Oxylog

Emergency Ventilator
Technical
Documentation
Copyright by Dräger Medical AG & Co. KGaA, Lübeck, Federal Republic of Germany.

Copying is prohibited for commercial purposes.

Attention is to be paid to the Operating Manual.

This technical documentation does not replace the Operating Manual.

The warranty and liability conditions of the general terms and conditions for business transac-
tions of Dräger Medical AG & Co. KGaA are not extended by this technical documentation.

Safety Regulations:
Reference is hereby made to the observance of the relevant safety provisions, such as in Ger-
many the Medical Equipment Ordinance (Medizingeräteverordnung), the Pressure Container
Ordinance (Druckbehälterverordnung), the Technical Rules for Pressurised Gases (Techni-
sche Regeln Druckgase) or the Occupational Health and Safety Provisions (Unfallverhü-
tungsvorschriften).

Insofar as reference is made to laws, regulations or standards, these are based on the legal
system of the Federal Republic of Germany.

Follow your local laws and regulations.


Contents

General

1 About This Service Manual 4

1.1 Definitions .........................................................................................................................5

2 For Your Safety and that of Your Patients 7

2.1 Strictly follows the Instructions for Use ......................................................................7

2.2 Maintenance .....................................................................................................................7

2.3 Accessories ......................................................................................................................7

2.4 Liability for proper function or damage .......................................................................7

3 Intended use 8

3.1 Oxylog with spontaneous breathing attachment (option) .......................................8

4 Inspection intervals 9

5 Checking readiness for use 10

5.1 Checking ventilation frequency ................................................................................. 10

5.2 Checking safety valve .................................................................................................. 10

5.3 Checking minute volume and compressed-gas supply .........................................11

6 Technical data 12

6.1 Materials used ................................................................................................................14

6.2 Dimensions and weights ..............................................................................................14

6.3 Oxylog with spontaneous breathing attachment ....................................................14


For internal use only. Copyright reserved.

7 Cleaning/disinfection/sterilisation 15

7.1 Cleaning ......................................................................................................................... 15

7.2 Disinfection .................................................................................................................... 16


GBK5503140IECIVZ.fm

7.3 Sterilisation .....................................................................................................................17

5503.140 Oxylog 04.97 Contents Page I


Contents

8 Prior to renewed use with patient 17

Functional Description

1 Gas supply 18

2 Pneumatic control 19

2.1 Oxylog ............................................................................................................................. 19

2.2 Oxylog Samsomatik ..................................................................................................... 20

Test List

1 Test Equipment 29

2 Checking general condition 30

3 Checking O2 pressure reducer 30

4 Functional test 31

4.1 Checking I:E ratio ..........................................................................................................31

4.2 Checking frequency setting ........................................................................................31

4.3 Checking O2 concentration ....................................................................................... 32

4.4 Checking minute-volume flow setting ...................................................................... 33

4.5 Checking safety valve .................................................................................................. 34

4.6 Checking ventilation pressure gauge ...................................................................... 34


For internal use only. Copyright reserved.

5 Put functional device at the customer’s disposal. 35

6 Test performed 35
GBK5503140IECIVZ.fm

5503.140 Oxylog 04.97 Contents Page II


Contents

Fault-Cause-Remedy

1 Fault - Cause - Remedy 36

Changes

1 Type of Changes 37

Appendix

1 Abbreviations 38

2 Spare parts list Oxylog / Oxylog Samsomatic 39


For internal use only. Copyright reserved.
GBK5503140IECIVZ.fm

5503.140 Oxylog 04.97 Contents Page III


General

1 About This Service Manual

This service manual conforms with the International Standard IEC 601-1.

This service manual does not replace the Instruction for Use.

Read through each step in every procedure thoroughly before beginning any test. Always use
the proper tools and specified test equipment. If you deviate from the instructions and/or
recommendations in this service manual, you may cause the equipment to operate improperly
or unsafely, or damage the equipment itself.

Use only genuine spare parts manufactured or sold by Dräger Medical AG & Co. KGaA as
listed in the spare parts list.

The test list of this service manual does not replace inspections and servicing by
Dräger Medical AG & Co. KGaA.
For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 4


1.1 Definitions

DANGER:

DANGER indicates an imminently hazardous situation which, if not


avoided, will result in death or serious injury.

WARNING:

WARNING indicates a potentially hazardous situation which, if not


avoided, could result in death or serious injury.

CAUTION:

CAUTION indicates a hazardous situation which, if not avoided, may result


in minor or moderate injury. CAUTION is also used to alert against unsafe
practices.

J
NOTICE:

NOTICE is used to provide information that can help avoid damaging the
product or property as a result of misuse.

IMPORTANT:

IMPORTANT is used to provide additional information, operating tips or


maintenance suggestions.
For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 5


Inspection = examination of actual condition
Service = measures to maintain specified condition
Repair = measures to restore specified condition
Maintenance = inspection, service, and repair, where necessary
For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 6


2 For Your Safety and that of Your Patients

2.1 Strictly follows the Instructions for Use

Any use of the apparatus requires full understanding and strict observation of these
instructions. The apparatus is only to be used for purposes specified here.

2.2 Maintenance

The apparatus must be inspected and serviced regulary by trained service personnel at 2
years intervals (and a record kept).

Repair and general overhaul of the apparatus may only be carried out by trained service
personnel. We recommend that a service contract be obtained with DrägerService® and
that all repairs also be carried out by them. Only authentic Dräger spare parts may be used
for maintenance.

Observe chapter “Maintenance Intervals”.

2.3 Accessories

Do not use accessory parts other than those in the order list.

2.4 Liability for proper function or damage

The liability for the proper function of the apparatus is irrevocably transferred to the owner or
operator to the extent that the apparatus is serviced or repaired by personnel not employed
or authorized by DrägerService® or if the apparatus is used in a manner not conforming to
its intended use.

Dräger Medical AG & Co. KGaA cannot be held responsible for damage caused by non-
compliance with the recommendations given above. The warranty and liabililty provisions of
the terms of sale and delivery of Dräger Medical AG & Co. KGaA are likewise not modified by
the recommendations given above.

Dräger Medical AG & Co. KGaA


For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 7


3 Intended use

The Oxylog is a ventilator for the controlled, time-cycled ventilation with constant volume of
infants as of a body weight of 5 kg and adults.

The unit is envisaged for mobile use in rescue operations, for transportation to a hospital in
rescue vehicles or by helicopter, for moving patients to a different hospital by road or air, for
ventilation of emergency admissions and for in-hospital transfer of patients receiving
ventilation, e.g. from operating theatre to intensive-care ward.

3.1 Oxylog with spontaneous breathing attachment (option)

Oxylog with spontaneous breathing attachment for time-cycled, constant-volume ventilation


and spontaneous breathing in a toxic atmosphere.

The Oxylog with spontaneous breathing attachment consists of the following:


• Oxylog with no Air Mix switch with demand regulator connected in parallel
• Ventilation valve with expiratory non-return valve
For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 8


4 Inspection intervals

IMPORTANT:

Before being serviced or sent in for repair, the unit (or the appropriate
components) must be cleaned and disinfected in line with hospital
regulations and the instructions given in this documentation.

Inspection intervals (Oxylog)

Inspection Every 6 months by specially trained personnel

Inspection intervals (Oxylog Samsomatik)

Inspection Every 2 years by specially trained personnel


For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 9


5 Checking readiness for use

• Readiness for use is always to be checked after unit upkeep. The unit must be fully
assembled before being tested.
• Performance of the tests is to be recorded in the logbook.

5.1 Checking ventilation frequency

• Make the following settings on the Oxylog:

Pneum. main switch I (on)


MV 3 L/min
Ventilation frequency 15 min-1
Switch No Air Mix

• Seal off ventilation valve at patient connection.


• Use stopwatch to measure time t for 10 full cycles.
• Determine ventilation frequency f as follows:

f = –––––– min-1

The Oxylog should switch at a ventilation frequency of between 13 and 17 min.-1.

5.2 Checking safety valve

• Retain the same unit settings.


• Seal off ventilation valve at patient connection.

The max. ventilation pressure should be 50 - 80 mbar.


For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 10


5.3 Checking minute volume and compressed-gas supply

• Insert catheter connector (size 5) in patient connection of ventilation valve.


• Use the following table to check the max. inspiration pressure on the airway pressure
gauge.

MV Airway pressure
7 L/min 4 - 8 mbar
15 L/min 15 - 24 mbar
20 L/min 28 - 38 mbar

The Oxylog should switch at regular intervals from inspiration to expiration.

The setting MV = 20 L/min. provides an indirect check on proper operating conditions (min.
2 bar with 60 L/min. on Oxylog).
• Remove catheter connector (size 5) from patient connection.

The unit is ready for operation.

60
t/10
For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 11


6 Technical data

Operating principle Flow chopper


Control Time-cycled, volume-constant
Ventilation frequency 10 - 35 min-1 ±20%, infinitely variable
I:E ratio 1:1,5 ±10%
Minute volume (MV) 2 - 20 L/min ±15%, infinitely variable

O2 concentration of ventilation gas using O2 as drive:

Switch set to Air Mix 55 Vol.% O2 ±10% (with MV > 7 L/min), with
MV < 7 L/min, O2 concentration increases
up to 80 vol.%
Switch set to No Air Mix 100 vol.% O2

Safety valve:

Opening pressure 50 mbar to 80 mbar


Airway pressure reading Pressure gauge -10 to +80 mbar ±2,5% of
full scale value

Drive gas O2 or air


Quality Dry, oil/dust-free, from central supply or
compressed-gas cylinders
Pressure at device inlet min. 2, max. 6 bar with a flow rate
of 60 L/min

Gas consumption:

Control approx. 0,8 L/min


MV (Air Mix) approx. 50% of set MV
MV (No Air Mix) approx. 100% of set MV
For internal use only. Copyright reserved.

Typical operating time With 3L cylinder/200 bar, MV = 10 L/min


(No Air Mix) 54 min
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 12


Pneumatic main switch I−0
Patient system Comprises 1.1 m silicone hose and
ventilation valve
Compressible volume approx. 0,3 mL/mbar
Inspiration resistance 3 mbar/L/s
Expiration resistance 3 mbar/L/s
Dead-space volume approx. 12 mL

Ambient conditions - operation:

Temperature −5 °C to +50 °C
Humidity 0 to 100% relative humidity
Ambient pressure 600 to 1200 mbar
Vibration Tested as per MIL STD 810 C 514.2-III, curve
M (helicopter)
In toxic atmosphere Switch set to No Air Mix, controlled ventilation
in the case of apnoea.

CAUTION:

Danger of poisoning! In the event of ongoing or resumed spontaneous


breathing, toxic air is drawn in with the Oxylog, which can result in
poisoning. For ventilation in a toxic atmosphere, use is to be made of the
special Oxylog with spontaneous breathing attachment.

Ambient conditions - storage:

Temperature −20 °C to +70 °C


Humidity 0 to 100% relative humidity
Ambient pressure 600 to 1200 mbar
For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 13


6.1 Materials used

Oxylog housing Oxylog Impact-proof ABS


(acrylonitrile-butadiene-styrene)
Ventilation hose Silicone rubber

Ventilation valve:
Housing PSU (polysulfone)
Control diaphragm Silicone rubber

6.2 Dimensions and weights

Dimensions (W x H xD) 200 x 80 x 200 mm


Weight approx. 2 kg

6.3 Oxylog with spontaneous breathing attachment

Spontaneous breathing attachment:

Opening pressure Opening pressure 0 to −4 mbar


Max. output 100 L/min at −7 mbar

Oxylog:

No Air Mix switch Ambient air is not sucked in


Pressure limitation valve 50+30 mbar
Ventilation valve With additional expiratory non-return valve
For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 14


7 Cleaning/disinfection/sterilisation

• Always clean and disinfect Oxylog after ventilation.

7.1 Cleaning

• Strip down Oxylog.


• Disassemble ventilation valve into its component parts.

IMPORTANT:

Do not remove red non-return valve from yellow control valve.

• Clean disassembled components in warm running water or in water containing


detergent.
• Thoroughly rinse components in running water and then dry them off.

The component parts can also be washed in the Dräger Purfactor“, which ensures that the
material is disinfected and dried.
• Wipe over the surface of the unit with a soft cloth soaked in detergent.

J
NOTICE:

Material damage! Petroleum, ether or similar solvents may damage the


plastic housing of the unit. Never use petroleum, ether or the like for
cleaning purposes.
For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 15


7.2 Disinfection

7.2.1 Use of liquid disinfectant

• Disinfect the cleaned and dried components of the patient system in a cold disinfectant
solution.
• Thoroughly rinse the components in water and then dry them off.

7.2.2 Disinfection by wiping

• Wipe over the surface of the Oxylog with liquid disinfectant.

Dräger Medical AG & Co. KGaA recommends consulting the "List of officially tested and
approved disinfectants and disinfection methods" as well as the current list of disinfection
procedures checked in line with the "Guidelines for testing chemical disinfectants" and
deemed suitable by the German Hygiene and Microbiology Association.

7.2.3 Disinfection in Dräger Aseptor®

• Disinfect the cleaned and dried components of the patient system and the Oxylog in line
with the Dräger Aseptor“ instructions for use.

J
NOTICE:

Material damage! Parts made of silicone rubber may be damaged in the


Dräger Aseptor®. Never disinfect ventilation hose and masks in the Dräger
Aseptor®.
For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 16


7.3 Sterilisation

• Sterilise the following in superheated steam at 134°C:


− Cleaned and dried components of patient system
− Component parts of ventilation valve
− Ventilation hose
• Plastic and rubber mouldings are always to be disassembled for sterilisation purposes.

J
NOTICE:

Equipment damage! Moisture ingresses into the unit on sterilising it in an


autoclave or immersing it in a disinfection bath. This can lead to damage.
Never sterilise Oxylog in an autoclave or immerse it in disinfectant
solution.

8 Prior to renewed use with patient

• Assemble unit.
• Check that unit is ready for operation.
For internal use only. Copyright reserved.
GBK5503140A01.fm 19.04.99

5503.140 Oxylog 04.97 General Page 17


Functional Description

1 Gas supply

The gas is either supplied from a central gas supply system (CS) or from a gas cylinder with
pressure reducer.

Operation of the Oxylog ventilator requires a pressure of between 2 and 6 bar (O2 or com-
pressed air) with a flow rate of 60 L/min.

J
NOTICE:

Equipment damage! Dust, oil and moisture may impair the function of the
unit or damage it. Make exclusive use of dry medical gases which are free
from dust and oil.
Schutzvermerk DIN 34 beachten. Copyright reserved.
GBF5503140T01.fm 19.04.99

5503.140 Oxylog 04/97 Functional Description Seite 18


2 Pneumatic control

2.1 Oxylog

4 7 K

12
1 2 3
5 6 8 16

L 10
17 18
9

A 14 B C D
13
11

E 15 F G H

Abb. 1: Block diagram of Oxylog

1 Gas connection 15 Vessel


2 Filter 16 Intake
3 Pressure reducer 17 Metering unit
4 On/off switch 18 Ventilation valve
5 Distributor O A Amplifier relay
6 AMV adjuster B Amplifier relay
7 Air Mix/No Air Mix switch C Amplifier relay
Schutzvermerk DIN 34 beachten. Copyright reserved.

8 Metering unit D Amplifier relay


9 Injector E Amplifier relay
10 Relief valve F Amplifier relay
11 Vent valve G Amplifier relay
12 Pressure gauge H Amplifier relay
13 Frequency adjuster K Amplifier relay
GBF5503140T01.fm 19.04.99

14 Vessel L 3/2-way valve

5503.140 Oxylog 04/97 Functional Description Seite 19


2.2 Oxylog Samsomatik

Microswitches 1 - 6 are grouped on an interconnection plate (pneumatic plate).

The internal connections in the block obviate the need for interconnection of the individual
microswitches with hoses. The microswitches are held in position in the block by a clamp.
Microswitches 2 and 5 were specially selected for their switching properties (switching pres-
sure).

Microswitches

Clamp

2 3 6 5 1 4
Block

Abb. 2: Interconnection plate


Schutzvermerk DIN 34 beachten. Copyright reserved.
GBF5503140T01.fm 19.04.99

5503.140 Oxylog 04/97 Functional Description Seite 20


S1 S2

9
WV1 D1
F1 DR MV WV2 DM
A1

SV PA
BV
F2 I
SS

VD
1 2 3
V1
FD EV
D2

D3

V2
4 5 6

Abb. 3: Block diagram of Oxylog (Samsomatik)

A Gas connection I Injector


F1 Filter DM Pressure gauge
F2 Filter PA Patient outlet
DR Pressure regulator SS Silicone hose
S1 On/off switch BV Ventilation valve
S2 Air-Mix/No Air Mix switch EV Vent valve
MV Minute volume valve V1 Vessel 1
WV1 3/2-way valve V2 Vessel 2
Schutzvermerk DIN 34 beachten. Copyright reserved.

WV2 3/2-way valve 1 Microswitch


VD Pre-metering unit 2 Microswitch
D1 Metering unit 3 Microswitch
D2 Metering unit 4 Microswitch
D3 Metering unit 5 Microswitch
FD Frequency valve 6 Microswitch
GBF5503140T01.fm 19.04.99

SV Safety valve 9 Microswitch

5503.140 Oxylog 04/97 Functional Description Seite 21


2.2.1 Unit switched off

Oxygen or compressed air flows via the connection A through the filter F1 to the pressure
regulator DR, which provides a constant gas pressure of 1.5 bar. This pressure is applied to
the on/off switch S1, to the distributor O and to the 3/2-way valve WV1.

S1 S2

9
WV1 D1
F1 DR MV WV2 DM
A

SV
PA BV
I
F2 SS

VD
1 2 3
FD V1
EV
D2

D3

V2 5 6
4

Abb. 4: Gas supply (unit switched off)

2.2.2 No Air Mix switch setting

The distributor applies the gas to all points marked O. This switches the time cycling to expi-
ration, the 3/2-way valve WV1 is disabled and the 3/2-way valve WV2 is set to No Air Mix.
Schutzvermerk DIN 34 beachten. Copyright reserved.
GBF5503140T01.fm 19.04.99

5503.140 Oxylog 04/97 Functional Description Seite 22


2.2.3 Unit switched on

After switching on the switch S1, gas flows to the distributor O and from there through the fil-
ter F2, the pre-metering unit VD and the frequency valve FD to the time-cycling microswitches
1 and 4. At the same time, gas is routed to the microswitches 2 and 3.

S1

WV1
F1 DR MV
A

F2

VD

1 2 3
V1
FD

V2
4 5 6

Abb. 5: Gas supply (unit switched on)


Schutzvermerk DIN 34 beachten. Copyright reserved.
GBF5503140T01.fm 19.04.99

5503.140 Oxylog 04/97 Functional Description Seite 23


2.2.4 Inspiration

Microswitch 4 causes the vessel V2 to be filled as far as the switching point of microswitch 5.
Microswitch 5 is vented. Microswitch 6 is also vented and its output thus depressurised (expi-
ration signal blocked).

Microswitch 3 causes the inspiration signal to be passed to the 3/2-way valve WV1, to the
vent valve EV and to the microswitch 6. Vent valve EV is closed.

The 3/2-way valve WV1 opens. The gas applied flows through the minute volume valve MV,
the metering unit D1, the 3/2-way valve WV2 and the injector I to the patient.

S2

9
WV1 D1
MV WV2 DM

SV PA
BV
I
SS

2 3
V1
EV

5 6

Abb. 6: Inspiration
Schutzvermerk DIN 34 beachten. Copyright reserved.

The minute volume valve MV operates on the pressure regulator principle. The setting
governs the flow. The output capacity of injector I is the same in both switch settings (Air
Mix/No Air Mix).
GBF5503140T01.fm 19.04.99

5503.140 Oxylog 04/97 Functional Description Seite 24


2.2.5 Air Mix switch setting

In the Air Mix switch setting, gas flows through the switch S2 to the 3/2-way valve WV2 and
to the microswitch 9. Microswitch 9 switches to ventilation.

The 3/2-way valve WV2 switches over; the intake to injector I switches to ambient air.

S2

9
D1
DM

WV2 SV
PA

Abb. 7: Air Mix switch setting


Schutzvermerk DIN 34 beachten. Copyright reserved.
GBF5503140T01.fm 19.04.99

5503.140 Oxylog 04/97 Functional Description Seite 25


2.2.6 Expiration

As soon as the reversing pressure is attained at microswitch 2 (delayed by filling time of ves-
sel V1), microswitch 2 switches to venting and thus causes microswitch 3 to be vented as
well. The expiration signal is passed to the microswitches 3, 4 and 5 and at the same time to
the 3/2-way valve WV1. Microswitches 3, 4 and 5 switch over. The 3/2-way valve WV1 is
reversed and the flow of gas to the patient is blocked. The vent valve EV relieves the tubing
system (unit tubing from 3/2-way valve WV1 and patient hose to ventilation valve). The patient
can exhale to the atmosphere through the ventilation valve BV.

S1 S2

9
WV1 D1
F1 DR MV WV2 DM
A

SV PA
BV
I
F2
SS
VD
1 2 3
V1
FD
EV
D2

D3

V2
4 5 6

Abb. 8: Expiration

At the same time, microswitch 4 fills the vessel V2 as far as the switching point of micro-
Schutzvermerk DIN 34 beachten. Copyright reserved.

switch 5. Switching to inspiration then takes place.


GBF5503140T01.fm 19.04.99

5503.140 Oxylog 04/97 Functional Description Seite 26


2.2.7 Pressure limitation

To protect the patient, a max. ventilation pressure of 80 mbar is set on the safety valve SV.

DM
WV2

SV

PA
BV
I
SS

Abb. 9: Pressure limitation


Schutzvermerk DIN 34 beachten. Copyright reserved.
GBF5503140T01.fm 19.04.99

5503.140 Oxylog 04/97 Functional Description Seite 27


Test List

Oxylog
Folder no.:
5503
Serial no.: —————————
Revision:
Installation site: ————————— 11.95
Copyright reserved. Reproduction allowed for non-commercial purposes only.
GBK5503140TL1.fm 19.04.99

This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA

Mappennr Gerätename Ausgabe Test List Page 28


1 Test Equipment
Copyright reserved. Reproduction allowed for non-commercial purposes only.
GBK5503140TL1.fm 19.04.99

This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA

Mappennr Gerätename Ausgabe Test List Page 29


2 Checking general condition

• Check the following components. There must not be any signs of


damage:
− Housing, general
− O2-compressed-air connecting hose, 1.5 m
− O2-compressed-air connecting hose, 3 m
− CS O2-compressed-air connecting hose, 3 m
− CS O2-compressed-air connecting hose, 5 m
− Ventilation valve
− Diaphragm
− Ventilation hose
− Masks
− Carrying strap
− Oxylog baseplate (if applicable)
− Oxylog case (if applicable)
− Oxylog back pack (if applicable)

3 Checking O2 pressure reducer

• Check seal. There must not be any signs of damage.


• Renew strainer every 2 years.
Copyright reserved. Reproduction allowed for non-commercial purposes only.
GBK5503140TL1.fm 19.04.99

This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA

Mappennr Gerätename Ausgabe Test List Page 30


4 Functional test

• Carry out the following measurements with O2 at a supply pressure of


2 - 6 bar.

4.1 Checking I:E ratio

• Make the following Oxylog settings:

No
Air Mix
1
10 10

• Create the following test set-up:

Oxylog
pp==22bis
- 66bar
bar

• Use the following formula to check the I:E ratio (check value):
Inspiration time
Check value =
Expiration time
The check value must be between 0.6 and 0.74.
Copyright reserved. Reproduction allowed for non-commercial purposes only.

4.2 Checking frequency setting

• Make the following Oxylog settings:

No
Air Mix
1
15 10

The test set-up remains the same.


GBK5503140TL1.fm 19.04.99

The frequency must be between 13 and 17 min -1

This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA

Mappennr Gerätename Ausgabe Test List Page 31


4.3 Checking O2 concentration

• Set frequency on Oxylog to 10 min-1.


• Create the following test set-up:

Oxydig

Test
Oxylog adapter
p = 2 - 6 bar
p= 2 bis 6 bar

4.3.1 Checking O2 concentration with Air Mix

• Set minute volume of 7 L/min. on Oxylog.

Oxydig must display 55 ± 9 % by vol. O2.

4.3.2 Checking O2 concentration with No Air Mix

• Set minute volume of 20 L/min. on Oxylog.

Oxydig must display 100 −2 % by vol. O2.


Copyright reserved. Reproduction allowed for non-commercial purposes only.
GBK5503140TL1.fm 19.04.99

This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA

Mappennr Gerätename Ausgabe Test List Page 32


4.4 Checking minute-volume flow setting

• Set frequency of 10 min.-1 on Oxylog.


• Create the following test set-up:

R Flowmeter

Oxylog
p =22bis
p= - 66 bar
bar

• With a minute volume of 2.5 L/min., use variable restrictor R to set a


resistance of 5 ± 0.5 mbar (check on unit pressure gauge).
• Check minute volume with No Air Mix.

MV 2,5 (left stop) 5,5–7,8 L/min

MV 7 15,5–19,5 L/min

MV 15 33–42 L/min

MV 20 44–56 L/min

• Check minute volume with Air Mix.

MV 7 15–20 L/min
Copyright reserved. Reproduction allowed for non-commercial purposes only.

MV 15 32–43 L/min

MV 20 42,5–57,5 L/min
GBK5503140TL1.fm 19.04.99

This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA

Mappennr Gerätename Ausgabe Test List Page 33


4.5 Checking safety valve

• Make the following Oxylog settings:

No
Air Mix
1
10 var

• Create the following test set-up:

Ventilation valve
Pressure gauge

Oxylog
pp==22bis
- 66bar
bar

Bellows K
Bellows E

IMPORTANT:

Use bellows K (infants) for testing with MV of 3 L/min.


Use bellows E (adults) for testing with max. MV.
Copyright reserved. Reproduction allowed for non-commercial purposes only.

The safety valve must open at 55 +22/-3 mbar.

4.6 Checking ventilation pressure gauge

• Take reference measurement.

Tolerance: ± 2.5 % of full scale valu


GBK5503140TL1.fm 19.04.99

This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA

Mappennr Gerätename Ausgabe Test List Page 34


5 Put functional device at the customer’s disposal.

6 Test performed

Date: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Name: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
Copyright reserved. Reproduction allowed for non-commercial purposes only.
GBK5503140TL1.fm 19.04.99

This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA

Mappennr Gerätename Ausgabe Test List Page 35


Fault-Cause-Remedy

1 Fault - Cause - Remedy

The following table is designed to provide assistance with detection and rectification of
possible faults.

Fault Error Remedy


Unit does not build up No gas in cylinder. Immediately connect unit to
pressure. a fresh gas cylinder.
Insufficient gas pressure Establish adequate supply
at unit inlet. pressure: 2 - 6 bar.
Yellow control diaphragm Open up ventilation valve
in ventilation valve twisted and assemble correctly.
or deformed.
Unit remains set on inspi- Insufficient supply pres- Establish adequate supply
ration. sure. pressure: 2 - 6 bar
Fault in Oxylog. Call in DrägerService.
Patient cannot exhale or Ventilation hose kinked. Ensure proper hose rou-
only with difficulty. ting.
Red non-return valve in Open up ventilation valve
yellow control diaphragm and assemble correctly.
missing or crumpled.
For internal use only. Copyright reserved.
GBK5503140F01.fm 19.04.99

5503.^40 Oxylog 04.97 Fault-Cause-Remedy Page 36


Changes

1 Type of Changes

This technical documentation is valid for the technical equipment type 8409520.

Type of Change Date


For internal use only. Copyright reserved.
GBK5503140AE1.fm 19.04.99

5503.140 Oxylog 04.97 Changes Page 37


Appendix

1 Abbreviations

ABS Acrylonitrile-butadiene-styrene copolymer


CGS Central gas supply
DIN German Industrial Standard
IEC International Electrotechnical Commission
I:E Inspiration/expiration ratio
MIL US-Military Standard
MV Minute volume
OP Operating room
O2 Oxygen
PSU Polysulfone
VDE Association of German Electrical Engineers
For internal use only. Copyright reserved.
GBK5503140AH1.fm 19.04.99

5503.140 Oxylog 04.97 Appendix Page 38


2 Spare parts list Oxylog / Oxylog Samsomatic
For internal use only. Copyright reserved.
GBK5503140AH1.fm 19.04.99

5503.140 Oxylog 04.97 Appendix Page 39


DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 1 von 36
OXYLOG

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 2 von 36
OXYLOG

Diese Ersatzartikelliste gilt für Sachnummer:


This spare parts list is valid for part no.:

Sach-Nr. Benennung
Part No. Description
8404400 OXYLOG
OXYLOG

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 3 von 36
OXYLOG

Inhaltsverzeichnis der Bilder


Summary of pictures

Bild Bezeichnung Sach-Nr. E-Liste


Picture Description Part No . Spare parts list
1 OXYLOG 8404440
BASIC UNIT OXYLOG

2 AMV-VENTIL 8404453
AMV-VALVE

3 MV-VENTIL 8409001
VALVE

4 VENTILGRUPPE 8404550
VALVE GROUP

5 INJEKTOR,KOMPL. 8403012
INJECTOR,CPL.

6 HALTER MIT KLAUE 8405009


HOLDER WITH CLAMP

7 BEATMUNGSVENTIL 8406600
RESPIRATION VALVE 2

8 VERSCHLAUCHUNGSSCHEMA OXYLOG 8404434


PIPING-SCHEMATIC OXYLOG

9 TRAGEGURT 8404773
CARRYING STRAP

10 DRUCKBEGRENZUNGSVENTIL 8405390
PRESSURE LIMITING VALVE

11 OXYLOG *FRANZOESISCH* 8405777


OXYLOG (FRENCH)

12 UMRUESTSATZ SPONTANATMUNG 8407896


MODIF.SET SPONT.BREATH.OXYLOG

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 4 von 36
OXYLOG

OXYLOG Bild/Picture 1
BASIC UNIT OXYLOG

Datei/File: 55031.001
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 5 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-59 OXYLOG-GRUNDEINHEIT 8404440
BASIC UNIT OXYLOG

1 GEHAEUSE,VOLLST. (OXYLOG) 8410975


HOUSING,CPL. (OXYLOG)

2 SENKSCHRAUBE AM4X16DIN963-A2 1329332


COUNTER SUNK AM4X16DIN963-A2

3 KURZBETRIEBSANLEITUNG (OXYLOG) 8404771


OPERATING INSTRUCTIONS(OXYLOG)

4 SIEBEINSATZ R22028 10
SCREEN INSERT

5 O-RING 2M12839 10
O-RING SEAL

6 DRUCKANSCHLUSS,UGR. 8404446
DELIVERY CONNECTION

DIN 914-M5X12-45H/155 1306383


ohne Abbildung
DIN 914 M5X12-45H/155
without illustration

7 FILTER,UGR. 8405272
FILTER

8 DICHTRING 2M01524 10
PACKING RING

9 PATIENTENANSCHLUSS,UGR. 8404443
PATIENT CONNECTION

10 VOLUMEN 3,2 CCM 8403026


VOLUME 3,2 CCM

11 VOLUMEN 4,8 CCM 8403025


VOLUME 4,8 CCM

12 DISTANZBOLZEN 2M07604
DISTANT BOLT

13 SCHRAUBE AM 4X8 DIN 84-A4 NG 1330659 25


CHEESE HEAD SCREW AM4X8 DIN84

14 FILTERGEHAEUSE 8404438
FILTER HOUSING

15-18 WINKELTUELLE M 6X0,5 8402398


ANGULAR PORCELAIN BUSH M 5X0,5

15 SCHEIBE 8402392 10
WASHER

16 TUELLE 8402397
SOCKET

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 6 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
17 SCHEIBE 8402395 10
WASHER

18 BOLZEN 8402396
BOLT

19 VENTILGRUPPE 8404550
VALVE GROUP

20 VERTEILER,UGR. 8404465
DISTRIBUTOR

21 SCHRAUBE 8403057 5
SCREW

22 RING 8404988 10
RING

23 DRUCKREGLER 8402745
PRESSURE REGULATOR

24 TUELLE 8404508
SOCKET

25 DIN 914-M5X12-45H/155 1306383


DIN 914 M5X12-45H/155

26 DIN 84-AM3X6-A4 1330810


SCREW M3X6 DIN 84-A4

27 SETZMUTTER CH17646
NUT

28 SCHRAUBE 8403020
SCREW

29 3/2-WEGEVENTIL 8403006
3/2-PORT DISTRIBUTING VALVE

30 MUTTER BM3 DIN 439-A4/051 1334743


NUT BM3 DIN 439-A4/051

31 DIN 84-AM3X30-A2 1333828


DIN 84-AM3X30-A2

32 ZYLINDERSCHRAUBE AM 3X12 DIN84 1304232 2


CHEESE HEAD SCREW AM3X12 DIN84

33 DOSIERUNG 8403795
DOSAGE

15,35 TUELLE M 6X0,5 8402394


SOCKET M 6X0,5

35 TUELLE 8402393
SOCKET

36 SETZMUTTER CH18043
NUT

37 EINSCHUB 8404449 10
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 7 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
PLUG-IN UNIT

38 SCHRAUBE 00003 1330802 100


SCREW 00003

39 FEINDOSIERVENTIL (OXYLOG) 8404454


PRECISION METERING VALVE

40 SCHLAUCHVERSCHRAUBUNG 8402032 5
THREADED HOSE COUPLING

41 SCHWENKVERSCHRAUBUNG 8402034
SWIVELLING SCREW-FITTING

42 AMV-VENTIL 8404453
AMV-VALVE

42 MV-VENTIL 8409001
MV-VALVE

43 3/2-WEGE-VENTIL 8402441
3/2-PORT DISTRIBUTING VALVE

44 SH-SCHELLE DIN 3017-AL50-70-W3 1336681


SH-CLAMP

45 MANSCHETTE 8404463
PACKING

46 DRUCKMESSER 7275094
PRESSURE GAUGE

47 BLECH 8404077
SHEET METAL

48 SENKSCHRAUBE AM3X6 DIN963-A4 B 1287400


COUNTERS.SCREW AM3X6DIN963-A4

49 MUTTER 8402484
NUT

50 VERSTAERKERRELAIS M22260
AMPLIFIER RELAIS

51 VERSTAERKERRELAIS M25716 10
AMPLIFIER RELAY

52 FRONTSCHILD 8404770
FRONT PANEL

53-57 DREHKNOPF CO,SCHWARZ M24403


CONTROL KNOB CO,BLACK

53 KAPPE CO,SCHWARZ M24629 10


CAP CO,BLACK

54 MUTTER M24628
NUT

55 EINLAGE M24418
INSERT

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 8 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
56 DREHKNOPF-GEHAEUSE CO SW,UGR. M24624
CONTROL KNOB HOUSING CO SW

57 SPANNZANGE TV M19191
COLLET CHUCK

58 BOLZEN 8404076
BOLT

59 E-SET DRUCKREGLER (KNOCKS) 8406678


REP.SET PRESS.REGULAT.(KNOCKS)

59 E-SET DRUCKREGLER (LORCH) 8411142


REP.SET PRESSURE REGULAT(LORCH

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 9 von 36
OXYLOG

AMV-VENTIL Bild/Picture 2
AMV-VALVE

Datei/File: 55031.002
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 10 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-19 AMV-VENTIL 8404453
AMV-VALVE

1 DIN 84-AM3X30-A2 1333828


DIN 84-AM3X30-A2

2 ZYLINDERSCHRB.AM3X25 DIN84 NG 1333801


CHEESE HEAD SCREW AM3X25 DIN84

3 SCHRAUBE NG D21045
SCREW

4 SCHEIBE 8402392 10
WASHER

5 FEDER 2M08869
SPRING

6 BOLZEN 8404456
BOLT

7 GEHAEUSE 8404455
HOUSING

8 KUGEL 4 MM III DIN 5401 1270621


BALL 4 MM III DIN 5401

9 MEMBRAN,UGR. 8404457
DIAPHRAGM

10 BUCHSE 8403492
SOCKET

11 FEDER 8405189
SPRING

12 GEHAEUSE 8403489
HOUSING

13 EINSATZ NG E22265
INSERT

14 GEWINDESTIFT 1323253 30
THREADED PIN

15 3M6X30 DIN 7-A4 1290150


3M6X30 DIN 7-A4

16 M3X10 DIN 427-A4 1334190


M3X10 DIN 427-A4

17 SPINDEL 8403491
SPINDLE

18 SCHAFTSCHRAUBE M3X6DIN427-A4TV 1300059


HEADLESS SCREW M3X6 DIN427-A4

19 DIN 913-M4X12-45H/155 1328298


DIN 913-M4X12-45H/155

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 11 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
20-33 FEINDOSIERVENTIL (OXYLOG) 8404454
PRECISION METERING VALVE

20 ZYL.SCHRB.AM3X8 DIN84-A4/051 1330799 25


CH.HEAD SCR.AM3XDIN84-A4/051

21 VERSCHLUSS 8403493
LOCK

22 RUNDSCHNURRING M11995 5
TOROIDAL SEALING RING

23 FEDER 2M03494
SPRING

24 O-RING E20566 5
O-RING SEAL

25 KONUSTEIL 8403494
CONICAL PART

26 GEHAEUSE 8404459
HOUSING

27 EINSATZ NG E22265
INSERT

28 GEWINDESTIFT 1323253 30
THREADED PIN

29 M3X10 DIN 427-A4 1334190


M3X10 DIN 427-A4

30 3M6X30 DIN 7-A4 1290150


3M6X30 DIN 7-A4

31 SPINDEL 8404460
SPINDLE

32 SCHAFTSCHRAUBE M3X6DIN427-A4TV 1300059


HEADLESS SCREW M3X6 DIN427-A4

33 DIN 913-M4X12-45H/155 1328298


DIN 913-M4X12-45H/155

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 12 von 36
OXYLOG

MV-VENTIL Bild/Picture 3
VALVE

Datei/File: 55031.013
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 13 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-20 MV-VENTIL 8409001
MV-VALVE

1 SCHRAUBE 8408977
SCREW

2 SCHEIBE 8408967
WASHER

3 FEDER ZUM SI-VENTIL D06169


SPRING FOR SAFETY VALVE

4 STOESSEL 8408999
TAPPET

5 DIN 84-AM3X30-A2 1333828


DIN 84-AM3X30-A2

6 ZYLINDERSCHRB.AM3X25 DIN84 NG 1333801


CHEESE HEAD SCREW AM3X25 DIN84

7 KUGEL 4 MM III DIN 5401 1270621


BALL 4 MM III DIN 5401

8 GEHAEUSEKOPF 8408963
HOUSING TOP

9 MEMBRAN,UGR. 8404457
DIAPHRAGM

11 BUCHSE 8403492
SOCKET

12 FEDER 8405189
SPRING

13 GEHAEUSE 8403489
HOUSING

14 SPINDEL 8403491
SPINDLE

15 3M6X30 DIN 7-A4 1290150


3M6X30 DIN 7-A4

16 DIN 913-M4X12-45H/155 1328298


DIN 913-M4X12-45H/155

17 SCHAFTSCHRAUBE M3X6DIN427-A4TV 1300059


HEADLESS SCREW M3X6 DIN427-A4

18 M3X10 DIN 427-A4 1334190


M3X10 DIN 427-A4

19 EINSATZ NG E22265
INSERT

20 GEWINDESTIFT 1323253 30
THREADED PIN

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 14 von 36
OXYLOG

VENTILGRUPPE Bild/Picture 4
VALVE GROUP

Datei/File: 55031.003
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 15 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-24 VENTILGRUPPE 8404550
VALVE GROUP

1 ELEMENTENSOCKEL 8404439
ELEMENT-BASE

2 DICHTUNG 8403433 10
GASKET

3 VENTILKOERPER 8403018
VALVE BODY

4 VENTILTELLER NG 8403017
VALVE DISK

5 MEMBRAN 8403014 10
DIAPHRAGM

6 DECKEL,UGR. 8404757
COVER

7 ZYLINDERSCHRB.DIN84-AM3X20-A2 1333798
CH.HEAD SCREW DIN84-AM3X20-A2

8 GEWINDESTIFT 1323253 30
THREADED PIN

9 3/2-WEGEVENTIL 8403006
3/2-PORT DISTRIBUTING VALVE

10 ZYLINDERSCHRB.AM3X25 DIN84 NG 1333801


CHEESE HEAD SCREW AM3X25 DIN84

11-16 DOSIERTUELLE NG 8403320


METERING NOZZLE

11 SCHEIBE 8402392 10
WASHER

12 TUELLE 8402397
SOCKET

13 SCHEIBE 8402395 10
WASHER

14 BOLZEN 8403321
BOLT

15 SCHRAUBE 8403322
SCREW

16 O-RING RM07823 10
O-RING SEAL

17 INJEKTOR, KOMPL. (OXYLOG) 8403012


INJECTOR, CPL. (OXYLOG)

18 ANSCHLUSS 8404437
CONNECTION

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 16 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
11-13,19 WINKELTUELLE M 6X0,5 8402398
ANGULAR PORCELAIN BUSH M 5X0,5

19 BOLZEN 8402396
BOLT

8,20-23 SICHERHEITSVENTIL 8404749


SAFETY VALVE

20 GEHAEUSE,UGR. 8404750
HOUSING

21 FEDERAUFNAHME 2M17674
SPRING SEAT

22 FEDER M08902 5
SPRING

23 SCHRAUBE 2M17675
SCREW

24 DIN 84-AM3X30-A2 1333828


DIN 84-AM3X30-A2

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 17 von 36
OXYLOG

INJEKTOR,KOMPL. Bild/Picture 5
INJECTOR,CPL.

Datei/File: 55031.004
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 18 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-9 INJEKTOR, KOMPL. (OXYLOG) 8403012
INJECTOR, CPL. (OXYLOG)

1 SPANNSCHRAUBE 8402481
STRAINING SCREW

2 DRUCKDUESE NG 8403010
PRESSURE NOZZLE

3 MISCHDUESE 8403011
MIXING NOZZLE

4 SCHEIBE 8403009 2
WASHER

5 VENTILTELLER 8408899
VALVE PLATE

6 RUNDSCHNURRING M18445 10
TOROIDAL SEALING RING

7 BUCHSE 8403007
SOCKET

8 PRAEZISIONS-O-RING M14545 10
PRECISION O-RING SEAL

9 RUNDSCHNURRING D16558 10
TOROIDAL SEALING RING

10-14 3/2-WEGEVENTIL 8403006


3/2-PORT DISTRIBUTING VALVE

10 SCHRAUBE 00003 1329987 50


SCREW 00003

11 DECKEL,UGR. 8403074
COVER

12 MEMBRAN 8403004 2
DIAPHRAGM

13 M3X10 DIN 551-A4 1334239


M3X10 DIN 551-A4

14 GEHAEUSE 8403002
HOUSING

15-17 DOSIERUNG 8403795


DOSAGE

15 GEHAEUSE,UGR. 8403798
HOUSING

16 O-RING 8403962 5
O-RING SEAL

17 DOSIERKEGEL 8403797
METERING CONE

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 19 von 36
OXYLOG

HALTER MIT KLAUE Bild/Picture 6


HOLDER WITH CLAMP

Datei/File: 55031.005
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 20 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-18 HALTER MIT SCHIENENKLAUE 8405009
SUPPORT WITH CLAMP

1-11 HALTER 8404560


HANDLE

1 WANNE NG 8405008
TROUGH

2 BOLZEN 8404572
BOLT

3 SCHRAUBE AM4X6 DIN963-A4/051 1315781


SCREW AM4X6 DIN 963-A4/051

4 FEDERRING B 4 DIN 127 1331345 50


SPLIT WASHER B4 DIN 127

5 6KT.MUTTER M4 DIN934-A4/051 1328956 20


HEXAGON NUT M4 DIN 934-A4/051

6 SCHEIBE A 4,3 DIN 125-A2 1327542 20


WASHER A 4,3 DIN 125-A2

7 SCHEIBE B 4,3 DIN 9021-A4 NG 1331914


WASHER B 4,3 DIN 9021-A4

8 B 3,9X22 DIN 7971-A2 1334964


B 3,9X22 DIN 7971-A2

9 ZYL.-BLECHSCHRB.B3,5X16 NG 1334956
CHEESE HEAD B3,5X16

10 ZYLINDERSCHRB.AM4X20DIN84-A4 1328980 10
CYLIND.HEAD SCREW AM4X20 DIN84

11 SCHRAUBE AM 4X40 DIN 84 1298712


SCREW AM 4X40 DIN 84

12 AM5X12 DIN 963-A2 1330225


AM5X12 DIN 963-A2

13-16 SCHIENENKLAUE (GESCHRAUBT) M20650


CLAW

13 PUFFER NG M20646
BUFFER

14 KLEMMSCHRAUBE M20647
ADJUSTING SCREW

15 HALTER M20651
SUPPORT

16 SICHERUNGSSCHEIBE 5 DIN 6799 1326406 50


THRUST WASHER

17 FEDERRING B 5 DIN 127 NG 1300849


SPLIT WASHER B5 DIN 127

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 21 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
18 M5 DIN 546-MS 1256637
M5 DIN 546-MS

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 22 von 36
OXYLOG

BEATMUNGSVENTIL Bild/Picture 7
RESPIRATION VALVE 2

Datei/File: 55031.006
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 23 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-4 BEATMUNGSVENTIL 2 8406600
RESPIRATION VALVE 2

1 DECKEL 8406585
COVER

2,3 MEMBRAN,UGR. 8403552 2


DIAPHRAGM

2 MEMBRANE 8403206
DIAPHRAGM

3 MEMBRANE 8403395
DIAPHRAGM

4 VENTILGEHAEUSE 8406595
VALVE BOX

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 24 von 36
OXYLOG

VERSCHLAUCHUNGSSCHEMA OXYLOG Bild/Picture 8


PIPING-SCHEMATIC OXYLOG

Datei/File: 55031.007
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 25 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-9 VERSCHLAUCHUNGSSCHEMA OXYLOG 8404434
PIPING SCHEME OXYLOG

1 T-STUECK 8401083
T-PIECE

2 KREUZSTUECK 8401071
CROSS PIECE

3 SCHLAUCH 4X1,5-SI 50 SH A NF 1190520


HOSE 4X1,5-SI 50 SH A NF

4 ROHR 6X1 DIN73378-PA 12 W NF. 1202618


TUBE 6X1 DIN73378-PA 12 W NF.

5 SCHLAUCH 2X1 SI NF 1180614 25


HOSE 2X1 SI NF

6 SCHLAUCH 2X1-SIGN RT 1204793 5


HOSE 2X1-SIGN RED

7 SCHLAUCH 2X1-SIGN BL 1204785 5


HOSE 2X1-SIGN BL

8 SCHLAUCH 2X1-SIGN GN 1204807 5


HOSE 2X1-SIGN GREEN

9 SCHLAUCH 2X1-SIGN GE 1204831 5


HOSE 2X1-SIGN YELLOW

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 26 von 36
OXYLOG

TRAGEGURT Bild/Picture 9
CARRYING STRAP

Datei/File: 55031.008
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 27 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1+2 TRAGERIEMEN 8404773
CARRYING STRAP

1 LASCHE 8405179
BRACKET

2 GURT 8404078
BELT

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 28 von 36
OXYLOG

DRUCKBEGRENZUNGSVENTIL Bild/Picture 10
PRESSURE LIMITING VALVE

Datei/File: 55031.009
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 29 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-7 DRUCKBEGRENZUNGSVENTIL 8405390
PRESSURE LIMITING VALVE

1 BUEGEL 8405392
CLIP

2 SENKSCHRAUBE AM 3X4 DIN 63 1308459


COUNTERSUNK SCREW AM 3X4DIN963

3 SCHRAUBE 8405393
SCREW

4 SPRENGRING RM08289
XYZ

5 FEDERAUFNAHME,UGR. 8405457
SPRING RECEIVER

6 STOESSEL 8405394
TAPPET

7 AUFNAHME,UGR. 8405391
RECEIVER

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 30 von 36
OXYLOG

OXYLOG *FRANZOESISCH* Bild/Picture 11


OXYLOG (FRENCH)

Datei/File: 55031.010
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 31 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-9 OXYLOG (FRANZOESISCH) 8405777
OXYLOG (FRENCH)

1-4 OXYLOG 8404400


OXYLOG

1 OXYLOG-GRUNDEINHEIT 8404440
BASIC UNIT OXYLOG

2 VERBINDUNGSSCHLAUCH -SILIKON- 8404063


SILICONE HOSE ADULT 1,1M

3 BEATMUNGSVENTIL 2 8406600
RESPIRATION VALVE 2

4 TRAGERIEMEN 8404773
CARRYING STRAP

5-10 UMRUESTSATZ OXYLOG (FRANZ.) 8405304


MODIF.SET OXYLOG (FRENCH)

5 DRUCKBEGRENZUNGSVENTIL 8405390
PRESSURE LIMITING VALVE

6 FRONTSCHILD,FRANZ. 8403028
FRONT PANEL, FRENCH

7 UMRUESTANWEISUNG 8405297
CONVERSION INSTRUCTIONS

8 KURZ-BETRIEBSANLEITUNG, FRZ. 8403029


OPERATING INSTR.,FRENCH

9 BA 5503.1 F OXYLOG 9027742


BA 5503.1 F OXYLOG

10 ADAPTER-AIR-LIQUIDE 8405301
ADAPTER-AIR-LIQUIDE

1-4,6A, OXYLOG 30 (FRANZOESISCH) 8405778


7A,8-10 OXYLOG 30 (FRENCH)

6A,7A, UMRUESTSATZ OXYLOG 30(FRANZ.) 8405299


8-10 MODIF.SET OXYLOG 30 (FRENCH)

6A FRONTSCHILD OXYLOG 30 (FRZ.) 8405298


FRONT PANEL OXYLOG 30 (FRENCH)

7A UMRUESTANWEISUNG OXYLOG 30*FRZ 8405302


CONVERSION INSTR.OXYLOF 30,FR.

1-4,8A OXYLOG (ENGLISCH) 8405790


OXYLOG (ENGL.)

8A KURZ-BA OXYLOG (ENGLISCH) 8405446


ABBR.OP.MANUAL OXYLOG(ENGL.)

1-4, 6B- OXYLOG (SPANISCH) 8405800


9B OXYLOG (SPAN.)

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 32 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
6B-9B UMRUESTSATZ OXYLOG (SPANISCH) 8405801
MODIFICATION SET OXYLOG(SPAN.)

6B FRONTSCHILD, SPANISCH 8403030


FRONT PANEL, SPANISH

7B UMRUESTANWEISUNG *SPANISCH* 8405789


CONVERSION INSTRUCTIONS, SPAN.

8B KURZ-BETRIEBSANLEITUNG,SPAN. 8403031
OPERATING INSTRUCTIONS,SPANISH

9A BA 5503.1 S OXYLOG 9027743


OPERAT.INSTR. 5503.1 S OXYLOG

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 33 von 36
OXYLOG

UMRUESTSATZ SPONTANATMUNG Bild/Picture 12


MODIF.SET SPONT.BREATH.OXYLOG

Datei/File: 55031.012
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 34 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-45 RS SPONTANATMUNG-DRELOBA 8407896
MODIF.SET SPONT.BREATH.DRELOBA

1-38 LUNGENAUTOMAT (UGR.) NG 8408505


LUNG DEMAND VALVE

1 SCHRAUBE 00001 1330055 20


SCREW 00001

2 B4 DIN 137-NICHTROST.ST 1335308


B4 DIN 137-STAINLESS STEEL

4 DECKEL 8407848
COVER

6 E-SET MEMBRAN (R27929) R23692


REP.SET DIAPHRAGM (R27929)

7 ZYLINDERSCHRB.AM3X25 DIN84 NG 1333801


CHEESE HEAD SCREW AM3X25 DIN84

8 GEHAEUSE,UGR. 8408037
HOUSING

9 GEWINDESTIFT M3X6 DIN551-A2 1334220


SET SCREW M3X6 DIN551-A2

10 SCHRAUBTUELLE 8407834
SCREW SOCKET

11-15 SICHERHEITSVENTIL NG 8407320


SAFETY VALVE

11 VENTILSCHRAUBE 8401739
VALVE SCREW

12 FEDER NG 2M03627
SPRING

13 FEDERAUFNAHME 8401737
BUFFER BLOCK

14 DICHTSCHEIBE 8401738
SEALING WASHER

15 VENTILGEHAEUSE 8407319
VALVE HOUSING

16-22 VENTILSTUECK,VOLLST. R50520


VALVE PIECE, CPL.

16-22,31 E-SET VENTILSTUECK,VOLLST. R23694


REP.SET VALVE PIECE

16 KRATEREINSATZ R50521
CRATER INSERT

17 KIPPHEBEL R27999
LEVER, CPL.

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 35 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
18 FEDER R28133 10
SPRING

19 GEHAEUSE FUER PA80 R50522


HOUSING FOR PA 80

20 RUNDSCHNURRING R27810 10
TOROIDAL SEALING RING

21 TUELLE R50479
SOCKET

22 UEBERWURFMUTTER R50478
UNION NUT

23 SCHLAUCHKLEMME R20476
HOSE CLAMP

24-30 DRUCKGASANSCHLUSS,KPL.UGR. NG 8408303


COMPRESSED GAS CONNECTION

24 SCHLAUCH 6X3 - PVC 1181106


HOSE 6X3 - PVC

25 FASSUNG SCHLAUCHHUELSE M26950 10


HOLDER FOR HOSE SOCKET

26 UEBERWURFMUTTER NG M19098
UNION NUT

27 SCHLAUCHTUELLE 8408293
HOSE SOCKET

28 DICHTRING M05128 20
PACKING RING

29,30 DRUCKGASANSCHLUSS, UGR. 8408301


COMPRESSED GAS CONNECTION

30 O-RING 2M12839 10
O-RING SEAL

31 O-RING R20771 10
O-RING SEAL

32 SCHLAUCH 12X3 1198912


HOSE 12X3

33,34 PATIENTENGASANSCHLUSS NG 8407893


PATIENT GAS SUPPLY

34 DICHTRING 2M01524 10
PACKING RING

35 KABELBINDER (2,4X92LG) 8712007 25


CABLE CLIP (2,4X92)

36 DICHTRING R22078 10
WASHER

38 GEWINDEKONUS 8407832
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.1 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG Seite/Page 36 von 36
OXYLOG

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
THREADED CONE

39-42 BEATMUNGSVENTIL M.R-VENTIL 8408568


RESPIRATION VALVE

39 DECKEL 8406585
COVER

40,41 MEMBRAN,UGR. 8403552 2


DIAPHRAGM

40 MEMBRANE 8403206
DIAPHRAGM

41 MEMBRANE 8403395
DIAPHRAGM

42 VENTILGEHAEUSE, KOMPL. 8408569


VALVE BOX, CPL.

43 DRAEGER-MASKE GR.3-ISO M25574


DRAEGER MASK SIZE 3-ISO

44 SCHELLE M21610 10
CLIP

45 ANSCHLUSS M27691
CONNECTION

46 VERBINDUNGSSCHLAUCH -SILIKON- 8404063


SILICONE HOSE ADULT 1,1M

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 1 von 44
OXYLOG (SAMSOMATIC)

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 2 von 44
OXYLOG (SAMSOMATIC)

Diese Ersatzartikelliste gilt für Sachnummer:


This spare parts list is valid for part no.:

Sach-Nr. Benennung
Part No. Description
8409520 OXYLOG
OXYLOG

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 3 von 44
OXYLOG (SAMSOMATIC)

Inhaltsverzeichnis der Bilder


Summary of pictures

Bild Bezeichnung Sach-Nr. E-Liste


Picture Description Part No . Spare parts list
1 OXYLOG M.BEATMUNGSZUBEHOER 8409520
OXYLOG

2 OXYLOG SAMSOMATIC 8408500


OXYLOG S

3 VENTILGRUPPE 8409597
VALVE ASSEMBLY

4 MV-VENTIL 8409001
MV-VALVE

5 FEINDOSIERVENTIL 8409090
PRECISION METERING VALVE

6 INJEKTOR,KOMPL. 8403012
INJECTOR,CPL.

7 SCHLAEUCHE SAMSOMATIC 8408694


HOSES SAMSOMATIC

8 BEATMUNGSVENTIL 8406600
RESPIRATION VALVE 2

9 ZUBEHOER FUER TRAGEPLATTE 8410221


ACCESSORIES CARRYING PLATE

10 ZUBEHOER FUER RUCKSACK 8410222


ACCESSORIES BACKPACK

11 ZUBEHOER FUER KOFFER 8410223


ACCESSORIES CASE

12 ZUBEHOER FUER KOFFER 8410223


ACCESSORIES CASE

13 SUJECTOR 2000 SEKRETABSAUGUNG 2120232


SUJECTOR 2000 ASPIRATOR

14 RESUTATOR 2000 ERWACHSENE 2120046


RESUTATOR 2000 ADULTS

15 OXIDEM 2000 DEMAND-VENTIL 2120070


OXIDEM 2000 DEMAND-VALVE

16 UMRUESTSATZ SPONTANATMUNG 8407896


MODIF.SET SPONT.BREATH.OXYLOG

17 TRAGEPLATTE MIT OXYLOGHALTER 8412148


CARRYING PLATE WITH OXYLOG SUPPORT

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 4 von 44
OXYLOG (SAMSOMATIC)

OXYLOG M.BEATMUNGSZUBEHOER Bild/Picture 1


OXYLOG

Datei/File: 550314.001
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 5 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-4 OXYLOG 8409520
OXYLOG

1 TRAGERIEMEN 8404773
CARRYING STRAP

2 OXYLOG S 8408500
OXYLOG S

3 VERBINDUNGSSCHLAUCH -SILIKON- 8404063


SILICONE HOSE ADULT 1,1M

4 BEATMUNGSVENTIL 2 8406600
RESPIRATION VALVE 2

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 6 von 44
OXYLOG (SAMSOMATIC)

OXYLOG SAMSOMATIC Bild/Picture 2


OXYLOG S

Datei/File: 550314.002
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 7 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-59 OXYLOG S 8408500
OXYLOG S

1 GEHAEUSE,VOLLST. (OXYLOG) 8410975


HOUSING,CPL. (OXYLOG)

2 SENKSCHRAUBE AM4X16DIN963-A2 1329332


COUNTER SUNK AM4X16DIN963-A2

3 KURZ-BETRIEBSANLEITUNG NG 8409596
OPERATING INSTRUCTIONS

4 SIEBEINSATZ R22028 10
SCREEN INSERT

5 O-RING 2M12839 10
O-RING SEAL

6 DRUCKANSCHLUSS,UGR. 8404446
DELIVERY CONNECTION

6.1 DIN 914-M5X12-45H/155 1306383


DIN 914 M5X12-45H/155

7 FILTER,UGR. 8405272
FILTER

8 DICHTRING 2M01524 10
PACKING RING

9 PATIENTENANSCHLUSS,UGR. 8404443
PATIENT CONNECTION

10 VOLUMEN, VOLLST. 2,3 ML NG 8409099


VOLUME, CPL. 2,3 ML

11 VOLUMEN,VOLLST.4,8 ML 8409111
VOLUME, CPL. 4,8 ML

12 DISTANZBOLZEN 2M07604
DISTANT BOLT

13 SCHRAUBE AM 4X8 DIN 84-A4 NG 1330659 25


CHEESE HEAD SCREW AM4X8 DIN84

14 FILTERGEHAEUSE 8404438
FILTER HOUSING

15-18 WINKELTUELLE M 6X0,5 8402398


ANGULAR PORCELAIN BUSH M 5X0,5

15 SCHEIBE 8402392 10
WASHER

16 TUELLE 8402397
SOCKET

17 SCHEIBE 8402395 10
WASHER

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 8 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
18 BOLZEN 8402396
BOLT

19 VENTILGRUPPE 8409597
VALVE ASSEMBLY

20 VERTEILER,UGR. 8404465
DISTRIBUTOR

21 SCHRAUBE 8403057 5
SCREW

22 RING 8404988 10
RING

23 DRUCKREGLER M29561
PRESSURE REGULATOR

24 TUELLE 8404508
SOCKET

25 DIN 914-M5X12-45H/155 1306383


DIN 914 M5X12-45H/155

26 DIN 84-AM3X30-A2 1333828


DIN 84-AM3X30-A2

27 EINSCHUB 8408688 10
PLUG-IN UNIT

28 SCHRAUBE 8403020
SCREW

29 3/2-WEGEVENTIL 8403006
3/2-PORT DISTRIBUTING VALVE

30 MUTTER BM3 DIN 439-A4/051 1334743


NUT BM3 DIN 439-A4/051

31 DIN 84-AM3X30-A2 1333828


DIN 84-AM3X30-A2

32 SETZMUTTER CH18043
NUT

33 FEINDOSIERVENTIL NG 8409090
PRECISION METERING VALVE

34 SCHWENKVERSCHRAUBUNG 8402034
SWIVELLING SCREW-FITTING

35 MV-VENTIL 8409001
MV-VALVE

36 FEDERSCHEIBE B 3 DIN 137 1335294


SPRING WASHER B3 DIN 137

37 ZYLINDERSCHRAUBE AM 3X12 DIN84 1304232 2


CHEESE HEAD SCREW AM3X12 DIN84

38 BUCHSE 00001 M11895


Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 9 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
SLEEVE 00001

15,39 TUELLE M 6X0,5 8402394


SOCKET M 6X0,5

39 TUELLE 8402393
SOCKET

40 KLEBESOCKEL 8406927
BASE

41 MIKROSCHALTER 8407324
MICROSWITCH

42 SCHWENKVERSCHRAUBUNG 8402034
SWIVELLING SCREW-FITTING

43 3/2-WEGE-VENTIL 8402441
3/2-PORT DISTRIBUTING VALVE

44 SENKSCHRAUBE AM3X6 DIN963-A4 B 1287400


COUNTERS.SCREW AM3X6DIN963-A4

45 BUEGEL 8403058 5
BOW

46 MUTTER 8402484
NUT

47 BZ2,2X6,5 DIN 7971-A4 1336819


BZ2,2X6,5 DIN 7971-A4

48 VERSCHALTUNGSPLATTE,VOLLST. 8408576
CONNECTION BOARD, CPL.

49 FRONTSCHILD NG 8409594
FRONT PANEL

50-54 DREHKNOPF CO,SCHWARZ M24403


CONTROL KNOB CO,BLACK

50 KAPPE CO,SCHWARZ M24629 10


CAP CO,BLACK

51 MUTTER M24628
NUT

52 DREHKNOPF-GEHAEUSE CO SW,UGR. M24624


CONTROL KNOB HOUSING CO SW

53 SPANNZANGE TV M19191
COLLET CHUCK

54 EINLAGE M24418
INSERT

55 BOLZEN 8404076
BOLT

56 MUTTER BM3 DIN 439-A4/051 1334743


NUT BM3 DIN 439-A4/051

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 10 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
57 SH-SCHELLE DIN 3017-AL50-70-W3 1336681
SH-CLAMP

58 MANSCHETTE 8404463
PACKING

59 DRUCKMESSER 7275094
PRESSURE GAUGE

60 E-SET DRUCKREGLER (KNOCKS) 8406678


REP.SET PRESS.REGULAT.(KNOCKS)

60 E-SET DRUCKREGLER (LORCH) 8411142


REP.SET PRESSURE REGULAT(LORCH

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 11 von 44
OXYLOG (SAMSOMATIC)

VENTILGRUPPE Bild/Picture 3
VALVE ASSEMBLY

Datei/File: 550314.003
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 12 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-23 VENTILGRUPPE 8409597
VALVE ASSEMBLY

1 INJEKTOR, KOMPL. (OXYLOG) 8403012


INJECTOR, CPL. (OXYLOG)

2,21-23 VENTILGRUPPE 8407595


VALVE ASSEMBLY

3 DICHTUNG 8403433 10
GASKET

4 VENTILKOERPER 8403018
VALVE BODY

5 VENTILTELLER NG 8403017
VALVE DISK

6 MEMBRAN 8403014 10
DIAPHRAGM

7 DECKEL,UGR. 8404757
COVER

8 ZYLINDERSCHRB.DIN84-AM3X20-A2 1333798
CH.HEAD SCREW DIN84-AM3X20-A2

9 ZYLINDERSCHRB.AM3X25 DIN84 NG 1333801


CHEESE HEAD SCREW AM3X25 DIN84

10 3/2-WEGEVENTIL 8403006
3/2-PORT DISTRIBUTING VALVE

11-16 DOSIERTUELLE NG 8403320


METERING NOZZLE

11-13,18 WINKELTUELLE M 6X0,5 8402398


ANGULAR PORCELAIN BUSH M 5X0,5

11 SCHEIBE 8402392 10
WASHER

12 TUELLE 8402397
SOCKET

13 SCHEIBE 8402395 10
WASHER

14 BOLZEN 8403321
BOLT

15 ANSCHLUSS 8403332
CONNECTION

16 O-RING RM07823 10
O-RING SEAL

18 BOLZEN 8402396
BOLT

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 13 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
19 ANSCHLUSS 8404437
CONNECTION

20 GEWINDESTIFT 1323253 30
THREADED PIN

21 VENTILTELLER 8407605
VALVE DISK

22 FEDER V06499
SPRING

23 SCHRAUBE 8407606
SCREW

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 14 von 44
OXYLOG (SAMSOMATIC)

MV-VENTIL Bild/Picture 4
MV-VALVE

Datei/File: 550314.004
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 15 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-20 MV-VENTIL 8409001
MV-VALVE

1 SCHRAUBE 8408977
SCREW

2 SCHEIBE 8408967
WASHER

3 FEDER ZUM SI-VENTIL D06169


SPRING FOR SAFETY VALVE

4 STOESSEL 8408999
TAPPET

5 DIN 84-AM3X30-A2 1333828


DIN 84-AM3X30-A2

6 ZYLINDERSCHRB.AM3X25 DIN84 NG 1333801


CHEESE HEAD SCREW AM3X25 DIN84

7 KUGEL 4 MM III DIN 5401 1270621


BALL 4 MM III DIN 5401

8 GEHAEUSEKOPF 8408963
HOUSING TOP

9 MEMBRAN,UGR. 8404457
DIAPHRAGM

11 BUCHSE 8403492
SOCKET

12 FEDER 8405189
SPRING

13 GEHAEUSE 8403489
HOUSING

14 SPINDEL 8403491
SPINDLE

15 3M6X30 DIN 7-A4 1290150


3M6X30 DIN 7-A4

16 DIN 913-M4X12-45H/155 1328298


DIN 913-M4X12-45H/155

17 SCHAFTSCHRAUBE M3X6DIN427-A4TV 1300059


HEADLESS SCREW M3X6 DIN427-A4

18 M3X10 DIN 427-A4 1334190


M3X10 DIN 427-A4

19 EINSATZ NG E22265
INSERT

20 GEWINDESTIFT 1323253 30
THREADED PIN

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 16 von 44
OXYLOG (SAMSOMATIC)

FEINDOSIERVENTIL Bild/Picture 5
PRECISION METERING VALVE

Datei/File: 550314.005
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 17 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-14 FEINDOSIERVENTIL NG 8409090
PRECISION METERING VALVE

1 FEDER 2M03494
SPRING

2 RUNDSCHNURRING M11995 5
TOROIDAL SEALING RING

3 VERSCHLUSS 8405979
LOCK

4 ZYL.SCHRB.AM3X8 DIN84-A4/051 1330799 25


CH.HEAD SCR.AM3XDIN84-A4/051

5 O-RING E20566 5
O-RING SEAL

6 KONUSTEIL 8405978
CONICAL PART

7 GEWINDESTIFT 1323253 30
THREADED PIN

8 EINSATZ NG E22265
INSERT

9 GEHAEUSE 8409089
HOUSING

10 M3X10 DIN 427-A4 1334190


M3X10 DIN 427-A4

11 SCHAFTSCHRAUBE M3X6DIN427-A4TV 1300059


HEADLESS SCREW M3X6 DIN427-A4

12 SPINDEL 8409088
SPINDLE

13 ZYLINDERSTIFT 2M6X30X12 DIN 7 1290185


CYLINDRICAL PIN 2 M 6X30X12

14 AM3X8 DIN 914-NICHTROST.ST 1299506


AM3x8 DIN 914 -STAINLESS STEEL

15-18 VERSCHALTUNGSPLATTE,VOLLST. 8408576


CONNECTION BOARD, CPL.

15 VERSCHALTUNGSPLATTE 8409046
CONNECTION BOARD

16 MIKROSCHALTER 8407265
MICROSWITCH

17 MIKROSCHALTER 8409233
MICROSWITCH

18 BUEGEL 8409047
CLIP

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 18 von 44
OXYLOG (SAMSOMATIC)

INJEKTOR,KOMPL. Bild/Picture 6
INJECTOR,CPL.

Datei/File: 550314.006
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 19 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-9 INJEKTOR, KOMPL. (OXYLOG) 8403012
INJECTOR, CPL. (OXYLOG)

1 SPANNSCHRAUBE 8402481
STRAINING SCREW

2 DRUCKDUESE NG 8403010
PRESSURE NOZZLE

3 MISCHDUESE 8403011
MIXING NOZZLE

4 SCHEIBE 8403009 2
WASHER

5 VENTILTELLER 8408899
VALVE PLATE

6 RUNDSCHNURRING M18445 10
TOROIDAL SEALING RING

7 BUCHSE 8403007
SOCKET

8 PRAEZISIONS-O-RING M14545 10
PRECISION O-RING SEAL

9 RUNDSCHNURRING D16558 10
TOROIDAL SEALING RING

10-14 3/2-WEGEVENTIL 8403006


3/2-PORT DISTRIBUTING VALVE

10 SCHRAUBE 00003 1329987 50


SCREW 00003

11 DECKEL,UGR. 8403074
COVER

12 MEMBRAN 8403004 2
DIAPHRAGM

13 M3X10 DIN 551-A4 1334239


M3X10 DIN 551-A4

14 GEHAEUSE 8403002
HOUSING

15-17 DOSIERUNG 8403795


DOSAGE

15 GEHAEUSE,UGR. 8403798
HOUSING

16 O-RING 8403962 5
O-RING SEAL

17 DOSIERKEGEL 8403797
METERING CONE

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 20 von 44
OXYLOG (SAMSOMATIC)

SCHLAEUCHE SAMSOMATIC Bild/Picture 7


HOSES SAMSOMATIC

Datei/File: 550314.007
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 21 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-13 SCHLAEUCHE SAMSOMATIC 8408694
HOSES SAMSOMATIC

1 SCHLAUCH 4X1,5-SI 50 SH A NF 1190520


HOSE 4X1,5-SI 50 SH A NF

2 ROHR 6X1 DIN73378-PA 12 W NF. 1202618


TUBE 6X1 DIN73378-PA 12 W NF.

3 SCHLAUCH 2X1 SI NF 1180614 25


HOSE 2X1 SI NF

4 SCHLAUCH 2X1-SIGN RT 1204793 5


HOSE 2X1-SIGN RED

5 SCHLAUCH 2X1-SIGN BL 1204785 5


HOSE 2X1-SIGN BL

6 SCHLAUCH 2X1-SIGN GN 1204807 5


HOSE 2X1-SIGN GREEN

7 SCHLAUCH 2X1-SIGN GE 1204831 5


HOSE 2X1-SIGN YELLOW

8 DOSIERUNG (0,5/1,5) 8406292


DOSAGE (0,5/1,5)

9 DOSIERUNG NG 8408048
DOSAGE

10 KREUZSTUECK 8401071
CROSS PIECE

11 T-STUECK 8401083
T-PIECE

12 SCHLAUCHANSATZ, MAENNL. 8405244


HOSE NOZZLE, MALE

13 FILTRATIONSVORSATZ 8405241 5
FILTERING ADAPTER

14 3/2-WEGEVENTIL 8403006
3/2-PORT DISTRIBUTING VALVE

15 VERTEILER,UGR. 8404465
DISTRIBUTOR

16 DRUCKREGLER NG 8408590
PRESSURE REGULATOR

17 VOLUMEN, VOLLST. 2,3 ML NG 8409099


VOLUME, CPL. 2,3 ML

18 VOLUMEN,VOLLST.4,8 ML 8409111
VOLUME, CPL. 4,8 ML

19 FILTERGEHAEUSE 8404438
FILTER HOUSING

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 22 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
20 INJEKTOR, KOMPL. (OXYLOG) 8403012
INJECTOR, CPL. (OXYLOG)

21 VENTILGRUPPE 8407595
VALVE ASSEMBLY

22 VERSCHALTUNGSPLATTE,VOLLST. 8408576
CONNECTION BOARD, CPL.

23 FEINDOSIERVENTIL NG 8409090
PRECISION METERING VALVE

24 MV-VENTIL 8409001
MV-VALVE

25 3/2-WEGE-VENTIL 8402441
3/2-PORT DISTRIBUTING VALVE

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 23 von 44
OXYLOG (SAMSOMATIC)

BEATMUNGSVENTIL Bild/Picture 8
RESPIRATION VALVE 2

Datei/File: 550314.008
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 24 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-4 BEATMUNGSVENTIL 2 8406600
RESPIRATION VALVE 2

1 DECKEL 8406585
COVER

2,3 MEMBRAN,UGR. 8403552 2


DIAPHRAGM

2 MEMBRANE 8403206
DIAPHRAGM

3 MEMBRANE 8403395
DIAPHRAGM

4 VENTILGEHAEUSE 8406595
VALVE BOX

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 25 von 44
OXYLOG (SAMSOMATIC)

ZUBEHOER FUER TRAGEPLATTE Bild/Picture 9


ACCESSORIES CARRYING PLATE

Datei/File: 550314.009
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 26 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-14 ZUBEHOER TRAGEPLATTE 8410221
ACCESSORIES CARRYING PLATE

1 ALDUK M I 2120208
ALDUK M I

2 SUJECTOR 2000 SEKRETABSAUGUNG 2120232


SUJECTOR 2000 ASPIRATOR

3 O2-FLASCHE AG 2,5/200-500L B03580


O2-CYLINDER AG 2,5/200-500L

4 MASKENKRUEMMER F.Y-STCK.-ISO M25649


MASK ELBOW F.Y-PIECE-ISO

5,5A O2-KUPPLUNGSSCHLAUCH 0,15M M23874


O2-COUPLING HOSE 0,15M

5A DICHTRING M07152 10
PACKING RING

KRUEMMER M.TUELLE NG 8410805


ohne Abbildung
ELBOW WITH SOCKET
without illustration

6,6A O2-DR.-ANSCHLUSS-SCHLAUCH 3M 8410219


O2-COMPR.AIR CONNECT.HOSE 3M

6A DICHTRING M05128 20
PACKING RING

7 O-RING (3ER-SET) F.ALDUK 2120364


O-RING SEAL(SET OF 3PCS.)ALDUK

7 O-RING 2121107
O-RING SEAL

8 GUMMIKAPPE (ALDUK) 2120682


RUBBER CAP (ALDUK)

9 DRUCKMESSER (ALDUK) 2121085


PRESSURE GAUGE (ALDUK)

10 UNTERLEGSCHEIBE 2120093
PLAIN WASHER

11 ZV-SCHLAUCH OXYPLATTE 8410884


CG-HOSE OXYPLATE

12,5A VERTEILER OXYPLATTE 8410883


DISTRIBUTOR OXYPLATE

O2-ANSCHLUSS-SCHLAUCH 0,25M 8410977


ohne Abbildung
O2-CONNECTING HOSE 0,25M
without illustration

14 RUECKSCHLAGVENTIL 8410885
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 27 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
NONRETURN VALVE

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 28 von 44
OXYLOG (SAMSOMATIC)

ZUBEHOER FUER RUCKSACK Bild/Picture 10


ACCESSORIES BACKPACK

Datei/File: 550314.010
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 29 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-11 ZUBEHOER RUCKSACK 8410222
ACCESSORIES BACKPACK

1 ALDUK M I 2120208
ALDUK M I

2 SUJECTOR 2000 SEKRETABSAUGUNG 2120232


SUJECTOR 2000 ASPIRATOR

3 O2-FLASCHE AG 2,5/200-500L B03580


O2-CYLINDER AG 2,5/200-500L

4 MASKENKRUEMMER F.Y-STCK.-ISO M25649


MASK ELBOW F.Y-PIECE-ISO

5 INHALATIONSEINRICHTUNG,ERW. 2182300
INHALATION DEVICE,ADULTS

6,6A,6B O2-ANSCHLUSS-SCHLAUCH 3M 8410224


O2 CONNECTION HOSE 3M

6A DICHTRING M05128 20
PACKING RING

6B DICHTRING M07152 10
PACKING RING

7 O-RING (3ER-SET) F.ALDUK 2120364


O-RING SEAL(SET OF 3PCS.)ALDUK

7 O-RING 2121107
O-RING SEAL

8 GUMMIKAPPE (ALDUK) 2120682


RUBBER CAP (ALDUK)

9 DRUCKMESSER (ALDUK) 2121085


PRESSURE GAUGE (ALDUK)

10 UNTERLEGSCHEIBE 2120093
PLAIN WASHER

11 RUECKSCHLAGVENTIL 8410885
NONRETURN VALVE

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 30 von 44
OXYLOG (SAMSOMATIC)

ZUBEHOER FUER KOFFER Bild/Picture 11


ACCESSORIES CASE

Datei/File: 550314.011
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 31 von 44
OXYLOG (SAMSOMATIC)

ZUBEHOER FUER KOFFER Bild/Picture 12


ACCESSORIES CASE

Datei/File: 550314.012
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 32 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-13 ZUBEHOER KOFFER 8410223
ACCESSORIES CASE

1 ALDUK M I 2120208
ALDUK M I

2 SUJECTOR 2000 SEKRETABSAUGUNG 2120232


SUJECTOR 2000 ASPIRATOR

3 INHALATIONSEINRICHTUNG,ERW. 2182300
INHALATION DEVICE,ADULTS

4 MASKENKRUEMMER F.Y-STCK.-ISO M25649


MASK ELBOW F.Y-PIECE-ISO

5,5A O2-KUPPLUNGSSCHLAUCH 0,15M M23874


O2-COUPLING HOSE 0,15M

5A DICHTRING M07152 10
PACKING RING

6,6A O2-DR.-ANSCHLUSS-SCHLAUCH 3M 8410219


O2-COMPR.AIR CONNECT.HOSE 3M

6A DICHTRING M05128 20
PACKING RING

7 O2-FLASCHE AG 2/200-400L B02352


O2-CYLINDER AG 2/200-400L

8 RESUTATOR 2000 ERWACHSENE 2120046


RESUTATOR 2000 ADULTS

9 O-RING (3ER-SET) F.ALDUK 2120364


O-RING SEAL(SET OF 3PCS.)ALDUK

9 O-RING 2121107
O-RING SEAL

10 GUMMIKAPPE (ALDUK) 2120682


RUBBER CAP (ALDUK)

11 DRUCKMESSER (ALDUK) 2121085


PRESSURE GAUGE (ALDUK)

12 UNTERLEGSCHEIBE 2120093
PLAIN WASHER

13 RUECKSCHLAGVENTIL 8410885
NONRETURN VALVE

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 33 von 44
OXYLOG (SAMSOMATIC)

SUJECTOR 2000 SEKRETABSAUGUNG Bild/Picture 13


SUJECTOR 2000 ASPIRATOR

Datei/File: 550314.013
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 34 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-6 SUJECTOR 2000 SEKRETABSAUGUNG 2120232
SUJECTOR 2000 ASPIRATOR

1 DICHTUNG F SUJECTOR-SEKRETFLA. 2120372


GASKET F COLLECTION-JAR ASPIR.

2 SEKRETFLASCHE F SUJECTOR 2120259


COLLECTION-JAR F ASPIRATOR

3 O2-ANSCHLUSSCHLAUCH F SUJECTOR 2120240


O2 ASPIRATOR HOSE F.SUJECTOR

4 ABSAUGEKATHETER 5ER SET 2120267


CATHETER F ASPIRATOR SET W5

5 DICHTUNG SEKRETFLASCHE 5ER SET 2120348


GASKET F COLLECTION-JA SET O,5

6 O-RING 2120690
O-RING SEAL

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 35 von 44
OXYLOG (SAMSOMATIC)

RESUTATOR 2000 ERWACHSENE Bild/Picture 14


RESUTATOR 2000 ADULTS

Datei/File: 550314.014
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 36 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-9 RESUTATOR 2000 ERWACHSENE 2120046
RESUTATOR 2000 ADULTS

1-11 KINDER-RESUTATOR 2000 2120984


CHILD-RESUTATOR 2000

BABY-RESUTATOR 2000 2120941


BABY-RESUTATOR 2000

1 ERSATZBEUTEL-KIND 2120992
REPLACEMENT BAG INFANT

2 ADAPTER F.UEBERDRUCKV.45MBAR 2120976


ADAPTER F.OVERPRESSURE 45MBAR

3 HALTERING 2120178
RETAINING RING / CLAMP

4-7 PATIENTENVENTIL KOMPL. 2120097


PATIENT VALVE ASSEMBLY

4 VENTIL-OBERTEIL PS 2120135
VALVE-HOUSING BAGSIDE

5 FLATTERVENTIL 5ER SET 2120151


INLET DIAPHRAGM

6 LIPPENVENTIL 2120143
DRUCKBILL DIAPHRAGM

7 VENTIL-UNTERTEIL PS 2120127
VALVE-HOUSING/PATTENSIDE

8 SILIKONMASKE GR.2,RUND 2121034


SILICONE MASK SIZE 2, ROUND

9 SILIKONMASKE GR.1,RUND 2121026


SILICONE MASK SIZE 1, ROUND

10 SILIKONMASKE GR.0, RUND 2121018


SILICONE MASK SIZE 0, ROUND

11 ERSATZBEUTEL-BABY 2120968
REPLACEMENT BAG BABY

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 37 von 44
OXYLOG (SAMSOMATIC)

OXIDEM 2000 DEMAND-VENTIL Bild/Picture 15


OXIDEM 2000 DEMAND-VALVE

Datei/File: 550314.015
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 38 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-8 OXIDEM 2000 DEMAND-VENTIL 2120070
OXIDEM 2000 DEMAND-VALVE

1 STEUERMEMBRAN (OXYDEM 2000) 2120704


CONTROL DIAPHRAGM (OXYDEM 2000

2 MEMBRAN FUER OXIDEM 2120542


DIAPHRAGM FOR OXIDEM

3 ADAPTER OXIDEM M12X1 8410547


ADAPTER OXIDEM M12X1

4 O2-SCHLAUCH OXIDEM 2120712


O2-TUBE OXIDEM

5 DICHTUNG SEKRETFLASCHE 5ER SET 2120348


GASKET F COLLECTION-JA SET O,5

6 ADAPTER AMBU MORK 3 2120488


ADAPTER AMBU MORK 3

7 ADAPTER LAERDAL 2120496


ADAPTER LAERDAL

8 ADAPTER RESUTATOR 2000 2120461


ADAPTER RESUTATOR 2000

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 39 von 44
OXYLOG (SAMSOMATIC)

UMRUESTSATZ SPONTANATMUNG Bild/Picture 16


MODIF.SET SPONT.BREATH.OXYLOG

Datei/File: 550314.016
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 40 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1-45 RS SPONTANATMUNG-DRELOBA 8407896
MODIF.SET SPONT.BREATH.DRELOBA

1-38 LUNGENAUTOMAT (UGR.) NG 8408505


LUNG DEMAND VALVE

1 SCHRAUBE 00001 1330055 20


SCREW 00001

2 B4 DIN 137-NICHTROST.ST 1335308


B4 DIN 137-STAINLESS STEEL

4 DECKEL 8407848
COVER

6 E-SET MEMBRAN (R27929) R23692


REP.SET DIAPHRAGM (R27929)

7 ZYLINDERSCHRB.AM3X25 DIN84 NG 1333801


CHEESE HEAD SCREW AM3X25 DIN84

8 GEHAEUSE,UGR. 8408037
HOUSING

9 GEWINDESTIFT M3X6 DIN551-A2 1334220


SET SCREW M3X6 DIN551-A2

10 SCHRAUBTUELLE 8407834
SCREW SOCKET

11-15 SICHERHEITSVENTIL NG 8407320


SAFETY VALVE

11 VENTILSCHRAUBE 8401739
VALVE SCREW

12 FEDER NG 2M03627
SPRING

13 FEDERAUFNAHME 8401737
BUFFER BLOCK

14 DICHTSCHEIBE 8401738
SEALING WASHER

15 VENTILGEHAEUSE 8407319
VALVE HOUSING

16-22 VENTILSTUECK,VOLLST. R50520


VALVE PIECE, CPL.

16-22,31 E-SET VENTILSTUECK,VOLLST. R23694


REP.SET VALVE PIECE

16 KRATEREINSATZ R50521
CRATER INSERT

17 KIPPHEBEL R27999
LEVER, CPL.

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 41 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
18 FEDER R28133 10
SPRING

19 GEHAEUSE FUER PA80 R50522


HOUSING FOR PA 80

20 RUNDSCHNURRING R27810 10
TOROIDAL SEALING RING

21 TUELLE R50479
SOCKET

22 UEBERWURFMUTTER R50478
UNION NUT

23 SCHLAUCHKLEMME R20476
HOSE CLAMP

24-30 DRUCKGASANSCHLUSS,KPL.UGR. NG 8408303


COMPRESSED GAS CONNECTION

24 SCHLAUCH 6X3 - PVC 1181106


HOSE 6X3 - PVC

25 FASSUNG SCHLAUCHHUELSE M26950 10


HOLDER FOR HOSE SOCKET

26 UEBERWURFMUTTER NG M19098
UNION NUT

27 SCHLAUCHTUELLE 8408293
HOSE SOCKET

28 DICHTRING M05128 20
PACKING RING

29,30 DRUCKGASANSCHLUSS, UGR. 8408301


COMPRESSED GAS CONNECTION

30 O-RING 2M12839 10
O-RING SEAL

31 O-RING R20771 10
O-RING SEAL

32 SCHLAUCH 12X3 1198912


HOSE 12X3

33,34 PATIENTENGASANSCHLUSS NG 8407893


PATIENT GAS SUPPLY

34 DICHTRING 2M01524 10
PACKING RING

35 KABELBINDER (2,4X92LG) 8712007 25


CABLE CLIP (2,4X92)

36 DICHTRING R22078 10
WASHER

38 GEWINDEKONUS 8407832
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 42 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
THREADED CONE

39-42 BEATMUNGSVENTIL M.R-VENTIL 8408568


RESPIRATION VALVE

39 DECKEL 8406585
COVER

40,41 MEMBRAN,UGR. 8403552 2


DIAPHRAGM

40 MEMBRANE 8403206
DIAPHRAGM

41 MEMBRANE 8403395
DIAPHRAGM

42 VENTILGEHAEUSE, KOMPL. 8408569


VALVE BOX, CPL.

43 DRAEGER-MASKE GR.3-ISO M25574


DRAEGER MASK SIZE 3-ISO

44 SCHELLE M21610 10
CLIP

45 ANSCHLUSS M27691
CONNECTION

46 VERBINDUNGSSCHLAUCH -SILIKON- 8404063


SILICONE HOSE ADULT 1,1M

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 43 von 44
OXYLOG (SAMSOMATIC)

TRAGEPLATTE MIT OXYLOGHALTER Bild/Picture 17


CARRYING PLATE WITH OXYLOG SUPPORT

Datei/File: 550314.017
Druckdatum: 19.04.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division  
Ersatzartikelliste 5503.14 Ausgabe/Edition
Spare parts list 04.06.96
OXYLOG (SAMSOMATIK) Seite/Page 44 von 44
OXYLOG (SAMSOMATIC)

Position Benennung Sach-Nr. Bestell-Nr. Packung


Item No. Description Part No. Order-Code Quantity
1 TRAGEPLATTE MIT OXYLOGHALTER 8412148
CARRYING PLATE W. OXYLOG SUPP.

2 VERSCHLUSSZAPFEN NG AB40013
LOCKING TAPPET

3 HALTESCHEIBE NG AB40028
WASHER

4 TRAGEGURT, KOMPL. 8412151


NECK STRAP, CPL.

5 GURT (FLASCHE) 8412150


STRAP (CYLINDER)

6 FLASCHENADAPTER(F.1,5L-FL.) NG 8412152
ADAPTER (FOR CYLINDER 1,5L)

7 GURT OXYLOG 8412149


STRAP FOR OXYLOG

Druckdatum: 19.04.1999 Änderungen vorbehalten


Printed Subject to change without notice
Dräger Medical AG & Co. KGaA

Moislinger Allee 53 – 55
D-23542 Lübeck
Federal Republic of Germany

T (++49)451/ 882 - 3127


FAX (++49)451/ 882 - 4294

5503.140

2nd Edition Sptember 1997

Subject to modification.
No replacement in event of modification.