Startup Instructions
Register 4 English
Print No.: RXB1-140.034.11.01.02 Doc. Gen. Date: 02.98
Replaces: n.a.
0-2 Revision
Page
1 _______Requirements/Notes ____________________________________________ 1- 1
Safety Notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1- 1
Tolerances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1- 1
Required Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1- 1
Required Measuring Instruments, Tools and Aids. . . . . . . . . . . . . . . . . . . . 1- 1
Text Emphasis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1- 2
Important Remarks Regarding Startup . . . . . . . . . . . . . . . . . . . . . . . . . 1- 3
Abbreviations Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1- 3
Definition of Terms for Cassettes and X-radiation . . . . . . . . . . . . . . . . . . . . 1- 4
Remark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Checking the Operating Voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Generator Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Page
Coincidence of Radiation Field and Center of Film (Bucky B) . . . . . . . . . . . . . 5 - 11
Checking the IONTOMAT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 12
Checking the Diamentor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 12
Checking the Grid Blur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 12
Measuring Ground Wire Resistance . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 14
Safety Notes 1
While carrying out all work steps and checks, the product-specific safety notes contained
in the instructions, as well as the general safety notes contained in Register 2 of the TI
binder must be observed.
Tolerances 1
These tolerances apply to all dimensions indicated in the present instructions, if no other
tolerance is stated expressly behind the value.
A tolerance of ± 10% is permissible for torque data.
Required Documentation 1
• Installation certificate
• Test Protocol for Direct Exposure Equipment: RXD0-000.037.05...
NOTICE All tools, measuring instruments and aids are listed and specified
in the ARTD (Part 3), with the exception of the "standard installa-
tion tools".
Text Emphasis 1
WARNING All texts marked with "WARNING" call attention to potential risks
to health or life.
All values marked with the symbol ” ” must be transcribed into the supplied MULTIX
PRO Test Protocol for Direct Exposure Equipment (the test protocol is located in
Register 9 of the Subsystem Binder).
These measurement values can be used for the Acceptance Test per Röntgenverordnung
§16 which is mandatory in Germany or other country-specific regulations.
CAUTION Checks or adjustment that must be carried out under radiation are
labeled with the radiation warning symbol .
While adjustment labeled with this symbol are being carried out,
radiation protective clothing must be worn.
Abbreviations Used 1
The terms ”length” and ”width” used in these instructions, as well as the orientation of
inserted cassettes should be understood as illustrated in the sketch:
Length
Tabletop
24 cm x 30 cm cassette
*1
Width X
30 cm x 24 cm cassette
Patient
*1 The first number always
indicates the width!
Fig. 1
Remark 2
From the Test Protocol, it can be seen at which line voltage the MULTIX PRO system was
tested. The standard connection is 2-phase, 400V, 50/60 Hz. If an adaptation needs to be
made to the local line power, proceed according to the points in the wiring diagram
marked with an asterisk (*). The Test Certificates are filed in Register 9 of the System
Binder.
Generator Startup 2
• Put older generators (24 V - KK connection) into operation per the on-site
documentations in the same sequence as above.
• After startup of the generator has been completed, continue with Chapter 3 of these
instructions.
Fig. 1
Select Drive: A
Window
"MULTIX Service Installation
is displayed 2 x
Continue
Create
OK
1
2
0041
0042
0043
0044
5
• Close the File Manager.
0000
2
0001
/DQJXDJH
1a
• Exit the MULTIX service software and reselect the MULTIX service software.
• Continue with Point 2.
3
5
4
0002
6 7
0003
10
11
0005
A
12 13
B
14 15 16 17 18
C
19 20
D
21 22 23 24 25
E
26 28
27 29
F
30 31 32
33
A System model:
12 MULTIX TOP 13 MULTIX PRO
14 Planigraphy (Tomography)
C Generator interface:
19 XCS interface 20 KK - 24 V interface
29 Catapult Bucky with fixed grid (only with CX Bucky wall stand)
F Ceiling stand:
30 Motorized lift - telescope
34
NOTICE Every time the system is switched on again during startup, the
system configuration must be confirmed with "ok" to allow the
Adjustment Mode to be selected.
35
0008
Generator Table 4
NOTICE The check of the generator delay time that has been entered is
carried out during the "Checking Grid Blur".
The kick switch for the table can be configured as a "dead man’s switch" or as a sensor.
• To do this, select the following procedure per the service software:
- Function groups
- Host table
- Adjustment of the table key type
When the tabletop brake is released, a buzzer sound can be activated or deactivated.
• To do this, select the following procedure per the service software:
- Function groups
- Host table
- Adjustment of the buzzer status with floating tabletop
The types of cassettes used on-site and the SID displays can be uniformly programmed in
"cm" or "inch".
• To do this, select the following procedure per the service software:
- Function groups
- Collimator
- Adjustment of the display
• Configure the Bucky wall stand in accordance with the installation location. Select the
following procedure as prompted by the service software:
- Function groups: Stand
- Adjustment: Adjustment of the Bucky Wall Stand alignment
- Mark the Bucky wall stand configuration and exit the window with "OK".
Logoff 4
CAUTION After selecting logoff, "F1" must appear in the SID display. While,
this is being displayed, the system may not be switched off.
0037
0039
Fig. 1
Release all the brakes described below at the control unit (1/Fig.1).
• Press the key; the longitudinal brake is released.
The movement forces must be: ≤ 40 N
• Press the key; the vertical brake is released.
The movement forces must be: ≤ 40 N
• Press the key; the rotation brake is released.
Rotation movement must be smooth and easy (no noise).
• Carry out the checks of table movements per the MULTIX PRO Operating Instructions.
• The EMERGENCY STOP button (not configured with the MULTIX PRO-FIX) blocks all
motorized movements.
- To release it, turn the button 1/4 turn to the right.
Collimator
centring
cross
Light field
Fig. 2
• Check the coincidence of the line light localizer to the centring cross.
• Set a vertical SID = 115 cm.
• Switch on the line light localizer and open the collimator all the way.
• Check the coincidence of the line light localizer to the centering cross (see Fig. 2). Move
the tube assembly between the min. and max. SID, when this is done, the line light
localizer alignment to the collimator centering cross must remain centered over the
entire SID range.
Fig. 3 Fig. 4
• Position the tube assembly with a vertical beam path approx. 1 m above the floor.
• Switch on the light localizer and mark the position of the light field on the floor.
• Use the knobs (1/Fig. 3) to collimate a narrow light field in the longitudinal direction.
• Switch on the light localizer and move the tube unit to the max. SID; when doing this,
check the drift of the light field. Drift may be a max. of 5 mm.
Correction:
• Remove the tube assembly housing (1/Fig. 4) (2 paneling screws).
• Loosen the 2 Allen screws (2/Fig. 4) approx. 2 turns.
• Loosen the two set screws (3/Fig. 8).
• Adjust the tube assembly on the longitudinal axis.
• Retighten the 2 Allen screws and set screws.
• Check the light field drift.
• Reinstall the tube assembly housing.
3 2
Fig. 5 Fig. 6
• Position the tube assembly with a vertical beam path approx. 1 m above the floor.
• Switch on the light localizer and mark the light field on the floor.
• Use the knobs (2/Fig. 5) to collimate a narrow light field in the transverse direction.
• Switch on the light localizer and move the tube assembly to the max. SID; when doing
this, check the drift of the light field. Drift may be a max. of 5 mm.
Correction:
- Remove the cover panel on the tube assembly support arm (1/Fig. 6).
- Loosen the M6 nut (2/Fig. 6).
- Adjust the eccentric bearing (3/Fig. 6); to do this, release the rotation brake at the
keyboard.
- Retighten the nut (2/Fig. 6).
- Reinstall the cover panel.
• Mark the center of the tabletop in the longitudinal direction (e.g. using a piece of string
stretched across the table).
• Switch on the light localizer.
• Adjust the line localizer centering cross to the mark.
• Move the column over its complete movement range and determine if there is any drift in
the longitudinal direction (tolerance: less than/equal to 5 mm).
Measuring edge
for tape measure
15
1 m
Fig. 7 Fig. 8
NOTICE Document the following adjustments in the MULTIX PRO Test Pro-
tocol (Point 5.2).
• Use the tape measure in the collimator to set an SID of 109.6 cm to the top surface of the
MULTIX PRO tabletop. (The distance from the film to the top surface of the tabletop is
5.4 cm).
• Affix the SID adhesive label (1.15 m) and the marker line from the accessory pack as
shown in Fig. 7.
• Adjust the tube assembly to the 1.15 m SID marker as shown in Fig. 7.
• Use the tape measure in the collimator to measure to the top surface of the table top and
add 5.4 cm (distance: top surface of tabletop- film).
• Enter the display and the measured values in the Test Protocol under Point 5.2, Bucky
A, (MULTIX table).
- Determined the difference and enter this value in the Test Protocol.
- The difference may not exceed 17 mm.
• In the Test Protocol, Point 5.2 check off the field "Checked during startup".
2m
2 00
m
Fig. 9 Fig. 10
• Position the tube assembly to the Bucky wall stand as shown in Fig. 9.
• Use the tape measure in the collimator to set an SID of 195.9 cm to the front panel of the
Bucky wall stand (the distance of the film to the top surface of the front panel is 4.1 cm).
• Affix 2.0 SID adhesive label and the line marker (from accessory pack) as shown in
Fig. 10.
• If there are other SID’s, carry out the adjustment as described above.
NOTICE Document the following adjustments in the MULTIX PRO Test Pro-
tocol (Point 5.3).
• Set the tube assembly to the 2.0 m SID marker as shown in Fig. 9.
• Use the tape measure in the collimator to measure to the front panel of the Bucky wall
stand and add 4.1 cm (distance: front panel- film plane).
• Enter the value in the Test Protocol under Point 5.3, Bucky B (Bucky wall stand) under
FFA1/SID1.
Max. Difference
SID
(corresponds to 1.5% of SID)
100 cm 15 mm
115 cm 17 mm
150 cm 22 mm
200 cm 30 mm
300 cm 45 mm
36’’ 14 mm
40’’ 15 mm
48’’ 18 mm
72’’ 27 mm
• In the Test Protocol, Point 5.2 check the field "Checked during startup".
Length
Light field
Marker
Side marker
Fig. 11
NOTICE The beam geometry has already been completely adjusted at the
factory and documented in the MULTIX PRO Test Protocol using
the supplied test exposures. By making test exposures, make
sure that none of the settings has changed. Document the follow-
ing adjustments in the MULTIX PRO Test Protocol.
Y1
Film
Light field
X2 X1
Radiation field
Y2
Fig. 12
Film format
Radiation field
∆
Metal washer
Center point of cassette
Side marker
Fig. 13
Procedure:
• Position the central beam so it is vertical, center the tube assembly to the cassette tray
marker and set the SID to 115 cm.
• Place a center marker on a 24x30 cm (10’’x12’’) cassette (tape on a metal washer) and a
second marker as a side marker (Fig. 13).
• Load the marked cassette with film, insert it into the cassette tray and insert it up to the
stop.
• Use the knobs on the collimator to set a format of 18 cm x 24 cm (8’’ x 10’’) per the scale
(SID = 115 cm).
• Trigger an exposure ( ; 50 kV; 5 mAs) and develop the film.
Evaluation:
• Evaluate the center deviation ∆
per Fig. 12 and document this value in the Test
Protocol, Point 4.2 "Checked during startup" (Checking Center of Radiation Field Center
and Film Center, Bucky A).
- The deviation may not exceed 17 mm at an SID of 115 cm.
• In the Test Protocol, Point 4.2 check off the field "Checked during startup".
Film format
Radiation field
∆
Metal washer
Center point of cassette
Side marker
Fig. 14
• Carry out the check of the IONTOMAT per the supplied startup instructions for the
particular generator
POLYDOROS LX 30/50 RX63-020.034.13...
POLYDOROS SX 65/80 RX63-055.034.01...
POLYDOROS LITE RX63-020.034.16...
System Configuration with the Service PC for
POLYDOROS LX 30/50 LITE and POLYDOROS SX 65/80 RX63-020.034.15...
per the Chapter "Startup and Function Check" and document this in the Test Protocol for
Direct Exposure Equipment, RXD0-000.037.05...
• After the check of the IONTOMAT is completed, continue with the following paragraph in
these instructions.
With the POLYDOROS SX 65/80, carry out the check of the Diamentor per Startup
Instructions, RX63-055.034.01..., Chap. "DIAMENTOR MS".
With the POLDOROS LX 30/50 and POLYDOROS LITE, only an external Diamentor can
be connected, and carry out the check per the manufacturer’s instructions.
After completing the check of the DIAMENTOR, continue with the following paragraph in
these instructions.
For each application device with a moveable grid, check at which exposure time tg grid
lines can no longer be seen on the film.
Procedure:
• At the generator, select small focus, 80%, 40 kV and middle IONTOMAT chamber.
• Set an SID of 115 cm.
• Insert a 18x24cm cassette without film.
• Trigger exposures beginning at 40 kV until the exposure times of ≥ 3 ms result
(increase the kV value -> exposure time becomes briefer).
• Insert a 18x24cm cassette with film.
• Trigger the exposure with the previously determined kV value.
• Use a magnifier at the light box to determine at which switch time grid lines can no longer
be seen on the film.
• If the grid delay needs to be changed, carry out the test setup and the measurement as
indicated in Fig. 15. The generator delay time can be changed using the Service PC per
Chapter "Programming the System".
• Document the switch time in Chapter 8 of the "Test Protocol for Direct Exposure
Equipment", RXD0-000.037.05... for the particular unit.
- Grid lines for exposure times of ≥ 3 ms are not allowed.
NOTICE The generator delay time must be determined for the generator
being used and the catapult Bucky programmed with this value;
see bottom diagram. Only then are exposures free of grid lines
beginning at 3 ms over all time ranges.
Top diagram (Fig. 15): Time of grid movement with the time point of radiation.
Bottom diagram (Fig. 15): Release signal relative to grid movement.
Exposure !
Programmed generator delay time
corresponds to the actual measured
Grid path generator delay time of the generator
being used
Turn-around
point
40 ms
Middle position
Grid start
Park position
180 ms 260 ms t
Channel 1 Channel 2
Radiation release
Generator delay time
U1 measured
Hardware signal
radiation release Physical exiting of
radiation
Channel 1: Test point on backplane D54 X109 (U1) and X110 (ground)
Channel 2: Measure radiation with the measuring chamber, Part No.: 96 60 754 RE999.
Fig. 15
Unit
Monitor
Cabinets
Transformers
Consoles
Tube assembly
Collimator
etc.
Fig. 16
Löscher/TD PS 21
Bernhagen, Tropia/TD SD 31