0% found this document useful (0 votes)
256 views23 pages

Spacelabs 91370 Service Manual

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
256 views23 pages

Spacelabs 91370 Service Manual

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 23

91370 Service Manual

070-1315-00 Rev. E
Maintenance

Contents
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Mechanical Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Electrical Safety Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Functional Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Assembly/Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22

Overview
The following information describes the requirements and tests necessary for safety and performance
verification of the monitor.

Caution:
Observe precautions for handling electrostatic-sensitive devices!

Note:
• Never touch electrostatic-sensitive electronic components without following proper anti-static
procedures, including the use of an ESD wrist band and mat. An electrostatic discharge from your
fingers can permanently damage electronic components and cause latent failures.
• All static-sensitive electronic components are packaged in static-shielding bags. Retain the bag for
repackaging the component should you need to store it or return it to Spacelabs Healthcare for any
reason.
• Cleaning, preventive maintenance, and safety checks should be performed annually and following any
product disassembly/assembly. Preventive maintenance and safety checks must be performed by
trained personnel only.

Required Test Equipment


• Electrical Safety Analyzer — Dynatech Nevada 232C or equivalent
• Patient Simulator — Dynatech Nevada 300B or equivalent

Note:
• Before testing, ensure that the module housing(s) and its DC power supply(s) have their associated
cables attached, but are not interconnected.
• Ensure that the Ethernet cable and AC power sources are removed.
• The AC line is auto detected; no user selection is necessary.
• Ensure that the leakage test equipment is calibrated.

91370 Service Manual 4-1


Maintenance

Mechanical Inspection
Verify that:
• The monitor and all optional equipment are clean.
• All screws are tight.
• The case and connector pins are not damaged.
• There are no frayed or pinched wires or cables.

Electrical Safety Testing


Safety testing protects the patient from electrical shock, especially micro-shock. It has been determined
experimentally that current values in the microampere (μA) range may cause fatal arrhythmias in electrically
susceptible patients. A patient is deemed electrically susceptible when connected to monitoring equipment.

Definitions
Classification — IEC/EN/UL 60601-1 Safety standard designation for the class of equipment and type of
patient applied parts that indicate the degree of protection provided against electrical shock.
Leakage Current — Current that is not functional. It includes patient leakage, ground leakage, and enclosure
(or chassis) leakage.
Patient Lead Leakage — Current that flows from the applied part of the patient lead to ground.
Chassis Leakage — Current flowing from the enclosure (or from conductive parts accessible to the operator)
through the ground conductor.
Normal Condition — Condition in which all means provided for protection are intact. Includes, ground
connections, insulation, creepage and clearance distances.
Single Fault Condition — Open ground, open neutral, line voltage on a patient connection, or any single state
other than normal condition that could compromise patient safety.
UUT — Unit Under Test.
Spacelabs Healthcare does not endorse standards to the exclusion of others. Therefore: BE SURE TO
CHECK YOUR LOCAL REQUIREMENTS TO ENSURE YOUR EQUIPMENT SAFETY TESTS COMPLY
WITH LOCAL STANDARDS. Generally accepted standards for medical monitoring equipment, such as the
Underwriters Laboratory (UL) and the National Fire Protection Association (NFPA) standards, are summarized
in Table 1.

91370 Service Manual 4-2


Maintenance

Table 1: Summary of Standards for Medical Monitoring Equipment

International Mains to U.S. (120 V) Mains to


Mains Resistance
Chassis Leakage Chassis Leakage

100 μA - normal condition, ground 300 μA - normal condition, ground


500 milliohms*
attached (AC connector to chassis) attached (AC connector to chassis)

500 μA - single fault condition, open 300 μA - single fault condition, open
500 milliohms*
ground or reverse polarity ground or reverse polarity

* Measured from the AC Power cord third wire ground to the most distant ground attachment

Equipment Required
Electrical Safety Analyzer, Fluke model 232D or equivalent.
These tests should be performed according to the hospital’s scheduling requirements, at least annually or after
repair or modification.

Warning:
Before starting safety tests, ensure that no patient is connected to the device under test. If safety
tests must be performed on equipment currently monitoring a patient, obtain permission to
disconnect the cables from the monitor and patient.

Note:
All tests must be performed according to the safety analyzer’s operations manual, and any local
requirements.

Ground Resistance
1 Attach the power cord to the monitor under test, then measure the resistance from the AC power cord third
wire ground to a chassis location, such as the equipotential post on the rear of the monitor.
2 Verify that the resistance is less than 500 milliohms (0.5 ohms).

Chassis Leakage Current Tests


1 Plug the leakage analyzers into mains power.
2 Plug the equipment into the analyzer’s AC receptacle.
3 Verify that the leakage current from the chassis to ground is less than the values in Table 2.

91370 Service Manual 4-3


Maintenance

Table 2: Enclosure Leakage Test Conditions and Limits

Neutral Ground International


Polarity U.S. Limit
Condition Condition Limit

Closed neutral Closed ground Normal polarity 100 μA 300 μA

Open neutral Open ground Normal polarity 500 μA 300 μA

Closed neutral Open ground Normal polarity 500 μA 300 μA

Patient Lead Leakage Current Tests (Patient Modules)


Refer to the service manual of the specific module(s) you are using for patient lead leakage test instructions.

Preventive Maintenance
A Spacelabs Healthcare Field Service Engineer or qualified hospital biomedical technician should check the
monitor and optional equipment for acceptable performance and electrical safety to ensure they operate
according to current requirements.

Touchscreen Calibration
In the event the touchscreen becomes difficult to use or a replacement has been installed, you may need to
calibrate it. This can be performed using the calibration function in the Biomed menu.
To perform the calibration, attach a mouse (in the event the touchscreen is unusable) and perform the
following:
1 Power ON the monitor with the mouse attached and wait for the monitor to pass the self tests.
2 Click (or touch) the MONITOR SETUP key and then click (or touch) the PRIVILEGED ACCESS key.
3 Enter the Biomed password (default is Biomed).
4 Click (touch) MONITOR CALIBRATION.
5 Click (touch) TOUCHSCREEN CALIBRATION and wait for the Reset Monitor dialog box to display. Click
(touch) the RESET MONITOR key. The monitor will reset.
6 After 15 to 20 seconds, a lighted box will appear in the upper left corner of the screen. Touch the lighted box
for one or two seconds until it disappears.
7 Two more lighted boxes will appear, one at a time, in the upper right and lower right corners. Touch them
as they appear.
8 After touching these three corners one at a time as the boxes appear, the touchscreen is calibrated.

91370 Service Manual 4-4


Maintenance

Functional Tests

Monitor Self-Test
A power-ON self-test is performed each time the monitor is turned ON.
• The power ON/OFF switch (refer to Figure 2-1 on page 2-2) illuminates indicating +5 V is present.
• If the monitor is operating on external AC power, the power LED on the front bezel also illuminates.
• If the monitor is operating on AC power and batteries are being charged, the battery A and battery B status
LEDs on the front bezel flash ON and OFF at a constant rate. Once the batteries are fully charged, these
LEDs remain ON. If no battery is present, one or both of the battery status LEDs will remain OFF.

Monitor Functional Tests


This procedure verifies operation of the recorder assembly, network connectivity, alarm relay output, and
external SDLC connections. It assumes that a 91496 multiparameter module (ECG, RESP, two PRESS
channels, and two TEMP channels) is available for testing purposes. In the event that these parameters are
not available, similar modules may be substituted.

Verifying Monitor Functions

Note:
To begin the test, make sure no modules are inserted and no Flexports or other SDLC devices are
connected.
1 Plug in the monitor and verify that the front bezel power LED is lit. If one or two batteries are installed, verify
that the appropriate battery LED(s) are ON solid or flashing.
2 Press the power ON/OFF switch to power ON the monitor and verify the following sequence of events:
a The power ON/OFF switch illuminates.
b The embedded alarm light LEDs illuminate briefly. Verify that all LEDs are functional.
c The Diagnostic menu displays. (If errors are noted during power ON, contact your Spacelabs
Healthcare Field Service Engineer.)
d If batteries are installed, the fan begins operating within 60 seconds after turning ON the monitor. The
fan will then turn OFF unless the internal temperature requires additional cooling by the fan.
e Monitor keys display along the right side of the screen.
3 Insert the ECG module without a patient cable connected. Verify that the ECG parameter key appears on
the screen with ??? and the message LEADS OFF.
4 Connect a patient simulator to the ECG input with a patient cable. Set the simulator to a known rate and
verify that the ECG count and the lead being monitored are displayed to the right of the ECG parameter key.
Also verify that the ECG waveform appears on the screen.
5 Connect a patient simulator to the invasive pressure inputs.
6 Zero the pressures and verify that the numerics and waveforms are accurate.

91370 Service Manual 4-5


Maintenance

Optional Recorder Assembly


If the monitor is configured with the optional recorder assembly, verify that the monitor initiates a recording by
touching the RECORD key and then touching one of the flashing parameter keys.

Note:
The printed waveform should be free from defects such as gaps, extra lines, etc. If not, notify a Spacelabs
Healthcare Field Service Engineer for servicing, or replace the recorder (refer to Recorder on page 5-19).

Alarm Relay Output


If the monitor is configured to use the alarm relay output to interface to an external alarm device:
1 Plug the cable into the external alarm connector and initiate a high-priority alarm. Verify that the external
alarm responds appropriately.
2 If an external alarm light capable of displaying multiple colors is attached, turn the high-priority alarm OFF
and generate a medium-priority alarm. Verify that the external alarm responds appropriately.
3 Repeat step 2 for a low-priority alarm.
An alternate method is to connect the cable and turn the monitor ON and then OFF. Each color of the
embedded alarm light will momentarily illuminate. An external alarm light will also briefly illuminate.

Ethernet (Wired)
1 Attach a 10/100BaseT cable from an active network onto the side panel connector.
2 Touch the SPECIAL FUNCTIONS key and verify that other monitors appear in the Remote View menu.

Assembly/Disassembly Procedures
Caution:
Before beginning any disassembly procedures, power OFF the monitor, disconnect the AC cord
from the AC power receptacle, and remove the batteries.

Note:
The external power supply is not designed for disassembly.

Required Tools and Parts


• Anti-static mat with wrist strap
• #1 and #2 Phillips-head screwdriver
• 5/16-inch nut driver
• 3/16-inch nut driver

91370 Service Manual 4-6


Maintenance

Setup for Disassembly


1 Remove the external power supply cable and battery(ies) if present.
2 Remove the patient parameter module.

Installing or Replacing the Optional Recorder Assembly


Note:
Unless purchased separately for a monitor already in the field, the recorder option is installed at the
factory. Order an upgrade from your local Sales Representative.
The recorder option consists of two components: the recorder module and the recorder CPU PCBA.
The recorder CPU PCBA and recorder module fit into the upper slot on the monitor’s left side (refer to Figure 4-
1). The recorder CPU PCBA must be installed first, followed by the recorder module. Refer to Table 1 on page
6-1 for the recorder CPU and recorder assembly part numbers.

Installing the Recorder

Warning:
A safety hazard is created if the mylar insulator is not properly installed over the recorder CPU
PCBA.
1 Power OFF the monitor.
2 Remove the printer “dummy panel” by removing the screw located in the top portion of the battery
compartment.
3 Remove the recorder hold cover and insert the recorder CPU into the connector located at the bottom of the
recorder compartment.
4 Tighten the thumbscrew to secure it to the bottom of the recorder compartment after fully inserting the PCBA
into the connector.
5 Install the insulator sheet between the bottom of the printer and newly installed recorder CPU PCBA.
6 Insert the recorder assembly, ensuring that it fully engages the connector at the back of the recorder
compartment. Press the release bar to open the recorder assembly and tighten the two phillips-head screws
located at the rear of the recorder.

91370 Service Manual 4-7


Maintenance

'

(
Figure 4-1: Recorder assembly and CPU PCBA

' Release bar


! Recorder assembly
( Recorder CPU PCBA, installed

Replacing the Recorder


1 Remove the recorder assembly (if installed) by first opening the printer door, loosening the two captive
screws, and then pulling out the assembly at the top.

'

! (
Figure 4-2: Optional recorder assembly

' Release bar


! Recorder assembly

91370 Service Manual 4-8


Maintenance

( Recorder CPU PCBA, installed


2 If replacing the recorder CPU PCBA, loosen the thumbscrew and pull it forward until it is free of the
connector.

Figure 4-3: Recorder CPU PCBA

3 Reverse the procedure to re-assemble the recorder.

Inserting Recorder Paper


Two rolls of thermal recorder paper have been included with the optional recorder assembly.
To insert a roll of paper:
1 Press the right side of the release bar at the top of the recorder assembly. The front of the recorder assembly
is hinged at the bottom and will drop open, revealing two spoon-shaped arms that hold the paper roll spindle
between them.
2 Unroll a short length of paper from the roll, and orient the roll so that the paper feeds from the bottom.
3 Slip the paper roll spindle between the plastic arms, and close the front of the assembly so that the end of
the paper roll protrudes out of the recorder assembly just below the release bar.

91370 Service Manual 4-9


Maintenance

Removing the Bezel Assembly


1 Remove the bezel by removing the five screws along the bottom of the enclosure, and then stand the
monitor upright and remove the four screws along the top of the bezel.

'

Figure 4-4: Bezel screw locations

' Five screws along the bottom of the bezel


! Four screws along the top of the bezel
2 With the monitor standing upright, gently move the top of the bezel and disconnect the flat cable running
from the display to the CPU PCBA.

91370 Service Manual 4-10


Maintenance

'

Figure 4-5: Removal of the front bezel

' Front panel power/LEDs cable


! Video cable

Caution:
• Never pull on the flex cable. This may damage the cable end or the connector.
• Take care when handling cables and connectors. They are fragile and can be damaged.

Note:
• Due to the short length of the video cable, remove it prior to fully opening the case.
• Proper operation of the monitor depends upon correct cable re-insertion following disassembly. Ensure
that the flex cable is aligned properly and fully inserted into the connector prior to locking the connector
tab down.
3 Remove the touchscreen five-wire cable from the CPU PCBA.
4 Remove the power switch/monitor assembly by sliding the “ZIF” connector locking mechanism outward,
then removing the cable from the socket.
5 Unlatch the touchscreen cable and remove the eight 6-32-inch × 1/4-inch screws (three along the top, three
in the center, and two at the bottom of the PCBA).
If you are replacing the front bezel, proceed to the next section (Replacing the Display or Touchscreen
Assembly, Backlight Inverter, and Power Monitor Assemblies) for removal of the remaining front bezel
assembly components.

91370 Service Manual 4-11


Maintenance

Replacing the Display or Touchscreen Assembly, Backlight Inverter, and


Power Monitor Assemblies
1 Remove the front bezel assembly as described in Removing the Bezel Assembly on page 4-10.
2 Remove the four screws securing the display assembly to the front bezel and carefully lift the assembly from
the bezel.

Figure 4-6: Removing the touchscreen display bracket

Caution:
The glass touchscreen may stick to the front of the display bracket. Do not allow it to fall off of the
bracket during removal.
3 Remove the four screws securing the display to the brackets.

Note:
Be careful not to leave finger prints or smudges on the front of the display or on the inside of the front bezel
surface.
4 Reassemble the bezel assembly. Pay close attention to placement of the touchscreen, assuring that it is
placed within the raised portion of the bezel prior to securing the display assembly. If alignment is not
correct, the touchscreen can be permanently damaged.

91370 Service Manual 4-12


Maintenance

Figure 4-7: Touchscreen placement

5 To remove the backlight inverter, remove both cable assemblies and the two screws attaching the inverter
to the display bracket.
6 When replacing the backlight inverter, be sure to replace the mylar shield (Figure 4-8) between the PCBA
and the chassis.

'

Figure 4-8: Backlight inverter

' Backlight inverter

91370 Service Manual 4-13


Maintenance

Replacing the Connector PCBA

Removing the Rear Panel


1 Remove the five 6/32-inch × 1/2-inch screws.

Figure 4-9: Rear panel replacement

2 Remove the two 10/32-inch × 1/2-inch screws.


3 Carefully remove the rear panel and set it aside for further disassembly.

Removing the Connector PCBA


1 Remove the six 3/16-inch jack screws that secure the SDLC, video, and RS-232 connectors.
2 Remove the two 6/32-inch × 1/4-inch screws that secure the SDLC PCBA to the rear panel.
3 Reinstall the replacement connector PCBA in reverse order.

Replacing the CPU PCBA


1 Remove the batteries, if present.
2 Remove the bezel (refer to Removing the Bezel Assembly on page 4-10).
3 Remove the nine screws that hold the CPU PCBA to the chassis (refer to Figure 4-10).

91370 Service Manual 4-14


Maintenance

"

1
/

$ *
Figure 4-10: Nine screws holding the CPU PCBA to the chassis

4 Place your fingers into the space on either side of the upward protrusion of the CPU PCBA (located near
the two top mounting screws) and pull. This will free the CPU PCBA-to-Interconnect PCBA connection.
5 Pull the CPU PCBA from the top carefully until the cables along the bottom of the assembly are visible.
6 Disconnect the three cables and set the CPU PCBA into a static-protected area.
7 Reinstall the replacement CPU PCBA in reverse order.

Replacing the Wireless Radio Card


If the monitor is equipped with optional wireless functionality (option Z), refer to the following instructions if
radio card replacement is necessary.
To remove the wireless radio card:
1 Remove the batteries, if present.
2 Remove the bezel (refer to Removing the Bezel Assembly on page 4-10).
3 Hold the PCMCIA card adapter in place and press the eject bar ('). The radio card (!) releases from
adapter. Remove the card and detach the antennas.

91370 Service Manual 4-15


Maintenance

'

Figure 4-11: Radio card released from the PCMCIA card adapter

To install the wireless radio card:


1 Remove the replacement radio card (P/N 010-1644-xx) from the packaging.
2 Reattach the antennas (refer to Figure 4-12).

Caution:
Be very careful to properly align the antenna connector and the receptacle before applying
pressure to mate the two.
If the connector cannot be seated using finger pressure, use the blunt end of a small tool (such as a
5/32-inch hex wrench) to apply pressure to the connector.

Antenna

X
No!
Yes

Radio Card

Figure 4-12: Antenna connection to the radio card.

3 Insert the radio card into the Compact Flash card adapter on the main CPU PCBA. When the card is fully
seated in the Compact Flash card adapter, the card eject bar (') extends. Refer to Figure 4-13.

91370 Service Manual 4-16


Maintenance

'

Figure 4-13: Radio card fully seated into the PCMCIA card adapter

Replacing the Radio Card Antennas


1 Remove the radio card as described in Replacing the Wireless Radio Card on page 4-15.
2 Detach the antennas from the radio card.
3 Detach the antennas from the side bezel of the monitor. Use a thin bladed knife, if necessary, to remove
any remaining foam tape adhesive.
4 Remove the replacement antennas (P/N 117-0165-xx) from their packaging.
5 Carefully connect the two antennas by inserting the antenna connectors into the micro-coaxial connector
receptacles on the radio card (either antenna can be connected to either connector). Ensure that the
connector and the receptacle in the radio card are properly aligned before applying pressure to connect
them (refer to Figure 4-12). Do not stress the antenna cables.
6 Refer to Figure 4-14 for the correct antenna placement (').
To attach the antennas:
a Remove the peel-off covering over the foam tape adhesive.
b Apply the antenna’s adhesive pad to the tip of a thin-bladed knife or small flat screwdriver.
c Use the knife or screwdriver to guide the antenna to the correct installation point (refer to Figure 4-15).
Ensure that the antenna cable is not stressed or kinked.

'

Figure 4-14: Antenna placement

91370 Service Manual 4-17


Maintenance

7 Press down firmly, and hold the adhesive pad in place while removing the screwdriver or knife.

Figure 4-15: Antenna installation, using a knife blade as a guide

91370 Service Manual 4-18


Maintenance

Removing the Chassis Assembly


The chassis assembly includes the sheet metal, Interconnect PCBA, and the fan/battery connection assembly.

'

Figure 4-16: Chassis assembly

' Fan/battery connection assembly


! Interconnect PCBA
( Battery enclosure
To remove the chassis assembly:
1 Remove the rear panel, bezel, CPU PCBA, module, and optional recorder and recorder CPU as previously
described.
2 Place the monitor face down with the bottom facing you.
3 Remove the screw near the ground terminal area.
4 Remove the two screws with washers from the rear panel.

Note:
These washers must be used when replacing the screws. If they are left out, the screws will penetrate into
the battery chassis and interfere with the battery clearance.
5 Remove the three 6/32-inch × 0.375-inch screws from the bottom of the monitor.
6 Turn the assembly over and remove the two top screws to secure the wall of the module compartment to
the rear housing.
7 To lift the chassis out of the case, place your left hand through the battery door and your right hand into the
module access opening. Work the assembly upward, allowing the top to move ahead of the bottom.
8 The chassis is now ready for further disassembly.

91370 Service Manual 4-19


Maintenance

Replacing the Fan/Battery Contact Assembly


1 Remove the chassis assembly from the housing as previously described (refer to Removing the Chassis
Assembly on page 4-19).
2 Separate the chassis assembly from the module wall assembly by removing the two screws above the
battery ejection springs and the two screws near the bottom.

Interconnect PCBA

Interconnect frame

Module door

Module wall
assembly

Battery divider
Chassis assembly

Battery ejection
springs

Fan/battery contact assembly


(includes fan)
Battery ejectors

Figure 4-17: Replacing the fan/battery contact assembly and Interconnect PCBA

3 Pry up the center retainer on the battery contact assembly (near the battery ejection springs). This should
free the fan/battery contact assembly.
4 Remove the screw between the chassis and the Interconnect PCBA.

91370 Service Manual 4-20


Maintenance

Replacing the Interconnect PCBA and Frame


1 Remove the chassis assembly from the housing as previously described.
2 Separate the chassis assembly from the module wall assembly by removing the two screws above the
battery ejection springs and the two screws near the bottom.
3 Remove the screw between the chassis and the Interconnect PCBA.
4 Separate the Interconnect PCBA and frame.
5 Reassemble the monitor in reverse order.

Replacing/Aligning the Module Door


1 To remove the door, remove the two nuts fastening the hinge to the wall of the module compartment
assembly. This will free the door assembly. When installing a new door, do not tighten the two nuts until the
chassis assembly is completely secured to the rear housing.
2 After the chassis and rear housing are mated, position the door and tighten the two nuts.

front view rear view

Figure 4-18: Module door

Replacing the Embedded Nurse Alert Assembly


1 Remove the front bezel carefully.
2 Slide out the embedded nurse alert LED assembly.
3 Reverse the procedure to install the replacement.

91370 Service Manual 4-21


Maintenance

Cleaning
Clean the case by washing it with mild soap and water or use Plast-N-Glas cleaner. Use TF solvent for
cleaning the electronic connectors and contacts as necessary.

Caution:
• Do not autoclave.
• Never use solvents, acetone, abrasive cleaning agents, or abrasive cleaning pads.
• Use only approved cleaning agents including 70% alcohol, soap and water, green soap, or 10%
bleach solution.

Note:
Avoid directly spraying liquids into the recorder openings.

91370 Service Manual 4-22

You might also like