0% found this document useful (0 votes)
15 views14 pages

FS300 Datsheet en

FS300 Datasheet
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
15 views14 pages

FS300 Datsheet en

FS300 Datasheet
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 14

Spectrum Analyzer

¸FS300/FS315
9 kHz to 3 GHz

y
t famil
n e w produc
Th e
from arz
d e & Schw
Roh

2006
i t i o n March
E d
Fourth
Professional test equipment
for laboratory, service and production

The ¸FS 300 is a highly accurate ­spectrum ana-


High-quality measurement characteristics
lyzer with a frequency range of 9 kHz to 3 GHz.
Owing to its modern, digital ­frequency processing Resolution bandwidths from 200 Hz to 20 MHz
technique, it offers high ­ measurement quality at (¸FS315)

a favorable price. The ¸FS315 is additionally Frequency counter with 1 Hz resolution


equipped with a built-in tracking generator from
9 kHz to 3 GHz for scalar network analysis; the Maximum input level 33 dBm
tracking generator is also suitable for ­ generating
fixed-frequency signals. Plus, the ¸FS315 Ergonomic user interface

includes various detectors for evaluating mea-


Remote control via USB interface
surement results and allows electric field strength
measurements taking into account the antenna AM/FM audio frequency demodulator output
factors. (¸FS315)

Condensed data
¸FS300 ¸FS315
Frequency range 9 kHz to 3 GHz
Resolution bandwidths (–3 dB) 200 Hz to 1 MHz 200 Hz to 20 MHz
Video bandwidths 10 Hz to 1 MHz 10 Hz to 20 MHz
Displayed average noise level < –110 dBm, typ. –115 dBm (300 Hz)
Intermodulation-free range < –70 dBc at –36 dBm input level
SSB phase noise, 10 kHz offset < –90 dBc (1 Hz)
Level uncertainty < 1.5 dB, typ. 0.7 dB
Detector peak max/min peak, sample, average, RMS
Measurement functions TOI, TDMA power, frequency counter, noise marker TOI, TDMA power, frequency counter, noise marker, oc-
cupied bandwidth (OBW), return loss, transmission,
channel power

Tracking generator – 9 kHz to 3 GHz


Audio frequency demodulator – AM /FM
Measurement with antenna factors – yes

 Spectrum Analyzer ¸FS300/FS315


The softkey area displays
The menu area displays
different instrument
menus for setting ­
functions as a function of
parameters and functions
the selected menu

Rotary knob
for gradually varying
instrument parameters

Numeric keypad
for entering
numeric parameters

Arrows
for navigating in the menus

Connector
Tracking generator output
(¸FS315)

Buttons
for confirming entered values

System key
for calling system and
service functions

Ergonomic user interface


Operation is menu-guided enabling even untrained users to quickly obtain correct results. Clear
structures simplify navigation within the menus.
The bright TFT color display allows traces to be read even at odd angles or when the incidence of
light is unfavorable.

Spectrum Analyzer ¸FS300/FS315 


Application ranges
The ¸FS300/FS315 is a versatile spectrum
analyzer for comprehensive measurements in
­laboratory, service and production.

Measurement of RF spectrum (level and frequency)

Measurement of radiated interference (EMC)

Time domain measurements

Radiomonitoring remote-controlled via USB


PC software
Scalar network analysis (only ¸FS315)

A powerful software package for remote control


from a PC is supplied with the ¸FS300/FS315.
The software enhances the ¸FS300/FS315
functions and supports the generation of test
reports on the PC.

Characteristics

■ Windows 2000/XP-compatible ■ Practically unlimited memory capacity for storing traces


■ PC linked to ¸FS300/FS315 via USB ­interface and measurement information (comparison of current and
■ Fast and simple transfer of measurements between previous measurements)
¸FS300/FS315 and PC ■ Export of trace values (900 points) in txt format for import
■ Permanent sweep and transmission of ongoing sweeps into MS Excel
to the PC with evaluation capabilities (marker, zoom, etc) ■ Export of displayed data (screenshots) in JPEG format
■ Extended range of functions (limit lines, log file) ■ Output of results to standard printer

 Spectrum Analyzer ¸FS300/FS315


High-quality measurement characteristics
The RF characteristics of the ¸FS300/FS315 are setting new standards in the
lower price class. Since the displayed average noise level is typically –115 dBm
(300 Hz), even weak signals can be reliably detected. Owing to the wide dynamic
range, this is also possible when strong carrier signals are present.

The points in the traces are displayed with an accuracy unrivalled in this price
class. This is an essential prerequisite for any measurement task.

Resolution bandwidths from 200 Hz to 1 MHz


With 16 digitally implemented resolution bandwidths from 200 Hz to 1 MHz, the
¸FS300 can be optimally adapted to the measurement task at hand. The
¸FS315 additionally covers the range up to 20 MHz. Wide resolution band-
widths for overall measurements ensure short sweep times, whereas narrow
bandwidths are ideal for high frequency resolution and a low noise level. The
¸FS300 and ¸FS315 fulfill every requirement in between.

Scalar network analysis


The ¸FS315 with built-in tracking generator is the perfect solution for cost-
efficient testing of the transmission behavior of filters, cables, amplifiers and
so forth. Equipped with an additional VSWR bridge, reflection measurements
can also be performed. Entering any frequency offset between 0 Hz and 3 GHz
allows measurements on frequency-converting DUTs. With simple applications,
the tracking generator can be used as a signal generator with a permanently set
frequency.

Locating EMC weak spots


The ¸HZ-15 near-field probes are diagnostic tools used for locating EMC weak
spots on printed boards, integrated circuits, cables, shieldings and other trouble
spots. The Near-Field Probe Set ¸HZ-15 is adequate for emission measure-
ments from 30 MHz to 3 GHz. The Preamplifier ¸HZ-16 up to 3 GHz, with
approximately 20 dB gain and a noise figure of 4.5 dB, increases sensitivity for
measurements. In combination with the ¸FS300/FS315, the preamplifier and
near-field probe set are a cost-effective means of analyzing and locating sources of
interference during development.

Spectrum Analyzer ¸FS300/FS315 5


 Spectrum Analyzer ¸FS300/FS315
Ready for the future –
the new instrument family

The ¸FS300 and the ¸FS315 are part of a new family of analyzers and generators for
development, service and production applications. The platform on which this family is based
– with its compact design, powerful processor system, fast internal bus and ergonomic user interface –
provides optimum conditions for professional, favorably priced instruments.

Compact housing with adjustable Remote control via USB interface


­handle

The ¸FS300 and the ¸FS315 are notable for their The ¸FS300 and the ¸FS315 can easily be operat-
compact and robust design. They require only a minimum of ed from a PC via the USB remote-control interface. Simply
space on your desktop or in the rack. Even two instruments connect the PC via hot plug & play, start the supplied soft-
of this family can easily be accommodated next to each oth- ware and that’s it. The supplied drivers for Windows 2000/
er in a 19-inch rack. The handle, which can be turned and XP make system software integration mere child’s play.
shifted, can be used to carry the instrument during transport
and as a fold-out support to ensure an ideal angle. The han-
dle can be conveniently tilted to the side if it interferes with
smooth operation.

Spectrum Analyzer ¸FS300/FS315 


Specifications

Our products are continuously enhanced and upgraded. For the latest on the ¸FS300 and the ¸FS315,
check out the Internet at www.fs300.rohde-schwarz.com

Frequenz

¸FS300 ¸FS315
Frequency range 9 kHz to 3 GHz
Frequency resolution 0.1 Hz
Reference frequency 10 MHz, nominal
Aging 2 × 10–6/year
Temperature drift 5 °C to 30 °C 1 × 10–6
External reference 10 MHz
Frequency counter
Resolution 1 Hz, 10 Hz
Count accuracy S/N >25 dB ±(marker frequency × reference error + ½ (last digit))
Frequency span 1 kHz to 3 GHz, 0 Hz
Span accuracy <1 %
Spectral purity
SSB phase noise 9 kHz ≤ f ≤ 3 GHz
10 kHz carrier offset <–90 dBc (1 Hz), typ. 95 dBc (1 Hz)
100 kHz carrier offset typ. –100 dBc (1 Hz)
1 MHz carrier offset typ. –110 dBc (1 Hz)
Residual FM 1 kHz resolution bandwidth, <100 Hz
1 kHz video bandwidth
9 kHz ≤ f ≤ 3 GHz, weighting
in line with CCITT
Sweep time
Span >1 kHz 100 ms to 1000 s 30 ms to 1000 s
(steps depending on RBW and
span)
Max. deviation 5% 1%
Span = 0 Hz 100 µs to 20 s 5 µs to 10 s
Resolution 150 ns 20 ns

 Spectrum Analyzer ¸FS300/FS315


Frequency

¸FS300 ¸FS315
Bandwidths
Resolution bandwidths (–3 dB) in 1/2/3/5 sequence 200 Hz to 1 MHz 200 Hz to 20 MHz
Bandwidth accuracy RBW ≤ 1 MHz 5% <1 %
2 MHz ≤ RBW ≤ 10 MHz – <5 %
RBW 10 MHz, 20 MHz – <10 %
Shape factor 60 db/3 dB RBW ≤ 1 MHz <4.6:1
Video bandwidths in 1/2/3/5 sequence 10 Hz to 1 MHz 10 Hz to 20 MHz

Amplitude
¸FS300 ¸FS315
Display range displayed average noise level to + 33 dBm
Display scaling 80 dB, 40 dB, 16 dB, 8 dB, linear
Display units
Logarithmic dBm, dBµV, dBmV
Linear V, W
Maximum input level
DC voltage 30 V
step from –30 V to +30 V 1200 V/µs
CW RF power RF attenuation <20 dB +13 dBm
RF attenuation ≥20 dB
50 MHz to 3 GHz +33 dBm
20 MHz to 50 MHz +26 dBm
9 kHz to 20 MHz +20 dBm
1 dB compression point of 1st mixer
f >100 kHz, –10 dBm nominal
RF attenuation 0 dB
Linearity
Harmonics input level –40 dBm, <–60 dBc
RF attenuation 0 dB
Intermodulation-free dynam- two-tone signal with level <–70 dBc
ic range for third-order inter- 2 × –30 dBm,
modulation RF attenuation 6 dB
Displayed average noise level
9 kHz to 3 GHz, <–110 dBm, typ. –115 dBm
RF attenuation 0 dB,
300 Hz RBW, 10 Hz video
bandwidth

Spectrum Analyzer ¸FS300/FS315 


Amplitude

¸FS300 ¸FS315
Spurious
Inherent spurious RF attenuation 0 dB, <–85 dBm
input terminated
Other spurious 10 MHz to 3 GHz, <–60 dBc
level at 1st mixer –35 dBm
Level settings
Setting range of reference –110 dBm to +36 dBm
level
Resolution 0.1 dB
RF attenuation range manual selection or automat- 0 dB to 70 dB
ically coupled to ­reference
­level
Resolution 2 dB
Traces 1 active trace and 1 stored trace
Trace detectors max peak max peak, min peak, sample,
­average, RMS,
Trace functions clear/write, max hold, min hold, average
Max. uncertainty of level measurement
Frequency response 9 kHz to 3 GHz, – <1.0 dB
RF attenuation 0 dB to 70 dB
Reference level uncertainty – <0.3 dB
Display nonlinearity 0 dB to –60 dB – <0.3 dB
–60 dB to –70 dB – <1 dB
Bandwidth switching <0.2 dB <0.3 dB
­uncertainty
Total measurement 0 dB to –60 dB below ref. <1.5 dB 1.5 dB, typ. 0.7 dB
­uncertainty ­level, RBW ≤5 MHz
Markers
Number of markers and ­delta 1 marker and 1 delta marker
markers
Marker functions peak, next peak left, next peak right,
center frequency = marker frequency,
reference level = marker level
Marker displays normal (level), noise marker, frequency counter, n dB down (bandwidth)
Audio demodulation zero span only, RBW ≤1 MHz – AM and FM

Spectrum Analyzer ¸FS300/FS315 10


Trigger

¸FS300 ¸FS315
Span ≥ 1 kHz
Trigger source free run, external
Trigger offset sweep time > 100 ms 0 ≤ trigger offset ≤ 100 ms, resolution 25 ns
Span = 0 Hz
Trigger source free run, external, video
Trigger offset negative offset limited by –100 ms ≤ trigger offset ≤ 100 ms –100 ms ≤ trigger offset ≤ 10 s
sweep time

Tracking generator

only ¸FS315
Frequency
Frequency range 9 kHz to 3 GHz
Frequency offset
Setting range 0 Hz to 3 GHz
Resolution 0.1 Hz
Spectral purity
SSB phase noise 10 kHz carrier offset <–90 dBc (1 Hz)
9 kHz ≤ f ≤ 3 GHz
Level
Level setting range 0 dBm to –50 dBm
Resolution 0.1 dB
Max. deviation of output level 9 kHz to 3 GHz, <1 dB
20 °C to 30 °C
50 kHz ≤ RBW ≤ 1 MHz
Spurious
Harmonics output level –10 dBm <–20 dBc
Nonharmonics output level 0 dBm <–30 dBm

Interfaces

¸FS300 ¸FS315
USB host device-specific command set, A plug, protocol version 1.1
remote control via supplied
Windows driver (Windows
XP/2000)
USB device B plug, protocol version 1.1
Connector for external moni- 15-pin D-Sub female
tor (VGA)
Keyboard connector PS/2 female

11 Spectrum Analyzer ¸FS300/FS315


Inputs

¸FS300 ¸FS315
RF input
Connector N female (front panel)
Impedance 50 W
VSWR RF attenuation 20 dB <1.5
External trigger input
Connector BNC female (rear panel)
Trigger voltage TTL
Reference frequency input
Connector BNC female (rear panel)
Reference frequency 10 MHz ± 50 Hz
Impedance 50 W
Input level 0 dBm to 20 dBm

Outputs

¸FS300 ¸FS315
RF output (tracking generator)
Connector – N female (front panel)
Impedance – 50 W
VSWR – <1.6
Reference frequency output
Connector BNC female (rear panel)
Reference frequency 10 MHz
Impedance 50 W
Output level 7 dBm nominal
AF output
Connector – 3.5 mm mini jack for headphones
(rear panel)
Impedance – 15 W nominal

12 Spectrum Analyzer ¸FS300/FS315


General data

¸FS300 ¸FS315
Display
Type 5.4“ active TFT color display
Resolution 320 × 240 pixel
Max. refresh rate 10 pictures/s, nominal
Power supply
Input voltage range autoranging 100 V to 240 V (AC), 50 Hz to 60 Hz
Power consumption <45 W <60 W
Ambient conditions
Operating temperature range meets EN 60068-2-1/2 +5°C to +45°C
Storage temperature range –20°C to +70°C
Relative humidity meets EN 60068-2-78 95 % at +40 °C
Mechanical resistance
Sinusoidal vibration meets EN 60068-2-6, 5 Hz to 150 Hz, max. 2 g at 55 Hz,
EN 61010-1 and MIL-T-28800D 55 Hz to 150 Hz: 0.5 g constant
class 5
Random vibration meets EN 60068-2-64 10 Hz to 500 Hz: 1.9 g
Shock meets EN 60068-2-27 and shock-spectrum
MIL-STD-810
Electromagnetic meets EN 55011 class B and EN 61326
­compatibility (EMC Directive of EU (89/336/EEC))
EMI field strength 10 V/m
Safety EN 61010-1/IEC61010-1, UL3111-1, CSA C22.2 No. 1010.1
Dimensions (W × H × D) 219 mm × 147 mm × 350 mm
Weight 8.5 kg 9 kg

Spectrum Analyzer ¸FS300/FS315 13


Spectrum Analyzer ¸FS300/FS315
Designation Type Order No.
Spectrum Analyzer ¸FS300 1147.0991.03
Spectrum Analyzer with Tracking Generator ¸FS315 1147.1000.03
Rack Adapter ¸ZZA-300 1147.1281.00
Transit Case ¸ZZK-300 1147.2542.02
Accessories supplied for the ¸FS300/FS315:
User manual (German/English), CD-ROM with PC software and documentation,
USB cable for PC connection, power cable

Recommended extras for the ¸FS300/FS315:


Near-Field Probe Set ¸HZ-15 1147.2736.02
Preamplifier for ¸HZ-15 ¸HZ-16 1147.2720.02
SWR Bridge 5 MHz to 3 GHz ¸ZRB2 0373.9017.52
SWR Bridge 5 MHz to 2.5 GHz ¸ZRB2 0373.9017.53
Spare Short/Open Calibration Standard for VSWR Calibration ¸FSH-Z30 1145.5773.02

¸is a registered trademark of Rohde & Schwarz GmbH & Co. KG · Trade names are trademarks of the owners · Printed in Germany (Pe as)
PD 0758.0297.32 · ¸FS 300/FS315 · Version 04.00 · March 2006 · Data without tolerance limits is not binding · Subject to change
Certified Quality System Certified Environmental System

ISO 9001
DQS REG. NO 1954 QM
ISO 14001
DQS REG. NO 1954 UM

More information at
www.rohde-schwarz.com
(search term: Smart Instruments, FS300, FS315)

www.rohde-schwarz.com
Europe: +49 1805 12 4242, [email protected]
USA and Canada: +1-888-837-8772, [email protected]
Asia: +65 65 130 488, [email protected]

You might also like