100% found this document useful (1 vote)
55 views

Calibration Slider Diagnosis Programs - 95587956

Uploaded by

muneeb
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
55 views

Calibration Slider Diagnosis Programs - 95587956

Uploaded by

muneeb
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 42

Diagnosis Programs

Diagnosis Programs
Page

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Detector lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Line scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Line signal with variance indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Further input signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Shutter and calibration slider . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Line test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Line module pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
X-ray generator test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Calibration slider test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
I/O test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Light barriers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Test of discrete light barriers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Test of light curtain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Power supply and temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Power supply test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Temperature test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Test of UPS unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Data transfer test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Ping an address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Transfer rates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Interface configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Routing table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Image transmission test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Test of material detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Functional test by means of the adjustment suitcase (34436851) . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Functional test by means of the adjustment suitcase (34432743) . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Indicator test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Conveyor test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Continuous test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Configuration display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Error report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Login report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Diagnosis report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
PI 4 Ethernet module report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

95587956 13/08/09 © Smiths Heimann 1


Diagnosis Programs

2 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Introduction

The EDt S software has several local diagnosis programs that you can reach through the Diagnosis menu. It
is possible with these programs to check all individual subsystems independently of one another.

All work may be performed by trained service staff only!


Observe the safety instructions at the front of this manual!

In some diagnosis programs it is possible to intervene deeply into the functional sequence of the X-
ray inspection unit and the communication to the BHS. Safety agreements are indispensable here!

The first step of the diagnosis represents the status display of the system. All important status information
and current error messages are displayed here in running operation.

Depending on the particular project there may be different ways of displaying the system status.
The most important information is always displayed.

The unit being in a correct operating condition, all signals show green.
If a signal is missing, or if an operating condition which momentarily does not permit an inspection has been
set, then the displays are either red for the case of an error or yellow for warning conditions.
In a communication with a BHS, relevant signals for the automatic operating sequence are displayed in fields
with a grey background.

95587956 13/08/09 © Smiths Heimann 3


Diagnosis Programs

Signal name Possible Description


display

POWER ON The power supply of the X-ray inspection unit is


switched on.

The optional overvoltage protection has tripped. The


defective module has to be replaced immediately!

The power supply of the X-ray inspection unit is


switched off.

MACHINE OK The X-ray inspection unit has no errors.

There is an error in the X-ray inspection unit.

XRAY READY The X-ray inspection unit indicates readiness for


operation. After switching on it lasts max. 5 minutes
until the display lights up green.

The X-ray inspection unit is not ready.

EVALCLIENT 2 of 2 The evaluation computers are ready.

EVALCLIENT 1 of 2 At least one evaluation computer is not ready.

All evaluation computers are not ready.

EMERGENCY STOP There is no emergency stop signal.

An emergency stop signal is present.

IL UNIT All interlock switches of the X-ray inspection unit are


closed.

At least one interlock switch of the X-ray inspection unit


is opened.

IL IO CONVEYOR All interlock switches of the pre-tunnel are closed.

At least one interlock switch of the pre-tunnel is


opened.

PRESSURE OK The system pressure of the pneumatic system is


sufficient.

The system pressure of the pneumatic system is below


the minimum limit.

4 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Signal name Possible Description


display

SCREENING The X-ray inspection unit is in normal operation.


Baggage is examined by X-ray.

NO SCAN Baggage is not examined by X-ray. The baggage


handling system controls the operation of the X-ray
inspection unit.

TRANSIT The X-ray inspection unit is switched to conveyor


operation, only the conveyor belt is controlled.

SERVICE_MODE* The service switch on the interface card stands in the


“SERVICE” position.

UPS The uninterruptible power supply of the X-ray inspection


unit is in normal condition.

The uninterruptible power supply of the X-ray inspection


unit indicates an error condition.

The uninterruptible power supply of the X-ray inspection


unit has been separated from the power supply and
feeds the connected systems through the internal
batteries.

BHS There is no fault in the communication to the baggage


handling system.

Communication to the baggage handling system is


disturbed.

CONV STATE 0
The frequency converter is not switched on and there is
no error message.

SPEED OK The frequency converter is switched on and the SPEED


OK signal is present.

The frequency converter is switched on and there is no


SPEED OK signal.

CODE_REQ A baggage code is requested by the baggage handling


1
system.

0
No baggage code is requested by the baggage handling
system.

CONV_ON The BHS starts the conveyor of the X-ray inspection


1
unit.

0
The BHS switch of the conveyor of the X-ray inspection
unit.

95587956 13/08/09 © Smiths Heimann 5


Diagnosis Programs

Diagnosis

You reach all diagnosis programs via the Diagnosis menu.

For safety reasons, the conveyor is also switched on in some test programs that work with the X-ray
generator switched on. The message shown, which must be confirmed, then appears before the X-ray
generator and the conveyor are activated.

Detector lines
When calling up the diagnosis program Detector lines the menu items Line scope and Line test will be
available.

6 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Line scope

It is possible with the Line scope diagnosis program to display graphically the individual line signals, the
supply voltage of the individual lines and the test signal of the lines.

An oscillogram showing the offset voltages of the individual detector channels as percentages appears firstly
on the screen.

The permissible range is between 2% and 15%.

Í If you switch on the X-ray generator with the first menu item X-ray generator, you receive the
maximum possible output voltages of the detector channels (full scale values).

You see initially the signal of a line with the gain adapted for the individual detectors for compensati-
on of the different detector sensitivities. The adapted gain is the result of the detector line calibrat-
ion. The permissible range is 25% to 95%.
After selection of a fixed gain factor (see below) you can check the mechanical adjustment of the detector
line and of the X-ray generator by the shape of the signal.

The circle in the upper left corner of the display indicates the direction in which the line signal will
change during the line adjustment.
If the circle rotates to the left, the total of the line signal will be reduced.
If the circle rotates to the right, the total of the line signal will be raised.

Í If you switch off the Update display function you freeze the signal. At the same time the X-ray
generator and the conveyor are also switched off.

Í In the third menu item Detectors you have the possibility of limiting the display to the HIGH or the
LOW detectors (upper/lower radiation spectrum).

Í With selection of a Zoom factor in the fourth menu item you can display details of the Line scope
magnified.
If the number of detector channels exceeds the available space on the screen (with very large units
and in the magnified display), you can select the bar located under the oscillogram with the navigator

95587956 13/08/09 © Smiths Heimann 7


Diagnosis Programs

and then scroll through the signal.

Í With Sections you can limit the display of the Line scope to the signal section which is generated by
one of the line interface boards in each case.

Í With the last but one menu item Gain you can switch the gain factors of all detector channels over to
fixed values. You select a gain factor at which the individual output signals do not exceed the 100%
line.

Line signal with variance indication

Í In addition to the normal line scope this display indicates the variance between the individual
measured values.

Í In this case the scaling of the indication will also change. A value of variance between 0 and 1000 //00
is indicated and not a value between 0 and 100%. The permissible value of variance must not exceed
the 10 % limit. If the limit is exceeded, then a mechanical adjustment of the line is required.

Í After modifying the line position the display will only stabilize after about 20 seconds.

The variance represents the difference between two raw data values measured within a very short interval
of time. A small variance is a sign of an optimum alignment of X-ray beam and detectors.
A variance exceeding 10 % on the average means that the detector lines have to be readjusted in the
respective section!
Due to the fast evaluation of data, it is quite possible that differences of the belt thickness affect the variance
value. You will notice this as a peak occurring in the horizontal section of the detector line! If a peak is
noticed over the whole detector line, then we have to assume that the X-ray generator is not stable. In such
case you have to check the adjustment of the respective XRC board! If an adjustment of the XRC board does
not improve the situation, then you have to exchange the generator!

8 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Further input signals

Í The last menu item enables you to switch the input signal of the line interface boards LIF from the
Line scope to a test signal or to display the supply voltage of the detector modules connected to
the line interface board.

Test signal

Í The test signal indicates whether the A/D converter of the line interface board is working correctly, or
the order of the connections of the line interface boards to the acquisition controller is correct.
You select the line interface board of interest via Sections (see above).

Test signal for the lines A, D and E

Test signal for the lines B and C

95587956 13/08/09 © Smiths Heimann 9


Diagnosis Programs

Detector supply

Í The supply voltage of the detector modules is displayed in the range of 4.75V (=0%) to 5.25V (=100%);
10% correspond to 50mV. The supply voltage must be within this range, if it is 0% or 100%, the
detector modules are not correctly supplied.

Detector supply of the lines A, D and E

Detector supply of the lines B and C

10 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Shutter and calibration slider

To test the correct operation of shutter and calibration slider, the following functions can be activated
separately with the generators being active:

- open shutter (line signal in line scope shows normal shape)

- close shutter (line signal in line scope must fall down to the zero line since shutters
are closed)

- insert calibration samples (line scope shows the characteristic shape of a line signal with the
calibration samples inserted, see following figure).

The changes of the respective line signal in line scope indicate the correct operation of the selected
shutter and calibration slider

The normal operation of shutter and calibration slider can be restored via the „normally controlled“
function.

95587956 13/08/09 © Smiths Heimann 11


Diagnosis Programs

Line test

The test offers you a list of faulty detector channels. These are channels the output signal of which lies
outside the range intended for the A/D conversion (below 0% and above 100%) and channels the difference
between offset and full scale value of which is less than 10%.

However, defective channels are listed only if at least three neighboring detector channels have an
error.

Line module pattern

The detector modules’ limits are projected into the existing X-ray image. This display indicates the position of
faulty modules.

12 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

X-ray generator test


This test shows you initially the current values for the heater current of the relevant X-ray generators.

The display accuracy is


±10% and therefore can-
not be used for adjusting purpo-
ses!

Í If you switch on one of the X-ray generators with menu item X-ray generator, you receive the
further values for the anode current, the high voltage and the heater current with high voltage
switched on.

The occurring error message about the wrong positioning of the shutters must be acknowledged.
This message is not relevant in this function.

Í All shutters are opened with the All shutters open menu item.

Í If you switch on with the All X-ray generators menu item you receive the values for the anode
current, the high voltage and the heater current of all X-ray generators with high voltage
switched on.

Í If you switch off the Update display function you freeze the measured values. At the same time the
X-ray generator and the conveyor are also switched off.

95587956 13/08/09 © Smiths Heimann 13


Diagnosis Programs

Calibration slider test


The calibration slider test enables the speed with which the relevant calibration sliders move into the beam
path to be adjusted.

The calibration slider test is started with the Start test and conveyors function. In this case the conveyor is
started, the shutters opened and the calibration sliders moved cyclically into the beam path. In this case the
views for the calibration samples shown adjacently arise.

In this test the time between the output of the START_CALI signal and reaching the end position of the
calibration slides is shown in percent. The values are determined anew each time the calibration slides move
in.

A positive sign of the percent display indicates that the inward movement speed is too high.

A negative sign of the percent display indicates that the inward movement speed is too low.

The adjustment of the shutter speed is satisfactory if the displayed values are within a range of + 5%.

The throttle adjustments of the shutters influence one another mutually, so that the adjustments
may have to be repeated.
Changes of the system pressure also lead to fluctuating results.

If a problem occurs with regard to the calibration sliders then the current X-ray image of the calibration sliders
can be saved on a storage medium indicated. This function will also be available if the X-ray unit is in the
normal operating mode. In such case you have to log in as „Service“ and then call up „ Calibration slider
test“ from the Diagnosis menu. Then you can store the X-ray image of the calibration sliders.

On the following page you find an example showing a perfect calibration slider image.

14 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Calibration slider of an EDtS right version Calibration slider of an EDtS


left version

95587956 13/08/09 © Smiths Heimann 15


Diagnosis Programs

I/O test
The I/O test enables the input data bits sent from the PI interface board to the HiTraX electronics to be
checked and the output data bits sent from the HiTraX electronics to the PI interface board to be controlled.

Í If you exit the I/O test with after you have changed output data bits, the status set by you is
retained.

Í You receive the original status of all data bits if you select the I/O test again and exit with .

The PI-internal inputs or outputs with the standard signal assignments are listed in the following tables. Refer
to the project-specific circuit diagrams for the exact signal assignment of the inputs and outputs.

16 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Inputs Signal PI connection Display 0 means:


plug
00 LS1 X23: 1,2,3 The light barrier is free.
01 LS3 X24: 4,5,6 The light barrier is free.
02 LS2.1 X23: 4,5,6 The light barrier is free.
03 LS2.2 X24: 1,2,3 The light barrier is free.
04 PM_OK X25: 4,5 Phasenmonitor zeigt keinen
Fehler.
05 LS ERROR X22: 1,2,3 There are no errors of the light
barriers.
06 LS5 X26: 1,2,3 The light barrier is free.
SV Position of the service switch on the The service switch is not in the
card PI service position.
08 SERVICE_MODE X37: 3,4 The input contact is opened.
09 NO_SCAN X47: 1 NO_SCAN is not selected by the
BHS.
10 INTERN
11 X-RAY_ON2E X62: 4 Acknowledgment 2 of generator
E.
Generator does not radiate.
12 CONV_ON X47: 2 CONV_ON is not selected by the
BHS.
13 INTERN
14 X-RAY_ON E X62: 6 Acknowledgment from generator
E.
Generator does not radiate.
15 INTERN
16 REG-END-E X62: 7 REC-END-E signal is not present.
17 LCM-ERROR There is no lamp current error.
18 RESERVE X47: 6
19 RESERVE X47: 7
20 RESERVE X47: 8
21 EMERGENCY_STOP X32: 3,4 There is no emergency stop.
22 INTERLOCK X31: 1,2 All interlock switches are closed.
23 LS0 X25: 1,2,3 The light barrier is free.
24 RESERVE X46: 1
25 INTERN
26 24 Volt Error There is no 24V error.
27 RESERVE X46: 2
28 RESERVE X46: 3
29 RESERVE X46: 4
30 READY_FOR_CAL X47: 3 No READY_FOR_CAL signal is
present.
31 INTERN
32 CONV-GO (Control unit) X19: 1,2 There is no CONV-GO signal from
the manual control lever.

95587956 13/08/09 © Smiths Heimann 17


Diagnosis Programs

Inputs Signal PI connection Display 0 means:


plug
33 NO-AUTO-STOP (Control unit) X 20: 1,2 There is no NO-AUTO-STOP
signal from the manual control
lever.
34 Shutter_AB_CLOSED X3: 1,2,3 The shutter is open or there is no
acknowledgment.
35 Shutter_CD_CLOSED X3: 4,5,6 The shutter is open or there is no
acknowledgment.
36 Shutter_E_CLOSED X2: 1,2,3 The shutter is open or there is no
acknowledgment.
37 OVP_OK X18: 3,4 The overvoltage protection has
not tripped.
38 REPAIR_SWITCH X37: 1,2 The auxiliary contact of the main
switch is closed.
39 CONV_STOP (Control unit) X18: 1,2 There is no CONV_STOP signal
from the manual control lever.
40 RESERVE X46: 5
41 RESERVE X46: 6
42 RESERVE X46: 7
43 X-RAY_ON2AB X60: 4 Acknowledgment 2 of generator
AB. The generator does not
radiate.
44 RESERVE X46: 8
45 PRESSURE_OK X26: 4,5,6 The system pressure is sufficient.
46 X-RAY_ON AB X60: 6 Acknowledgment from generator
AB.
The generator does not radiate.
47 X-RAY_ON CD X61: 4 Acknowledgment from generator
CD.
The generator does not radiate.
48 REG-END-AB X60: 7 REC-END-AB signal is not
present.
49 EXT_CONV_ERROR X29: 1,2,3,4
50 X-RAY _ON2CD X61: 6 Acknowledgment 2 from
generator CD. The generator does
not radiate.
51 POWER_GOOD X29: 5,6 The POWER_GOOD signal is
available.
52 REG-END-CD X61: 7 REC-END-CD signal is not
present.
53 IL_IO_CONVEYOR X28: 1,2,3,4 The interlock switches of the pre-
tunnel are closed.
54 FREQUENCY_OK X19: 3,4 The output frequency of the
frequency converter is not yet
reached or the frequency
converter is switched off.
55 FU_ERROR X20: 3,4 The frequency converter has an
internal error.

18 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Inputs Signal PI connection Display 0 means:


plug
56 IL_UNIT X28: 5,6,7,8 The interlock switches of the core
system are closed.
57 RESERVE IN4 X27: 1,2,3,4
58 NO_SCAN_MS X35: 3,4 The NO_SCAN signal from the
operating console is not present.
59 TRANSIT X32: 1,2 The TRANSIT signal from the
operating console is not present.
60 REMOTE_CON X36: 1,2 The REMOTE_CON signal from
the operating console is not
present.
61 POW_ON X47: 5 The POW_ON signal from the
operating console is not present.
62 XRAY_SENSOR X30: 5,6,7,8 The input is open.

95587956 13/08/09 © Smiths Heimann 19


Diagnosis Programs

Out- Signal PI connection Input 0 leads to:


puts plug
00 GEON_E_ON_1 X62: 2 The generator E is switched on together
with the GEON_E_ON_2 signal.
01 GEON_E_ON_2 X62: 3 1->0
02 AP2_E X62: 5 The XRC is switched over into the working
point 2.
03 LS2.1_OUT X48: 3 Output of the signal on the VT-IF.
04 LS2.2_OUT X48: 4 Output of the signal on the VT-IF.
05 LS3_OUT X49: 3 Output of the signal on the VT-IF.
06 LS4_OUT X48: 5 Output of the signal on the VT-IF.
07 LS5_OUT X48: 6 Output of the signal on the VT-IF.
08 NEXT_BAG_OUT X48: 1 Output of the signal on the VT-IF.
09 LS_SCALE X21: 5,6,7,8
10 LS_TEST X21: 1,2,3,4 The discrete light barriers of the type
LBT/LBR are switched into the test mode.
11 BFOR X16: 2,3,4 The main conveyor is switched on in
forwards direction.
12 BREV X16: 1,2,4 The main conveyor is switched on in
reverse direction.
13 P3_BAGGAGE_COUNTER X13: 1,2,3 Output of a counting pulse.
14 P2_XRAY_TIME X14: 1,2,3 Output of a counting pulse.
15 P1_RUN_TIME X14: 4,5,6 Output of a counting pulse.
16 NON_AUTO_STOP X11: 1,2 Switching over to AUTO_STOP in the
stand-alone mode.
17 P4_MARKING_COUNTER X13: 4,5,6 Output of a counting pulse.
18 RESERVE X48: 8
19 ES_RESET_ACTIVE X49: 5 Pulse: Output of the signal on the VT-IF.
20 SHUTTER OPEN X8 : 3,4 All shutters are open.
21 OVERSIZE_1 X49: 6 Output of the signal on the VT-IF.
22 OVERSIZE_2 X49: 7 Output of the signal on the VT-IF.
23 RESERVE X49: 8
24 RESERVE_OUT1 X41: 1,2,3,4
25 RESERVE_OUT2 X41: 5,6,7,8
26 PC_ON X59: 6
27 RESERVE_OUT8 X44: 5,6,7,8
28 RESERVE_OUT9 X39: 1,2
29 RESERVE_OUT10 X39: 3,4
30 NO_SCAN_ACTIVE X49: 2 Activation of the NO_SCAN mode.
31 INTERN
32 GEON_AB_ON_1 X60: 2 The generator AB is switched on together
with the GEON_AB_ON_2 signal.
33 GEON_AB_ON_2 X60: 3 1->0
34 AP2_AB X60: 5 The XRC is switched over into the working
point 2.
35 RESERVE_OUT3 X42: 1,2,3,4

20 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Out- Signal PI connection Input 0 leads to:


puts plug
36 ES_RESET_PI X40. 1,2,3,4 ES_RESET from the PI to reset the
emergency stop relay.
37 GEON_CD_ON_1 X61: 2 The generator CD is switched on together
with the GEON_CD_ON_2 signal.
38 GEON_CD_ON_2 X61: 3 1->0
39 AP2_CD X61: 5 The XRC is switched over into the working
point 2.
40 MODE_A2 X60,61,62: 14 Inquiry of the AD converter on the XRC for
inquiring the operating data. The signal is
constantly toggled in the normal case.
41 MACHINE_OK X50. 1 Output of the signal on the VT-IF.
42 CODE_REQ X50: 2 Output of the signal on the VT-IF.
43 START_CALI X8: 1,2 Output of the signal on the VT-IF.
44 MAINS_FAILURE X12: 1,2,3 Output of the signal on the server and the
analysis stations if present.
45 BHS_ERROR X48: 7 Output of the signal on the VT-IF.
46 MODE_A0 X60,61,62: 12 Inquiry of the AD converter on the XRC for
inquiring the operating data. The signal is
constantly toggled in the normal case.
47 MODE_A1 X60,61,62: 13 Inquiry of the AD converter on the XRC for
inquiring the operating data. The signal is
constantly toggled in the normal case.
48 XRAY_READY X50: 3 Output of the signal on the VT-IF.
49 SPEED_OK X50: 4 Output of the signal on the VT-IF.
50 REG_START Output of the signal on the XRC (no
function).
51 RESERVE_OUT5 X43: 1,2,3,4
52 CAL_REQ X50: 5 Output of the signal on the VT-IF.
53 IN_CONV_FOR X17: 6,7,8 Start of the inlet conveyor in forwards
direction.
54 OUT_CONV_FOR X17: 2,3,4 Start of the outlet conveyor in forwards
direction.
55 UPS_STOP Output of the signal at the UPS.
56 RESERVE_OUT4 X42: 5,6,7,8
57 LS1_OUT X48: 2 Output of the signal on the VT-IF.
58 RESERVE_OUT6 X43: 5,6,7,8
59 SERVICE_MODE X50: 8 Output of the signal on the VT-IF.
60 IN_CONV_REV X17: 5,6,8 Start of the inlet conveyor in reverse
direction.
61 OUT_CONV_REV X17: 1,2,4 Start of the outlet conveyor in reverse
direction.
62 RESERVE_OUT7 X44: 1,2,3,4
63 RESERVE OUT11 X38: 3,4

95587956 13/08/09 © Smiths Heimann 21


Diagnosis Programs

Light barriers

It is possible in the Light barriers diagnosis menu to perform a static test of the discrete light barriers or also
to select the test of the light curtain which indicates simultaneously a dynamic test of the light barriers.

Test of discrete light barriers

Normally light barriers of the type LBT/LBR are built into the X-ray inspection unit. Their transmitters are
switched on and off during the test through the test input to test the correct switching behavior of the
receivers.

The display has the following meaning for the LBT/LBR light barriers:

Free: The light barrier is in order and not interrupted.


Interrupted: The receiver has a constant output potential log0, i.e. the light barrier is interrupted or the
transmitter or the receiver is defective or the connection has a short to ground.
Error: The receiver has a constant output potential log1, even if the transmitter is switched off, i.e.
the receiver is defective or the connection is open.
not existing: The light barrier is not installed.

If other light barriers which cannot be tested dynamically are installed, the display has the following
meaning:

Free: The light barrier output is at log1 potential.


Interrupted: The light barrier output is at log0 potential.

22 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Test of light curtain

The conveyor must be switched on to test the light curtain. For starting use the switches
and or the corresponding keys on the keyboard.

If a bag is now placed on the conveyor, all signals of the light barriers and of the light curtain are displayed on
the display. The height information of the bag placed on the conveyor is shown in the upper display.
On the other hand, the static light barriers show their status with their high level, whether they are
interrupted or not.

This dynamic test offers the possibility of displaying dynamic errors of the static light barriers.

95587956 13/08/09 © Smiths Heimann 23


Diagnosis Programs

Power supply and temperature


You can call up information about the different supply voltages of the PI interface board, temperature
information on the PI and system information of the UPS in this submenu.

Power supply test

Here you receive all information about the different supply voltages on the PI. The information about the
existence of these voltages can also be checked on the control LEDs of the PI.

The display accuracy is


±3%.

Temperature test

The values of the PI-internal sensor, or of an optionally connected external sensor are queried in this test.

24 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Test of UPS unit

This test queries all status information of the used UPS.


If the displayed values deviate strongly from the manufacturer’s requirements on the operational readiness of
the UPS, then the intended maintenance work, such as battery changing, must be performed as quickly as
possible.

With a call of button „1", Set battery exchange date, you take over the current system date as the new
battery exchange date.

If there is no status information of the UPS available check the function of the serial interface of the
UPS by means of the manufacturer’s diagnostic software.

95587956 13/08/09 © Smiths Heimann 25


Diagnosis Programs

Data transfer test


With its different menu items the data transfer test offers the possibility of checking the communication of
the individual connected systems among one another.

Ping an address

In this submenu a ping is sent by the acquisition controller to the subscriber with the preselected IP address.

The test is started or interrupted with the START/STOP function. As soon as the test is stopped, a summary
of the transmission and reception protocols is displayed in the display.
This test provides information about the physical availability of a data network and of the addressed
subscriber.

The input mask of the IP address is selected with the IP address function. Here the IP address to which a
ping should be sent must be set.

26 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Transfer rates

This test offers the possibility of checking the transfer rates in the individual network segments. Bottlenecks
in data transfer can thus be shown.
If the test is stopped, a summary of the data transfer in the relevant network segment is displayed.

Ethernet 1: network between Acquisition controller and Eval PCs.


Ethernet 0: network to MatriX Server.

Interface configuration

The interface configuration offers an overview of the individual network segments.

95587956 13/08/09 © Smiths Heimann 27


Diagnosis Programs

Routing table

The routing table contains the information of the individual network segments.

Image transmission test

If the transmission test is switched on, then the images stored in the acquisition controller are transferred to
the relevant selected directory on an available evaluation station.
In the image transmission interval the time in which further images should be transferred after reception of
an evaluation result of the image evaluation station is preselected.
An unmistakable identification of the X-ray images in the status display of the evaluation station can be set
with the bag ID prefix.

The image transmission is activated only if the main menu is exited.


No real bag can be X-rayed during the image transmission test!

28 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Test of material detection


The recorded image of the test case that still has to be produced is evaluated with the test of material
detection.

Passing this test is the major acceptance criterion for the X-ray unit!
The adjustment suitcase is the tool destined to check the adjustment and geometry data of the unit.

The program returns to the overview image and states in the set frame the determined detection probability
and the density of the determined material.

For the test of the material detection you can use adjustment suitcase 34436851 or 34432743. Both
adjustment suitcases are suited to check the geometry data of the HS 10080 EDt S. The basic difference is
the different positioning of the adjustment suitcase on the belt. Due to the different orientation of the
material probes, you will get two different X-ray images.

In the following paragraph you find a description of the different bag contents and the orientation of the
material probes. Evaluation and troubleshooting are the same for both test suitcases.

95587956 13/08/09 © Smiths Heimann 29


Diagnosis Programs

Functional test by means of the adjustment suitcase (34436851)

The suitcase contains four foam layers.

Í The first layer is used to accommodate the test materials.

Í The second foam layer accommodates 2 vertical aluminium plates


and one vertical iron plate.

Í The third foam layer contains 2 Delrin balls (orange mark) and 3
aluminium balls (green mark). There are also 3 small steel balls
(yellow mark) in this layer.

Í The fourth foam layer contains 2 horizontally placed aluminium


Four of the six steel balls are situated on the back of the foam layer.

Start the unit and wait for the material calibration to be carried out.

Í Put the suitcase on the input belt (in the middle) so that TOP can be
read and the suitcase's handle is
directed to the system monitor!

Í Start the scanning of the calibration suitcase in service mode.

Í When the suitcase has left the tunnel, stop the scanning and call up
the function Test of material detection in the Diagnosis menu.

Í For the test the baggage item scanned last is used; the test results are displayed on the unit's monitor.

30 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Functional test by means of the adjustment suitcase (34432743)

By means of the adjustment suitcase the technical system functions are checked.
It is a tool suitable for checking the adjustment and geometry of the system, i.e. you can only determine, by
means of the adjustment suitcase, if the unit has been properly adjusted.

The suitcase contains three foam layers.


The first layer is used to accommodate the test materials.
The second foam layer accommodates 3 Delrin balls and 3
aluminium balls.
The third layer contains 2 small steel plates, 2 small aluminium
plates, one big aluminium plate and nine lead balls (see yellow
mark in the picture beside).

The second layer is placed into the suitcase in such a way that
one Delrin ball and one aluminium ball lie over the steel plates.
No ball must lie on top of the big aluminium plate!

Í Start the unit and wait for the material calibration being carried out.

Í Put the suitcase on the input belt (in the middle) so that the big aluminium plate enters the tunnel first.

Í Start the scanning of the calibration suitcase in service mode.

Í When the suitcase has left the tunnel, stop the scanning and call the function Test of material detection in
the Diagnosis menu.

Í For the test the baggage item scanned last is used; the test results are displayed on the unit's monitor.

95587956 13/08/09 © Smiths Heimann 31


Diagnosis Programs

Evaluation

Adjustment and geometry data ok

In all views the big balls are marked by a red frame and at least 5 small balls are marked by a blue frame.

Adjustment suitcase (34436851)

Adjustment suitcase (34432743)

In system versions ES-03-00-G, ES-03-00-K, ES-03-00-O the small balls are not detected.

The percentages indicated on the evaluated image serve the purpose of internal statistics.
In the Diagnosis report the graphical data of the test suitcase's evaluation is stored as numerical information.
This data serves to archive the evaluation results.

32 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Errors

If not all materials are detected and marked when the test suitcase is scanned for the first time, you have to
have the suitcase scanned once more. If not all marks are set after the second scanning either, you have to
take the following into account:

No frames in any of the views

Possible cause: Wrong adjustment of lines

Correction: Go through all the following points as described in the technical manual:

Í Set lines to maximum by means of the linescope.


Í Calibration of detector lines (offset and range)
Í Geometry calibration by means of calibration sheet
Í Geometry calibration by means of calibration body. Position the calibration body in such a way that it is
located on center of the conveyor when entering the basic unit!
Í Wait for the complete material calibration being completed before scanning the test suitcase once more.

Not each of the six balls is marked

Possible cause: Wrong adjustment of lines


Wrong position of calibration body.
Correction: Go through all the following points as described in the technical manual:

Í Set lines to maximum by means of the linescope.


Í Calibration of detector lines (offset and range)
Í Geometry calibration by means of calibration sheet
Í Geometry calibration by means of calibration body. Position the calibration body in such a way that it is
located on center of the conveyor when entering the basic unit!
Í Wait for the complete material calibration being completed before scanning the test suitcase once more.

If not each of the six balls is marked by a frame when repeating the test, you have to repeat the
scanning 10 times. If all balls are detected in at least 5 attempts, then the unit is correctly adjusted.

95587956 13/08/09 © Smiths Heimann 33


Diagnosis Programs

Not at least 5 out of 9 small balls are marked by frames

Possible cause: Wrong adjustment of lines


Wrong position of calibration body.

Correction: Go through all the following points as described in the technical manual:

Í Geometry calibration by means of calibration sheet


Í Geometry calibration by means of calibration body. Position the calibration body in such a way that it is
located on center of the conveyor when entering the basic unit!
Í Wait for the complete material calibration being completed before scanning the test suitcase once more.

If, again, not at least 5 small balls are marked by frames, you have to repeat the scanning of the
test suitcase 10 times. If more than 5 small balls are detected in at least 5 attempts, then the unit
has been correctly adjusted.

TEST ERROR (Timeout)

Possible cause: The evaluation of the test suitcase is impossible because a


timeout has occurred.

Correction: Place the test suitcase out of the belt middle and start the
evaluation once more.
Switch off the communication with the HMC during this test!

TEST ERROR (out of memory)

Possible cause: Instead of the provided test suitcase with the specified
content, a different suitcase was scanned.

Correction: Use the orignal Smiths Heimann test suitcase (material


number: 34432743 or 34436851). Carry out the test once
more.

34 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Indicator test
The X-RAY ON indicators of the X-ray inspection unit can be checked with the indicator test.

Conveyor test

With the conveyor test you can operate the conveyor, e.g. for its adjustment, without X-ray radiation having to
be switched on. For starting use the switches and or the
corresponding keys on the keyboard.

After starting you receive an indication of the nominal conveying speed, i.e. the speed at which the
conveyor must run so that the software can correctly compute the position of bags in the normal operating
sequence, activate the X-ray generator on time and synchronize image processing.

95587956 13/08/09 © Smiths Heimann 35


Diagnosis Programs

Continuous test
The continuous test enables you to automatically test the X-ray unit over a longer period.

Í After you have started the test with Start test, the conveyors and the X-ray generators are operated. In
intervals the shutter is opened and the calibration body pushed into the beam path.

If error messages occur during the continuous test, these are acknowledged automatically and counted in the
Error field. If fatal errors occur, then the test is interrupted.
It is also possible to stop the test in the case of simple errors.

The error messages can be seen in the Error report in the Diagnosis menu.

36 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Configuration display
The configuration display provides you with an overview of the current configuration of the system.

Unit type: Set type, right or left version of the EDt S

Variant: Project variant selected.

Preset software
configuration: Default configuration for EDtS units

Serial number: Serial number of the current unit. This number is required in the case of
inquiries to Smiths Heimann.

Software version: Currently available software version.

Detection characteristic: Currently available detection characteristic.

PI4 server version: Currently available software version of the PI4 server.

Detection characteristic Active detection characteristic

Available options: Available, released options in the dongle used.

Free IMS memory: Available storage capacity for recording X-ray images.

Unit network name: Name for identification of the X-ray inspection unit in the Smiths Heimann
network.

95587956 13/08/09 © Smiths Heimann 37


Diagnosis Programs

Host address: IP address of the X-ray inspection unit.

Net mask/ gate address: Set network parameters

Dongle ID: Identification of the dongle used.

38 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

Reports
Different reports can be selected in the reports menu item.

Error report

All occurring errors are stored in the error report with date, time and error code. The user can scroll through
the file using the navigation keys.
The error report can also be used for remote diagnosis through the network connection.
If the error report is deleted with the key “1", then the corresponding report will be ended and a new error
report is generated.

Login report

You can see in the login report who was registered at which time for the unit and when login procedures
have been interrupted because of wrong IDs or passwords.

95587956 13/08/09 © Smiths Heimann 39


Diagnosis Programs

Diagnosis report

The diagnosis report is created with each selection of this menu. This can take some time. The newly
generated file is filed analogously to the error report and is then ready for transmission.

When calling up the diagnosis report, the following message will be displayed:

During this time the conveyor belts and generators are switched on several times. The diagnosis being
terminated, the diagnosis report is created.

The diagnosis report contains all important system information which can also be retrieved via the submenus
of the diagnosis menu:

- Unit type / serial number


- Date / time of the unit
- Data of the configuration display
- Hardware settings
- Light barrier test results
- Keyboard test result
- Line interface test results
- Line test results
- Line scopes (with and without X-radiation)
- Operating voltage values
- X-ray generator measured values (with and without X-radiation)
- Nominal conveyor speeds
- Status of the baggage counter (software counter)
- Calibration data
- Geometry data
- Results of the „Tests of material detection"

40 95587956 13/08/09 © Smiths Heimann


Diagnosis Programs

PI 4 Ethernet module report

If this report is called up, a list will be generated showing who has logged in on the Acquisition Controller via
the network.

95587956 13/08/09 © Smiths Heimann 41


Diagnosis Programs

42 95587956 13/08/09 © Smiths Heimann

You might also like