Manual Agilent
Manual Agilent
Operator’s Manual
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2011 Warranty (Technical Data) and 12.212 (Computer
No part of this manual may be reproduced Software) and, for the Department of
The material contained in this doc- Defense, DFARS 252.227-7015 (Techni-
in any form or by any means (including
ument is provided “as is,” and is cal Data - Commercial Items) and DFARS
electronic storage and retrieval or transla-
tion into a foreign language) without prior
subject to being changed, without 227.7202-3 (Rights in Commercial Com-
agreement and written consent from Agi- notice, in future editions. Further, puter Software or Computer Software
lent Technologies, Inc. as governed by to the maximum extent permitted Documentation).
United States and international copyright by applicable law, Agilent disclaims
laws. all warranties, either express or Safety Notices
implied, with regard to this manual
Manual Part Number and any information contained
herein, including but not limited to CAUTION
70-9068 the implied warranties of mer-
chantability and fitness for a par- A CAUTION notice denotes a
Edition ticular purpose. Agilent shall not be hazard. It calls attention to an
Rev A, July 2011 liable for errors or for incidental or operating procedure, practice, or
consequential damages in connec-
Printed in USA
tion with the furnishing, use, or the like that, if not correctly per-
Agilent Technologies, Inc. performance of this document or of formed or adhered to, could result
3501 Stevens Creek Blvd. any information contained herein. in damage to the product or loss
Santa Clara, CA 95052 USA Should Agilent and the user have a of important data. Do not proceed
separate written agreement with beyond a CAUTION notice until
warranty terms covering the mate-
rial in this document that conflict the indicated conditions are fully
with these terms, the warranty understood and met.
terms in the separate agreement
shall control.
1 Safety 7
Electrical Hazards 8
Warning 9
Caution 9
Note 9
Information Symbols 10
General 11
WEEE Directive 12
2 Introduction 13
Available Configurations 15
Conventions Used in this Manual 16
USP Suitability Test 17
USP Physical Parameters 17
Index 43
This page was intentionally left blank, except for this message.
1
Safety
Electrical Hazards 8
WEEE Directive 12
The Friability Tester has been carefully designed so that when used
properly you have an accurate, fast, flexible, and safe instrument.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
The Friability Tester is operated in conjunction with equipment that uses
aqueous liquids. Unskilled, improper, or careless use of this instrument
can create shock hazards, fire hazards, or other hazards which can cause
death, serious injury to personnel, or severe damage to equipment and
property.
Information on safety practices is provided with your instrument and
operation manuals. Before using your instrument or accessories, you
must thoroughly read these safety practices.
Observe all relevant safety practices at all times.
Agilent Technologies 7
1 Safety
Electrical Hazards
The Friability Tester contains electrical circuits, devices, and
components operating at dangerous voltages. Contact with these
circuits, devices, and components can cause death, serious injury, or
painful electric shock.
Panels or covers that are retained by fasteners which require the use of a
tool for removal may be opened only by Agilent-trained,
Agilent-qualified, or Agilent-authorized service engineers. Consult the
manuals or product labels supplied with the Friability Tester to
determine which parts are operator-accessible.
Application of the wrong supply voltage, connection of the instrument
to an incorrectly wired supply outlet, or lack of proper electrical
grounding can create a fire hazard or a potentially serious shock hazard
and could seriously damage the instrument and any attached ancillary
equipment.
Always use a three-wire outlet with ground connection which is
adequately rated for the load. The installation must comply with local,
state, and federal safety regulations.
Do not connect the instrument to the main power supply until you have
made sure that the operating voltage is correctly set for the main power
supply in the specific outlet in your laboratory to which the equipment
will be connected.
Warning
WA RNING
Caution
CAUTION
Note
NOT E
A ‘Note’ appears in the manual to give advice or information.
Information Symbols
I Switches main power on
0 Switches main power off
General
CE Compliant Products
The Friability Tester has been designed to comply with the requirements
of the Electro-magnetic Compatibility (EMC) Directive and the Low
Voltage Directive (LVD) of the EU.
Agilent, Inc. has confirmed that each product complies with the relevant
directives by testing a prototype against the prescribed European Norm
(EN) standards.
Proof that a product complies with the directives is indicated by:
• the CE marking appearing on the rear of the product.
• the documentation package that accompanies the product containing
a copy of the declaration of conformity. This declaration is the legal
declaration by Agilent, Inc. that the product complies with the
directives and also shows the EN standards to which the product was
tested to demonstrate compliance. The declaration of conformity is
signed by the representative of the manufacturing plant.
WEEE Directive
All Agilent products that are subject to the WEEE directive shipped
after August 13, 2005 are compliant with the WEEE marking
requirements. Such products are marked with the “crossed out wheelie
bin” WEEE symbol shown on page 10 in accordance with European
Standard EN 50419.
This symbol on the product or on its packaging indicates that this
product must not be disposed of as unsorted municipal waste. The
separate collection and recycling of your waste equipment at the time of
disposal will help to conserve natural resources and ensure that it is
recycled in a manner that protects human health and the environment.
For more information on collection, reuse, and recycling systems,
please contact your local/regional waste administration, your local
distributor, or Agilent, Inc.
2
Introduction
Available Configurations 15
Conventions Used in this Manual 16
USP Suitability Test 17
USP Physical Parameters 17
Agilent Technologies 13
2 Introduction
Panels or covers that are retained by fasteners which require the use
CAUTION
of a tool for removal may be opened only by Agilent-trained,
Agilent-qualified, or Agilent-authorized service engineers.
Available Configurations
The Friability Tester is available in a single-drum or dual-drum
configuration. The dual-drum configuration turns synchronously on
opposite ends of the same shaft. This option allows two tests to run
simultaneously.
Three types of drums are available:.
Figure 3 Agilent USP drum with two arched bridges which saves
time by allowing for two tests in one drum.
This page was intentionally left blank, except for this message.
3
Setting Up the Friability Tester
Initial Setup 20
Setting the Date and Time 22
Hidden Key Functions 23
Agilent Technologies 19
3 Setting Up the Friability Tester
Initial Setup
Complete the following sections to initially set up the Friability Tester
and all other system components.
Unpacking Procedure
Complete these steps to safely unpack and set up your equipment:
1 Open each carton and check the contents for damage which may
have occurred during shipping. Shipping damage rarely occurs, but if
it does contact both the carrier who delivered the instruments and the
Dissolution Systems Service Department. Though claims for damage
should be filed with the carrier, we can help you file a claim.
2 Carefully remove the Friability Tester base unit from the shipping
carton. Place it on a clean, dry, level area of your benchtop within
four feet (122 cm) of a grounded electrical outlet.
CLEAR > 6 Use this key sequence to enable the PCB display.
CLEAR > 8 Use this key sequence to indicate the Friability Tester.
CLEAR > 9 Use this key sequence to enable the PRINT key.
This page was intentionally left blank, except for this message.
4
Operating the Friability Tester
Operating the Friability Tester without a Printer 26
Operating the Friability Tester with a Printer 28
Agilent Technologies 25
4 Operating the Friability Tester
Running a Test
Complete the following steps to run a test using a Friability Tester not
equipped with a printer:
1 Ensure the PRINT key is disabled. See “Hidden Key Functions” on
page 23 for instructions on how to disable the PRINT key.
2 Select the mode of operation, either timer or counter, by pressing
TIME COUNT. This key toggles the unit between the two operating
modes. The Timer LED illuminates to indicate the timer option is
selected. The Counter LED illuminates to indicate the counter option
is selected.
3 Enter either the number of rotations (up to 999,999) or the time
duration (in hh:mm:ss format) as applicable and press ENTER.
Once entered, the test duration is saved by the unit and does not need
to be reset unless you want to change it. To change the duration, enter
the new value and press ENTER.
4 Locate the drums supplied with your unit. Hold the drums firmly so
the covers do not fall off. Place the drums on the benchtop and
carefully remove the covers and set them aside.
The right- and left-side drums are not interchangeable. The vanes curve
NOT E in opposite directions. Be sure to identify them properly before
mounting. Drums supplied with dual-drum units have right- and
left-side configurations. Check the label on the cover of each to identify
them. Left-side drums mount on the left side of the unit as viewed from
the front. Right-side drums mount on the right side as viewed from the
front.
Stopping a Test
When running a test on a Friability Tester without a printer, you can stop
the test by pressing START STOP. The drums stop immediately and the
display returns to the original test length setting.
Running a Test
If your unit is equipped with the optional printer, you can print a report
at the end of each test to document the date, starting time, mode
selected, and the set duration. If the Printer On LED is illuminated, the
printer is enabled. Press PRINT to enable or disable the printer.
The right- and left-side drums are not interchangeable. The vanes curve
NOT E in opposite directions. Be sure to identify them properly before
mounting. Drums supplied with dual-drum units have right- and
left-side configurations. Check the label on the cover of each to identify
them. Left-side drums mount on the left side of the unit as viewed from
the front. Right-side drums mount on the right side as viewed from the
front.
9 For each drum, hold the cover firmly in place and slide the drum
body onto the shaft. There are two pins at the end of the shaft which
hold the drum while the shaft turns. Match the pins with the slots in
the hub.
10 For each drum, place the locking nut on the end of the shaft and
tighten. Do not overtighten or damage to the drum may result.
11 Press START STOP. You are prompted to enter the initial weight (in
grams) of the product in the drum. Up to four weights can be entered
depending on the number of chambers or drums used (see
“Configuring for the Number of Chambers or Drums Used” on
page 28). The following is a summary of what displays after pressing
START STOP but before the drums begin to turn:
• The number 1 displays on the display screen alternating with the
last initial weight entered for the product in chamber or drum 1.
Press ENTER to accept the displayed value or enter the new
initial weight and press ENTER. If configured for one chamber
or drum, the drum begins to turn.
• If configured for two chambers or drums, the number 2 flashes on
the display screen alternating with the last initial weight entered
for the product in chamber or drum 2. Press ENTER to accept the
displayed value or enter the new initial weight and press ENTER.
If configured for two chambers or drums, the drums begin to turn.
• If configured for three chambers, the number 3 flashes on the
display screen alternating with the last initial weight entered for
the product in chamber 3. Press ENTER to accept the displayed
value or enter the new initial weight and press ENTER. If
configured for three chambers, the drums begin to turn.
• If configured for four chambers, the number 4 flashes on the
display screen alternating with the last initial weight entered for
the product in chamber 4. Press ENTER to accept the displayed
value or enter the new initial weight and press ENTER. If
configured for four chambers, the drums begin to turn.
When the drums have stopped turning at the conclusion of the test, the
NOT E screen prompts you to enter ending weights for the products in each
chamber or drum as it did for starting weights.
13 Remove the tablets and brush away any loose dust. Per the USP, if
obviously cracked, cleaved, or broken tablets are present in the tablet
sample after tumbling, the sample fails the test.
14 Re-weigh the tablets and enter the requested values.
15 After the ending weight for the last chamber or drum has been
entered, the printer prints a test report including the calculated
percentage of weight loss for all samples.
Stopping a Test
1 When running a test on a Friability Tester equipped with a printer,
you can stop the test by pressing START STOP. The drums stop
immediately and the display prompts you to enter the ending weights
for the products as described above.
2 Enter the ending weights and press ENTER for each chamber as
applicable. The printer prints a test report and calculates the
percentage of weight loss for the products even though the test was
stopped prematurely.
This page was intentionally left blank, except for this message.
5
Troubleshooting and Maintenance
Maintenance 34
Fuse Replacement 36
Troubleshooting 37
Agilent Technologies 33
5 Troubleshooting and Maintenance
Maintenance
The Friability Tester contains electrical circuits, devices, and
WA RNING
components operating at dangerous voltages. Contact with these
circuits, devices, and components can cause death, serious
injury, or painful electric shock.
Periodic Maintenance
Periodic maintenance needs may vary depending on frequency of
instrument usage.
• Immediately wipe all spills or errant materials from the exterior of
the instrument or any exposed part using a clean cloth and plain
water.
• Do not use cleaners containing ammonia or organic solvents, such
as alcohol, on the acrylic drums. These compounds can attack the
plastic, causing cracks and fractures which will not be covered by
the warranty.
• Store all Friability Tester drums properly.
Monthly Maintenance
• Check the drums for cracks or chips.
• Wipe the shaft clean.
• Ensure the pins which hold the drums while the shaft turns are not
bent.
• Clean the outside of the tester with a damp cloth.
Annual Maintenance
The Friability Tester contains electrical circuits, devices, and
WA RNING
components operating at dangerous voltages. Contact with these
circuits, devices, and components can cause death, serious
injury, or painful electric shock.
Panels or covers that are retained by fasteners which require the
use of a tool for removal may be opened only by Agilent-trained,
Agilent-qualified, or Agilent-authorized service engineers.
Acrylic Care
Do not use cleaning compounds containing ammonia or abrasive
CAUTION
cleaners on your drums.
Fuse Replacement
The Friability Tester contains electrical circuits, devices, and
WA RNING
components operating at dangerous voltages. Contact with these
circuits, devices, and components can cause death, serious
injury, or painful electric shock.
Panels or covers that are retained by fasteners which require the
use of a tool for removal may be opened only by Agilent-trained,
Agilent-qualified, or Agilent-authorized service engineers.
Troubleshooting
The Dissolution Systems Service Department can assist you if you
experience problems or have questions concerning your Friability
Tester. Many problems can be traced to simple sources and are easily
solved.
Following is a troubleshooting guide which may help you. The
Dissolution Systems Service Department can be reached at
800.229.1108 (within the USA) or 919.677.1108 (outside the USA).
Optionally, you can send a fax to 919.677.1138. You can also e-mail the
Dissolution Systems Service Department at
[email protected].
Table 2 Troubleshooting
The LED on the front panel There is no power at the outlet. Ensure the outlet has power going to it.
display screen does not
illuminate. The unit is not plugged in. Ensure the unit is securely connected to
an electrical outlet.
The drums do not turn. Program was not started. Press START STOP.
The test length is incorrect. The wrong operating mode is Set the proper mode (counter or timer).
selected. See “Running a Test” on page 26 and
page 28.
This page was intentionally left blank, except for this message.
6
Service and Warranty
Service and Warranty Information 40
Agilent Technologies 39
6 Service and Warranty
Warranty Limitations
Agilent Technologies, Inc. makes no other warranty, either express or
implied, with respect to this product. Specifically disclaimed are any
implied warranties of merchantability and fitness for a particular use. In
no event will Agilent Technologies, Inc. be liable for any indirect,
incidental, or consequential damages arising from the use of this
product. This warranty gives you specific legal rights which may vary
from state to state or province to province, so you may have other rights
and some of these exclusions may not apply to you.
Exclusive Remedies
The remedies provided herein are the customer’s sole and exclusive
remedies. In no event shall Agilent Technologies, Inc. or its
representatives be liable for any direct, indirect, special, incidental, or
consequential damages, whether based on contract, tort, or any other
legal theory. Some states or provinces do not allow the exclusion or
limitation of incidental or consequential damages, so the above
limitation or exclusion may not apply to you.
This page was intentionally left blank, except for this message.
This page was intentionally left blank, except for this message.