Stephanie Technical Manual V2.2 - en
Stephanie Technical Manual V2.2 - en
Pediatric Ventilator
Technical Manual
Vorwort
F. Stephan GmbH
- Medizintechnik -
Kirchstrasse 19
56412 Gackenbach
Table of contents
Table of contents
Table of contents ........................................................................................3
1 Basic principles..................................................................................7
1.1 Structure...................................................................................7
1.1.1 Control unit module ....................................................8
1.1.2 Electronic module .......................................................8
1.1.3 PC module .................................................................12
1.1.4 Electric power supply module...................................13
1.1.5 Battery module ..........................................................14
1.1.6 Pneumatics module ...................................................15
1.1.7 Patient component .....................................................16
1.1.8 Test module ...............................................................16
1.1.9 Valve actuation..........................................................17
1.2 Functional operation ..............................................................18
1.2.1 Gas supply .................................................................18
1.2.2 Electric power supply................................................20
1.2.2.1 General description ..........................................20
1.2.2.2 Transformer......................................................20
1.2.2.3 Generating supply voltages ..............................20
1.2.2.4 Charging circuit, switch-over to battery
operation...........................................................21
1.2.2.5 Heating control.................................................22
1.2.2.6 Control .............................................................22
1.2.2.7 Power failure ....................................................24
1.2.2.8 Safety controls..................................................24
1.2.3 Generation of the ventilation pattern.........................24
1.2.4 Heating and humidifying ventilatory gases...............25
1.2.5 Monitoring.................................................................25
1.2.6 Control.......................................................................26
2 Mechanical structure........................................................................27
2.1 General...................................................................................27
2.2 Assembly and disassembly ....................................................28
3 Pneumatics .......................................................................................31
3.1 General...................................................................................31
Table of contents
1.1 Structure
1 Basic principles
1.1 Structure
Using the modular structure, this section will briefly deal the individual
functional modules and their interaction.. Detailed descriptions and
explanations of the individual functionalities of the modules are given in
the sections that follow.As shown in the Appendix (modular display,
Stephanie), theStephanie consists of several components or modules :
The control unit module is positioned on the front side of the Stephanie
unit. It is lectern-shaped and is comprised of all the adjustment and
settings elements in the form of switches, potentiometers and keys (push-
buttons). Furthermore, these are grouped for the various therapy settings
and for monitoring. The control unit module, ie. the control elements,
transmits the information to the central controller in the electronic block
module.
1.1 Structure
The sensor circuit board has two pressure sensors for regulating and
Sensor circuit board monitoring ventilatory pressure, one differential pressure sensor that,
together with the pneumotachograph head, carries out volume
measurement, and three amplifiers for evaluating thermistor signals for
ventilatory-gas humidifying.
Microcomputer The Microcomputer (MC) is the essential core of the unit controlling all
therapy functionalities provided by the Stephanie unit. It processes the
information given via the control module received from the adjustment
and settings elements and transfers these on to the serial port interface RS
232 to the PC module.
Fig. 6: Microcomputer
1.1 Structure
Power amplifier The power amplifier amplifies the analogue signal generated by the
microcomputer which, in turn, activates the valve actuation. By means of
feedback of the path signal of a positional transducer coupled with the
valve actuation, a path control loop is created, which greatly improves the
dynamic of the valve actuation.
1.1.3 PC module
1.1 Structure
Fig. 9: PC module
The sensor signals generated by the sensor circuit board are evaluated
independent of the microcomputer and the monitoring parameters are
alphanumerically displayed in a dedicated field on the monitor display
screen.. It coordinates the visual alarm with the alarm statuses, that are
recognized either by the PC module itself or by the microcomputer. In
addition, the PC module generates the screen display of ventilatory
pressure, volume and volume flow as a function of time.
The electric power supply module supplies the electric current necessary
for the operation of the ventilation unit. It contains all monitoring
switching devices as well as power transistors for switching on/off the
heating elements of the respiratory-gas humidifier and ventilation hoses.
1.1 Structure
The pneumatics module contains all the pneumatic components that are
necessary for processing the respiratory gas, including respiratory gas
mixer, valves, pressure regulator, etc. It is pneumatically connected to the
patient component via a plug coupling.
The test module contains an electromagnet via which the safety valve of
the patient component is activated; it also contains a lever for switching
the ventilator over to manual operation, as well as for calibration for
Compliance and Resistance, for testing the entire system during the unit
system selftest.
1.1 Structure
The gas supplies, compressed air (CA) and oxygen (O2) are fed in the
pneumatics module via magnetic valves (solenoids) to the oxygen/air
mixer and can be used either for mechanical or manual ventilation. The
current (actual) oxygen concentration is monitored by the oxygen sensor
and evaluated by the PC module.
Technical modifications The F. Stephan GmbH has carried out advanced technical modifications
on pediatric-ventilator STEPHANIE, affecting the handling/operation of
the ventilator.
1.2.2.2 Transformer
The supply voltages necessary for the STEPHANIE (5V / 4A and +/-
12/2A) are taken from the intermediate circuit voltages with switching
regulators (see circuit diagram, sheet 1). Types LM2679 from National
Semiconductor (NSC) were used.
As the intermediate circuit voltage from the mains can lie below as well
as above this value, a regulator is used for battery charging that operates
both upwards and downwards (component at U5, Circuit Diagram,
sheet 1). The circuit is similar to an application of the company Linear
Technologies.
The circuit breaker for the two heating systems of the STEPHANIE are
also built in to the power pack module. They are Profets from the
company, Infineon: (U6, U7, Circuit Diagram, sheet 1). They are Mosfets
with protective and monitoring functions.
1.2.2.6 Control
Monitoring of the power pack and the generation of the control signals
for the controller is via the microprocessor (Motorola 68HC11, ICI,
Circuit Diagram, sheet 2).
Outputs: Two power OK signals (POK1/2) These signals are set when all supply
voltages lie within the default limits.These are:
5V :4.8 – 5.2V
12V :11 – 13V
-12V :-11 – -13V
Intermediate circuit voltage : 18- 37V
The voltages are led through voltage multipliers to the A/D converter
inputs of the processor and the processor controls the compliance with the
limit values. The POK signals are logical HIGH (5V) for correct
functioning, and LOW (0V) for malfunctioning. POK1 is used as RESET
for the controller of the STEPHANIE. A correct start of the controller
was only achieved, when this signal was set with a delay of 200 ms.
The central control regulates the pressure, volume flow regulating circuits
for generating the ventilation pattern desired. It receives from the control
elements of the control unit the desired ventilation parameters and from
these generates, via the controller program, all the analogue and digital
control signals necessary for carrying out the tasks involved in the
therapy.
For this purpose, actual values of ventilation pressure and volume flow
are collected by the analogue-digital converter (ADUs) of the central
control (microcomputer) and are via control algorithms converted into
an analgoue control signal for the power amplifier. This regulates, via the
valve actuation, the control valve in the patient loop.Mandatory
monitoring of the ventilation pressure is carried out by the PC module
via a second pressure sensor.
For reasons of safety, the positional transducer is a dual system and the
position signal is fed to the central control by System 2. By comparing
the pre-set (target) valve (output DAC1 of the central control) with the
actual position, a complex monitoring of the ADU – DAU path of the
central control is obtained including power amplifier, actuator and control
valve.
1.2.5 Monitoring
Together with the display screen, the PC module serves and supports:
The PC is coupled with the central control via a serial interface port and
receives the setting parameters of the control unit, including those of the
alarm statuses of the central control, and also transmits its own alarm
statuses as well as the monitoring parameters to the central control..
1.2.6 Control
The central control and PC module take on, independent of one another,
the monitoring of safety-relevant parameters and statuses in line with
current standards.Both units classify alarm statuses via independent alarm
systems, into alarms of high and medium priority and then feed these data
to the circuit board. This triggers an audible alarm and, should the
situation or error be endangering, switches off the emergency air valve
for reasons of patient safety.
2.1 General
2 Mechanical structure
2.1 General
The control unit serves as the front cover of the housing unit.
Con0nections between the individual electronic units are in the form of
wiring backplane, ie circuit boards, and connection cables. A specific of
the Stephanie is the drive of the oxygen/air mixer positioned in the
pneumatics module. As the setting element belonging to it is positioned
on the control unit, a toothed-belt drive serves as a mechnical coupling to
both components.
All functional units of the patient circuit such as valve actuation, patient
component and the test module are mounted to a robust mounting plate
on the right-hand sidewall of the Stephanie. The valve actuation is
positioned inside the patient component and the test module on the
exterior of the housing.
Below the lowest tier insert, there are two blowers mounted for providing
cooling of the inside of the unit. RPMs of the blowers and therewith the
cooling performance, are regulated by thermistors that are positioned at
temperature-critical points in the unit.
To remove the cover, loosen the two long hexagon socket screws (3mm)
that are accessible through the large bores on the rear wall of the cover.
The cover is carefully pulled forward approximately 3 mm, so that the
tongues located on the underside of the cover can be freed from the pins
mounted to the housing. Following this, the cover can be lifted from the
housing. Make sure that the control unit does not tip forward. After
loosening M4 countersunk screws on the lowerside of the control unit,
the control unit is carefully pulled forwards and hung in the hing pins of
the unit. Now the control unit can be swung forward and all components
of the backplane as well as the control unit are accessible.
Assembly is carried out in reverse order. Only make sure that while
inserting the control unit, the potentiometer axis of the FiO2- drive must
be inserted into the correct sleeve.
3.1 General
3 Pneumatics
3.1 General
Technical modifications The F. Stephan GmbH has carried out advanced technical modifications
on Pediatric-ventilator STEPHANIE, affecting the handling/operation of
the ventilator. (See chapter 1.2.1)
O2-Zelle M5
R cksclagventil OV10/DN8
V2
O,3
NC
NW 0.8
Drucksensor
V1 M5 2.8 bar V3
Air
Duese NC
R1/8 NW 0.8
Injektor M5
Air
V1
NC V4 NC
O2-Kalb.21%
NW 2
D1 NW 0.8
R1/8 O2
P=min ca.2.8bar
Mischerausgang
R1/8 V5 NC
Druckregler 150 mbar R1/8
NW 2
V1 V2 V3 V4 V5 V6 Ventile stromlos geschlossen
V7 Ventil stromlos ge”ffnet
24 Volt 2 WATT
Druckbereich 2.5 - 6bar
Maást ab
Wer k s t of f / Hal bz eug
Hal bf er t i gt ei l
Roht ei l
Dat um Name ( Benennung)
ng)
Bear b. 6.10.00 Mainusch Pneumatic-Block
Gepr .
Nor m Stephanie
A NA E S T H E S I E ( Ar t i kel nummer ) Bl at t
S t e ph an P Ž DI A T R I E
A T E MT H E R A P I E
1 035 65 053
Bl .
Zus t . Žnder ung Dat um Name ( Ur spr . : ( Er s . f : )
Operating voltages
SV_P12LV
SV_P12
SV_N12
Control signals
/ÜW_VPL Actuation of compressed-air valve (-12V)
/ÜW_VO2 Actuation of oxygen valve (-12V)
/ÜW_VINJ1 Actuation of injector valve 1(-12V)
/ÜW_VINJ2 Actuation of injector valve 2 (-12V)
/ÜW_VSubst Actuation of substitution valve (-12V)
/ÜW_VAerosol Actuation of aerosol valve (-12V)
Outlet signals
PM_UP120 Pressure sensor 120 mbar admission pressure
PM_UPl Pressure sensor, compressed-air admission
pressure
PM_UO2 Pressure sensor, oxygen admission pressure
PM_UFiO2 FiO2 sensor signal
PM_UTemp Temperature-sensor signal
Tab. 1: Signal destination
All sensors apart from the oxygen sensor and their respective electronic
components are fitted to a circuit board. This circuit board is located
inside the pneumatics module in such a way, that a SUB-D plug
connector mounted to the rear side of the circuit board – hermetically
sealed – protrudes through the rear side of the pneumatics module. This
allows all electrical connections to be effected outside the pneumatics
module.
The pressure sensors for the admission pressure of the compressed air and
oxygen as well as the ventilatory-gas mixture are sensors with integrated
electronic, that directly provide the pressures of the assigned currents
PM_UPl, PM_UO2 and PM_UP120 at the outlet.
After loosening the screws and removing the cover, all components of the
module are now accessible.
The so-called test block is mounted on the right sidewall of the basuc
unit. On the underside of the test block there is a screw-on bottle attached
filled with copper wool that serves as test Compliance. Via a pneumatic
resistance, it is connected to a connecting union (pipe) found on the rear
side of the test block. When the unit is not in operation or when running
the test program, the Y-piece of the ventilation hose is attached to this
connecting union.
Using the lever „Hand / Resp“, the unit can be switched over between
manual ventilation and ventilator operation. When the lever is in „Resp“
position, the ventilatory-gas mixture reaches the patient via the patient
component and the patient is mechanically ventilated. When the lever is
in the „Hand“ position, a tappet rod found in the pneumatics module
switches the unit over to manual ventilation. The ventilatory-gas mixture
is then provided at the connecting union below this lever and is led via a
hose to the manual ventilation bag. By using a screwdriver, the volume
flow for manual ventilation can be adjusted at the flow control valve.
This becomes visible when the lever is in the „Hand“ position.
Otherwise, it is covered by the lever.
When re-assemblying, make sure that the tappet rod of the switch-over
"Hand/Ventilator" is centrally positioned on the tappet of the pneumatics
module.When replacing the electromagnet (solenoid), loosen the ring nut
between the patient component and the test block by using box or socket
spanner (wrench).
3.4.1 General
All pneumatic components of the patient loop are found in the patient
component; all these components must undergo sterilisation.
Valves V1, V2, V3, V4 and V5 form the safety system within the patient
component and prevent, in the event of mechanical malfunctioning, any
endangering of the patient.
By means of the excess pressure safety valve V1, the maximum possible
excess pressure at the inspiration outlet is mechanically limited. Here, the
opening pressure of valve V1 is adjustable in the range from 20 mbar to
approximately 80 mbar by using the toggle found on the front side of the
patient component.
The inspiration outlet and the expiration inlet are connected to the
emergency air valve via the check valves V2 and V3. During normal
operation of the ventilator, a lifting magnet (solenoid) in the test block
presses a valve head against the outlet of V5. This disables V2 and V3. In
the event that the unit detects an operational disturbance, the solenoid is
deactivated, the valve head opens the emergency air valve, and the patient
can inhale via valve V2 and exhale via valve V3, whereby the effective
dead space is less than 2.5 ml.
The inspiration valve Vin and the expiration valve Vex structurally form
a single unit, in that the inspiration valve opens when the expiration valve
closes, and vice versa. This valve unit is actuated by a valve actuation
that is found in the basic unit. The mechanical connection between
actuation and valve unit is by means of magnetic forces. Here, a small
magnetic clamp is attached to the actuation and the side of the valve
pistion facing the actuation is equipped with a thin soft iron plate. When
the patient componenet is locked onto the right side of the housing, the
magnetic clamp attracts the valve piston above the soft iron plate and in
this way forms a tight (free of play) connection between actuation and
valve unit. Through a slight crowned design of the soft iron plate, slight
tilting of the patient component can be compensated for, without
endangering the safety of the mechanical connection.
When the patient component is separated from the basic unit, the moving
part of the valve actuation found in the basic unit is pulled against a stop
until it overcomes the maximum magnetic force, and the connection
between actuation and valve is mechanically broken.
Here, it must be ensured that the red hose sleeve coupling of the
inspiration hose is on the red inspiration outlet. At the end of the
inspriation hose, a nebulizer pot can be attached, which then can be
connected via an additional hose with the Y-piece.
At the patient end, the temperature sensor is also inserted in to the bore
marked red. The plugs for heating and temperature sensor are pluged into
the respective sockets located in the connection block for heating,
pressure measurement and aerosol.
The measurement of pressure and volume flow takes place near the
patient between Y-piece and tube connector. For this, a
pneumotachograph head is inserted between both parts. The differential
pressure resulting above the resistor of the PNT-head is a rate for the
volume flow. The differential pressure is led, via two thin tubes, to the
pressure sensors (ventilation pressure) of the sensor circuit board found in
the basic unit, and to the differential pressure sensor (volume flow). The
mechanical connection of these hoses to the basic unit is via a Luer Lock
coupling.
On the right sidewall there is also a connection union (pipe) for aerosol
(see also Pneumatics Module). When the nebulizer function is activated
the therapy ventilatory-gas mixture at this connection is under a pressure
of approximately 2 bar. It is fed to the nebulizer chamber via a hose
attached to this connection. This is to be placed, near the patient, on the
end of the inspiration hose (see also Hose System).
4 Electronic
4.1 General
The entire electronic system is housed in the basic unit. It can be broken
down in to the following.
Control unit,
Monitor,
Electric power supply,
Pneumatics module,
Actuator
The control unit is the communications level (interface) between the user
and the equipment.
On the one hand it contains all control elements necessary for the
therapeutic operation of the Stephanie (Therapy Control Module), on the
other hand, it also contains the monitor, consisting of a display screen and
its control elements (Monitor Control Module), via which the user can
access all information pertaining to the patient and equipment.
The switches are equipped with resistor networks, so that their position,
including that of the potentiometer, is recorded and transmitted to the
microcomputer by an analogue-digital converter via a 16 ibt databus
(MC_D0..MC_D15).
The switching state of the keys is stored in a switching circuit and also
transmitted via the databus.
4.3 PC module
The PC circuit boards supply a PC (486 -33 DX2) and contain all
necessary drives and interfaces for the keyboard and storage systems.
The VGA card contains the entire electronics necessary to drive TFT-
VGA display screen. For detailed information, refer to manufacturer's
technical data..
4.3.3 PC interface
4.3 PC module
Operating voltages
SV_P5, SV_0V
Input signals
AIN0...AIN7 8 Analogue inputs
DI0...DI15 16 digital inputs
IGRa, IGRb Inputs for an incremental
transmitter
Output signals
DAC0...DAC3 4 Analogue outputs
D01...D15 16 digital outputs
Bus signals
D0...D15 16 Databus signals
A0...A10 11 Addressbus signals
OSC,/IOW, /IOR, BALE, AEN, Controlbus signals
RESET.DRV, /IOCS16
The PC interface card provides the connection between the address bus
and the databus of the PC as well as analogue and digital input/output
signals of the processor.
Via circuits D6 and D7, 16 digital inputs can be read and via D8 and D11
an additional 16 digital outputs can be recorded.
The sequential and combinational control logic for these processes is
provided by PALs D3, D4 and D5.
Bus signals
MC_D0...MC_D15 16 Databus signals
MC_A0...MC_A7 8 Address bus signals
MC_RD, MC_WR, MC_CLK, Control bus signals
MC_SEL
Tab. 8: Bus signals
The analogue input signals are collected by the two ADUs and together
with the digital input signals further processed by the microcomputer in
line with the program. At its outputs, it furnishes the corresponding
digital control signals and provides the DAU the information for its
analogue output data.
Address bus and databus signals are transmitted via the respective drivers
to the outside. Coordination of the internal processes is ensured by the
processor, and by the PAL. In particular, it generates control signals for
the peripheral components.
By means of the adjusting rheostat R1, the reference voltage of the DAU
and by means of R6 that of the 10 bit ADU can be set.
Bus signals
MC_D0...MC_D15 16 data bus signals
MC_A7 Address bus signal
MC_RD, MC_WR, MC_SEL Control bus signals
Tab. 14: Bus signals
Internal power supply For reasons of safety, the internal voltages for 5V are doubled and the
DC-DC transformer converts the externally supplied voltage RV_P12
into IC21 (5VB) and IC22 (5VB1).
Watchdog monitoring The PC module and the controller independently transmit, in the course
of properly processing their respective program and at intervals of approx
100..200 ms, watchdog signals PC_WD and MC_WD, that are evaluated
by the watchdog for the PC signal (IC8) and that for MC signal (IC9). In
the event of errors in the program run, the watchdog signal cease and the
PC and/or MC are restarted. During this time a HP (high priority) alarm
is triggered, and the patient has the possibility to spontaneously breathe
ambient air via the emergency air valve.
The precise time for running this routine can be read from the status
graphs for the watchdog automaton. It is identical for both the PC and
MC systems.
Generating the audible The audible alarm signal is also generated in an alarm automaton. The
alarm signal signal sequence is stipulated according EN 475 and for alarms of medium
priority (PC_MP, MC_MP) is distinct from those alarms of high priority
(PC_HP, MC_HP). The alarm automaton is made up of the PALs IC1
and IC2 in connection with a clock generator (IC5).
The output signal of the automaton activates the signal generator SPK1
via the power amplifier (IC10). This is found in the basic unit and is
electrically switched between outputs ÜW_SPK1_1 and ÜW_SPK1_2.
This allows, via the signal (ÜW_SPK1_2) to determine whether current
is actually flowing through the signal generator SPK1 on request/prompt
for audible signaling.. In this respect, signal ÜW_SPK1_2 is actually an
input signal and serves towards monitoring the audible signal routing
from input of watchdog switching down to signal generator SPK1.
In the event of any error within this routing, alarm automaton 2 kicks in
and triggers the audible alarm (ÜW_SPK2) via signal generator 2.
Valve actuation The magnetic valves (solenoids) in the pneumatics system are actuated by
bus signals (MC_D0..MC_D10) and the respective control. Both registers
IC11 and IC12 store the databytes sent by the MC and control, on their
part, the MOSFET power transistors T2..T11 via the optical coupler
IC15..IC17. Via their outputs, they switch (ÜW_V0.../ÜW_V10) the
solenoids to - 12V (SV_N12LV). In high-impedance state of the
transistors, the valves are switched off.
The sensor module is a circuit board on which all sensors necessary for
the proper functioning of the Stephanie are mounted, including their
respective electronic components. They are the following:
Input signals
MC_UDAC2 MC: DAC-Signal 2
/ÜW_VPNT ÜW: zero point valve PNT ON
GG_UIn1 GG: analogue input 1
GG_UIn2 GG: analogue input 2
GG_Ures1 GG: analogue reserve input 1
GG_Ures2 GG: analogue reserve input 2
GG_Rth11 GG: Thermistor 1, terminal1
GG_Rth12 GG: Thermistor 1, terminal2
GG_Rth21 GG: Thermistor 2, terminal1
GG_Rth22 GG: Thermistor 2, terminal2
GG_Rth31 GG: Thermistor 3, terminal1
GG_Rth32 GG: Thermistor 3, terminal2
Tab. 15: Input signals
Output signals
SM_UP1 SM: voltage, pressure 1
SM_UP1Out SM: k. fixed voltage, pressure 1
SM_MP1 SM: measuring pt. 1 (refer. voltage 1.23 V)
SM_UP2 SM: voltage, pressure 2
SM_UFlow SM: voltage, volume flow
SM_UFlowOut SM: k. fixed voltage volume flow
SM_UOut1 SM: output voltage 1
SM_UOut2 SM: output voltage 2
SM_UTemp1 SM: voltage, temperature sensor 1
SM_UTemp2 SM: voltage, temperature sensor 2
SM_UTemp3 SM: voltage, temperature sensor 3
SM_Ures SM: voltage, reserve
In the same way the distal supply pressure P2 reaches, via a second
distributing manifold, the distal pressure sensor (sensor 3) and via the
solenoid valve (valve 1), the second pressure inlet of the differential
pressure sensor.
Solenoid valves Valve1 and Valve 2 are electrically actuated in order and
switch in deactivated state a respective pressure inlet of the differential
pressure sensor to ambient air. During normal operation the valves are
switched on (/ÜW_VPNT= -12V) and connect the pressure inlets of the
differential pressure sensor with the pressure port for the proximal and
distal pressure.
During a periodical zero point correction, the valves, via the signal
/ÜW_VPNT, are switched off for the duration of approximately 50 ms
and the pressure inlets of the differential pressure sensor are switched
over to ambient air. The microcomputer measures the ouput voltage
SM_UFlow and in a compensating procedure, via the control signal
MC_UDAC2, the bridge zero point is modified until SM_UFlow Null
reaches zero.
Pressure sensors for The output signal of the proximal pressure sensor (sensor 2) is amplified
ventilatory pressure via amplifier A2B ( V=2) to signal SM_UP1 and via a second amplifier
A2A (V=1) short circuit-proof to signal SM_UP1Out.
The distal pressure signal of sensor 3, however, is only converted via
amplifier A2C (V=2) to signal SM_UP2.
Voltages
SV_P12, SV_N12, SV_P12LV, Operating voltages
SV_N12LV
Tab. 17: Voltages
Input signals
WM_I1PSD1 WM: current 1 from PSD 1
WM_I2PSD1 WM: current 2 from PSD 1
WM_I1PSD2 WM: current 1 from PSD 2
WM_I2PSD2 WM: current 2 from PSD 2
MC_UDAC1 MC: Analogue voltage 1
MC_LVEin MC: power amplifier ON
Tab. 18: Input signals
Output signals
LV_USum1 LV: cumulative voltage of
positional transd. 1
LV_USum2 LV: cumulative voltage of
positional transd. 2
LV_Us1 LV: voltage, path 1
LV_Us2 LV: voltage, path 2
LV_UEDA1 LV: output voltage 1 for EDA
LV_UEDA2 LV: output voltage 2 for EDA
Tab. 19: Output signals
Explanatory note on the The actual positional transducer system is found on the right sidewall of
positional transducer the Stephanie. It consists of two position measurement paths, each of
system which are formed by a luminescence diode (LED), an optic system and a
position sensitive device (PSD).
The lightband generated by the LED and optic is covered by the PSD
opposite a crevice located on the valve piston, so that at the point of the
crevice only a narrow light spot appears on the PSD. This spot of light
generates two photoelectric currents, the strength of which both in
intensity as well as the position of the spot of light on the active surface
of the PSD can be determined.
When the quotients are formed from Id and Is, one receives
S = Id/Is = 1 - 2*x. (Eq4)
Power amplifier The actual power amplifier (LV) is realised by a TDA 2050 (A6). In
connection with the feedback resistors (R33, R41) it has for the input
signal an amplification of V=13.6. Input resistor R33 is, via relay REL1
in its activated state, switched to control input MC_UDAC1 and in
deactivated state, via R40, switched to the output of A6.
This means that the LV can no longer be activated via its input, and the
output current through EDA sinks to values less than 30 mA, which no
longer create thermal load.
Travel feedback The output signal of positional transducer 1 is led via a transfer link with
PD behaviour (A5A, A5B) to a further control input for the power
amplifier. Der P-part carries back a travel-proprtional part, and the D-part
a speed-proportional part. Via the travel-proportional part, the overall
transfer factor of LV and EDA, concerning travel, and via the speed-
proportional part (ER3) the attenuation can be stipulated. Due to the high
loop amplification of this system, the transfer function LV-EDA is
practically only dependent on that of the positional transducer system,
and results
Safety observations Since the positional transducer system is designed as a dual system, one
system is for the feedback, a second is for checking the system.. Thanks
to the good coordination between the control voltage of the MC
(MC_UDAC1) and the position of the control valve, a check can be run,
via the second positional transducer system, as to whether the desired
position has actually been achieved. This comparison check takes place in
the MC. This monitoring check is very complex and entails the entire
path MC - LV - actuator – control valve - MC. Its operational function
is: within a certain temporal and voltage error span, the MC checks
whether the generated control voltage of the MC agrees with the voltage
of positional transducer system 2. In the case of deviation, an alarm with
high priority is triggered along with activation of the emergency air
system.
Voltages:
SV_P5, SV_P12, SV_N12 Operating voltages for the
computers
SV_P12LV, SV_N12LV Operating voltages for power
amplifier and valves,
SV_P24 Intermediate circuit voltage
(24...28V)
SV_HEIZ1 Activated voltage for heating and
humidifier,
SV_Heiz2 Activated voltage for hose heating
Tab. 20: Voltages
Monitoring signals:
SV_PG Power good signal of computer
voltages
SV_PGLV Power good signal of LV voltages
SV_Netzok Mains voltage is OK
SV_Akkok Battery voltage is OK
SV_AkkLow Battery voltage is less than 20 V
Tab. 21: Monitoring signals
Control voltage:
MC_SD MC: Shut down
/BE_SBER BE: switch position not ready
MC_HEIZ1 MC: Heating 1 ON
MC_HEIZ2 MC: Heating 2 ON
/UW_NL ÜW: emergency air valve not
activated
Tab. 22: Control voltage
Switch signals
SV_S1a, SV_S1b Switch contact of mains power
switch
Tab. 23: Switch signals
The input voltage of 220 V is led, via the fuses and the mains power
Voltage generation switch to a primary voltage regulator. If the signal is MC_SD=LOW, it
generates an intermediate circuit voltage of 24 V...28 V (otherwise it is
zero). At the same time the charging current for the 24 V lead battery is
generated and the battery is charged.In the case of /BE_SBER =HIGH,
the secondary voltage regulators for the computer voltages and the power
amplifier are switched on. A contact of the mains power switch is on the
outside.
Substitution switch of If there is a power failure during operation on mains power supply
the battery voltage (SV_Netzok=LOW), the intermediate circuit voltage through the primary
regulator becomes zero, and the battery voltage (24V) is switched via a
MosFet to the intermediate circuit. This ensures that there continue, with
no break, all output voltages of the power supply. Only the heating is not
supplied with current during a power failure.
Heating voltage The battery voltage is monitored for compliance within certain voltage
generation parameters and its state is determined via the respective signals.
The voltages for respiratory gas humidifier and hose heating are obtained
via the power-MosFets directly from the intermediate circuit voltages.
Prerequisite for provision of these heating voltages is that the mains
power is on and the systems monitor does not detect an emergency air
situation (SV_Netzok * /UW_NL=HIGH). If this is the case, heating
voltages 1 and 2 from the MC can be switched via signals MC_Heiz1,2.
When the patient component is connected to the housing, both parts come
into magnetic contact and form a tight hysteresis-free mechanical
coupling.The executed positional feedback, in connection with the valve
actuator, has already been explained in Section 4.3.4. In addition, it
allows to check the integrity of the connection between the valve
actuation and the valve piston.The assembly of the valve actuation is
carried out within the unit, whereby using an adjustment gauge, the
adjustment of the position of the positional transducer and the ventilation
valve unit can be conducted.
5.1 General
5 Safety concept
5.1 General
5.4.2 Pressure
Setting Limit Value Setting the limit value is carried out manually.
Setup: Pgob = 30 mbar,
Pgun = 10 mbar.
5.4.5 Temperature
Error corrective actions Visual: " Respiratory gas temperature too high!"
„ Respiratory gas temperature too low!“
Audible: MP - Alarm for a temperature >= 40°C
The safety system responds by switching off the emergency air valve (
/ÜW_NL), the two valves for compressed air (/ÜW_VPL) and oxygen
(/ÜW_VO2), the central gas supply as well as the injector valve
(/ÜW_VInj1, /ÜW_VInj2). In this event the patient can at least
spontaneously breathe ambient air.
MPALneu MPALneu
MP-OptAnzeige MP-OptDisplay
Gelb Yellow
Grün green
MP-Signalton MP-Signal tone
Kein Alarm No alarm
MPALneu MPALneu
Tab. 24: Alarm automation graphic glossar
In detail:
The various responses of the safety system are listed in the last column.
Definitions:
Safety status 1:
the switching off of all valves, including the emergency air valve, so that
the patient can breathe ambient air or be manually ventilated.
Safety status 2:
For CMV ventilation with standard values for CPAP mit PEEP = 2mbar
Ventilatory parameters
7. „Peak pressure“ When Plim_max MC switches after
is reached for 0.1 s to PEEP level.
longer than 0.1 s PC switches off
for MC and for emergency air after
longer than 0.2s 0.2s
for PC
8. „Cont. high pressure“ HP MC PC Pmean > Pmean /V_NL for 1sec.
lim for longer than After that start of
15 s new inspiration.
Max. 3 attempt, then
/V_NL,/V_PL,
/V_O2,
9. „pressure low, MHP MC PC Pendinsp < Pmax
Disconn ?“ during machine
inspiration
10. „AMV high“ MP MC PC Amv > AMV lim
max
11. „AMV to low“ MHP MC PC AMV < AMV lim
min
12. „Apnea“ MHP MC PC no inspiration start
within 15 s
Hardware error
20. „Controller error “ HP ÜW PC watchdog error Safety status 1
21 „PC error „ almost HP ÜW MC watchdog error Safety status 1
imposs..
21. „Valve error n “ MHP MC No electrical
response of valve
n
22. „Potentiometer n“ HP MC Potentiometer Safety status 2
makes no contact
23. „power supply error“ HP MC, PC /SV_PG + Safety status 1
ÜW /SV_PGLV
By means of this test, all errors possibly occuring in the control unit can
be detected.
7 List of figures
7 List of figures
8 List of tables
8 List of tables
10 Notices
10 Notices