F Please read the CIRRUS photo user manual carefully before using this product.
Move instrument towards patient. The focusing aid and working distance dots (two
white dots) will appear together with the fundus image.
false correct
Bring the working distance dots roughly into focus.
If AutoFocus fails, use knob on side of instrument (bars should be aligned
horizontally).
Bring the working distance dots into focus by turning and moving the joystick and
C
align the working distance dots within the horizontal lines.
CM
Acquire image(s)
MY Instruct patient to blink once and to look directly at the fixation target .
CY
Click joystick button to capture image(s).
Advanced tip:
CMY
To modify OCT scan pattern location, press arrow () on keypad before capturing
K images. Use controls on OCT fundus image to change the scan location and
rotation/size (HD Raster only).
Grab and drag scan Grab edge of scan pattern to Press <!> on keypad or the
pattern to preferred scan resize the raster scan length Optimize button in display.
location. and spacing. For 1 Line Raster, Use the scroll wheel of your
drag spacing to zero. mouse to center the scan.