CRP Kit
Mea
AutoQuant
in i lite zs
ceneaqeor | Buller oo catbrator
Reagent
Packsize | 2xa0m1 210m adel
INTENDED USE
‘Autouant CRP Kits intended for the quantitative determination of
Reactive protein in human serum by turbidimetric immunoassay
fon AutoQuant 400i/ 400 Supremus/ 400 automated nical
chemistry analyzer
iTropucTiON
reactive protein (CRP) is an acute phase _ protein
synthesized in the liver Its rate of synthesis increases within hours
of acute injury or the onset of inflammation and may reach as high
as 20 times the normal levels, A rapid fll of CRP indicates recovery.
‘The degree of elevation of CAP level directy reflects the mass of
activity of inflamed tissue. Its ability to fall to normal levels on
resolution of the condition renders quantified CRP values to be a
‘good indicator to allow rapid selection of appropriate
antiinflammatory therapy in several rheumatic cseases which are
Clinically dificult to assess. Apart from indicating inflammatory
disorders, CRP levels helps in differential diagnosis, in the
‘management of neonatal septicemia and meningitis where standard
‘lcrobioiogical investigations are cicult. CRF levels se invariably
after major surgery, but fall to normal within 7-30 days. Absence of
this fall is indicative of septic or inflammatory postoperative
‘complications, Serum CRP concentration provides useful
information in patients with myocardial infarction there being an
‘excellent correlation between peak levels of CRP and creatine
phosphokinase,
PRINCIPLE OF THE TEST
‘AutoQuant CRP is @turbidimetri assay for determination of
‘Greactive protein In human serum and is based on the principle of,
agglutination reaction, When the activation buffer mixed with
sample is allowed to react with the AutoCuant CRP reagent, an
Inzoluble complexis formed inthe presence of CRP imparting
turbiulty whl cere Le viceaureu at 240 rn, The concentration of
‘CRP is dependent on the intensity of turbidity.
KIT COMPONENTS
Composition
Ri - Buffer : Phosphate buffered saline (pH 7.43), Polyethylene
lycol (40 g/l), Sodium azide (0.09 %)
R2.- Ant serum Reagent : Phosphate buffered saline (pH 7.43),
Polyclonal goat anti-human CRP (variable), Sodium azide (0.09 %)
3 - Calibrator : Dilution of human plasma and pleural fluid
containing high levels of CRP with phosphate buffered saline, liquid
and stablized, Contains 0,09 % sodium aide, Concentration Is lot
specific, see vial label,
[MATERIALS REQUIRED BUT NOT PROVIDED
41. 0.9% Saline
2, Instrumentation: fully auto analyzer
SPECIMEN:
Fresh Serum Sample
STORAGE OF TEST KIT
Unopened test kts snoutd be stored at 2-8 °C upun etept The teat
kit may be used throughout the expiration date on the kit (18,
Months from the date of manufacture). Do not freeze.
Diagnostics
PRECAUTIONS AND WARNINGS
4. For in vito use only. 7
2,This pack insert must be completely underst0o
‘operation. Do not modify the test procedure or substitute
reagents from other manufacturers or other lots unless the
reagent Is stipulated as interchangeable. It Is recommended to
handle carefully by entitled and professionally educated person,
3, Do not pipette by mouth, Use disposable gloves while performing
the assay, Wash hands thoroughly when finished.
4. Follow good laboratory practice to avoid microbial
contamination of reagents 4s Uns may reduce the lie ut he
product and cause erroneous results.
Ensure that the assay Is incubated at the mentioned
‘temperature in the protocol
Do not use reagents beyond the expiry date.
Incase of skin contact with any of the reagents, wash thoroughly,
with running water.
8, Shake the RZ: Antiserum reagent well before use to Improve test
performance.
9. The calibration curve must be validated periodically with known
controls,
(rorinwwo degree vo
Preparation of calibration curve
The concentration of R3: Calibrator is as mentioned on the vial
Iobel
Dilute the calibrator serially as mentioned below for preparation of
calibration curve.
‘The above five dilutions ofthe calibrator including the highest (D1)
and lowest (D6) concentrations of measuring range must be used
for the preparation of the calibration curve.
Tet Tube]i [2 ]3 ]4 ]5 |e
No.
Callbrator [01 [02 [os [oa | ps | 06
dilution No.
(0.9%Saline | | 100, | 100, | 100 _| 100_ | 100
sama tome
cama — ts So Oe
othe Ne P|
"The ealration eurve so obtained is valid only for the same lot of
reagents.
PERFORMANCE CHARCTERISTICS
3. Unearity
“The measuring ange I 0 to 32 mel.
2, Senstivity/ Limit of detection (LOD)
“The limit of detection Is 0.08 me/dl
3. Accuracy
Control ‘Assigned Measured
(me/al) (mesa)
eta 2.63 (2.1-3.16) 25
larg 5.26 (421-634) 5.02
Levels 7.15 (5,728.58) 7.05;
Precision
Intraassay/ Within run precision
20 | Mean 30 ow
(rg/at)_| _trng/ay (,
sample | 0.68 0.02, 3.25
sample2 | 5.62 0.03 057
tual ert eons Pt, Sz on, DD, Me Pa Sure, ES 8 74, eran Mr, Chea, Vph9t6
india, T+9'-260:3052100, F +91-260-0062125, W wirmmsife.com
ltome} Obets sa, Bd General Wanss 83, 1060, Busses, Belgum. T-+(G2) 2732-595, F +(32) 2 732-60-03, E mal@obéis.not
GujaratInter-assay/Run to run precision
Programme Parameter for AutoQuart400
n=29 | Mean 2 w “estitem CRP
(mefay_ | (mgsay_| (0 Decimal dt 3
sample [O71 0.08 325 Wait maga
sample? [5.51 oz zit Assay 2 Pont End
5. Hook Effect 5 fete as
Hook effect is > 100 ma/d re eg 2
REFERENCE VALUES ‘Second wavelength 700)
“The reference values of CRP in normal population are between 0-1 | Normal volume ef sample iG
mefd. The reference values vary with regards to the time of [absorbance lit 33
collection of the serum sample. Each laboratory should establish ts esate =
‘own reference value. ene a0
BIBLIOGRAPHY R2 0
1. MB.Pepys, G.MHirschfeld, reactive protein: acrtcal update, RS 30
the journal of CnicalInvestigstion, volume 121, 2003 Position .
2. L.Wu, THan, Xan eta, Serum C-reactive protein asa possible [Calibration method Logi Toes
marker to predict delayed hemorrhage aftr colonoscople Point .
polypectomy, Med Sci Monit 18(8): CR480-485, 2012, Galbration soltion 1106 + (ealibrator value)
3. Kseo, Lee, Lahn etal, active protein as an indicator of Sse dati
inflammatory respanses to experimentally induced cyettsin dogs
Vet Sei, 13(2):79-185, 2013,
4, Yl, 2M. Y.Mia eta, Serum C-reactive protein (CRP) 2s a
simple and independent Prognostic Factor in Exranodal Natural
Killer/-Cell ymphorra, Nasal Type, PLOS ONE, volume 8, issue 5,
264158,2013.
5. Sd.Semple, reactive protein: biological functions,
cardiovascular disease and physical exercise, SAJSM, vol 18 No. 1,
2006.
6. Data on file: Merl Diagnostics.
Ipu/cRPAQao1/01
22-07-2019
‘Sols ved on Mel Diagnosto lbs
EVEnRg
A
— Aten Se stn or Use
Balch No {ta} into bogrostes
Sxpiy0ae EAM) Conat tason bx Use
Monvcer — Stage Tongans
Kp Ory AS aap Any tom Sung
Mandactwing Dato @) Do nal sei package is damaged
(ETRE) viz Exogean Raveseraiva inthe Exopcn Comiy
Programme Parameter for AutoQuant 400i/ 400 Supremus
._ BASIC PARAMETER:
Method End Point
Primary Fiter 340
Secondary Fer NO
Unit mg/dl
Decimal z
RLVolume: 200
Ri Position *
Incubation Time 1 (Sec) 0
2 Volume 50
Incubation THe 2 (See) 20
‘Sample Volume 16
Reading Time (Sec) 36
Linearity 2
2.__ REFERENCE RANGE
Normal Low o
Normal High a
3,_ CALIBRATION:
No: of Calibration é
Calibration Rule Polynomial
Calibration Cup z
* User defined
jr lsenaelined = eel
Bar Code i valid for AutoCuvant 400 400 Supremus analyzer.