1.UT Coursenote
1.UT Coursenote
NDT4
Contents
Section Subject
Preliminary pages
Contents
Standards and Associated Reading
COSHH, H&S, Cautions and Warnings
Introduction to NDT Methods
NDT Certification Schemes
1 Physical Principles
1.1 Frequency
1.2 Wavelength
1.3 Resolution
1.4 Signal amplitude
2 Modes of Sound Energy
2.1 Compressional waves
2.2 Shear waves
2.3 Rayleigh or surface waves
2.4 Lamb waves
3 Generating Ultrasound
3.1 Piezo-electric crystals
3.2 Quartz or silicon oxide (SiO2)
3.3 Lithium sulphide (Li2SO4)
3.4 Barium titanate (BaTiO3)
3.5 Lead metaniobate (PbNb2O6)
3.6 Lead zirconate titanate (PbZrO3, PbTiO3)
3.7 Electromagnetic acoustic transducers
4 Pulse Length and Damping
5 The Sound Beam
5.1 Near zone (or near field)
5.2 Far zone
6 Total Attenuation Loss
6.1 Attenuation due to beam spread
7 Acoustic Impedance
8 Snell's Law
8.1 Critical angle calculation
8.2 Mode conversion
9 Probe Design
9.1 Compression wave probes
9.2 Angle probes
9.3 Twin crystal probes
9.4 Other probe types
NDT4-51015
Contents Copyright © TWI Ltd
10 Test Techniques
10.1 Pulse echo
10.2 Through-transmission
10.3 Tandem scanning
10.4 Contact scanning
10.5 Gap scanning
10.6 Immersion testing
10.7 Presentation
11 Ultrasonic Flaw Detector
11.1 Principles
11.2 Cathode ray tube
11.3 Pulse generation
11.4 Range control
11.5 Delay
11.6 Calibrated gain/attenuator control
11.7 Reject/suppressor control
11.8 The decibel
12 Calibration and Sensitivity
12.1 Finding the probe index
12.2 Checking the probe angle
12.3 Calibration of shear waves for range
13 Flaw Location
14 Flaw sizing
14.1 The 6dB drop sizing method
14.2 The 20dB drop sizing metho
14.3 Construction of a beam edge plot - 20dB
14.4 Constructing an angle beam plot
14.5 Proving the beam plot
14.6 Modified near zone angle probes
14.7 Horizontal beam plot
15 Sensitivity Setting
15.1 The Institute of Welding (IOW) block
15.2 Distance amplitude correction (DAC) curves
15.3 Flat-bottomed holes (FBH)
15.4 Using noise
15.5 Transfer correction
15.6 The distance gain size (DGS) method
15.7 Signal-to-noise ratio (SNR)
16 Ultrasonic Equipment Checks
16.1 Linearity of time base
16.2 Linearity of equipment gain
16.3 Probe index and beam alignment
16.4 Beam angle
16.5 Sensitivity and signal-to-noise ratio
16.6 Pulse duration
16.7 Resolving power (resolution)
NDT4-51015
Contents Copyright © TWI Ltd
17 Practical Weld Inspection
17.1 Identifying flaws in butt welds
18 Root Flaws
18.1 Excess penetration
18.2 Root concavity
18.3 Root crack
18.4 Lack of root fusion
18.5 Misalignment
19 Face and Body Flaws
19.1 Lack of fusion
19.2 Crack
19.3 Gas pore
19.4 Porosity
19.5 Linear inclusion (slag)
20 Plate Inspection
21 Inspection Procedure
21.1 Information required prior to testing
21.2 Compression scan
21.3 Root scan
21.4 Weld scan
21.5 Transverse scan
21.6 Double V welds
21.7 Pipes
21.8 Reporting
22 Ultrasonic Thickness Measurement
22.1 Measurement modes
22.2 Requirements – test object, instruments, probes and reference blocks
Glossary
Product Technology Notes
NDT4-51015
Contents Copyright © TWI Ltd
Preface
These notes are provided as training reference material and to meet the study
requirements for examination on the NDT course to which they relate.
They do not form an authoritative document, nor should they be used as a reference for
NDT inspection or used as the basis for decision making on NDT matters. The standards
listed are correct at time of printing and should be consulted for technical matters.
NOTE: These training notes are not subject to amendment after issue.
NTD4-51015
Prelims Copyright © TWI Ltd
Standards and Associated Reading
NDT4-51015
Prelims 1 Copyright © TWI Ltd
BS EN 12668-3 Non-destructive testing – Characterization and verification
of ultrasonic examination equipment.
Part 3: Combined equipment
Associated Reading
NDT Ed.org – Introduction to ultrasonic testing
http://www.ndt-
ed.org/EducationResources/CommunityCollege/Ultrasonics/cc_ut_index.htm
NDT4-51015
Prelims 2 Copyright © TWI Ltd
COSHH, H&S, Caution and Warnings Relevant to TWI Training & Examination
Services
Introduction
The use of chemicals in NDT is regulated by law under the Control of Substances
Hazardous to Health (COSHH) Regulations 2005. These regulations require the School to
assess and control the risk of health damage from every kind of substance used in
training. Students are also required by the law to co-operate with the School’s risk
management efforts and to comply with the Control Measures adopted.
The School holds Manufacturers Safety Data Sheets for every substance in use. Copies
are readily available for students to read before using any product. The Data Sheets
contain information on:
What is Exposure?
Exposure to a substance is uptake into the body. The exposure routes are:
Many thousands of substances are used at work but only about 500 substances have
workplace exposure limits (WELs). Until 2005 it had been normal for HSE to publish a
new edition of EH40, or at least an amendment, each year. However, with increasing use
of the website facilities, the HSE no longer always publishes a revised hardcopy edition
or amendment.
The web-based list applicable from 1 October 2007 can now be found at
http://www.hse.gov.uk/coshh/table1.pdf
NDT4-51015
Prelims 3 Copyright © TWI Ltd
Safety and Environmental Requirements
Ultrasonic testing requires the use of couplant and cleaning fluids, some of which may be
hazardous to health. Extended or repeated contact of such materials with the skin or
mucous membranes shall be avoided.
NDT4-51015
Prelims 4 Copyright © TWI Ltd
Introduction to Non-Destructive Testing
Non-destructive testing (NDT) is the ability to examine a material (usually for
discontinuities) without degrading it or permanently altering the article being tested, as
opposed to destructive testing which renders the product virtually useless after testing.
Other advantages of NDT over destructive testing are that every item can be examined
with no adverse consequences, materials can be examined for conditions internally and
at the surface and, most importantly, parts can be examined whilst in service, giving a
good balance between cost effectiveness and quality control. NDT is used in almost
every industry with the majority of applications coming from the aerospace, power
generation, automotive, rail, oil & gas, petrochemical and pipeline markets, safety being
the main priority of these industries. When properly applied, NDT saves money, time,
materials and lives. NDT as it is known today has been developing since around the
1920s, with the methods used today taking shape later and vast technological
advancements being made during the Second World War. The basic principal methods
are:
In all NDT methods, the interpretation of results is critical. Much depends on the skill and
experience of the technician, although properly formulated test techniques and
procedures will improve accuracy and consistency.
Visual testing begins with the eye; however, the first boroscopes used a hollow tube and
a mirror with a small lamp at the end to investigate the bores of rifles and cannons for
problems and discontinuities. In the 1950s, the lamps were replaced by glass fibre
bundles which were used to transmit the light. These became known as fibrescopes
which were also less rigid, increasing the capabilities of testing. With usage expanding,
many users began to suffer from eye fatigue which led to the development of video
technology. This was first used in the 1970s and relies on electronics to transmit the
images rather than fibreoptics.
Further enhancements to video technology include pan, tilt and zoom lenses, and
mounting cameras to platforms and wheels, all allowing more parts to be tested and
better images for improved inspection. Video devices also allow recordings of inspections
to be taken, meaning permanent records can be kept. This has a number of advantages
such as enabling other inspectors to observe the test as it was performed and allowing
further review and evaluation.
NDT4-51015
Prelims 5 Copyright © TWI Ltd
Advantages Disadvantages
Able to test large parts with a portable kit Careful surface preparation required
A different (though related) method was introduced in the 1940s. The surface under
examination was coated with a lacquer, and after drying, the sample was caused to
vibrate by the tap of a hammer. The vibration causes
the brittle lacquer layer to crack generally around
surface defects. The brittle lacquer (stress coat) has
been used primarily to show the distribution of
stresses in a part and not for finding defects.
NDT4-51015
Prelims 6 Copyright © TWI Ltd
Magnetic particle testing (MT)
Magnetic particle testing is used to locate surface and slightly sub-surface discontinuities
in ferromagnetic materials by introducing a magnetic flux into the material.
Advantages Disadvantages
This form of NDT became much more common after the First World War, in the 1920s,
when William Hoke discovered that flaws in magnetised materials created distortions in
the magnetic field. When a fine ferromagnetic powder was applied to the parts, it was
observed that they built up around the defects, providing a visible indication of their
location.
Magnetic particle testing superseded the oil and chalk method in the 1930s as it proved
far more sensitive to surface breaking flaws. Today it is still preferred to the penetrant
method on ferromagnetic material and much of the equipment being used then is very
similar to that of today, with the only advances coming in the form of fluorescent coating
to increase the visibility of indications and more portable devices being used. In the early
days, battery packs and direct current were the norm and it was some years before
alternating current proved acceptable.
Magnetism
The phenomenon called magnetism is said to have been discovered in the ancient Greek
city of Magnesia, where naturally occurring magnets were found to attract iron.
The use of magnets in navigation goes back to Viking times or maybe earlier, where it
was found that rods of magnetised material, when freely suspended, would always point
in a north-south direction. The end of the rod which pointed towards the North Pole star
became known as the North Pole and consequently the other end became the South
Pole.
Hans Christian Oersted (1777-1851) discovered the connection between electricity and
magnetism, followed by Michael Faraday (1791-1867), whose experiments revealed that
magnetic and electrical energy could be interchanged.
NDT4-51015
Prelims 7 Copyright © TWI Ltd
Historical perspective
Electromagnetic testing – the interaction of magnetic fields with circulating electrical
currents - had its origin in 1831 when Michael Faraday discovered electromagnetic
induction. He induced current flow in a secondary coil by switching a battery on and off.
D E Hughes performed the first recorded eddy current test in 1879. He was able to
distinguish between different metals by noting a change in excitation frequency resulting
from effects of test material resistivity and magnetic permeability.
Advantages Disadvantages
Can detect through surface coatings Will not detect defects parallel to surface
Accurate conductivity measurements Not suitable for large areas and/or complex
geometries
Portability
NDT4-51015
Prelims 8 Copyright © TWI Ltd
History of eddy current testing
The principles of eddy currents arose in 1831 with Faraday’s discovery of
electromagnetic induction; eddy current testing methods have their origins in a period
just after the First World War, when materials with a high magnetic permeability were
being developed for electrical power transformer cores and motor armatures. Eddy
currents are a considerable nuisance in electrical engineering – they dissipate heat and
efforts to reduce their effect led to a discovery that they could be used to detect material
changes and cracks in magnetic materials. The first eddy current testing devices for NDT
were in 1879 by Hughes, who used the principles of eddy currents to conduct
metallurgical sorting tests and the stray flux tube and bar tests.
It was left to Dr Friedrich Förster in the late 1940s to develop the modern day eddy
current testing equipment and formulate the theories which govern their use. The
introduction by Förster of sophisticated, stable, quantitative test equipment and of
practical methods for analysis of quantitative test signals on the complex plane was by
far the most important factor contributing to the rapid development and acceptance of
electromagnetic induction and eddy current testing. Förster is rightly identified as the
father of modern eddy current testing.
By 1950, he had developed a precise theory for many basic types of eddy current tests,
including both absolute and differential or comparator test systems and probe or fork coil
systems used with thin sheets and extended surfaces.
Applications for microcomputer chips abound, from giving lift-off suppression in simple
crack detection to providing signal processing for immediate analysis of condenser tube
inspection. As with other testing methods, improvements to the equipment have been
made to increase its portability and computer-based systems now allow easy data
manipulation and signal processing. Eddy current testing is now a widely used and
understood inspection method for flaw detection as well as for thickness and conductivity
measurements.
NDT4-51015
Prelims 9 Copyright © TWI Ltd
Advantages Disadvantages
The origin of modern ultrasonic testing (UT) is the discovery by the Curie brothers in
1880 that quartz crystals cut in a certain way produce an electric potential when
subjected to pressure - the piezo-electric effect, from the Greek piedzein (to press or
strike). In 1881 Lippman theorised that the effect might work in reverse, and that quartz
crystals might change shape if an electric current was applied to them. He found that
this was so and experimented further. Crystals of quartz vibrate when alternating
currents are applied to them. Crystal microphones in a modern stereo rely on this
principle.
When the Titanic sank in 1912, the Admiralty tried to find a way of locating icebergs by
sending out sound waves and listening for an echo. They experimented further with
sound to detect submarines during the First World War. Between the wars, marine echo
sounding was developed and in the Second World War ASDIC (Anti-Submarine Detection
Investigation Committee) was extensively used in the Battle of the Atlantic against the
U-boats.
In the years after the Second World War, researchers in Japan began to experiment on
the use of ultrasound for medical diagnostic purposes. Working largely in isolation until
the 1950s, the Japanese developed techniques for the detection of gallstones, breast
masses, and tumours. Japan was also the first country to apply Doppler ultrasound, an
application of ultrasound that detects internal moving objects such as blood coursing
through the heart for cardiovascular investigation.
The first flaw detector was made by Sproule in 1942 while he was working for the
Scottish firm Kelvin & Hughes. Similar work was carried out by Firestone in the USA and
by German physicists. Sproule went on to develop the shear-wave probe.
NDT4-51015
Prelims 10 Copyright © TWI Ltd
Initially UT was limited to testing aircraft, but in the 1950s it was extensively used in the
building of power stations in Britain for examining thick steel components safely and
cheaply. UT was found to have several advantages over radiography in heavy industrial
applications:
No health hazards were associated with radiography, and a UT technician could work
next to welders and other employees without endangering them of holding up work.
It was efficient in detecting toe cracks in boilers – a major cause of explosions and
lack of fusion in boiler tubes.
It could find planar defects, like laminations, which were sometimes missed by
radiography.
A UT check on a thick component took no more time than a similar check on a thin
component as opposed to long exposure times in radiography.
Over the next twenty years, improvements focused on accurate detection and sizing of
the flaws with limited success, until 1977 when Silk first discovered an accurate
measurement and display of the top and bottom edges of a discontinuity with the time-
of-flight diffraction (TOFD) technique. Advances in computing technology have now
expanded the use of TOFD as real time analyses of results are now available.
It was also during the 1970s that industries focused on reducing the size and weight of
ultrasonic flaw detectors and making them more portable. This was achieved by using
semiconductor technology and during the 1990s microchips were introduced into the
devices to allow calibration parameters and signal traces to be stored. LCD display
panels and digital technology have also contributed to reducing the size and weight of
ultrasonic flaw detectors. With the development of ultrasonic phased array and increased
computing power, the future for ultrasonic inspection is very exciting.
In UT a sound pulse is sent into a solid object and an echo returns from any flaws in that
object or from the other side of the object. An echo is returned from a solid-air interface
or any solid-non-solid interface in the object being examined. We can send ultrasonic
pulses into material by making a piezo-electric crystal vibrate in a probe. The pulses can
travel in a compression, shear or transverse mode. This is the basis of ultrasonic testing.
However, the information from the returning echoes must be presented for
interpretation. It is for this purpose that the UT set, or flaw detector as it is frequently
called, contains a cathode ray tube.
In the majority of UT sets, the information is presented on the screen in a display called
the A Scan. The bottom of the CRT screen is a time base made to represent a distance -
say 100mm. An echo from the backwall comes up on the screen as a signal, the
amplitude of which represents the amount of sound returning to the probe. By seeing
how far the signal comes along the screen we can measure the thickness of the material
we are examining.
NDT4-51015
Prelims 11 Copyright © TWI Ltd
If that material contains a flaw, sound is reflected back from the flaw and appears on the
screen as a signal in front of the backwall echo (BWE) as the sound reflected from the
flaw has not had so far to travel as that from the backwall.
BWE
BWE
BWE
BWE
Defect
Defect
Ultrasonic signals
Anything that sends back sound energy to a probe to cause a signal on the screen is
called a reflector. By measuring the distance from the edge of the CRT screen to the
signal, we can calculate how far down in the material the reflector lies.
Advantages Disadvantages
NDT4-51015
Prelims 12 Copyright © TWI Ltd
History of radiographic testing
X-rays were discovered in 1895 by Wilhelm Conrad
Roentgen (1845-1923) who was a Professor at Wϋrzburg
University in Germany. Whilst performing experiments in
which he passed an electric current through a Crookes
tube (an evacuated glass tube with an anode and a
cathode), he found that when a high voltage was applied,
the tube produced a fluorescent glow. Roentgen noticed
that some nearby photographic plates became fogged.
This caused Roentgen to conclude that a new type of ray
was being emitted from the tube. He believed that
unknown rays were passing from the tube and through the
plates. He found that the new ray could pass through most
substances. Roentgen also discovered that the ray could
pass through the tissue of humans, but not bones and
metal objects. One of Roentgen's first experiments late in
1895 was a film of the hand of his wife.
While working in France at the time of Becquerel's discovery, Polish scientist Marie Curie
became very interested in his work. She suspected that a uranium ore known as pitch-
blende contained other radioactive elements. Marie and her husband, French scientist
Pierre Curie, started looking for these other elements. In 1898, the Curies discovered
another radio-active element in pitchblende, and named it polonium in honour of Marie’s
native homeland. Later that year, the Curies discovered another radioactive element
which they named ‘radium’, or shining element. Both polonium and radium were more
radioactive than uranium. Due to her lifelong research in this field, Marie Curie is widely
NDT4-51015
Prelims 13 Copyright © TWI Ltd
credited with the discovery of gamma radiation and the introduction of the new term:
radio-active.
Since these discoveries, many other radioactive elements have been discovered or
produced. Radiography in the form of NDT took shape in the early 1920s when H H
Lester began testing on different materials. Radium became the initial industrial gamma
ray source. The material allowed castings up to 10 to 12 inches thick to be radiographed.
During the Second World War, industrial radiography grew tremendously as part of the
Navy's shipbuilding programme. In 1946, man-made gamma ray sources from elements
such as cobalt and iridium became available. These new sources were far stronger than
radium and much less expensive. The man-made sources rapidly replaced radium, and
the use of gamma rays increased quickly in industrial radiography.
Direct radiography (DR) systems are also used based upon complementary metal oxide
sensor (CMOS) technology and TFT (thin film transistors). These systems have the
ability to directly convert light into digital format; additionally, they may be coupled with
a scintillator which coats CMOS and charged couple device (CCD) sensors. The
scintillator converts photon energy to light before the sensor and ADC converts to digital
format. Systems which use scintillators in this way are often referred to as indirect
systems.
Quality issues of any digital system are based upon the effective pixel size and the
signal-to-noise ratio (SNR). The benefits of using digital systems are the speed of
inspection and the absence of chemical processing requirements and wet film; however,
the initial equipment costs will be high.
NDT4-51015
Prelims 14 Copyright © TWI Ltd
NDT Certification Schemes
NDT4-51015
Prelims 15 Copyright © TWI Ltd
PCN – Personal Certification in Non-destructive testing
Managed and marketed by the British Institute of Non-Destructive
Testing (BINDT) which owns and operates the PCN Certification Scheme,
it offeres a UKAS accreditied certification of competence for NDT and
condition monitoring in a variety of product sectors.
Both schemes offer NDT certification conforming to BS EN ISO 9712; Qualification and
Certification of NDT personnel, this superseding EN473.
The points below cover extracts from this document which are major items, the full
document can be viewed on the BINDT website – www.bindt.org/certification/PCN.
NDT4-51015
Prelims 16 Copyright © TWI Ltd
References
PCN documents
PSL/4 Examination availability
PSL/8A PCN documents – issue status
PSL/30 Log of pre-certification experience
PSL/31 Use of PCN & UKAS logo
PSL/42 Log of pre-certification on-the-job training
PSL/44 Vision requirements
PSL/49 Examination exemptions for holders of certification other than PCN
PSL/51 Acceptable certification for persons supervising PCN candidates gaining
experience prior to certification
PSL/57C Application for certification, experience gained post examination
PSL/67 Supplementary 56 day waiver
PSL/70 Request for L2 certificate issue to a L3 holder
CP9 Requirements for BINDT authorised qualifying bodies
CP16 Renewal and recertification of PCN Levels 1 & 2 certificates
CP17 Renewal and recertification of PCN Level 3 certificates
CP19 Informal access to authorised qualifying bodies by third parties
CP22 Marking and grading PCN examinations
CP25 Guidelines for the preparation of NDT procedures and instructions in PCN
examinations
CP27 Code of ethics for PCN certificate holders
Set up equipment.
Carry out the test.
Record and classify the results in terms of written criteria.
Report the results.
Level 1 personnel have not demonstrated competence in the choice of test method or
technique to be used, nor for the assessment, characterisation or interpretation of test
results.
NDT4-51015
Prelims 17 Copyright © TWI Ltd
Level 2 personnel have demonstrated competence to perform and supervise non-
destructive testing according to established or recognised procedures. Within the scope
of the competence defined on the certificate, Level 2 personnel may be authorised by the
employer to:
Level 3 personnel are qualified to direct any NDT operation for which they are
certificated and may be authorised by the employer to:
Assume full responsibility for a test facility or examination centre and staff.
Establish, review for editorial and technical correctness and validate NDT instructions
and procedures.
Interpret codes, standards, specifications and procedures.
Designate the particular test methods, techniques and procedures to be used.
Within the scope and limitations of any certification held carry out all Level 1 and 2
duties and;
Provide guidance and supervision at all levels.
Level 3 certificated personnel may be authorised to carry out, manage and supervise
PCN qualification examinations on behalf of the British Institute of NDT.
Where Level 3 duties require the individual to apply routine NDT by a method(s) within a
particular product or industry sector, the British Institute of NDT strongly recommends
that industry demand that this person should hold and maintain Level 2 certification in
the applicable methods and sectors.
NDT4-51015
Prelims 18 Copyright © TWI Ltd
Training
ET 40 40 40
PT 16 24 24
MT 16 24 32
RT 40 80 72
RI N/A 56 N/A
UT 40 80 72
VT 16 24 24
Note 1. Direct access to Level 2 requires the total number of hours shown in Table 1 for Levels 1
and 2, and direct access to Level 3 requires the total number of hours shown in Table 1 for
Levels 1-3. Up to one third of the total specified in this table may take the form of OTJ training
documented using form PSL/42 provided it is verifiable and covered practical application of the
syllabus detailed in CEN ISO/TR 25107:2006.
NDT4-51015
Prelims 19 Copyright © TWI Ltd
Table 2 Minimum duration of experience for certification.
Experience, months
NDT method
Level 1
Level 2 Level 3
ET 3 9 18
MT 1 3 12
PT 1 3 12
RT 3 9 18
UT 3 9 18
RI N/A 6 N/A
VT 1 3 12
Work experience in months is based on a nominal 40-hour week or the legal week of work.
When an individual is working in excess of 40h/week, he may be credited with experience
based on the total hours, but he shall be required to produce evidence of this experience.
Direct access to Level 2 requires the total number of hours shown in Table 2 for Levels 1 and 2,
and direct access to Level 3 requires the total number of hours shown in Table 2 for Levels 1-3
Qualification examination
Table 3 Numbers of general questions.
NDT method Level 1 Level 2
ET 40 40
PT 30 40
MT 30 40
RT 40 40
RI N/A 40
UT 40 40
VT 30 40
BRS 30 N/A
NDT4-51015
Prelims 20 Copyright © TWI Ltd
Re-examination
a A candidate who fails to obtain the pass grade for any examination part (general,
specific or practical) may be re-examined twice in the failed part(s), provided the re-
examination takes place not sooner than one month, unless further training
acceptable to BINDT is satisfactorily completed, nor later than twelve months after
the original examination.
b A candidate who achieves a passing grade of 70% in each of the examination parts
(general, specific or practical) but whose average score is less than the required 80%
may be re-examined a maximum of two times in any or all of the examination parts
in order to achieve an overall average score of 80%, provided the re-examination
takes place not sooner than one month, unless further training acceptable to BINDT
is satisfactorily completed, nor later than twelve months after the original
examination.
c A candidate who fails all permitted re-examinations shall apply for and take the initial
examination according to the procedure established for new candidates.
d A candidate whose examination results have not been accepted for reason of fraud or
unethical behaviour shall wait at least twelve months before re-applying for
examination.
Summary
The PCN scheme is managed and administered by the British Institute of NDT (BINDT)
on behalf of its stakeholders. It meets or exceeds the criteria of BS EN ISO 9712.
1 Aerospace.
2 Castings.
3 Welds.
4 Wrought Products and Forgings.
5 Pre and in-service inspection (multi sector).
6 Railway.
There are many additional supporting documents varying from vision requirements
PSL44 to renewal and recertification (Levels 1 and 2 – CP16; Level 3 – CP17) and so on.
The document defines many terms used in certification of NDT personnel (PCN Gen
Section 3)
The certification body (BINDT) meets the requirements of BS EN ISO 17024 (PCN Gen
section 5)
NDT4-51015
Prelims 21 Copyright © TWI Ltd
BINDT approves authorised qualifying bodies (AQBs) to carry out the examinations (PCN
Gen Section 5)
a The document sets out the Levels of PCN certification and what each level of
personnel is qualified to do (PCN Gen section 6). There are 3 Levels of PCN
certification.
b Candidates for examination must have successfully completed a BINDT validated
course of training at a BINDT authorised training organisation (PCN Gen Section 7).
c Table 1 shows the minimum required duration of training for all Levels and methods
plus a section of notes.
d Table 2 gives the minimum duration of experience for each Level and method.
e A candidate is required to have a vision test of colour perception and a near vision
test (Jaeger Number 1 or N4.5). PCN Gen Section a – the near vision test to be taken
annually.
f Examination applications are made directly with the AQB.
g PCN Level 1s and 2 initial exams comprise general; specific and practical parts.
h Table 3 shows the number of general questions at Levels 1 and 2 examinations.
i There are 30 specific questions on the Level 1 papers.
j There are 36 questions on the Level 2 specific papers.
k A variety of practical samples are tested depending on the method and sector.
l A Level 3 examination comprises a basic and a method examination – however the
basic examination needs to be passed only once. Table 4 shows the number of basic
examination questions. Table 5 shows the number of Level 3 examination questions.
NDT4-51015
Prelims 22 Copyright © TWI Ltd
Table 5 Main method examination.
m A pass is obtained where each part is 70% or over with an average grade of 80% or
over.
n A PCN certificate is valid for 5 years.
o Renewal and recertification requirements are covered in CP16 for Level 1 and Level 2
and CP17 for Level 3.
NDT4-51015
Prelims 23 Copyright © TWI Ltd
Section 1
Physical Principles
1 Physical Principles
Sound is generated when something vibrates. You can twang a ruler on a table
or flick a stretched elastic band to verify this. The stretched surface of the
rubber band or the ruler vibrates and sets up a series of vibrations, sound
waves, in the air. As the surface of the band or ruler pushes into the air, the air
molecules are forced together and a region of high pressure forms; this process
is called compression. As the surface moves back, the air molecules move
apart, forming a low pressure area (rarefaction). As the surface vibrates,
alternate compressions and rarefactions are set up in the air and travel out
from the surface to form a sound wave. The air molecules don’t move with the
wave - they vibrate to and fro in time with the vibrating surface.
If we plot the displacement of the particle against time, it will produce a sine
wave as shown below.
D
i
s
p
l
Particle a
Motion c Time
e
m
e
n
t
One cycle
Figure 1.1 Ultrasonic vibration.
The sound wave thus produced travels through the air at a speed of about
332m/sec, at 0oC, at sea level. We hear the sound when it hits a membrane in
our ear and causes it to vibrate.
Sound will travel through any medium that has molecules to move but it travels
faster in more elastic materials because the vibrations are passed on more
quickly. Sound travels faster in water or metal than it does in air because
liquids and solids are more elastic than air. The speed of sound in a material
increases with its stiffness (elasticity) and decreases with its density; more
precisely, the square root of the stiffness divided by the density gives the speed
of sound.
1.1 Frequency
As sound is a series of vibrations, one way of measuring it is to count the
number of vibrations per second - the frequency. Frequency is measured in
hertz (Hz). One vibration in one second is one hertz. Two vibrations in one
second is two hertz. Ten vibrations in one second is 10 hertz and 1000
vibrations in one second is 1000 hertz or one kilohertz (kHz). One million
vibrations in a second is one megahertz (MHz).
NDT4-51015
Physical Principles 1-1 Copyright © TWI Ltd
The higher the frequency - the higher the note sounds - the higher the pitch. If
you twang the ruler or the rubber band hard, the noise is louder, it has greater
amplitude but the note remains the same. If, however, you shorten the ruler or
tighten the rubber band, they vibrate more quickly and the note given out is
higher, the frequency is greater. To raise the pitch of their instrument, guitar
players move their fingers down the frets, thus shortening the string and
making it vibrate more quickly.
We can only hear sounds between certain frequencies - more than 20 and less
than 20,000Hz. If you were able to move your arm up and down 20 times a
second, it would sound like a very low hum. You cannot move your arm this
fast, so you cannot hear the vibrations in the air caused by your moving arm. A
dog whistle vibrating at 25,000Hz cannot be heard by humans but it can be
heard by the sensitive ears of a dog.
Testing
0.5 - 50MHz
It rarely occurs to us that there is a whole world of sound that we cannot hear.
Some other animals can hear sounds at higher frequencies - bats can hear
sound at 100,000Hz - and some animals, like snakes, have worse hearing than
we have.
A sound with frequencies above the upper range of human hearing is called
ultrasound. Sound below about 16Hz is called infrasound. Therefore the
definition of ultrasound is sound with a frequency greater than 20kHz.
There is an advantage to using lower frequencies: The lower the frequency, the
more penetrating a sound wave is - that is why foghorns give out very low
notes and why the low throbbing notes from your neighbour’s stereo set come
through the wall rather than the high notes. Elephants and hippos can
communicate over distances of up to 30 kilometres using ultrasound, while
whales can communicate through water across an ocean!
1.2 Wavelength
A wave in the sea is a vibration of energy. As the wave passes a fixed point it
produces a constant rise and fall of energy. A complete vibration is a change in
energy from maximum to minimum and back to maximum. The distance over
which one complete vibration of energy occurs is called a wavelength.
NDT4-51015
Physical Principles 1-2 Copyright © TWI Ltd
Wavelength Velocity
λ =V
f
Frequency
332
1.66m
200
5,920,000
2.96mm
2,000,000
An easy way to remember how this formula works is to split it down within a
triangle - with the velocity, wavelength and frequency at the corners. The
velocity must be placed at the top (note how it forms a diamond shape) and the
wavelength and frequency at either of the bottom two corners.
V
λ f
NDT4-51015
Physical Principles 1-3 Copyright © TWI Ltd
If we want to work out wavelength we cover the wavelength symbol - this
leaves V over . If we need to find the velocity, cover the V which gives us x
. Covering the frequency () will leave V over .
V λ=
V
f
λ f
V = λf
V
λ f
V f=
V
λ
λ f
1.3 Resolution
Resolution is the ability of an equipment/probe combination to distinguish
between two echoes from reflectors that are close together. To have good
resolution, a probe must present two signals on a CRT screen from two
separate reflectors: if it has poor resolution, the echoes from the two reflectors
appear as one signal on the screen.
NDT4-51015
Physical Principles 1-4 Copyright © TWI Ltd
In the early days of ultrasonic testing, we used 100, 91 and 85mm steps at the
radius end of the V1 block to test resolving power. However, today this is
regarded as too crude a test and BS 4331 part 3 (now obsolete and superseded
by BS EN 12668-3: Methods of assessing the performance characteristics of
ultrasonic flaw detection equipment Part 1: Overall performance on-site
methods) recommends that we should be able to recognise two discrete echoes
less than two wavelengths apart. Discrete echoes means split by more than 6dB
(see 1.4) or to more than half the total height of the signals.
100%
50% - 6dB
Accept Reject
The relative amplitude of ultrasonic signals is expressed using the decibel (dB),
a logarithmic unit of comparison. When we compare the height of two signals
on the CRT screen, we are in fact comparing the electric voltage that is being
sent to the Y plates; electric voltage is proportional to the square of the current.
To compare two signals we must use a formula that takes account of this fact:
H1
Differencein dB 20 x log10
H2
For example, if we want to compare a signal of 40mm with one of 20mm on the
CRT screen:
Difference in dB 20 log 10
40
20
20 log 10 2
20 x 0.301 = 6.02dB
NDT4-51015
Physical Principles 1-5 Copyright © TWI Ltd
So the answer is 6dB and this can be tested on a CRT screen by obtaining a
signal from a backwall echo on a test block and increasing or decreasing the
gain until the signal touches the top of the screen. Take out 6dB with the gain
control and the signal should drop to 50% full screen height (FSH). If it does
not, the vertical linearity of the UT set is out or inaccurate; the signal height is
not changing in accordance with energy from the probe.
12dB difference means that one signal is 4 times bigger than another.
10dB difference means that one signal is 3 times bigger than another.
20dB difference means that one signal is 10 times bigger than another.
Remember that decibels are only a means of comparing signals. All UT sets are
different, so a defect may be at FSH with a gain control reading of, say, 36dB
on one set and be at FSH on another set with a gain control reading of only
28dB on another set. The gain control allows us to set sensitivities and forms
the basis of ultrasonic sizing techniques.
NDT4-51015
Physical Principles 1-6 Copyright © TWI Ltd
Section 2
Modes of Sound Energy
2 Modes of Sound Energy
Sound waves propagate due to the vibrations or oscillatory motions of particles
within a material. Within a freely vibrating medium each particle is subject to
both inertial and elastic forces. These forces cause particles to exhibit oscillatory
motions comparable to the free vibration of a system of masses and springs.
The elastic restoring forces in a material can be described as microscopic spring
forces as shown below.
Figure 2.1 Representation of sound wave propagation using partial mass and
microscopic restoring spring forces.
This theory agrees with both Hooke’s Law and Newton’s second law. Hooke’s
Law states that, within the elastic limit of any body, the ratio of the stress to
the strain produced is constant; therefore, the more stress or force is placed on
an object, the more it will strain or deform. Newton’s second law of motion
states that the force (F) equals the mass (m) times the acceleration (a).
F = ma
The spring theory makes accurate predictions for the propagation of sound. The
propagation of a sound wave velocity is determined by the elastic properties
and density of the material. The velocity of a longitudinal wave is described by
the following equation:
V = fλ
NDT4-51015
Modes of Sound Energy 2-1 Copyright © TWI Ltd
Propagation
Particle vibration
E 1
VL
1 1 2
Note: Sound can only travel through air and water in the compressional mode.
Sound can travel through Perspex, steel and aluminium in modes other than the
compressional mode.
NDT4-51015
Modes of Sound Energy 2-2 Copyright © TWI Ltd
In the shear or transverse mode, molecules of a solid move rather like beach
balls floating on the surface of the sea - they move up and down as a wave
passes.
Particle vibration
Propagation
This type of sound travel can only happen when the molecules through which it
propagates are joined together - as in a solid. A solid has rigidity as well as
elasticity. Air and water, like other gases and liquids, do not have rigidity. Shear
or transverse waves cannot travel in gases of liquids for this reason.
E 1 G
Vs
2 1
The speed of sound in the shear or transverse mode is less than it is in the
compressional or longitudinal mode. The shear speed of sound in steel is
3250m/sec and in aluminium 3130m/sec. There is no shear or transverse speed
for air or water, as shear waves cannot be supported in these media.
NDT4-51015
Modes of Sound Energy 2-3 Copyright © TWI Ltd
Table 2.1 Comparison of compression wave and shear wave velocities in
different materials.
Applying these values for one velocity of steel to the formula used previously
for determining the wavelength, it can be seen that for a given frequency that
the wavelength of the shear wave is less than that of the compression wave.
NDT4-51015
Modes of Sound Energy 2-4 Copyright © TWI Ltd
2.4 Lamb waves
Another mode of sound travel is Lamb or plate waves which propagate in thin
plate materials when the plate thickness is about the same as the wavelength.
Lamb or plate waves travel at velocities which vary with the plate thickness and
the wavelength. Particle motion is elliptical, as with surface waves.
Asymmetrical
Symmetrical
NDT4-51015
Modes of Sound Energy 2-5 Copyright © TWI Ltd
Section 3
Generating Ultrasound
3 Generating Ultrasound
Sound is created when something vibrates. It is a stress wave of mechanical
energy. The piezo-electric effect changes mechanical energy into electrical
energy. It is reversible, so electrical energy - a voltage - can be changed into
mechanical energy or sound, which is the reverse piezo-electric effect. The first
people to observe the piezo-electric effect were the Curie brothers who
observed it in quartz crystals.
- + + - - +
It was later discovered that by varying the thickness of crystals and subjecting
them to a voltage, they could be made to vibrate at different frequencies. The
frequency depends on the thickness of the piezo-electric crystal, according to
the following formula:
V
t
2f
Where:
t = Crystal thickness.
V = Velocity of sound in crystal.
f = Frequency.
NDT4-51015
Generating Ultrasound 3-1 Copyright © TWI Ltd
3.2 Quartz or silicon oxide (SiO2)
Found in granite as a natural crystal, quartz can produce compressional or
shear waves according to the way the crystal is cut. An X-cut crystal is cut in a
direction that directly crosses the axis joining two angles of the crystal. A Y-cut
crystal is cut in a direction parallel to the axis joining two angles of the crystal
(see Figure 3.2).
x x x
z
y
z
Figure 3.2 X-cut quartz crystal.
Quartz is not much used now. Several types of quartz crystal can be produced,
each with its advantages and disadvantages, some of which are listed below.
Advantages
Resistant to wear.
Insoluble in water.
Resistant to ageing.
Easy to cut to give the required frequencies.
Disadvantages
Needs a lot of electrical energy to produce a small amount of ultrasound,
which means it is inefficient.
Quartz crystals are susceptible to mode change.
High voltage is needed to give low frequency sound.
For these reasons quartz has been largely superseded by other piezo-electric
materials.
NDT4-51015
Generating Ultrasound 3-2 Copyright © TWI Ltd
3.3 Lithium sulphate (Li2SO4)
Crystals grow as a solution of lithium sulphate is evaporated.
Advantages
Most efficient receiver of ultrasound.
Very low electrical impedance.
Operate well at low voltages.
Do not age.
Very good resolution.
Crystals are easily damped to give short pulse lengths (to give good
resolution).
Disadvantages
Soluble in water.
Break easily.
Decompose at temperatures above 130°C.
Advantages
Efficient generator of sound.
Only needs a low voltage.
Good sensitivity.
Disadvantages
Its Curie temperature, at which the crystal depolarises, is only about 120°C,
which makes it susceptible to heating.
Deteriorates over time.
Advantages
Heavy internal damping.
Gives out very narrow pulses of ultrasound, which gives good resolution.
Disadvantages
Much less sensitive than lead zirconate titanate (PZT).
NDT4-51015
Generating Ultrasound 3-3 Copyright © TWI Ltd
3.6 Lead zirconate titanate (PbZrO3, PbTiO3)
Lead zirconate titanate (PZT) crystals have the best all-round performance for
industrial testing.
Advantages
A high Curie point, up to 350C.
Good resolution.
Does not dissolve in water.
Tough and resistant to ageing.
Easily damped.
EMAT probes are used for the detection of flaws and the determination of
material properties such as the precise velocity as well as attenuation
measurements. They do not require the use of couplant and as such can
operate without contact at elevated temperatures and in remote locations.
EMAT probes are, however, inefficient and require strong magnetic fields and
large currents to produce ultrasound that is often weaker than that produced by
piezo-electric transducers. Rare earth materials such as samarium-cobalt and
neodymium-iron-boron are often used to produce sufficiently strong magnetic
fields, which may also be generated by pulsed electromagnets.
EMAT probes generate ultrasonic waves due to the interaction between a static
magnetic field of a magnet and the high frequency magnetic field generated by
a coil. The eddy currents produced in the material due to the coil create a
Lorentz force, causing the atomic lattice of the material to oscillate and produce
an ultrasonic wave. A magnetic structure component is also generated by the
EMAT and although not very efficient in terms of energy, the ultrasonic
proportion can have useful in-service applications (see below).
NDT4-51015
Generating Ultrasound 3-4 Copyright © TWI Ltd
Figure 3.3 Method of ultrasonic wave generation from an EMAT probe.
NDT4-51015
Generating Ultrasound 3-5 Copyright © TWI Ltd
Section 4
Pulse Length and Damping
4 Pulse Length and Damping
A pulse of ultrasound from a piezo-electric crystal has a length or width of
several vibrations or wavelengths. When you strike a bell it continues to ring for
several seconds as the metal continues to vibrate. The vibrations get steadily
weaker and the sound dies away. If you put your hand on the bell you stop the
vibrations and the sound dies away more quickly - you dampen the sound.
Pulse length (or width) is also sometimes called wave train length. It is defined
in a number of ways but not even the standards always agree. We choose the
one in EN 1330 Part 4 NDT terminology – Part 4: Terms used in ultrasonic
testing, which defines it as the leading and trailing edges of a pulse measured
at a defined level below the peak amplitude.
Maximum
10% of maximum
Pulse length
NDT4-51015
Pulse Length and Damping 4-1 Copyright © TWI Ltd
1-2 cycles
10-12 cycles
5 cycles
NDT4-51015
Pulse Length and Damping 4-2 Copyright © TWI Ltd
Section 5
The Sound Beam
5 The Sound Beam
The spread of sound waves from a piezo-electric crystal has been likened to the
beam of a torch, ie an elongated cone. Just as the intensity of light from a torch
diminishes with distance, so sound pulses get weaker the further they travel
from the crystal. An acoustic sound wave has also previously been described as
being a single sinusoidal wave propagating through a material. These analogies
do not however present a totally true picture. The sound produced from an
ultrasonic crystal does not originate from a single point but rather it is derived
from many points along the surface of the piezo-electric crystal. This results in
a sound field with many waves interacting or interfering with each other (see
Figure 5.1).
When waves interact. they overlay each other and the amplitude of the sound
pressure or particle displacement at any point of interaction is the sum of the
amplitudes of the two individual waves. When the waves are fully in phase, the
result is additive or constructive and the intensity is doubled. When completely
out of phase, the result would be the amplitudes cancelling each other out. The
interaction can vary between these two extremes and the wave produced will
equal the sum of the amplitudes at all points with peaks of intensity referred to
as nodes. In an ultrasonic probe the situation is further complicated as sound
originates from not just two but many points on the crystal surface.
NDT4-51015
The Sound Beam 5-1 Copyright © TWI Ltd
In an ultrasonic probe, one would expect the sound intensity to be highest at
the probe face and to fall away gradually as the distance from the probe
increases. Due to interactions near the face of the probe, however, the sound
field is very uneven in this region with peaks and troughs in sound intensity.
This area of intensity variation is known as the near field or Fresnel zone. As
one moves farther away from the probe these variations are eliminated and the
sound field behaviour becomes more uniform. This region of the sound beam is
referred to as the far field, or Fraunhofer zone. In the far field, the intensity
behaves as expected and is reduced exponentially with distance. The beam
spreads out as a circular wave front.
Intensity of the
sound beam on its
central axis Exponential decay
After a time, the shock waves or pulses even out to form a continuous front.
The area between the crystal and the point where the wave front evens out is
what we call the near or Fresnel zone. Inside the near zone, signals from a
reflector bear no accurate relation to the size of the reflector, as the sound
vibrations are going in all directions. This affects the accuracy of flaw sizing of
small reflectors inside the near zone.
Far zone
Near zone
NDT4-51015
The Sound Beam 5-2 Copyright © TWI Ltd
The near zone of a crystal varies with the material being tested, but it can be
worked out by a formula:
D2
Near zone =
4λ
D 2f
Or
4V
D = Diameter
f = Frequency
= Wavelength
V = Velocity
For example, the near zone of a 5MHz compression probe with a 10mm
diameter crystal will be, in steel:
10 2 x 5,000,000
21.1mm
4 x 5,920,000
The near zone of a 2.5MHz probe with a 20mm diameter crystal will be:
20 2 x 2,500,000
42.2mm
4 x 5,920,000
In twin crystal or angle beam probes with a Perspex stand-off component in the
probe body, some or even all of the near zone is contained in the Perspex shoe.
This must be taken into account when the calculation for near zone is applied.
We will need to know the length of the beam path within the Perspex, the
velocity of the Perspex and the velocity of the test material, as well as the near
zone within the test material.
The first step is to calculate the near zone in the test material; then the path
length in the Perspex is multiplied by the velocity in the Perspex before dividing
by the velocity in the test material; this is finally subtracted from the near zone
length.
NDT4-51015
The Sound Beam 5-3 Copyright © TWI Ltd
Ratio of near zone
in the wedge
Nrs = Ns Tw
1 -Nw
Ratio
remaining
Figure 5.5 Near zone path distance in the steel test object and probe wedge.
I = Sound intensity.
r = Distance from the crystal.
The higher the frequency of the crystal, the less the beams spread out. The
angle of beam spread can be found using the formula below:
2
NDT4-51015
The Sound Beam 5-4 Copyright © TWI Ltd
Kλ KV
sin θ/2 or
D Df
Where:
Edge, K = 1.22
20dB, K = 1.08
6dB, K = 0.56
Beam axis
Where the sound intensity drops by 6dB (half the intensity), K is 0.56.
If we take the edge of the beam to be where the sound energy is 10% (20dB)
of the energy at the beam centre, K is 1.08.
If we take the extreme edge of the sound beam to work out beam spread
angles, then K is 1.22.
Example
The beam spread of a 10mm 5MHz probe in steel is calculated as follows:
5920
sin /2 = 1.08 or 1.22 x
5000 x10
So the angle of beam spread is 14.7 if you take the edge of the beam to be
where the energy is 10% of the main energy or 16.6 taking the extreme edge
of the beam.
This is one of the reasons why low frequency probes have large diameter
crystals.
NDT4-51015
The Sound Beam 5-5 Copyright © TWI Ltd
Section 6
Total Attenuation Loss
6 Total Attenuation Loss
EN 1330 defines attenuation as the decrease in sound pressure that occurs
when a wave travels through a material arising from absorption and scattering.
The two components, absorption and scatter, are defined as:
Absorption
Component of the attenuation resulting from transformation of ultrasonic
energy into other types of energy (eg thermal).
Absorption occurs as the sound pulse hits the molecules of the test material and
makes them vibrate. The energy lost in vibrating the molecules turns to heat.
The rate of absorption varies from one material to another and even from one
type of steel to another. It is very high in Perspex, nylon and lead and is low in
aluminium.
Scatter
Randomly reflected energy caused by grain structure and/or by small
discontinuities in the beam path.
Scatter occurs as sound energy is reflected from grains in the test material. The
larger the grains, the more scatter occurs. The grass at the bottom of the CRT
screen is caused by reflections from grain boundaries in the test material. More
grass arises from cast iron or brass than from small grained materials like
refined steel or annealed aluminium.
The longer the wavelength of a sound pulse, the less energy is scattered.
Where the wavelength is smaller than the grain size, a sound pulse is scattered
very quickly. It is for this reason that a low frequency probe, with its longer
wavelength, has greater penetration in a given material than a high frequency
probe.
Scatter
NDT4-51015
Total Attenuation Loss 6-1 Copyright © TWI Ltd
6.1 Attenuation due to beam spread
The amount of energy reflected back depends on whether the reflector is bigger
in area than the sound beam at that distance. If the reflector is bigger in area
than the sound beam, the signal on the screen varies according to the law of
large reflectors. If the reflector is smaller in area than the sound beam at that
distance, it obeys the law of small reflectors.
Both of these laws only apply beyond a distance of three times the length of the
near zone.
80% FSH
20mm
40mm
40% FSH
NDT4-51015
Total Attenuation Loss 6-2 Copyright © TWI Ltd
6.1.2 Law of the small reflector
Small reflectors outside three near zones obey the inverse square law.
20mm
80%
FSH BWE
BWE
40mm
20%
FSH
Figure 6.3 Law of the small reflector (ie smaller than the beam width).
NDT4-51015
Total Attenuation Loss 6-3 Copyright © TWI Ltd
Measurement of material attenuation
a Place a compression probe on a piece of the test material and turn the
backwall echo to FSH. Note: If the back wall echo (BWE) is within three
near zones of the probe, use the first BWE outside the distance of three
near zones.
b Obtain an echo from twice the distance of the BWE used and increase the
signal height until it is at FSH. Note: The dB difference.
c To remove the effects of beam spread we subtract 6dB (due to beam spread
loss) from the dB difference and divide the remainder by the distance the
sound has travelled between the two echoes. This is twice the distance
shown on the CRT, as the sound has to travel to the backwall and back to
trigger the probe.
d The answer will give the number of decibels lost per millimetre by
attenuation. This can give an assessment of plate quality and heat
treatment.
If the difference between the first and second backwall echoes from a 75mm
thick block of steel was 9dB, what is the attenuation of the material?
This gives 3dB due to attenuation only as the sound travels through the block.
The 75mm block gives a sound path for the pulse echo of 150mm; the sound
has to travel to the backwall, then back to the probe:
3/150 = 0.02
Attenuation is 20dB/m.
NDT4-51015
Total Attenuation Loss 6-4 Copyright © TWI Ltd
Section 7
Acoustic Impedance
7 Acoustic Impedance
When a sound pulse arrives at an interface between different materials at right
angles, some sound is reflected back into the material from whence it came.
The rest of the sound, however, is transmitted into the second material. This is
due to the difference in acoustic impedance of the two materials and is known
as acoustic impedance mismatch or sometimes as interface behaviour.
We can calculate how much sound is transmitted and how much sound is
reflected back if we know the acoustic impedance of both materials.
Z V
Where (the Greek letter rho) is the density and V is the sound velocity
(compressional or shear, depending on the case).
Once we know the acoustic impedances of two materials, we can use a formula
to work out how much sound will be reflected back. The formula is:
2
Z1 Z 2
x100 % reflected
Z1 Z 2
Where Z1 is the acoustic impedance of the first material and Z2 is the acoustic
impedance of the second material.
Example
To calculate the amount of energy reflected back at a steel-water interface, we
must find out the acoustic impedances of steel and water. They are:
2
46.7 1.48
x100
46.7 1.48
2
45.22
x100
48.18
0.938562 x 100
88% of the sound energy is reflected back at the interface. This means that
12% of the energy is transmitted at the interface. Using the same formula, the
figures for other media can be worked out. At a steel/oil interface, 91% of
sound energy is reflected back; at a glycerine/steel interface, 90% of energy is
reflected back.
NDT4-51015
Acoustic Impedance 7-1 Copyright © TWI Ltd
These substances, water, oil and glycerine, are used as couplants in ultrasonics
to transmit sound energy from the probe into the test materials. So in fact, only
about 10% of the energy generated by the probe crystal actually gets through
the couplant into the test material.
Aluminium 17
Brass 36
Copper 41
Lead 27
Magnesium 93
Nickel 50
Steel 46.7
Glass 18
Polystyrene 29
Oil 1.3
Water 1.4
Air 0.0041
NDT4-51015
Acoustic Impedance 7-2 Copyright © TWI Ltd
Section 8
Snell’s Law
8 Snell’s Law
When sound waves pass obliquely (not at 90°) between materials having
different acoustic velocities, the direction of sound propagation is changed on
passing through the interface and the sound wave is said to have been
refracted.
Light is also refracted when passing from one medium to another with a
different velocity; this means that objects seen across an interface appear to be
shifted relative to where they really are.
Incident
Transmitted
sin I V1
I
Material 1
sin R V2
Where:
I = Incident angle
R = Refracted angle
V1 = Velocity in material 1
Material 2
V2 = Velocity in material 2
R
NTD4-51015
Snell’s Law 8-1 Copyright © TWI Ltd
If we want to make a probe transmitting a shear wave at a certain angle, we
have to transpose the formula:
sin R V1
sin I
V2
For example, if we want a probe giving a 45 shear wave in steel, we must
calculate the angle at which to cut the Perspex wedge - the incident angle. The
compression speed of sound in Perspex is 2730m/s, the shear speed of sound in
steel is 3250m/s and the refracted angle we need is 45.
sin45 o x2730
sin I
3250
0.7071x2730
sin I
3250
sin I 0.594
I 36.44 o
However, when the incident angle in the Perspex shoe is less than 27, both
compression and shear waves are generated in the steel. This makes
interpretation very confusing. To get a shear wave on its own, the angle of
incidence must be more than 27.4, called the first critical angle. This gives a
shear wave of 33 (the lowest standard angle probe manufactured is 35).
C 20º
Perspex
Steel 48.3º
C
24º
S
NTD4-51015
Snell’s Law 8-2 Copyright © TWI Ltd
If the incident angle is above 57.14, the shear wave is replaced by a surface
wave. This angle is called the second critical angle.
C C C
27.4
57.14
Perspex Perspex
C
S
Steel Steel
33
S
The largest probe angle below a surface (90°) wave probe available from
manufacturers without a special order is 80.
Shear waves on their own in steel are only possible when the incident angles
are between 27.4 and 57.14. This is worked out by the probe manufacturers
and it must be borne in mind that a probe which gives a refracted angle of 45
in steel will give a different refracted angle in other materials.
The first critical angle is the incident angle at which the compression wave in
the test material is generated at 90. So using Snell’s law:
sin I 2730
o
sin 90 5960
sin 90 o 1
2730
sin I
5960
NTD4-51015
Snell’s Law 8-3 Copyright © TWI Ltd
The second critical angle is the incident angle at which a shear wave is
generated in the material at 90. Use Snell’s law again:
2730
sin I
3240
sin I 0.8245
I 57.4 o
NTD4-51015
Snell’s Law 8-4 Copyright © TWI Ltd
Section 9
Probe Design
9 Probe Design
In the US, a probe is usually called a search unit or formerly a transducer.
However, we generally now understand the transducer to be the crystal. There
are a number of probe designs and configurations. We shall deal with those
most commonly used in weld, aerospace and general ultrasonic testing.
Electrical
connectors
Housing
Damping
Transducer
NDT4-51015
Probe Design 9-1 Copyright © TWI Ltd
9.2 Angle probes
An angle probe is a piezo-electric crystal mounted on a Perspex wedge at an
angle calculated to generate a shear (transverse) wave in the test material.
I
C
S R
If there is no couplant on the bottom surface of the Perspex wedge, all the
energy is reflected back into the probe. If there is couplant and if the probe is
placed on a test material, sound energy passes into the test material and
generates a shear wave. Angle probes use compression probes mounted on a
wedge of Perspex. The wedge of such a probe is cut to a particular angle to
enable the beam to refract into the test material at a chosen angle.
Angle probes usually transmit a shear or sometimes a surface wave into test
materials and are used largely in weld testing, casting and forging inspection
and in aerospace applications.
For a single crystal probe, the length of the initial pulse is the dead zone and
any signal from a reflector at a shorter distance than this will be concealed in
the initial pulse.
NDT4-51015
Probe Design 9-2 Copyright © TWI Ltd
In twin crystal probes, the initial pulse is deliberately delayed beyond the left of
the time base by mounting the transducers of a twin (or double) crystal probe
onto plastic wedges. This, in addition to the focusing of the crystals, reduces
the dead zone considerably and flaws can be assessed anywhere except where
the transmission and receptive beams do not overlap.
A twin or double crystal probe is designed to minimise the problem of the dead
zone. A twin crystal probe has two crystals mounted on Perspex shoes, angled
slightly inward to focus at a set distance in the test material. If the crystals
were not angled, the pulse would be reflected straight back into the
transmitting crystal.
The Perspex shoes hold the crystals away from the test surface so that the
initial pulse does not appear on the CRT screen. The dead zone is greatly
reduced to the region adjoining the test surface, where the transmission and
reception beams do not overlap.
Disadvantages:
Good contact is difficult with curved surfaces.
Difficult to size small defects accurately as the width of a double crystal
probe is usually greater than that of a single crystal probe.
The amplitude of a signal decreases the further a reflector is situated from
the focal distance - a response curve can be made out.
NDT4-51015
Probe Design 9-3 Copyright © TWI Ltd
9.4 Other probe types
Immersion probes are designed for use where the test part is immersed in
water. They are typically used inside a water tank or as part of a squirter or
bubbler system in scanning applications. Immersion transducers usually have
an impedance matching layer that helps to get more sound energy into the
water and thus into the component being inspected. Immersion transducers can
be purchased in a flat, cylindrically or spherically focused lens. A focused
transducer can improve sensitivity and axial resolution by concentrating the
sound energy to a smaller area.
Delay line probes, as the name implies, introduce a time delay between the
generation of the sound wave and the arrival of any reflected waves. This
allows the crystal to complete its transmission function before it begins to
receive returning signals. Delay line transducers are recommended for
applications that require a contact transducer with good near-surface resolution
and are designed for use in applications such as high-precision thickness
gauging of thin materials and delamination checks in composite materials. They
are also useful in high temperature measurement applications since the delay
line provides some heat insulation to the piezo-electric element.
NDT4-51015
Probe Design 9-4 Copyright © TWI Ltd
Section 10
Test Techniques
10 Test Techniques
10.1 Pulse echo
The major disadvantage is that pulse echo inspection is reliant upon the defects
having the correct orientation relative to the beam in order to generate a
returning signal to the probe and is therefore not considered fail safe (see
Figures 10.1 and 10.2). If the sound pulse hits the flaw at an angle other than
90°, much of the energy will be reflected away and not return to the probe with
the result that the flaw will not show up on the screen.
Figure 10.1 Schematic drawing of the pulse echo technique showing specular
reflection from a discontinuity: Normal compression probe.
Figure 10.2 Schematic drawing of the pulse echo technique showing specular
reflection from a discontinuity: Shear wave angle probe.
10.2 Through-transmission
Through-transmission was used in the early days of UT and is still used in plate
and bar production. A probe on one side of a component transmits an ultrasonic
pulse to a receptor/receiver probe on the other side. The absence of a pulse
arriving at the receiver indicates a defect.
NDT4-51015
Test Techniques 10-1 Copyright © TWI Ltd
T No defect
Transmission
signal
T Defect
Transmission signal
reduced
Advantages
Less attenuation of sound energy.
No probe ringing.
No dead zone on the screen.
Orientation of a defect does not matter as much as on the pulse echo
display.
Disadvantages
Defect cannot be located.
Defect cannot be identified.
Component surfaces must be parallel.
Vertical defects do not show.
Process must be automated.
Requires access to both sides of the component.
NDT4-51015
Test Techniques 10-2 Copyright © TWI Ltd
10.3 Tandem scanning
Tandem scanning is used mainly to locate defects lying perpendicular to the
surface. It involves the use of two or more angle probes of the same angle of
incidence and facing the same direction with one probe acting as the
transmitter and the others as receivers.
T R R R R R
NDT4-51015
Test Techniques 10-3 Copyright © TWI Ltd
10.6 Immersion testing
Immersion testing involves the test object being submerged in a liquid, usually
water and the probe being scanned at a fixed distance above the component.
The water serves to provide constant coupling conditions and amounts to a long
fluid delay line. Although the probe itself requires a compression wave, shear
waves can be produced within the sample by angulation of the probe. This
technique frequently uses high frequency probes (25-50MHz) and focused
probes for automated inspections and is suited to the inspection of complex
components, see eg BS M36: Ultrasonic Testing of Special Forgings By an
Immersion Technique.
Water path
distance
10.7 Presentation
C – Plan view
B – End View
D – Side View
NDT4-51015
Test Techniques 10-4 Copyright © TWI Ltd
Figure 10.8 P scan image.
10.7.1 A scan
The flaw detector or UT set sends ultrasound energy into test materials. Some
of this energy returns to the set to be presented as information on a cathode
ray tube (CRT) screen. This is an A scan display with the amplitude of signals
displayed as a function of time or distance.
10.7.2 B scan
This gives an end or cross-sectional view of the component being examined,
with the position of the probe displayed on one axis and the distance from the
surface to the signal on the other (see Figure 10.7). The B scan is used in
hospitals and on aircraft components. It is often used with specimens immersed
in water and with an automated scanning device.
NDT4-51015
Test Techniques 10-5 Copyright © TWI Ltd
10.7.3 C scan
The C scan gives a plan view of a defect (see Figure 10.7). It is often used as
an automated process to map out laminations in plate. It gives the area of a
defect, so it is good for plotting the extent of laminations in metal sheets.
10.7.4 D scan
The D scan gives a side view of the defect seen from a viewpoint normal to the
B scan (see Figure 10.7). It is usually automated and shows the length, depth
and through thickness of a defect. The D scan should not be confused with the
delta technique.
NDT4-51015
Test Techniques 10-6 Copyright © TWI Ltd
Section 11
Ultrasonic Flaw Detector
11 Ultrasonic Flaw Detector
11.1 Principles
The ultrasonic flaw detector, which is part of the UT set, sends a voltage down a
coaxial cable (sometimes called the lead) to a probe. The piezo-electric crystal
in the probe is hit by the voltage and vibrates. The vibration creates an
ultrasonic pulse which enters the test material. The pulse travels through the
material until it strikes a reflector and is reflected back to the probe.
It re-enters the probe, hits the crystal and vibrates it, causing it to generate a
voltage. The voltage causes a current which travels back to the flaw detector
along the cable. The set displays the time the pulse has taken through the test
material and therefore the distance travelled back and the strength amplitude
of the pulse as a signal on the CRT screen.
In essence, a UT set transmits energy into a material via a probe and measures
the time in microseconds that the sound pulse takes to return to the probe. The
controls on the UT set are almost entirely concerned with presenting a display
on the CRT screen for the operator to interpret.
Sweep
Timer generator
Pulser
H
Amplifier
V
Probe
NDT4-51015
Ultrasonic Flaw Detector 11-1 Copyright © TWI Ltd
A positive potential electric charge is positioned further down the vacuum tube
and the negatively charged electrons from the filament are attracted towards it.
The electrons pass through a negatively charged focusing ring which pushes
them towards the centre of the tube, forcing them into a fine stream. This
stream of electrons hits a phosphor-covered screen at the end of the tube. The
electron bombardment forces the phosphor to give out light and a green dot
appears on the screen.
The X and Y plates above, below and beside the electron stream carry
potentials that move the electron stream from side to side and up and down,
moving the green dot on the screen.
The X plates control horizontal movement and the Y plates control vertical
movement. By altering the potential of the X and Y plates, the dot can be
moved on the screen.
The PRF on most sets is about 1000 pulses/s, though this can be varied on
most sets from 50 for thick specimens to 1250 pulses/s for thinner specimens.
The pulse generator sends the pulse to the time base generator on the CRT and
to the pulse transmitter.
The time base generator sends the green dot moving across the CRT screen by
generating a charge in the X plates in the tube.
Simultaneously the pulse transmitter sends an electric voltage down the coaxial
cable to the piezo-electric crystal in the probe. The crystal vibrates, transmitting
the pulse of sound into the test material.
At the end of each pulse, the green dot on the CRT screen flies back to the left-
hand side of the screen to await the next pulse.
If the test material is thick, the dot must travel across the screen fairly slowly,
as the pulse repetition rate is lowered. Only one pulse may be in the test
material at any one time or confusing echoes will result. For this reason the PRF
is lowered when thicker specimens are examined.
If the speed of the dot across the screen is not even, as a result of equipment
failure, we say the time base is not linear. Flaw detectors should be checked
frequently to assess time base linearity.
NDT4-51015
Ultrasonic Flaw Detector 11-2 Copyright © TWI Ltd
11.5 Delay
The delay control makes the time base generator wait before sending the green
dot moving across the screen.
Twin crystal and angle probes have Perspex blocks or wedges between the
crystal and the test material. This need not be shown on the CRT screen, so we
adjust the delay to move it sideways off the display, so the passage of the
ultrasound through the Perspex in the probe does not appear on the screen.
You can also use the delay control to wait until the sound has travelled part of
the way through the test piece itself before representation on the screen. For
example, if you only want to look at the bottom 25mm of a 200mm specimen,
you can adjust the delay so that the green dot begins to travel across the
screen at 175mm. During thickness checks, this can make for more accurate
readings for thicker specimens.
The current returning to the set goes to an amplifier which increases it and
filters out irrelevant signals. The returning current is alternating (AC) and must
therefore pass through the rectifier before going to the CRT.
The rectified current now goes to the attenuator, which uses a variable
resistance to control the current passed on to the CRT. The greater the
resistance, the smaller the current. This attenuator is controlled by the
calibrated gain/attenuator control on the set.
From the attenuator, the current goes to the Y plates in the CRT. When the
current hits the Y plates, they pull the electron stream upwards and the green
dot jumps from the bottom of the screen to make a signal. The height of the
signal is increased or decreased by turning the gain up or down.
We can find the latter by comparing a signal from the reflector in the test piece
with a signal from an artificial reflector in a reference block.
NDT4-51015
Ultrasonic Flaw Detector 11-3 Copyright © TWI Ltd
11.8 The decibel
Gain is measured in decibel (dB) - tenths of a unit called a bel. When we
compare the height of two signals on the CRT screen, we are in fact comparing
the electric voltage that is being sent to the Y plates; electric voltage is
proportional to the square of the current. To compare two signals, we must use
a formula that takes account of this fact (see Section 1.4).
NDT4-51015
Ultrasonic Flaw Detector 11-4 Copyright © TWI Ltd
Section 12
Calibration and Sensitivity
12 Calibration and Sensitivity
Angle probes, initial checks and calibration
Before we can start to use an angle probe, we need to find out more about it.
For instance:
So we must check the probe before we can calibrate the time base for use. The
following paragraphs also describe a number of other performance checks which
should be carried out at specified intervals.
To find the probe index, place the probe on a Calibration Block No 1 (see BS EN
ISO 2400 Ultrasonic Testing Specification for Calibration Block No 1), also
referred to as a V1 Block, and obtain an echo from the 100mm radius and
establish it at more than 50% FSH using the gain control. Maximise the echo by
moving the probe backwards and forwards. Mark a line on each side of the
probe directly above the slots which indicate the centre of the 100mm radius.
This is the probe index, where the axis of the beam leaves the Perspex shoe.
NDT4-51015
Calibration and Sensitivity 12-1 Copyright © TWI Ltd
12.2 Checking the probe angle
For a 45 or 60 probe, place it on the Calibration Block No 1, approximately
adjacent to where the appropriate angle is inscribed, directed towards the
plastic insert. Obtain a signal on the screen from the plastic insert and
maximise it. Find the position where the probe index coincides with the angle
indicated on the side of the No 1 Block and this will tell you the probe angle.
This procedure can be repeated for a 70 probe, but reflecting the energy from
the plastic insert radius is unreliable. Therefore we suggest you use the 1.5mm
hole as a target.
It is possible on some flaw detectors to calibrate the time base to 100mm range
from the Calibration Block No 1. However, this involves delaying the signal by
100mm and not all equipment can do this on the appropriate scale expansion
setting, so we will confine ourselves to calibrating for 200mm full screen width.
Place the probe on the Calibration Block No 1 and obtain a boundary echo from
the 100mm radius. Establish this signal to more than 50%FSH using the gain
control. Further maximise the echo by moving the probe backwards and
forwards, then keep the probe stationary.
NDT4-51015
Calibration and Sensitivity 12-2 Copyright © TWI Ltd
Place the signal from 100mm at 5 (half scale) on the time base and the one
from 200mm at 10 (full scale), using the delay and range controls. The time
base is calibrated for 200mm; longer ranges can be catered for in multiples of
100mm.
However, the Calibration Block No 1 is bulky and inconvenient for site work, so
it is not always possible to calibrate for 100mm and we tend to use the
Calibration Block No 2 (also referred to as the V2 Block).
5 10
Figure 12.3: Calibration of the time base (range) of the flaw detector:
a Schematic drawing of shear wave probe positioned on Block No 2;
b ‘A’ scan display.
The time base now represents 100mm. Check this by turning the probe around
and pointing it at the 50mm arc. If you have calibrated correctly, the signal
when maximised will come up exactly in the middle of the screen at 5.
NDT4-51015
Calibration and Sensitivity 12-3 Copyright © TWI Ltd
2.5 10
a c
2.5 6.25 10
a b
NDT4-51015
Calibration and Sensitivity 12-4 Copyright © TWI Ltd
Calibration for 250mm
Point the probe at the 25mm radius arc on a Calibration Block No 2 (V2 Block)
and adjust the set until you get four echoes. These represent 25, 100, 175 and
250mm. Maximise these signals by sliding the probe forward and backward.
Adjust range and delay until the first echo comes one tenth of the way across
the screen at 1 and the fourth echo comes at the extreme edge of screen at 10.
Check on a Calibration Block No 1. On the 100mm arc you should get one echo
4/10 across the screen and the other 7/10 across the screen.
1 4 7 10
a b
NDT4-51015
Calibration and Sensitivity 12-5 Copyright © TWI Ltd
Section 13
Flaw Location
13 Flaw Location
You can calculate the location of a flaw by using trigonometric formulas as
shown below. You need to know the angle of the probe and the stand-off
measured from the centre of a weld.
Stand-off
Depth
Range
Figure 13.1 Reflector depth and probe stand-off as a function of probe angle.
A general rule of thumb used to calculate the depth of an indication from the
range on the screen is:
It is quicker and easier, however, to use a flaw location slide and a beam plot or
even a piece of clear plastic film with the probe angle drawn on it (see Figure
13.2). Use the slide as follows:
a Draw a cross section of the weld on the transparent outer envelope of the
slide.
b Draw a mirror image of the weld cross section immediately under it if the
sound energy is going to bounce off the backwall, ie using full skip.
c Use the printed datum line on the plastic envelope as the centre of the weld
and measure all stand-offs from it.
d Maximise the echo from a defect and mark where the index point falls on
the parent metal. Measure its distance from the centre of the weld.
Note: The defect on a sketch as well as the stand-off and range of the centre of
the defect.
e Move the weld datum line on the plastic envelope to the stand-off distance.
f Look along the centre of the beam plot until you come to the range shown
on the screen.
g Make a mark on the envelope; this represents the centre of the defect. It
shows the position of the defect in the weld body.
NDT4-51015
Flaw Location 13-1 Copyright © TWI Ltd
Stand-off
Range
Stand-off
Range
Figure 13.2 Determination of flaw location using the flaw location slide:
a Probe position for ‘full skip’ into the weld for detection of side wall defect;
b Flaw location slide showing graduated range, stand-off and defect depth.
NDT4-51015
Flaw Location 13-2 Copyright © TWI Ltd
Section 14
Flaw Sizing
14 Flaw Sizing
14.1 The 6dB drop sizing method
This method is used for sizing large reflectors. If the probe is moved until the
signal amplitude from a reflector drops to half its original screen height, then it
can be said that the sound beam is half on and half off the reflector. So by
moving the probe until the signal from the end of a large reflector halves in
height, we can estimate that the edge of the reflector is immediately below the
centre of the probe.
This method is called the 6dB drop sizing method because the amplitude of the
signal drops by half, which corresponds to 6dB, when the probe is moved to the
edge of a large reflector.
Note: The last peak on the screen before the probe goes off the end of the
reflector is usually considered as the peak of the reflector, rather than the
maximum signal from the reflector.
a c
Probe
Defect
b d
6dB
BWE
Defect
signal
NDT4-51015
Flaw Sizing 14-1 Copyright © TWI Ltd
14.2 The 20dB drop sizing method
We can use a beam plot to find the edge of a defect by using the edge of the
sound beam.
If we know the width of a beam at a certain distance from the crystal, we can
mark the distance across a defect from where the extreme edges of the beam
touch each end of the defect and then subtract the beam width to get the
defect size.
When the signal from the defect drops by 20dB from its peak, we judge that the
edge of the beam is just touching the end through-thickness extremity of the
defect. We can find the width of the sound beam at that range by consulting the
beam plot we have made.
Note: The last peak on the screen before the probe goes off the end of the
defect is usually considered as the peak of the defect, rather than the maximum
signal from the defect.
100%
10%
a b c
NDT4-51015
Flaw Sizing 14-2 Copyright © TWI Ltd
14.3 Construction of a beam edge plot - 20dB
Find the hole at a depth of 13mm on an IOW block with a 0 probe and
maximise the signal. Move the probe until you get the highest signal you can
from the hole, then turn the signal to FSH using gain. Mark the position of the
middle of the probe on the side of the block.
13mm
Move the probe to one side until the signal drops to 10% FSH (-20dB) and mark
the centre of the probe on the side of the block.
Move the probe to the other side of the hole until the signal drops to 10% FSH
(-20dB) and mark the centre of the probe on the block.
Use the distances between the marks on the block to plot the beam on a piece
of graph paper. Measure 13mm depth on the paper then mark the distances of
the probe centre at -20dB from the beam centre at 100% FSH on either side.
13mm
Now find the 25mm hole and maximise the signal, turning it to 100% FSH.
Move the probe to either side of the hole, marking the centre of the probe on
the side of the block where the signal drops by 20dB.
Measure 25mm on the paper and use the distances on the block to plot the
beam dimensions at 25mm.
NDT4-51015
Flaw Sizing 14-3 Copyright © TWI Ltd
13mm
25mm
Repeat using the 32mm hole. Join up the points marking the probe centre at
20dB to obtain a beam plot.
13mm
25mm
32mm
Note: We have only drawn the beam width in one plane, so the probe must be
marked accordingly and used to measure defects in this plane. We use
knowledge of the beam spread to size defects, find their edges and hence their
width, length and sometimes orientation.
NDT4-51015
Flaw Sizing 14-4 Copyright © TWI Ltd
Figure 14.4 IOW reference block with 1.5mm side-drilled holes at different
depths.
We will start with a 60 probe. Using the probe, find the hole which is 13mm
below the top surface and maximise the signal to 100% FSH. Mark where the
index point comes on the block with a pencil or crayon. Move the probe forward
until the signal drops to one tenth screen height (20dB drop). Make a second
mark on the block where the index point on the probe stands on the block.
100%
10%
Move the probe backwards until the signal maximises and then drops down to
1/10 screen height. Mark where the index point now stands and draw a vertical
line on the block from the hole to the upper edge. Measure the distances of the
three index point marks from the top of the line and note them down.
Now find an echo from the 19mm deep hole and repeat the process, noting the
distances and repeat the process a third time using the 25mm hole.
Take the slide out of a beam plotting chart and draw three faint lines across it
at depths of 13, 19 and 25mm.
Transfer the distances of the index points from the vertical lines to the relevant
pencil lines on the chart. Join the marks up. The centre line represents the main
NDT4-51015
Flaw Sizing 14-5 Copyright © TWI Ltd
energy of the beam and the other two marks represent the leading and trailing
edges of the beam.
With a 45 probe, use the 19, 25 and 32mm depth holes as the 13mm hole may
be in the probes near zone.
Use a 10dB drop with a 70 probe and instead of dropping the signal to 1/10
FSH for the leading and trailing edges, use the 3/10 line on the screen.
On the cover of the beam plotting chart, use the corner of the block to
represent the centre line.
Obtain a signal from the top hole of the six, maximise it, then push the probe
towards the block corner until the signal drops to 1/10 FSH (3/10 for a 70
probe). Mark where the index point occurs on the block and measure the stand-
off. Note the range of the reflector on the screen.
Use the stand-off and the range to plot the defect along the trailing edge of
your beam spread. Mark it on the slide cover.
Now obtain a signal from the bottom hole of the six, maximise it and turn it up
to FSH on your screen. Pull the probe back until the signal drops to 1/10 FSH
(3/10 for a 70 probe). Plot the bottom of the defect on your slide cover using
the leading edge of your beam plot.
Lay the transparent slide over the IOW block; the top and bottom of the drilled
holes should coincide with the marks on the slide. If they do not, your beam
plot is off or you are going wrong somewhere. If they do, you have just sized a
defect by the 20dB drop method (10dB drop for the 70 probe).
NDT4-51015
Flaw Sizing 14-6 Copyright © TWI Ltd
Figure 14.6 Example of bespoke reference block to check beam characteristics
including beam edge resolution and near-zone approximation.
However, the beam starts to travel in the plastic wedge and is then refracted
and carries on in the material being tested. We are only concerned with the part
of the beam near zone registering later than zero on the time base, ie in the
test material. This is called the modified near zone.
Here is an example:
A 5MHz shear probe has a 10mm diameter crystal. The beam travels in Perspex
for 10mm. What is the modified near zone?
D2 f
NZ, if totally developed in steel =
4V
10 2 x 5 x 1,000,000
= mm = 38.46mm
4 x 3250 x 1000
We must now subtract the Perspex wedge part of the beam which is 10mm,
multiplied by the ratio of the Perspex and steel velocities which is
10 x 2730/3250 = 8.4mm
NDT4-51015
Flaw Sizing 14-7 Copyright © TWI Ltd
Method:
a Place the probe to pick up the 1.5mm hole at ½ skip and maximise the
signal from the intersection of the hole and the opposite face. Mark the
straight edge adjacent to the near centre of the probe to indicate the beam
centre.
b Position a straight edge either in front of or behind the probe to hold the
probe in the fixed transverse position. Scan the probe laterally (sideways)
until the hole signal drops by 20dB. Mark on the straight edge adjacent to
the rear centre position of the probe. This registers half a beam at the ½
skip range.
c Scan the probe laterally the other way, through the maximum signal
position, until the hole signal again drops by 20dB. Mark the straight edge
as before.
d You now have three marks on the straight edge to indicate the beam width
at that range. Transfer these to the beam plotting chart as appropriate.
e Repeat steps a-d, but at full skip and 1½ skip for a 45 probe (only at full
skip for a 60 probe). Note that mode conversion reduces the 1½ skip signal
on a 60 probe to too low a level to be reliable.
f Join up the three points on either side of the centre line to complete the
beam. Only take the lines back to the near zone because the edge is not
reliable beyond that.
NDT4-51015
Flaw Sizing 14-8 Copyright © TWI Ltd
Section 15
Sensitivity Setting
15 Sensitivity Setting
Setting a sensitivity level is essential to providing reproducible results when the
same inspection is carried out by different operators, using different probe set
combinations and maybe working in different locations. They must all see the
same flaw giving the same signal height and therefore have the same data on
which to base their accept/reject decisions.
There are several systems for setting sensitivity. We have already encountered
the method of setting the first backwall echo (BWE) to FSH for lamination
checks. However, when checking plates adjacent to a weld, the second BWE
should be set to FSH.
When setting the sensitivity, we must be sure that a signal from a defect will be
visible on the CRT screen and that we will be able to distinguish the defect
signal from background noise or grass. All UT sets differ slightly, so we cannot
say, ‘Set the sensitivity to xdB’, as different probes and equipment will give
entirely different signals from the same reflector. The methods of setting
sensitivity have evolved to attain some uniformity.
Different methods are used in different places. At TWI, the IOW block is used as
the recommended method for PCN examinations. On North Sea contracts,
either the distance amplitude correction (DAC) curve or the American Society of
Mechanical Engineering (ASME) curve is used. The DAC method is
recommended in BS EN ISO 17640 (Non-destructive testing of welds –
ultrasonic testing techniques, testing levels and assessment), while in Germany
the distance gain size (DGS) system is usually applied, especially when
evaluating small reflectors.
The purpose of sensitivity setting is to find a gain level sufficient to find a flaw
and depends on the:
Find a hole on the block that approximately coincides with the thickness of the
material you are testing. Double the thickness if you are examining at full skip,
ie bouncing your sound beam off the backwall.
Obtain a signal from the hole and turn the gain control until the signal is at
FSH.
You have now set the sensitivity and can be assured that flaws having the
equivalent reflectivity of 1.5mm side-drilled holes will appear on the screen.
NDT4-51015
Sensitivity Setting 15-1 Copyright © TWI Ltd
This method has several advantages:
Simple to use.
Provides a uniform system of reference.
A fairly large and visible echo is assured from small flaws.
Side-drilled hole reflectors are independent of angle.
Figure 15.1 The IOW reference block containing a series of 1.5mm SDH at
different depths. A 60° angle probe is shown in position on the block.
NDT4-51015
Sensitivity Setting 15-2 Copyright © TWI Ltd
The procedure described in BS EN ISO 17640 for constructing a DAC curve is:
Calibrate the time base for the maximum sound path length to be used.
Adjust the gain so that the amplitude from the series of reflectors falls
between 20 and 80% FSH.
Without altering the gain setting, maximise the amplitude of each reflector
in turn and mark the tip of the signal, either on the screen or on a
transparent overlay.
Record the gain setting used for plotting the DAC curve and reference this to
some other reflector, such as the radius in a V calibration block. This action
enables the gain to be reset without the reference block.
1 2 3 4 5
Figure 15.3 DAC construction from a series of reflector SDHs to cover the
thickness range of the test component.
Note: Should the difference in height between the largest and the smallest
echoes exceed the range of 20-80%FSH, the line shall be split and separate
curves plotted at different gain settings. The difference in gain between the two
curves shall be noted.
Advantages
A quick way of accepting or rejecting discontinuities without too much time
consumed in sizing reflectors.
Some idea can be gained of the size of the discontinuities in relation to the
reference holes.
Uniformity provided by all technicians constructing their curves from the
same test block.
Disadvantages
Curves must be constructed for each probe in conjunction with each UT set.
Transfer correction must be worked out.
When setting sensitivities, the specification or technique will specify the block to
be used and the amplitude of the signal to be obtained from the FBH.
NDT4-51015
Sensitivity Setting 15-3 Copyright © TWI Ltd
Blocks are cut for use with 0 probes or angle probes in different materials. This
method is mostly used in aerospace applications.
Advantages
Easy to use.
Uniformity assured when different technicians use the same blocks.
Blocks can be made from different materials.
Disadvantages
Fairly rigid system for specific applications.
Large number of blocks needed for different settings.
Transfer correction usually needed.
Blocks for angle probes are rarely cut exactly normal to the beam.
You may hear these blocks called HITT or ALCOA blocks, after the originators.
Advantages
Quick and easy.
No reference block is needed.
Any defect larger than the material grains will show up.
No transfer correction needed.
Disadvantages
No accurate sizing of the defect.
Discontinuities near the surface of the test material may be hidden in the
grass.
Therefore, the artificial defects in a reference block may give higher amplitude
signals (anything up to 6dB or even more) than signals from similarly sized
discontinuities in the test materials.
Allowances have to be made for this and corrections made for different
surfaces. This allowance is named transfer correction or transfer loss.
NDT4-51015
Sensitivity Setting 15-4 Copyright © TWI Ltd
There are several methods of determining the transfer correction, some
requiring the construction of separate DAC curves and some requiring
calculation according to formulae. Two simple methods are explained below.
Place the two probes opposite each other on the reference block with one probe
transmitting and the other receiving, so that the sound energy is bounced off
the backwall and caught by the receiving probe (pitch and catch).
Figure 15.4 Schematic drawing of the transfer correction method for angle
probes.
Place the two probes on a piece of test material of the same thickness as the
reference block and repeat the process.
Note the difference between the two gain settings. This is the transfer
correction needed.
Reflector sizes are expressed in terms of the probe diameter and distances from
the probe are expressed as multiples of the near zone.
If you have a signal from a flaw at a certain depth, you can compare the signal
size to what the signal of a BWE should be at that depth and estimate the size
of a flat-bottomed hole that would give such a signal at that depth. The defect
can then be sized according to its flat-bottomed hole equivalent.
NDT4-51015
Sensitivity Setting 15-5 Copyright © TWI Ltd
The attenuation factor for the test material must be taken into consideration
when using the DGS system.
Example
You are using a 5MHz 10mm diameter compression probe on a 100mm steel
plate and you find a defect at 60mm depth which gives a signal at FSH with a
30dB gain setting. What is its flat-bottomed hole equivalent?
First, work out the probe near zone. It is 21mm, so the defect is at a distance
of three near zones.
Now get a BWE and find what the dB reading is. Say it is 20dB when the BWE is
at FSH. 100mm is five near zones. What will it be at 60mm – three near zones?
Refer to the DGS curves. If the BWE is at FSH with 20dB at 100mm, by the law
of large reflectors and according to the BWE line on the DGS curves, a BWE at
60mm should reach FSH at 16dB - 4dB less than at 100mm.
The signal height from the flaw is 30dB, which is 14dB more than the BWE.
Look down the scale 14dB at three near zones from the BWE and you find that
the nearest line is at 0.5 of the probe diameter. The probe diameter is 10mm so
the nearest equivalent flat bottomed hole to the flaw had a diameter of 5mm.
Advantages
Can choose a gain level for sizing.
Tells you the smallest defect you can find at a given range.
Provides the basis for an accept/reject system.
Gives a rough equivalent to the size of a flaw.
Uniformity between results from different technicians.
Disadvantages
Operators must keep referring to a chart and making calculations.
Attenuation must be taken into account.
No account is taken of the flaw orientation.
Most effective on small defects.
An equivalence system, not a sizing system.
Flaw surfaces and shapes are not ideal reflectors; therefore, the signal
amplitudes are not the same as those of a comparable flat-bottomed hole.
For angle probes, plastic slides have been manufactured by Krautkramer to fit
over the CRT screen. The set is calibrated and gain setting is performed by
bringing the BWE or the echo from the 1.5mm hole on the V2 block up to marks
on the slide. Flat-bottomed hole equivalents for flaws can then be read straight
off the slide. The DGS system is widely used in Germany.
NDT4-51015
Sensitivity Setting 15-6 Copyright © TWI Ltd
15.7 Signal-to-noise ratio (SNR)
It has been mentioned elsewhere that frequency and wavelength have a major
influence on flaw detection. However, the detection of a defect is also
influenced by many other factors. The amount of sound that reflects from a
defect is, for example, dependent on the acoustic impedance mismatch between
the flaw and the surrounding material. A gas-filled defect such as a lack of
fusion is generally a better reflector than a metallic inclusion because the
difference in acoustic impedance is greater between air and metal than between
a metal and another metal.
The nature of the surrounding material also greatly affects the detection of
defects with coarse-grain structures, reducing defect detectability. A measure of
detectability of a flaw and the effect of the many factors involved is its signal-
to-noise ratio (SNR). The SNR is a measure of how the signal from the defect
compares to other background reflections (categorised as noise). An SNR of 3-1
is often required as a minimum.
The absolute noise level and the absolute strength of an echo from a small
defect depend on a number of factors:
General factors to consider with respect to SNR and therefore defect detection:
NDT4-51015
Sensitivity Setting 15-7 Copyright © TWI Ltd
Section 16
Ultrasonic Equipment Checks
16 Ultrasonic Equipment Checks
16.1 Linearity of time base
General
This check may be carried out using a standard calibration block, eg Block 1
(see BS EN ISO 2400), and a compression wave probe. The linearity should be
checked over a range at least equal to that which is to be used in subsequent
testing.
Method
a Place the probe on the 25mm thickness of Calibration Block 1 and adjust the
controls to display ten BWEs.
b Adjust the controls so that the first and last BWEs coincide with the scale
marks at 1 and 10.
c Increase the gain to bring successive backwall echoes to 80% FSH. The
leading edge of each echo should line up with the appropriate graticule line.
d Record any deviations at approximately half screen height. Deviations
should be expressed as a percentage of the range between the first and last
echoes displayed (ie 225mm).
Tolerance
Unless otherwise specified by the testing standard, a tolerance of ±2% is
considered acceptable.
Frequency of checking
This check shall be carried out at least once per week.
Figure 16.1 ‘A’ scan flaw detector showing signal amplitude vs distance (transit
time) for linearity of time base checks.
NDT4-51015
Ultrasonic Equipment Checks 16-1 Copyright © TWI Ltd
Figure 16.2 ‘A’ scan (echo amplitude at approximately 80% vs depth distance)
presentations.
NDT4-51015
Ultrasonic Equipment Checks 16-2 Copyright © TWI Ltd
16.2 Linearity of equipment gain
General
This is a check on both the linearity of the amplifier within the set and the
calibrated gain control. It can be carried out on any calibration block containing
a side-drilled hole; the probe should be the same that is used in subsequent
testing. Reject/suppression controls shall be switched off.
Method
Position the probe on a calibration block to obtain a reflected signal from a
small reflector, eg the 1.5mm hole in Calibration Block No 1.
Adjust the gain to set this signal to 80% FSH and note the gain setting (dB).
- Increase the gain by 2dB and record the amplitude of the signal.
- Remove the 2dB again and return the signal to 80% FSH.
- Reduce the gain by 6dB and record the signal amplitude.
- Reduce the gain by a further 12dB (18 in total) and record the signal
amplitude.
- Reduce the gain by a further 6dB (24 in total) and record the signal
amplitude.
Tolerance
Table 16.1 Expected screen height results and limits for equipment gain check.
Gain, Expected screen Recorded Limits
dB height (%) amplitude
0 80 (Reference line)
-6 40 37-43%
-12 20 17-23%
-18 10 8-12%
NDT4-51015
Ultrasonic Equipment Checks 16-3 Copyright © TWI Ltd
Frequency of checking
The check shall be carried out at least once per week.
Figure 16.3 ‘A’ scan presentations showing respective echoes set at 80% FSH.
Method
a Position the probe on the appropriate side of the block to obtain a reflection
from the quadrant.
b Move the probe backwards and forwards to maximise the amplitude of the
reflected signal.
c When the signal is at maximum, the probe index will correspond to the
engraved line on the block. Mark this position on the side of the probe.
NDT4-51015
Ultrasonic Equipment Checks 16-4 Copyright © TWI Ltd
Tolerance
The tolerance depends on the application, but for plotting of defects it is
recommended that the probe index be accurate to within 1mm.
Frequency of checking
When a probe is in continuous use, it is recommended that the check be carried
out every few hours; otherwise, a daily check is recommended.
NDT4-51015
Ultrasonic Equipment Checks 16-5 Copyright © TWI Ltd
Figure 16.4 Amplifier and gain control checks at 0db to 24db.
NDT4-51015
Ultrasonic Equipment Checks 16-6 Copyright © TWI Ltd
16.4 Beam angle
General
The beam angle can be checked on several calibration blocks, eg Calibration
Block No 1 (BS EN ISO 2400) or Beam Calibration Block (BCB, A5 Block). The
beam angle check shall preferably be made on a probe in conjunction with the
flaw detector to be used in subsequent testing.
Method
a Place the probe in such a position as to receive a reflected signal from the
selected transverse hole in the calibration block (eg the 19mm deep hole in
the BCB).
b Maximise the signal from the hole and mark the index point of the probe on
the block.
c Measure the distance from the marked point on the block to the edge of the
block. Knowing the position of the drilled hole will allow the beam angle to
be calculated (see below).
Note If only a rapid check is required, maximise the signal from the 50mm hole
in Calibration Block No 1. The angle can then be assessed by visual
interpolation between the reference markings on the block.
Tolerance
The accuracy achieved by the described method is 1.5.
The accuracy in this case can only be assumed to be 3.
Frequency of checking
When a probe is in continuous use, it is recommended that the check be carried
out at least every few hours; otherwise, a daily check is recommended.
Note 1: A demonstration of the sensitivity of the probe and the flaw detector
on a calibration block does not guarantee that the same size of reflector could
be detected in the workpiece.
NDT4-51015
Ultrasonic Equipment Checks 16-7 Copyright © TWI Ltd
Tolerance
The tolerance depends on the application. Any deterioration in the sensitivity
value indicates a problem with the probe or flaw detector. A low SNR would be
typical of a coarse-grained material.
Frequency of checking
Unless otherwise agreed, the check shall be carried out once per probe per day.
Method
a Calibrate the time base in millimetres to a range that is to be used in
subsequent testing.
b Maximise the signal from the 1.5mm side-drilled hole for shear wave probes
or a BWE for compression wave probes and set its peak to 100% screen
height.
c Measure the width of the signal in millimetres at the 10% screen height
position.
d If desired, the measurement in millimetres can be converted into
microseconds by dividing it by the relevant sound velocity.
Tolerance
The tolerance depends upon the application. A long pulse duration will limit
range resolution and indicate the need for a resolution check*, while a short
pulse duration may indicate that the flaw detector has built-in suppression that
could prevent the observation of small signals.
NDT4-51015
Ultrasonic Equipment Checks 16-8 Copyright © TWI Ltd
Frequency of checking
Unless otherwise agreed, the check should be carried out daily.
20%
Method
a Calibrate the time base to a range of 0-100mm for either the compression
or the shear wave probe.
b Place the probe so that the axis of the beam impinges upon the 2mm step in
the A7 ‘resolution’ calibration block for shear wave probes, or the 3mm step
for compression wave probes.
c Adjust the position of the probe so that the echoes from the two targets are
of the same height (approximately half the full graticule height).
d The steps are said to be resolved when their echoes are clearly separated at
half maximum echo height or lower.
Note: The 3mm steps between the 9mm and 3mm drilled holes in the A6
calibration block may also be used when checking compression probes.
NDT4-51015
Ultrasonic Equipment Checks 16-9 Copyright © TWI Ltd
Frequency of checking
This check shall be carried out monthly, or when too long a pulse duration is
suspected.
a b
c d
Figure 16.6 Flaw detector and probe system resolving power (resolution) check
to give separate indications of closely situated reflections (note how
discontinuities a and b above give clear echo resolution).
NDT4-51015
Ultrasonic Equipment Checks 16-10 Copyright © TWI Ltd
Section 17
Practical Weld Inspection
17 Practical Welding Inspection
Equipment (instrumental and probe) sensitivity is set in accordance with one of
the methods described in Section 15, with calibration covered in Section 12 and
flaw location and sizing covered in Sections 13 and 14, respectively.
BS EN ISO
Ultrasonic testing Characterisation of indications in welds
23279
BS EN ISO Characterisation and sizing of
Ultrasonic testing
16827 discontinuities
Echo amplitude.
Directional reflectivity.
Echostatic pattern (A scan).
Echodynamic pattern.
The first stage of assessing the echo amplitude involves comparing the
amplitude of an indication to DAC level and classifying it into one of the four
categories shown in the table below.
NDT4-51015
Practical Weld Inspection 17-1 Copyright © TWI Ltd
Indications falling into the S1 category would be immediately discounted. All
other indications would then proceed to be assessed for directional reflectivity,
which is defined as the variation in echo amplitude from a discontinuity in
relation to the angle at which the ultrasonic beam is incident upon it. A
spherical indication would show the same echo amplitude over a wide range of
incident angles, eg 45o, 60o and 70o, and is said to have low directional
reflectivity. A large smooth planar reflector would show a great variation in
echo amplitude and would therefore be said to have high directional reflectivity.
The next two stages of the process analyse firstly the shape of the signal as
displayed on the A scan equipment and finally the behaviour of the signal when
the probe is scanned at 90o to the discontinuity (traversing). Echostatic patterns
are categorised as:
a b c
NDT4-51015
Practical Weld Inspection 17-2 Copyright © TWI Ltd
With respect to echodynamic patterns, indications fall into one of five categories
dependent upon the changes observed in the signal on the A scan in response
to probe movement. To aid the identification of defects, there are four basic
probe movements:
Lateral
Traversing
a b
c d
Orbital Rotational
NDT4-51015
Practical Weld Inspection 17-3 Copyright © TWI Ltd
Variations in signal amplitude
Pattern 1
Pattern 2
Pattern 3a
NDT4-51015
Practical Weld Inspection 17-4 Copyright © TWI Ltd
Variations in signal amplitude
Pattern 3b
NDT4-51015
Practical Weld Inspection 17-5 Copyright © TWI Ltd
Section 18
Root Flaws
18 Root Flaws
The signal from a root flaw will appear on the time base while you are scanning
laterally along a straight edge, at a fixed position from the root. Once the signal
is maximised by getting the best reflection from the flaw, it can be assumed
that the centre of the beam is hitting the bottom of the flaw. Fine adjustment of
the straight edge will perhaps be necessary.
C/L C/L
C/L C/L
NDT4-51015
Root Flaws 18-1 Copyright © TWI Ltd
18.3 Root crack
Usually high amplitude response with fir tree appearance.
Probe movement - orbit, echo held over large angle.
Lateral, echo held with multi-range signals and variations on time base.
Measurement - 6dB for length. Traverse forward with 20dB for height.
C/L C/L
C/L C/L
18.5 Misalignment
Large single echo from one side. No echo from opposite side.
Probe movement - traverse back echo falls rapidly.
Measurement - lateral for 6dB drop.
C/L C/L
NDT4-51015
Root Flaws 18-2 Copyright © TWI Ltd
Section 19
Face and Body Flaws
19 Face and Body Flaws
A degree of additional flexibility can be applied when flaws are located at the
face or in the body of a weld. The diagrams below illustrate the characteristic
shape of the screen presentation, but they are ideal rather an actual.
Probe movement:
Rotate or orbit - echo falls rapidly.
Lateral or traversing - echo height held.
Measurement: For depth use 20dB. For length use 6 or 20dB.
19.2 Crack
Multiple peak reflector: usually high amplitude, but dependent on type of
crack and size; echo with fir tree appearance.
Probe movement:
Orbit - echo over larger angle than with fusion defects.
Lateral - signal held with varying height.
Measurement: For length use 6 or 20dB. For depth use 20dB.
NDT4-51015
Face and Body Flaws 19-1 Copyright © TWI Ltd
19.3 Gas pore
Spherical even reflector: single peak echo, narrow profile, similar to drilled
hole or radius of calibration block intensity (approximately 50% CRT).
Probe movement:
Rotate, lateral and traversing - echo falls rapidly.
Orbit - echo height remains.
Measurement: Impractical to measure height and length. Report as isolated
reflector. Equate reflectivity against disc area or DGS.
19.4 Porosity
Multiple peak echo: Low intensity (20% CRT), broad at time base due to
numerous ranges.
Probe movement: Orbit - echoes held with amplitude variations.
Measurement: Indicate area by pinpointing last maximum signal from
traversing and lateral scans.
NDT4-51015
Face and Body Flaws 19-2 Copyright © TWI Ltd
19.5 Linear inclusion (slag)
Echo may be wide at time base and will be multi-faceted, due to having more
than a single range. Height will vary between 20-90%.
Probe movement:
Orbit (traversing is similar) - echo held with various maxima and minima.
Rotational - echo will drop quickly.
Lateral - will produce large variations in height, perhaps with total loss of
signal for distances shorter than the beam width.
NDT4-51015
Face and Body Flaws 19-3 Copyright © TWI Ltd
Section 20
Plate Inspection
20 Plate Inspection
EN 10160 describes methods for inspecting uncoated steel plates between 6
and 200mm in thickness for internal discontinuities. It details:
NDT4-51015
Plate Inspection 20-1 Copyright © TWI Ltd
Section 21
Inspection Procedure
21 Inspection Procedure
BS EN ISO 17640 (NDT of welds – Ultrasonic testing – Techniques, testing,
levels and assessment) describes the procedure for examining welds and details
beam paths to be used for welds of different configurations, eg plates, pipes,
nozzles and nodes. To standardise methods of examination, BS EN ISO 17640
and EN ISO 16811 (NDT - Ultrasonic testing – Sensitivity and range setting)
recommend the use of DACs and different sensitivity settings to match the
criticality of different examinations. The sensitivity is set higher for examining,
say, a high-pressure steam pipe in a chemical plant than it is for a
comparatively low-pressure line in a refinery.
The following is the procedure laid down in BS EN ISO 17640, recommended for
the inspection of examination test pieces:
Compression scan to check the parent metal on either side of the weld for
laminations and to check through-thickness dimensions.
Root scan to check the root for longitudinal defects such as lack of
penetration, lack of root fusion, cracking or mismatch.
Weld body scan with shear probes to check the sides of the weld and the
weld body for longitudinal defects like lack of fusion, cracks, slag and
porosity.
Transverse weld scan to check the weld for transverse and chevron
cracking.
Reporting.
NDT4-51015
Inspection Procedure 21-1 Copyright © TWI Ltd
EN 17640 also states that it is essential that the operator performing an
inspection on a welded joint shall have access to the following:
EN 17640 itself would, in many cases, satisfy the requirement for a written test
procedure.
Find the centre of the weld. The stand-off must be measured from there. If the
root is detectable, find the root from both sides with a 60 probe. Mark where
the index point falls on both sides of the weld when the root signal is
maximised; the weld centre is midway between these two marks. Note the
thickness of the parent metal on either side of the weld and ensure there is no
counter boring.
C/L
Examine the parent metal on either side of the weld with a 0 compression
probe with sensitivity set at either 2nd BWE to FSH or the sensitivity given in
the procedure. Look for changes in thickness and lamination. You should cover
the parent metal on either side of the weld to full skip distance for the highest
angle probe, that is for a 60 or 70 probe. Mark any discontinuities found in the
parent metal.
NDT4-51015
Inspection Procedure 21-2 Copyright © TWI Ltd
Full skip
Probe
movement
Full skip
Draw a cross section of the weld on the slide of your beam plotting chart once
you are sure of the weld dimensions.
If the cap of the weld has been dressed ground flush with the parent metal,
examine the body of the weld with the 0 probe. If a backing bar has been
used, check bonding at the root if possible. Check the root on a double sided
weld if the cap has been ground flush. Lack of penetration at the root will
sometimes show with a 0 probe.
If you are examining a single V weld with a dressed cap, a 0 probe scan of the
weld can reveal lack of inter-run fusion and large pockets of slag.
Measure out the half skip distance on the parent metal on either side of the
weld, measuring from the centre line of the weld. Draw lines on either side of
the weld at this distance. Note the range of the root centre along the main
beam.
Stand-off
Index
Range
Set the gain from an IOW block so that a hole at the depth of the root will give
an echo to FSH. (Or set the gain according to the set procedure.)
Place the probe index point on the line and put a magnetic ruler or strip behind
it as a guide. Move the probe laterally along the weld. Signals from a good root
should be small, while defects will give large signals. Signals from a fully
penetrated root will usually appear just beyond the range of the root centre.
NDT4-51015
Inspection Procedure 21-3 Copyright © TWI Ltd
Stand-off
Stand-off
Straight
edge
Mark any defects on the parent metal or the magnetic guide behind your probe.
Find the height of the defect using a 20dB drop on the trailing edge of your
beam. Record the defect on a sketch and note the stand-off distance and range.
If a 60 probe is used first, follow up with a 70 probe if the plate thickness is
less than 25mm and a 45 probe if it is over 20mm. There is a bit of an overlap.
Use your beam plot and flaw location slide to work out the half skip distance to
the root centre and full skip distance plus half the weld width to the top of the
weld. Mark these stand-off distances on the parent metal on either side of the
weld. Note the range of the root can cap.
Figure 21.5 Angle probe raster scanning from the weld cap edge to the full skip
position.
Set the gain relative to DAC according to the sensitivity setting given in the
procedure.
Move the 60 probe backwards and forwards along the weld on either side so
that the beam covers the sidewalls and centre body of the weld (see Figure
21.5). Defects should maximise between the root and cap signals. If any signal
NDT4-51015
Inspection Procedure 21-4 Copyright © TWI Ltd
occurs near the cap, use a couplant-covered finger to damp any signal that may
be coming from the cap, thus identifying its position.
Make a second scan of the weld body with a 45 probe for materials with more
than 20mm thickness or a 70 probe for less than 25mm thickness.
Check the area just under the cap with a 45 probe if you are examining thicker
materials. You may find porosity that the 60 probe has not revealed. Check
any defect already noted and see if it is longer than shown with the 60 probe.
On thinner materials, a 70 probe will confirm defects already noted and may
help you distinguish between signals from a defect and signals from, say, the
weld cap or mode change.
Figure 21.6 Shear wave angle probe scan to detect transverse defects.
If the weld has been dressed, push the probe along the centreline of the weld in
both directions and then push it along both edges of the weld in both directions.
Examine the root with a 70 probe straight into the root at 1/4 skip (see Figure
21.7). Run the probe laterally along the weld with sensitivity set from a hole in
the IOW block at a suitable range; this should show any lack of penetration at
the root.
NDT4-51015
Inspection Procedure 21-5 Copyright © TWI Ltd
Make body scans with 60 and 45 probes from both sides of the weld if the
plate is thick (over 25mm) using ¼ to ½ skip stand-off.
Figure 21.8 Weld body scan between a ¼ and ½ skip stand-off using 45° and
60° probes.
On thinner plates, use a 60 and a 70 probe from one side of the weld only
between ¼ and full skip and ½ weld thickness stand-off positions. Take great
care in locating the defects and reporting, as you will have to use a mirror
image on your plotting chart and this can lead to confusion. You may find a
defect and place it on the wrong side of the weld.
21.7 Pipes
Circumferential pipe welds can be examined in the same way as butt welds in
plates are examined, that is:
Compression scan to check the parent metal on either side of the weld for
laminations and to check through-thickness dimensions.
Root scan using a flexible strip at the back of the probe will help to ensure
that the centre of intensity of the beam goes into the centre of the root.
Weld body scan with shear probes to check the sides of the weld and the
weld body for longitudinal defects like lack of fusion, cracks, slag and
porosity.
Transverse weld scan - difficulty may arise here as you must be sure that
the beam reaches the bore of the pipe.
This flexible guide is convenient for marking defects. All defects must be
measured from a datum point. The button at the top of the weld is a convenient
datum point. When examining a longitudinal weld on a pipe or when doing a
transverse check on a circumferential weld, you must choose a probe angle that
will reach the bore of the pipe (see Figure 21.8). The formula below provides
this angle.
NDT4-51015
Inspection Procedure 21-6 Copyright © TWI Ltd
Inside diameter of pipe a
Sine of max probe angle
Outside diameter of pipe
So a 12cm OD pipe with a 1cm wall will need what probe angle?
10
= 0.8333 which is the sine of 56.
12
A 60 probe will not reach the bore of this pipe, so a 45 probe is advised.
NDT4-51015
Inspection Procedure 21-7 Copyright © TWI Ltd
21.8 Reporting
A detailed test report will normally be produced for each item of test and will
cover all of the salient parameters that affect the quality and integrity of the
test as laid out in the test procedure, which must be made available to the test
technician/operator prior to starting the test along with written instructions
detailing the components to be tested, the specific test procedure,
specifications/standards and acceptance criteria to be applied along with any
special instructions that might apply (eg PPE to be used, use of photographs
etc.). The test report will normally include an assessment of the condition of the
component against the specified acceptance criteria.
Specific details that may be included in the test procedure are as follows:
What follows is a sample NDT report. The headings within it are taken from BS
EN ISO 17640, (Ultrasonic Testing – Techniques, testing levels and assessment)
which specifies these as the minimum content.
NDT4-51015
Inspection Procedure 21-8 Copyright © TWI Ltd
Ultrasonic Test Report
Reference
standard
Item Inspected
Material Configuration
Product form Stage of
manufacture
Dimensions Surface
condition
Location of
weld
Specification Date of test
Operator
certification
Equipment
Flaw Detector
Maker Type Serial number
Ultrasonic Probes
Maker Type Frequency Serial number
Couplant
NDT4-51015
Inspection Procedure 21-9 Copyright © TWI Ltd
Technique
Testing levels
Extent of test
Reference point
Identification of probe
position
Time base range
Sensitivity level
Reference level
Parent metal
Acceptance level
Results
NDT4-51015
Inspection Procedure 21-10 Copyright © TWI Ltd
Section 22
Ultrasonic Thickness Measurement
22 Ultrasonic Thickness Measurement
Standards to be used for reference include:
Four ultrasonic measurement modes are given below and illustrated in Figure
22.1:
a Mode 1: Measure the transit time from an initial excitation pulse to a first
returning echo, minus a zero correction to account for the thickness of the
transducer wear surface and the couplant layer (single echo mode).
b Mode 2: Measure the transit time from the end of a delay line to the first
back wall echo (single echo delay line mode).
c Mode 3: Measure the transit time between back wall echoes (multiple
echoes).
d Mode 4: Measure the transit time for a pulse travelling from the transmitter
to a receiver in contact with the back wall (through transmission mode).
NDT4-51015
Ultrasonic Thickness Measurement 22-1 Copyright © TWI Ltd
A
A3
C C
D D
E1
F F1
Mode 1 B Mode 2
B
G
E1 A1 A2
E2
A E3
D C
D H
F E1 E2 E3
Key
A transmit/receive probe B test object F interface echo
(dual element probe) C sound path travel time G delay path
A1 transmit probe D transmission pulse indication H received pulse
A2 receive probe E1-E3 back wall echoes
NDT4-51015
Ultrasonic Thickness Measurement 22-2 Copyright © TWI Ltd
22.2 Requirements – test object, instruments, probes and reference blocks
a Test Object:
The object to be tested shall enable ultrasonic propagation/transmission and
have access to apply the probe to the test surface that will be free of all
dirt, grease, scale or any material that could interfere with the examination.
If the surface to be tested is coated, the coating will have tight adhesion to
the base parent material, otherwise it must be removed.
b Ultrasonic Instruments
The following Ultrasonic measurement instruments can be used:
c Probes
Generally compression (longitudinal) wave probes are used based on either
of the following configurations:
d Reference Blocks
The ultrasonic measurement system (ie instrument and probe) must be
calibrated on a sample(s) or reference block that closely represent the
component whose thickness is to be measured in terms of dimensions,
material and structure and the reference sample thickness range should
cover the range of component thicknesses to be measured. Either the
thickness or the sound velocity of the reference sample block shall be
known.
NDT4-51015
Ultrasonic Thickness Measurement 22-3 Copyright © TWI Ltd
Glossary
Glossary
Acoustic emission (AE). Method of flaw detection which uses an array of accurately
positioned transducers to listen to the structure under stress. The high-frequency sound
given out by a growing crack would be detected by the transducer array.
Angle of incidence. Angle which the axis of an ultrasonic beam makes with the normal
to a tangent plane of a surface at its point of impingement as it travels towards that
surface.
Angle of reflection. Angle which the axis of an ultrasonic beam makes with the normal
to the tangent plane of a reflecting interface, at the point of impingement of the incident
wave, as it travels away from that interface in the same medium.
Angle of refraction. Angle which an ultrasonic beam makes with the normal to a
tangent plane of an interface as it travels away from that interface into the second
material. Note: A synonymous term is beam angle.
Angle of squint. Angle between the side edge of the probe and the projection of the
beam axis on the plane of the probe face. Note: For angle probes this normally relates
to deviation in the lateral direction.
Angle probe. Contact probe from which the main lobe of waves propagates at any angle
other than 0° or 90° to the normal to the tangent plane of the surface at the place
where the probe is positioned.
A-scan. Ultrasonic flaw detector display in which the pulse amplitude is represented as a
displacement along one axis (usually the Y axis) and the travel time of the ultrasonic
pulse is represented as a displacement along the other axis (usually the X axis).
ASME. Cross-drilled holes with diameter and position as required by the ASME pressure
vessel code. Note: ASME = American Society of Mechanical Engineers.
Automatic scanning. Systematic relative displacement of the ultrasonic beam and the
material under test by other than manual means.
AVG/DGS diagram. Series of curves which show the relationship between distance
along the beam to gain in dB compared with a particular back wall echo and the size of a
particular flat-bottom hole reflector.
NDT4-51015
Glossary 1 Copyright © TWI Ltd
Back wall echo (BWE). Pulse of ultrasonic energy reflected from the boundary of a
body directly opposite to the surface on which the probe(s) is/are positioned and
returned to that surface by the shortest possible path. Note: The term is generally
restricted to normal compressional waves and is sometimes referred to as bottom echo
(first).
Beam axis. Locus of points of maximum intensity in the far field in a beam of ultrasonic
waves, and its geometrical prolongation into the near field.
Beam index. Point on the surface of a body through which the beam axis passes (cf.
probe index).
Beam spread. Divergence of the main lobe of an ultrasonic beam in the far field. Note:
The beam spread is proportional to the ratio of the wavelength to the diameter of the
ultrasonic crystal.
Boundary echo (first). Pulse of ultrasonic energy reflected from any boundary of a
body to the surface on which the probe(s) is/are positioned, and returned to that surface
by the shortest possible path. Note: Term is generally restricted to shear or surface
waves.
B-scan display. Two-dimensional graphical plot showing the apparent size and position
of reflectors in the test piece on a cross-sectional plane which is normal to the test
surface and contains the beam axis of the probe during a single line scan.
Compressional wave. Form of wave motion in which the particle displacement at each
point in a material is parallel to the direction of propagation. Note: A synonymous term
is longitudinal wave; also sometimes referred to as dilatational wave or irrotational
wave.
Constructive interference. When two positive or negative sound waves from two
sources (Huygens's Principle) meet at a point at the same instant, the wave is reinforced
and assists the sound to propagate into the material under test. This is constructive
interference and takes place in the near field. See also destructive interference.
Contact scan. Scanning carried out by means of ultrasonic probe(s) in contact with the
body under examination.
NDT4-51015
Glossary 2 Copyright © TWI Ltd
Convergence point. Point of intersection of the axes of the transmitting and receiving
sound fields in a double crystal probe.
Corner effect. Reflection of ultrasonic energy back to a point coincident with or very
close to its point of origin, after impinging successfully on two or three orthogonal
surfaces.
Couplant. Liquid or pliable medium interposed between two solids to assist the passage
of ultrasonic waves between them. Note: In the majority of cases, the couplant is a
liquid interposed between the probe and the body under examination. Synonymous
terms are coupling film and coupling medium.
Coupling monitor. Probe operating in the receive mode and positioned such that it
detects ultrasonic energy originating from a second probe or multiple probes, thereby
monitoring that successful coupling is taking place between the second probe and the
body.
Cross-drilled holes. Cylindrical holes drilled parallel to the test surface and at right
angles to the vertical plane of the probe, the cylindrical surfaces of which form the
ultrasonic reflectors.
Cross talk. Acoustical or electrical signal leakage across an intended barrier. Note:
Sometimes referred to as cross noise.
Crystal array. Single housing containing an orderly assembly of crystals which may be
energised together in groups, with or without time delay, to give directional effects,
focused beams or variable angle beams.
Crystal backing. Material attached to the rear surface of a crystal to increase damping.
Crystal loading. Mechanical power per unit surface area delivered by a crystal to a
medium acoustically coupled to it.
Curie point. Temperature above which a ferromagnetic material loses its polarisation.
NDT4-51015
Glossary 3 Copyright © TWI Ltd
Cylindrical reflector. Reflecting surface in the form of a circular cylinder.
Damped train. Wave train in which the amplitudes of successive waves diminish.
Dead zone. Region in a material adjoining the surface of entry from which no direct
echoes from discontinuities can be detected due to the characteristics of the ultrasonic
equipment in association with the material under test and its surface condition.
Decibel (dB). Unit used to express the magnitude of change in the amplitude of an
ultrasonic signal, defined by the equation: dB = 20 log10(A1/A0), where A0 is a reference
amplitude.
Delayed time-base sweep. Cathode ray tube display in which the initial part of the
time scale is not shown.
Depth scan. Manipulation of an ultrasonic shear wave probe over the surface of a body
so as to cause an oblique beam to traverse a particular plane section of the body.
Destructive interference. When a positive sound wave from one point source
(Huygens's Principle) passes a negative sound wave from a second point source at the
same instant, the pressure sound wave is nullified at that point. This is destructive
interference, which results in areas of maximum and minimum pressure giving spurious
indications from any reflectors in this area. This area of interference is the near field.
See also constructive interference.
Diffuse reflection. Reflection of an ultrasonic wave from a rough surface so that the
reflected energy is detectable over a range of angles on either side of the theoretical
angle of specular reflection, ie reflection in a non-specular manner.
Directional sensitivity. Relationship between the angle made with the normal to the
surface of a reflector by a beam of ultrasonic waves and the amplitude to the resultant
echo.
Dispersive medium. Material in which the phase velocity of an ultrasonic wave varies
with frequency.
Distance amplitude curve. Curve constructed from the peak amplitude responses from
reflectors of equal area at different distances in the same material.
NDT4-51015
Glossary 4 Copyright © TWI Ltd
Double bounce technique. See triple traverse technique.
Double crystal probe. Probe comprising two separate crystals in a single housing, one
acting as a transmitter and the other as a receiver. Note: Synonymous terms are twin
crystal probe and combined double probe.
Double probe technique. Ultrasonic testing technique using one probe for transmission
and the other for reception.
Double skip technique. Ultrasonic testing technique where the distance between the
point where the waves enter the body and the region under examination is twice the skip
distance.
Double transceiver technique. Ultrasonic testing technique involving the use of two
probes, each used simultaneously as a transmitter and a receiver.
D-scan. Two-dimensional graphical projection onto a plane normal to the test surface
and to the projection of the beam direction on the test surface, showing the apparent
size and position of reflectors in the volume inspected by scanning an area of test
surface.
Echo. Distinct pulse of ultrasonic energy reflected from any surface or discontinuity.
Electronic noise. Unwanted random signals that vary rapidly with time, caused by
electronic pick-up and thermal noise in the amplifier of the flaw detector.
Expanded time-base sweep. Increased speed of time-base spot sweep which enables
echoes from a selected region within the thickness or length of a body to be displayed in
greater detail on the screen of the ultrasonic flaw detector.
Far field. Region in an ultrasonic beam where the intensity is inversely proportional to
the square of the distance. Note: Sometimes referred to as the Fraunhofer region.
First critical angle. Angle of incidence of a longitudinal wave in one medium such that
the refracted longitudinal wave is 90° in the second medium, ie along the surface, only
the transverse wave being transmitted into the second medium.
Flat-bottom hole (FBH). Cylindrical blind hole with a flat bottom, the flat bottom being
used as the ultrasonic reflector.
NDT4-51015
Glossary 5 Copyright © TWI Ltd
Flaw location scale. Specially graduated device that can be attached to a shear wave
probe which, in conjunction with the position of the flaw echo on the screen of the
cathode ray tube, gives a direct reading of the location of the discontinuity within the
body.
Gap scanning. Form of scanning in which the probe carrier follows the contour of the
material under examination but the probe, whilst not in direct contact with its surface, is
coupled to it through a layer or jet of liquid which is maintained between the surfaces of
the probe and the material.
Gate. Electronic means of monitoring a selected region of the cathode ray tube display
of an ultrasonic flaw detector.
Grass. Spatially random signals arising from the reflection of ultrasonic waves from
grain boundaries and/or microscopic reflectors in a material.
Half skip technique. Ultrasonic testing technique in which the inspection of a surface
region of a body is accomplished by using a shear wave beam entering from the opposite
surface at a point corresponding to the half skip-distance.
Hard face probe. Probe in which the contact surface is of a hard material, such as steel
or ceramic, to minimise wear.
Head wave. Shear wave generated by mode transformation when a compressed wave
travels at a grazing angle on a free solid surface. Note: In steel, the head wave is at
33°.
Holography (ultrasonic). Ultrasonic image from two transducers, the beams of which
are positioned to produce an interface pattern, usually on a liquid surface, which when
illuminated by laser light produces a visible indication of ultrasonic wave intensity
distribution.
Immersion testing. Ultrasonic testing technique in which the material under test and
the probe(s) are immersed in a tank of water or other liquid.
NDT4-51015
Glossary 6 Copyright © TWI Ltd
Indexing. Automatic measuring of probe position, usually electrically, to generate probe
position data that can be recorded.
Indirect scan. Use of surface(s) of a body to direct an ultrasonic beam into a particular
region of the body.
Interface signal. Displayed ultrasonic signal arising from the part reflection of an
incident pulse at an interface.
Interface trigger. Interface signal used as the initiating point from which other timing
sequences (eg gate position) are referenced.
Lamb wave. Term applied to those modes of vibration which propagate in a plate. Note
1: A synonymous term is plate wave. Note 2: In general, both compressional and shear
elasticity are involved, together with plate thickness and frequency; also, the
propagation shows dispersion.
Logarithmic decrement. For a damped train, the natural logarithm of the ratio of the
peak values of the amplitudes of two successive cycles.
Love wave. Acoustical wave which propagates along a stratum bounded on both sides
by layers of material which differ from the stratum in their elastic properties. Note: The
particle displacement of the wave is parallel to the wave front and to the stratum.
Maximum working range. Total distance over which a probe will transmit sufficient
energy to find the smallest reflector to be detected.
Multiple echo. Repeated reflection of an ultrasonic pulse between two or more surfaces
or discontinuities in a body.
NDT4-51015
Glossary 7 Copyright © TWI Ltd
N-distance. Distance from the probe to the N-point (see BS M36).
N-point. Position in an ultrasonic beam where the intensity of sound on the beam axis
reaches a final maximum before beginning a uniform reduction with distance (ie far
zone; see BS M36).
Normal probe. Probe from which waves propagate at 90° to its contact surface.
Opacity technique. Ultrasonic shear wave technique for the examination of thin plate
which makes use of the principle that if the plate thickness is less than a minimum
value, ultrasonic waves at a fixed angle and frequency are unable to propagate.
Pitch and catch technique. Ultrasonic testing technique involving the use of two
separate probes, one being used to transmit the ultrasonic energy into the body and the
other positioned so as to receive the reflected energy from a discontinuity. Note: In
variations of the technique, more than two probes may be used.
Plane wave. Wave in which points of the same phase lie on parallel plane surfaces.
Probe array. Array of probes which may comprise: (1) Probes in a mechanical holder
which scan together and are used sequentially, individually and/or in pairs. (2) A single
unit comprising probes used as in (1).
Probe face. Part of a probe through which ultrasonic waves are transmitted and
received.
Probe index. Point on a shear wave or surface wave probe through which the emergent
beam axis passes. Note: The index can vary slightly depending on the method of
measurement.
NDT4-51015
Glossary 8 Copyright © TWI Ltd
Probe shoe. Shaped piece of solid material interposed between the probe and the
material under examination for the purpose of improving acoustical contact.
Probe shoe delay. Time taken for the transmitted ultrasonic wave to traverse the probe
shoe and to be reflected back to the ultrasonic crystal.
Pulse amplitude. Pulse height of a signal, usually base to peak, when displayed in an
A-scan.
Pulse length. Time interval between the leading and trailing edges of a pulse, usually
measured at the half-amplitude value. Note: Synonymous terms are pulse duration and
pulse width.
Rayleigh waves. Particular type of surface wave which propagates on the surface of a
body with effective penetration of less than a wavelength.
Reference piece. An aid to interpretation in the form of a test piece of the same
nominal composition, significant dimensions and shape as a particular object under
examination. Note: Such pieces may or may not contain natural or artificial
imperfections.
NDT4-51015
Glossary 9 Copyright © TWI Ltd
Reflection factor. See reflection coefficient.
Refracting prism. A prism, usually of plastic material, which when placed in acoustical
contact between an ultrasonic crystal and a body will cause ultrasonic waves to be
refracted at a known angle into that body.
Refractive index. Ratio of the velocity of an incident wave in one material to the
velocity of a refracted wave in a second material in acoustical contact with the first
material.
Ringing time. Time during which the mechanical vibrations of a crystal continue after
the electrical pulse has stopped.
Scan pitch. Pitch or distance between lines of scan during passage of the probe(s) over
the scan area.
Scanning. Systematic relative displacement of the ultrasonic beam and the material
under test.
Scatter (ultrasonic). Energy reflected in a random way by small reflectors in the path
of a beam of ultrasonic waves (eg grain boundaries).
Screen marker. Small electronically generated pulses following one another at a preset
time interval, presented on a time-base sweep to provide a calibration less dependent on
the linearity of the time base.
Second critical angle. Angle of incidence of a longitudinal wave in one medium such
that the refracted transverse wave is at 90o, ie along the surface of the second medium.
NDT4-51015
Glossary 10 Copyright © TWI Ltd
Sequence number. In an automatic testing system, the order of connection of channels
and probes required to perform defined scans.
Shear wave. Form of wave motion in which the particle displacement at each point in a
material is at right angles to the direction of propagation. Note: Synonymous terms are
distortional wave, rotational wave and transverse wave.
Short pulse. Pulse which has few (usually less than 1.5) cycles in the time interval over
which its amplitude exceeds half of its maximum amplitude.
Side lobe. Peak or pronounced shoulder in an ultrasonic beam lying to one side of the
main beam.
Signal-to-noise ratio (SNR). Ratio of the amplitude of an ultrasonic echo arising from
a discontinuity in a material to the amplitude of the average background noise.
Single probe technique. Technique involving the use of a single crystal probe for both
transmitting and detecting ultrasonic waves.
Skip distance. For a beam of shear waves entering a body, the distance measured over
the surface of the body between the probe index and the point where the beam axis
impinges on the surface after a single reflection from the opposite surface.
Snell’s law. States that the angle of refraction is a function of the angle of the incident
beam and the change in relative velocity between the two materials.
Soft-faced probe. Probe in which the contact surface is a flexible membrane and a
space between the crystal and membrane is filled with a liquid couplant.
Soft-tipped probe. Probe in which a thick flexible medium forms the coupling between
its crystal and the surface of the material under test.
Sound attenuation. Reduction in the level of sound intensity due to distortion, scatter
and beam spread.
NDT4-51015
Glossary 11 Copyright © TWI Ltd
Specular reflection. A mirror-like reflection of an ultrasonic beam such that the angle
of incidence is equal to the angle of reflection.
Spherical reflector. Surface of spherical or near spherical form, separating two media
of different characteristic impedance.
Spherical wave. Wave in which points of the same phase lie on the surfaces of
concentric spheres.
Spurious echo. Term used for any indication not obviously associated with a
discontinuity or boundary. Note: A synonymous term is parasitic echo.
Stand-off. Block, usually of plastic material, which serves to separate the ultrasonic
crystal(s) from the surface of the test piece. Note: The use of such blocks is generally
confined to compression wave probes.
Surface noise. Unwanted signals at very short range, produced by ultrasonic waves
being reflected within the coupling film and from irregularities of the surface.
Surface wave probe. Probe for generating and/or detecting surface waves.
Swivel scan. Shear wave technique used to provide information about the form of a
previously located discontinuity, the probe(s) being positioned at a constant distance
from and directed at the discontinuity and rotated by an angle of up to 360o.
Tandem probe technique. Technique involving the use of two probes, one transmitting
the ultrasonic energy into the body and the other positioned to pick up any energy
reflected from a discontinuity. Note: The probes are usually scanned together at a fixed
separation and the technique is mostly used for the detection of vertically oriented,
through-wall, planar defects.
Test block. Piece of material capable of propagating ultrasonic waves and suitable for
assessing particular features of ultrasonic equipment performance.
Test surface. Surface of a piece of material through which ultrasonic waves pass.
Time base. Trace on the screen of a cathode ray tube which is generated in such a way
that distance along it is proportional to time.
NDT4-51015
Glossary 12 Copyright © TWI Ltd
Time base range. Maximum ultrasonic path length that can be displayed on a particular
time base.
Time corrected gain. Facility of flaw detectors to represent flaws of equal reflective
size with the same screen amplitude, irrespective of their depth in the material.
Toe-in-semi-angle. Half the angle between the normals to the crystal faces in a twin
crystal probe.
Total internal reflection. Reflection which occurs when the angle of incidence is
greater than the critical angle and the reflection coefficient is unity.
Transmission point. Point on the time base which corresponds to the instant at which
ultrasonic energy enters the material under examination.
Ultrasonic frequency. Any frequency of vibration greater than the range of audibility of
the human ear, generally taken as greater than 20kHz.
Ultrasonic mode changer. Device which causes vibrations of a particular mode (eg
compressional) in one body to produce vibrations of another mode (eg shear) in another
body.
Wavelength (λ). Perpendicular distance between two wave fronts with a phase
difference of one complete period.
NDT4-51015
Glossary 13 Copyright © TWI Ltd
Ultrasonic wave. Disturbance which travels through a material at ultrasonic frequency
by virtue of the elastic properties of that material.
Wedge (ultrasonic). Device placed between the probe and the test surface for the
purpose of causing ultrasonic waves to pass between the two at a particular angle.
Wetting Agent. Substance added to a coupling liquid to decrease its surface tension.
NDT4-51015
Glossary 14 Copyright © TWI Ltd