to the computer, last module
(bottom of list) is last on the network (with address = 1). Modules are
automatically assigned addresses by the program according the order of
this list, so it is important that new module being added to the network
are placed at the proper location in the list.
All modules (a total of six) needed for a TruePath system are already
stored in the configuration file and shown on the ‘Module List’. These
modules cannot be removed. If for some reason additional modules need
to be added to the TruePath system TestNet network, use the following
procedure.
After clicking Add Module the ‘Add TestNet Module’ window will
appear.
Next, enter values for ‘Gain Factor’, ‘10V Excitation’, and ‘5V
Excitation’. For older ADIO modules, these values are contained on the
calibration sheet supplied with each module in the three ring binder and
need to be entered manually. For modules with serial numbers greater
than 1400, these settings are entered automatically. These modules
contain a non-volatile memory programmed with the calibration
information.
Next, select an ‘Input Range’ for each channel. The input range for any
channel should be selected to accommodate the maximum output voltage
from the sensor connected to that channel. For best resolution, use the
smallest input range available. The 0.1 V range will be suitable for most
Perform this procedure for the Load frame ADIO and both DigiFlow
pump ADIO modules.
Once all the ADIO modules settings have been checked, make sure all
cables are connected properly, network power is turned on, and
termination switches are in ON position only for the last module on the
network. Next, click Done to close the ‘TestNet Setup’ window. A
prompt will appear requiring the network to be reset.
Command buttons along the window bottom allow for adding, removing,
editing, or testing sensors. These commands can also be invoked by
right clicking on a sensor name in the list box or choosing the ‘Sensor’
menu item.
is a program feature that links the sensor list to a separate data base that
can be created and can contain information on all sensors that may be
used with TruePath or any other GeoTAC program. This makes it
possible to switch out sensors (e.g. switching load cells to accommodate
a different load range) without having to reenter all the calibration data.
Instructions for this feature are contained in a separate document.
Check the sensor information included in the three ring binder and enter
it as needed based on the item descriptions below. Note that some of the
information cannot be changed (e.g. sensor name) and that some of the
dummy information loaded at initial startup is correct and does not need
to be changed.
Click OK. Sensor information appears in the list box in the ‘Sensor
List’ window. If any sensor information is incorrect, click the sensor
name and then Edit or right click the sensor name and select Edit from
the pop-up menu. The ‘Edit <sensor name>’ window will appear.
Make required changes and click OK. Note that sensor ID cannot be
edited. If this is entered incorrectly, remove the sensor from the
database and reenter the correct information.
The new sensor is placed directly after the sensor name highlighted
when Add is clicked. Note that for new sensors, the user can enter a
‘Name’. Enter the appropriate sensor information and then click ‘OK’.
If any sensor information is incorrect, click the sensor name and then
Edit or right click the sensor name and select Edit from the pop-up
menu. The ‘Edit <sensor name>’ window will appear. Make required
changes and click OK. Note that sensor ID cannot be edited. If this is
entered incorrectly, remove the sensor from the database and reenter the
correct information.
3.4.1.3.Actuators
Load frame and pump information is added to the Configuration file by
selecting ‘Setup | Actuator’. The ‘Actuator Setup’ window shown below
should appear. It contains a tab for each actuator. The first tab is for the
cell pump.
Note: If the load frame is not a standard GeoTAC model, such as older
load frames that GeoTAC converted, check the three ring binder or call
GeoTAC for the appropriate ‘Load Frame’ parameters. Severe damage
can occur if the incorrect values are used.
Click the ‘x’ in the upper right hand corner to dismiss the ‘Actuator
Setup’ window when all the tabs have been selected and the appropriate
entries have been made.
3.4.1.4.Reading Schedules
Click ‘Reading Schedules’ and the following window will appear.
# of Interval
Readings (min)
1 0
20 .5
40 1
40 5
# of Interval
Readings (strain)
1 0
40 0.025
40 .1
40 .25
3.4.1.5. Preferences
The window also contains entry boxes for specimen diameter and height.
Enter the most commonly used values. These values will appear in data
entry boxes in the ‘Specimen Data’ window eliminating the need to type
them in.
If an internal load cell is present, the piston diameter and friction are
irrelevant. If an internal load is not being used. ‘Piston Friction’ and
‘Diameter’ are used to adjust the calculated the vertical load applied to
the specimen.
3.4.1.6.Setup ‘Save’
Note that any changes made using any Setup Menu item are not saved
until you quit TruePath. If the program terminates prematurely,
changes will be lost. Therefore, it is good practice to save through
Setup | Save immediately after any changes have been made.
When Test Data is selected, the ‘Test Data Setup’ window appears
Both of these windows are used to create a data file. Details are
discussed in a later section.
When Manual Mode menu item is selected, the ‘Manual Mode’ window
appears. The window consists of tab panels, one for each actuator. Click
the ‘Load Frame’ tab and note the two options available on the lower
tab.
The entry boxes can be used to move the platen or piston at a desired
speed and distance. Once entries are made, click one of the inner arrows
then ‘Start’. The platen or piston will move the specified distance, at the
entered speed, in the direction selected of the selected arrow. The space
bar can be used to start and stop platen movement. Note that the
distance the platen or piston will travel is reset each time the ‘Stop’ and
‘Start’ button is clicked (or space bar). In other words, it will not
complete an interrupted move; it will start a new one.
Clicking the ‘Reset’ button sets the position display box in the load
frame shown in the Status Panel to zero
In the ‘Load Control’ option, a load is entered along with a time. Upon
clicking ‘Start’, the load will ramp to the desired load in the time
entered. Alternatively, the constant load option can be selected in which
case the load frame will move as quickly as possible to achieve the
desired load Start’ (the ‘Constant Load’ option should be used with
caution to avoid over loading).
The entry boxes can be used to move the piston at a desired rate and
volume. Once entries are made, click one of the inner arrows then
‘Start’. The piston will move the specified volume, at the entered rate,
in the direction selected of the selected arrow. The space bar can be
used to start and stop piston movement. Note that the volume expelled
or drawn in is reset each time the ‘Stop’ and ‘Start’ button is clicked (or
space bar). In other words, an interrupted move will not be completed; a
new move will be initiated instead.
Clicking the ‘Reset’ button sets the volume display box next to the pump
shown in the Status Panel to zero
The calibration is performed by first filling the cell with confining fluid
(water or silicone oil), placing it in the load frame, unlocking the piston,
and then applying a small cell pressure such that the piston raises and
makes contact with the external load cell.
Click ‘Yes’ to cancel the calibration, click ‘No’ to continue. Once the
target pressure is reached, the ‘Done’ button becomes active.
Allow the cell pressure to reach its initial value before starting another
calibration or moving on to another procedure. Any subsequent piston
uplift calibrations will overwrite the previously stored calibration
factors.
Note that data collection will enter the ‘hold’ mode when ‘Maximum
Readings’ is reached. The data file remains open and data collection can
be reinitiated by choosing ‘Tools |Data Acquisition | Edit Task’ and
entering a new set of data collection parameters. Subsequent data will
be appended to the open data file. Also note that no additional readings
will be taken unless ‘number of readings’ has been increased.
To end data collection, choose ‘Tools |Data Acquisition | Close’, the
following message will appear.
Click ‘Yes’, the data file will close, and data collection will terminate.
A new data collection task can be started at any time.
3.4.4.1.Help Topics
3.4.4.2.About TruePath
Air trapped in the pumps, pressure sensor, tubing, or specimen drains can cause errors in system
volume calculations. These errors result when pump pressure is raised and air trapped in the
pump compresses and slowly dissolves in solution. The resulting volume change due to air
compression or dissolution is recorded as flow in or out of the specimen. Saturating the pumps
and associated tubing can minimize errors due to air volume change. A procedure for saturation
is outline below.
Note that this procedure involves operating the pump with the port valve closed. DigiFlow
pumps have a shutdown feature that turns off the pump motor if pressure exceeds ‘Maximum
Cutoff’ pressure entered on the ‘Edit <sensor name>’ window (accessed through ‘Setup | Sensors
| Edit’). This gives some degree of protection. However, caution is required. It takes some time
for the program to become aware of over an over pressure situation. The pressure sensor is read
only so often. If the system is very stiff (i.e. closed valve condition) pressure can exceed the
maximum sensor pressure before pump shuts down.
4.1. Connections
A typical pump setup is shown below.
If the volume reading slowly increases with time and does not stabilize, there is most
likely a leak. Check all connections and snug any fittings where water is found.
Click ‘Stop’ at the end of the pressure test, enter zero for pressure, and then click ‘Start’.
Click ‘Stop’ when the pump pressure reaches zero.
A stress path triaxial test using TruePath software consists of four stages: Seating, Back
Pressure, Consolidation, and Shear. Tabs on the right-hand side of the screen represent each
stage. Tabs become active after creating a ‘Specimen’ and ‘Test’ data file under the ‘File’ menu
bar item. Each stage has a series of instructions that guide the user through a test. A detailed
discussion of each stage along with instructions for creating a data file and test file is given
below.
Next choose ‘Test Data’ under the ‘Edit’ menu and the following widow will appear:
This window provides a convenient means for entering test control parameters in one
place. Most of these parameters appear on the tab panels and can be changed at a later
time. If a series of tests is being performed with all or mostly the same parameters, they
can be saved using the ‘Default’ menu item
.
Once saved, they will automatically appear in the entry boxes the next time this window
is opened. A description of the entry boxes is given below.
5.1.3. Consolidation.
Three types of consolidation can
be selected. If ‘Anisotropic’ is
selected, ‘Target K’ must be
entered. ‘Vertical Stress’ is the
target value for consolidation,
while ‘Strain Rate’ controls the
rate of loading. The initial
movement of the load frame is
controlled by the ‘Direction’
option.
A data file for the tests is created after the ‘Save’ button has been clicked on either the
‘Specimen Data’ or ‘Test Data’ windows. A standard Window “Save As’ window will
appear. The file is written to the hard disk when the ‘Save’ button from the remaining
window is clicked. Once the data file is written to the hard disk, the type of
consolidation and drain conditions cannot be changed. Also note that calibration
information cannot be changed once the data file is saved on the hard disk. Therefore,
all sensors should be zeroed before the data file is created.
Once the data file has been saved to the hard disk, the ‘Seating” tab becomes active.
Pressure, then remove the quick connect from the top port. Click the ‘Next’ button
to move to the next stage.
This technique must be used with caution however so that the bottom pump is not
over pressurized. Continue flowing water through the top drains until bubbles
cease to emerge. Click ‘Stop’ and close both the top inlet and vent valves. Switch
Attach the pore pressure sensor to either the top or bottom vent drain valve. Open
the valve, hold a paper towel under the pressure sensor bleed port then loosen the
setscrew.
(Note: do not remove the set screw, just loosen it a couple turns so that water and
air can flow around it). Leave the setscrew loose, open the corresponding inlet
drain valve, then click ‘start’ to initiate upward movement of the pump piston.
Stop the piston when air ceases to emerge from around the setscrew. Be sure the
pump is stopped and tighten the setscrew. Just snug the setscrew, do not over
tighten.
If the pump reaches the upper limit during flushing, switch the ‘Pore’ pump valve
to the drain/refill container and click ‘down’. The piston will move down and
water from the container will be drawn into the pump. Use this procedure to refill
the pump when flushing is complete. When flushing is complete, click ‘Done’
then click the next step.
Note that the response time for volume change can be long. This stage of seating
may be as short as 15 minutes for soft specimens that are normally consolidated
or may be as long as 24 hours for specimens that have a low hydraulic
conductivity and are highly over consolidated. Sufficient time should be spent at
each cell pressure to ensure that the plots show the correct trend.
The Backpressure Tab is used for saturating the specimen using backpressure. The
process involves raising the specimen pore pressure (backpressure pump pressure) while
simultaneously maintaining a constant difference (effective stress) between the cell
pressure and pore pressure. By maintaining a constant effective stress, the specimen
should not undergo a volume change. An important criterion for this process is that the
pore pressure be increased at a rate slow enough to allow pore pressure equalization
throughout the specimen.
Saturation is achieved by two processes. The first process involves compressing air
bubbles due to the pressure increase, Boyle’s Law. This process occurs in a matter of
seconds. The second process is the dissolving of air into solution, Henry’s Law. This
process is time dependent.
Once pore pressure increase with time ceases, click ‘Done’ to return to the
‘Backpressure’ tab. The cell pressure will return to its pre ‘B’ check value (this
Note: The ‘B Value Check’ window can be dismissed by clicking ‘x’ in the
upper right corner. However, this procedure is not recommended. If the cell
pressure has not returned to the pre ‘B’ check value, any action that is initiated,
(stepping to a higher back pressure value or initiating consolidation) will
maintain the current value of effective stress which will be different from that at
the start of the ‘B’ check.
Step loading can be achieved by using a fast ‘Stress Rate’. Consolidation can be
stopped and restarted at any time. Parameters can be changed after the ‘Stop’
button is clicked.
Once the ‘Target Effective Stress’ has been reached, the system enters a creep
mode. Cell pressure and backpressure are maintained while specimen volume
change can continue. Specimen equilibrium is determined by viewing plots
available through the ‘Graph’ button. Additional consolidation or unloading can
be accomplished by clicking ‘Stop’, entering a new ‘Effective Stress’ and
clicking ‘Restart’.
To end the consolidation stage, click ‘Stop’, then ‘Done’ to activate the ‘Shear’
Once the target vertical effective stress has been reached, the system enters a
creep mode. Cell pressure, backpressure, and vertical load are maintained while
specimen volume change can continue. Specimen equilibrium is determined by
viewing plots available through the ‘Graph’ button. Additional consolidation or
To end the consolidation stage, click ‘Stop’, then ‘Done’ to activate the ‘Shear’
5.4.3. K0 Consolidation
If ‘K0’ was selected, the following window will appear.
Once the target effective stress has been reached, the system enters a creep
mode. Cell pressure, backpressure, and vertical load are maintained while
specimen volume change can continue. Specimen equilibrium is determined by
viewing plots available through the ‘Graph’ button. Additional consolidation or
unloading can be accomplished by clicking ‘Stop’, entering a new ‘Vertical
Effective Stress’ and clicking ‘Restart’.
To end the consolidation stage, click ‘Stop’, then ‘Done’ to activate the ‘Shear’
TruePath allows for drained (CD) or undrained (CD) conditions during the shear stage.
For each case, shearing can be accomplished through compression or extension. Details
of the various options are discussed below.
Click ‘Graph’ to view plots and monitor shear. Or click ‘Stop’ and the buttons
change to:
When shear is stopped, the platen stops and cell pressure is held constant. Entry
box parameters can be change and shear reinitiated by clicking ‘Restart’.
Testing can be terminated by clicking ‘Done’.
Check entry boxes to make sure proper the ‘Strain Rate’ has been entered.
Select Enter limits for both the ‘Maximum Vertical Effective Stress’ and
‘Maximum Strain’. The load frame platen or piston will stop when either of
these limits are reached. Open any valves between the specimen drain lines and
the backpressure or pore pump. Check that the valve between the pore pressure
sensor and drain line is open.
Click ‘Graph’ to view plots and monitor shear. Or click ‘Stop’ and the buttons
change to:
To import the file in Microsoft Excel, launch Microsoft Excel, go to File | Open and then
select ‘All Files (*. *)’ in the file selection box. Browse to the folder that contains the data file.
The Open dialog box should list all available files. Select the data file you want to import and
click Open, the text import wizard will pop up. In step 1, for “Original Data Type” check
Delimited, in step 2 select Tab as “Delimiters”, in step 3 check General as “Column Data
Format”, and click Finish.
Refer to the section concerning calibrations factors for instruction as to how to convert the raw
data to engineering units.
To see plots, click ‘View Plots’ and the ‘View Graphs’ window will appear.
Menu Items
Two floating cursors on the graph can
be manipulated using the mouse to
obtain coordinates of any two points in
the plot space. The cursors can be free
floating or can me made to snap to data
points for precise measurements by
selecting appropriate options from
‘Cursor Behavior’ menu.
Plot properties window allows user to set axis limits, number of divisions, and plot
results on a logarithmic or inverted axis. A convenient auto-scale feature automatically
sizes any axis to fit available data. Settings selected through ‘properties window’ are
stored in a settings file named ‘Graphs.ini’ in the program directory and are recalled next
time the program is launched.
The ‘View Graphs’ window can be dismissed by clicking the ‘x’ in the upper right hand
corner or by clicking any other button on the lower task bar.
The calibration factor (CF) for GEOTAC systems is defined as the magnitude of physical input
applied to a sensor that would produce 1 volt of signal output while using 1 volt of excitation. In
terms of an equation CF can be expressed as follows:
CF = P /(Vs /Ve)
The following is a typical example of how to calculate CF. A load cell manufacturer provides a
specification sheet that lists the sensitivity of a 2000-pound capacity load cell as 3.296 mV/ V.
This means that the output signal from the load cell will equal 3.296 mV when 1000 pounds is
applied to the load cell that has an excitation voltage of 1 volt. CF in this case is determined as
follows:
P = 2000 lbs.
Vs = 3.296mVs x 1 Vs / 1000 mVs = 0.003296 Vs
Ve = 1 Ve
CF = 2000lbs / (0.003296 Vs / 1 Ve)
CF = 60,6796.116 lbs / Vs / Ve
So in this case, applying 60,6796.116 pounds to the load cell would produce 1 V of output if an
excitation voltage of 1 volt were used.
In this case, calibration factor can be obtained by performing linear regression analysis on the
data set using Microsoft Excel® or similar spreadsheet program. Select sensor output as the X
input range and applied load as the Y input range for regression analysis. The calibration factor
is calculated by multiplying the slope (X Variable 1 from regression output), by the excitation
used during calibration. The slope for the above example is 107,321.6 lbs / Vs, giving a
calibration factor of 1,082,252.702 lbs/(Vs/Ve).
P / Vs = 107,321.6 lbs / Vs
Ve = 10.0842 Ve
CF = 107,321.6 lbs / (1 Vs / 10.0842 Ve)
CF = (107,321.6 lbs/ 1 Vs) x 10.0842 Ve
CF = 1,082,252.702 lbs/(Vs/Ve)
The following observations may be helpful in understanding CF. First, CF may seem much larger
than other conversion constants you may have used in the past. The reasons are twofold. One is
A second observation is that in calculating CF, the signal voltage is normalized by the excitation
voltage, Ve. This normalization makes it convenient to account for changes in excitation voltage
for ratiometric sensors. A ratiometric sensor produces a signal that is directly proportional to
excitation voltage. Most strain gauge sensors are ratiometric. A load cell that is excited by 5 V
and produces a 0.012 V signal for a particular load would produce a 0.024 V signal if excitation
was increased to 10 V for the same load.
The following is an example of how CF is used to calculate the corresponding physical excitation
for a measured signal voltage. The above load cell will be used in this example.
A signal voltage (Vs) of 0.027663 volts is measured from the above load cell while using an
excitation voltage of 9.9546. A ‘zero’ sensor signal voltage (Vso) of 0.000087 volts was
measured previously when no physical loading was applied to the load cell. The load (P) applied
to the load cell is calculated as follows: