Eppendorf Reference® 2
Eppendorf® and the Eppendorf Brand Design are registered trademarks of Eppendorf AG, Germany.
Registered trademarks and protected trademarks are not marked in all cases with ® or ™ in this manual.
Table of contents
1 Operating instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1 Using this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Danger symbols and danger levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.1 Danger symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.2 Danger levels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Purpose and scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.4.1 Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.4.2 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.5 Version overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Product overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.1 Pipette models. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3 Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 User profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Information on liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1 Setting the volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.2 Attaching the pipette tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.3 Optimum immersion depths . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.4 Forward pipetting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.4.1 Aspirating liquid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.4.2 Dispensing liquid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.4.3 Ejecting the pipette tip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.5 Reverse pipetting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.5.1 Aspirating liquid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.5.2 Dispensing liquid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.5.3 Ejecting the pipette tip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.6 Storing the pipette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
5 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.1 Control button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.2 Dispensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.3 Pipette tip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.4 Tip cone. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6 Disassembly/assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.1 Disassembling the single-channel pipette lower part up to 1000 μL . . . . . . . . . . . . . . . . . . . . . 19
6.2 Assembling the single-channel pipette lower part up to 1000 μL . . . . . . . . . . . . . . . . . . . . . . . 20
6.3 Disassembling the lower part of the single-channel pipette from 2 mL upward . . . . . . . . . . . . 21
6.4 Assembling the lower part of the single-channel pipette from 2 mL upward . . . . . . . . . . . . . . 22
6.5 Disassembling the lower part of the multi-channel pipette . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.6 Assemble the lower part of the multi-channel pipette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.7 Removing the upper part . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.8 Installing the upper part . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Table of contents
4/72 Eppendorf Reference® 2
English (EN)
7 Alignment/adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.1 Adjusting the pipette. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.1.1 General information on user adjustment and the factory adjustment . . . . . . . . . . . . . 31
7.1.2 Changing the user adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
7.1.3 Changing the factory adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.1.4 Reference 2 - Changing the factory adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.2 Changing the volume for changed user adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7.2.1 Changing volume for single-channel pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7.2.2 Volume change for multi-channel pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
8 Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
9 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
9.1 Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
9.1.1 Single-channel pipettes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
9.1.2 Multi-channel pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
9.2 Check after repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
9.2.1 Checks overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
9.2.2 Perform checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
9.3 Shipping the device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
10 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
10.1 Increments of the single-channel and multi-channel pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . 45
10.2 Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
10.3 Single-channel pipette with fixed volume. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
10.4 Single-channel pipette with variable volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
10.5 Multi-channel pipette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
10.6 Errors of measurement in accordance with ISO 8655-2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
10.7 Test conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
11 Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
11.1 Single-channel lower parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
11.1.1 Ejector sleeves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
11.1.2 Piston, piston spring and piston mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
11.1.3 Protective filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
11.2 Multi-channel lower parts with fixed cone spacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
11.2.1 O-rings and replacement channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
11.3 Spare parts for single-channel pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
11.3.1 Ejector sleeves for single-channel pipettes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
11.3.2 Additional spare parts for single-channel pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
11.4 Spare parts for multi-channel pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
11.5 Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
11.6 Service spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
11.6.1 Adjustment plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
11.6.2 Lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
11.6.3 Stroke system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
11.6.4 Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
11.6.5 Housing case, multi-channel lower part . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
11.6.6 Housing lid, multi-channel lower part . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
11.6.7 O-ring tip cone, multi-channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Table of contents
Eppendorf Reference® 2 5/72
English (EN)
12 Technical information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Table of contents
6/72 Eppendorf Reference® 2
English (EN)
Operating instructions
Eppendorf Reference® 2 7/72
English (EN)
1 Operating instructions
1.1 Using this manual
Make sure that the service manual and the operating manual are available in the latest versions. To do
so, compare the version numbers. The latest versions of the manuals can be downloaded from
www.eppendorf-support.com.
Read the service manual before commencing work on the device.
Read the chapters "Installation" and "Operation" of the operating manual.
Observe the safety instructions in the operating manual.
The safety precautions in these instructions have the following danger symbols and danger levels.
Depiction Meaning
1. Actions in the specified order
2.
Actions without a specified order
• List
Direction of movement
Text Display or software texts
Additional information
Operating instructions
8/72 Eppendorf Reference® 2
English (EN)
This document describes all prerequisites and actions necessary to perform installation, diagnosis, repair
or maintenance of the product named on the title page.
1.4.2 Scope
This document applies to all service organizations of Eppendorf AG, service providers commissioned by
Eppendorf AG, as well as certified Eppendorf AG service technicians.
2 Product description
2.1 Product overview
Abb. 2-1: Single-channel pipette and multi-channel pipette
1
2
3
4
17
Reference 2
5 AD
J
Reference 2
AD
J
6
7
000
0000
000
000
0
8
9
10 14
18
15
11 19
16
12
13
2.2 Features
The Reference 2 pipette is a piston-stroke pipette for aspirating and dispensing liquids. The pipette
operates according to the air cushion principle. A suitable pipette tip must be fitted prior to use. The
dispensing, volume setting ring and ejector are controlled with the control button. Depending on the
model, volumes between 0.1 μL and 10 mL can be dispensed.
3 Safety
3.1 User profile
The specialist entrusted with maintenance, repair or testing must meet the following prerequisites:
• Successful participation in service trainings with certification by Eppendorf AG for the product that is to
be maintained, repaired or tested.
• Qualification as a specialist with knowledge of the applicable local and international standards.
• Qualification to evaluate the work assigned to him.
• Recognition of potential hazards and how to prevent or correct them.
The Eppendorf service organization responsible for servicing, repair, or testing is liable for ensuring that all
work is performed in a professional manner.
The constructive condition of the device must not be altered as a result of repair, servicing, or testing.
The safety of the original condition must be maintained.
Only accessories and original spare parts recommended by Eppendorf as well as test and measuring
equipment recommended by Eppendorf may be used for servicing, repair, and testing.
When handling infectious liquids and pathogenic germs, observe the national regulations,
the biosafety level of your laboratory, and the manufacturers' Safety Data Sheets and
application notes.
Wear your personal protective equipment.
For comprehensive regulations about handling germs or biological material of risk group II
or higher, please refer to the "Laboratory Biosafety Manual" (source: World Health
Organization, Laboratory Biosafety Manual, in the currently valid version).
Find out more about the risk of contamination before starting work.
Check the device decontamination certificate of the device.
Work on decontaminated devices only.
Wear your personal protective equipment (protective gloves, protective goggles).
Safety
12/72 Eppendorf Reference® 2
English (EN)
4 Operation
4.1 Setting the volume
Set the volume from a high to a low value. If necessary, exceed the desired volume and then
go back.
1. Keep the volume catch pressed.
2. Turn the control button.
0 0 0 0
-
The pipette tip can be attached by hand or mounted directly from a tip holder (tray) with the pipette.
The control button and the trays are color-coded. The color marks the pipette size and the
volume of the pipette tip (epT.I.P.S.).
1. Insert the tip cone into the pipette tip using light
force.
Operation
14/72 Eppendorf Reference® 2
English (EN)
Prerequisites
• The pipette tip has been attached.
To ensure maximum precision and accuracy, we recommend initially wetting each new tip by
aspirating and dispensing the liquid one to three times.
If necessary, wipe the pipette tip against the tube inner wall.
An additional volume (blow out) is aspirated when performing reverse pipetting. This can improve
dispensing results for viscous or foaming liquids. Volume limits may occur when using filter tips.
If necessary, wipe the pipette tip against the tube inner wall.
1. Place the pipette tip on the tube inner wall at a steep angle.
2. Slowly press down the control button to the first stop.
The liquid is dispensed.
3. Wait until the flow of liquid stops.
4. Hold down the control button and wipe the pipette tip on the tube inner wall.
Residual liquid remains in the pipette tip.
During dispensing, the additional volume (blow out) does not belong to the dispensing volume.
The pipette can be stored in a pipette carousel, a wall mount or in horizontal position.
Operation
16/72 Eppendorf Reference® 2
English (EN)
Troubleshooting
Eppendorf Reference® 2 17/72
English (EN)
5 Troubleshooting
5.1 Control button
5.2 Dispensing
6 Disassembly/assembly
6.1 Disassembling the single-channel pipette lower part up to 1000 μL
PUSH TO RELEASE
1 2 3 4
4
Disassembly/assembly
20/72 Eppendorf Reference® 2
English (EN)
1. Carefully guide the piston and the piston spring into the cylinder.
1 Make sure that the piston is guided correctly in the piston spring
and the cylinder. There must not be any perceptible resistance.
Stop pushing as soon as any resistance is felt. This may indicate
2
that the piston is not positioned correctly in the cylinder. There is
a risk of bending the piston if too much force is applied. Carefully
pull out the piston and repeat the process correctly. In the case of
piston springs with double windings (4), these windings must
3 point downwards.
2. Keep the piston and the piston spring pressed.
3. Keep the stop pins on the piston mount pressed with the other
4 hand.
4. Mount the piston mount so that the two locking mechanisms
engage in their holders.
5. Gently press a pipette tip against the inserted piston. The piston
must move down in the cylinder without any noticeable
resistance.
6. Insert the pre-assembled lower part of the pipette into the upper
part until it audibly engages.
7. Place the ejector sleeve on the ejector rod. If correctly fitted, it
will engage slightly.
Disassembly/assembly
Eppendorf Reference® 2 21/72
English (EN)
Referen
Referen
ADJ ADJ
0000
000
0000
000 1
1 2 3
1 2 3 4 5 6
1. Insert piston mount, piston spring and piston into the upper part of the lower part.
The further assembly corresponds to the figures indicated in chapter "Disassembling the lower part of the
single-channel pipette from 2 mL upward". Carry out the procedure in reverse order.
10
3
11 12
9
4
8
5
6
7
1 2 3
1. To unlock the lower part, slide the lever to the left or right.Remove the lower part of the pipette. (fig. 1 +
2)
2. Hold the lower part in such a way that the lever is facing downwards.
3. Slide the left and right latches upward. (Fig. 3)
4. Remove the housing lid with integrated ejector rail.
1 2 3 4 5
Remove the piston from the piston attachment (fig. 1 + 2) and remove the tip cone from the housing case
(fig. 3 - 5)
Remove all tip cones to continue with the disassembly of the housing case.
3. Press the piston mount and turn it 90° to the left or right. Remove the piston mount.
10
3
11 12
9
4
8
5
6
7
3. Push up the piston attachment. Assemble the piston spring and fix it onto the piston mount.
4. Insert all tip cones
5. Insert all pistons.
First, insert the piston in the tip cone. Then insert the piston in the piston attachment.
6. Attach the housing lid. To fix the housing lid, insert the right and left latch and push them up for
locking.
Gravimetric testing is required after opening the upper part of the Reference 2.
Prerequisites
The pipette lower part has been removed.
Required tools
Required tools
• Assembly tool
Gravimetric testing is required after opening the upper part of the Reference 2.
Alignment/adjustment
Eppendorf Reference® 2 31/72
English (EN)
7 Alignment/adjustment
7.1 Adjusting the pipette
NOTICE! Incorrect dispensing volume with special liquids and due to temperature
differences.
Solutions with physical data very different to that of water or temperature differences between
the pipette, pipette tip and liquid may result in incorrect dispensing volumes.
Avoid temperature differences between the pipette, pipette tip and liquid.
The random and systematic errors recorded upon delivery can be found on the Eppendorf
Certificate. This certificate is included in the delivery. If the factory adjustment is changed, the
certificate becomes invalid.
The Reference 2 was adjusted and checked before delivery. The adjustment opening is sealed with a gray
plastic adjustment seal. The adjustment seal bears the abbreviation "ADJ". The adjustment display shows
"0".
Changing the adjustment of the Reference 2 is recommended for solutions whose density, viscosity, surface
tension and/or vapor pressure differs greatly from that of water. If the density of an aqueous solution
changes by approx. ±10 %, for example due to its salt concentration, its volume changes by approx.
±10 %. This does not apply if other relevant properties of the liquid also change.
If the location the pipette is used at is at an extreme altitude, correction due to atmospheric pressure is
required. At 1 000 m above sea level, a volume error of approx. -0.3 % occurs for a 100 μL pipette.
If special tips with shapes different from standard tips are used, changing the adjustment can improve
dispensing accuracy (systematic bias).
Observe the recommendations for adjusting epT.I.P.S. long in the User adjustment on the Reference 2 CD.
Aside from changes to the user adjustment, a Reference 2 can be changed permanently by changing the
factory adjustment.
Changes to the user adjustment or to the factory adjustment have no influence on dispensing precision
(random error). Precision can be improved by replacing worn parts. Precision is also greatly influenced by
the handling of the device.
Before changing the adjustment of factory calibration, check the existing dispensing rate.
Alignment/adjustment
32/72 Eppendorf Reference® 2
English (EN)
At 20 °C, the density of distilled water is approx. 0.9982 mg/μL, and 0.9965 mg/μL at 27 °C.
If there is a difference between the actual volume and the set volume of distilled water, please check the
following:
Only if all questions can be answered with "Yes" can adjustment be performed. In any other case, the
problems corresponding to the questions answered with "No" must first be remedied. If replacing the
complete lower part or other volume-determining parts, proper assembly must be gravimetrically checked.
The systematic and random errors to be fulfilled can be found in the technical data.
Alignment/adjustment
Eppendorf Reference® 2 33/72
English (EN)
If the adjustment is changed, the volume is changed by a specific value. Strictly speaking, this change only
applies for the testing volume.
Example:
You are adjusting a 10 – 100 μL pipette by 1 μL at a volume setting of 100 μL (1 μL ≙ 1 %). The pipette is
also adjusted by 1 μL at a volume setting of 10 μL (≙ 10 %).
Auxiliary equipment
• Pin
• Gray adjustment seal (ADJ) made of plastic
1. Stab into the center of the gray adjustment seal (ADJ) with the pin.
2. Remove the adjustment seal.
3. Insert the adjustment tool.
4. Rotate the adjustment tool until the adjustment display shows the desired value.
5. Read the set value without distortion via the bearing aid in the viewing window.
6. Weight to check accuracy and precision.
1000 µL
ADJ
1
- 1
+
If the adjustment only applies for a specific liquid, mark the pipette correspondingly. Use the labeling area
on the pipette to note the liquid and the volume. Gravimetrically check any change to the adjustment.
Observe the test specifications in EN ISO 8655-2 and 8655-6. A Standard Operation Procedure (SOP) and
more information on user adjustment and factory adjustment can be found on the CD or our website
www.eppendorf.com.
Alignment/adjustment
34/72 Eppendorf Reference® 2
English (EN)
Auxiliary equipment
• Gray safety plug made of plastic
• Pin
AD
J
Refe
The random and systematic errors recorded upon delivery can be found on the Eppendorf Certificate. This
certificate is included in the delivery. If the factory adjustment is changed, the certificate becomes invalid.
If the dispensing accuracy is only changed temporarily, changing the user adjustment is the correct
method. Before changing the user adjustment or the factory adjustment, observe the general information
provided (see General information on user adjustment and the factory adjustment on p. 31) and the
corresponding gravimetric checks.
The volume display is set to correspond to the piston stroke and the detected actual volume during factory
adjustment. If a deviation to be corrected is detected during the gravimetric check and factory adjustment
is required, proceed as follows:
ADJ
1
- 1
+
Alignment/adjustment
Eppendorf Reference® 2 35/72
English (EN)
2. Pierce the safety plug with the pin and remove it.
AD
J
Refe
AD
J
Re
AD
J
Refere
The volume values (Δ V) are theoretical values and used only for orientation. They apply for any volume set
for pipettes with variable volume setting. Deviations to the values may result depending on the operation
and other conditions (temperature, density, etc.).
2.5 μL
5 μL medium gray 0.2 0.15 0.1 0.05 -0.05 -0.1 -0.15 -0.2
10 μL
20 μL light gray 0.4 0.3 0.2 0.1 -0.1 -0.2 -0.3 -0.4
10 μL yellow 0.4 0.3 0.2 0.1 -0.1 -0.2 -0.3 -0.4
20 μL
25 μL yellow 2 1,5 1 0.5 -0.5 -1 -1.5 -2
50 μL
100 μL
200 μL yellow 4 3 2 1 -1 -2 -3 -4
300 μL orange 6 4.5 3 1.5 -1.5 -3 -4.5 -6
200 μL blue 20 15 10 5 -5 -10 -15 -20
250 μL
500 μL
1000 μL
2 mL red 50 37.5 25 12.5 -12.5 -25 -37.5 -50
2.5 mL
5 mL violet 100 75 50 25 -25 -50 -75 -100
10 mL turquoise 200 150 100 50 -50 -100 -150 -200
Alignment/adjustment
38/72 Eppendorf Reference® 2
English (EN)
8 Diagrams
k
an
bl
ft
le
lly
na
io
nt
te
In
Diagrams
40/72 Eppendorf Reference® 2
English (EN)
Maintenance
Eppendorf Reference® 2 41/72
English (EN)
9 Maintenance
9.1 Service
For instructions on how to disassemble and assemble the pipette, read the corresponding
chapter "Disassembly/assembly".
If the O-rings cannot be cut, the tool is worn and must be replaced.
2. Check if the tip cones are scratched. Replace scratched tip cones.
3. Place the new O-rings on the tip cones.
4. Check the piston along with the seal. Replace defective pistons with seals.
5. Clean the piston using an alcohol-based solution, then lubricate the piston.
6. Use the pipette tester to check each channel for leakage.
7. Assemble the pipette. Set the nominal volume. Check to see if the piston runs smoothly and evenly by
slowly pressing the control button several times.
Maintenance
42/72 Eppendorf Reference® 2
English (EN)
Transport packing can be ordered via your Eppendorf service organization. The order number
can be found in the chapter Ordering information.
Maintenance
44/72 Eppendorf Reference® 2
English (EN)
Technical data
Eppendorf Reference® 2 45/72
English (EN)
10 Technical data
10.1 Increments of the single-channel and multi-channel pipettes
Model Increment
0.1 μL – 2.5 μL 0.002 μL
0.5 μL – 10 μL 0.01 μL
2 μL – 20 μL 0.02 μL
2 μL – 20 μL 0.02 μL
10 μL – 100 μL 0.1 μL
20 μL – 200 μL 0.2 μL
30 μL – 300 μL 0.2 μL
100 μL – 1000 μL 1 μL
0.25 mL – 2.5 mL 0.002 mL
0.5 mL – 5 mL 0.005 mL
1 mL – 10 mL 0.01 mL
Test conditions and test evaluation in compliance with ISO 8655, Part 6. Tested using a standardized
analytical balance with evaporation protection.
The three largest testing volumes per tip (10 %, 50 %, 100 % of the nominal volume)
correspond to the specifications in accordance with ISO 8655, Part 2 or Part 5. The test is to
be carried out with these three testing volumes for testing of the systematic and random error
in compliance with the standard. The smallest adjustable volume serves to provide additional
information.
• Number of determinations per volume: 10
• Water according to ISO 3696
• Inspection at 20 °C – 27 °C
Maximum temperature variation during measurement ±0.5 °C
• Dispensing onto the tube inner wall
Technical data
50/72 Eppendorf Reference® 2
English (EN)
Ordering information
Eppendorf Reference® 2 51/72
English (EN)
11 Ordering information
11.1 Single-channel lower parts
Abb. 11-1: Lower parts for single-channel pipettes
2
4
1 2
2
1 4
5
1 2 3
2 3
9 mm
3
4
5
1 3
Fig. 11-9: Example: Replacement channel for 10 μL, 100 μL und 300 μL multi-channel lower part
11.5 Tools
Abb. 11-10: IMAGE CAPTION ??? Tools for single- and multi-channel pipettes
1 2 3 4 5 6
1
6
ADJ
1 2
11.6.2 Lid
0 +1 +2
11.6.4 Counter
1 O-ring
1 2
1 Sealing washer
12 Technical information
Technical information
70/72 Eppendorf Reference® 2