Yoo PC Mri Fourier Velocity Encoding
Yoo PC Mri Fourier Velocity Encoding
Original Research
Once a cross-sectional image through a vessel of in- The maximum velocity of each FVE spectrum was
terest was obtained, a circular region of interest (ROI) defined as the largest velocity component above a
could be graphically prescribed in the vessel to specify threshold spectral amplitude (25% of maximum), as
the sample volume for the FVE scan (see Fig. 2). The shown in Fig. 3. This 25% value was chosen empirically
application of the cylindrical excitation as a spatial- to be greater than observed background spectral arti-
saturation pulse during imaging could also be toggled fact. The maximum velocity was measured at each ser-
on and off. This allowed the true position and size of the vo-motor speed, with an uncertainty equal to the veloc-
sample volume to be visualized as a darkened region of ity resolution of each acquisition.
the image. High resolution axial PC MR images were obtained
The lower portion of the interface provided control using an ungated sequence with the following parame-
over the duration, aliasing velocity, and flip angle of the ters: flip angle ⫽ 30°, slice thickness ⫽ 10 mm, field of
real-time flow measurement. The aliasing velocity was view ⫽ 20 cm, matrix ⫽ 512 ⫻ 512, number of aver-
set manually based on the expected peak velocity in ages ⫽ 8, bandwidth ⫽ ⫾ 15.625 kHz, and TE/TR ⫽ 6
each vessel, and increased if aliasing was observed. msec/18 msec. Magnitude masking of the phase im-
Imaging paused automatically during the acquisition of ages was not performed. Spatial-maximum velocities
velocity data but resumed immediately after the velocity were measured in each PC image by averaging the ve-
spectra were displayed. locities from a small ROI (25% of tube area) positioned
In Vitro Experiments
The FVE method was first evaluated using an experi-
mental flow apparatus. This was constructed from a
gear pump (AMBV1A, Hypro Corporation, New Brigh-
ton, MN) and computer-controlled servo motor
(BSM80N-250AA, Baldor Electric Company, Fort
Smith, AR), which pumped water through reinforced
nylon tubing into the scan room. To ensure a symmetric
and developed parabolic flow profile over the measure-
ment volume, a rigid Plexiglas tube of inner-diameter
9.5 mm and length 2 m carried the pumped water
through the bore of the magnet.
Velocity data were acquired using the FVE method
and compared to the data from PC MR. Data acquisition
was performed under steady flow conditions using the Figure 3. Experimental FVE spectrum of constant flow
body coil, with the speed of the servo motor increased through the in vitro apparatus (flow waveform inset). To char-
incrementally between acquisitions. All data were saved acterize maximum flow velocities reproducibly in the presence
for offline analysis in MATLAB (Mathworks, Natick, of spectral artifact, the velocity at which the spectrum dropped
MA). to 25% of its maximum amplitude was measured (vertical line).
300 Macgowan et al.
at the point of highest flow. The correlation and velocity profile with an approximate area 25% that of
weighted linear fit between the FVE and PC MR mea- the vessel.
surements were then calculated. The FVE scans were conducted using the parameters
described above (see In Vitro Experiments section). The
MR sample volume was intentionally extended beyond
In Vivo Experiments the walls of the studied vessel during initial scans to
include signal from the surrounding static tissue. This
In vivo studies were performed in eight normal adult
static tissue appeared as a horizontal band in the wave-
volunteers using spectral Doppler ultrasound, a seg-
form that could act as a zero-velocity marker for sub-
mented PC MR imaging method, and the FVE method to
sequent acquisitions. As with the ultrasound measure-
produce 24 measurements of pulsatile flow. Approval of
ments, three FVE measurements of peak velocity were
all procedures was obtained from our institutional re-
averaged from each vessel for analysis.
view committee, and informed consent was obtained
An experienced radiologist (C.J.K.) qualitatively eval-
from each subject prior to examination. Abdominal ves-
uated the flow waveforms measured by each method
sel locations were chosen for this study because they according to 1) the number of waves shown, 2) the
were easy to access by ultrasound and easy to land- relative width and timing of the waves, and 3) the rela-
mark using either ultrasound or MR imaging. Flow tive amplitude of each wave for each vessel. Based on
waveforms were obtained in the suprarenal and infra- the number of characteristics that were equivalent be-
renal aorta and the hepatic segment of the inferior vena tween two waveforms, the agreement between FVE, ul-
cava of each subject. Waveforms were compared quali- trasound, and PC MR was rated as excellent (three
tatively according to the flow pattern and quantitatively matching characteristics), good (two matching charac-
according to the peak velocity. teristics), fair (one matching characteristic), or poor (0
Ultrasound measurements were performed by an ex- matching characteristics). Quantitatively, the correla-
perienced radiologist (C.J.K.) using a commercial du- tion and linear fit between the real-time and PC mea-
plex ultrasound system (Acuson Sequoia, Mountain surements of peak velocity were calculated and a
View, CA) with a 3-MHz transducer. All subjects had Bland-Altman analysis of the data was performed. The
fasted for over 4 hours prior to being scanned to reduce velocities measured using the two MR methods were
artifacts from intestinal gas. Pulsed-wave Doppler re- also correlated and fitted to values measured using
cordings were performed during quiet-breathing. The Doppler ultrasound.
site of sampling was guided by grey scale imaging to
position the sample gate (4 mm) in the center of the
vessel and to minimize the angle of insonation between RESULTS
the targeted vessel and the Doppler beam (less than In Vitro Evaluation of the FVE Method
60°). Recordings were made in the hepatic segment of
Presented in Fig. 4 is a comparison between the FVE
the inferior vena cava (2–3 cm inferior to the junction
and PC MR measurements of velocity obtained using
with the hepatic veins), in the suprarenal aorta (1–2 cm
the experimental flow apparatus. These measurements
proximal to the celiac trunk), and in the infrarenal aorta
were highly correlated (R2 ⫽ 0.99, P ⫽ 2.10⫺6) and had
(1 cm above the aortic bifurcation) with the subject
a unitary slope (slope ⫽ 0.97 ⫾ 0.05) over a wide veloc-
supine on an MR trolley bed outside the MR scan room.
ity range (0 –3.3 m/second).
The maximum velocity in each vessel was measured
after angle correction using built-in calipers. Three
measurements from each vessel were averaged for data Qualitative Comparison of Velocity Waveforms
analysis. Velocity waveforms were also stored for later The validated FVE method was then applied to human
qualitative analysis. volunteers. Representative velocity waveforms acquired
Each subject was then transferred into the MR scan- using spectral Doppler ultrasound and FVE are pre-
ner and velocity measurements were performed using sented in Figs. 5 and 6. These figures correspond to
the PC and FVE methods. Imaging planes were pre- measurements in the suprarenal aorta and inferior
scribed to target the same vessel locations previously vena cava, respectively.
targeted by Doppler ultrasound. Data were acquired Qualitatively, the velocity waveforms acquired using
during quiet breathing to ensure the MR and ultra- the two methods were similar. In the suprarenal aorta,
sound measurements were performed under similar waveforms from both methods exhibited systolic and
physiologic conditions. The body coil was used for all diastolic forward flows separated by a brief period of
MR measurements. retrograde flow (Fig. 5). The relative timing and ampli-
The PC measurements were performed using a seg- tude of the peaks were also consistent between these
mented k-space acquisition (17) with the following pa- waveforms. The main difference between the two meth-
rameters: flip angle ⫽ 20°, slice thickness ⫽ 6 mm, ods was that the spectrum of flow velocities displayed
matrix ⫽ 256 ⫻ 160, number of averages ⫽ 2, band- on FVE was consistently narrower than on ultrasound,
width ⫽ ⫾ 15.625 kHz, TE/TR ⫽ 6.3 msec/18 msec, resulting in clearer definition of peaks by real-time MR.
views-per-segment ⫽ 2, and cardiac phases recon- A similar agreement was obtained between measure-
structed ⫽ 20. Scans lasted approximately 160 sec- ments of venous flow in the inferior vena cava; however,
onds. Magnitude masking of the phase images was not the deflections of the triphasic flow pattern appeared
performed. Peak velocities were obtained by averaging better defined by FVE. As shown in Fig. 6, both methods
the velocities from an ROI centered manually over the depicted characteristic features of the venous waveform
In Vivo Fourier Velocity Encoding 301
DISCUSSION
including the S-wave, D-wave, and a-wave. The respi-
ratory variation of the venous flow pattern was also This study presents the first quantitative in vivo evalu-
apparent. ation of real-time FVE. The strong agreement obtained
The PC MR flow waveform resolved fewer peaks than between these MR measurements and those of Doppler
either Doppler ultrasound or FVE, and those resolved ultrasound demonstrates the clinical potential of real-
peaks exhibited different relative amplitudes and tim- time MR for hemodynamic assessment and provides
ings compared with the other two methods. For this justification for future studies of pathologic flow using
reason, the flow curves produced by ultrasound and FVE, particularly in vessels difficult to access using
FVE compared more closely with each other than they ultrasound such as the pulmonary arteries and veins.
did with the curves produced by PC MR. The qualitative Such hemodynamic measurements will complement
agreement of the ultrasound and FVE waveforms was the anatomic and other physiologic information already
excellent in 14 cases, good in seven, and fair in three. provided by MR imaging.
FVE measurements of peak velocity were shown to scanner and processed using an objective algorithm, as
correlate strongly with those from both ultrasound and was done by Steinman et al (21).
PC MR. The better agreement between the MR methods A sampling rate of 20 samples per heartbeat is gen-
was likely a result of the fact that both methods erally considered sufficient to visualize cardiac rhythms
scanned the same vessel location and the cardiovascu- (22). Therefore, the acquisition window used in this
lar physiology changed little between the two sequential study (30 msec per TR) should not introduce dramatic
MR measurements. The lower correlation between FVE temporal averaging for heart rates of at least 100 beats/
and Doppler ultrasound was attributed to differences in minute. At higher heart-rates, the aliasing velocity can
the anatomic locations of the measurements, potential be increased by narrowing each trapezoid in the read-
changes in hemodynamic state, and the different out gradient instead. As explained in the Methods sec-
sources of random error between the two modalities tion, this makes optimal use of the gradient hardware
(e.g., constrained insonation angle affects ultrasound and improves temporal resolution but requires the
accuracy). post-processing parameters to be measured separately
The shape of the sample volume from which ultra- at each aliasing velocity. This complication has moti-
sound velocity measurements were acquired also dif- vated the future development of automated methods to
fered from that of the FVE measurements, because of calibrate the data-processing parameters (i.e., k-space
fundamental differences between the two modalities. trajectory) on a per-acquisition basis. However, such
The ultrasound sample volume was defined by the pro- optimization will only improve the method and does not
file of the ultrasound beam and the gating of the re- detract from the presented results.
corded signal, while the MR sample volume was deter- Over the past decade, several advances have been
mined by the radiofrequency pulse and the spatial made in the field of real-time flow assessment by MR.
resolution of the MR acquisition. Because the velocity Real-time flow imaging has been developed by different
measurements depend on the intersection of the sam- groups to visualize flow in large vessels and the heart
ple volume and the velocity profile of the vessel, the two using rapid MR imaging methods at low spatial resolu-
methods could produce differing velocity waveforms tion (e.g., echo-planar or spiral imaging at approxi-
(18). Despite this potential difference, the measured mately 2.5–3.5 mm in-plane resolution) to achieve tem-
waveforms showed excellent qualitative agreement in poral resolutions of approximately 50 –180 msec
most cases. (5,6,23). While these methods may have sufficient res-
Absolute velocities measured using MR were consis- olution to acquire volumetric flows through major ves-
tently lower than those from Doppler ultrasound. This sels, they are limited by intra-voxel velocity averaging
bias was attributed to the known sources of velocity when velocities vary rapidly across a vessel or valve
overestimation by ultrasound related to spectral broad- (e.g., stenotic jet). Because the FVE method is not lim-
ening at large insonation angles and the Doppler gain ited by intra-voxel averaging, it may provide more ac-
settings. For example, a 40% overestimation of velocity curate and reproducible peak velocities under such
has been reported by Winkler and Wu (19) due to in- conditions. The performance of the FVE method under
trinsic spectral broadening at a 60° insonation angle, more complex flow conditions is the subject of on-going
while Hoskins (20) found peak velocity overestimated work at our institution.
by an average of 18% using standard imaging parame- Other real-time MR methods have also been devel-
ters and clinical ultrasound systems. More consistent oped to assess flow in one vessel at a time, similar to
ultrasound values for peak velocity may be expected if FVE, and have been used to measure hemodynamic
the Doppler spectra could be extracted from the clinical properties such as the velocity spectrum, flow acceler-
In Vivo Fourier Velocity Encoding 303
REFERENCES
Figure 7. Quantitative comparison between FVE and PC MR
velocity measurements in vivo. a: Peak systolic velocities mea- 1. Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow mea-
sured using the FVE method vs. using PC MR. b: The corre- surement with phase-contrast MR imaging: basic facts and imple-
mentation. Radiographics 2002;22:651– 671.
sponding Bland-Altman analysis. As observed during the in
2. Pelc LR, Pelc NJ, Rayhill SC, et al. Arterial and venous blood flow:
vitro evaluation, the velocities measured by the two methods
noninvasive quantitation with MR imaging. Radiology 1992;185:
were highly correlated (R2 ⫽ 0.87, P ⫽ 2.10⫺10) and statisti- 809 – 812.
cally equivalent (fitted slope ⫽ 1.04 ⫾ 0.08). No bias between 3. Powell AJ, Geva T. Blood flow measurement by magnetic resonance
the two measurements was observed in the Bland-Altman imaging in congenital heart disease. Pediatr Cardiol 2000;21:47–
analysis (mean ⫽ – 4 cm/second, 2 䡠 SD ⫽ 23 cm/second). 58.
304 Macgowan et al.
4. Kondo C, Caputo GR, Semelka R, Foster E, Shimakawa A, Higgins 16. Farb RI, McGregor C, Kim JK, et al. Intracranial arteriovenous
CB. Right and left ventricular stroke volume measurements with malformations: real-time auto-triggered elliptic centric-ordered 3D
velocity-encoded cine MR imaging: in vitro and in vivo validation. gadolinium-enhanced MR angiography—initial assessment. Radi-
AJR Am J Roentgenol 1991;157:9 –16. ology 2001;220:244 –251.
5. Wetzel SG, Lee VS, Tan AG, et al. Real-time interactive duplex MR 17. Pelc NJ, Herfkens RJ, Shimakawa A, Enzmann DR. Phase contrast
measurements: application in neurovascular imaging. AJR Am J cine magnetic resonance imaging. Magn Reson Q 1991;7:229 –254.
Roentgenol 2001;177:703–707. 18. Macgowan CK, Wood ML. Fast measurements of the motion and
6. Nayak KS, Pauly JM, Kerr AB, Hu BS, Nishimura DG. Real-time velocity spectrum of blood using MR tagging. Magn Reson Med
color flow MRI. Magn Reson Med 2000;43:251–258. 2001;45:461– 469.
7. Holen J, Aaslid R, Landmark K, Simonsen S. Determination of 19. Winkler AJ, Wu J. Correction of intrinsic spectral broadening er-
pressure gradient in mitral stenosis with a non-invasive ultrasound rors in Doppler peak velocity measurements made with phased
Doppler technique. Acta Med Scand 1976;199:455– 460. sector and linear array transducers. Ultrasound Med Biol 1995;21:
8. Pat GT, Pauly JM, Hu BS, Nishimura DG. One-shot spatially re- 1029 –1035.
solved velocity imaging. Magn Reson Med 1998;40:603– 613. 20. Hoskins PR. Accuracy of maximum velocity estimates made using
9. Hu BS, Pauly JM, Nishimura DG. Localized real-time velocity spec- Doppler ultrasound systems. Br J Radiol 1996;69:172–177.
tra determination. Magn Reson Med 1993;30:393–398. 21. Steinman AH, Tavakkoli J, Myers Jr JG, Cobbold RS, Johnston
10. Earls JP, Ho VB, Foo TK, Castillo E, Flamm SD. Cardiac MRI: KW. Sources of error in maximum velocity estimation using linear
recent progress and continued challenges. J Magn Reson Imaging phased-array Doppler systems with steady flow. Ultrasound Med
2002;16:111–127. Biol 2001;27:655– 664.
11. Hardy CJ, Cline HE. Broadband nuclear magnetic resonance 22. McDonald DA, Nichols WW, O’Rourke MF. McDonald’s blood flow
pulses with two-dimensional spatial selectivity. J AppL Phys 1989; in arteries: theoretic, experimental, and clinical principles. Phila-
66:1513–1516. delphia: Lea & Febiger; 1990. xviii, 456 p.
12. Pauly JM, Nishimura DG, Macovski A. A k-space analysis of small- 23. Nayak KS, Hu BS. Triggered real-time MRI and cardiac applica-
tip-angle excitation. J Magn Reson 1989;81:43–56. tions. Magn Reson Med 2003;49:188 –192.
13. Hashemi, Bradley WG. MRI: the basics. Bradley WG, editor. New 24. Itskovich VV, Kraft KA, Fei DY. Rapid aortic wave velocity measure-
York: Lippincott, Williams & Wilkins; 1997. 307 p. ment with MR imaging. Radiology 2001;219:551–557.
14. Takahashi A, Peters T. Compensation of multi-dimensional selec- 25. Macgowan CK, Wood ML. Motion measurements from individual
tive excitation pulses using measured k-space trajectories. Magn MR signals using volume localization. J Magn Reson Imaging 1999;
Reson Med 1995;34:446 – 456. 9:670 – 678.
15. Noll DC, Nishimura DG, Macovski A. Homodyne detection in mag- 26. Macgowan CK, Henkelman RM, Wood ML. Pulse-wave velocity
netic resonance imaging. IEEE Trans Med Imaging 1991;10:154 – measured in one heartbeat using MR tagging. Magn Reson Med
163. 2002;48:115–121.