Manual Sample Stage PEEM
Manual Sample Stage PEEM
Sample Stage
Control for FE-LEEM P90
©2013. All rights reserved. No part of this manual may be reproduced without the prior permission of SPECS GmbH. Other
product and company names mentioned in this document may be the trademarks or registered trademarks of their respective
owners and are used for identification purposes only.
Chapter 2 Installation 5
Chapter 5 Maintenance 15
5.1 Opening the Analysis Chamber 15
5.2 Removing the Sapphire Tube 17
5.3 Dismounting the Sample Stage 18
Index 19
Chapter 1
LEEM Sample Stage
The FE-LEEM P90 sample stage is equipped with high precision linear positioners that can position the sample with high
precision. It is able to move the sample in X,Y,Z directions as well as tilting the sample without moving the center point.
Stepper motors move the sample stage. These are controlled using the SmarAct Control Unit and SSC-5. This manual
describes the installation and use of these items to position the sample stage.
SmarAct supply a manual for the Control Unit and SSC-5. This is included in the delivery and provides additional tech-
nical information about the equipment.
The photograph below shows the sample stage as viewed from the sample chamber. There are a total of eight motors
arranged around the stage. For more details about these motors, please see "Normal Operation" on page 7.
1.3 SSC-5
The SSC-5 gives direct control of the sample stage positioning system. The device is pre-configured so that it is ready for
use without further changes.
1.4 Safety
Please note the following warnings and safety instructions carefully when using the product.
Caution!
The MCS controller must be installed horizontally with 3 cm air circulation area behind the ventilator.
Insufficient air flow can cause overheating, which can result in a limited functionality of the controller.
Caution!
The system is NOT hot-pluggable. Always make sure to power down the device before connecting or dis-
connecting any plugs!
Note: You can safely remove or attach the USB cable during operation. On removing the USB cable, all positioners will
be stopped immediately as a safety precaution.
Chapter 2
Installation
The sample chamber has two DN40CF feedthroughs, each with three electrical feedthroughs, as shown in Figure 4.
These are labeled A and B. All cables are labeled for easy connection.
Figure 5 shows the connection scheme for the sample stage and controllers. Please note the following points for
the connections:
■ Before starting, make sure that the power is off (0).
Note: There is an RS-232 connector for computer control. If you are using computer control with your own programs, use
a normal (not crossed) RS-232 cable for the connection.
Chapter 3
Normal Operation
The FE-LEEM P90 sample stage is fitted with eight motors to achieve sample movement. The positions of the motors
are shown in the drawing below.
■ There are four piezo motors that set the Z position and the tilt. There is an angle of 90º between the motors.
■ There are four piezo motors that set the X–Y position. Again, there is an angle of 90º between the motors; the axis
is rotated by 45º compared to the X and Y axes. Thus all four motors are in operation when moving in the X or Y dir-
ections. This arrangement distributes the weight of the sample stage over the motors.
Normal operation mode for the on the SSC-5 controller allows you to change motor position and thereby move the
sample. This chapter also describes common actions such as changing the step size for movement, zeroing the motor
position and setting the correct motor mode.
The following sections describe the groups and their corresponding axes.
Z1 -7 50µm
----
Z1 -11 50µm
Z1 7 50µm
Z1 -3 50µm
Figure 7: Group 1—Z position
T2 -5 100µm
----
T2 8 100µm
t2 -7 100µm
t2 1 100µm
Figure 8: Group 2—Tilt settings
----
X3 -81 100µm
Y3 ,
197 590 100µm
Y3 -81 100µm
Figure 9: Group 3—X,Y positions
Step size
-81 100µm
,
197 590 100µm
-81 100µm
Figure 10: Display of step size
Note: This procedure actually uses Menu Mode, described in the following chapter. However, it is a relatively common
operation, and so is included with the other normal operating procedures.
SENSOR OPTIONS
-
0.Exit>
.1.Zero All
>
-
2.Zero Channel -
>
Figure 11: Zeroing the motor positions
If the "μm" symbols is not shown, you can restore it for correct operation:
1. Press and hold the Mode Adjust button. Cursors will be shown next to the step size setting.
2. Turn one of the dials until the "μm" symbol returns next to the corresponding step size value.
3. Release the Mode Adjust button.
Y3 .Spd11
-50>
Figure 12: Changing the motor mode back to μm
Chapter 4
Editing the Configuration
It is possible to change the configuration of the SSC-5. However, the settings on delivery are optimized for the specific
task of controlling the sample stage. As such, it is strongly recommended that you do not change the settings. The
instructions in this section explain some of the more important settings and how to restore settings.
Note: The SmarAct manual contains more details about changing the configuration.
Note: In order to change these settings, you need to leave User Mode in the Control Unit. For further instructions, please
refer to the SmarAct Modular Control System User Manual.
Group 1
Group 2
Group 3
Chapter 5
Maintenance
This chapter covers the following maintenance topics:
■ Removing the sapphire tube for cleaning.
■ Removing the sample stage from the chamber, e.g. for replacing or repairing motors.
In order to perform these tasks, you need to vent and open the analysis chamber. This procedure is also described.
Caution!
Do not perform any of the procedures in this chapter without contacting SPECS first.
It is possible to damage the equipment or lose the calibration—in such cases, the guarantee is voided.
■ GV Loadlock
5. Switch off the following:
■ IGP Transfer
■ IGP Sample
■ TSP Sample
■ PG Sample
■ TP Sample
6. Switch off TP Loadlock. This will vent the chamber.
7. Remove all connectors from the top side of the analysis chamber. Unscrew the HV connector and remove each
pin individually.
You may be able to perform your tasks by removing the DN100CF viewport on the chamber and working through the
opening. However, you may need to remove the top flange.
Note: The gasket for the top flange is a special part only available from SPECS.
Figure 13: Removing the top flange from the analysis chamber
Use a new gasket when fitting the flange. This is a special gasket, available from SPECS. To tighten the gasket:
The cleaning method depends on the type of contamination and there are no "guaranteed" procedures. Experience has
shown that immersing the tube in a dilute acid (e.g. nitric acid) can successfully clean the tube.
You can remove the sapphire tube without dismounting the sample stage from the chamber:
1. Make sure that all power supplies are switched off and the cables are removed from the analysis chamber.
2. Disconnect the cables from the feedthroughs inside the chamber.
3. Remove the four bolts on the mounting plate highlighted in the figure below.
4. Carefully pull the sample holder socket away from the sample stage and remove it from the chamber.
5. Loosen the screw that clamps the mounting plate to the sapphire tube and slide the mounting plate away.
Fitting the sapphire tube back into the chamber is the reverse of the above procedure. Do not overtighten the
clamping screw.
Caution!
Do not remove the screws marked with a red cross (on the brown part)! These secure the sample stage
to the objective lens. It can be difficult to realign the sample stage if it is removed from the objective.
3. Remove the sample stage from the chamber, being careful not to get the cables tangled.
I Z Movement 8
Zero Positions 9, 11
Installation 5
Menu Mode 3, 9
Motors 7, 18
Moving 8
Piezo 7
Reference Channel 11
RS-232 2, 6
Safety 3
Sample Stage
Connectors 5
Dismounting 18
Sapphire Tube 17