Sop Test
Sop Test
July 2005
Contents
Acknowledgement 3
Introduction 4
Serology 5
Clinical Chemistry 21
Hematology 43
2
Acknowledgements
3
Introduction
The Ministry of Health recognizes the need to develop a National Public Health
Laboratory system to support the Antiretroviral Treatment (ART) program as high
priority. The care and treatment of HIV/AIDS requires the provision of laboratory
testing services at all levels of health facility, including HIV diagnosis, CD4 count,
clinical chemistry and hematology. The success of ART program is dependent on
the ability to diagnose and qualify HIV-positive patients for therapy and to monitor
treatment efficacy, toxicity and drug resistance.
While some existing laboratory facilities in the country are able to provide the
necessary tests, currently there are several initiatives underway to ensure that a
range of laboratory tests necessary for ART program are available to all
Rwandans.
In order to provide quality laboratory services, the National Reference Laboratory
in collaboration with Management Sciences for Health / Rational Pharmaceutical
Management Plus have developed Standard Operating Procedures (SOPs) for
laboratory tests in support of ART program.
This SOPs manual is useful in improving and maintaining quality laboratory
services by providing laboratory personnel with standardized written information
on how to perform tests. The SOPs will also provide guidance in the procurement
of laboratory reagents and supplies.
4
Serology
SOP SERO 001/ 2005: Rapid HIV screening test (Determine Method) 6
SOP SERO 002/ 2005: Rapid HIV screening test (Unigold Method) 9
HIV screening 13
SOP SERO 005/2005: Rapid Plasma Reagin (RPR) test for Syphilis 15
5
SOP TITLE: RAPID HIV SCREENING TEST (DETERMINE METHOD)
SOP No.: SERO 001 / 2005
DATE: July 2005
Clinical significance
Requirements
Equipment
Automatic precision pipettes
Reagents
Abbott Determine test kit
6
Procedure
Quality Control
Results
Interpretation of Results
• Reactive (positive) (two bars): Red bars appear in both the control window and
the patient window of the strip. Any visible red colour in the patient window
should be interpreted as reactive.
• Nonreactive (negative) (one bar): One red bar appears in the control window of
the strip and no red bar appears in the patient window of the strip.
• Invalid (no bar): If there is no red bar in the control window of the strip (even if a
red bar appears in the patient window), the result is invalid and should be
repeated.
7
Test Limitations and Sources of Error
o Infection with a variant of the virus that is less detectable by the Determine
HIV assay configuration
o HIV antibodies in the patient that do not react with specific antigens
utilised in the assay configuration specimen-handling conditions, resulting
in loss of HIV multivalency
References
User’s manual for Determine HIV 1/2. Abbott Laboratories, Illinois, USA.
8
SOP TITLE: RAPID HIV SCREENING TEST (UNIGOLD METHOD)
SOP No.: SERO 002 / 2005
DATE: July 2005
Clinical significance
Whole blood or serum may be used. If the sample cannot be tested on the same
day of collection it should be stored at 2 to 8(C. Serum samples not required for
testing within 48 hours should be stored at -15 to -25(C. The use of haemolysed
samples, incompletely clotted blood, or samples contaminated with bacteria must
be avoided. Multiple freeze-thaw cycles of samples must also be avoided.
Requirements
Refer to the product literature.
Procedure
Refer to the product literature.
Quality control
This test has an in-house quality control of both positive and negative samples.
9
Interpretation of results
A negative result does not exclude the possibility for a patient being in the
window period or having been exposed to HIV infection. All positive results must
be repeated using another rapid test (CAPILLUS). If the results are discordant,
perform the ELISA.
Refer to the product literature
References
10
SOP TITLE: RAPID HIV SCREENING TEST (CAPILLUS METHOD)
SOP No.: SERO 003 / 2005
DATE: July 2005
Clinical significance
The latex agglutination test uses latex beads coated with recombinant or
synthetic antigen. Small amounts of whole blood or serum are mixed with latex
beads and rotated on a card. After several minutes, the tests are evaluated by
eye in bright light for level of agglutination relative to that of the negative control.
Whole blood or serum may be used. If the sample cannot be tested on the same
day of collection it should be stored at 2 to 8(C. Serum samples not required for
testing within 48 hours should be stored at -15 to -25(C. The use of haemolysed
samples, incompletely clotted blood, or samples contaminated with bacteria must
be avoided. Multiple freeze-thaw cycles of samples must also be avoided.
Requirements
Refer to the product literature.
Procedure
Refer to the product literature.
Quality control
This test has an in-house quality control of both positive and negative samples.
Interpretation of results
Refer to the product literature.
11
Disposal of test materials
References
12
SOP TITLE: ENZYME-LINKED IMMUNOSORBENT ASSAY (ELISA)
FOR HIV SCREENING
SOP No.: SERO 004 / 2005
DATE: July 2005
Clinical significance
The assay is used for the detection of antibodies against HIV infection.
Principle
Serum or plasma may be used. If the sample cannot be tested on the same day
of collection it should be stored at 2 to 8(C. Samples not required for testing
within 48 hours should be stored at -15 to -25(C. The use of haemolysed sample,
incompletely clotted blood, or samples contaminated with bacteria must be
avoided. Repeated freezing and thawing of samples should also be avoided.
Requirements
1. ELISA reader
2. ELISA kit
3. ELISA washer
4. Water bath of incubator with relative humidity
5. Automatic pipettes (single and multi-channel)
6. Glassware
7. Timer
13
Procedure
Interpretation of results
Samples giving an absorbance value of less than the cut-off value are
considered negative. Samples giving an absorbance equal to or greater than the
cut-off value are considered initially reactive in the assay and should be
confirmed by another, preferably a specific, method.
Quality control
The assay has an in-house quality control system of both positive and negative
samples. Manufacturers always specify the lower and upper limits for cut-off
validity.
Test plates and other materials used in the test should be disinfected, preferably
with 1% sodium hypochlorite solution (household bleach), before disposal.
The plates should then either be burnt or buried with other waste materials.
References
14
SOP TITLE: RAPID PLASMA REAGIN (RPR) TEST
SOP No.: SERO 005 / 2005
DATE: July 2005
Principle
Serum and plasma samples may be used. If the sample cannot be tested on the
day of collection it should be stored at 4 to 8°C. Samples not required for testing
within 48 hours should be stored at -15 to -25°C. The use of haemolysed
samples or samples suspected to be contaminated with bacteria must not be
used to avoid false positive results. Repeated freezing and thawing of samples
should be avoided
Requirements
1. RPR test kit (containing a bottle of carbon coated antigen, test cards, sample
dispenser/mixers, a dropper for antigen dispensing, positive and negative
control sera).
2. Mechanical rotator with a horizontal stage operating at 100rpms
3. Centrifuge
4. Timer
Procedure
4. Mix the sample and antigen and spread the mixture to the full area of the
15
circle.
5. Rotate for 8 minutes using the mechanical rotator.
6. Immediately after 8 minutes, read the results by visual inspection in good light.
7. Record the result
It is important to refer to the product literature provided in the RPR test kit for
specific instructions.
Quality control
Each run of the test should include control samples (which came with the kit, or
known samples). These controls must give the expected pattern for the test run
to be valid.
Interpretation of results
A positive result is indicated by the presence of clearly visible clumps of the black
particles. In a negative result the carbon particles remain in even suspension, no
aggregates are visible. Results should be reported as “Reactive” or “Non-
reactive”. All positive results should be confirmed by specific tests to rule out
false-positive results.
RPR is not a specific antibody test against Treponema pallidum, the causative
agent for syphilis. The test has been shown to give positive results in other
conditions such as acute malaria, leprosy and in pregnancy. Specific tests like
the Treponema Pallidum HaemAgglutination (TPHA) or Fluorescent Treponema
Antibody-Absorption (FTA-Abs) tests must be used to confirm positive results.
Uneven mixing of the antigen suspension results in the carbon particles settling
at the bottom of the bottle and only the watery part of antigen being dispensed.
Failure to observe specified time for reaction. The results must be read
immediately after 8 minutes for them to be valid.
Test kits must not be used after expiry. Attention must be paid to expiry dates.
Any visible or particulate matter must be removed from the sample prior to
testing
The test cards are not re-usable. The disposal of the cards is very easy as they
can be burnt, or dispose of with other rubbish after adequate disinfection,
preferably with 1% sodium hypochlorite (household bleach)
16
References
Fortress Diagnostics Limited, BT41 1QS, United Kingdom, Reviewed April 2000
17
SOP TITLE: PREGNANCY TESTING
SOP No.: SERO 006/ 2005
DATE: July 2005
Principles
Direct latex agglutination. The direct latex agglutination tests fix antibody against
hCG to latex particles. The presence of hCG in the test sample will produce
agglutination. A positive test, is therefore, associated with a visible agglutination
reaction. As a general rule, the end point of these assays is more difficult to
detect than that of the agglutination-inhibition reactions.
18
on the membrane and are subsequently immobilized by the gaot anti-rabbit
antibodies coated on the membrane at the control region, forming a pink band.
This control band services to validate thee test results.
Urine/serum samples collected at any time of the day may be used. However, the
first morning specimen of urine is recommended as it contains the highest
concentration of hCG. If there is any delay in testing, samples may be stored at 2
to 8C but the testing should preferably be done as soon as the samples are
collected. Please refer to the product literature for specific instructions.
Requirements
Procedure
Quality control
Each test should include control samples. These are usually provided in the test
kits. Please refer to the product literature.
Interpretation of results
This list is not necessarily complete. Thus, the product literature must be
consulted for each test used. Every commercial product is somewhat unique and
changes are introduced from time to time. It is important for the laboratory
technician to become familiar with the specified details of any technique he/she
proposes to use.
References
20
CLINICAL CHEMISTRY
21
SOP TITLE: SERUM UREA (UREASE BERTHELOT METHOD)
SOP No.: CHEM 001/ 2005
DATE: July 2005
Urea is the main end product of protein metabolism in the body. Removal of the
amino group from Amino acids from which urea is formed takes place in the liver.
It is excreted by the kidney in urine. Therefore, its measurement in serum and
urine can be used to assess kidney function.
Principle
Urea is hydrolyzed by the action of the urease to produce ammonia and carbon
dioxide. The ammonia reacts with hypochlorite and phenol in the presence of
nitroprusside to form indophenol, which in alkaline medium gives an intense blue
colour. The intensity of the colour formed is directly proportional to concentration
of urea in the sample.
Requirements
Equipment
Reagents
• Urea testing kit
• Low and high controls
Procedures
For kit methods, follow manufacturer’s instructions.
Quality Control
Include the low and high controls with each run. The values obtained for low and
high controls must be within ± 2 SD of the given range. C.V.% must be ¡Ü5%.
22
Result / Calculations of results
Linearity
Test is linear up to 33.3 mmol/L for plasma or serum. Serum samples with urea
concentration higher than this limit should be diluted 1 in 5 with normal saline and
the result multiplied by 5.
Urine samples must be diluted 1 in 20 with distilled water and the result multiplied
by 20.
Reference Range
Interpretation of results
References
23
SOP TITLE: SERUM CREATININE (JAFFE’S REACTION)
SOP No.: CHEM 002/ 2005
DATE: July 2005
Clinical significance
Principle
Creatinine in alkaline solution reacts with picric acid to form a coloured complex.
The intensity of the colour formed is directly proportional to the creatinine
concentration in the sample.
Requirement
Equipment
• Spectrophotometer
• Timer
• Water bath/heat block
Reagents
• Creatinine testing kit
• Low and high controls
Note: The working reagent is made by mixing equal volumes of picric acid and
sodium hydroxide solutions.
Procedure
Deproteinisation
24
Quality Control
The integrity of the reaction will be monitored by use of low and high control sera.
The control values should lay within ± 2 SD of the expected range. If the results
of the quality control materials are outside this range, the run must be repeated.
Linearity
Reference Ranges
References
www.biomerieux.com
25
SOP TITLE: SERUM BILIRUBIN TOTAL AND CONJUGATE (JENDRASSIK
AND GROF)
SOP No.: CHEM 003/ 2005
DATE: July 2005
Clinical significance
Bilirubin test is used to investigate the causes of liver diseases and jaundice and
to monitor patient progress e.g an infant with serious neonatal jaundice (high
level of unconjugated bilirubin).
Principle
Total bilirubin is determined by its reaction with diazotised sulphanilic acid in the
presence of caffeine to give a red compound, the intensity of which is directly
proportional to the concentration of bilirubin.
Direct (conjugated) bilirubin is determined by the same method in the absence of
caffeine.
Requirements
Equipment
• Spectrophotometer
• Timer
Procedure
Quality Control
Include the low and high controls with each run. The values obtained for low and
high controls must be within ± 2 SD of the given range. C.V.% must be ¡Ü5%.
26
Results / Calculations of results
Linearity
Reference Ranges
Interpretation of results
Bilirubin is light sensitive, so samples must be kept in a cool, dark place. Analyze
bilirubin samples on the same day of collection. Haemolysis interferes with the
test.
References
27
SOP TITLE: SERUM TOTAL PROTEINS (BIURET REACTION)
SOP No.: CHEM 004/ 2005
DATE: July 2005
Principle
Cupric ions in an alkaline solution interact with peptide bonds, resulting in the
formation of a coloured complex. The intensity of the colour formed is directly
proportional to the concentration of protein in the sample.
Requirements
Equipment
• Spectrophotometer
• Timer
Reagents
• Protein testing kit
• Low and high controls
Procedure
Quality Control
The values obtained for low and high controls must be within ± 2 SD of the given
range. C.V.% must be Ü5%.
28
Results and calculations
Linearity
The method is linear up to 150 g/L. If the total protein concentration exceeds 150
g/L, dilute the sample with an equal volume of normal saline and multiply the
result by 2.
Reference Range
Interpretation of results
Disposal of waste
References
bioMerieux, Produits et reactifs de laboratoire, Marcy l’Etoile 2003, Total Proteins Kit,
www.biomerieux.com
29
SOP TITLE: SERUM ALANINE AMINOTRANSFERASE (ALT)
SOP No.: CHEM 005/ 2005
DATE: July 2005
Clinical Significance
Principle
ALT
2-oxoglutarate + L-alanine ¨ L-glutarate + Pyruvate (1st
method)
ALT
2-oxoglutarate + L-alanine ¨ L-glutarate + Pyruvate (2nd
method)
LDH
Pyruvate + NADH, H+ Lactate + NAD+
The rate at which the NADH is consumed is measured at 340 nm. This rate is
proportional to ALT catalytic activity.
Requirement
Equipment
• Water bath
• Spectrophotometer
• Stopwatch or timer
30
Reagents
Refer to literature
• ALT testing kit
• Sodium hydroxide (0.4 mol/L, 16 g in 1 L distilled water)
• Calibrator
• Low and high controls
Procedure
Refer to manufacturer instructions
Quality Control
The values obtained for low and high controls must be within ± 2 SD of the given
range. C.V.% must be ¡Ü5%.
Results / Calculations
Refer to manufacturer instructions
Linearity
The reagent is linear up to 290 IU/L.
Reference Range
Interpretation of results
31
References
www.biomerieux.com
32
SOP TITLE: SERUM ASPARTATE AMINOTRANSFERASE (AST)
SOP No.: CHEM 006/ 2005
DATE: July 2005
Clinical significance
Principle
AST
2-Oxoglutarate + L-aspartate glutamate + Oxaloacetate (1st
method)
AST
2-Oxoglutarate + L-aspartate glutamate + Oxaloacetate (2nd
method)
MDH
Oxaloacetate + NADH, H+ Malate + NAD+
The rate at which the MDH is consumed is measured at 340 nm. This rate is
proportional to AST catalytic activity.
Requirement
Equipment
• Spectrophotometer
• Timer
• Water bath
33
Reagents
• AST testing kit
• Sodium hydroxide (0.4 mol/L, 16 g in 1 L distilled water)
• Calibrator
• Low and high controls
Procedure
Quality Control
Include both low and high controls when performing the test as shown above.
Values obtained for low and high controls must be within ± 2 SD of the given
range. C.V.% must be ¡Ü8%.
Results / Calculations
Obtain the activity of AST in the serum from the following calculation:
Factor = AST activity of Calibrator (Standard) / Absorbance of Calibrator
(Standard)
AST (TEST) = Factor × Absorbance (TEST) (U/L)
Linearity
If absorbance exceeds 0.170, dilute sample 1 in 10 with 0.9% NaCl solution and
re-assay.
Multiply the result by 10.
Reference Range
Interpretation of results
34
Disposal of waste materials
References
Product insert user’s manual for urease test kit, Randox Laboratories, UK.
www.biomerieux.com
35
SOP TITLE: BLOOD AND CSF GLUCOSE (OXIDASE METHOD)
SOP No.: CHEM 007/ 2005
DATE: July 2005
Clinical significance
Glucose is the chief source of energy in the body. The levels of this compound
are balanced by digestion and absorption of carbohydrates in the intestine, its
storage and release in the liver and its utilisation in the muscle.
Apart from the screening for and monitoring of diabetes, glucose is measured in
cases of pancreatic, metabolic or endocrinic disorders.
Principle
Glucose oxidase
Glucose + O2 gluconic acid + H2O2
Requirements
Equipment
• Spectrophotometer
• Water bath (optional) at 37ºC
Reagents
• Glucose testing kit
• Low and high controls
36
Procedure
Quality Control
Include both low and high controls when performing test. The values obtained for
low and high controls must be within ± 2 SD of the given range. C.V.% must be
Ü5%.
Results / Calculations
Linearity
Test is linear up to 22.2 mmol/L. Samples above 22.2 mmol/L should be diluted
1:1 with distilled water and the result multiplied by 2.
Reference Ranges
Interpretation of results
37
SOP TITLE: DETERMINATION OF ALPHA AMYLASE ACTIVITY
SOP No.: CHEM 008/ 2005
DATE: July 2005
Clinical Significance
Principle
Alpha- amylase
5 CNPG3 3 CNP + maltose + 2 CNPG2 + 2 glucose
Serum or urine
Requirements
Equipment
• Timer
• Spectrophotometer
• Water bath
Reagents
• Amylase testing kit
• Calibrator
• Low and high controls
Procedure
Mix reagents to make the working reagent as recommended by the reagent kit
manufacturer.
38
Quality Control
Include both low and high controls. The values obtained for low and high controls
must be within ± 2 SD of the given range. C.V.% must be ¡Ü8%. Refer to the
quality control literature in the kit.
Obtain the activity of alpha-amylase in the serum from the following calculation:
Linearity
For absorbance readings exceeding 0.160, repeat the assay with sample diluted
1 in 10 using normal saline and multiply the result by 10.
Reference Ranges
Interpretation of results
• Levels three times higher than normal are indicative of acute pancreatitis.
• High levels are also found in severe glomerular impairment, severe diabetic
ketoacidosis, and perforated peptic ulcers.
39
References
www.biomerieux.com
40
SOP TITLE: SERUM TOTAL CHOLESTEROL
SOP No.: CHEM 009/ 2005
DATE: July 2005
Clinical significance
Principle
Free cholesterol and cholesterol released from its esters are oxidised after
enzymatic hydrolysis.
The indicator quinoneimine is formed from hydrogen peroxide and 4-
aminoantipyrine in the presence of phenol and peroxide.
Requirements
Equipment
• Spectrophotometer
• Water bath
• Timer
Reagents
• Cholesterol testing kit
• Low and high controls
Procedure
Quality Control
Include both low and high controls as shown in the table above. The values
obtained for low and high controls must be within ± 2 SD of the given range.
C.V.% must be 5%.
Results / Calculations
41
Cholesterol = Absorbance (TEST) × Concentration of standard (mmol/L) /
Absorbance (STANDARD)
Linearity
The method is linear up to 19.2 mmol/L. Samples with concentrations above this
limit should be diluted 1 in 3 with normal saline and the result multiplied by 3.
Reference Range
Interpretation of results
References
www.biomerieux.com
42
HEMATOLOGY
SOP HAEM 003/2005: Automated full blood count (QBC Autoread Plus) 51
43
SOP TITLE: HAEMOGLOBIN ESTIMATION
(CYANMETHAEMOGLOBIN METHOD)
SOP No.: HAEM 001/ 2005
DATE: July 2005
Clinical significance
Used for the estimation of haemoglobin in whole blood for diagnosis of anemia.
Used as baseline to assess for anaemia before starting treatment and to assess
possible bone marrow suppression during ARV treatment, especially with
zidovudine-containing regimens.
Principle
Requirements
Equipment
• Colorimeter or spectrophotometer
• Timer
Reagents
• Haemoglobin standard (commercially obtained)
• Drabkin’s solution pH 7.0–7.4 prepared as follows:
44
stable at room temperature and should be stored in a brown bottle. If ambient
temperature is above 30ºC, solution should be kept in refrigerator. DO NOT
freeze.
Procedure
Note: If cloudiness appears in the diluted blood, centrifuge the fluid before
reading the colorimeter.
Note: Every effort must be made to use the calibration curve method because it
saves on the standard.
1. Label five tubes as tubes 1, 2, 3, 4, and 5 dilute the standard with Drabkin’s
solution as follows:
1 4.0 mL 0 mL 100%
2 3.0 mL 1.0 mL 75%
3 2.0 mL 2.0 mL 50%
4 1.0 mL 3.0 mL 25%
5 0 mL 4.0 mL 0%
2. Mix well and allow to stand for 5 minutes.
3. Zero the instrument with Drabkin’s solution.
4. Read the absorbance at 540 nm against reagent blank (if using one cuvette,
start with tube 5).
5. Prepare a graph plotting the absorbance readings of the diluted standards (Y
axis) against their concentrations (X axis).
(Example: The reference solution concentration is calculated as 18 g/dL.)
Tube Standard Drabkin’s % Haemoglobin Haemoglobin Absorbance
Quality Control
• If possible, all specimens should be tested in duplicate; results should not differ
by more than± 0.2 g/dL.
• The photometer must be checked each day by means of the haemoglobin
cyanide reference material with known haemoglobin values. The value obtained
must read within 2% of the reference material’s stated value.
• A calibration curve must be prepared with each new batch of reagents.
Results / Calculations
Haemoglobin concentration =
Absorbance (TEST) × Concentration of standard × Dilution factor of standard /
Absorbance (STANDARD)
Reference Range
Interpretation of Results
46
Test Limitations and Sources of Error
References
Lewis, S. M., B. J. Bain, and I. Bates (eds.). 2001. Dacie and Lewis Practical
Haematology. 9th edition. New York: Churchill Livingstone, pp. 19–23.
47
SOP TITLE: HAEMATOCRIT METHOD
SOP No.: HAEM 002 / 2005
DATE: July 2005
Clinical significance
Haematocrit is used to calculate the mean cell haemoglobin concentration and
cell volume. It measures the proportion of red blood cells to plasma. It is useful
as a screening test for anaemia and to diagnose polycychaemia vera and to
monitor its treatment.
Principle
The method for determining haematocrit by centrifugation is the micromethod.
Requirements
• Microhaematocrit centrifuge
• Microhaematocrit tubes plain and heparinised (75mm long with 1.5mm bore.)
• Specially designed scale for reading results e.g. Critocap Tube reader.
• Sealant e.g. cristaseal
Procedures
Immerse the tip of the capillary tube just below the surface of the blood.
Tilt the tube slightly to permit the blood to move rapidly up the capillary
tube. The capillary tube should be filled by capillary action until it is about
three-quarters filled.
Seal one end of the tube with a sealant
Place the capillary tubes in the numbered slots, making sure that the
number on the slot corresponds with specimen number.
The sealed ends of the tube should point away from the center.
Centrifuge at (RCF 10, 000 -15,000xg) rpm for 3 - 5 minutes. The tube
will show 3 distinct layers: plasma at the top, buffy layer of WBC in the
middle, and RBC at bottom.
48
Read the percentage of packed red cells by placing the spun capillary
tube on the provided scale.
Quality control
Known samples (low and high) should be run at frequent intervals and
specimens should also be run in duplicate and must agree within ± 0.01.
Results / calculations
How to use the scale
Hold the tube against the scale and align the base of the blood in the
column with zero and the bottom of the meniscus of the plasma with 100
as shown in figure 1. The value of the haematocrit is taken directly from
the reader - 50% as shown in figure 1.
The line passing through the top of the column of red cells gives the
haematocrit. If the top of the column of cells is not on a line but between
lines, its position can be estimated to the nearest digit.
100
90
80
70
60
50
40
30
20
10
0
Figure 1: Microhaematocrit reading scale showing positioning of tube
Reference ranges
49
Interpretation of results
High values are found in cases of plasma depletion e.g. severe burns and
dehydration due to diarrhoea and/or vomiting. Low values are found in patients
suffering from anaemia and in fluid overload.
References
Lewis, S. M., B. J. Bain, and I. Bates (eds.). 2001. Dacie and Lewis Practical
Haematology. 9th edition. New York: Churchill Livingstone, pp. 19–23.
50
SOP TITLE: AUTOMATED FULL BLOOD COUNT
(QBC AUTOREAD Plus )
SOP No.: HAEM 002/ 2005
DATE: July 2005
Clinical significance
Principe
Requirements
Equipment
QBC Autoreader Plus
QBC Centrifuge
QBC Printer
QBC pipette
Reagents
QBC ACCUTUBES
Procedure
51
- Fix the stopper of the other end of the ACCUTUBE
- Insert the floater
- The floater should not be touched with the fingers
- Slip the open end of the ACCUTUBE on the floater up to a partial insertion
- Finish the insertion by pushing the floater against the cover
- Mark the tube
- Centrifuge the tubes for five minutes
Quality control
QBC ACCUTUBE control or use in house sample
Reference ranges
52
Test limitation and sources of error
- Haematocrit from 15 to 65 %
The results beyond these intervals flicker and are preceded by asterisk sign
- The results apart from the intervals accepted by the machine must be
found by other techniques
- The ACCUTUBES are made to optimise the results in case of layers of
normal cells
- The errors due to technician or material used can lead to a non display of
results
- The QBC AUTOREAD PLUS cannot detect the qualitative abnormalities of
blood cells
- The QBC AUTOREAD PLUS cannot replace the leukocytic formula done
manually
- The QBC does not calculate the number of red blood cells
References
Lewis, S.M; Bain, B.S; Bates, I; Dacie and Lewis practical Hematology; 9th Ed.
2003
53
SOP TITLE: MAKING A THIN BLOOD FILM
SOP No.: HAEM 004/ 2005
DATE: July 2005
Clinical significance
The thin blood film is important in the investigation and management of anemia
infections and other conditions which produce changes in the appearance of
blood cells and differencial white cell count. The test examine the cell
morphology all blood cells and perform a differential blood cell count.
Principle
A drop of blood is placed on one end of a slide. It is spread into a thin film while
holding a spreader at a 45º angle.
Equipment
• Glass slides
• Spreader
• Orange stick
• Cotton wool
Reagents
• Methanol
Procedure
54
Results / Calculations
A well-made slide will have three distinct regions: a head, body, and tail.
Dispose of specimen containers and orange sticks in a plastic bag fixed in a bin
ready for incineration.
References
55
SOP TITLE: MANUAL DIFFERENCIAL LEUKOCYT COUNT
(THIN BLOOD FILMS)
Clinical significance
To determine the proportions of different white blood cells types in blood and look
at the morphology of white blood cells, red blood cells and platelets.
Principle
A drop of blood is spread on a slide and then fixed, stained, and examined under
the microscope. In this way, red blood cells, leucocytes, and platelets may be
studied.
The Romanowsky stains contain eosin Y, which is an acidic anionic dye, as well
as azure B and other thiazine dyes, which are basic cationic dyes. When these
dyes are diluted in buffered water, ionization occurs. Eosin stains the basic
components of blood cells; for example, haemoglobin stains pink-red, and other
methylene blue–derived dyes stain the acidic components of the cells. Nucleic
acids and nucleoproteins stain various shades of mauve-purple and violet, the
granules of basophils stain dark blue-violet, and the cytoplasm of monocytes and
lymphocytes stains blue or blue-grey. The staining reactions of Romanowsky
stains are pH dependent, which is why the stains are diluted in buffered water
with a specific pH.
Requirements
Equipment
• Differential counter
• Timer
• Microscope
• Staining racks/staining jars
Reagents
• Romanowsky stains (e.g., May-Grunwald-Giemsa, Leishman)
• Absolute methanol
• Buffered water at pH 6.8
• Immersion oil
• Thin blood films
56
Procedure
Note: Methanol should be free of water, and films should not be allowed to dry
between steps and
Quality Control
Repeat counts on selected slides on subsequent days because this will give an
indication of the range in the variation of the results. (Include note on reagent
quality.)
Procedure
Examine with 10× objective to check if the film is well made and for any
extracellular organisms.
Examine with the 40× objective and with the 100× objective if necessary.
Examine the body of the smear where the cells are lined singly and are
evenly distributed.
Count a total of 100 leucocytes and record the number of each type of
leucocyte seen. Present as a percentage of each type.
Comment on any abnormalities seen in the WBC (e.g., blastoid features of
lymphocytes, neutrophil shift left or right, toxic granulation of polymorphs,
vacuolation of cytoplasm).
Comment on the shape, size, and colour of red blood cells
.Comment on the number and morphology of platelets seen.
Reference ranges
Adults
57
Neutrophils 45–75% 2.0–7.5 x 109/L
Lymphocytes 15–45% 1.5–4.0 x 109/L
Monocytes 2–10% 0.2–0.8 x 109/L
Eosinophils 1–6% 0.04–0.4 x 109/L
Basophils 0–1% <0.01–0.1 x 109/L
Children (2-6years)
Interpretation of results
Interpretation of results
Morphologic Reports
• Red cell colour: normochromic, hypochromic, hyperchromic
• Red cell size (anisocytosis): normocytic, macrocytic, and microcytic
• Red cell shape (poikilocytosis): normal or abnormal; report the presence of
sickle cells, target cells, spherocytes, etc.
• Red cell inclusion: RNA, polychromasia, punctate basophilia, Howell-Jolly
bodies, etc.
• Report the presence of parasites such as sporazoa, nematodes, and
trypanosomes as well as bacteria such as spirochetes
• Platelets: numbers and morphology
58
Test Limitations and Sources of Error
• Clean oil off slide with xylene. Store cleaned slides for 7 days before specimen
disposal or keep slides of interest in slide racks for future reference.
• Autoclave specimen containers and dispose of them.
References
Lewis, S. M., B. J. Bain, and I. Bates (eds.). 2001. Dacie and Lewis Practical
Haematology. 9th edition. New York: Churchill Livingstone.
59
SOP TITLE: ENUMERATION OF CD4+ T LYMPHOCYTES
( FACSCount FLOW CYTOMETER )
SOP No.: HAEM 006 / 2005
DATE: July 2005
Clinical significance
The test is mainly used for baseline assessment and monitoring response to
treatment.
Principle
A single test requires one convenient, ready-to-use reagent tube pair. When
whole blood is added to the reagents, fluorochrome-labelled antibodies in the
reagents bind specifically to lymphocyte surface antigens. After a fixative solution
is added to the reagent tubes, the sample is run on the instrument. Here, the
cells come in contact with the laser light, which causes the fluorochrome labelled
cells to fluoresce. This fluorescent light provides the information necessary for
the instrument to count cells. In addition to containing the antibody reagent, the
tubes also contain a known number of fluorochrome-integrated reference beads.
These beads function as a fluorescence standard for locating the lymphocytes
and also as a quantification standard for enumerating the cells.
Requirements
Equipment
• BD FACScount instrument
• Automatic electronic pipette and tips
• Vortex mixer
• Coring station
• Cleaning tubes
• FACScount workstation
• Disposable clothing
• Biohazard waste container or bag
• Safety Cabinet class II (optional)
60
Reagents
• BD FACScount reagent kit
• BD FACScount control kit
• BD multicheck control
• BD FACScount sheath fluid
• BD FACScount rinse
• BD FACScount clean
Procedures
Label the tab of one reagent tube pair with patient laboratory number.
Vortex the reagent tube pair upside down for 5 seconds, then upright for 5
seconds.
Open the reagent tube pairs with the coring station.
Transfer the reagent tube pair from the coring station to the workstation,
keeping the tubesupright.
Close the workstation cover to protect the reagents from light.
Mix the whole blood by inverting the BD Vacutainer tubes five times.
Pipette 50 µL of blood into each of the four reagent tubes. Change the tips
between each tube.
Cap the reagent tube pairs and vortex upright for 5 seconds.
Replace the reagent tube pairs in the FACScount workstation, close the
cover to protect reagent from light, and incubate for 60–120 minutes at
room temperature (20–25ºC).
After the incubation step is complete, uncap the tubes and pipette 50 µL of
fixative solution into each reagent tube. Change tips between tubes.
Seal the reagent tube pair with new caps and vortex upright for 5 seconds.
(Fixed samples can be held up to 12 hours before adding the control
beads.)
Run the tubes on the FACScount instrument within 2 hours of adding
control beads to the reagent tubes.
Store samples at room temperature in the workstation until they are run on
the instrument. Vortex upright for 5 seconds immediately before running
and run on the BD FACScount instrument following the instructions in the
user’s manual.
Quality Control
Reference ranges
Interpretation of Results
• The lower the CD4 count, the more the disease has progressed. Treatment with
ARVs will be initiated when counts are below 500 cells/mL.(For Rwanda policy is
below 350 cells/mL)
• AIDS is diagnosed when CD4 cell counts are below 200 cells/mL.
References
62