MODEL A5B40

MILLIVOLT-LEVEL ISOLATION AMPLIFIER WITH FILTER


OPERATING INSTRUCTIONS

1. GENERAL 3. PHYSICAL CONNECTIONS


Campbell Scientific's Model A5B40 Isolation Connections to the datalogger for the A5B40
Amplifier is a peripheral that provides electrical are as follows:
isolation for the analog inputs of CSI
dataloggers. The A5B40 is designed around A5B40 Datalogger
the Analog Devices model 5B40-03 Millivolt
Input Module. A fourth-order, low-pass filter, IN HI Sensor HI or Mux Com HI
with a 50 Hz cutoff frequency, was added to IN LO Sensor LO or Mux Com LO
reject noise while maintaining measurement +12V 12V
rates of up to 10 samples per second. GND GND
OUT HI Analog Channel HI
The A5B40 is powered by the datalogger's 12 OUT LO Analog Channel LO
volt output, (automatically switched to conserve ENABLE Control Port or Excitation Ch.
power), and isolates one differential analog
input. In most applications, a single A5B40 is 4. MAKING MEASUREMENTS
used between an AM32 Multiplexer and the
datalogger, thus providing isolation for up to 32 The maximum and minimum signal inputs to the
analog channels with a single amplifier. A5B40 are +100 mV and -100 mV respectively.
The A5B40 applies a gain of 50 to these signal
When measuring thermocouples, the model and outputs them to the datalogger as +5 V and
A5B40 requires special datalogger PROMs. -5 V respectively.
These PROMs contain special instructions used
to linearize the thermocouple signals and The control port connected to the enable
compensate for the reference junction terminal must be switched on prior to the
temperature. Instruction making the measurement.

2. PROM INSTALLATION Measuring Millivolt Signals - Use the same


instruction that you would normally use to
21X(L): To install OSXA5-0.1, OSXA5-1.1, or measure the signal but remember to use an
OSXA5-2.1 operating system software, simply input range of 5000 mV and to apply a multiplier
follow the instructions as outlined in the of 0.02 to convert to the true mV signal.
attached document entitled "Installation of 21X
Micrologger OSX PROM Update" substituting Measuring Thermocouples - Use Instruction
the following PROM assignments in Instruction 14 as usual, but be sure to insert the digit "8" in
4: front of the thermocouple type code. For
example, to measure a type T thermocouple,
Location PROM Signature enter "81" for the type code (Parameter 4).
With the preceding 8 in the code, the datalogger
8 6704 24486 will automatically apply the multiplier of 0.02 to
7 6705 42295 the signal before converting to degrees C.
OSXA5-0.1: 6 6706 13044
OSXA5-1.1: 6 6707 33417
OSXA5-2.1: 6 6708 6820

NOTE: PROM numbers are labeled as


Item No on each individual PROM.
This is a blank page.