1 Configuration Changes the settings in the following items:

Language, Highcut Filter, Alarm Off Message,


Battery Insert Message, DSI Option, Pacing
Option, Freeze Function, Use Other Power,
Charge Time Clear, Default Setting, Flash Save
2 Adjust AD Changes the A/D values for ECG waveform
acquisition, biphasic HV unit, battery voltage and
paddle contact impedance threshold.
3 Check Hardware Checks the operation of the key switches, LEDs,
recorder, memories, sounds, etc
4 AD View Checks the A/D values that the main CPU
currently receives.
5 Operation Time Displays the count of external discharges and
operation time.
6 Version Up You can update the system program and data
7 Debug mode Displays the characters used in this defibrillator
and checks the flash memory.
8 Alarm Setting Select “Off” to turn off the ECG, SpO 2 and CO 2
alarms.
Configuration Language Select the language or country.
Selection list: English, Japanese, American, Chinese, Spanish, French,
Germany, Italian, Norwegian, Finnish, Other 1 to 5
Highcut Filter Select “On” to turn the high-cut filter for ECG measurement.
Selection list: On, Off
Alarm Off Message Select “On” to display a message when any of the vital alarm settings
is set to off.
Selection list: On, Off
Battery Insert Message Select “On” to display a message when the battery is not in the
defibrillator.
Selection list: On, Off
DSI Option Select “On” when using the optional DSI interface unit and/or
DSI/AUX interface unit.
Selection list: On, Off
Pacing Option On: For the TEC-5531 series defibrillators.
Off: For the TEC-5521 series defibrillators
Freeze Function Select “On” to enable the freeze function. The ECG second trace can
be frozen when the key is pressed.
Selection list: On, Off
Use Other Power Select “On” to supply power from other equipment.
Selection list: On, Off
Charge Times Clear Deletes the count of external discharges and operation time.
Selection list: Off, Exec
Default Setting Select “Exec” to return all settings including the alarm settings to the
factory default settings. Before doing this operation, write down the
necessary settings for each item.
Selection list: Off, Exec
Flash Save Select “Exec” to save the current settings in the System Maintenance
screen in the flash ROM.
Selection list: Off, Exec
Adjust AD Gain Adjust the A/D value for the ECG sensitivity.
Adjust ECG AD Setting range: 0 to 31 in of 1
Offset Adjust the A/D value for the ECG offset at gain 1/4, 1/2, 1, 2 and
4.
Setting range: -99 to 99 in steps of 1
Adjust AD CHARGE AD Adjust the A/D value to measure the charged energy in the HV
Adjust HV AD capacitor.
Setting range: -25 to +25 in steps of 1
TTR AD Adjust the A/D value to measure the TTR (transthoracic resistance).
Setting range: -25 to +25 in steps of 1
DELIVERED ENERGY Adjust the A/D value to measure the delivered energy from the HV
capacitor.
Setting range: -25 to +25 in steps of 1
Charge Time (AC) Adjust the energy charging time in AC operation.
Setting range: -5 to +5 in steps of 1
Charge Time (Batt) Adjust the energy charging time in battery operation.
Setting range: -5 to +5 in steps of 1
Adjust AD BATTERY VOLTAGE Adjust the A/D value of the voltage that is applied to the battery.
Adjust Battery AD Setting range: -25 to +25 in steps of 1
Adjust AD 100 , 200 , 350 Adjust the A/D value to measure the skin-paddle contact impedance
Adjust Paddle for 100, 200 and 350 .
Contact AD
The default setting of the following items may differ depending on the defibrillator:
Adjust ECG A/D, CHARGE AD, TTR AD, DELIVERED ENERGY, Charge Time,
BATTERY VOLTAGE, Paddle Contact AD.

English, Japanese, Chinese, Sync (cardioversion) YYYY/MM/DD


Spanish, French, Germany,
Italian, Norwegian, Finnish,
Other 1 to 5
American Def (defibrillation) MMM/DD/YYYY

“Language” sets the language which is used in the defibrillator. The default setting
of “Sync mode after CV” (the mode after cardioversion) and “Date format” in the
System Setup screen changes according to the selected language.
2J ± 0.5 J
3J ±1J
5, 7, 10 J ±2J
20 J or more ± 10%
Pacing Rate Up/Down Highlighted when the PACING RATE Up or Down key on the top
panel is pressed. (TEC-5531 series only).
Pacing Current Up/Down Highlighted when the PACING OUTPUT Up or Down key on the
top panel is pressed. (TEC-5531 series only).
Pacing Start Highlighted when the START/STOP key on the top panel is pressed
(TEC-5531 series only).
Record Highlighted when the record key is pressed.
Event Highlighted when the event key is pressed.
Paper Exit or Out of Paper Shows whether the recording paper is set or not.
Battery or No Battery Shows whether the battery pack is installed or not.
AC AC: Displayed when the AC power cord is connected to the
defibrillator
DC: Displayed when the AC power cord is not connected to the
defibrillator.
SetUp Shows the status of the Energy/Mode Select control.
Card exist or No Card Shows whether an SD card is inserted into the SD card slot.
Panel Sync Highlighted when the SYNC button on the front panel is pressed.
Panel Charge Highlighted when the CHARGE/AED button on the front panel is
pressed.
Panel Discharge Highlighted when the DISCHARGE button on the front panel is
pressed.
Paddle Charge Highlighted when the CHARGE button on the apex external paddle
is pressed.
Paddle Dis L Highlighted when the left DISCHARGE button on the sternum
external paddle is pressed.
Paddle Dis R Highlighted when the right DISCHARGE button on the apex
external paddle is pressed.
External Shows the type of the connected paddle. Displays "No Connect"
when no paddle is connected.
DSI 1, DSI 2 or DSI 1 2 DSI 1, DSI 2: Displayed when SpO2 adapter or CO2 sensor kit is
connected to the left or right SpO2/CO2 connector.
DSI 1 2: Displayed when SpO2 adapter and CO2 sensor kit are
connected to the SpO2/CO2 connectors.
Leads Highlighted when the ECG lead key is pressed.
Sens Highlighted when the ECG sensitivity key is pressed.
Suspend Highlighted when the Silence alarms key is pressed.
Alarm Highlighted when the key is pressed, then the System
Maintenance - Check Hardware screen is displayed.
Program card Instrument < Version up card Versionup Start
Instrument > Version up card Versiondown Confirm
Data card ------- Versionup Start