Colado CAD CoCr4 ®

Cobalt-chromium disc for CAD/CAM technology


Colado® CAD CoCr4 is a precious metal-free cobalt-chromium alloy especially designed for
­ceramic and composite veneers.

Co Cr Mo Fe Mn Nb Si
63.0 ± 2.0 29.0 ± 2.0 6.0 ± 1.0 < 1.0 < 1.0 < 1.0 < 1.0

Advantages
– Wide range of indications owing to the type 4 classification
– Excellent milling properties thanks to a fine grain structure
– Excellent veneering properties with Ivoclar Vivadent veneering materials
such as IPS Style®
– Tested biocompatibility (high corrosion resistance)
– Coordinated compatibility with the Ivoclar Vivadent product system

Indications
Colado CAD CoCr4 is suitable for thin frameworks, which are exposed to very high
loads e.g. clasps, veneered crowns, large spanned bridges or bridges with thin
cross-sections, bars, fixtures, implant-supported superstructures according to
ISO 22674.

Technical data
Material properties according to ISO 22674:2016

Specification
Alloy type Type 4
Density [g/cm ]
3
8.4 ± 0.2
CTE (25 – 500 °C) [10-6/K] 14.4 ± 0.5
0.2 % proof stress [MPa] ≥ 360
Elongation [%] ≥2
Modulus of elasticity [GPa] ≥ 190
Solidus/Liquidus temperature [°C] ≥ 1300 / ≥ 1360
Resistance to corrosion Yes
Biocompatibility Testing

Certificate

Manufacturer Ivoclar Vivadent AG, 9494 Schaan/Principality of Liechtenstein

Cytotoxicity Cytotoxicity tests were conducted on cells of the mouse cell line L929
with extracts of Colado CAD CoCr4. The extracts did not reveal any
cytotoxic effect.

Skin irritation In-vitro test were conducted on human skin models with extracts of
Colado CAD CoCr4. The extract did not cause any skin irritations.

Genotoxicity The AMES reversion mutation test was conducted on bacterial cells
with extracts of Colado CAD CoCr4. The extracts did not show any
mutagenic effects.

Corrosion resistance Static immersion tests were conducted with Colado CAD CoCr4. The
cumulative substantial loss after 7 days was significantly lower than
the limit value of 200 µg/cm2 according to ISO 22674.

Schaan, August 2017

Dr Thomas Völkel
Head of Department Scientific Services

Manufacturer and EU representative: Distribution Germany: Notified Body:


Ivoclar Vivadent AG Ivoclar Vivadent GmbH TÜV Süd Product Service GmbH
Bendererstrasse 2 Dr. Adolf-Schneider-Strasse 2 Riedlerstrasse 65
FL-9494 Schaan D-73479 Ellwangen, Jagst D-80339 Munich/Germany
EN/2017-11-16

Principality of Liechtenstein Tel. +49 (0) 79 61 / 8 89-0


Tel. +423 235 35 35 Fax +49 (0) 79 61 / 63 26
Fax +423 235 33 60 [email protected]
www.ivoclarvivadent.com www.ivoclarvivadent.de

You might also like