ID-006 Test Equipment
ID-006 Test Equipment
QC Check:
BP CASPIAN SEA
workshop skills
workshop skills instruments
test equipment (id-006)
Duty: Instrumentation
Course Title: Test Equipment (ID-006)
Duration: 2 Days
OBJECTIVES
Comparator
Digital multimeter
Decade Box
Millivolt source
Smart calibrator
Milliamp source
1.1. INTRODUCTION
1.3. SENSITIVITY
1.4. TRANSMISSION
1.5. STABILITY
1.6. RELIABILITY
2. MULTIMETER
4. COMPARATOR
6. DECADE BOX
7. MV SOURCE
8. SMART CALIBRATOR
9. MA SOURCE
1.1. INTRODUCTION
150 160
150 160
Figure 2
Level Transmitter
Figure 3
If the accuracy specification is made tight, the cost will increase greatly. It
must not be forgotten that the maintenance of an instrument of high accuracy
will also increase if the accuracy is to be retained. Frequent and careful
calibration will be required. Thus greater accuracy than is really required
must not be specified. Equally, the narrowest practical range should be used,
increasing accuracy without any cost penalty.
1.4. TRANSMISSION
Electrical signals are now most common, largely because of the advent of
solid state electronics, hand in hand with development of intrinsic safety
techniques. The common transmission signal is 4 to 20 mA, but 0 – 10 volts
and 1 – 5 volts are used for short distances, that is behind control room
panels.
1.5. STABILITY
1.6. RELIABILITY
Impulse lines will block up, samples will be unrepresentative, separation will
occur causing the variations and so on.
Precautions can be taken in most cases such as purges for impulse lines,
short sample lines, heat tracing and so on. This makes the installation and
maintenance more expensive and less reliable.
We known that there are three types of error from which a pressure
instrument may suffer. They are illustrated by the graphs shown in figure 4.
Indicated Pressure
Correct Calibration
Actual Calibration
True Pressure
Figure 4
Graph (a) shows a zero error, which is a constant error over the whole span
of the instrument. In the case of the C-type Bourdon tube gauge this could be
corrected by removing the gauge pointer and replacing it in the correct
position (normally with 50% pressure applied).
Graph (c) shows an angularity or linearity error. This means that the
calibration is correct at zero and full-scale but out at mid-scale. Correcting
this type of error involves a process called squaring the links. This is done
by applying 50% pressure to the gauge and adjusting the length of the link
until the angle between the link and the quadrant is 90° as shown in figure 5.
Figure 5
This is an error that is seen when progressive rising and falling pressures are
applied to the instrument. If we take the example of the C-type gauge again,
hysteresis error is caused by wear of the teeth on the quadrant or pinion.
C B
Indicated Pressure
Figure 6
Primary standard instruments are the more accurate of the two groups, been
accurate to within plus or minus 0.05% (written ± 0.05%).
For example, if a pressure of 100 bar is indicated, the true value of pressure
is within the range of 100 bar ± 0.05%.
This means the true value of the applied pressure lies between 100 + 0.05%
and 100 – 0.05%.
Figure 7
Figure 8
Before fitting the gauge the DWT must be primed to ensure there is no air in
the system. With the isolation valve closed, the screw press is turned until the
oil rises up the gauge mounting connection. Fit gauge and apply a little
pressure by turning the screw press. Open the oil reservoir valve and check
for the release of any air bubbles.
When the isolation value is closed and the screw press wound in, the oil
pressure increases.
To accurately measure the input oil pressure applied to the gauge under test,
a finely machined piston and cylinder unit is fitted to the system.
The applied oil pressure is compared to the pressure exerted by the piston
inside the cylinder. When these two pressures are equal a point of balance is
reached. When this occurs the piston lifts in the cylinder and is supported by
the oil pressure. This is illustrated in figure 8 above, where the exposure of a
coloured band indicates that the balance point has been achieved and that
the pressure of the oil equals the pressure exerted by the piston inside the
cylinder.
When the calibration is complete, wind the screw press right out and open the
isolation value before removing the weights from the DWT table.
Well type manometers and U-tube manometers are often used for calibration
purpose.
Figure 9
In this case the gauge is calibrated using air pressure from the regulator. The
pressure to the connecting pipework is increased in increments and the
manometer reading (the true value) and the pressure gauge reading (the
indicated value) are noted at each stage.
Figure 10
Also standard test gauges are often of large format, having a scale of
between 200mm and 250mm as opposed to 150mm for a plant gauge.
The readings on the gauge been tested are compared with those on the
standard test gauge. They are recorded and any error identified.
T Transfer
TransferStandard
Standard Dead Weight Tester
ransfer Standard e.g.
Figure 12
Each element in the ladder should be significantly more accurate than the
element below it.
If the test equipment is out of calibration, it must not used until a valid
calibration has been carried out.
The third party calibrations use a transfer standard that is traceable against a
national ultimate standard. When the test equipment has been calibrated, a
calibration certificate is issued to that piece of equipment.
% PSI / Bar ma
0
25
50
75
100
75
50
25
Figure 13
Figure 14 – Multimeter
Voltage
Current
Resistance
Frequency sometimes
If you are checking a circuit, you must know whether the circuit is:
AC (alternating current)
DC (direct current)
You also need to know how much current and voltage you’re potentially
measuring.
Resistance (ohms W)
V DC (volts DC)
V AC (volts AC)
A DC (amps DC)
Resistance (W)
Display
Voltage DC Voltage AC
Current
Resistance
Meter Lead Terminals
Figure 15
The next step in the process is to select an appropriate range. There are a
variety of ranges available for volts, amps and resistance. For example, you
would not use a two-volt setting to measure 600 volts.
The previous slide shows the multimeter measuring a voltage of 1.999V. The
2v range is selected. If a voltage of 10v was to be measured, the next range
(20v) would have to be selected. If the voltage exceeded 20v, then the 200v
scale would be required etc. If you’re unsure of the approximate voltage
range, a good technique is to start with the maximum range and work your
way down. If you are not sure if you were measuring two volts or 200 volts,
you should initially measure the voltage with the highest possible range (in
this case 1000 volts)
Voltage
Resistance
This means the meter probes are placed directly across the circuit to be
measured.
Note
Multimeters have the ability to measure current directly. This portion of the
meter is usually fused to protect against excessive current loads. If high
current measurements are required, such as starting or charging circuit tests,
current clamps are available.
Figure 18
The only quantity measured with a multimeter in series with the circuit. Circuit
must be modified in the following way:
Figure 19
Sometimes, you just need to find out if power is able to get from point A to
point B. This is called a continuity test. A continuity test involves using the
Ohms section of your Multimeter. Make sure power to wire is off. Place one
probe at one end of the wire that you’re testing and place the other probe on
the other end. This involves the multimeter sending a tiny amount of current
into the circuit to be measured. If the resistance level is very low, then the
current can get through the wire. The circuit is then ‘closed circuit’. If the
resistance level is infinite, then the circuit is ‘open’. The circuit being
measured has a break in it from point A to point B. Many meters produce an
audible signal when continuity is present. If the circuit is continuous, you will
hear the tone.
You can also use the Ohm section of your multimeter to test resistance. First,
make sure that there’s no power going through the circuit. Place the red
probe on one side of the object to be measured for resistance. Place the
When measuring AC, it doesn’t matter which probe goes to which wire.
When measuring DC, you should always observe polarity by attaching the red
probe to the positive and the black probe to the negative. Measuring current
and voltage are the only operations that should be performed with the power
on. All other measurements should be taken with the power off.
Pressure Outlet
Keypad and Display
Hand Pump
Figure 20
Instrument Connection
The pressure input or pressure output from the calibrator is made through the
instrument connection.
To measure pressure simply connect the signal line to the calibrator through
the instrument connector, select the correct range and monitor the reading on
the digital display.
Parameter selection, zero adjustment and all other operating commands via
the key-pad and the powerful microprocessor provide a wide range of display
capabilities.
For example the Druck DPI 601 Series model can measure pressures to an
accuracy of + 0.05% and can supply 10v to energise transducers or
transmitters under test and measure the output response (except IS version).
A 2 volt analogue output is also provided for use with chart recorders or other
data recording systems. The microprocessor controlled circuit provides many
advanced features such as peak reading, signal filtering, square root and 0 to
100% readout.
Safety
Oil reservoir
Gauge ports
Hand pump
Gauges
Pressure switches
Pressure Transmitters
6. Open oil reservoir slowly and observe pressure release for any air
bubbles.
8. Close oil reservoir isolation valve and repeat steps 6 to 8 until no air is
released.
Safety
Ensure that the correct fittings and sealing washers are used.
the gauges
Weights
Instrument under test
Piston
Oil Reservoir
Adjustable feet
Piston Handle
Figure 24
Piston
Hand Pump
Oil
Figure 25
As previously stated, the mass off the weights is known and equal to a known
pressure:
No Pressure applied:
Figure 26
10 KG Weight added:
Figure 27
Figure 28
The instrument under test has now been checked against a known pressure
reference source. If the instrument indication does not match the pressure
applied to the dead weight tester, zero or span adjustments must be made to
the instrument under test.
Safety
Ensure that the DWT table is level by checking the spirit level on the DWT
and adjusting the feet on the DWT to bring the table into the horizontal plane.
Ensure there is sufficient oil in the reservoir before carrying out test
Ensure connections, fittings and seals are used before carrying out test.
Figure 29
Coil 1 = increments of 1 W
Coil 2 = increments of 10 W
Up to increments of 100,000 W
3
7
0
8
9
5
Figure 30
Uses
Output terminals
Thumbwheel switches
Range switches
3 Ranges up to 1V
0-999.9mV in 3 ranges
0.05% Accuracy
20 mA Output Current
Portable
The 404S is an accurate low cost millivolt source suitable for voltage injection
applications.
Three output ranges are provided to give adjustable output values from 1µV
to 1V with a basic 0.05% accuracy.
0-999.9mV in 3 ranges
For signal injection, the operator needs only to switch on, check the battery
condition, select the range and set the required voltage using the thumbwheel
switches.
The 404S uses a precision reference diode and low temperature co-efficient
resistors to give a highly stable output.
The 404S has up to 20 mA drive current and is short circuit and overload
protected. A normal /off / reverse output polarity switch is provided.
Uses
A/D converter
Figure 32
SMART Transmitter
HART Communicator
Figure 33
Digital display
Figure 34
Figure 36
8mA = 25%
12mA = 50%
Trip amplifiers
Indicators
Loop power
The calibrator varies the current within the instrument loop. Current of up to
24mA can be generated. The current can be generated in ‘steps’. This is
useful for checking at set intervals:
A valve can be checked for operation (or ‘stroked’) using the mA calibrator.
The position of the valve can be controlled by varying the mA signal to the
control valve. Valves are commonly checked at fixed intervals to ensure they
are fully opening and closing. See figure 37
8mA = 25%
12mA = 50%
Safety
Temperature Display
Figure 38
Temperature sleeve
Temperature Bath
Figure 39
The instrument should be a tight fit in the sleeve to ensure good heat transfer
Figure 40
Liquid temperature baths are another temperature standard also used. They
are similar in principle to the dry block temperature bath, except liquid is
heated as the reference. The instrument to be tested is inserted into the
liquid at known temperature.
Liquid Reservoir
Temperature Display and Controller
Figure 41
Liquid bath’s can achieve much lower temperatures than a dry block bath. A
typical liquid bath can control at temperatures as low as –80 DegC. Care
must be taken not to come into contact with liquid at low temperatures.
Tongs