0% found this document useful (0 votes)
8 views

Current Sensors Design Guide Application Note Melexis

Uploaded by

cyberjan2015
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
8 views

Current Sensors Design Guide Application Note Melexis

Uploaded by

cyberjan2015
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 33

CURRENT SENSORS

REFERENCE DESIGN GUIDE

CURRENT SENSORS
REFERENCE DESIGN GUIDE
Application Note

REV015 Page 1 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

TABLE OF CONTENTS

INTRODUCTION ................................................................................................................................ 3
HALL-EFFECT CURRENT SENSORS ..................................................................................................... 3
HALL EFFECT ..................................................................................................................... 3
HALL EFFECT CURRENT SENSORS ........................................................................................... 4
TYPES OF HALL SENSORS USED WITH FERROMAGNETIC MATERIALS .............................................. 5
Conventional Hall sensors.............................................................................................. 5
Planar IMC-Hall® sensors .............................................................................................. 5
CONVENTIONAL HALL SENSORS........................................................................................................ 6
INTRODUCTION ................................................................................................................. 6
QUICK SELECTION GUIDE .................................................................................................... 7
IMC-HALL® SENSORS ...................................................................................................................... 10
IMC VERSIONS ................................................................................................................ 11
MAGNETIC DESIGN........................................................................................................... 12
Magnetic field estimation............................................................................................ 12
Magnetic field estimation with shield ......................................................................................................... 13
Magnetic field estimation without shield ................................................................................................... 14
Conductor with neck-down.......................................................................................... 14
Stray field immunity .................................................................................................... 15
Stray field reduction ..................................................................................................................................... 15
Stray field rejection ...................................................................................................................................... 16
Use of laminated shields.............................................................................................. 17
Mechanical tolerances and vibrations ......................................................................... 17
Cancelling stray field without shield ............................................................................ 18
Avoiding cross-talk without shield ............................................................................... 18
PCB DESIGN ................................................................................................................... 19
REFERENCE DESIGNS ........................................................................................................ 20
PCB application, 2-10A range, multi-layer/multi-turn solution.................................... 21
PCB application, 2-10A range, ferromagnetic shield solution...................................... 22
PCB application, 10-50A range .................................................................................... 22
Bus bar application, 50-250A range ............................................................................ 23
Bus bar application, 300-700A range .......................................................................... 24
Bus bar application, dual range 5A/200A .................................................................... 25
Cable application, 10-100A range ............................................................................... 25
QUICK SELECTION GUIDE .................................................................................................. 26
FERROMAGNETIC MATERIALS......................................................................................................... 27
FERROMAGNETIC SUPPLIERS .............................................................................................. 27
FERROMAGNETIC SHIELDS ................................................................................................. 28
U-Shield ....................................................................................................................... 28
Mechanical assembly .................................................................................................. 28
FERROMAGNETIC CORES ................................................................................................... 29
FERROMAGNETIC MATERIALS COMPARISON .......................................................................... 31
END-OF-LINE CALIBRATION............................................................................................................. 31
ADDITIONAL INFORMATION ........................................................................................................... 32
CONTACT US ................................................................................................................................... 32
DISCLAIMER .................................................................................................................................... 33

REV015 Page 2 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Introduction
Current sensing has become more and more important in the modern world. Many applications, such
as traction inverters, electric motors, ACDC and DCDC converters, on-board chargers, battery
management systems, and many others, are relying on current sensing to improve efficiency and
safety. Melexis offers a wide portfolio of current sensors. In this document, we are going to present
the high current sensing solutions of melexis: IMC-Hall® and Conventional Hall sensors, together with
examples of assembly structures, and suggestions on how to design and build a current sensing
module. This guide is divided in 5 main sections:
1. Hall-effect current sensors: an introduction to hall effect sensors, with some basic theory.
2. Conventional Hall sensors: a description of Conventional Hall sensors, with some insight on
how to obtain an optimal design.
3. IMC-Hall® sensors: a description of IMC-Hall® sensors, an advanced technology enabling
simple assembly. This section will also focus on examples and tips to reach the optimal design.
4. Ferromagnetic materials: a description of ferromagnetic materials used in a current sensing
module, with also a focus on some assembly solutions.
5. End-of-line calibration: a description on how to reduce at minimum errors, to obtain the best
design for your application.

Hall-effect current sensors


Hall effect
The Hall effect is named after the physicist Edward Hall who discovered it in 1879. It is based on the
voltage difference arising between two extremities of a current conductor, when a current is flowing
in it and a magnetic field is applied perpendicularly (Figure 1).
The charges flowing inside the current conductor (Hall element) are subjected to the Lorentz force
caused by the magnetic field 𝐵, and drifted perpendicularly to the current flow 𝐼𝐻 . Due to charges
accumulation on the sides of the conductor, a voltage (Hall voltage 𝑉𝐻 ) is generated, that is
proportional to the magnetic field, following Equation 1, where 𝑛 is the charge density, 𝑞 is the
electron charge and 𝑑 is the material thickness.
𝐼𝐻
𝑉𝐻 = 𝐵 = 𝑆𝐻 𝐵 Equation 1
𝑛𝑞𝑑
𝐼
𝐻
𝑆𝐻 = 𝑛𝑞𝑑 is the Hall sensitivity. Equation 1 is based on a metal conductor. A semiconductor will need
a different equation, since there are 2 different charge carriers: holes and electrons. However, the
general conclusions can be drawn: the Hall sensitivity increases with the increase of the Hall current,
and with the decrease of the charge carriers density and of the thickness of the material.

REV015 Page 3 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Figure 1: Scheme showing the Hall effect.

Hall effect current sensors


In Melexis current sensors, the current conductor is a silicon surface, integrated on the silicon IC. When
it approaches a wire with a current flowing inside, it is subjected to the resulting magnetic field, that
can be calculated thanks to Equation 2.
𝜇0 𝐼
𝐵= Equation 2
2𝜋𝑟

Where 𝑟 is the distance from the conductor and 𝜇0 is the vacuum permeability. This field can be
substituted into Equation 1, obtaining the link between the Hall voltage and the current inside the
wire (Equation 3).
𝜇0
𝑉𝐻 = 𝑆𝐻 𝐼 = 𝑆𝐻 ∙ 𝐹𝐹 ∙ 𝐼 Equation 3
2𝜋𝑟
𝜇0
The parameter 𝐹𝐹 = 2𝜋𝑟 is the field factor. For geometries different than a wire, the field factor is
computed with different equations, and in general it is equal to the ratio between the field and the
generating current (𝐹𝐹 = 𝐵/𝐼). Normally, the farther is the sensor from the conductor, the smaller
the signal will be.
Signal processing plays a fundamental role in the quality of the sensor. Thanks to the integration of
the Hall element on a CMOS processed silicon IC, we can take advantage of integrated circuits for
signal variable amplification (to obtain the desired sensitivity), offset and sensitivity correction and
temperature compensation.
To concentrate the magnetic field on the sensor, and also to reduce the influence of external fields, it
is possible to use ferromagnetic elements (cores and shields).

REV015 Page 4 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Types of Hall sensors used with ferromagnetic materials


Conventional Hall sensors

Conventional Hall current sensors (Figure 2) are sensitive to the magnetic field perpendicular to the
chip surface. They are meant to be used in combination with a ferromagnetic core. In a typical
application, the core is wrapped around the current conductor and concentrates the magnetic flux on
a small air gap (typically 2-5mm) where the sensor is inserted.

Pros
▪ Strong magnetic gain from the core
▪ Very robust against cross-talk
▪ Suitable for medium to very high currents

Cons
▪ Performance limited by the core (geometry
and material): saturation, hysteresis,
frequency response and thermal drift
▪ Bigger footprint (size, weight) than solutions
based on IMC-Hall® sensors

Figure 2: Conventional Hall sensor

Planar IMC-Hall® sensors

The IMC-Hall® sensor (Figure 3) includes on top of the silicon IC a ferromagnetic layer (the Integrated
Magnetic Concentrator, IMC) whose purpose is to concentrate the magnetic field on the sensing
elements. Thanks to this technology, IMC-Hall® sensors are sensitive to magnetic fields parallel to the
chip surface. Thus, the sensors can directly measure the current flowing in a bus bar or a PCB trace
below the package, without the need for a core.

Pros
▪ Sensitive to magnetic field parallel to the chip
surface, enabling an easy SMD assembly,
vertical stacking and minimum footprint.
▪ IMC is made of magnetic material featuring
very high permeability and very low
hysteresis
▪ Magnetic gain from IMC

Cons
▪ Requires magnetic shielding or specific design
to avoid cross-talk and/or noise from external
fields
Figure 3: IMC-Hall® sensor

In the following document, we will discuss how to design at best a system with these 2 types of
sensors.

REV015 Page 5 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Conventional Hall sensors


Introduction
Conventional hall sensors are typically enclosed in the air gap of a ferromagnetic core (circular or
squared), wrapped around the current conductor (as shown in Figure 4. For a description of possible
ferromagnetic materials, see section 5).

I
d

Figure 4: Example of mounting of conventional hall sensors.


The magnetic field 𝐵 at the sensor position (in the center of the air gap), for a current 𝐼 and a
ferromagnetic core with air gap 𝑑, can be approximated as in Equation 4.

𝐼 [𝐴]
𝐵 [𝑚𝑇] = 1.25 × Equation 4
𝑑 [𝑚𝑚]

The naming convention for ferromagnetic cores is C followed by the air gap dimension (for instance,
C5 for a 5mm air gap shield). A development kit (link to application note on website) is available for
designs testing.

REV015 Page 6 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Quick Selection Guide


Melexis provides a list of sensors that can be selected for different applications (Table 1). Each sensor
can be bought with a factory trimmed sensitivity, using a specific option code. Sensitivity can be tuned
according to customer needs.

91209 91217 91219


Sensitivity [mV/mT] * 5-150 5-150 7-150
Thermal sensitivity drift [%S] ±1.5 ±1 ±1
Thermal offset drift [mV] ±10 ±5 ±5
Response time [µs] 2 2 2
Bandwidth [kHz] 250 250 400
Noise [mVrms] 10 10 2
Analog output Yes Yes Yes
Programmable Yes Yes Yes
Diagnostic functions
▪ Over/Under-voltage detection Yes Yes No
▪ Broken-track detection No Yes No
▪ Clamping No Yes No
▪ Over Current Detection No No Yes
Possible supply voltages [V] 5 5 5, 3.3
Package SIP4-VA SIP4-VA SOIC-8/SIP4-VA
Operating temperature range [°C] -40 to 150
Table 1: Main features and specifications of conventional Hall effect current sensors. Values are
typical. See datasheets for maximum limits.
* Programmable.

REV015 Page 7 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Standard configurations are listed in Table 2. Updated list is available on datasheets.

Sensor Option code Sensitivity range (typical) [mV/mT]


CAA-000 5-150 (50)
CAA-001 5-150 (15)
MLX91209
CAA-002 5-150 (7.3)
CAA-003 5-150 (19)
ACA-000 5-150 (10)
ACA-001 5-150 (15)
MLX91217 ACA-002 5-150 (17)
ACA-003 5-150 (9)
ACA-005 5-150 (13)
AAA-500 6.5-22.5 (7)
AAA-501 6.5-22.5 (10)
MLX91219 AAA-503 6.5-22.5 (15)
AAA-504 16-55 (25)
Generic part 5-105
Table 2: Option code and sensitivity range of conventional Hall effect current sensors. Please contact
your local sales representative for customized versions.
For VA packages, different lead bending options are also available, to better match the mounting
needs (see Figure 5). The shape can be selected by changing the third letter of the option code (for
instance, to select a planar leads shape, the option conde for a MLX91219 should be AAZ instead of
AAA). Table 3 resumes the length of the leads, and the reference points from which it is measured
(described in Figure 6).

Option code: xxA Option code: xxR Option code: xxS Option code: xxT Option code: xxZ
(a) (b) (c) (d) (e)

Figure 5: Different leads bending options: (a) straight leads, (b, c, d) bent leads, (e) and planar leads.

REV015 Page 8 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Option code Name Reference point Lead length [mm]


A VA straight leg Package 18
R STD-2 Dambar 5.34
S STD-3 Dambar 3.7
T STD-4 Dambar 1.68
Z SMD style Dambar 2.15
Table 3: Dimensions for each different lead bending option.

Figure 6: example of reference distances for VA package.


In case customized dimensions are needed, please contact your local sales representative.

REV015 Page 9 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

IMC-Hall® Sensors
IMC-Hall® sensors are based on an Integrated Magnetic Concentrator (IMC, a ferromagnetic layer
placed on top of the silicon IC) that locally converts the horizontal magnetic field (𝐵𝑥 ) into a vertical
component (𝐵𝑧 ) that can be measured by high performing silicon Hall elements. To concentrate the
field on the sensor, and to protect it from external fields, ferromagnetic shield are normally used (see
Figure 7).

Figure 7: IMC-Hall® sensor. The yellow regions show the approximate position of the IMCs.
Figure 8 shows how a horizontal field is bent to be converted into a vertical field at the hall plates
position (2 sensing points are present). This enables a differential sensing, rejecting any other field
that is not parallel to the sensor plane. IMC are available in different types, as explained in section 4.1.

Figure 8: IMC and field bending. Any field in the Z direction is rejected by the differential sensing.

REV015 Page 10 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

IMC versions
IMC-Hall® sensors are available in 4 different versions covering a broad range of sensitivities and
magnetic field ranges: Low Field (LF), High Field (HF), Very High Field (VHF), Extra High Field (XHF).
With its strong magnetic gain, the LF IMC is ideally suited for applications with low currents, requiring
high magnetic sensitivities (up to 700mV/mT). At the other end of the scale, the XHF IMC can linearly
sense strong magnetic fields up to ±90mT, for current sensing applications with very high-power
densities. In the following pages, Figure 9 shows the different sensitivity ranges that IMCs can reach.
Figure 10 shows the linearity ranges for each one of the IMCs, and Figure 11 shows the current ranges
in which IMCs can be used.

Figure 9: Sensitivity range of each IMC version.

Figure 10: Linearity range for different IMC configurations.

REV015 Page 11 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Figure 11: Optimal IMC configuration for different current ranges. Saturation of shields should also be
considered. Melexis offers an online simulator tool, to choose the most optimized solution for a
current sensing application, the current sensor simulator.
The design of the magnetic system around the sensor is important to be able to take advantage of the
best performances of Melexis sensors.

Magnetic design
The next sections provide a guide for an optimal magnetic design of an IMC-Hall® sensing system. First,
simple rules for magnetic field estimation will be described (section 4.2.1). Then, solutions will be
described in order to improve signal to noise ratio and stray field immunity (sections 4.2.2 and 4.2.3),
to increase frequency response and linearity (section 4.2.4), to remove the effect of mechanical
tolerances (section 4.2.5), and finally to implement the IMC-Hall® sensor without ferromagnetic
shields (sections 4.2.6 and 4.2.7). For a focus on the ferromagnetic materials, see section 5.

Magnetic field estimation

To compute the needed sensitivity, the magnetic field at the sensor location can be estimated,
considering the configuration in which the IMC-Hall® sensor is used: with or without shield.

REV015 Page 12 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Magnetic field estimation with shield

In a typical application, a U-shaped ferromagnetic shield is wrapped around the current conductor to
protect the sensor from external fields and improve the overall robustness of the sensing solution (see
Figure 12).

Figure 12: Shield and sensor assembly, with standard spacings.


The most important dimension of the shield is the inner width. In this configuration, the magnetic field
B measured by the sensor for a current I and an inner width W can be estimated as:

𝐼 [𝐴]
𝐵 [𝑚𝑇] = 1.25 × Equation 5
𝑊 [𝑚𝑚]

Figure 13 shows how dimensions are measured on a shield.

Figure 13: definition of dimensions of shields.


In order to have an increased signal-to-noise ratio, the sensor should be mounted at a certain depth
inside the opening of the shield, where the field lines are quasi-parallel to each other, and therefore
homogeneous. We suggest to follow the dimensions shown in Figure 12.
Typically, the shield thickness T ranges between 0.8 and 3 mm. The height H and length L are between
12 and 15 mm, in order to properly surround the sensor in each direction. The naming convention of
ferromagnetic shields is U (for standard shields) or LU (for laminated shields) followed by the width
number (for instance, L20). NiFe and SiFe are two available materials. Different properties are
described in section 5.4.

REV015 Page 13 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Magnetic field estimation without shield

Depending on application environment and requirements, the ferromagnetic shield is not necessarily
required. However, this configuration would lead to a drastic reduction of stray field immunity.

Figure 14: dimensions on sensor without shield.

Without shield, the magnetic field B measured by the sensor for a current I, a trace width W and a
vertical position H can be approximated as:

𝐼[𝐴] 𝑊[𝑚𝑚]
𝐵 [𝑚𝑇] = 0.4 × ∗ atan ( ) Equation 6
𝑊[𝑚𝑚] 2𝐻[𝑚𝑚]

Conductor with neck-down

In order to limit the cost, size and weight of the shield on systems with wide bus bars, we recommend
reducing its cross-section locally, as illustrated in Figure 15. Such a neck-down has minimal impact on
the electrical resistance and allows for a much more compact current sensing solution.

Figure 15: Necked-down bus bar example.

REV015 Page 14 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Stray field immunity

IMC Hall® sensors are protected from external or stray fields (i.e. fields that are not generated by the
current under measurement) by shields. Stray fields can be generated by external sources or, in case
of an inverter application, by the other phases, that should be measured by other sensors (cross talk).
Stray fields can affect both offset (in case of a constant field, not related to the measured current) and
sensitivity (when the external field changes with the measured current).

Stray field reduction

Stray field from an external source can be reduced first of all by increasing its distance with the sensor.
Figure 16 shows how cross talk decreases with the increase of the distance. These simulated values
are calculated for distances between 20mm and 170mm, with 2 cases: one with the sensor perfectly
centred in the middle of the shield, and one with the sensor displaced in the horizontal direction by
1mm. It is possible to notice that a centred sensor represents the best configuration to minimize the
stray field.

Figure 16: Cross talk error and stray field dependence with the increase of bus bar spacing.

REV015 Page 15 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Stray field rejection

When the distance cannot be increased enough to reduce the error from a stray field source, it is
necessary to use the ferromagnetic shield properties. We saw that by decreasing the shield width, the
field intensity on the sensor increases (Equation 5). Moreover, this also reduces the quantity of
external field that reaches the sensor, or, in other words, increases the ratio between the external
field 𝐵𝑒𝑥𝑡 and its component inside the shield 𝐵𝑖𝑛𝑡 , called the Shielding factor 𝑆𝐹 = 𝐵𝑒𝑥𝑡 /𝐵𝑖𝑛𝑡 (Figure
17).

Figure 17: Shield factor definition

Figure 18 shows how increasing the shield width reduces the shielding factor, leading to a larger
influence from stray field sources. In case the shield width is limited by the bus bar dimensions, it is
possible to adopt the necking down solution (see Figure 15), to locally reduce the bus bar thickness to
be able to use small shields.

Figure 18: Shielding factor versus shield width.

REV015 Page 16 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Use of laminated shields

Laminated shields are made of a stack of thin sheets separated by insulators. This allows to reduce the
effect of eddy currents (that are generated in the shield by time oscillating fields, and cause gain
reduction and phase shift of the sensor signal), boosting the sensing performances. Moreover, for very
high currents (typically above 800A), it is often more efficient and cost-effective to use a laminated
shield so that the in-plane thickness can be increased, decreasing the depth and overall footprint for
the same performance.

Mechanical tolerances and vibrations

The mechanical assembly of the sensor and shield has tolerances, that may affect the measurements
of the system. Any static error can be compensated by end of line calibration (see section 6). However,
vibrations will introduce a time variable error of the sensitivity, since movements of the sensor with
respect to bus bar and shield affect directly the field factor. The influence of displacements in the
three Cartesian directions is shown in Figure 19, while Figure 20 shows how the reference system is
chosen on the bus bar and shield assembly.

Figure 19: Effect of mechanical vibrations on the sensitivity.

Figure 20: definition of reference system on the bus bar and shield assembly
From Figure 19 we can observe that displacements in the Z direction affect the most the sensitivity of
the sensor. Displacements in X and Y directions influence less the error.
We can conclude that the IMC-Hall technology is very robust against mechanical tolerances in the X
and Y directions. For an optimal design it is recommended to strengthen the fixation along the Z
direction by placing the fixation screws close to the sensor and shield assembly.

REV015 Page 17 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Cancelling stray field without shield

In AC applications, external stray fields can be cancelled out by the microcontroller. Computing the
difference between max and min sensor output values provides a signal independent of any parasitic
DC field (see Figure 21).

Figure 21: Stray field cancelling without shield.

Avoiding cross-talk without shield

Even without ferromagnetic shield, cross-talk between adjacent current tracks can be avoided by
design. Figure 22 illustrates a concept of current trace layout with slots to force the current to flow
perpendicular to the main track axis. The sensors are rotated by 90° with their sensitive axis (blue
arrow) parallel to the current trace. With such a configuration, there is virtually no cross-talk between
phases.

Figure 22: Current trace layout with slots and rotated sensors to avoid cross-talk between phases.
PCB design is also important to have an optimal current sensing solution. Section 4.3 shows some
suggestions on how to optimally integrate the sensor and to design a PCB.

REV015 Page 18 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

PCB design
The PCB design and layout play an important role in the final performances of the current sensing
module. More specifically, two different aspects are to be considered when designing the ground layer
on the PCB.
When an application implies high voltage switching (for instance in motor control applications), an
expanded ground layer, as depicted in Figure 23, will help reducing the parasitic coupling capacitance
generated by voltage time transients (dV/dt).

Figure 23: Current sensing structure. Expanded Ground Layer on PCB in order to reduce the parasitic
coupling capacitance (coming from voltage transients).
At the same time, the ground layer can have a big impact on the response time.
If the ground layer covers all the surface of the PCB and surrounds the two legs of the shield, Eddy
currents will start to flow circularly around them, generating a counter-magnetic field which slows
down the response time of the sensor.
In order to avoid increased response time, the ground layer should be divided such that it interrupts
the circulations of Eddy currents around the shield (see Figure 24).

Figure 24: Current Sensing Structure – Designing the Ground Layer on the PCB such that Eddy Currents
circulations is interrupted

REV015 Page 19 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Reference Designs
Table 4 shows an overview of reference designs based on IMC-Hall® current sensors for different
target applications.

Application Solution Illustration


Multi-turn and multi-layer PCB or
PCB trace current
single layer PCB with C-shaped
ferromagnetic shield (section 4.4.1
2-10A
and 4.4.2).

PCB trace current Single layer PCB with or without


ferromagnetic shield (section
10-50A 4.4.3).

Bus bar High field sensor with 12mm U-


shaped ferromagnetic shield
50-250A (section 4.4.4).

Bus bar Very high field sensor with 12mm


U-shaped ferromagnetic shield
300-700A (section 4.4.5).

High dynamic range Dual range sensor with U-shaped


and C-shaped shields (section
50mA to 250A 4.4.6).

Non-intrusive current
Simple PCB with clamp-on shield
sensing from cable
wrapped around the cable
(section 4.4.7).
10-100A

Table 4: Overview of the reference designs based on planar IMC-Hall® current sensors.
In the following sections, each solution is illustrated and described in details.

REV015 Page 20 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

PCB application, 2-10A range, multi-layer/multi-turn solution

This solution is based on PCB with multiple layers and trace windings (current loops) for very high
sensitivity. It can be used with or without ferromagnetic shield, depending on sensitivity and accuracy
requirements. Figure 25 and Table 5 show the different designs and sensitivities achievable.

Figure 25: PCB layout example for very high sensitivity with 6 windings on 3 layers.

3 windings 3 windings 6 windings 6 windings


without with shield without with shield
shield shield
Sensitivity (max) [mV/A] 210 350 420 700
Table 5: Maximum achievable sensitivities with the design shown in Figure 25.

REV015 Page 21 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

PCB application, 2-10A range, ferromagnetic shield solution

This solution is based on a PCB with one layer and a single current trace (no windings). A closed
ferromagnetic shield for high magnetic gain is used to boost the current sensitivity. Figure 26 shows
the full system and some assembly details.

(a) (b)

(c) (d)
Figure 26: Top (a) and side (b) view of the full system. (c) Shield in one piece inserted through slots on
the PCB edge. and (d) shield in two parts inserted in PCB slits and assembled together.

PCB application, 10-50A range

This solution is based on a PCB with one layer and a single current trace. It can be used with or without
ferromagnetic shield (U-shaped). Two different sensitivity ranges are available: up to 60mV/A (without
shield) or up to 170mV/A (with shield). Figure 27 shows the PCB with (a) and without shield. Figure
27.c shows how to assemble the shield on the PCB.

(a) (b) (c)


Figure 27: Single-layer evaluation board (a) without and (b) with shield. (c) Assembly instructions for
shield.

REV015 Page 22 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Bus bar application, 50-250A range

This solution is based on the HF IMC sensor, that is mounted directly above the conductor on the PCB.
A simple, low-cost and compact U-shaped shield is mounted around the sensor to protect it from stray
fields and ensure good signal robustness against vibrations and displacements. With the dimensions
shown here, the linearity error is lower than ± 1.5A for currents up to ±250A. Figure 28 shows the
assembly detail, while Figure 29 shows the shield dimensions. Plot in Figure 30 shows an example of
output voltage, with the linearity error.

Figure 28: Demonstrator based on MLX91206 HF sensor and U12 shield with 0.8mm thickness.

Figure 29: Shield dimensions.

20mV/A

Figure 30: Typical output and non-linearity of a sensor calibrated for ±100A.

REV015 Page 23 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Bus bar application, 300-700A range

This solution is based on the use of a VHF IMC-Hall® sensor. The measurement range can be extended
to 700A while keeping the same inner width than the one of the design described in 0, i.e. 12mm. To
obtain a linearity up to 650A, the shield thickness must be adapted from 0.8 to 1.5mm. Figure 31
shows the assembly and the shield dimensions. This compact solution allows to measure up to 700A
with a footprint of less than 2cm2. Figure 32 shows the linearity error lower than ±5A for currents up
to ±650A (i.e. no saturation of the signal).

12
1.5

Figure 31: Demonstrator based on MLX91208 VHF sensor and U12 shield with 1.5mm thickness.

Figure 32: Sensor output and linearity error versus current.

REV015 Page 24 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Bus bar application, dual range 5A/200A

This solution allows applications with a wide dynamic range. One sensor with C-shaped (closed) shield
is used for high accuracy at small currents (typical ±5A), while another sensor with U-shaped shield is
used for high saturation limit (typical ±200A). Other combinations of ranges are possible depending
on the application requirements. Figure 33 shows the assembly and shields geometry for this solution.

±5A range ±200A range

(a)

(b)

(c)
Figure 33: (a) Assembly. Images and sections for (b) closed shield and (c) U-shaped shield.

Cable application, 10-100A range

This solution is based on the clamp-on shield, that gathers the magnetic field around the cable and
concentrates it above the sensor package. Small air gap ensures high magnetic gain. The shield
geometry can be adapted to match various cable diameters and current ranges. Figure 34 shows the
designs, while Figure 35 shows the real implementation of this concept.

Figure 34: Cable clamp concepts (left: monolithic shield, right: two-part shield in plastic housing).

REV015 Page 25 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Figure 35: Implementation examples (monolithic and two-part shield in plastic housing).
Melexis has different IMC-Hall® sensors in its portfolio. The following section will introduce them,
describing their specifications.

Quick Selection Guide


Melexis provides a list of sensors that can be selected for different applications (see Table 6). Each
sensor can be bought with a factory trimmed sensitivity, using a specific option code. Sensitivity can
be tuned according to customer needs.

91208 91216 91218


Sensitivity [mV/mT] *
▪ Extra High Field version (XHF) 20-125 12-115
▪ Very High Field version (VHF) 30-200 30-200 18-165
▪ High Field version (HF) 50-300 50-350
▪ Low Field version (LF) 100-700
Thermal sensitivity drift [%S] ±1.5 ±1 ±1.5
Thermal offset drift [mV] ±10 ±5 ±5
Bandwidth [kHz] 250 250 400
Noise 10 mVrms 6.5 mVrms 2 mVrms
Analog output Yes Yes Yes
Programmable Yes Yes Yes
Diagnostic functions
▪ Over/Under-voltage detection Yes Yes No
▪ Broken-track detection No Yes No
▪ Clamping No Yes No
▪ Over Current Detection No No Yes
Possible supply voltages [V] 5 5 5, 3.3
Package SOIC-8 SOIC-8 SOIC-8
Temp. range [°C] -40 to 150
Table 6: Main features and specifications of planar IMC-Hall® current sensors. Values are typical. See
datasheets for maximum limits.
* Programmable
Standard sensors are available in different option codes (Table 7). Updated list is available on
datasheets.

REV015 Page 26 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Sensor Option code IMC version Sensitivity range (typical) [mV/mT]


CAL-000 LF 100-700 (250)
CAH-000 HF 50-300 (100)
MLX91208
CAV-000 VHF 30-200 (40)
CAV-001 VHF 30-200 (60)
ACH-000 HF 50-350 (100)
ACV-000 VHF 30-200 (40)
ACV-001 VHF 30-200 (60)
MLX91216 ACV-002 VHF 30-200 (30)
ACX-000 XHF 20-125 (25)
ACX-001 XHF 20-125 (30)
ACX-002 XHF 20-125 (20)
ARV-500 VHF 33.5-71 (40)
ARX-501 XHF 20-40 (30)
ARV-303* VHF 22-35 (30)
MLX91218
AFV-204** VHF 47.5-165 (80)
Generic VHF VHF 18-165
Generic XHF XHF 12-115
Table 7: Option code and sensitivity range of planar IMC-Hall® current sensors (supply voltage equal
to 5V unless otherwise specified). Please contact the Melexis sales department to have customized
versions.
* Supply voltage 3.3V
** Fixed mode, supply voltage 3.3V
Melexis provide a development kit for IMC-Hall® sensors. The application note is available on the
website, at this link.
After the presentation of the two main technologies available with Melexis, the next sections will focus
on the ferromagnetic materials selection, and on the calibration of sensors.

Ferromagnetic materials
This section introduces Melexis’ main partners for ferromagnetic materials supply. Moreover, it
introduces standard dimensions for shield and core ordering.

Ferromagnetic suppliers
Melexis partnered with MagLab for ferromagnetic components supply (

www.maglab.ch
Figure 36).

www.maglab.ch

REV015 Page 27 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Figure 36: Logo and website of MagLab.


Section 5.2 is dedicated to shields, while section 5.3 is dedicated to cores. Section 5.4 shows the
difference between the available ferromagnetic materials.

Ferromagnetic shields
U-Shield

Standard (U) and laminated (LU) shields can be ordered from the Maglab website using the following
order codes convention (valid for both types). Figure 37 shows shields image and cross sections. Table
8 shows an example of shield ordering, and Table 9 shows different shield thicknesses.

U (or LU) -Shield – Width – Length – Height – Thickness (– Ni)


Figure 37: Ordering information for the standard shield from MagLab.

Order code example W [mm] L [mm] H [mm] T [mm]


U (or LU) -Shield – 12 – 13 – 12.5 – 0.8 12 13 12.5 0.8
Table 8: Shield dimension convention for U-shields and LU-shields.

Material Specification T [mm] Ni [%]


Standard material 0.8 48
Other Thickness options 0.35 / 0.5 / 1 / 1.2 …
Table 9: Different thicknesses for U-shields.
We always suggest to use the standard shield dimensions, as this minimizes the fabrication costs and
does not influence the final module performances.

Mechanical assembly

Ferromagnetic shields can be assembled by crimping, screwing or bonding (glue or tape). They can
also be encapsulated in a pre-molded plastic case. The optimal solution depends on the application.
In any case, care should be taken to avoid mechanical stress on the part of the shield involved in the
magnetic measuring circuit. Figure 38 shows different solutions for shields assembly.

Figure 38: Several solutions for shield assembly.

One of the most common solutions is to use a pre-molded plastic case, with slots to insert the shields,
as illustrated in Figure 39.

REV015 Page 28 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Figure 39: Assembly with pre-molded plastic case.

Ferromagnetic cores
Amorphous ferromagnetic cores can be ordered with the order codes convention shown in Figure 40.
Table 10 shows an example of core order number.

AMC1R5 – L
Figure 40: Amorphous ferromagnetic core sections, with ordering convention.
Order code example L [mm]
AMC1R5 – 5 5

REV015 Page 29 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Table 10: Ordering example for amorphous ferromagnetic core.

Laminated cores can be ordered using the order codes convention shown in Figure 41. Table 11 shows
an example of core order number.

Ca – b – c
Figure 41: Ordering convention for laminated cores.

Order code example a [mm] b [mm] c [mm]


C2.5 – 4 – 3.8 2.5 4 3.8
Table 11: Order code example for laminated cores.
Cores and shields are available in different ferromagnetic materials.

REV015 Page 30 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Ferromagnetic materials comparison


The performance of the current sensing solution relies on the careful selection of a proper core or
shield material and manufacturing conditions (annealing, lamination, etc.).
Table 12 displays the main features of the most common material types.

Saturation field Hysteresis


Material Price
density BSAT[T] [%FS]
SiFe $$ 1.5 <0.25%
50% NiFe $$$ 1.3 <0.1%
ferrite $ 0.5 0.1%

Table 12: Features of most common ferromagnetic core materials.

End-of-line Calibration
Each current sensor is individually tested and calibrated over temperature in the Melexis production
line. However, in order to achieve optimal accuracy, a final calibration is required at customer-side
after assembly to compensate for mechanical tolerances (sensor position deviations, shield
dimensions, etc.)
This final calibration can be done in two ways: either by using the Melexis tools to directly program
the sensor EEPROM, or by adjusting the gain/offset at microcontroller level.
(a) (b) (c) (d)

Figure 42 resumes how the error are added. We can see that the mechanical assembly introduces an
error in the field factor, that affect the total error. A final calibration will lower the error down to the
case of the sensor after fabrication.Figure 1

(a) (b) (c) (d)

Figure 42: Sum of error of sensor and mechanical assembly. We can see that the mechanical assemble
(a) introduces an error that will affect the total error (c). An end of line calibration (d) will lower again
the error to the value of the sensor after fabrication (b).

REV015 Page 31 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Additional information
Please refer to the following document for additional information on specific topics:

Typical cores and shield geometries


Standard designs of laminated and un-laminated U-shields and C-cores.

Current sensors programming and calibration


Different options available for customers in terms of sensor calibration.

Contact us

To get in contact with our current sensors application team, please fill and submit the technical inquiry
form.

REV015 Page 32 of 33 Sept-23


CURRENT SENSORS
REFERENCE DESIGN GUIDE

Disclaimer
The content of this document is believed to be correct and accurate. However, the content of this document is furnished "as
is" for informational use only and no representation, nor warranty is provided by Melexis about its accuracy, nor about the
results of its implementation. Melexis assumes no responsibility or liability for any errors or inaccuracies that may appear in
this document. Customer will follow the practices contained in this document under its sole responsibility. This documentation
is in fact provided without warranty, term, or condition of any kind, either implied or expressed, including but not limited to
warranties of merchantability, satisfactory quality, non-infringement, and fitness for purpose. Melexis, its employees and
agents and its affiliates' and their employees and agents will not be responsible for any loss, however arising, from the use
of, or reliance on this document. Notwithstanding the foregoing, contractual obligations expressly undertaken in writing by
Melexis prevail over this disclaimer.

This document is subject to change without notice, and should not be construed as a commitment by Melexis. Therefore,
before placing orders or prior to designing the product into a system, users or any third party should obtain the latest version
of the relevant information.
Users or any third party must determine the suitability of the product described in this document for its application, including
the level of reliability required and determine whether it is fit for a particular purpose.

This document as well as the product here described may be subject to export control regulations. Be aware that export might
require a prior authorization from competent authorities. The product is not designed, authorized or warranted to be suitable
in applications requiring extended temperature range and/or unusual environmental requirements. High reliability
applications, such as medical life-support or life-sustaining equipment or avionics application are specifically excluded by
Melexis. The product may not be used for the following applications subject to export control regulations: the development,
production, processing, operation, maintenance, storage, recognition or proliferation of:
1. chemical, biological or nuclear weapons, or for the development, production, maintenance or storage of missiles for such
weapons;
2. civil firearms, including spare parts or ammunition for such arms;
3. defense related products, or other material for military use or for law enforcement;
4. any applications that, alone or in combination with other goods, substances or organisms could cause serious harm to
persons or goods and that can be used as a means of violence in an armed conflict or any similar violent situation.

No license nor any other right or interest is granted to any of Melexis' or third party's intellectual property rights.

If this document is marked “restricted” or with similar words, or if in any case the content of this document is to be reasonably
understood as being confidential, the recipient of this document shall not communicate, nor disclose to any third party, any
part of the document without Melexis’ express written consent. The recipient shall take all necessary measures to apply and
preserve the confidential character of the document. In particular, the recipient shall (i) hold document in confidence with at
least the same degree of care by which it maintains the confidentiality of its own proprietary and confidential information,
but no less than reasonable care; (ii) restrict the disclosure of the document solely to its employees for the purpose for which
this document was received, on a strictly need to know basis and providing that such persons to whom the document is
disclosed are bound by confidentiality terms substantially similar to those in this disclaimer; (iii) use the document only in
connection with the purpose for which this document was received, and reproduce document only to the extent necessary for
such purposes; (iv) not use the document for commercial purposes or to the detriment of Melexis or its customers. The
confidentiality obligations set forth in this disclaimer will have indefinite duration and in any case they will be effective for no
less than 10 years from the receipt of this document.

This disclaimer will be governed by and construed in accordance with Belgian law and any disputes relating to this disclaimer
will be subject to the exclusive jurisdiction of the courts of Brussels, Belgium.

The invalidity or ineffectiveness of any of the provisions of this disclaimer does not affect the validity or effectiveness of the
other provisions.
The previous versions of this document are repealed.

Melexis © - No part of this document may be reproduced without the prior written consent of Melexis. (2022)

IATF 16949 and ISO 14001 Certified

REV015 Page 33 of 33 Sept-23

You might also like