1935 Chauvin Arnoux Navod K Obsluze CA 6116n CA 6117 Anglicky Main
1935 Chauvin Arnoux Navod K Obsluze CA 6116n CA 6117 Anglicky Main
C.A 6116N
C.A 6117
E N G L I S H User's manual
Thank you for purchasing a C.A 6116N or C.A 6117 installation tester. To obtain the best service from your unit:
read this user manual carefully,
comply with the precautions for use.
WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.
The voltage on the terminals must not exceed 550 V. Equipment protected by double insulation.
The product is declared recyclable following a life cycle analysis in accordance with standard ISO 14040.
The CE marking indicates conformity with European LVD and EMC directives.
Chauvin Arnoux has adopted an Eco-Design approach in order to design this appliance. Analysis of the complete
lifecycle has enabled us to control and optimize the effects of the product on the environment. In particular this ap-
pliance exceeds regulation requirements with respect to recycling and reuse.
The rubbish bin with a line through it means that in the European Union, the product must undergo selective disposal
in compliance with Directive WEEE 2002/96/EC.
This device is protected against accidental voltages of not more than 600V with respect to earth in measurement category III or
300V with respect to earth in measurement category IV (under shelter). The protection provided by the device may be compro-
mised if it is used other than as specified by the manufacturer.
Do not exceed the maximum rated voltage and current and the measurement category.
Never exceed the protection limits indicated in the specifications.
Comply with the conditions of use, namely the temperature, the humidity, the altitude, the degree of pollution, and the place
of use.
Do not use the device or its accessories if they seem damaged.
Do not use the device if the battery compartment cover is missing or incorrectly installed.
To recharge the battery, use only the mains adapter unit provided with the device.
To replace the battery, disconnect everything connected to the device and set the switch to OFF.
Do not use a battery with a damaged jacket.
Use connection accessories of which the overvoltage category and service voltage are greater than or equal to those of the
measuring device (600 V Cat. III or 300 V Cat. IV).
Troubleshooting and metrological checks must be done only by accredited skilled personnel.
Wear the appropriate protective gear.
2
CONTENTS
3
1. FIRST START-UP
1.1. UNPACKING
➀ 11
➂ ➄
12
➅
14
➈
➆
➃
➇
➁
13
➉
4 One tripod cable with mains plug (adapted to the country of sale.
4
1.2. CHARGING THE BATTERY
Before the first use, start by fully charging the battery. The charging must be done between 10 and 45°C.
> 90 VAC
< 264 VAC
50 / 60 Hz
After prolonged storage, the battery may be fully discharged. In this case, the first charge may take longer and the indicator on
the device flashes for the first few minutes.
Set the switch to OFF, but charging is possible when the device is not off,
The 4-point hands-free strap will let you use the device while leaving your
hands free. Snap the four fasteners of the strap onto the four lugs on the
device.
Adjust the length of the strap, then the tilt of the device.
5
To withdraw the strap, slide a flat screwdriver under the tab of the fastener to lift it, then slide the fastener down.
+ or
sustained
press
6
1.6. CHOICE OF LANGUAGE
Before using the device, first choose the language in which you want the device to display messages.
Set the switch to SET-UP. Use the directional keypad to select the languages icon:
SET UP
OK
OFF
11-21-2013 11:17
OK
Press the OK key to validate your
choice.
SET-UP
Select your language, from among those proposed, using the keys and validate by pressing the OK key again.
You can download other languages from our site’s support space (see §10.2).
7
2. PRESENTATION OF THE DEVICES
Connection terminals.
TEST
Four function
keys. /
SET UP
OFF
Indicator light.
OK
Directional keypad:
Stud for fixing on the four navigation keys
4-point hands-free Brightness adjustment key. and one validation
strap. Help key. key.
8
2.1. FUNCTIONS OF THE DEVICES
C.A 6116N and C.A 6117 installation testers are portable measuring devices with a colour graphic display. They are powered by
a rechargeable battery with a built-in charger and external power supply unit.
These instruments are intended to check the safety of electrical installations. It can be used to test a new installation before it is
powered up, to check an existing installation, whether in operation or not, or to diagnose a malfunction in an installation.
Controls one thirteen-position switch, one five-key navigator, one keypad with four function keys, one context-
sensitive help key, one brightness key, and one TEST button.
Display 5.7” (115 x 86mm) colour graphic display unit, 1/4 VGA (320 x 240 points).
The only difference between the C.A 6116N and the C.A 6117 is that the C.A 6117 can test type B RCDs.
2.2. KEYPAD
The actions of the 4 function keys are indicated on the display unit by adjacent icons. They depend on the context.
The help key can be used in all functions. The help function is context-sensitive: it depends on the function.
The directional keypad comprises four navigation keys and one validation key.
9
2.3. DISPLAY UNIT
➁ ➂ ➃ ➄
➀ 11-21-2013 10:47 50 . 0 Ω 50 . 1 Hz
L
➆
6 mA
L-PE 230.3 V
➅ L-N 230.4 V ➇
N-PE 0.8 V
6%
➈ LOOP ZS
➉ 11
The USB port can also be used to update the device’s internal software (see §10.6).
The USB cord and the associated software are supplied with the device.
10
3. USE
3.1. GENERAL
When it leaves the plant, the device is configured so that it can be used without changing the parameters. For most meas-
urements, simply select the measurement function by turning the switch and press the TEST button.
The device is not designed to operate when the charger is connected. The measurements must be made using battery
power.
3.1.1. CONFIGURATION
When configuring the measurements, you can always choose between:
validating by pressing the OK key,
or exiting without saving by pressing the key.
3.1.2. HELP
In addition to an intuitive interface, the instrument provides complete help in use and analyses and appraisals. Three types of
help function are available:
Help before the measurement can be accessed using the key. It indicates the connections to be made for each function
and important recommendations.
Error messages appear, as soon as the TEST button is pressed, to report connection errors, measurement parameterizing
errors, out-of-range values, defective installations tested, etc.
Help associated with the error messages. Messages containing the icon invite you to look up the help for ways to
eliminate the error found.
The user is assumed to be at the reference earth potential. He/she must therefore not be insulated from earth: must not
wear insulating shoes or insulating gloves and must not use a plastic object to press the TEST button.
For measurements made at the mains voltage, the device checks that the connection is correct and displays the position of the
phase on the socket outlet. It also checks the presence of a protective conductor on the PE terminal by means of the contact the
user makes with his/her finger by touching the TEST button.
11
3.2.2. MAKING A MEASUREMENT
Connect the lead to the device to be tested. As soon as the device is powered up, it measures the voltages present on its terminals
and displays them, whatever the setting of the switch.
In the ZS (RA/SEL.) and RCD, settings, the device also indicates the position of the phase on the display unit using the symbol.
The mains socket outlet of the measuring cable is marked with a white reference spot.
: the phase is on the right-hand contact of the mains plug when the white spot is up.
: the phase is on the left-hand contact of the mains plug when the white spot is up.
: the device cannot determine where the position of the phase, probably because the PE is not connected or the L and
PE conductors are interchanged.
The L symbol is displayed as soon as the voltage is high enough (> UL programmable in SET-UP). The terminal identified
as L is the one that has the highest voltage with respect to PE.
12
3.3. RESISTANCE AND CONTINUITY MEASUREMENT
3.3.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
For continuity measurements, the device generates a DC current of 200 or 12 mA, at the user’s discretion, between the W and
COM terminals. It then measures the voltage present between these two terminals and from it deduces the value of R = V/I.
For resistance measurements (current chosen = kW), the device generates a DC voltage between the W and COM terminals. It
then measures the current between these two terminals and from it deduces the value of R = V/I.
In the case of a measurement at high current (200 mA), at the end of one second, the device reverses the direction of the current
and makes another measurement for one second. The result displayed is the mean of these two measurements. It is possible to
make measurements with either the positive or the negative polarity of the current disabled.
For measurements at low current (12 mA or kW), the polarity is positive only.
When you make continuity measurements that are not contractual, prefer a current of 12 mA. Even though the results cannot be
regarded as those of a normative test, this significantly increases the life of the device between charges and forestalls untimely
tripping of the installations if there is a connection error.
The permanent mode is used to chain measurements without having to press the TEST button each time.
If the object to be measured is inductive, it is better to switch to pulse mode at 200 mA and make a measurement at positive
polarity, then a measurement at negative polarity, manually, in order to leave time for the measurement to settle.
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is below threshold, making it un-
necessary to look at the display unit to check this point.
Set the switch to W . Use the leads to connect the device to be tested between the W and COM ter-
minals of the device. The object to be tested must not be live.
SET UP
R
/
OFF
To correct for the resistance of the measurement leads (leads and probe tips or crocodile clips), for measurements
at 12 and 200 mA (see §3.16).
Pressing the TEST button starts only one measurement (pulse mode).
Pressing the TEST button starts the continuous measurement (permanent mode). To
stop it, you must press the TEST button again.
13
Automatic reversal of polarity for a measurement at 200 mA.
Ω 002.00 To set the alarm threshold (see §3.17). The default threshold is 2W.
kΩ
Before the measurement: to display the measurements already recorded.
After the measurement: to record it.
6% The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
Once the parameters have been defined, you can start the measurement.
If you selected the pulse mode, press the TEST button once and the measurement stops automatically when it is over.
If you selected the permanent mode, press the TEST button once to start the measurement and a second time to
Measurement result:
R = (R+) + (R-)
200 mA 2
Measurement current.
1%
CONTINUITY
Measurement with reversal of polarity.
Permanent mode.
14
To see the next display page.
200 mA
1%
CONTINUITE
1%
CONTINUITY
The polarity of the current is positive.
Pulse mode.
15
In the case of a resistance measurement (kW), there is no current reversal and no compensation for the measurement leads.
11-22-2013 10:47 - - .- Hz
kΩ
1 . 58 k Ω Measurement result.
1%
RESISTANCE
Permanent mode.
In this case, the measurement is not enabled. Eliminate the cause of the interference voltage and start the measurement over.
Another possible error is measurement of an overly inductive load that prevents the measurement current from stabilizing. In this
case, start the measurement in permanent mode with only one polarity and wait for the measurement to stabilize.
For help with connections or any other information, use the help function.
16
3.4. INSULATION RESISTANCE MEASUREMENT
3.4.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The device generates a DC test voltage between the COM and MW terminals. The value of this voltage depends on the resistance
to be measured: it is greater than or equal to UN when R ≥ RN = UN /1 mA, and less otherwise. The device measures the voltage
and current present between the two terminals and from them deduces the value of R = V / I.
The COM terminal is the voltage reference point. The MW terminal therefore provides a negative voltage.
Set the switch to MW. Use the leads to connect the device to be tested between the COM and MW
terminals of the device. The object to be tested must not be live.
SET UP
R
/
OFF
To avoid leakage during the insulation measurement, which would
throw off the measurement, do not use the measuring cable when
you make this type of measurement, but two simple leads.
Generally, an insulation measurement on an installation is made between the interconnected phase(s) and neutral, on the one
hand, and earth, on the other.
L1
L2
L3
PE
If the insulation is not sufficient, you must then make the measurement between each of the pairs to locate the fault. It is for this
reason possible to index the recorded value with one of the following values:
L-N, L-PE, N-PE, L1-PE, L2-PE, L3-PE, L1-N, L2-N, L3-N, L1-L2, L2-L3 or L1-L3
The remoted TEST button of the optional remote control probe makes it easier to trigger the measurement. To use the remote
control probe, refer to its operating data sheet.
C.A 6113/16/17
with
Operation only
L1
L2
L3
PE
17
3.4.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
To choose the nominal test voltage UN: 50, 100, 250, 500 or 1000 V.
k Ω 0500.0 To set the alarm threshold (see §3.17). As default, the threshold is set to
R (kW) = UN / 1 mA.
MΩ
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
6% The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
Once the parameters have been defined, you can start the measurement.
Keep the TEST button pressed until the measurement is stable. The measurement stops when the TEST button
is released.
TEST
Before disconnecting the leads or starting another measurement, wait a few seconds for the device tested to be discharged
3 1 . 0 6 MΩ Measurement result.
INSULATION
To change display pages.
18
To see the next display page.
INSULATION
The measurement may be unstable, probably because of an overly capacitive load or an insulation fault. In this case, read the
measurement on the bargraph.
For help with connections or any other information, use the help function.
19
3.5. 3P EARTH RESISTANCE MEASUREMENT
This function is the only one that can measure an earth resistance when the electrical installation to be tested is not live (new in-
stallation, for example). It uses two auxiliary rods, with the third rod being constituted by the earth electrode to be tested (whence
the name “3P”).
It can be used on an existing electrical installation, but the power must be cut off (main RCD). In all cases (new or existing instal-
lation), the earthing strip of the installation must be open during the measurement.
It is possible to make a rapid measurement and measure only RE or else to make a more detailed measurement by also measur-
ing the resistances of the rods.
To measure the resistances of the RS and RH rods, the device internally reverses the E and S terminals and makes a measurement.
It then does likewise with the E and H terminals.
Plant the H and S rods in line with the earth electrode. The distance between the S rod
Set the switch to RE 3P.
and the earth electrode must be approximately 62% of the distance between the H rod
and the earth electrode.
In order to avoid electromagnetic interference, we recommend paying out the full length of
SET UP the cables, placing them as far apart as possible, and not making loops.
earth
strap
H S
OFF
62% d
d
/
Connect the cables to the H and S terminals. Power down the installation and disconnect the earth strap. Then connect the E
terminal to the earth electrode to be checked.
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it un-
necessary to look at the display unit to check this point.
Choice of type of measurement: rapid, to measure RE only (icon crossed out), or detailed, to measure also rod
resistances RS and RH. This last case is useful if the ground is dry, making the resistance of the rods high.
To compensate for the resistance of the lead connected to the E terminal, for measurements of low values (see §3.16).
20
To activate the alarm.
Ω 050.00 To set the alarm threshold (see §3.17). As default, the threshold is set to 50W.
kΩ
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
6% The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
If the measurement must be made in a damp environment, remember to change the value of maximum contact voltage
UL in SET-UP (see §5) and set it to 25 V.
Press the TEST button to start the measurement. The measurement stops automatically.
TEST
Do not forget to reconnect the earth strap at the end of the measurement before powering the installation back up.
21
3.5.5. VALIDATION OF THE MEASUREMENT
To validate your measurement, move the S rod towards the H rod by 10% of d and make another measurement. Then move the
S rod, again by 10% of d, but towards the earth electrode.
H S
52% d
62% d
72% d
/
The 3 measurement results must be the same to within a few percent. If this is the case, the measurement is valid. If not, it is
because the S rod is in the zone of influence of the earth electrode.
If the resistivity of the ground is homogeneous, it is necessary to increase distance d and repeat the measurements. If the re-
sistivity of the ground is inhomogeneous, the measurement point must be moved either towards the H rod or towards the earth
terminal until the measurement is valid.
S
d2
H S
E
d1
If you find the same values, your measurement is reliable. If the measured values are substantially different, it is probable that
they were influenced by earth currents or a groundwater artery. It may be useful to drive the rods deeper.
If the in-line configuration is not possible, you can plant the rods in a triangle. To validate the measurement, move the S rod on
either side of the line HE.
H E
Avoid routing the connecting cables of the earth rods near or parallel to other cables (transmission or power supply), metal pipes,
rails, or fences, this in order to avoid the risk of cross-talk with the measurement current.
22
3.5.7. ERROR INDICATION
The commonest errors in the case of an earth measurement are the presence of an interference voltage or rod resistances that
are too high.
In these two cases, the earth measurement is not enabled. Move the rods and start the measurement over.
To reduce the resistance of the rods RH (RS), you can add one or more rods, two metres apart, in the H (S) branch of the circuit.
You can also drive them deeper and pack the earth around them, or wet it with a little water.
For help with connections or any other information, use the help function.
23
3.6. LOOP IMPEDANCE MEASUREMENT (ZS)
In a TN or TT type installation, the loop impedance measurement is used to calculate the short-circuit current and to size the
protections of the installation (fuses or RCDs), especially their breaking capacity.
In a TT type installation, the loop impedance measurement makes it easy to determine the earth resistance without planting any
rods and without cutting off power to the installation. The result obtained, ZS, is the loop impedance of the installation between
the L and PE conductors. It is barely greater than the earth resistance.
From this value and the conventional touch voltage limit (UL), it is then possible to choose the rated differential operating current
of the RCD: IDN < UL / ZS.
This measurement cannot be made in an IT type installation because of the high earthing impedance of the supply transformer,
which may even be completely isolated from earth.
It then applies a low current, 6, 9 or 12 mA at the user’s discretion, between the L and PE terminals. This low current serves to
avoid tripping residual current devices of which the nominal current is greater than or equal to 30 mA. This second measurement
is used to determine ZPE.
The device then calculates loop resistance ZS = ZL-PE = ZL+ZPE , and short-circuit current Ik = ULPE/ZS.
The value of Ik serves to check the proper sizing of the protections of the installation (fuses or RCDs).
For greater accuracy, it is possible to measure ZS with a high current (TRIP mode), but this measurement may trip the RCD of
the installation.
Set the switch to ZS (RA/SEL.). Connect the measuring cable to the device, then to the socket outlet of the instal-
lation to be tested.
SET UP At the time of connection, the device first checks that the voltages present
on its terminals are correct, then determines the position of the phase (L) and
of the neutral (N) with respect to the protective conductor (PE) and displays
it. If necessary, it then automatically switches the L and N terminals so that
the loop measurement can be made without modifying the connections of
the device.
OFF If possible, first disconnect all loads from the network on which you make the loop
measurement.
It is possible to eliminate this step if you use a measurement current of 6 mA, which
allows a leakage current of up to 9 mA for an installation protected by a 30 mA
residual current device.
RN N RN N
PE RE PE
Rb Ra Rb
/ /
24
For a more accurate measurement, you can choose a high current (TRIP mode), but the RCD that protects the installation may trip.
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it un-
necessary to look at the display unit to check this point.
The signal can be smoothed to produce a mean of several values. But the measurement then takes longer.
To compensate for the resistance of the measurement leads, for measurements of low values (see §3.16).
The device proposes choosing the voltage for the Ik calculation from among the following values:
ULN (the measured voltage value),
(Ik) the voltage of the old standard (for example 220 V),
the voltage of the current standard (for example 230 V).
Depending on the voltage ULN measured, the device proposes the following choices:
if 170<ULN<270 V: ULN, 220 V, or 230 V.
if 90<ULN<150 V: ULN, 110 V or 127 V.
if 300<ULN<500 V: ULN, 380 V or 400 V.
Z-R To activate the alarm on ZLPE (in TRIP mode) or on RLPE (in non-tripping mode).
Press the TEST button to start the measurement. The measurement stops automatically.
When the TEST button is pressed, the device checks that the contact voltage is less than UL. If not, it does not
make the loop impedance measurement.
25
3.6.4. READING OF THE RESULT
In the case of a non-tripping measurement, with smoothing:
Value of the alarm threshold.
11-22-2013 10:47 50 .0 Ω 50 . 1 Hz
R s 1.36 Ω
To change display pages.
L s 2.2 mH
Value of the reference voltage for the
4% calculation of Ik.
L s 9.6 mH
Value of the inductance.
4%
LOOP Zs
26
3.7. EARTH MEASUREMENT ON LIVE CIRCUIT (ZA, RA)
This function is used to make an earth resistance measurement in a place where it is impossible to make a 3P earth measurement
or to disconnect the earth connection strap, often the case in an urban environment.
This measurement is made without disconnecting the earth, with only one additional rod, saving time with respect to a conven-
tional earth measurement with two auxiliary rods.
In the case of a TT type installation, this measurement is a very simple way to measure the earth of frame grounds.
In the case of a TN type installation, to determine the value of each of the earths put in parallel, it is necessary to perform a se-
lective earth measurement on live circuit using a current clamp (see §3.8). Without this clamp, what you find is the value of the
global earth connected to the network, which is rather meaningless.
It is then more useful to measure the loop impedance to size the fuses and RCDs, and to measure the fault voltage to check the
protection of persons.
For greater accuracy, it is possible to make the measurement with a high current (TRIP mode), but this measurement may trip
the RCD of the installation.
Set the switch to ZS (RA/SEL.). Connect the measuring cable to the device, then to the socket outlet of the instal-
lation to be tested.
At the time of connection, the device detects the positions of the phase (L)
SET UP and of neutral (N) with respect to the protective conductor (PE) and displays
them. If necessary, it then automatically switches the L and N terminals so
that the loop measurement can be made without modifying the connections
of the terminals of the device.
If possible, first disconnect all loads from the network on which you make the earth
OFF
measurement on line circuit.
It is possible to eliminate this step if you use a measurement current of 6 mA, which
allows a leakage current of up to 9 mA for an installation protected by a 30 mA re-
sidual current device.
Plant the auxiliary rod at a distance of more than 25 metres from the earth electrode and
connect it to the (RA SEL) terminal of the device. The symbol is then displayed.
RN N RN N
PE RE PE PE
Rb Ra Rb Ra
> 25 m > 25 m
/ /
27
To make this measurement, you can choose:
either a low current which avoids any untimely tripping out of the installation but gives only the earth resistance (RA).
or a high current (TRIP mode), which yields a more accurate earth impedance (ZA) with good measurement stability and can
also be used to calculate the short-circuit fault voltage, UFK, in accordance with standard SEV 3569.
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it un-
necessary to look at the display unit to check this point.
The signal can be smoothed to produce a mean of values. But the measurement then takes longer.
Before starting the measurement, you can configure it by modifying the parameters displayed:
To compensate for the resistance of the measurement leads, for measurements of low values (see §3.16).
The device proposes choosing the voltage for the Ik calculation from among the following values:
ULN (the measured voltage value),
(Ik) the voltage of the old standard (for example 220 V),
the voltage of the current standard (for example 230 V).
Depending on the voltage ULN measured, the device proposes the following choices:
if 170<ULN<270 V: ULN, 220 V or 230 V.
if 90<ULN<150 V: ULN, 110 V or 127 V.
if 300<ULN<500 V: ULN, 380 V or 400 V.
Z-R To activate the alarm on ZA (in TRIP mode) or on RA (in non-tripping mode).
28
Press the TEST button to start the measurement. The measurement stops automatically.
TEST
11-25-2013 10:47 50 . 0 Ω 50 . 1 Hz
L
Value of the short-circuit current.
UFk
U FK 0.6 V
above the alarm threshold.
UFk is calculated only in earth measurement on live circuit with a high current (TRIP mode). UFk = Ik x ZA.
29
To see the next display page.
11-25-2013 10:47 50 . 0 Ω 50 . 1 Hz
L
UFk
EARTH 1P (Ra)
The third page displays the values of ZS, RS, LS. The fourth page displays the voltages ULN, ULPE, UNPE and on the rod ( ) before
the measurement.
In the case of a measurement with a low current and smoothing, the first display screen is the following:
11-25-2013 10:47 50 . 0 Ω 50 . 1 Hz
L Measurement result.
6 mA
RA 2 5.1 0 Ω
Case where the measurement is
below the alarm threshold.
30
3.7.5. VALIDATION OF THE MEASUREMENT
Move the rod ± 10% of the distance from the earth electrode and make two more measurements. The 3 measurement results
must be the same to within a few percent. In this case the measurement is valid.
If this is not the case, this means that the rod is in the zone of influence of the earth electrode. You must then move the rod away
from the earth electrode and redo the measurements.
31
3.8. SELECTIVE EARTH MEASUREMENT ON LIVE CIRCUIT
This function is used to make an earth measurement and to select one earth from among others, in parallel, and measure it. It
requires the use of an optional current clamp.
Connect the measuring cable to the device, then to the socket outlet of the instal-
Set the switch to ZS (RA/SEL.). lation to be tested.
At the time of connection, the device detects the positions of the phase (L)
SET UP and of neutral (N) with respect to the protective conductor (PE) and displays
them. If necessary, it then automatically switches the L and N terminals so
that the measurement can be made without modifying the connections of
the terminals of the device.
Plant the auxiliary rod at a distance of more than 25 metres from the earth electrode
and connect it to the (RA SEL) terminal of the device. The symbol is then dis-
OFF played.
Connect the clamp to the device; the symbol is displayed. Then place it on the
earth circuit to be measured.
Case of a TN installation
RL L
RN N
RE PE PE
> 25 m
/
Case of a TT installation
RL L
RN N
PE
> 25 m
/
For a more accurate measurement, you can choose a high current (TRIP mode), but the RCD that protects the installation may trip.
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it un-
necessary to look at the display unit to check this point.
The signal can be smoothed to produce a mean of several values. But the measurement then takes longer.
32
In the selective earth measurement on live circuit, it is essential to do a compensation of the measurement leads and to
redo it if it has not been done recently or if you have changed leads.
To compensate for the resistance of the measurement leads (see §3.16). It is essential for the selective earth meas-
urement on live circuit.
The device proposes choosing the voltage for the Ik calculation from among the following values:
ULN (the measured voltage value),
(Ik) the voltage of the old standard (for example 220 V),
the voltage of the current standard (for example 230 V).
Depending on the voltage ULN measured, the device proposes the following choices:
if 170<ULN<270 V: ULN, 220 V or 230 V.
if 90<ULN<150 V: ULN, 110 V or 127 V.
if 300<ULN<500 V: ULN, 380 V or 400 V.
kΩ
Ik To activate the alarm on Ik (in TRIP mode only).
kA
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
6% The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
Press the TEST button to start the measurement. The measurement stops automatically.
33
3.8.4. READING OF THE RESULT
11-25-2013 10:47 50 . 0 Ω 50 . 1 Hz
Measurement result.
7% L a 2.6mH
Value of the reference voltage for the
EARTH Ra Sel. calculation of Ik.
The second page is used to see the value of short-circuit current Ik, f loop impedance ZS, of loop resistance RS nd of loop induct-
ance LS.
The third page is used to see the value of the voltages ULN, ULPE, UNPE and on the rod ( ) before the measurement.
3.8.5. ERROR INDICATION ((LOOP, EARTH ON LIVE CIRCUIT, AND SELECTIVE EARTH ON LIVE CIRCUIT)
The commonest errors in the case of a loop impedance measurement or earth measurement on live circuit are:
A connection error.
An earth rod resistance that is too high (>15 kW): reduce it by packing the earth around the rod and moistening it.
A voltage on the protective conductor that is too high.
A voltage on the rod that is too high: move the rod out of the influence of the earth electrode.
Tripping in the non-tripping mode: reduce the test current.
A current measured by the clamp in selective earth on live circuit that is too low: the measurement is not possible.
The user may have picked up a charge of static electricity, for example by walking on a carpet. In this case, when he/she
presses the TEST button, the device displays the error message «earth potential too high». The user must then be dis-
charged by touching an earth before making the measurement.
For help with connections or any other information, use the help function.
34
3.9. MEASUREMENT OF THE LINE IMPEDANCE (Zi)
The loop impedance measurement Zi (L-N, L1-L2, or L2-L3 or L1-L3) is used to calculate the short-circuit current and size the protec-
tions of the installation (fuse or RCD), whatever type of neutral the installation uses.
The device then calculates the short-circuit current Ik = ULN/Zi the value of which serves to check the proper sizing of the protec-
tions of the installation.
Set the switch to Zi. Connect the measuring cable to the device, then to the socket outlet of the instal-
lation to be tested.
At the time of connection, the device first checks that the voltages present
SET UP
on its terminals are correct, then determines the position of the phase (L) and
of the neutral (N) with respect to the protective conductor (PE) and displays
it. If necessary, it then automatically switches the L and N terminals so that
the line impedance measurement can be made without modifying the con-
nections of the terminals of the device.
If you use the measuring cable that is terminated by three leads, you can connect
OFF
the PE lead (green) to the N lead (blue). Otherwise, the device cannot display the
position of the phase. But this does not prevent making the measurement.
RN N RN N
PE RE PE
Rb Ra Rb
/ /
Case of an IT installation
RL L
RN N
PE
Z Ra
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it un-
necessary to look at the display unit to check this point.
The signal can be smoothed to produce a mean of values. But the measurement then takes longer.
35
3.9.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
To select measurement of Zi (line impedance measurement) or of DV (measurement of the voltage drop in the cables,
for the C.A. 6117 only). Here, you must select Zi.
To compensate for the resistance of the measurement leads, for measurements of low values (see §3.16).
The device proposes choosing the voltage for the Ik calculation from among the following values:
ULN (the measured voltage value),
(Ik) the voltage of the old standard (for example 220 V),
the voltage of the current standard (for example 230 V).
Depending on the voltage ULN measured, the device proposes the following choices:
if 170<ULN<270 V: ULN, 220 V, or 230 V.
if 90<ULN<150 V: ULN, 110 V or 127 V.
if 300<ULN<500 V: ULN, 380 V or 400 V.
The device then deduces Isc from this and displays it.
36
Press the TEST button to start the measurement. The measurement stops automatically.
When the TEST button is pressed, the device checks that the contact voltage is less than UL. If not, it does not
make the loop impedance measurement.
If Ik is less than Isc, the fuse is not suited to the installation it protects and must be replaced.
11-25-2013 10:47 50 . 0 Ω 50 . 1 Hz
R i 0.15 Ω
To change display pages.
L i 0.8 mH
Value of the reference voltage for the
calculation of Ik.
5%
Programmed maximum contact
LOOP Zi voltage.
37
3.10. MEASUREMENT OF THE VOLTAGE DROP IN THE CABLES (DV)
For the C.A. 6117 only. The voltage drop in the cables is measured to check that the cross section of the cables is sufficient for the
installation. A voltage drop that is too large (> 5%) means that the cross section of the cables is too small.
This measurement can be made whatever the type of neutral used in the installation.
At the time of connection, the device first checks that the voltages present on
its terminals are correct, then determines the position of the phase (L) and of
the neutral (N) with respect to the protective conductor (PE) and displays it.
If necessary, it then automatically switches the L and N terminals so that the line
OFF impedance measurement can be made without modifying the connections of the
terminals of the device.
RL L F
RN
PE
Rb Ra
For the second measurement, connect the tripod cable to the instrument and to one of the outlets of the installation.
RL L F
RN
PE
Rb Ra
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it un-
necessary to look at the display unit to check this point.
The signal can be smoothed to produce a mean of values. But the measurement then takes longer.
38
3.10.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
To select the measurement of Zi (line measurement impedance) or of V (measurement of the voltage drop in the
cables). Here, you must select DV.
To compensate for the resistance of the measurement leads, for measurements of low values (see §3.16).
The device proposes choosing the voltage for the Ik calculation from among the following values:
ULN (the measured voltage value),
(Ik) the voltage of the old standard (for example 220 V),
the voltage of the current standard (for example 230 V).
Depending on the voltage ULN measured, the device proposes the following choices:
if 170<ULN<270 V: ULN, 220 V, or 230 V.
if 90<ULN<150 V: ULN, 110 V or 127 V.
if 300<ULN<500 V: ULN, 380 V or 400 V.
% 5.00
Pour régler le seuil d’alarme (voir § 3.17 ). Par défaut le
seuil est fixé à 5%.
The percentage indicates the quantity of memory already used.
Press the TEST button to start the measurement. The measurement stops automatically.
When the TEST button is pressed, the device checks that the contact voltage is less than UL. If not, it does not
make the loop impedance measurement.
TEST This symbol invites you to wait while the measurement is in progress.
If Ik is greater than Isc, the fuse is unsuited to the installation it protects and must be replaced.
39
3.10.4. READING OF THE RESULT
After the first measurement:
11-25-2013 10:47 50 . 0 Ω 50 . 1 Hz
Value of the short-circuit current.
L
1.00% Value of the reference impedance.
11-25-2013 10:47 50 . 0 Ω 50 . 1 Hz
Z ref 0.836 Ω
I n 16.00 A
5%
LOOP dV
40
3.11. TEST OF RESIDUAL CURRENT DEVICE
The device can be used to perform three types of test on RCDs:
a tripping test in ramp mode,
a tripping test in pulse mode,
a non-tripping test.
The test in ramp mode serves to determine the exact value of the tripping current of the RCD.
The test in pulse mode serves to determine the tripping time of the RCD.
The non-tripping test serves to check that the RCD does not trip at a current of 0.5 IDN. For the test to be valid, the leakage current
must be negligible with respect to 0.5 IDN and, to ensure this, all loads connected to the installation protected by the RCD that is
being tested must be disconnected.
It then calculates UF = ZS x IDN (or UF = ZS x 2 IDN or UF = ZS x 5 IDN depending on the type of test requested), which will be the
maximum voltage produced during the test. If this voltage is greater than UL, the device does not perform the test. The user can
then reduce the measurement current (to 0.2 IDN) so that the test current combined with the leakage current present in the instal-
lation will not lead to a voltage greater than UL.
For a more accurate measurement of the fault voltage, we recommend planting an auxiliary rod, as for earth measurements on
live circuits. The device then measures RA and calculates UF = RA x IDN (or UF = RA x 2 IDN or UF = ZS x 5 IDN depending on the type
of test requested).
Once this first part of the measurement has been made, the device goes on to the second part, which depends on the type of test.
For the ramp mode test, the device generates a sinusoidal current of which the amplitude increases gradually from 0.3 to 1.06
IDN between the L and PE terminals for type AC and A RCDs and from 0,2 to 2,2 IDN for type B RCDs. When the RCD opens
the circuit, the device displays the exact value of the tripping current and the tripping time. This time is an indication and may
differ from the trip time in pulse mode, which is closer to the operational reality.
For the pulse mode test, the device generates a sinusoidal current at the mains frequency, having an amplitude of IDN, 2 IDNor
5 IDN between the L and PE terminals for type AC and A RCDs and 2 IDN or 4 IDN for type B RCDs, lasting at most 500 ms. And
it measures the time the RCD takes to open the circuit. This time must be less than 500 ms.
For the non-tripping test, the device generates a current of 0.5 IDN for one or two seconds, depending on what the user has
programmed. Normally, the tripping must not trip.
In the ramp and pulse mode tests, if the RCD does not trip, the device sends a current pulse between the L and N terminals. If
the RCD trips, it is because it was incorrectly installed (N and PE reversed).
Set the switch to RCD . Connect the measuring cable to the device, then to a socket outlet included in the circuit
protected by the RCD to be tested.
SET UP At the time of connection, the device detects the positions of the phase (L) and
of neutral (N) with respect to the protective conductor (PE) and displays them. If
necessary, it then automatically switches the L and N terminals so that the test
can be done without modifying the connections of the terminals.
OFF
41
RCD
RL L
RN
PE
Rb Ra
If possible, first disconnect all loads from the network on which you test the RCD. This prevents interference with the test
by any leakage currents due to these loads.
If you have a current clamp, you can measure the leakage current (see §3.12) at the RCD and so make allowance for it during
the test.
To make a more accurate measurement of the fault voltage, plant the auxiliary rod at a distance of more than 25 metres
from the earth electrode and connect it to the (RA SEL) terminal of the device. The symbol is then displayed.
RCD
RL L
RN
PE
Rb Ra
> 25 m
/
Particular case:
To test a residual current device located downstream of another residual current device having a smaller nominal current, you
must use the measuring cable terminated by 3 leads and make the connections shown opposite (upstream-downstream method).
RCD RCD
RL L
RN
30 mA 300 mA
PE blue green red
Rb Ra
42
3.11.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
Choice of the nominal current of the residual current device IDN: VAR. (variable: the user programs a value be-
30 mA tween 6 and 999 mA; this choice does not exist for type B RCDs), 10 mA, 30 mA, 100 mA, 300 mA, 500 mA,
650 mA, or 1000 mA.
Choice of type of residual current device: STD (standard), S or G (the S type is tested with a current of 2 IDN
as default).
Choice of the form of the test signal:
To restore the factory adjustment parameters: IDN = 30 mA, STD and signal types.
To perform a prior check of voltage UF, choose a test current: 0.2, 0.3, 0.4, or 0.5 IDN .
To make a faster measurement by eliminating the prior check of voltage UF, choose: --x--.
0.3 I∆N
To activate or deactivate the audible voltage alarm (the threshold being equal to UL).
This function makes it possible to locate, on the distribution panel, using the audible signal, the RCD
protecting a remote current socket outlet (typical case of a panel at a distance from the socket outlet)
without being in the immediate vicinity of the device.
Press the TEST button to start the measurement. The measurement stops automatically.
In the case of type S or G circuit-breakers, the device counts 30 seconds between the prior test of UF and the test of
the RCD itself, in order to allow its demagnetization. This wait can be cut short by pressing the TEST button again.
43
3.11.4. READING OF THE RESULT
11-25-2013 10:47 50 . 1 Hz
L
30 mA
UF = ZS x IA or RA x IA.
RCD : Ia
Type of signal.
Type of RCD.
Programmed maximum contact
voltage.
44
3.11.5. MAKING A TEST IN PULSE MODE
Connect the measuring cable to the device, then to a socket outlet included in the circuit
Set the switch to RCD . protected by the circuit-breaker to be tested.
At the time of connection, the device detects the positions of the phase (L) and
of neutral (N) with respect to the protective conductor (PE) and displays them. If
SET UP
necessary, it then automatically switches the L and N terminals so that the test
can be made without modifying the connections of the terminals of the device.
RCD
RL L
OFF N
RN
PE
Rb Ra
For a more accurate measurement of the fault voltage, plant the auxiliary rod at a distance of more than 25 metres from
the earth electrode and connect it to the (RA SEL) terminal of the device. The symbol is then displayed.
RCD
RL L
RN
PE
Rb Ra
> 25 m
/
Particular case:
To test a residual current device located downstream of another residual current device having a smaller nominal current, you
must use the measuring cable terminated by 3 leads and make the connections shown opposite (upstream-downstream method).
RCD RCD
RL L
RN
30 mA 300 mA
PE blue green red
Rb Ra
If it is active, the alarm on the tripping time informs the user by an audible signal, that the measurement is outside the range limits,
so there is no need to look at the display unit.
45
3.11.6. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
Choice of the nominal current of the residual current device IDN: VAR. (variable: the user programs a value between
30 mA 6 and 999 mA; this choice does not exist for type B RCDs), 10 mA, 30 mA, 100 mA, 300 mA, 500 mA, 650 mA or
1,000 mA.
Choice of type of residual current device: STD (standard), S or G (the S type is tested with a current of 2 IDN as
default)
Choice of pulse current in multiples of IDN : x1, x2, x4, x5, x0,5/1s, x0,5/2s or UF . The 2 values at 0.5 IDN are used
to perform a non-tripping test.
Choice of the form of the test signal:
Depending on the type of fuse and the form of the test signal, only some values of the pulse current are possible.
To restore the factory adjustment parameters: IDN = 30 mA, STD type RCD, pulse current = IDN and
signal .
To perform a prior check of voltage UF, choose a test current: 0.2, 0.3, 0.4, or 0.5 IDN.
To make a faster measurement by eliminating the prior check of voltage UF, choose: --x--.
0.3 I∆N measure ZS or calculate UF . This mode allows a faster test of the RCD.
46
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
6% The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
Press the TEST button to start the measurement. The measurement stops automatically.
In the case of type S or G RCD, the device counts 30 seconds between the prior test of UF and the test of the RCD
itself, in order to allow its demagnetization. This wait can be cut short by pressing the TEST button again.
L
UF = ZS x IA or RA x IA.
30 mA for type AC and A RCDs. UF is not
calculated for type B RCDs.
Ta 1 7 1.6 m s
Case where: TAmin < TA < TAmax.
8% Type of signal.
47
In the case of a non-tripping test in pulse mode:
11-25-2013 10:47 50 . 1 Hz
L
30 mA UF = ZS x IA or RA x IA.
Ta > 1.0 0 s
8% Type of signal.
For help with connections or any other information, use the help function.
48
3.12. CURRENT AND LEAKAGE CURRENT MEASUREMENT
This measurement requires the use of a specific optional current clamp.
It can measure very low currents (of the order of a few mA) like fault currents or leakage currents, and high currents (of the order
of a few hundred Amperes).
Two of the four points of connection of the clamp serve to identify the type of clamp (x 1,000 or x 10,000) and the other two to
measure the current. Knowing the ratio of the clamp, the device displays a direct reading of the current.
Set the switch to . Connect the clamp to the terminal on the device. The symbol is then dis-
played. Actuate the trigger to open the clamp and encircle the conductor to be meas-
ured. Release the trigger.
SET UP
I
OFF /
The current measurement can be made on different conductors of an installation. This is why it has been made possible to index
the value recorded with one of the following values:
1, 2, 3, N, PE, or 3L (sum of the phase currents or phase and neutral currents, to measure the leakage current).
m A 200.0 To set the alarm threshold (see §3.17). As default, the threshold is set to 200 A.
A
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
6% The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
Press the TEST button once to start the measurement and a second time to stop it.
TEST
49
3.12.4. READING OF THE RESULT
CURRENT
The clamp is connected.
Here, the conductor passes through the clamp 4 times. You will
have to divide the measured current by 4 to know the true value
of I.
I /
green
red
The frequency is too unstable for the measurement. In this case connect the cor-
blue
responding mains voltage between L and PE. The device will then synchronize to
the frequency of the voltage and will be able to measure the current at this same
frequency. I
/
The current measured by the clamp is too high. Use a clamp having a higher ratio.
For help with connections or any other information, use the help function.
50
3.13. DIRECTION OF PHASE ROTATION
This measurement is made on a three-phase network. It is used to check the phase order of the network.
Connect the measuring cable terminated by 3 leads to the device and to each of the phases:
Set the switch to .
the red to L1, the blue to L2, and the green to L3.
SET UP red
L1
blue
L2
green
L3
OFF PE
Press the TEST button once to start the measurement and a second time to stop it.
TEST
11-25-2013 10:47 50 . 1 Hz
U12: 379 V
U23: 380 V Voltages between the phases.
U31: 378 V
10 %
PHASE SEQUENCE
51
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
6% The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
For help with connections or any other information, use the help function.
52
3.14. POWER MEASUREMENT
This measurement requires the use of the optional specific MN77 current clamp. It can be made on a single-phase network or
on a three-phase network that is balanced in voltage and in current.
For a three-phase network balanced in voltage and in current, the device measures one of the three phase-to-phase voltages, mul-
tiplies it by the current of the third phase, then multiplies the result by √3. Example: P3f = U12 x I3 x √3
N
green
blue
PE
OFF
In the case of a three-phase network balanced in voltage and in current, connect the measuring cable terminated by 3 leads
to the device and to two of the three voltages U12, U23 or U31 using the red and green leads. Then, with the clamp, surround the
conductor of the third phase I3 (for U12), I1 (for U23) or I2 (for U31).
red
L1
green
L2
L3
blue
The power measurement can be made on different phases of an installation. This is why it has been made possible to index the
recorded power value with one of the following values: 1, 2, or 3 (single-phase measurements on a three-phase network).
Press the TEST button once to start the measurement and a second time to stop it.
TEST
53
3.14.4. READING OF THE RESULT
11-25-2013 10:47 50 . 1 Hz
Measurement result.
The + sign indicates power con-
sumed. The - sign indicates power
+ 64 W returned.
If the phase of the current with respect to the voltage is not correct, turn the clamp around, with the help of the arrow marked on
the jaws, in order to reverse the phase by 180°.
Press this function key to display the voltage and current curves, as on an oscilloscope. If the clamp is not con-
nected, only the voltage curve is displayed. The current curve cannot be displayed alone.
11-25-2013 10:47 50 . 1 Hz
URMS = 232.5 V
Numerical values.
UMAX = 328.8 V
IRMS = 278.1 mA
IMAX = 393.3 mA
POWER
54
3.14.5. POWER FACTOR
In the case of sinusoidal signals, the sign of cos j indicates whether the measurement is being made on a generator (cos j <
0) or on a receiver (cos j > 0). The power factor, PF, can be regarded as equivalent to cos j but generalized to non-sinusoidal
signals, which is often the case with currents.
However, on the instrument, the sign of the PF is treated conventionally, meaning that it indicates only the phase advance or delay
(inductive or capacitive load) and not whether a receiver or a generator is involved.
The phase angle is counted algebraically. It represents the angular difference of the voltage vector with respect to the current
vector, taken as reference.
i(t)
ϕ i
time
V(t)
+ ϕ
V
For help with connections or any other information, use the help function.
55
3.15. HARMONICS
This function is used to display the harmonic analysis of a voltage or current of which the signal is steady-state or quasi-steady-
state. It is used to establish a first diagnostic of the harmonic pollution of an installation.
The current analysis requires the use of the MN77 current clamp (optional).
Connect the measuring cable terminated by 3 leads to the device and to a socket outlet of the
Set the switch to . installation to be tested, using the red and green leads.
Or connect the MN77 clamp to the device and encircle the phase.
SET UP L
red
N
green
blue
PE
OFF
linear scale,
LIN
logarithmic scale,
LOG
Choice of calculation of the level of distortion with respect to the fundamental (THD-F) or of the distortion factor
F with respect to the RMS amplitude (THD-R or DF).
Press the TEST button once to start the measurement and a second time to stop it.
TEST
56
3.15.4. READING OF THE RESULT
11-25-2013 10:47 50 . 0 Hz
log
10
1
Representation of the harmonics.
F
HARMONICS U
The frequency and amplitude of the selected harmonic (in black) are indicated at the bottom of the graph. To select another
harmonic, use the keys. The device then shifts from the fundamental (H1) to harmonic H2, then to harmonics (H3, H4, ...,
H25). And on the next page it sweeps the harmonics from H26 to H50.
11-25-2013 10:47 50 . 0 Hz
You must scroll through 6 other screens using the key to display the values of all 50 harmonics.
57
3.15.5. ERROR INDICATION
The commonest errors in the case of an analysis of a signal into harmonics are:
The voltage is outside the measurement range.
The frequency is outside the measurement range.
The current is too low to be measured.
The signal is not steady-state.
For help with connections or any other information, use the help function.
58
3.16. COMPENSATION FOR THE RESISTANCE OF THE MEASUREMENT LEADS
Compensation for the resistance of the measurement leads serves to neutralize their values and obtain a more accurate meas-
urement when the resistance to be measured is low. The cords are already compensated in the plant; you must perform a new
compensation if you use cords other than those provided.
The device measures the resistance of the accessories (leads, probe tips, crocodile clips, etc.) and subtracts this value from the
measurements before displaying them.
Compensation for the resistance of the measurement leads is possible in continuity, 3P earth, and loop tests. It is different for
each of these functions. It must be renewed at each change of accessories.
The current value(s) of the compensation is(are) displayed at top right. A value of zero indicates that no compensation has
been determined. The R∆
0 symbol, present on the bottom strip of the display unit, reminds you that the resistance of the
leads is compensated.
3.16.1. IN CONTINUITY
R∆: 0.052 Ω
Connect the two leads that you are going to use for the measurement to
the W and COM terminals, short-circuit them, then press the TEST button.
The device measures the resistance of the leads and displays it. Press OK
to use this value or to keep the old value.
3.16.2. IN 3P EARTH
R∆: 0.750 Ω
Connect the lead that you are going to use to connect the E terminal to the
earth between the H and E terminals, then press the TEST button.
The device measures the lead and displays its value. Press OK to use this
value or to keep the old value.
The device measures each of the three leads and displays their values.
Press OK to use this value or to keep the old values.
59
3.16.4. ELIMINATING THE COMPENSATION
Proceed as for compensation, but rather than short-circuiting the leads, leave them disconnected. Then press the TEST button.
R∆
The device removes the compensation, then returns to voltage measurement. The 0 symbol disappears from the display unit
and the icon is crossed out.
3.16.5. ERROR
If the resistance of the measurement leads is too high (>2.5 W per lead), compensation is impossible. Check the connections
and any junctions and leads that might be open-circuit.
If, during a continuity, 3P earth, or loop impedance measurement, you obtain a negative measurement result, you must have
changed the accessories without redoing the compensation. In this case, perform a compensation with the accessories you
are now using.
60
3.17. ADJUSTMENT OF THE ALARM THRESHOLD
The device makes an audible signal and the indicator flashes:
in continuity, resistance and insulation measurement, if the measurement is below threshold;
for earth and loop measurements and measurements of the voltage drop in the cables, if the measurement is above threshold;
for short-circuit current measurements, if the measurement is below threshold;
in test of residual current device, if the measurement is not between the two thresholds (Tmin and Tmax).
The alarm threshold is adjusted in essentially the same way for all measurements.
Start by entering the alarm function by pressing the or key.
kΩ
Ω 002.00 Using the keys, choose the unit of the alarm threshold you want to set: W or kW. Depending
on the function chosen, MW, mA, A, kA, and ms are also possible.
kΩ
Ω 002.00 Using the key, move the cursor to the value of the threshold.
kΩ
Using the keys, modify the selected digit. Then move the cursor to the next digit to
Ω 042.00 modify it, and so on.
kΩ
61
4. ERROR INDICATION
OK
Press the OK key or the key to exit from the help function.
62
4.1. NO CONNECTION
>40.0W
< 5.0 V } The value is outside the measurement range of the device. The minimum and maximum
values depend on the function.
The voltage is regarded as dangerous from 25, 50, or 65V, depending on the value of UL programmed in SET-UP.
For measurements made without voltage (continuity, insulation, and 3P earth), if the device detects a voltage, it
disables starting of the measurement by the pressing of the TEST button and displays an explanatory error message.
For measurements that are made on live circuits, the device detects the absence of voltage, the absence of a protective conduc-
tor, a frequency or voltage outside the measurement range. When the TEST button is pressed, the device then disables starting
of the measurement by the pressing of the TEST button and displays an explanatory error message
If the device detects an error in the measurement configuration or in the connection, it displays this symbol and a
corresponding error message.
63
4.6. CHECK OF INTERNAL PROTECTION DEVICES
The device includes two internal protection devices that cannot be reset and that the user cannot replace. These devices act only
under extreme conditions (e.g. a lightning strike).
SET UP
OFF
/
If the internal protection devices are intact, the display should indicate:
11-25-2013 10:47 50 .0 Ω 50 . 1 Hz
6 mA
L-PE V
L-N V
N-PE V
6%
LOOP Zs
If UL-PE does not display – x --, the If UN-PE does not display – x --, the Case where both protections have
protection in the L terminal has been protection in the N terminal has been been activated.
activated. activated.
6 mA 6 mA 6 mA
L-PE 0.0 V L-PE V L-PE 0.0 V
In these last three cases, the device must be sent in for repair (see §10.5).
64
5. SET-UP
11-21-2013 11:17
Set the switch to SET-UP.
SET UP
OFF
SET-UP
OK Use the directional keypad to select an icon, select a field, and modify it.
This key is used to exit from the current screen without saving.
next page
To set the date and time and choose the display format.
65
Adjustment of the time to automatic switching off of the device: 5 min (default), 10 min, 30 min, or ∞ (per-
manent operation).
To return to the factory configuration (compensation for resistance of measurement leads and all adjustable
parameters in the various measurements). The device requests confirmation before executing.
General configuration
Audible signal: activated
UL = 50 V
Duration of lighting of the backlighting: 2 min.
Duration of operation of the device before automatic switching off: 5 min.
Date and time format: DD/MM/YYYY and 24 h.
Language: English.
Insulation measurement
Test voltage: 500 V.
Alarm activated.
Alarm threshold: 1 MΩ.
Measurement of loop impedance (ZS), of earth on live circuit, and of selective earth resistance on
live circuit
Measurement current: 6 mA.
Compensation of the cords: 30 mΩ, 30 mΩ, 30 mΩ respectively for R∆L, R∆N, R∆PE (measuring cable with
mains plug).
UREF = UMEAS.
Alarm deactivated.
No smoothing of the measurement.
66
Test of RCD
Nominal range I∆N = 30 mA.
Type of circuit-breaker: Standard (STD).
Test waveform: sinusoidal signal that begins with a positive half-wave.
Test current for determination of UF = 0.3 I∆N.
Alarm deactivated.
Audible RCD identification function: deactivated.
Power measurement
Single-phase network.
Harmonics
No default configuration. Each time the device is started up, the configuration is:
Voltage harmonics.
Display in bar-chart form with linear ordinates.
Calculation of the total distortion referred to the fundamental (THD-F).
67
6. MEMORY FUNCTION
SITE 1
ROOM 1
OBJECT 1
OBJECT 2
ROOM 2
OBJECT 1
SITE 2
ROOM 1
...
Navigation in the tree is done using the directional keypad. The titles of the SITES, ROOMS, and OBJECTS can be parameter-
ized by the user.
If a SITE or ROOM is preceded by the sign, it means that this level has sub-levels that can be expanded using the key or
the OK key. The sign is then replaced by the sign.
To compress the tree (change from the sign to the sign), use the or OK key.
Measurements are always recorded on an OBJECT. In the OBJECT, measurements are classified by TYPE OF TEST (continuity,
insulation, loop, etc.). Each OBJECT can contain up to nine TESTS belonging to the same TYPE OF TEST. Each TEST corresponds
to one measurement.
To see the tests contained in an OBJECT, go to the OBJECT and press the OK key.
A status symbol displayed to the right of the OBJECTS, of the TYPES OF TEST, and of the TEST indicates:
that the OBJECT has not yet been tested,
R that all TESTS of the OBJECT are OK,
T that at least one TEST of the OBJECT is not OK.
If you want to record the measurement you have just made, press the key corresponding to the record icon.
For a measurement to be «recordable», the TEST button must have been pressed. It is not possible to record voltage
measurements alone.
68
The following screen then appears:
To delete an element.
INSULATION
Erase a letter.
INSULATION
You can then rename the SITE. Start by erasing the existing text. Then move about on the keyboard using the directional keypad
( ) and validate each letter by pressing the OK key.
69
To add a ROOM to a SITE, place the cursor on the chosen SITE and press the key. Give the ROOM a name and validate it.
Then press the key again to create an OBJECT in the ROOM. This results in the following tree:
Site1
Room1
Object1
House
Entry
Switchboard
INSULATION
To avoid losing time when you are making the measurements, you can prepare your tree in advance.
For insulation, loop impedance, line impedance, current, and power measurements and the harmonic analysis, the device pro-
poses indexing your measurement, because several measurements are possible.
INSULAT. LN - PE
INSULAT. L-N
INSULAT. L - PE
INSULAT. N - PE
INSULAT. L1 - PE
INSULAT. L2 - PE
INSULAT. L3 - PE
INSULAT. L1 - N
INSULAT. L2 - N
INSULAT. L3 - N
INSULAT. L1 - L2
INSULAT. L2 - L3
INSULAT. L1 - L3
INSULATION
Using the arrows, select the type of insulation measurement you have just made and validate by pressing the OK key.
70
You can in this way make several insulation measurements on the electrical panel. And then move on to another type of measure-
ment, still on the electrical panel, for example a loop impedance measurement.
LOOP Zs
LOOP L - PE
As in the case of insulation, you can
LOOP L1 - PE
index the measurement.
LOOP L2 - PE
LOOP L3 - PE
LOOP Zs
To see all of the measurements made on the selected OBJECT, press the OK key.
INSULATION
71
Press the OK key to expand a TYPE OF TEST.
INSULATION
OK
11-25-2013 10:47 500 kΩ - - .- Hz
7 s
2%
72
6.6. ERASURE
You can erase a SITE, a ROOM, an OBJECT or a record either when creating the tree or while reading in memory.
Move the cursor onto the element to be erased using the keys of the directional keypad ( ).
Site1
Room1
Object1
House
Entry
Switchboard
Socket1
Kitchen
Socket1
Socket2
Socket3
INSULATION
Press the key to erase ROOM1. The device asks you to confirm by pressing the OK key or abort by pressing the key.
If the number of measurements recorded is large, the erasure may last several minutes.
6.7. ERRORS
The commonest errors during storage are the following:
The name given already exists. Change the name or index it (ROOM1, ROOM2, etc.)
The memory is full. You must eliminate at least one OBJECT to be able to record your new measurement.
It is not possible to record a measurement in a SITE or a ROOM. You must create an OBJECT in a ROOM or access an exist-
ing OBJECT and record the measurement there.
73
7. DATA EXPORT SOFTWARE
Start by installing the driver and the software pro- Then connect the device to the PC using the USB cord
grams using the CD provided with the C.A. 6116. provided with the instrument and removing the cover
that protects the USB port of the device.
Then, turn the device on by turning the switch to When the device is in communication with a PC, it
any setting. does nothing else and its keys are inactive. It then
displays the following message:
SET UP
OFF
To use the data export software, refer to the help functions of the software.
Once the USB cord has been disconnected, the device restarts after a few seconds.
74
8. TECHNICAL CHARACTERISTICS
The intrinsic uncertainty is the error defined under the reference conditions.
The operating uncertainty includes the intrinsic uncertainty plus the effects of variation of the quantities of influence (supply
voltage, temperature, interference, etc.) as defined in standard IEC-61557.
The device is not designed to operate when the charger is connected. The measurements must be made using the battery.
75
Measurements of potential of the voltage probe
The characteristics are the same as in the voltage measurements except that the input impedance is 200 kW.
This voltage must normally be between 0 and UL.
200 mA current
12 mA current
76
8.2.4. RESISTANCE MEASUREMENTS
Particular reference conditions:
External voltage on the terminals: zero.
Inductance in series with the resistance: zero.
50V range:
Value for guidance
Uncertainty of operation ± (12% + 3 ct) ± (10% + 2 ct) ± (10% + 2 ct)
Other ranges:
± (10% + 2 ct)
I = 1 mA
UN
RN = UN / 1 mA
77
Typical measurement settling time as a function of the elements tested
These values include influences due to the capacitive component of the load, to the automatic range system, and to the regula-
tion of the test voltage.
Measurement range 0.50 - 39.99 Ω 40.0 - 399.9 Ω 400 - 3999 Ω 0.20 - 15.00 kΩ 1
Resolution 0.01 Ω 0.1 Ω 1Ω 10 Ω
Intrinsic uncertainty ± (2% + 10 ct) ± (2% + 2 ct) ± (10% + 2 ct)
Operating uncertainty ± (9% + 20 ct) ± (9% + 5 ct) -
Typical peak-to-peak measure-
4.3 mA 4.2 mA 3.5 mA -
ment current 2
Measurement frequency 128 Hz
No-load voltage 38.5 V peak-to-peak
1: the 40 kW display range is used only for measurements of the RH and RS rods.
2: current at mid-range with RH = 1000 W.
78
Characteristics in trip mode:
Measurement range 0.100 - 0.500 Ω 0.510 - 3.999 Ω 4.00 - 39.99 Ω 40.0 - 399.9 Ω
The duration of the measurement depends on the voltage of the installation, on the measured impedance value, and on the activa-
tion of the smoothing filter (SMOOTH).
If smoothing is activated (SMOOTH mode), the instability of the intrinsic uncertainty is then halved (for example: ±5 digits be-
comes ±2.5 digits).
Measurement range 0.20 - 0.99 Ω 1.00 - 1.99 Ω 2.00 - 39.99 Ω 40.0 - 399.9 Ω 400 - 3999 Ω
Resolution 0.01 Ω 0.1 Ω 1Ω
RMS measurement current choice of 6. 9. or 12 mA
Intrinsic uncertainty on the im-
± (15% + 10 ct) ± (15% + 3 ct) ± (10% + 3 ct) ± (5% + 2 ct)
pedance measurement 4
Intrinsic uncertainty on the
± (15% + 10 ct) ± (15% + 3 ct) ± (10% + 3 ct) ± (5% + 2 ct)
resistive part
Intrinsic uncertainty on the
± (10% + 10 ct) ± (10% + 3 ct) ± (10% + 3 ct) ± (5% + 2 ct)
inductive part
Operating uncertainty on the
± (20% + 10 ct) ± (20% + 3 ct) ± (12% + 3 ct) - -
impedance measurement
4: There is no measurement of the inductive in L-PE loop part with a low current.
The intrinsic uncertainty is defined for 0.1 ≤ RL / RN ≤ 10 with RL and RN ≥ 1 Ω.
The duration of the measurement depends on the voltage of the installation, on the measured impedance value, and on the ac-
tivation of the smoothing filter (SMOOTH).
If smoothing is activated (SMOOTH mode), the instability of the intrinsic uncertainty is then halved (for example: ±5 digits becomes
±2.5 digits) and the duration of the measurement is of the order of 30 s.
79
Characteristics in selective mode:
The duration of the measurement depends on the voltage of the installation, on the measured impedance value, and on the activa-
tion of the smoothing filter (SMOOTH).
The measurement current corresponds to the test current indicated in the table of characteristics in tripping mode divided by the
ratio RSEL/RA avec RSEL/RA ≤ 100. Beyond this, the maximum current, 20 mA peak, is reached.
80
8.2.10. VOLTAGE DROP IN THE CABLES
Particular reference conditions:
Voltage of the installation: 90 to 500 V.
Stability of the voltage source: <0.05%.
Frequency of the installation: 15.8 to 17.5 Hz and 45 to 65 Hz.
Resistance of the leads: zero or compensated.
Impedance of the inductive part: < 0.1 x the resistive part of the impedance measured
Limitation of the ranges accessible as a function of the voltage for type AC and A RCDs
Limitation of the test current as a function of the nature of the test signal for type AC and A RCDs
Depending on the range IDN chosen and the nature of the test signal, some test modes will be unavailable. This test of coherence
is performed when the test of RCDs is started..
Wave or
Ramp
IDN pulse
2 x IDN pulse IDN ≤ 500 mA
5 x IDN pulse IDN ≤ 200 mA
Wave or
81
Characteristics in pulse mode for type AC and A RCDs:
10 mA - 30 mA - 100 mA - 300 mA - 500 mA - 650 mA - 1000 mA
Range IDN
Variable (6 to 999 mA) 6
Tripping test (se-
Nature of the test Determination of UF Non-tripping test Tripping test Tripping test
lective)
Test current 0.2 x IDN … 0.5 x IDN 7 0.5 x IDN IDN 2 x IDN 5 x IDN
Intrinsic uncertainty on the test +0 -7% +0 -7% -0 +7% -0 +7% -0 +7%
current ± 2 mA ± 2 mA ± 2 mA ± 2 mA ± 2 mA
Maximum duration of applica- from 32 to 72
1000 or 2000 ms 8
300 ms 150 ms 40 ms
tion of the test current periods
6: the upper limit of the variable range (999 mA) depends on the nature of the test performed and on the type of test current (half
or full wave).
7: this current can be adjusted in steps of 0.1 IDN and must not be less than 4 mA. As default, this current is 0.3 IDN.
8: to be chosen when configuring the measurement.
Characteristics of the fault voltage calculation (UF) for type AC and A RCDs:
Calculation formula:
UF = ZLPE x IDN or ZA x IDN or RA x IDN or ZLPE x 2IDN if the test is at 2IDN
82
Limitation of the ranges available as a function of the voltage for type B RCDs
90 - 280 V
280 - 550 V
Limitation of the test current as a function of the nature of the test signal for type B RCDs
Depending on the range IDN chosen and the nature of the test signal, some test modes will be unavailable. This test of coherence
is performed when the test of RCDs is started.
Wave or
Ramp
2 x IDN pulse
4 x IDN pulse
83
8.2.12. CURRENT MEASUREMENT
Particular reference conditions:
Peak factor = 1,414
DC component< 0.1 %.
Frequency: 15.8 450 Hz.
When a voltage is connected between the L and PE terminals, the device synchronizes to the frequency of this voltage, allowing
current measurements from 1 mA.
Measurement range 5.0 - 399.9 mA 0.400 - 3.999 A 4.00 - 39.99 A 40.0 - 199.9 A
Resolution 0.1 mA 1 mA 10 mA 100 mA
Intrinsic uncertainty ± (2% + 5 ct) ± (1.5% + 2 ct) ± (1% + 2 ct) ± (1.2% + 2 ct)
When a voltage is connected between the L and PE terminals, the device synchronizes to the frequency of this voltage, allowing
current measurements from 5 mA.
Characteristics:
The phase order is «positive» if rotation L1-L2-L3 is anticlockwise.
The phase order is «negative» if rotation L1-L2-L3 is clockwise.
The three voltages are measured (see the characteristics in §8.2.1) and indicated as U12, U23 and U31.
84
8.2.14. POWER MEASUREMENTS
Particular reference conditions:
Sinusoidal voltage and current signals: cosj = 1.
Voltage ≥ 10 V.
Current ≥ 0.1 A (for the C177A clamp).
Frequency: 15.8 to 17.5 Hz and 45 to 65 Hz.
No DC component.
40.0 - 110.0 kW 13
Measurement range 5 - 3999 W 4.00 - 39.99 kW
40.0 - 330.0 kW
Resolution 1W 10 W 100 W
Intrinsic uncertainty ± (2% + 5 ct) ± (2% + 2 ct) ± (2% + 2 ct)
13: full scale is 110 kW (550V x 200A) in single-phase and 330 kW in three-phase.
The sign of the power factor depends on whether the phase of the voltage leads or lags that of the current. This can be used to
determine whether the load is inductive (+ sign) or capacitive (- sign).
8.2.16. HARMONICS
Particular reference conditions:
Signal without inter-harmonics, of which the fundamental is stronger than the other harmonic components and than the DC
component.
Frequency of the fundamental: 16.66 Hz, 50 Hz, or 60 Hz ± 0.05 Hz.
Peak factor of the signal ≤ 4.
Characteristics:
10 to 500 V, the display range being determined by the value of the
Characteristics of voltage display
strongest harmonic component.
1 to 200 A, the display range being determined by the value of the
Characteristics of current display
strongest harmonic component.
Stability of the current and voltage display ± 2 ct
Domain of use Harmonics of orders 1 to 50
Measurement range for the harmonic factor 0.2 - 399.9 %
Detection threshold for the harmonic factor 0.1 %
Measurement range in THD-F and THD-R 0.2 - 100 %
Resolution for the harmonic factor, THD-F and THD-R 0.1%
Factor > 10% and order < 13: 5 ct
Intrinsic uncertainty on the RMS value and the harmonic Factor <10% and order < 13: 10 ct
factor Factor > 10% and order > 13: 10 ct
Factor > 10% and order > 13: 15 ct
Intrinsic uncertainty on the THD-F and THD-R 10 ct
85
Method and definitions:
Determination of harmonics: Cooley-Tukey FFT algorithm on 16 bits
Sampling frequency: 256 times the frequency of the fundamental
Filtering window: rectangular, 4 periods
n=50
√ ΣH² n=2
n
THD-F =
H1
THD-R: Total distortion referred to the RMS value of the signal (also called DF: distortion factor).
n=50
√ ΣH² n=2
n
THD-R =
R[RMS]
86
8.3.3. RESISTANCE AND CONTINUITY MEASUREMENT
87
8.3.6. EARTH MEASUREMENT ON LIVE CIRCUIT, LOOP AND SELECTIVE EARTH
8.3.9. POWER
88
8.4. INTRINSIC UNCERTAINTY AND OPERATING UNCERTAINTY
The installation testers complie with standard IEC-61557, which requires that the operating uncertainty, called B, be less than 30%.
In loop measurement, B = ± ( |A| + 1,15 √ E1² + E2² + E3² + E6² + E7² + E8² )
with E6 = influence of the phase angle from 0 to 18°.
E7 = influence of the network frequency from 99 to 101% of the nominal frequency.
E8 = influence of the network voltage from 85 to 110% of the nominal voltage.
In earth measurement, B = ± ( |A| + 1,15 √ E1² + E2² + E3² + E4² + E5² + E7² + E8² )
with E4 = influence of the interference voltage in series mode (3 V at 16.6, 50, 60, and 400 Hz)
E5 = influence of the resistance of the rods from 0 to 100 x RA but ≤ 50 kW.
In test of residual current device, B = ± ( |A| + 1,15 √ E1² + E2² + E3² + E5² + E8² )
with E5 = influence of the resistance of the probes within the limits indicated by the manufacturer.
The instrument is not designed to operate when the charger is connected. The measurements must be made using battery
power.
The battery charger of the device is in two distinct parts: an external power supply and a charger built into the device.
The charger manages the charging current, the battery voltage, and the internal temperature of the battery simultaneously. This
optimizes charging while guaranteeing a long battery life.
The day before you use your device, check its charge condition. If the battery level indicator shows less than three bars, charge
the device overnight (see §1.2).
Following prolonged storage, the battery may be fully discharged. In this case, the first charge may take longer.
89
Set the switch to OFF; charging is possible when the device is not off, but will take longer.
The battery is at the bottom of the device, and the evacuation of the heat can be facilitated by placing the device upright while
charging. The battery temperature is then lower and it will be charged more fully.
This precaution is especially important when the air temperature is high (in summer).
How long your device can operate when the battery is fully charged depends on several factors:
The consumption of the device, which depends on the measurements you make,
The capacity of the battery. It is greatest when the battery is new, and declines as the battery ages.
14: If the device is to be left unused for more than 2 months, remove the battery. To keep it fully charged, recharge it every 4 to
6 months.
A: With automatic switching off after 10 minutes, one measurement every 30 minutes, 7 hours a day.
B: With one 5-second measurement every 25 seconds and a programmed automatic switching-off time.
C: With 5 consecutive 10-second measurements every 10 minutes and a programmed automatic switching-off time.
D: With 5 consecutive 5-second measurements every minute and a programmed automatic switching-off time.
E: With 5 consecutive 30-second measurements every 3 minutes and a programmed automatic switching-off time.
90
8.5.5. END OF BATTERY LIFE
The internal resistance of a battery at the end of its life is high. The result is an abnormally short charging time.
After a full charge, the device indicates “charging over”, but as soon as the charger is disconnected, the display unit loses its
contrast and goes off, meaning that the battery no longer holds a charge.
15: This range corresponds to the range of the operating uncertainty defined by standard IEC-61557. When the device is used
outside this range, it is necessary to add 1.5%/10°C and 1.5% between 75 and 85% RH to the operating uncertainty.
Protection class IP 53 per IEC-60 529 if the cover of the USB port is closed, IP 51 if it is open.
IK 04 per IEC-50102
91
9. DEFINITIONS OF SYMBOLS
Here is a list of the symbols used in this document and on the display unit of the device.
92
TA effective trip time of the residual current device.
THD-F level of harmonic distortion referred to the fundamental.
THD-R level of harmonic distortion referred to the RMS value of the signal.
TN type of link to earth defined in standard IEC-60364-6.
TT type of link to earth defined in standard IEC-60364-6.
U12 voltage between phases 1 and 2 of a three-phase network.
U23 voltage between phases 2 and 3 of a three-phase network.
U31 voltage between phases 3 and 1 of a three-phase network.
UC contact voltage between conducting parts when they are touched simultaneously by a person or an animal (IEC-
61557).
UF fault voltage appearing during a fault condition between accessible conducting parts (and/or external conducting
parts) and the reference frame ground (IEC-61557).
UFk fault voltage, in the event of a short-circuit, according to Swiss standard SEV 3569.
UFk = Ik x ZA = UREF x ZA/ZS.
UH-E voltage measured between terminals H and E.
UL conventional maximum contact voltage (IEC-61557).
UL-N voltage measured between the L and N terminals.
UL-PE voltage measured between the L and PE terminals.
UN nominal test voltage in insulation measurement, generated between the MW and COM terminals.
UN-PE voltage measured between the N and PE terminals.
UPE voltage between the PE conductor and the local earth measured when the user presses the TEST button.
UREF reference voltage for calculation of the short-circuit current.
US-E voltage measured between the S and E terminals.
ZA earth impedance in earth measurement on live circuit.
ZS impedance in the loop between the phase and the protective conductor.
Zi impedance in the loop between the phase and the neutral or between two phases (line loop impedance).
ZL-N impedance in the L-N loop.
ZL-PE impedance in the L-PE loop.
93
10. MAINTENANCE
Except for the battery, the instrument contains no parts that can be replaced by personnel who have not been specially
trained and accredited. Any unauthorized repair or replacement of a part by an “equivalent” may gravely impair safety.
10.1. CLEANING
Disconnect anything connected to the device and set the switch to OFF.
Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry rapidly with a dry cloth or forced air. Do not use
alcohol, solvents, or hydrocarbons.
To keep the device safe, replace the battery only with the original model. Do not use a battery with a damaged jacket.
Replacement procedure:
1. Disconnect anything connected to the device and set the switch to OFF.
MADE IN FRANCE
2. Turn the device over and insert a screwdriver into the hole in the battery pack.
TO AVOID ELECTRICAL SHOCK
DISCONNECT LEADS, PROBES AND POWER SUPPLY
BEFORE REMOVING LI-ION MODULE
3. Then push the screwdriver towards the rear and the battery slides out of its compartment.
Spent batteries must not be treated as ordinary household waste. Take them to the appropriate recycling collection
point.
The internal clock of the instrument continues to run for at least 60 minutes with the battery out.
4. Insert the new battery pack in its compartment and press it home.
94
10.3. RESETTING THE DEVICE
If the device crashes, it can be reset, like a PC.
Set the switch to ZS (RA/SEL.). Press the 3 keys indicated below simultaneously.
SET UP
/
SET UP
OFF
OFF
OK
Like all measuring or testing devices, the instrument must be checked regularly.
This instrument should be checked at least once a year. For checks and calibrations, contact one of our accredited metrology
laboratories (information and contact details available on request), at our Chauvin Arnoux subsidiary or the branch in your country.
10.5. REPAIR
For all repairs before or after expiry of warranty, please return the device to your distributor.
Our site:
http://www.chauvin-arnoux.com
Sign in and open your account.
Then go to “Software support space”, then “Freely available software”, then “C.A 6116N / C.A 6117”.
The update of the embedded software depends on its compatibility with the hardware version of the device. This version is,
indicated in SET-UP (see § 5).
The update of the embedded software overwrites the whole configuration. As a precaution save the data to a PC before
updating the embedded software.
95
11. WARRANTY
Except as otherwise stated, our warranty is valid for twelve months starting from the date on which the equipment was sold.
Extract from our General Conditions of Sale provided on request.
96
12. TO ORDER
12.1. ACCESSORIES
15m earth kit (red/blue/green) ..................................................................................................................................... P01102017
3P earth kit (50 m) ....................................................................................................................................................... P01102021
3P earth kit (100 m) ..................................................................................................................................................... P01102022
1P earth kit (30 m, black) ............................................................................................................................................. P01102018
C177A clamp (200 A) ................................................................................................................................................... P01120336
MN77 clamp (20 A) ...................................................................................................................................................... P01120460
Continuity pole ............................................................................................................................................................ P01102084A
Lithium Ion pack charging stand ................................................................................................................................. P01102130
Dataview Software ...................................................................................................................................................... P01102095
97
3 4mm-dia. probe tips (red, blue and green) ............................................................................................................... P01101921
3 crocodile clips (red, blue and green) ........................................................................................................................ P01101922
2 elbowed-straight safety leads (red and black) 3m long ........................................................................................... P01295094
Hand strap ................................................................................................................................................................... P01298057
98
99
01 - 2014
http://www.chauvin-arnoux.com