ABB - Pressure Transmitter
ABB - Pressure Transmitter
07
Functional specification
Measuring range limits and span limits
Sensor code Measuring range upper Measuring range lower Minimum measuring span
limit (URL) limit (LRL) Model 261GS Model 261AS
Model 261GS Gauge pressure Absolute pressure
C 6 kPa -6 kPa 0,3 kPa 0,3 kPa
60 mbar -60 mbar 3 mbar 3 mbar
24 inH2O 24 inH2O 1,2 inH2O 2,25 mm Hg
F 40 kPa -40 kPa 2 kPa 2 kPa
400 mbar -400 mbar 20 mbar 20 mbar
160 inH2O -160 inH2O 8 inH2O 15 mm Hg
L 250 kPa 0 abs 12,5 kPa 12,5 kPa
2500 mbar 125 mbar 125 mbar
1000 inH2O 50 inH2O 93,8 mm Hg
D 1000 kPa 0 abs 50 kPa 50 kPa
10 bar 500 mbar 500 mbar
145 psi 7,25 psi 375 mm Hg
U 3000 kPa 0 abs 150 kPa 150 kPa
30 bar 1,5 bar 1,5 bar
435 psi 21,7 psi 21,7 psi
R 10000 kPa 0 abs 500 kPa 500 kPa
100 bar 5 bar 5 bar
1450 psi 72,5 psi 72,5 psi
V 60000 kPa 0 abs 3000 kPa -
600 bar 30 bar
8700 psi 435 psi
IMPORTANT (NOTE)
The measuring range lower limit (LRL) for model 261AS is absolute 0 for all measuring ranges.
Change from one to two columns
Transmitter with 4 ... 20 mA output signal and HART communication Transmitter with 4 ... 20 mA output signal and HART communication
Certificate no. PTB 05 ATEX 2032 Certificate no. IECEx ZLM 10.0002
Labeling II 1/2 G Ex ia IIC T4 … T6 Labeling Ex ia IIC T6 bzw. T4 Ga/Gb
II 2 G Ex ib IIC T4 … T6 Ex ia IIIC T66°C bzw. T95°C Da/Db
Ex nA IIC T6 bzw. T4 Gc
Permissible ambient temperature range according to temperature Ex tb IIIC T66°C bzw. T95°C Db
class:
Ambient temperature Temperature class Maximum permissible ambient temperature range according to
-40 … 85 °C (-40 … 185 °F) T1 … T4 temperature class
-40 … 71 °C (-40 … 159 °F) T5 Surface
Ambient temperature Temperature class
-40 … 56 °C (-40 … 132 °F) T6 temperature
-40 … 85 °C
T4 95 °C (203 °F)
or (-40 … 185 °F)
-40 … 56 °C
T6 66 °C (151 °F)
Labeling II 1/2 D IP65 T95 °C Ex ia D (-40 … 133 °F)
II 2 D IP65 T95 °C Ex ib D
Electrical data according to designation Ex ia IIC T6 or T4
Permissible ambient temperature range: Ga/Gb and Ex ia IIIC T66°C or T95°C IP6X Da/Db
-40 … 85 °C (-40 … 185 °F)
Supply and signal circuit with "intrinsic safety" Ex ia or Ex ib type of
Supply and signal circuit with "Intrinsically safe Ex ia/ib IIB/IIC" protection, for connection to power supply units with the following
type of protection", with the following maximum values maximum values (terminal signal ±)
Ui = 30 V Ui = 30 V
Ii = 130 mA Ii = 130 mA
Pi = 0.8 W Pi = 0.8 W
Effective internal capacitance Ci = 10 nF Effective internal capacitance Ci = 10 nF
Effective internal inductance Li = 0.5 mH Effective internal inductance Li = 0.5 mH
IN ≤ 22,5 mA
UN ≤ 45 V
Transmitter with 4 ... 20 mA output signal and HART communication Transmitter with 4 ... 20 mA output signal and HART communication
Intrinsic Safety Class I; II and III; Division 1; Intrinsic Safety (Gas, order code X3)
Groups A, B, C, D; E, F, G Designation Ex ia II CT1~CT6
Class I; Zone 0; AEx ia
Group IIC T6; T4 Permissible ambient temperature range according to temperature
Non-incendive Class I, II, III, Division 2; class:
Groups A, B, C, D, F, G Ui max. = 30 V; li max = 130 mA; Pi = 0,8 W;
Degree of protection NEMA type 4X Ci = 10 nF; Li = 0,5 μH
(indoor and outdoor installation) Ex ia II CT1 … T6 T6 T5 T1 … T4
-40 … 56 °C -40 … 71 °C -40 … 85 °C
Canadian Standards Association (CSA) DIP A20 TA95 °C -40 … 85 °C
Voltage sup ply Minimum operating voltage ( VDC ) SIL: Functional safety (optional)
R(k )
23.6 mA According to IEC 61508/61511
Device with certificate of conformity for use in safety-related
A minimum resistance of 250 Ω is required for HART applications, up to and including SIL 2.
communication.
Unless otherwise stated, errors are specified as a % of the Temperature range Effect on zero signal and span
measuring span value. -10 … 60 °C (14 … 140 °F) Sensor code C, F:
The accuracy of the measurement in relation to the upper ± (0.15 % x TD + 0.15 %)
range limit (URL) is affected by the turndown (TD); i.e., the Sensor code L, D, U, R, V:
ratio of the upper range limit (URL) to the set span (URL/span). ± (0.05 % x TD + 0.05 %)
Vibration effect
±(0.10 x TD) % acc. to IEC 61298-3
Viewing window in cover (LCD display) Name plate and operating instruction language
Polycarbonate, Makrolon 6557
Communication plug connectors
Cover O-ring
EPDM Process connections
1/2-14 NPT female or male thread, DIN EN 837-1 G 1/2 B or
Plates G 1/2 B (HP) for convex seal, flush diaphragm, for installation
Plastic data plate attached to the electronics housing in ball valve.
Weight
(without options)
— Approx. 0.7 kg (1.54 lb)
— Additional 650 G (1.5 lb)
Packaging
Carton with dimensions of approx. 240 x 140 x 190 mm
(9.45 x 5.51 x 7.48 in.)
Mounting dimensions
(not design data) - dimensions in mm (inch)
Standard version
1 2 Ø 60 (2.36) 3 15 (0.59)1)
8 (0.31)2)
64 (2.52)
4
36 (1.42)
A
Ø 21 (0.83)
Ø 28 (1.1)3) 7
5
6
M10684-01
Dimension "A" is dependent on the process connection as per the following table:
Version with the options “LCD indicator” and “Harting Han plug”
72 (2.83)
A+8
36 (1.42)
Ø 21 (0.83)
Ø 28 (2.13)1)
M10685-01
Fig. 2: Dimensions — with options
1 Housing cover for LCD indicator option
Dimension "A" is dependent on the process connection as per the following table:
1 Wrench size 36
60 (2.36)
86,5 (3.4)
86,5 (3.4)
Ø 60 (2.36) (max. 63 (2.48))
12,5 15
(0.49)(0.59)
72 (2.83)
72 (2.83)
20
(0.79)
46 (1.81)1)
35 (1.38)2)
M10688-01
IMPORTANT (NOTE)
The bracket for wall or pipe mounting has four holes with a Ø of 10.5 mm. The holes are arranged in a square and spaced
72 mm away from one another.
Electrical connections
HART version
XX XX XX XX XX XX XX
Explosion Protection Certification
Factory Mutual (FM) - Intrinsically Safe (Note: 10) EA
Canadian Standard Association (CSA) - Intrinsically Safe (Note: 10) ED
ATEX Group II Category 1/2 G - Intrinsic Safety EEx ia EH
ATEX Group II Category 1/2 G and 1/2 D - Intrinsic Safety EEx ia (without Cable Gland) (Note: 10) EL
IECEx Ex ia IIC T6 + Ex nA IIC T6, Gas ER
IECEx Ex ia IIC T6 + Ex nA IIC T6 + Ex tb. Gas & Dust ES
GOST Russia - EEx ia W1
GOST Kazakhstan - EEx ia W3
GOST Ukraine - EEx ia WA
GOST Belarus - EEx ia WG
NEPSI Ex ia llC T-T6 Gas X3
NEPSI Ex ia llC T-T6 Gas & Dust X4
Integrated Digital Display (LCD)
With integrated LCD display L1
Electronic Housing: Ground Terminal
Housing with external grounding terminal AA
Electronic Housing: Cable Gland
Cable Gland M16 x 1.5 / M20 x 1.5 and atmosphere ventilation of metal AB
Mounting Bracket Shape / Material
For pipe mounting / AISI 304 SST (1.4301) B2
For wall mounting / AISI 304 SST (1.4301) B4
Applications: Oxygen
Oil- and grease-free, for oxygen applications (O2)
(Pmax = 21 MPa / 210 bar / 3045 psi, Tmax = 60 °C / 140 °F) (Note:11) P1
Operating Instruction Language
German M1
Italian M2
Spanish M3
French M4
English M5
Swedish M7
Finnish M8
Russian MB
Dutch MD
Unless otherwise specified prior to manufacture, the customer shall be responsible for the selection of suitable parts that make
contact with the medium and appropriate filling fluids in order to ensure compatibility with the relevant process medium.
XX XX XX XX XX XX XX
Explosion Protection Certification
Factory Mutual (FM) - Intrinsically Safe (Note: 10) EA
Canadian Standard Association (CSA) - Intrinsically Safe (Note: 10) ED
ATEX Group II Category 1/2 G - Intrinsic Safety EEx ia EH
ATEX Group II Category 1/2 G and 1/2 D - Intrinsic Safety EEx ia (without Cable Gland) (Note: 10) EL
IECEx Ex ia IIC T6 + Ex nA IIC T6, Gas ER
IECEx Ex ia IIC T6 + Ex nA IIC T6 + Ex tb. Gas & Dust ES
GOST Russia - EEx ia W1
GOST Kazakhstan - EEx ia W3
GOST Ukraine - EEx ia WA
GOST Belarus - EEx ia WG
NEPSI Ex ia llC T-T6 Gas X3
NEPSI Ex ia llC T-T6 Gas & Dust X4
Integrated Digital Display (LCD)
With integrated LCD display L1
Electronic Housing: Ground Terminal
Housing with external grounding terminal AA
Electronic Housing: Cable Gland
Cable Gland M16 x 1.5 / M20 x 1.5 and atmosphere ventilation of metal AB
Mounting Bracket Shape / Material
For pipe mounting / AISI 304 SST (1.4301) B2
For wall mounting / AISI 304 SST (1.4301) B4
Applications: Oxygen
Oil- and grease-free, for oxygen applications (O2)
(Pmax = 21 MPa / 210 bar / 3045 psi, Tmax = 60 °C / 140 °F) (Note:11) P1
Operating Instruction Language
German M1
Italian M2
Spanish M3
French M4
English M5
Swedish M7
Finnish M8
Russian MB
Dutch MD
Unless otherwise specified prior to manufacture, the customer shall be responsible for the selection of suitable parts that make
contact with the medium and appropriate filling fluids in order to ensure compatibility with the relevant process medium.
Trademarks
™ Hastelloy C-276 is a Cabot Corporation trademark
Notes
Notes
Notes