COWELL®

IMPLANT
SYSTEM

COWELL®
Implant System
More than 20 years of clinical outcomes
® 

INTRODUCTION

002 Process Flow Chart


004 COWELL® Warranty
005 Package System
011 SLA-SH® Surface Treatment
015 COWELL® CLASS 1000

INNO SLA-SH® IMPLANT SYSTEM

018 INNO-Fixture Design


020 Abutment Prosthetic Protocol
024 INNO Submerged Implant
027 INNO Submerged Short Implant
054 INNO Submerged Narrow Implant
®    070 INNO Internal Implant

rs of clinical outcomes 090 INNO External Implant

More than 20 yea 104 Surgical Kit

MINI PLUS® IMPLANT SYSTEM

121 Mini Plus® Implant


126 Surgical Kit
Process Flow Chart
CNC Machining Cleansing

Precise machining process using state of the art computer numerical control system Cleansing process by ultrasonic wave using the 3rd stilled distilled, vacuum dry and heating dry sterilization enable
fused to the COWELL® Class 1000, operated by world-class technical unit. ultimately sterilized products and no residues are left on products.

Surface Treatment Packing / Sterilization

The SLA-SH® Surface treatment with biologically active materials to achieve the ideal Osseointegration. Sanitarily packed products at Class 10,000 clean room are sterilized by gamma-ray using radiation isotope.

Inspection Shipping Warehouse

Absolutely accurate test and quality control system with cutting edge equipment such as optical profiling measurer, Finished products are sorted and stored in warehouse for immediate delivery.
stereoscopic microscope, micrometer scope and other specialized devices for dental implant manufacturing.

002 / 003 COWELL® Implant System


COWELL® Warranty Package System
* For more details, visit our website at [Link]
1. Color classification (Coding) by fixture type and external label marking
A. Color classification by fixture type
Submerged Submerged Internal External Submerged Mini Cement Mini Ball
Fixture type (Sub.) Short (Sub.) (Int.) (Ext.) Narrow (Sub-N.) (1P-C.) (1P-B.)

1. Guarantee beneficiary and scope

Products Period Conditions Remarks

The period Package


Implant Lifetime Replacement with equivalent Implant shall begin
from the sale date

2. Scope of Warranty SUB. INT. EXT. SUB-N.


HEXAGON OCTAGON HEXAGON HEXAGON
1) Quality benefits Connection SYSTEM SYSTEM SYSTEM SYSTEM
> In case the product material or the manufacture process is flawed. Blue Orange Green Emerald Pink
2) Surgical benefits
> In case implants fail to be grafted to the bone.
B. External label marking and color coding by fixture diameter & fixture type
>G Color coding by diameter on the external label.
3. Claim Procedure >G Reuse is prohibited after opening as the product is sterilized.
1) In case certain faults occur after transplanting implants (procedure), the staff in charge shall be contacted
>G After the ampule is opened, care should be taken from dropping which may cause by incomplete fastening.
within 30 days thereafter.
>G Care should be taken from infection after the product is opened and store at room temperature and in a dry place.
2) When such contact is made, the Customer Complaint Report shall be written out, and shall be submitted
together with the concerned product. >G Discard expired products.

20A060040A0004
4. Exclusions from Warranty Service ST4010SM
1) In case implants are transplanted onto patients with diabetics and alcohol addiction.
Ø4X10mm
Sterilized
INNO Sub. Fixture Caution Using Irradiation
2) In case implants are transplanted onto patients for whom surgical procedures are difficult to perform due to the
history of systematic disease.
Diameter Ø2.5 Ø3.0 Ø3.1 Ø3.3 Ø3.5 Ø4.0 Ø4.5 Ø5.0 Ø5.5 Ø6.0
3) In case implants are transplanted onto patients who depend on habitual medications.
INNO Fixture Fixture
4) In case the procedure is not conducted according to the protocol of the COWELLMEDI. (No-Mount) Type(abbr.) Bisque Green Burgundy Orange Yellow Emerald Red Pink Violet Gray

5) In case the procedure is not performed in compliance with biological indication : PRODUCT NAME  COWELL® INNO Implant System
CATALOG No.  ST4010SM Submerged (Sub.) - - - - -
(E.g. distance between buccal wall and implant should be at least 2mm). SIZE  Ø4.0X10mm([Link])
6) In case the procedure is conducted using contaminated surgical devices. LOT NO.  20A060040A Submerged - - - - -
7) In case implants are transplanted onto patients who sustain or are infected with cell issue contamination. DATE OF  2020-01-06 Short (Sub.)
MANUFACTURE
USE BY  2025-01-05
8) In case other materials from other companies are mix used with Implants, prosthetic parts and instruments
PACKING UNIT  1EA Internal (Int.) - - - - -
of the COWELLMEDI. STORAGE CONDITION  Store at room temperature and in a dry place.
9) In case the result of investigation by COWELL R&D Institute, Div. of QA and QC shows the issue is not related to the Do Not
Reuse Caution
30˚C
1˚C External (Ext.) - - - - -
products manufactured and provided by the COWELLMEDI. Sterilized
Using
Consult
Instructions
Irradiation for Use
10) In case the information hereby requested, especially, product Lot no., Serial no. or X-ray photos, is missing. Submerged
Manufacturer Cowellmedi Co.,Ltd. - - - - - - - -
11) In case that the concerned products are not returned. Narrow (Sub-N.)
48, Hakgam-daero 221beon-gil, Sasang-gu, Busan,
46986, Republic of Korea TEL.: +82-51-312-2027~8
12) In case the product is damaged due to negligence of handling. Website: [Link]
D/T : 218 Trianon LN Villanova PA 19085-1442 USA Mini Cement (1P-C.) - - - - - - - -
13) In case the product is opened and fails to remain sterilized. EC-Representative EC REP : Certification Experts B.V.
Amerlandseweg 7, 3621 ZC Breukelen, The Netherlands
14) In case that the expiry date of the concerned product (not opened products only) is not longer than 1/4.
(01) 08800016106725 Mini Ball (1P-B.) - - - - - - - -
(11) 200106
(10) 20A060040A
(21) 0004

Rx Only
MEDICAL DEVICE CWM-L-004 (Ver.3)

Ex.) INNO SLA-SH® Sub. Fixture (No-Mount)


Dimension : Ø4.0X10mm

004 / 005 COWELL® Implant System


2. Fixture user guide (Embedded in the packaging) 3. Fixture packaging opening and the sequence of the product extraction

COWELL® IMPLANT SYSTEM Instructions for Use Upper Upper

1. Device Description status, and adequacy of bone. Lateral cephalometric radiographs, CT scans and
The COWELLMEDI implant system includes a variety of precision-machined tomograms may also be beneficial. Adequate radiographs, direct palpation and $PWFS4DSFX $PWFS4DSFX
fixtures manufactured from titanium. These implants are surgically inserted into a visual inspection of the fixture site are necessary prior to treatment, planning and
Middle Middle
mandible (the lower jawbone) or a maxillary bone (the upper jawbone) and serve as use of the COWELLMEDI fixtures.
a replacement for patient's tooth root providing a stable foundation for restoration.
7. Adverse Effects
2. Intended for use Some of the complications (loss of fixture anchorage, prosthesis etc.) are possible
To support dental prosthesis as a dental device, which is implanted into alveolar occurrences after surgery. Lack of quantity or poor quality of remaining bone, .PVOU'JYUVSF
bone to recover masticatory function and give better esthetics in patients with infections, poor patient oral hygiene or cooperation, patient discomfort, fixture 'JYUVSF

partially or full edentulous jaws. mobility, local soft tissue degeneration, and unfavorable fixture placement or
alignment are some potential causes for loss of anchorage.
No-Mount Pre-Mount
3. Directions for use
1) Surgery - The first stage 8. Surgical complications
Lower Lower
a. According to the patient's condition, appropriate dental cleaning operations may The implant procedure has risk, including localized swelling, dehiscence,
be performed and preventive antibiotics may be administered prior to implant tenderness of short duration, edema, hematoma or bleeding. Numbness of the
operation. lower lip and chin region following lower jaw surgery, and of the tissue beside the
b. Clean and disinfect the operative site, administer local anesthesia in the area nose following upper jaw surgery, is a possible side-effect of the surgery. Though it
and expose the alveolar bone by making appropriate incisions and reflecting the would most probably be of a temporary nature, in very rare cases, the numbness
gingival tissues along the alveolar crest in the area from where teeth were has been permanent. Gingival-mucosal (gum tissue) ulceration, tissue reaction or
extracted. infection may occur, but generally responds to local care.
c. Drill into the gum in order to implant a fixture into the planned place with
various dental operation tools. The speed of revolution of the drill should be 9. How to manage after implanting the Product
adjusted by the condition of the bone and the kinds of operation tools. Saline a. The upper jaw requires a healing period of 6~8 months depending on the bone
solution should be poured onto the area so that necrosis doesn't occur by heating quality and the lower jaw requires a healing period of 3~5 months depending on Taking out the ampule
of the bone (The speed for all drilling should be less than 1,200 rpm). the bone quality. If pressures such as mastication would be applied to the fixture
d. Remove the external sterile package cover sheet : open the cap of the ampule: during the healing period, early fixation may not be obtained or osseointegration
affix the Fixture Driver (in case of No-mount Fixture) or the Mount Driver (in case may not occur for the fixture within the healing period.
of Pre-mount Fixture) to the Hand-piece and connect it to the fixture : move the b. Once the operator clinically determines that sufficient osseointegration has been
assembled piece to the osteotomy site for the implant using care to prevent the achieved, he/she should begin the stages to produce the dental prosthesis.
assembled piece from being separated or contaminated with foreign materials. c. The Lot number identification tag should be attached to a patient's chart and
e. A fixture is implanted into the bone as planned depth by turning (25~30 rpm) a X-ray film in order to pursue the product when needs be.
hand-piece clockwise with 15~50 [Link] torque. In event that it is hard to insert, d. The operator should determine the osseointegration status of the implant
extend width of bone by Tap Drill or Countersink (less than 1,200 rpm) in order to through X-ray and clinical methods such as percussion and/or reverse torquing.
facilitate better implantation.
f. After finishing implantation, the treated part should be sutured by using a hex 10. Storage / Sterilization and Handling

1 2
driver to connect to the Cover Screw with torque 5 [Link] to prevent the intrusion a. Store at room temperature and in a dry place.
of a foreign substance in the fixture. b. The fixture, fixture mount and cover screw have been cleaned and sterilized
To open, press the upper dotted area Remove the moisture-resistant paper on the
2) Surgery - The second stage by radiation (gamma irradiation) and are ready to use. and take out the sterilized blister. back of the blister, and drop the ampule
a. Incise gingival of upper part of fixture subsequent to bone fusion and remove c. Product packages should be opened just before their use during operation lightly on the palm of a practitioner or
Cover Screw, tighten up Healing Abutment and start gingival curing for prosthesis. and any product beyond the expiration date should not be used.
b. In general, surgery is done by a method that makes prosthesis. d. Only appropriate sterilized surgical tools made specifically for dental implants surgical clothes.
should be used during the operation.
4. Contraindication
The operation should be reconsidered when the patient has any of the following 11. Expiration date
Fixture separation
conditions. Expiration date of the product is 5 years from manufacturing.
a. Patient with oral infection or inflammation.
b. In the case of low quality bone which will result in an unstable implant. 12. Cleaning & Sterilization
c. Patients who have a drinking problem or mental disease or substance or Cleaning of surgical instruments supplied non-sterile should be performed
medicine abuse. according to current dental standard practices. Select a suitable method of
d. Internal disease such as hematodyscrasia or diabetes and undernourishment. cleaning that removes all visible contamination from the product in sterilized and
e. Any patient who is not suitable for operation. distilled water. After cleaning, package the product appropriately and then
sterilized by autoclave at the minimum condition of 250°F (121°C/15 mins).
5. Warnings
Implant surgery and restoration involves complex dental procedures. For safe and 13. Caution
effective use of the COWELLMEDI fixtures, it is strongly suggested that specialized a. As this product is sterilized by Gamma radiation, should not be used

1 2
training be undertaken since the surgical techniques required to place dental under any circumstances if is open. Hold the ampule with both hands, twist Fasten in two ways.
implants are highly specialized and complex procedures. Improper patient b. Every product is disposable. It should not be reused.
selection and technique can contribute to fixture failure and/or loss of supporting it 45°, and separate the middle part. 1) No Mount -> Fasten the Fixture Driver. Y
bone. the COWELLMEDI fixtures are intended for use only in the indicated Care should be taken so that the fixture 2) Pre Mount -> Fasten the Mount Driver.
applications. Dental fixtures must not be altered in any way. The use of electro-sur-
COWELLMEDI Co., Ltd. will not fall.
gical instruments or lasers around metallic fixtures and their abutments is not
recommended due to the risk of electric shock and/or burns. Fixture mobility, bone 48, Hakgam-daero 221beon-gil, Sasang-gu, Busan, 617-801, Republic of Korea
loss, or chronic infection may indicate fixture failure. The treatment should be done Tel. +82-51-312-2027~8 Fax. +82-51-316-2628
in an aseptic condition by an operator who wears an aseptic costume. If the fixture [Link] Cover Screw separation
becomes contaminated by the patient's body fluids in any way, the fixture cannot be D/T
used in any other patient.
COWELLMEDI USA INC. 218 Trianon LN Villanova PA 19085-1442 USA
Tel. +1-623-939-1344 Fax. +1-623-939-1472
6. Precautions
The surgical techniques required to place endosseous dental fixtures require EC-REPRESENTATIVE
specialized and complex procedures. Formal training for placement of fixtures is Certification Experts B.V. Amerlandseweg 7, 3621 ZC Breukelen, The Netherlands
recommended.
Important : Determine local anatomy and suitability of the available bone for fixture
placement. Thorough screening of prospective fixture candidates must be
performed. Visual inspection as well as panoramic and periapical radiographs are
essential to determine anatomical landmarks, occlusal conditions, periodontal
2020. 01. 30 / CWM-I-007 (Ver.3)

1 Separate the upper part of the ampule. 2 Fasten the Hex Driver to the Cover Screw
completely. Care must be taken to prevent
patient from swallowing the Cover Screw at
the time of placing.

006 / 007 COWELL® Implant System


4. Abutment packaging and external label marking

Hybrid S Ti-Base
PRODUCT NAME  Dental Implnat
CATALOG No.  2SLH524
SIZE  Ø5X3.75mm C:2([Link])
LOT NO.  20A090004A
DATE OF
MANUFACTURE
 2020-01-09
USE BY  Not Applicable (N/A)
PACKING UNIT  1SET
STORAGE CONDITION  Store at room temperature and in a dry place.

Consult 30˚C
Do Not Caution Instructions
Reuse for Use 1˚C

Manufacturer Cowellmedi Co.,Ltd.


48, Hakgam-daero 221beon-gil, Sasang-gu, Busan,
46986, Republic of Korea TEL.: +82-51-312-2027~8
Website: [Link]
D/T : 218 Trianon LN Villanova PA 19085-1442 USA
EC-Representative EC REP : Certification Experts B.V.
Amerlandseweg 7, 3621 ZC Breukelen, The Netherlands

(01) 08800016104301
(11) 200109
(10) 20A090004A
(21) 0001

The Pioneers in Dental Implant & [Link]-2 Rx Only


MEDICAL DEVICE CWM-L-005(Ver.5)

5. Surgical Kit packaging and external label marking

INNO Sub. Full Kit


PRODUCT NAME  Reusable Surgical Instrument
CATALOG No.  KCA010F
SIZE  INNO (Full / Sub.)
LOT NO.  20A090004A
DATE OF
MANUFACTURE
 2020-01-09
USE BY  Not Applicable (N/A)
PACKING UNIT  1SET
STORAGE CONDITION  Store at room temperature and in a dry place.

Consult 30˚C
Caution Instructions
for Use 1˚C

Manufacturer Cowellmedi Co.,Ltd.


48, Hakgam-daero 221beon-gil, Sasang-gu, Busan,
46986, Republic of Korea TEL.: +82-51-312-2027~8
Website: [Link]
D/T : 218 Trianon LN Villanova PA 19085-1442 USA
EC-Representative EC REP : Certification Experts B.V.
Amerlandseweg 7, 3621 ZC Breukelen, The Netherlands
When INNOVATION meets Dental Implant.
(01) 08800016103113
(11) 20109
(10) 20A090004A
(21) 0001

Rx Only
MEDICAL DEVICE CWM-L-006 (Rev.4)

008 / 009 COWELL® Implant System


SLA-SH® Surface Treatment
by Nano/CaP soaking technology

SLA-SH®:
Sandblasted, Large-grit, Acid-etched and Super-Hydrophilised
> Hydrophilicity by activation with neutralization solution & Nano/CaP soaking technology.

Experience the > Sandblasted with biocompatible grits unlike majority of other implants in market are done with Alumina.
> Macro-pore & micro-pore of Ti-oxide layer mimicking the etched enamel rod of tooth.

difference of SLA-SH ®
> Even distribution of roughness through the whole portion of implant surface.
> No destruction or alternation of the surface are caused even with torque force of 120 [Link].
> Acceleration of osseointegration and maximization of BIC.

made by "Nano/CaP soaking technology" > The SLA-SH® is applied for all types of the COWELL® Implant Systems.

1. Evaluation using SEM (Scanning Electron Microscope) Images


A. SLA-SH® Surface magnified X300, 1,000 and 3,000

X300 X1,000 X3,000

B. Comparison to other SLA treated implants currently sold in the market

" #

$ SLA-SH®

Superhydrophilicity, Uniform micro-surface geometry,


> Surface treatment patterns were observed on electron microscope photographs of 5,000 magnifications for top parts of the
Maximized BIC and Acceleration of osseointegration implants.

Aspiring for 100% perfection with SLA-SH® > Sand-blasted surface conditions were observed in the product A, B and C due to insufficient acid etching patterns in deep parts
as the SLA-SH® is sandblasted with biocompatible grits with even particle size unlike others are done with alumina.
by Nano/CaP soaking technology > The entire surface of the SLA-SH® treated implant showed uniform acid etching patterns. This implies that the
acid etching of the SLA-SH® surface is perfect.

010 / 011 COWELL® Implant System


SLA-SH®

2. Evaluation using SSEM (Stereo Scanning Electron Microscope) 3D images 3. The surface activity increased due to the great surface wetness
A. SLA-SH® Surface A. Contact angle measurement evaluation result for the saline solution

Surface Stats 133.06°


Ra : 1.80 ༁ 36.33°
Rq : 2.27 ༁
Rt : 18.49 ༁

Measurement Info
Magnification : 19.93
Measurement Mode : VSI
Sampling : 496.74 nm
After hydrophilicity After superhydrophilicity
Array Size : 640 X 480 After SLA treatment (133.06°)
activation treatment (36.33°) activation treatment (9°)

After neutralization process and Nano/CaP soaking treatment, the sample became extremely hydrophilic
Surface Stats and the surface energy increased, which facilitated expedition of osteoblast activation to be fused to the bone faster.
Ra : 1.95 ༁
Rq : 2.52 ༁
Rt : 19.83 ༁

Measurement Info
Capillarity in the actual clinical setting, which accelerated the penetration of blood.
Magnification : 19.93 ୔ Quoted from the website of Cowellmedi Clinical Research Group ([Link])
Measurement Mode : VSI
Sampling : 496.74 nm
Array Size : 640 X 480 Blood Friendly

B. Relation between surface wetness and roughness

B. Comparison to other SLA treated implants currently sold in the market

A B C SLA-SH®
After SLA treatment (Ra: 1.78༁) After super-hydrophilicity activation treatment (Ra: 1.90༁)

> There was almost no difference of surface roughness and micro-geometry, and the difference of surface wetness took place
in the same physicochemical properties as surface energy increased by hydrophilicity activation treatment.

6QQFSRa : 2.47༁G 6QQFSRa : 1.07༁G 6QQFSRa : 2.65༁G 6QQFSRa : 1.80༁G


-PXFSRa : 3.11༁ -PXFSRa : 1.13༁ -PXFSRa : 2.09༁ -PXFSRa : 1.95༁
%FWJBUJPO༁ %FWJBUJPO༁ %FWJBUJPO༁ %FWJBUJPO༁ C. Physicochemical alternation of surface by hydrophilicity activation treatment

Name Start BE Peak BE End BE Name Start BE Peak BE End BE


> Uniform distribution of Macro-pore and micro-pore.
C1s 290 284.6 280.5 C1s 290.46 284.6 284.6
> Roughness of the SLA-SH® showed 1.90um while the others were 1.08~3.11um. O1s 535.3 530.42 525.6 O1s 538.8 533.73 529.3
Ti2p 468.1 458.78 450.4 Ti2p 468.2 456.76 453.4

After SLA treatment After hydrophilicity activation treatment

> Surface wetness was improved by increased surface energy of C1s, O1s and Ti2p after hydrophilicity activation treatment.
> To maintain and even to enhance surface wetness, superhydrophilicy activation treatment was carried out and contamination
by carbon in the atmosphere is prevented during packing and sterilization.

012 / 013 COWELL® Implant System


4. Its safety has been proven through perfect cleaning with an automated system COWELL® CLASS 1000
A. Comparison of surface element tests through X-ray diffraction A SUBTLE DIFFERENCE MAKES THE DENTAL IMPLANT OR NOT
" # $ SLA-SH®
1. Fixture manufacturing tolerance evaluation

Evaluation Item Manufacturing Tolerance

A1
a. The specimen was fixed in Jig.
Method b. Each dimensional difference of 3 inner hexagonal connection sides
> Cutting-edge automated system that produces the 3rd distilled water. (Hex-1, Hex-2, Hex-3) of 5 specimens was measured.
C4
C2
C1 Used Equipment Measuring Microscope and Jig

A2

B3
C3
Each dimensional difference of 3 inner hexagonal sides should be
Criteria
B. Comparison of surface element tests (X-ray Photo-electron Spectroscopy, XPS) no more than ±0.001mm (1.000༁) from 2.500mm.
Unit : % INNO SLA-SH® Submerged Fixture (5 Pieces of ST4510S)
Specimen
Sample C1s O1s Ti2p Si2p N1s #1 #2 #3 #4 #5

B2

B1
" 34.12 45.05 15.11 5.24 0.47 Hex-1 2.499 2.500 2.500 2.500 2.500
# 31.84 46.49 15.22 4.87 1.57 Hex-2 2.500 2.500 2.501 2.500 2.500

A4

A3
$ 32.19 47.58 17.58 2.65 N.D Hex-3 2.500 2.500 2.500 2.501 2.499
SLA-SH® 27.19 50.81 17.61 N.D N.D Average 2.500 2.500 2.500 2.500 2.500
Total Average 2.500
Result (Pass/Fail) Pass
> Quantitative analysis of each surface element found 30% carbon, 47% oxygen, 16% titanium and 4% silicon in all products.
Manufacturing No more than
> For the SLA-SH®, they only consisted of carbons(C1s), oxygen(O1s) and titanium(Ti2p). Tolerance ±0.001mm (1.000༁)

B4
> Sodium hydroxide, the main element of the alkali washing solution, combined with silicon(Si) to form water-soluble
Na2SiO2(OH)2·4H2O(water glass), which removed the other elements.
2. Prosthetic component manufacturing tolerance evaluation

Evaluation Item Manufacturing Tolerance


C. Comparison of elution tests using combustion ion chromatography
a. The specimen was fixed in Micro-Measuring Instrument.

A1
Unit : ppm Method b. Each dimensional difference of 3 outer hexagonal connection
Sample F- Cl- NO2- SO42- Br- NO3- PO43- sides (Hex-1, Hex-2, Hex-3) of 5 specimens was measured.

" N.D 0.024 0.027 0.002 N.D 0.031 N.D Used Equipment Micro-Measuring Instrument
# N.D 0.027 0.019 0.002 N.D 0.030 N.D C
Each dimensional difference of 3 outer hexagonal connection sides should be
C3
Criteria C2
$ C2
no more than ±0.001mm (1.000༁) from 2.490mm.

A2
N.D 0.071 0.020 N.D N.D 0.023 N.D C1 C1

SLA-SH® N.D N.D N.D N.D N.D 0.032 N.D

B3
C3

A2
INNO SLA-SH® Sub. Cemented Abutment (5 Pieces of 2SCH4515)
Specimen

A3
> Similar ions were detected in all the products, but they are not harmful to human because their elements and quantities #1 #2 #3 #4 #5

A4
Hex-1 2.489 2.490 2.490 2.490 2.490

B2
do not affect the human body and those have been proven in many studies.

B1
Hex-2 2.490 2.490 2.490 2.490 2.490
> For the SLA-SH®, no other elements except for NO3- were detected. Alkali washing completely removed the SO42- and

B1
Hex-3 2.490 2.490 2.490 2.490 2.491

B2 B4
Cl- ions of sulfuric acid and hydrochloric acid, which are used for heated acid etching because they form water-soluble salts

A3
Average 2.490 2.490 2.490 2.490 2.490
of Na2SO4 and NaCl.
Total Average 2.490
> No elements that interfere with osteo anagenesis were found from both the surface and elution elements, which showing Pass
Result (Pass/Fail)
that the cleansing process was perfectly carried out.

C1
Manufacturing No more than
C2
Tolerance ±0.001mm (1.000༁)

B4
014 / 015 COWELL® Implant System
COWELL® IMPLANT SYSTEM
Volume-up™ Healing Abutment
Help your daily practice superior Devised for preventing food penetration
and forming aesthetical cervical area by restoring
contracted buccal alveolar bone & gingiva to the
original shape and width.

Meta G UCLA Abutment


Modification to angulated Easy Temporary Abutment
abutment, customized Temporary restoration for anterior esthetic zone.
abutment and telescopic For simpler and speedier chair-side process.
abutment.

INNO Submerged Narrow Fixture


Designed for anterior esthetic zone with narrow alveolar ridge. Angulated Abutment
Double tapered thread acquires higher primary stability Simple solution for anterior esthetic zone.
through a wedge action.

Milling Abutment
Block abutment to customize contouring.

INNO Submerged Short Fixture


Designed for severe bone resorption. Wide and
deep upper thread prevents the compressive
necrosis of the cortical bone.
Multi S&A and Lock Abutment
Designed for both edentulous and
partially edentulous arches.
A broad range of prosthetic
Miniplus® options allows to meet individual
For mandible anterior spaces requirements of your patients.
and edentulous arch.
Semi-permanent or
temporary solution for
anterior spaces with extremely
narrow ridge.

INNO External Fixture


Stable engraftment of periosteum
in the boundary surface of bone and implant. Ball Abutment
Sonator 80’s S&A Abutment Designed for use with removable
Allows you to treat your patients with overdentures or partial
INNO Submerged Fixture minimum standard of care of an implant supported dentures retained in whole or
INNO Internal Fixture
Designed for all clinical cases including Cemented Abutment overdenture at affordable cost with great satisfaction. in part by endosseous implants
immediate implant placement, immediate loading, One stage restoration with state
Cutting surface is in the maxilla and mandible.
implant depth adjustment, maxillary sinus and etc. of the art design. 4 spiral round cutting
designed for anti-rotation
Simply doing more for your implant treatment. edges maximize the efficiency of Beauty-up™ Abutment
of prosthesis.
self tapping with a sharp edge and
Specially designed to solve esthetical
accommodates bone chips as ideal
and functional challenges
cutting edge pocket space.
when SCRP with angulated screw
channel is required in the anterior portion.
016 / 017 COWELL® Implant System
INNO-Fixture Design An advantageous design for all clinical cases including immediate
implant placement & immediate loading, implant depth adjustment,
maxillary sinus and etc.

Fixture type Submerged (Sub.) Submerged Short (Sub.) Internal (Int.) External (Ext.) Submerged Narrow (Sub-N.)

11°
Fixture Design

SEM Image, x47

11° Tapered and 2.5 Hex Connection


> Allows the ideal cold welding between the
SUB. INT. EXT. SUB-N.
fixture and the abutment. Connection HEXAGON OCTAGON HEXAGON HEXAGON
SYSTEM SYSTEM SYSTEM SYSTEM
> Prevents micro-sinking of the abutment.
> Excellent compatibility.

Platform Neck Simpler, Speedier and Safer Surgical Kit


> Stable engraftment of periosteum in the boundary
Wide and deep upper thread Sub. / Sub-N. / Int. / Ext. / provide different types of exclusive kits, respectively.
surface of bone and implant.
> Prevents the compressive necrosis of the > Prevents possible infections around the implant.
cortical bone.
> Minimizes the need for Countersink Drills. Sub. Full KCA010F Int. Full KCA010FI Ext. Full KCA010FE
> Reinforces mechanical strength by reinforcing
thickness. Open Thread
> Possible to place deeper even without additional
drilling.
Double Tapered Thread
> Secures early fixation even for an alveolar socket
or parts with weak bone quality.
> Allows placement to be completed with only
2-4 rotation with half the length of the fixture
inserted in the drill hole. 4 spiral round cutting edges
> Acquires higher primary stability through a wedge > Maximize the efficiency of self tapping with
action even with an additional half turn. a sharp edge.
> Accommodate bone chips as ideal cutting edge Sub. Smart KSA002 Sub. Short KSI001 Sub-N. Narrow KNA001
pocket space.
arrow K I T
> Allow soft placement but higher initial stability 12
Ratchet X-Long Machine X-Long

10

(refer to the test table below). 8


Drill
Stopper
Ø 3.3
Fixture
Fixture Driver

Ø 3.1
Fixture

Ø 2.6
Parallel Pin
Countersink
Ratchet Long
Ø 2.2

Spare
Point 1.2 Hex Driver
Depth Gauge

Apex Thread with a sharp cutting edge


Torque Wrench

> Prevents schneiderian membrane from being ripped.


> Allows soft placement but higher initial stability
(refer to the test table below).
All in One Drill : Minimal drilling frequency with Initial and Final Drill
※ Placement torque force (Unit: for 4pcs of Ø4.5X10mm implants Chair time for implantation is shortened because the fixture can be implanted with just three drillings for general bone quality
Shortens the placement time with 5mm from 4 different manufacturers in 5.0 & 5.5mm depth hole of type (when implanting Ø3.5-Ø4.5 fixtures).
2 bone quality test block.
or more of already entered depth as well
as double thread. Classification INNO A B C
Depth 5.0mm 26.2 [Link] 29.2 [Link] 26.8 [Link] 28.4 [Link]
Depth 5.5mm 44.0 [Link] 38.0 [Link] 34.4 [Link] 38.5 [Link]

018 / 019 COWELL® Implant System


2. Abutment Level Impression - Prosthesis Fabrication
Abutment Prosthetic Protocol * Two / One Piece Screw Retained Abutment
> For digital procedure, refer to the COWELL® Digital Products.
Submerged & Submerged Short : Multi S | Multi A | Lock | SFIT Absolute
Submerged Narrow : Multi S | Multi A
1. Fixture Level Impression - Prosthesis Fabrication
* Two Piece Screw Retained Abutment * One Piece Cemented Retained Abutment
Submerged & Submerged Short : Absolute | Straight (Direct)
Submerged & Submerged Short : Temporary | Easy Temporary | SFIT S | SFIT A Submerged Narrow : Straight
Internal : SFIT S | SFIT A Internal : Solid | Shoulder
External : Temporary External : Shoulder

* Two Piece Screw-Cement Retained / Cement Retained Abutment


* Two / One Piece Attachment Retained Abutment
Submerged & Submerged Short : Cemented | Angulated | Beauty-up™ | Milling | Meta G UCLA | Plastic UCLA Submerged & Submerged Short : Sonator S | Sonator A | Ball
Hybrid S | Hybrid L | Hybrid A | Ti-Block Internal : Sonator S | Ball
Submerged Narrow : Cemented | Angulated | Temporary | Meta G UCLA | Hybrid S | Hybrid L | Hybrid A External : Ball
Internal : Cemented | Angulated | Meta G UCLA | Hybrid S | Hybrid L
External : Cemented | Angulated | Temporary | Meta G UCLA | Plastic Sleeve Indirect Impression Technique
(No Abutment Modification Applied) Multi Pick-up Sonator Impression
Ø4.5 Abutment Ø5.5 Abutment Ø6.5 Abutment Impression Coping Multi Transfer Post Coping
(with Open Tray)
Pick up (Open Tray)

Remove the Connect the Fasten the Impression Cap Take impression Connect the Lab Analog
Healing Abutment. abutment with on the Abutment. with
t closed
c osed tray.
t ay. with the Impression
p Cap
Connect the Pick-up Impression Coping & Remove the Guide Pin. Connect the Replica with the the fixture. positioned in the imp
impression
ress
Guide Pin and take the impression Impression Coping positioned in tray.
with open tray. the impression tray, form the gum
and pour stone into the tray for
Single Crown : Hex / Octa Type working cast.
Bridge: Non-Hex / Non-Octa Type

Form the gum and pour The working cast Fasten the Plastic Coping Wax build-up Complete final Try-in final prosthesis
stone into the tray for with the gum. on the Lab Analog on the abutment. prosthesis. with the fixture.
PS working cast. (Absolute).

Remove the Connect the abutment Wax build-up Complete final Try-in final
Healing Abutment. with replica in the cast model. on the abutment. prosthesis. prosthesis
Direct Impression Technique (Abutment Modification Applied)
with the fixture.

Transfer (Closed Tray)

Remove the Connect the Modify


M odify the abutment for
fo Take impression Pure stone into the
en
nt.
Healing Abutment. abutment with pathway and aesthetics.
p aestheticss. with closed tray. tray for working cast.
the fixture.

Connect the Transfer Post & Guide Pin and Connect the Replica Interlock the Transfer Post
take the impression with closed tray. with the Transfer Post into the impression tray,
removed from form the gum and pour stone
Single Crown : Hex / Octa Type the impression tray. into the tray for working cast.
Bridge: Non-Hex / Non-Octa Type
The working cast Wax build-up Complete final Try-in final prosthesis
with the gum. on the abutment. prosthesis. with the fixture.

020 / 021 COWELL® Implant System


INNO SUBMERGED IMPLANT (Sub.)
System Flow

Fixture Abutment Impression

030p 030p 031p 031p 031p 032p 032p 032p 033p 034p 034p 034p

Fixture Level
Procedure I

Impression
Prosthetic

Cemented Angulated Beauty-up™ Milling Easy Temporary Temporary Meta G UCLA Plastic UCLA Replica Bite Impression Pick-up Impression Transfer Post
Coping Coping

036p 036p 037p 037p 037p 038p 038p 038p 039p 039p
Procedure II
Prosthetic

Multi S Multi A Multi Multi Pick-up Multi Multi Multi Meta G Multi Plastic Multi Titanium Multi Polishing
Protection Cap Impression Coping Transfer Post Lab Analog ULCA Cylinder UCLA Cylinder Cylinder Protector

Hex 2.5

042p 042p 042p 043p 043p 043p


Procedure III
Prosthetic

Length
7 / 8 / 10 Lock
12 / 14 Lock Lock Pick-up Lock Lock Meta G Lock Titanium
16 / 18mm Protection Cap Impression Coping Lab Analog UCLA Cylinder Cylinder

046p 046p 046p 046p 046p


Diameter
Procedure IV

Ø3.5 / 4.0 / 4.5 / 5.0 / 6.0


Prosthetic

Absolute Absolute Absolute Absolute Absolute


Protection Cap Impression Cap Lab Analog Plastic Coping
Hex 2.5

Abutment Level
047p
Procedure V

Impression
Prosthetic

Direct
Impression
1
2 Straight
Length
4mm

Diameter 049p 049p 050p 051p


Procedure VI

Ø4.0 / 4.5 / 5.0 / 5.5 / 6.0


Prosthetic

Sonator Impression Sonator


Sonator S Sonator A Coping Analog

052p 052p
Procedure VII
Prosthetic

Ball Ball Analog

022 / 023 COWELL® Implant System


No-Mount > Packing unit : 1 Fixture + 1 Cover Screw.
INNO Submerged Implant
Diameter
Ø3.5
Length

Submerged Fixture 7 -
SUB.
HEXAGON Surface Treatment : SLA-SH® 8 ST3508SM
Hex
8 10 12 14
SYSTEM > Interchangeable with hexagonal morse tapered fixture 10 ST3510SM L
> Internal hex connection (Taper 11°/ Hex 2.5)
12 ST3512SM
14 ST3514SM D

Hex 2.5 Diameter


Ø4.0
Length

7 ST4007SM
8 ST4008SM
7 8 10 12 14 16 18
10 ST4010SM
L
12 ST4012SM
14 ST4014SM
16 ST4016SM D

18 ST4018SM

Diameter
Ø4.5
L Length

7 ST4507SM
8 ST4508SM
7 8 10 12 14 16 18
10 ST4510SM L
12 ST4512SM
14 ST4514SM
D
16 ST4516SM
D 18 ST4518SM

Diameter
INNO Fixture Code Length
Ø5.0

S T 40 10 S M Ex.)
7 ST5007SM
Type body Diameter Length Surface Treatment Mount SLA No-Mount ST4010SM
Submerged Taper Ø4.0 10mm SLA No-Mount 8 ST5008SM 7 8 10 12 14
10 ST5010SM L

12 ST5012SM

S T 40 10 S Ex.) 14 ST5014SM
D
Type body Diameter Length Surface Treatment Mount SLA Pre-Mount ST4010S
Submerged Taper Ø4.0 10mm SLA Pre-Mount
Diameter
Ø6.0
Length

7 ST6007SM
7 8 10 12
8 ST6008SM
L
10 ST6010SM
12 ST6012SM
D
14 -

024 / 025 COWELL® Implant System


Pre-Mount

Diameter
> Packing unit : 1 Fixture + 1 Cover Screw + 1 Mount.
INNO Submerged Short Implant
Ø3.5
Length
8 10 12 14
7 - Submerged Short Fixture
8 ST3508S Hex SUB.
HEXAGON Surface Treatment : SLA-SH®
10 ST3510S L
SYSTEM > Interchangeable with Hexagonal Morse Tapered Fixture.
12 ST3512S > Internal hex connection (Taper 11°/ Hex 2.5).
14 ST3514S D

Diameter
Ø4.0
Hex 2.5
Length

7 ST4007S 7 8 10 12 14 16 18

8 ST4008S
10 ST4010S
12 ST4012S L

14 ST4014S 1(Machined Surface)


16 ST4016S
18 ST4018S
D
2(Sandblast)

Diameter
Ø4.5
L(SLA)
Length

7 8 10 12 14 16 18
7 ST4507S
8 ST4508S
D
10 ST4510S
L
12 ST4512S
Pre-Mount > Packing Unit : 1 Fixture + 1 Cover Screw + 1 Mount.
14 ST4514S
16 ST4516S D
Diameter 4 Diameter
18 ST4518S Ø4.0 Ø5.5 4
Length Length
1 1
2 2
Diameter 4 2ST4004S 4 2ST5504S
Ø5.0
Length L L
7 8 10 12 14
D D
7 ST5007S
8 ST5008S
Diameter
10 ST5010S L
Ø4.5
Length 4 Diameter 4
12 ST5012S Ø6.0
Length
14 ST5014S 1
1
D 2
4 2ST4504S 2
L
4 2ST6004S
L

D
Diameter D
Ø6.0
Length
7 8 10 12

7 ST6007S Diameter 4
Ø5.0
Length
8 ST6008S
L 1
10 ST6010S 2
4 2ST5004S
12 ST6012S L
14 - D
D

026 / 027 COWELL® Implant System


Fixture Mount
Prosthetic Procedure I
Length 5.4 Components Selection Guide for Cemented and UCLA Abutment
L
2SMHR001
> Packing unit : 1 Mount + 1 Mount Screw. 2.45 2.15

> Tightened with the 1.2 Hex Driver.


> Tightening torque force : 5~10 [Link].

Abutment Screw

Cover Screw D Diameter


Length Ø3.35 Ø3.75 Ø4.15
(15°) (25°) (15°) (25°)
L 2SCS000
3
4.2 2SCS001
5.2 2SCS002
> Packing unit : 1 Cover Screw. *Extra Product (Hex) (N-Hex) (Hex) (N-Hex) (Hex) (N-Hex) (Hex) (N-Hex) (Hex) (N-Hex) (Hex) (N-Hex)
> To seal the conical interface of fixture.
Cemented Angulated Beauty-up™ Milling
> The longer Cover Screw for deeply inserted fixture. Abutment Abutment Abutment Abutment
> Tightened with the 1.2 Hex Driver.
>Tightening torque force : 5~10 [Link].

Healing Abutment
D
Diameter Ø4.5 Ø5.5 Ø6.5 (Hex) (N-Hex) (Hex) (N-Hex) (Hex) (N-Hex) (Hex) (N-Hex)
L
C Length
Cuff 1 2 1 2 1 2 Easy Temporary Temporary Meta G UCLA Plastic UCLA
Abutment Abutment Abutment Abutment
1 2HS4511 2HS5511 2HS6511
2 2HS4522 2HS5522 2HS6522
3 2HS4532 2HS5532 2HS6532
4 2HS4542 2HS5542 2HS6542
5 2HS4552 2HS5552 2HS6552
7 2HS4572 2HS5572 2HS6572

Diameter Ø7.5 Ø8.5 Ø9.5 Replica


Length
Cuff 2 2 2
3 2HS7532 2HS8532 2HS9532

> Packing unit : 1 Healing Abutment.


> For remodeling gingival contour during soft tissue healing.
> Recommended to use with the Volume-up™ Guide System.
> Select according to gingival height and abutment type.
> Tightened with the 1.2 Hex Driver.
> Tightening torque force : 5~10 [Link].

(Hex) (Hex) (N-Hex) (Hex) (N-Hex)


Bite Impression Coping Pick-up Impression Coping Transfer Post
Volume-up™ Healing Abutment
D

L
Diameter Ø6.5 Ø7.5 Ø8.5
Length
H C Cuff 2 2 2
7.9
3 VUHN6532 VUHN7532 VUHN8532
> Packing unit : 1 Volume-up™ Healing Abutment (Inbuilt Abutment Screw).
> Used for an implant procedure to form the gingival tissue and alveolar bone in the form 1mm
of natural teeth and gums by prevention or minimizing the food penetration. off-centered
> Extremely effective when used with the COWELL® BMP.
> Recommended to use with the Volume-up™ Guide System.
> Select according to gingival height and abutment type.
> Tightened with the 1.2 Hex Driver.
> Tightening torque force : 25~35 [Link].
Healing
Fixture Cover Screw Abutment

028 / 029 COWELL® Implant System


Cemented Abutment Type Hex
Beauty-up™ Abutment Type Hex N-Hex Hex N-Hex
1.2 Hex. Diameter Ø4.5 Ø5.5 Ø6.5 A Diameter(Angle) Ø3.8 (15˚) Ø3.8 (15˚) Ø3.8 (25˚) Ø3.8 (25˚)
D Length Length
4 5.5 7 4 5.5 7 4 5.5 7 Cuff 5 5 5 5
2.15 Cuff L
L 1 2SCH4514 2SCH4515 2SCH4517 2SCH5514 2SCH5515 2SCH5517 2SCH6514 2SCH6515 2SCH6517 2 2SBH381525 2SBN381525 2SBH382525 2SBN382525
C
C 2 2SCH4524 2SCH4525 2SCH4527 2SCH5524 2SCH5525 2SCH5527 2SCH6524 2SCH6525 2SCH6527 > Packing unit : 1 Beauty-up™ Abutment > Tightened with the Angulated Screw Ratchet Drive
Angulated Screw
Ratchet Drive (Inbuilt Abutment Screw). and Torque Wrench.
3 2SCH4534 2SCH4535 2SCH4537 2SCH5534 2SCH5535 2SCH5537 2SCH6534 2SCH6535 2SCH6537
D > For Screw-Cement Retained Prosthesis > Tightening torque force : 30 [Link].
4 2SCH4544 2SCH4545 2SCH4547 2SCH5544 2SCH5545 2SCH5547 2SCH6544 2SCH6545 2SCH6547 with angulated screw channel. > Library available for EXOCAD®, 3Shape® & Others.
5 2SCH4554 2SCH4555 2SCH4557 2SCH5554 2SCH5555 2SCH5557 2SCH6554 2SCH6555 2SCH6557 > Solution for anterior esthetic zone. > Use the Scanbody for 3D Work.
Hex N-Hex
> The gingival line of the Beauty-up™ Abutment > Fixture level impression.
Hex N-Hex allows more esthetic prosthesis.
Type N-Hex > Oval design allows lower incisal application
(Mesiodistal diameter : 3.8mm).
Diameter Ø4.5 Ø5.5 Ø6.5
Length
Cuff 4 5.5 7 4 5.5 7 4 5.5 7 * Angulated Screw Ratchet Driver
1 2SCN4514 2SCN4515 2SCN4517 2SCN5514 2SCN5515 2SCN5517 2SCN6514 2SCN6515 2SCN6517 Height
Type Ratchet
2 2SCN4524 2SCN4525 2SCN4527 2SCN5524 2SCN5525 2SCN5527 2SCN6524 2SCN6525 2SCN6527 24(Short) KRBUD15
29(Long) KRBUD20
3 2SCN4534 2SCN4535 2SCN4537 2SCN5534 2SCN5535 2SCN5537 2SCN6534 2SCN6535 2SCN6537
H
> Stable to internal slip or fracture due to wide contact area of the Angulated Driver and
4 2SCN4544 2SCN4545 2SCN4547 2SCN5544 2SCN5545 2SCN5547 2SCN6544 2SCN6545 2SCN6547 the dedicated Stargrip Abutment Screw.
5 2SCN4554 2SCN4555 2SCN4557 2SCN5554 2SCN5555 2SCN5557 2SCN6554 2SCN6555 2SCN6557 > Tightening torque force : 30 [Link] (50 [Link] Max.).

> Packing unit : 1 Cemented Abutment + 1 Abutment Screw.


> For Screw-Cement or Cement Retained Prosthesis. Milling Abutment
> Cutting surface for anti-rotation of prosthesis. 1.2 Hex. Type Hex N-Hex
> Gold color for more translucent restoration. D
Diameter Ø4.5 Ø5.5 Ø6.5 Ø4.5 Ø5.5 Ø6.5
> Library available for EXOCAD®, 3Shape® & Others. 2.45 Length
> Connected with the Abutment Screw (2SSHR200). L Cuff 7 7 7 7 7 7
> Tightened with the 1.2 Hex Driver and Torque Wrench. 2 2SMH4527 2SMH5527 2SMH6527 2SMN4527 2SMN5527 2SMN6527
C
> Tightening torque force : 30 [Link].
> Use the Scanbody for 3D Work. 4 2SMH4547 2SMH5547 2SMH6547 2SMN4547 2SMN5547 2SMN6547
> Fixture level impression.
> Packing unit : 1 Milling Abutment + 1 Abutment Screw.
> For Screw-Cement or Cement Retained Prosthesis.
Hex N-Hex > Block abutment for customized contouring.
> Gold color for more translucent restoration.
> Connected with the Abutment Screw (2SSHR100).
Angulated Abutment Type Hex-A Hex-B > Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 30 [Link].
Diameter(Angle) Ø4.5(15°) Ø4.5(25°) Ø5.5(15°) Ø5.5(25°) Ø4.5(15°) Ø4.5(25°) Ø5.5(15°) Ø5.5(25°) > Fixture level impression.
A 1.2 Hex.
Length
Cuff 8 8 8 8 8 8 8 8
2.45 1 2SAH45151 2SAH45251 2SAH55151 2SAH55251 2SAH45151B 2SAH45251B 2SAH55151B 2SAH55251B
L
2 2SAH45152 2SAH45252 2SAH55152 2SAH55252 2SAH45152B 2SAH45252B 2SAH55152B 2SAH55252B
C
3 2SAH45153 2SAH45253 2SAH55153 2SAH55253 2SAH45153B 2SAH45253B 2SAH55153B 2SAH55253B
D 4 2SAH45154 2SAH45254 2SAH55154 2SAH55254 2SAH45154B 2SAH45254B 2SAH55154B 2SAH55254B
Hex N-Hex
Type N-Hex
Easy Temporary Abutment
Type Hex N-Hex
Hex-A D
Diameter(Angle) Ø4.5(15°) Ø4.5(25°) Ø5.5(15°) Ø5.5(25°) 1.2 Hex. Diameter Ø4.5 Ø5.5 Ø4.5 Ø5.5
Length
Cuff 8 8 8 8 Length
10 10 10 10
L Cuff
Hex-B 2.15
1 2SAN45151 2SAN45251 2SAN55151 2SAN55251 2 2STHA45C 2STHA55C 2STNA45C 2STNA55C
2 2SAN45152 2SAN45252 2SAN55152 2SAN55252 C
> Packing unit : 1 Easy Temporary Abutment + 1 Abutment Screw.
3 2SAN45153 2SAN45253 2SAN55153 2SAN55253 > For Screw Retained Prosthesis.
4 2SAN45154 2SAN45254 2SAN55154 2SAN55254 > For simpler and speedier chair-side process.
> Venerable polymer material.
> Packing unit : 1 Angulated Abutment + 1 Abutment Screw. > Temporary restoration for anterior esthetic zone.
Hex N-Hex
> For Screw-Cement or Cement Retained Prosthesis. > Titanium core for strength.
> Solution for anterior esthetic zone. > Connected with the Abutment Screw (2SSHR200).
> Connected with the Abutment Screw (2SSHR100). > Tightened with the 1.2 Hex Driver and Torque Wrench.
> Gold color for more translucent restoration. > Tightening torque force : 20 [Link].
> Select Hex-A or Hex-B according to case. > Fixture level impression.
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 30 [Link].
> Fixture level impression.

030 / 031 COWELL® Implant System


Temporary Abutment Abutment Screw
Diameter
D Height Ø2.45 Ø2.15
Type Hex N-Hex 1.2 Hex. 1.2 Hex.
1.2 Hex.
8.5 2SSHR100 2SSHR200
Diameter Ø4.5 Ø4.5
L 2.45 Length D D > Packing unit : 1 Abutment Screw.
Cuff 10 10
> 2SSHR100 : Hybrid Block / Scanbody / Angulated / Milling / Temporary / Meta G UCLA / Plastic UCLA Abutment.
C
1 2STHA45 2STNA45 H > 2SSHR200 : Hybrid S & L Ti-Base / Cemented / Easy Temporary Abutment.
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Packing unit : 1 Temporary Abutment + 1 Abutment Screw.
> For Screw-Cement Retained Prosthesis.
> For provisional restoration.
> Connected with the Abutment Screw (2SSHR100).
Hex N-Hex
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 20 [Link].
> Fixture level impression.

Meta G UCLA Abutment Replica D


D Type Hex N-Hex Diameter
Height Ø4
1.2 Hex.
Diameter Ø4.5 Ø4.5 12 2SRHR001
H
Length
2.45 Cuff 12 12 > Packing unit : 1 Replica.
L
1 2SGH45N 2SGN45N > Mimicking of conical interface of fixture.
> Analog of fixture for working cast.
C 2 2SGH452N 2SGN452N
3 2SGH453N 2SGN453N

> Packing unit : 1 Meta G UCLA Abutment + 1 Abutment Screw.


> For Screw-Cement or Screw Retained Prosthesis.
Hex N-Hex > Modification to angulated abutment, Customized abutment and telescopic abutment.
> CCM alloy core for precise connection.
> Cast with non-precious metal or gold alloy.
> Connected with the Abutment Screw (2SSHR100).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 30 [Link].
> Fixture level impression.

Plastic UCLA Abutment


D
Type Hex N-Hex
1.2 Hex.
Diameter Ø4.5 Ø5.5 Ø4.5 Ø5.5
Length
2.45 Cuff 11 11 11 11
L
3 2SPHR001 2SPHW001 2SPNR001 2SPNW001
C
> Packing unit : 1 Plastic UCLA Abutment + 1 Abutment Screw.
> Same purpose of use as the Meta G UCLA Abutment but low accuracy of connection during lab procedure.
> PMMA material.
> Connected with the Abutment Screw (2SSHR100).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
Hex N-Hex > Tightening torque force : Finger light force during wax pattern fabrication, 30 [Link] after casting.
> Fixture level impression.

032 / 033 COWELL® Implant System


Bite Impression Coping
D Type Hex(Short) Hex(Long) Hex(X-Long)
Prosthetic Procedure II
Diameter Ø4.5 Ø4.5 Ø4.5 Component Selection Guide for Multi S&A Abutment
L
Cuff
Length 2 4 6
C 4.0 2SBIC45S 2SBIC45L 2SBIC45X

> Packing unit : 1 Bite Impression Coping (Inbuilt Guide Pin).


Multi Polishing Protector
> Designed to simultaneously take bite and impression.
> For Closed Tray Impression (Bite Impression).
(Short) (Long) (X-Long) > Tightened with the 1.2 Hex Driver and Torque Wrench.

Hex Multi Cylinder Screw

(Hex) (N-Hex) (Hex) (N-Hex) (Hex) (N-Hex)


Pick-up Impression Coping Multi Meta G Multi Plastic Multi Titanium
UCLA Cylinder UCLA Cylinder Cylinder
D Type Hex N-Hex
Diameter
Length / Cuff Ø4.5 Ø5.5 Ø6.5 Ø4.5 Ø5.5 Ø6.5
14 (Short) / 4 2SIH454S 2SIH554S 2SIH654S 2SIN454S 2SIN554S 2SIN654S
L
14 (Short) / 2 2SIH45S 2SIH55S 2SIH65S 2SIN45S 2SIN55S 2SIN65S

C
16 (Long) / 4 2SIH45L 2SIH55L 2SIH65L 2SIN45L 2SIN55L 2SIN65L

> Packing unit : 1 Pick-up Impression Coping + 1 Guide Pin.


Lab Analog
> For open tray impression.
(Short) (Short) (Long) (Short) (Short) (Long)
> Connected with the Guide Pin (2SISR001SS / 2SISR001SL).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
2 Pieces > Tightening torque force: 12~15Ncm.
Hex N-Hex

(Hex) (N-Hex) (Hex) (N-Hex)


Multi Pick-up Multi Transfer Post
Impression Coping

Transfer Post
D
Type Hex N-Hex
Diameter Multi Protection Cap
Length / Cuff Ø4.5 Ø5.5 Ø6.5 Ø4.5 Ø5.5 Ø6.5
L
9 (Short) / 2 2STH45S 2STH55S 2STH65S 2STN45S 2STN55S 2STN65S
11 (Long) / 4 2STH45L 2STH55L 2STH65L 2STN45L 2STN55L 2STN65L
C
> Packing unit : Hex - 1 Transfer Post + 1 Guide Pin / N-Hex - 1 Transfer Post (Solid Type). (15°) (30°)

> For closed tray impression.


(Short) (Long) (Short) (Long)
> Connected with the Guide Pin (2STH001SS / 2STH001SL).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
2 Pieces 1 Piece > Tightening torque force: 12~15Ncm. (Hex) (N-Hex) (Hex) (N-Hex)

Multi S Abutment Multi A Abutment

Hex N-Hex

Healing
Fixture Cover Screw Abutment

034 / 035 COWELL® Implant System


Multi S Abutment Diameter Ø4.5 Ø5.5
Multi Protection Cap
Multi S & A
Abutment Diameter Ø4.5 Ø5.5
Length 2 2 D
Cuff Diameter
D
Height Ø5.2 Ø6.2
1 2SMS451 2SMS551
L H
5 2SMPC45 2SMPC55
C 2 2SMS452 2SMS552
1 Piece 2SMS453 2SMS553 > Packing unit : 1 Multi Protection Cap.
3
> Protection from cheek and tongue for gingival healing period.
4 2SMS454 2SMS554 > Gingival retraction for prosthodontic margin of abutment.
5 2SMS455 2SMS555 > Alternative usage for sub-structure of temporary prosthesis.

> Packing unit : 1 Multi S Abutment.


> For Screw-Retained Prosthesis.
> Titanium base for the cylinders.
> Gold color for more translucent restoration.
> Integrated with screw and abutment.
> Library available for EXOCAD®, 3Shape® & Others.
> Use the S Holder for more stable position.
> Tightened with the S Machine & S Ratchet Driver and Torque Wrench.
> Tightening torque force : 30 [Link].
> Abutment level impression.

Multi A Abutment Type Hex


1.2 Hex.
Multi Pick-up Impression Coping
Diameter(Angle) Ø4.5(15°) Ø4.5(30°) Ø5.5(15°) Ø5.5(30°)
D
L Length D
A 2.15 Cuff 2 2 2 2 Type Hex N-Hex
C 2 Multi S & A
2SMAH45152 Abutment Diameter Ø4.5 Ø5.5 Ø4.5 Ø5.5
3 2SMAH45303 2SMAH55153 2SMAH55303 Diameter
2SMAH45153 H Height Ø4.65 Ø5.65 Ø4.65 Ø5.65
4 2SMAH45154 2SMAH45304 2SMAH55154 2SMAH55304 14.8 2SMIH45 2SMIH55 2SMIN45 2SMIN55
5 2SMAH55155 2SMAH55305
2 Pieces > Packing unit: 1 Multi Pick-up Impression Coping + 1 Guide Pin.
Hex N-Hex Type N-Hex > For open tray impression.
> Connected with the Guide Pin (2SMGP012).
Diameter(Angle) Ø4.5(15°) Ø4.5(30°) Ø5.5(15°) Ø5.5(30°) > Tightened with the 1.2 Hex Driver and Torque Wrench.
Length Hex N-Hex > Tightening torque force: 12~15 [Link].
Cuff 2 2 2 2
2 2SMAN45152
3 2SMAN45153 2SMAN45303 2SMAN55153 2SMAN55303
4 2SMAN45154 2SMAN45304 2SMAN55154 2SMAN55304
5 2SMAN55155 2SMAN55305
> Packing unit: 1 Multi A Abutment + 1 Abutment Screw.
> For Screw-Retained Prosthesis.
> Titanium base for the cylinders.
> Gold color for more translucent restoration.
> Library available for EXOCAD®, 3Shape® & others.
> Use the A Holder for more stable position. Multi Transfer Post
> Connected with the Abutment Screw (2SSHR300 : ★ / 2SSHR400 : ● ). Type Hex N-Hex
> Tightened with the 1.2 Hex Driver and Torque Wrench. D
Multi S & A
Abutment Diameter Ø4.5 Ø5.5 Ø4.5 Ø5.5
> Tightening torque force: 30 [Link].
Diameter
> Use the Multi Scanbody for digital flow. Height Ø4.5 Ø5.5 Ø4.5 Ø5.5
H
> Abutment level impression. 8.5 2SMTH45 2SMTH55 2SMTN45 2SMTN55

Abutment Screw > Packing unit: 1 Multi Transfer Post + 1 Guide Pin.
1.2 Hex. Height > For closed tray impression.
Diameter 7.5 6.5 > Connected with the Guide Pin (2SMTHS100).
Hex N-Hex
D
2.15 2SSHR300 2SSHR400 > Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force: 12~15 [Link].
> Packing unit: 1 Abutment Screw.
H > To connect the Multi A Abutment.
> Tightened with the 1.2 Hex Driver and Torque Wrench.

036 / 037 COWELL® Implant System


Multi Lab Analog Multi Titanium Cylinder Type Hex N-Hex
D Multi S & A Multi S & A
Abutment Diameter Ø4.5 Ø5.5 1.2 Hex. Abutment Diameter Ø4.5 Ø5.5 Ø4.5 Ø5.5
L Diameter
Length Ø4.5 Ø5.5 Diameter Ø4.5 Ø5.5 Ø4.5 Ø5.5
L 2.25 Length
2 2SMA45 2SMA55 Cuff 8.5 8.5 8.5 8.5
> Packing unit : 1 Multi Lab Analog. 0.5 2STCH45 2STCH55 2STCN45 2STCN55
C
> Replacement of abutment shape in working cast. D
> Choose by abutment size. > Packing unit : 1 Multi Titanium Cylinder + 1 Multi Cylinder Screw.
> For Screw, Cement or Screw-Cement Retained Prosthesis.
> Connected with the Multi Cylinder Screw (2SMCS100).
Hex N-Hex > Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 20 [Link].

Multi Meta G UCLA Cylinder Type Hex N-Hex Multi Cylinder Screw
Diameter
1.2 Hex.
Multi S & A
Ø4.5 Ø5.5 Ø4.5 Ø5.5 Height Ø2.25
Abutment Diameter
1.2 Hex.
Diameter Ø4.5 Ø5.5 Ø4.5 Ø5.5 5 2SMCS100
L 2.25 Length
Cuff 10.9 10.9 10.9 10.9 D > Packing unit : 1 Multi Cylinder Screw.
0.5 2SCCH45 2SCCH55 2SCCN45 2SCCN55 > Connected with the Meta G UCLA, Plastic UCLA and Titanium Cylinder.
H > Tightened with the 1.2 Hex Driver and Torque Wrench.
C
D > Packing unit : 1 Multi Meta G UCLA Cylinder + 1 Multi Cylinder Screw. > Tightening torque force : 20 [Link].
> For Screw, Cement or Screw-Cement Retained Prosthesis.
> Modification to various types of abutments.
> CCM alloy core for precise connection.
Hex N-Hex > Cast with non-precious metal or gold alloy.
> Connected with the Multi Cylinder Screw (2SMCS100).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 20 [Link].

Multi Plastic UCLA Cylinder Type Hex N-Hex Multi Polishing Protector Type Hex
Multi S & A Multi S & A
Abutment Diameter Ø4.5 Ø5.5 Ø4.5 Ø5.5 D Abutment Diameter Ø4.5 Ø5.5
Diameter
1.2 Hex. Diameter Ø4.5 Ø5.5 Ø4.5 Ø5.5 L Length Ø4.5 Ø5.5
Length 2 2SMPP45 2SMPP55
Cuff 11.5 11.5 11.5 11.5
L 2.25
0.5 2SMPH45 2SMPH55 2SMPN45 2SMPN55
> Packing unit : 1 Multi Polishing Protector.
> Packing unit : 1 Multi Plastic UCLA Cylinder + 1 Multi Cylinder Screw. > For polishing work during lab procedure.
C
> For Screw, Cement or Screw-Cement Retained Prosthesis.
D
> Same purpose of use as the Meta G UCLA Cylinder but low accuracy of connection.
> PMMA material.
> Connected with the Multi Cylinder Screw (2SMCS100).
Hex N-Hex > Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 20 [Link].

038 / 039 COWELL® Implant System


Multi Holder Prosthetic Procedure III
H
Component Selection Guide for Lock Abutment
Type
Height Hand
20 KMHS01
S Holder > Packing unit: 1 Multi S Holder.
> To position the Multi S Abutment more stably.

H Lock Cylinder Screw


Type
Height Hand
32 KMHA01
A Holder
> Packing unit: 1 Multi A Holder.
> To position the Multi A Abutment more stably.

Lock Meta G UCLA Cylinder Lock Titanium Cylinder

˝ ˞ ˟ ˠ
Connect the Multi A Holder with the Hold the handle of the Multi A Connect the assembled body Tighten the Multi Abutment Lock Lab Analog
Multi A Abutment with its Abutment Holder and bend it according with the fixture. with the 1.2 Hex Driver and
Screw in it and match the direction of to the placement position in Torque Wrench.
holes of the abutment and the holder. the oral cavity.

Lock Pick-up Impression Coping


Multi S Machine Driver Type
Height Machine
H 27.5 KMMSD21L

> Packing unit: 1 Multi S Machine Driver.


> To install the Multi S Abutment by machine.
Lock Protection Cap

Multi S Ratchet Driver


Type
Height Ratchet
H
22 KRMSD15L Lock Abutment

> Packing unit: 1 Multi S Ratchet Driver.


> To install the Multi S Abutment by hand.

Healing
Fixture Cover Screw Abutment

040 / 041 COWELL® Implant System


Lock Abutment Diameter Ø3.5 Lock Lab Analog Lock Abutment
Diameter Ø3.5
Length D
D
Cuff 2.15 Diameter
L Length Ø3.5
L 0.5 2SLA400
2.15 2SLA45
C
1 2SLA410
2
> Packing unit : 1 Lock Lab Analog.
2SLA420
> Replacement of abutment shape in working cast.
3 2SLA430 > Tightened with the 1.2 Hex Driver and Torque Wrench.
4 2SLA440

> Packing unit : 1 Lock Abutment.


> For Screw-Retained Prosthesis.
> Titanium base for cylinder.
> Gold color for more translucent restoration.
> Integrated with screw and abutment.
> Tightened with the Lock Ratchet Driver and Torque Wrench.
> Tightening torque force : 30 [Link].
> Abutment level impression.

Lock Meta G UCLA Cylinder


Lock Protection Cap Lock Abutment Ø3.5
Lock Abutment
Diameter Ø3.5
Diameter 1.2 Hex.
D Diameter
Diameter Height Ø3.5
Ø3.5
Height
H 2.3 11.2 2SLCH45
H 2SLP45
4.3
> Packing unit : 1 Lock Protection Cap. > Packing unit : 1 Lock Meta G UCLA Cylinder + 1 Lock Cylinder Screw.
> Protection from cheek and tongue for gingival healing period. D > For Screw, Cement or Screw-Cement Retained Prosthesis.
> Gingival retraction for prosthodontic margin of abutment. > Modification to various types of abutments.
> CCM alloy core for precise connection.
> Cast with non-precious metal or gold alloy.
> Connected with the Lock Cylinder Screw (2SLCS200).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 30 [Link].

Lock Titanium Cylinder


Lock Pick-up Impression Coping
1.2 Hex.
Lock Abutment Lock Abutment
Diameter Ø3.5 Diameter Ø3.5
Diameter H 2.3 Diameter
Height Ø3.5 Height Ø3.5
H
16 2SLIH45 10 2SLTH45
D
> Packing unit : 1 Lock Pick-up Impression Coping + 1 Guide Pin. > Packing unit : 1 Lock Titanium Cylinder + 1 Lock Cylinder Screw.
D > Connected with the Guide Pin (2SLIH45S). > For Screw, Cement or Screw-Cement Retained Prosthesis.
> For open tray impression. > Connected with the Lock Cylinder Screw (2SLCS200).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 30 [Link].

042 / 043 COWELL® Implant System


Lock Cylinder Screw
1.2 Hex. Height
Diameter
Ø2.3
Prosthetic Procedure IV
4.8 2SLCS200 Component Selection Guide for Absolute Abutment
D
> Packing unit : 1 Lock Cylinder Screw.
H > Connected with the CCM Cylinder or Titanium Cylinder.
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 30 [Link].

(Crown) (Bridge)

Absolute Plastic Coping

Absolute Lab Analog


Lock Ratchet Driver
H
Type
Height Ratchet
14.2 KRLRD18

H
28.5 KRLRD28
> Packing unit : 1 Lock Ratchet Driver. Absolute Impression Cap
> To install the Lock Abutment by hand.

Absolute Protection Cap

Absolute Abutment

Healing
Fixture Cover Screw Abutment

044 / 045 COWELL® Implant System


Absolute Abutment
D
Diameter

Cuff
Length
4
Ø4.5
5.5 7 4
Ø5.5
5.5 7 4
Ø6.5
5.5 7
Prosthetic Procedure V
1 2SAC4514 2SAC4515 2SAC4517 2SAC5514 2SAC5515 2SAC5517 2SAC6514 2SAC6515 2SAC6517
L
1.2 Hex Driver
Component Selection Guide for Straight Abutment
2 2SAC4524 2SAC4525 2SAC4527 2SAC5524 2SAC5525 2SAC5527 2SAC6524 2SAC6525 2SAC6527
C
3 2SAC4534 2SAC4535 2SAC4537 2SAC5534 2SAC5535 2SAC5537 2SAC6534 2SAC6535 2SAC6537
4 2SAC4544 2SAC4545 2SAC4547 2SAC5544 2SAC5545 2SAC5547 2SAC6544 2SAC6545 2SAC6547
5 2SAC4554 2SAC4555 2SAC4557 2SAC5554 2SAC5555 2SAC5557 2SAC6554 2SAC6555 2SAC6557

> Packing unit : 1 Absolute Abutment + 1 Production Cap. > Tightened with the 1.2 Hex Driver /
> For Cement Retained Prosthesis. Absolute Ratchet Driverand Torque Wrench.
> Cutting surface for anti-rotation of prosthesis. > Tightening torque force : 30 [Link].
> Integrated with the Screw and Abutment. > Abutment level impression.
Straight Abutment

Absolute Ratchet Driver


Diameter Ø4.6 Ø5.6 Ø6.6
Length
Height 12 19 12 19 12 19
H 19 KRAD4512S KRAD5512S KRAD6512S
L
26 KRAD4519L KRAD5519L KRAD6519L
D

Absolute Protection Cap Absolute Abutment


Diameter Ø4.5 Ø5.5 Ø6.5 Healing
Fixture Fixture Cover Screw Abutment
D Diameter
Height Ø5.0 Ø6.0 Ø7.0
H 6 2SHPC454 2SHPC554 2SHPC654
7.5 2SHPC455 2SHPC555 2SHPC655
9 2SHPC457 2SHPC557 2SHPC657
> Packing unit : 1 Absolute Protection Cap . Straight Abutment
> Protection from cheek and tongue for gingival healing period.
> Gingival retraction for prosthodontic margin of abutment. D Diameter Ø3.5 Ø4.5
> Alternative usage for sub-structure of temporary prosthesis. Length
8 8
Cuff
7 5.5 L 0.5 2SSCM308 2SSCR408
4
Absolute Impression Cap 1 2SSCM318 2SSCR418
Absolute
[Link]
Dia. Ø4.5 Ø5.5 Ø6.5 C
D Diameter
Diameter 2 2SSCM328 2SSCR428
Height Ø5.5 Ø6.5 Ø7.5
3 2SSCM338 2SSCR438
H 10.3 2SIC45 2SIC55 2SIC65
4 2SSCM348 2SSCR448
> Packing unit : 1 Absolute Impression Cap.
> Confirm locking with abutment by rotation of clockwise and anti-clockwise direction. > Packing unit : 1 Straight Abutment.
> For Cement Retained Prosthesis.
> Integrated with screw and abutment.
> Tightened with the Shoulder Driver.
Absolute Lab Analog Absolute Abutment
Diameter Ø4.5 Ø5.5 Ø6.5 > Tightening torque force : 30 [Link].
D
Diameter Ø4.5 Ø5.5 Ø6.5 > Direct impression.
Length

L
4.1 2SHLA454 2SHLA554 2SHLA654
5.6 2SHLA455 2SHLA555 2SHLA655
7.1 2SHLA457 2SHLA557 2SHLA657
> Packing unit : 1 Absolute Lab Analog.
> Replacement of abutment shape in working cast. Shoulder Driver
> Choose according to width and length of abutment.
H Diameter
Height Ø4.5
Absolute Plastic Coping (Burn Out Cylinder) D 19 KRR19L
Type Crown Bridge > Packing unit : 1 Driver.
H H Absolute Abutment
Ø4.5 Ø5.5 Ø6.5 Ø4.5 Ø5.5 Ø6.5 > To install the Straight Abutment.
Diameter
Diameter
Height Ø5.1 Ø6.1 Ø7.1 Ø5.1 Ø6.1 Ø7.1
D
10 2SHBC45 2SHBC55 2SHBC65 2SHBB45 2SHBB55 2SHBB65
> Packing unit : 1 Absolute Plastic Coping.
Crown Bridge > Connected with the Lab Analog.
> Burn out and casting for metal framework.

046 / 047 COWELL® Implant System


Prosthetic Procedure VI Sonator S Abutment
Block-out
Black
Processing
Outer Cap Spacer Male H-Type Inner Caps
Component Selection Guide for Sonator Abutment
2.25mm

Ø5.4
Blue : 10N Pink :15N White : 22N
Black Carrier
Block-out Processing
Outer Cap Spacer Male

Diameter Ø4.0
Cuff 1 2 3 4 5 6
Length
1.5 SONS401 SONS402 SONS403 SONS404 SONS405 SONS406

> Packing unit : 1 Sonator S Abutment + 1 Carrier + 3 H-Type Inner Caps + 1 Outer Cap
Ratchet Driver + 1 Block-out Spacer + 1 Black Processing Male.
D
> For Implant Supported Over-denture Prosthesis.
L > Stable with low vertical height.
H-Type Inner Caps NH-Type Inner Caps
C > 6 kinds of the Inner Caps give various holding force (Both, H and NH-Type Inner Caps are used for the Sonator S Abutment).
> Path compensation up to 20° based on 2 implants.
> Carrier : Used for delivery of the abutment.
> Tightened with the Sonator S Ratchet Driver and Torque Wrench.
> Tightening torque force : 30 [Link].
> Abutment level impression.

Sonator A Abutment
Black
Block-out Processing
Outer Cap Spacer Male NH-Type Inner Caps

2.25mm
Analog
Ø5.4
Red : 10N Orange : 15N Green : 22N

Diameter Ø4.0
Carrier
Length 1.5 3.0
Impression Coping Cuff 3 3
Angle
15° SONA415 SONA430

> Packing unit : 1 Sonator A Abutment + 1 Abutment Screw + 1 Carrier + 3 NH-Type Inner Caps + 1 Outer Cap
D 1.2 Hex Driver + 1 Block-out Spacer + 1 Black Processing Male.
A > For Implant Supported Over-denture Prosthesis.
> Stable with low vertical height.
1.2 Hex. > 6 kinds of Inner Caps give various holding force (Both, H and NH-Type Inner Caps are used for the Sonator A Abutment).
L > Path compensation up to 40° based on 2 Implants.
C 2.15
> Connected with the Abutment Screw (2SSHR300).
Sonator S Abutment Sonator A Abutment
> Carrier : Used for delivery of the abutment.
> Tightened with the 1.2 Hex Driver and Torque wrench.
> Tightening torque force : 30 [Link].
> Abutment level impression.

Abutment Screw
1.2 Hex.

Healing
Fixture Cover Screw Abutment D
Diameter
Height Ø2.15
H 7.5 2SSHR300

048 / 049 COWELL® Implant System


Outer Cap Sonator Lab Analog
Diameter Diameter
D
Black Processing Male Height Ø5.4 Length Ø4
D
2.25 SONOC01 L 1.4 SONLA04
H
> Packing unit : 2 Outer Caps and 2 Black Processing Males. > Packing unit : 4 Sonator Lab Analogs.
> Black Processing Male: Inserted and removed with the I&R Driver. > Replacement of abutment shape in working cast.

H-Type Inner Cap Code SONIC01


> Packing unit : 3 Block-out Spacers + 3 Inner Caps (1 Blue, 1 Pink and 1 White).
Block-out
> Path compensation up to 20° based on 2 implants.
Spacer Inner Cap > Mainly used for the Sonator S Abutment.
> Inner Caps: Inserted and removed with the I&R Driver.

Blue : Pink : White : Sonator S Ratchet Driver


Retention Force Retention Force Retention Force
About 10N About 15N About 22N H
Type
Height Ratchet
18 SONRD19L
> Used to tighten and untighten the Sonator S Abutment.

NH-Type Inner Cap Code SONIC02


> Packing unit : 3 Block-out Spacers + 3 Inner Caps (1 Red, 1 Orange and 1 Green).
> Non-humped design.
Block-out
Spacer Inner Cap > Path compensation up to 40° based on 2 implants.
> Mainly used for the Sonator A Abutment.
> Inner Caps : Inserted and removed with the I&R Driver.

Red : Orange : Green :


Retention Force Retention Force Retention Force
About 10N About 15N About 22N
I & R Driver
Height 95.4
SONIR002

For Removal For Insertion

Sonator Impression Coping > Used to insert and remove the Inner Caps and Block Processing Male.
Diameter
Length Ø4.8
D
Black Processing Male
3 SONIP04

L
> Packing unit : 4 Sonator Impression Coping and 4 Black Processing Males.
> Connected over the Sonator Abutment after placing the Block-out Spacer.
> For close tray impression.

050 / 051 COWELL® Implant System


Prosthetic Procedure VII Ball Abutment
Outer Cap Block-out Spacer
Diameter
Length
Ø4.0
4
Cuff
Component Selection Guide for Ball Abutment 1 2SBAT414R
2.4mm

Ø3.4 2 2SBAT424R
3 2SBAT434R
4 2SBAT444R
5 2SBAT454R
Red : 10N Orange : 15N Green : 20N

Ball Inner Cap > Packing unit : 1 Ball Abutment + 3 Inner Caps (1 per each colour) + 1 Block-out Spacer + 1 Outer Cap.
D > For Implant Supported Over-denture Prosthesis.
> Tightened with the Ball Driver and Torque Wrench.
L
> Tightening torque force : 30 [Link].
Ball Outer Cap Block-out Spacer > Direct impression.
C

Ball Abutment

Red : 10N Orange : 15N Green : 20N


Ball Inner Cap Ball Outer Cap Diameter
D Height Ø3.4
2.4 BATC003C
H
> Packing unit : 2 Outer Caps.

Ball Inner Cap


Block-out Ball Inner Cap
Spacer BATC003I
> Packing unit: 2 Block-out Spacers + 6 Inner Caps (2 per each color).
Ball Analog > Retention force: Red 10N, Orange 15N & Green 20N.
Red : 10N Orange : 15N Green : 20N

Ball Analog
D
Diameter
Length Ø4.0
L
4 SBAL400

> Packing unit : 4 Lab Analogs.


> Replacement of abutment shape in working cast.

Ball Abutment

Ball Driver
Type
H Height Ratchet
19 KRB19L
> Packing unit : 1 Ball Driver. *Extra Product
> Used with the Torque Wrench to tighten and untighten the Ball Abutment.
Healing
Fixture Cover Screw Abutment

I&R Driver Height 100


KBIR01

For Removal For Insertion

> Packing unit : 1 I&R Driver.


> Used to insert and remove the Inner Caps into and out of the Outer Cap.

052 / 053 COWELL® Implant System


INNO SUBMERGED NARROW IMPLANT (Sub-N.)
System Flow

Fixture Abutment Impression

059p 059p 059p 060p 060p 061p 061p

Fixture Level
Procedure I

Impression
Prosthetic

Replica Pick-up Impression Coping Transfer Post


Cemented Angulated Temporary Meta G UCLA

Hex 2.1

063p 063p 064p 064p 064p 065p 065p 065p 066p 066p
Procedure II
Prosthetic

Length Multi Multi Pick-up Multi Multi Multi Meta G Multi Plastic Multi Titanium Multi Polishing
8 / 10 / 12 / 14mm Protection Cap Impression Coping Transfer Post Lab Analog ULCA Cylinder UCLA Cylinder Cylinder Protector
Multi S Multi A

Diameter
Ø3.1 / 3.3

Abutment Level
Impression
068p
Procedure III
Prosthetic

Direct
Impression

Straight

054 / 055 COWELL® Implant System


INNO Submerged Pre-Mount > Packing unit : 1 Fixture + 1 Cover Screw + 1 Mount.

Diameter

Narrow Implant (Sub-N.) Length


Ø3.1
8 10 12 14

8 SR3108NS
L
10 SR3110NS
Submerged Fixture 12 SR3112NS
SUB-N. ୔ Note
HEXAGON Surface Treatment : SLA-SH® 14 SR3114NS D
> The INNO Sub. Narrow System is not compatible with
SYSTEM > Interchangeable with hexagonal morse tapered fixture. the INNO Submerged System as hexagon size is different.
> Internal hex connection (Taper 11°/ Hex 2.1).
Diameter
Ø3.3
Length 8 10 12 14

Hex 2.1 8 SR3308NS


10 SR3310NS L

12 SR3312NS
14 SR3314NS D

Fixture Mount
Length 5.4
L
RSM001
> Packing unit : 1 Mount + 1 Mount Screw.
L > Tightened with the 1.2 Hex Driver.
> Tightening torque force : 5~10 [Link].

Cover Screw
D Diameter
Ø2.85 Ø3.25 Ø3.6
D Length
L 1.7 RCS000
2.7 RCS001
No-Mount > Packing unit : 1 Fixture + 1 Cover Screw.
3.7 RCS002

Diameter > Packing unit : 1 Cover Screw.


Ø3.1
Length > To seal the conical interface of fixture.
> The longer the Cover Screw for deeply inserted fixture.
> Tightened with the 1.2 Hex Driver.
8 SR3108NSM 8 10 12 14
> Tightening torque force : 5~10 [Link].
L
10 SR3110NSM
12 SR3112NSM Healing Abutment Diameter Ø3.5 Ø4.5
14 SR3114NSM D Length
D Cuff 1 2 1 2
0.5 HR3501
L
C 1 HR3511 HS4511N
Diameter
Ø3.3 2 HR3522 HS4522N
Length
3 HR3532 HS4532N

8 SR3308NSM 8 10 12 14 4 HR3542 HS4542N

10 SR3310NSM L 5 HS4552N

12 SR3312NSM 7 HS4572N

14 SR3314NSM D
> Packing unit : 1 Healing Abutment.
> For remodeling gingival contour during soft tissue healing.
> Select according to gingival height and abutment type.
> Tightened with the 1.2 Hex Driver.
> Tightening torque force : 5~10 [Link].

056 / 057 COWELL® Implant System


Prosthesis Procedure I Cemented Abutment
1.2 Hex.
Type Hex N-Hex

Components Selection Guide for Cemented and UCLA Abutment D Diameter Ø4.5 Ø4.5
2.25 Length
L Cuff 4 5.5 7 4 5.5 7

C
1 SCH4514N SCH4515N SCH4517N SCN4514N SCN4515N SCN4517N
2 SCH4524N SCH4525N SCH4527N SCN4524N SCN4525N SCN4527N
3 SCH4534N SCH4535N SCH4537N SCN4534N SCN4535N SCN4537N
4 SCH4544N SCH4545N SCH4547N SCN4544N SCN4545N SCN4547N
2.25 Hex N-Hex 5 SCH4554N SCH4555N SCH4557N SCN4554N SCN4555N SCN4557N
> Packing unit : 1 Cemented Abutment + 1 Abutment Screw.
> For Screw-Cement or Cement Retained Prosthesis.
> Cutting surface for anti-rotation of prosthesis.
Abutment Screw > Gold color for more translucent restoration.
> Library available for EXOCAD®, 3Shape® & Others.
> Connected with the Abutment Screw (SSHR100N).
(15°) (25°) > Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 20~25 [Link].
> Use the Scanbody for 3D Work.
> Fixture level impression.
(Hex) (N-Hex) (Hex) (Hex) (N-Hex) (Hex) (N-Hex)

Cemented Angulated Temporary Meta G UCLA


Abutment Abutment Abutment Abutment Angulated Abutment Type Hex

A Diameter(Angle) Ø4.5(15°) Ø4.5(25°)


1.2 Hex. Length
Cuff 8 8
L
2.25
1 SAH45151NA SAH45251NA
2 SAH45152NA SAH45252NA
C
3 SAH45153NA SAH45253NA
D 4 SAH45154NA SAH45254NA

> Packing unit : 1 Angulated Abutment + 1 Abutment Screw.


> For Screw-Cement or Cement Retained Prosthesis.
Hex > Solution for anterior esthetic zone.
Replica
> Gold color for esthetics.
> Connected with the Abutment Screw (SSHR100N).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 20~25 [Link].
> Fixture level impression.

(Hex) (N-Hex) (Hex) (N-Hex)


Pick-up Impression Coping Transfer Post Temporary Abutment
D Type Hex N-Hex
1.2 Hex.
Diameter Ø4.5 Ø4.5
Length
Cuff 10 10
2.25
L 1 STHA45N STNA45N
Open or Closed Tray Impression C > Packing unit : 1 Temporary Abutment + 1 Abutment Screw.
> For Screw-Cement Retained Prosthesis.
> For provisional restoration.
> Connected with the Abutment Screw (SSHR100N).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
Hex N-Hex > Tightening torque force : 20~25 [Link].

Healing
Fixture Cover Screw Abutment

058 / 059 COWELL® Implant System


Meta G UCLA Abutment Type Hex N-Hex Pick-up Impression Coping Type Hex N-Hex
D
Diameter Ø4.5 Ø4.5 Diameter
D 1.2 Hex. Length / Cuff Ø4.5 Ø4.5
Length
Cuff 12 12 14 (Short) / 2 SIH45SN SIN45SN
2.25 L
1 SGH45N SGN45N SIH45LN SIN45LN
2 Pieces 16 (Long) / 4
L 2 SGH452N SGN452N
C
3
> Packing unit : 1 Pick-up Impression Coping + 1 Guide Pin.
SGH453N SGN453N
C > For open tray impression.
> Packing unit : 1 Meta G UCLA Abutment + 1 Abutment Screw. (Short) (Long) (Short) (Long)
> Connected with the Guide Pin (SIS001SN / SIS001LN).
> For Screw Retained Prosthesis. > Tightened with the 1.2 Hex Driver and Torque Wrench.
> Modification to angulated abutment, customized abutment and telescopic abutment. > Tightening torque force: 12~15Ncm.

Hex N-Hex
> CCM alloy core for precise connection.
> Cast with non-precious metal or gold alloy. Hex N-Hex
> Connected with the Abutment Screw (SSHR100N).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force: 20~25 [Link].
> Fixture level impression.

Abutment Screw Transfer Post


D
1.2 Hex.
Diameter
Height 2.25
L Type Hex N-Hex
D 10.2 SSHR100N
Diameter
Length / Cuff Ø4.5 Ø4.5
> Packing unit : 1 Abutment Screw. C
H > Tightened with the 1.2 Hex Driver and Torque Wrench. 9 (Short) / 2 STH45SN STN45SN
> Tightening torque force : 20~25 [Link]. 11 (Long) / 4 STH45LN STN45LN
(Short) (Long) (Short) (Long)
> Packing unit : Hex - 1 Transfer Post + 1 Guide Pin / N-Hex - 1 Transfer Post (Solid Type).
> For closed tray impression.
> Connected with the Guide Pin (STS001SN / STS001LN).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
Hex N-Hex > Tightening torque force: 12~15Ncm.

2 Pieces 1 Piece

Replica D
Diameter
Height Ø4
H 12.1 SRHR001N
> Packing unit : 1 Replica.
> Mimicking of conical interface of fixture.
> Analog of fixture for working cast.

060 / 061 COWELL® Implant System


Prosthesis Procedure II Multi S Abutment Diameter Ø4.5
Length 2
Component Selection Guide for Multi S&A Abutment D Cuff
1 SMS451N
L
C 2 SMS452N
1 Piece 3 SMS453N

4 SMS454N

Multi Polishing Protector > Packing unit : 1 Multi S Abutment.


> For Screw-Retained Prosthesis.
> Titanium base for the cylinders.
> Gold color for more translucent restoration.
> Integrated with screw and abutment.
Multi Cylinder Screw > Library available for EXOCAD®, 3Shape® & Others.
> Use the S Holder for more stable position.
> Tightened with the S Machine & S Ratchet Driver and Torque Wrench.
> Tightening torque force : 20~25 [Link].
> Abutment level impression.

(Hex) (N-Hex) (Hex) (N-Hex) (Hex) (N-Hex)

Multi Meta G Multi Plastic Multi Titanium


UCLA Cylinder UCLA Cylinder Cylinder Multi A Abutment Type Hex
Diameter(Angle) Ø4.5(15°) Ø4.5(30°)
1.2 Hex.
Length 2 2
D Cuff
L
A 1.95 2 SMAH45152N

C 3 SMAH45153N SMAH45303N
4 SMAH45154N SMAH45304N

Lab Analog Type N-Hex


Diameter(Angle) Ø4.5(15°) Ø4.5(30°)
Hex N-Hex 2 Pieces Length
Cuff
2 2
2 SMAN45152N
3 SMAN45153N SMAN45303N
4 SMAN45154N SMAN45304N
(Hex) (N-Hex) (Hex) (N-Hex)
Multi Pick-up > Packing unit : 1 Multi A Abutment + 1 Abutment Screw.
Impression Coping Multi Transfer Post
> For Screw-Retained Prosthesis.
> Titanium base for the cylinders.
> Gold color for more translucent restoration.
> Library available for EXOCAD®, 3Shape® & Others.
> Use the A Holder for more stable position.
> Connected with the Abutment Screw ( SSHR200N : ★ / SSHR300N : ● ).
Multi Protection Cap > Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 20~25 [Link].
> Abutment level impression.

(15°) (30°)

Abutment Screw
(Hex) (N-Hex) (Hex) (N-Hex) Height
1.2 Hex.
Multi S Abutment Multi A Abutment Diameter 8.7 9.3
Ø1.95 SSHR200N SSHR300N
D

> Packing unit: 1 Abutment Screw.


H > To connect the Multi A Abutment.
> Tightened with the 1.2 Hex Driver and Torque Wrench.

Healing
Fixture Cover Screw Abutment

062 / 063 COWELL® Implant System


Multi Protection Cap Multi Lab Analog Multi S & A
Multi S & A D Abutment Diameter Ø4.5
Abutment Diameter Ø4.5
D Diameter
Diameter L Length Ø4.5
Height Ø5.2
H 2 2SMA45
5 2SMPC45

> Packing unit : 1 Multi Protection Cap. > Packing unit : 1 Multi Lab Analog.
> Protection from cheek and tongue for gingival healing period. > Replacement of abutment shape in working cast.
> Gingival retraction for prosthodontic margin of abutment.
> Alternative usage for sub-structure of temporary prosthesis.

Multi Pick-up Impression Coping Multi Meta G UCLA Cylinder


D Type Hex N-Hex
Type Hex N-Hex
1.2 Hex. Multi S & A
Multi S & A Abutment Diameter Ø4.5 Ø4.5
Abutment Diameter Ø4.5 Ø4.5 L
Diameter Diameter Ø4.5 Ø4.5
Height Ø4.65 Ø4.65 2.25
H Length
Cuff 10.9 10.9
14.8 2SMIH45 2SMIN45
C 0.5 2SCCH45 2SCCN45
D
> Packing unit: 1 Multi Pick-up Impression Coping + 1 Guide Pin.
> For open tray impression. > Packing unit : 1 Multi Meta G UCLA Cylinder + 1 Multi Cylinder Screw.
> Connected with the Guide Pin (2SMGP012). > For Screw, Cement or Screw-Cement Retained Prosthesis.
> Tightened with the 1.2 Hex Driver and Torque Wrench. Hex N-Hex > Modification to various types of abutments.
Hex N-Hex > Tightening torque force: 12~15 [Link]. > CCM alloy core for precise connection.
> Cast with non-precious metal or gold alloy.
> Connected with the Multi Cylinder Screw (2SMCS100).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 20 [Link].

Multi Transfer Post


Multi Plastic UCLA Cylinder
Type Hex N-Hex
D Type Hex N-Hex
Multi S & A
Abutment Diameter Ø4.5 Ø4.5 Multi S & A
1.2 Hex. Abutment Diameter Ø4.5 Ø4.5
Diameter
H Height Ø4.5 Ø4.5
Diameter Ø4.5 Ø4.5
L
8.5 2SMTH45 2SMTN45 2.25 Length
Cuff 11.5 11.5

> Packing unit: 1 Multi Transfer Post + 1 Guide Pin. 0.5 2SMPH45 2SMPN45
C
> For closed tray impression.
Hex N-Hex
> Connected with the Guide Pin (2SMTHS100).
D > Packing unit : 1 Multi Plastic UCLA Cylinder + 1 Multi Cylinder Screw.
> Tightened with the 1.2 Hex Driver and Torque Wrench. > For Screw, Cement or Screw-Cement Retained Prosthesis.
> Tightening torque force: 12~15 [Link]. > Same purpose of use as the Meta G UCLA Cylinder but low accuracy of connection.
Hex N-Hex
> PMMA material.
> Connected with the Multi Cylinder Screw (2SMCS100).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 20 [Link].

064 / 065 COWELL® Implant System


Multi Titanium Cylinder Multi Holder
Type Hex N-Hex
1.2 Hex.
Multi S & A
Abutment Diameter Ø4.5 Ø4.5 H
L Diameter Ø4.5 Ø4.5 Type
2.25 Height Hand
Length
Cuff 8.5 8.5 20 KMHS01
C
0.5 2STCH45 2STCN45 S Holder > Packing unit: 1 Multi S Holder.
D
> To position the Multi S Abutment more stably.
> Packing unit : 1 Multi Titanium Cylinder + 1 Multi Cylinder Screw.
> For Screw, Cement or Screw-Cement Retained Prosthesis.
Hex N-Hex > Connected with the Multi Cylinder Screw (2SMCS100).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
H
> Tightening torque force : 20 [Link].
Type
Height Hand
32 KMHA01
A Holder
> Packing unit: 1 Multi A Holder.
> To position the Multi A Abutment more stably.

Multi Cylinder Screw


Diameter
Height Ø2.25
1.2 Hex.
5 2SMCS100

D > Packing unit : 1 Multi Cylinder Screw.


> Connected with the Meta G UCLA, Plastic UCLA and Titanium Cylinder.
H > Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 20 [Link]. ˝ ˞ ˟ ˠ
Connect the Multi A Holder with the Hold the handle of the Multi A Connect the assembled body Tighten the Multi Abutment
Multi A Abutment with its Abutment Holder and bend it according with the fixture. with the 1.2 Hex Driver and
Screw in it and match the direction of to the placement position in Torque Wrench.
holes of the abutment and the holder. the oral cavity.

Multi S Machine Driver Type


Height Machine
H 27.5 KMMSD21L

> Packing unit: 1 Multi S Machine Driver.


> To install the Multi S Abutment by machine.

Multi Polishing Protector Type Hex


Multi S & A
D Abutment Diameter Ø4.5
Diameter
L Length Ø4.5
2 2SMPP45 Multi S Ratchet Driver
Type
Height Ratchet
> Packing unit : 1 Multi Polishing Protector. H
> For polishing work during lab procedure. 22 KRMSD15L

> Packing unit: 1 Multi S Ratchet Driver.


> To install the Multi S Abutment by hand.

066 / 067 COWELL® Implant System


Prosthesis Procedure III
Component Selection Guide for Straight Abutment

Straight Abutment

Healing
Fixture Cover Screw Abutment

Straight Abutment
D
Diameter Ø3.5
7 5.5 L Length [Cuff] 8 [0.5] 8 [1] 8 [2] 8 [3] 8 [4]
4
SR308 SR318 SR328 SR338 SR348
C
> Packing unit : 1 Straight Abutment.
> For Cement Retained Prosthesis.
> Integrated with screw and abutment.
> Tightened with the Shoulder Driver.
> Tightening torque force : 20~25 [Link].
> Direct impression.

Shoulder Driver
Diameter
Height Ø4.5
H
19 KRR19L
D > Packing unit : 1 Shoulder Driver.
> To install the Straight Abutment.

068 / 069 COWELL® Implant System


INNO INTERNAL IMPLANT (Int.)
System Flow

Fixture Abutment Impression

Fixture Level
077p 077p 077p 078p 078p 078p
Procedure I

Impression
Prosthetic

Cemented Angulated Meta G UCLA Replica Pick-up Transfer Post


Impression Coping

080p 080p 080p 081p 081p 081p


Procedure II
Prosthetic

Solid Solid Solid Solid Positioning Solid Solid


Protection Cap Impression Cap Cylinder Lab Analog Plastic Coping
Octa 3.1

Platform
Ø4.8 / 5.9

Cuff
083p 083p 083p 083p 083p
1.8 / 2.4mm
Procedure III
Prosthetic

Length Shoulder
7 / 8 / 10 / 12 / 14mm Shoulder Shoulder Shoulder
Positioning

Abutment Level
Shoulder Abutment Cap Impression Cap Lab Analog
Cylinder

Impression
Diameter
Ø3.5 / 4.0 / 4.5 / 5.0 / 6.0
085p 086p 086p
Procedure IV
Prosthetic

Sonator
S Abutment Impression Coping Sonator Analog

089p 089p
Procedure V
Prosthetic

Ball Analog
Ball

070 / 071 COWELL® Implant System


INNO Internal Impant (Int.) No-Mount Cuff 1.8mm fixture > Packing unit : 1 Fixture + 1 Cover Screw.

* Diameter
Ø3.5
Length
Internal Fixture * Platform : Ø4.8
INT. Surface Treatment : SLA-SH® 7 - C
OCTAGON Platform 8 10 12 14
> Interchangeable with 1 staged internal fixture.
SYSTEM 8 IPT3508SM
> Internal Octa Connection (Taper 8°/ Octa 3.1). L
> No-Mount type. 10 IPT3510SM Fixture
Diameter

12 IPT3512SM
14 IPT3514SM D
Octa 3.1
* Diameter
Ø4.0
Length
Platform * Platform : Ø4.8
7 IPT4007SM
C 7 8 10 12 14
8 IPT4008SM
10 IPT4010SM L
C 12 IPT4012SM
14 IPT4014SM
D

* Diameter
Ø4.5
Length

L 7 IPT4507SM
* Platform : Ø4.8
C 7 8 10 12 14
8 IPT4508SM
L
10 IPT4510SM
12 IPT4512SM
14 IPT4514SM D

* Diameter
Ø5.0
D Length

* Platform : Ø5.9
7 IPT5007SM C 7 8 10 12 14
8 IPT5008SM
INNO Fixture Code L
10 IPT5010SM

I P T 40 10 S M Ex.)
SLA Cuff 1.8 No-Mount IPT4010SM
12 IPT5012SM
Type P=Cuff 1.8 body Diameter Length Surface Treatment Mount 14 IPT5014SM D
Internal Taper Ø4.0 10mm SLA No-Mount

* Diameter
Ø6.0
I T 40 10 S M Ex.)
SLA Cuff 2.4 No-Mount IT4010SM
Length
* Platform : Ø5.9
Type Cuff 2.4 body Diameter Length Surface Treatment Mount C 7 8 10 12
7 IPT6007SM
Internal Taper Ø4.0 10mm SLA No-Mount
8 IPT6008SM L
10 IPT6010SM
12 IPT6012SM
D
14 -

072 / 073 COWELL® Implant System


No-Mount Cuff 2.4mm fixture > Packing unit : 1 Fixture + 1 Cover Screw. Cover Screw
D
Platform [Fixture Dia.] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.9 [Ø5.0 / Ø6.0]
Diameter
Height Ø5.0 Ø6.0
H
* Diameter
Ø3.5 6.5 ICVR002 ICVW002
Length
* Platform : Ø4.8 > Packing unit : 1 Cover Screw.
> To seal the conical interface of fixture.
7 - C
Platform 8 10 12 14 > Tightened with the 1.2 Hex Driver.
8 IT3508SM > Tightening torque force : 5~10 [Link].
L
10 IT3510SM Fixture
Diameter

12 IT3512SM
14 IT3514SM D

* Diameter
Ø4.0
Length Headless Screw Diameter
Height Ø3.5
D
* Platform : Ø4.8 6 ICVR001
7 IT4007SM
C 7 8 10 12 14
8 IT4008SM H > Packing unit : 1 Headless Screw.
> For narrow mesio-distal distance.
10 IT4010SM L
> Tightened with the 1.2 Hex Driver.
12 IT4012SM > Tightening torque force : 5~10 [Link].
14 IT4014SM
D

* Diameter
Ø4.5
Length

* Platform : Ø4.8
7 IT4507SM C 7 8 10 12 14
8 IT4508SM
Healing Abutment
L
Platform [Fixture Dia.] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.9 [Ø5.0 / Ø6.0]
10 IT4510SM Diameter
D Length Ø5.5 Ø6.6
12 IT4512SM L 2 IHCR020 IHCW020
14 IT4514SM D
3 IHCR030 IHCW030
4.5 IHCR045 IHCW045
* Diameter
Ø5.0 > Packing unit : 1 Healing Abutment.
Length > For remodeling gingival contour during soft tissue healing.
* Platform : Ø5.9 > Select according to gingival height and abutment type.
7 IT5007SM > Tightened with the 1.2 Hex Driver.
C 7 8 10 12 14 > Tightening torque force : 5~10 [Link].
8 IT5008SM
L
10 IT5010SM
12 IT5012SM
14 IT5014SM D

* Diameter
Ø6.0
Length
* Platform : Ø5.9
C 7 8 10 12
7 IT6007SM
8 IT6008SM L
10 IT6010SM
12 IT6012SM
D
14 -

074 / 075 COWELL® Implant System


Cemented Extension Abutment
Prosthetic Procedure I D
1.2 Hex.

2.5
Type
Platform [Fixture Dia.] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5]
Octa
Ø5.9 [Ø5.0 / Ø6.0]
L
Component Selection Guide for Cemented & UCLA Abutment Diameter Ø4.8 Ø5.8 Ø5.9 Ø6.9
C Length
Cuff 6 6 6 6
0.5 IECR406 IECW506
1 IECR416 IECW516
Octa N-Octa 2 IECR426 IECW526
2.5
2.5 3 IECR436 IECW536
2.5

Type N-Octa
Platform [Fixture Dia.] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.9 [Ø5.0 / Ø6.0]
Coping Screw
Diameter Ø4.8 Ø5.8 Ø5.9 Ø6.9
Length
Cuff 6 6 6 6
0.5 IENR406 IENW506

(15°) (25°)
1 IENR416 IENW516
2 IENR426 IENW526
3 IENR436 IENW536

> Packing unit : 1 Cemented Extension Abutment + 1 Abutment Screw.


> For Cement Retained or Screw-Cement Retained Prosthesis.
(Octa) (N-Octa) (Octa) (Octa) (N-Octa)
> Cutting surface for anti-rotation of prosthesis.
Cemented Extension Angulated Abutment Meta G UCLA Abutment > Connected with the Abutment Screw (ISHR110).
Abutment
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 30 [Link].
> Fixture level impression.

Angulated Abutment
A Type Octa
1.2 Hex. Platform [Fixture Dia.] Ø4.8 & Ø5.9 [Ø3.5 / Ø4.0 / Ø4.5 / Ø5.0 / Ø6.0]
Replica Diameter(Angle)
L 3.8 (15°) 3.8 (25°)
2.5 Length
8 IAAR158A IAAR258A

D > Packing unit : 1 Angulated Abutment + 1 Abutment Screw.


> For Cement Retained or Screw-Cement Retained Prosthesis.
> Solution for anterior esthetic zone.
> Connected with the Abutment Screw (ISHR100).
Octa
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 30 [Link].
> Fixture level impression.
(Octa) (Octa) (N-Octa)

Pick-up Impression Coping Transfer Post & Screw

Meta G UCLA Abutment


Type Octa N-Octa
D
1.2 Hex. Platform [Fixture Dia.] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.9 [Ø5.0 / Ø6.0] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.9 [Ø5.0 / Ø6.0]
Diameter
Length Ø5 Ø6 Ø5 Ø6
2.5
L
12 IGOR400N IGOW500N IGNR400N IGNW500N

> Packing unit : 1 Meta G UCLA Abutment + 1 Abutment Screw.


Healing > For Screw-Cement or Screw Retained Prosthesis.
Fixture Cover Screw Headless Screw Abutment > Modification to angulated abutment, customized abutment and telescopic abutment.
Octa N-Octa
> CCM alloy core for precise connection.
> Cast with non-precious metal or gold alloy.
> Connected with the Abutment Screw (ISHR120).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 30 [Link].
> Fixture level impression.

076 / 077 COWELL® Implant System


Abutment Screw Height
Diameter Ø2.5 Ø2.5 Ø2.5
1.2 Hex. 1.2 Hex. 1.2 Hex.
6.3
7.8
ISHR100
ISHR120
Prosthetic Procedure II
D D D
9.2 ISHR110 Component Selection Guide for Solid Abutment
H
> Packing unit : 1 Abutment Screw.
> ISHR110 : Cemented Abutment.
> ISHR100 : Angulated Abutment.
> ISHR120 : Meta G Abutment.
> Tightened with the 1.2 Hex Driver and Torque Wrench.

(Crown) (Bridge)

Solid Plastic Coping

Replica
D Platform [Fixture Dia.] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.9 [Ø5.0 / Ø6.0]
Diameter
Height Ø4.8 Ø5.9
H 12 IROR001 IROW001
> Packing unit : 1 Replica.
> Mimicking of conical interface of fixture.
> Analog of fixture for working cast.
Solid Lab Analog

Pick-up Impression Coping


D
Type Octa
Solid Impression Cap Solid Positioning Cylinder
Platform [Fixture Dia.] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.9 [Ø5.0 / Ø6.0]
L Diameter
Length Ø5.5 Ø6.6
2 Pieces
13.7 IIOR001 IIOW001

> Packing unit : 1 Pick-up Impression Coping + 1 Guide Pin.


> For open tray impression.
> Connected with the Guide Pin (IIOR001S).
> Tightened with the 1.2 Hex Driver and Torque Wrench.
Octa > Tightening torque force: 12~15 [Link]. Solid Protection Cap

Transfer Post
D

Type Octa N-Octa


L Solid Abutment
Platform [Fixture Dia.] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.9 [Ø5.0 / Ø6.0] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.9 [Ø5.0 / Ø6.0]
Diameter
Length Ø4.85 Ø5.95 Ø4.85 Ø5.95
11.6 ITOR400 ITOW500 ITNR400 ITNW500
2 Pieces 1 Piece
> Packing unit : Octa - 1 Transfer Post + 1 Guide Pin / N-Octa - 1 Transfer Post (Solid Type).
> For closed tray impression.
> Connected with the Guide Pin (Regular: ITOR400S / Wide: ITOW500S).
Octa N-Octa > Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force: 12~15 [Link].

Healing
Fixture Cover Screw Headless Screw Abutment

078 / 079 COWELL® Implant System


Solid Abutment Solid Positioning Cylinder
D Platform [Fixture Dia.] Ø4.8 & Ø5.9 [Ø3.5 / Ø4.0 / Ø4.5 / Ø5.0 / Ø6.0] Solid Abutment Ø3.5
D Diameter
Diameter Diameter
L Ø3.5 Height Ø5.7
Length 3 4 5.5 7 10.2 IPCR001
H
IASR030 IASR040 IASR055 IASR070 > Packing unit : 1 Solid Positioning Cylinder.
> Packing unit : 1 Solid Abutment + 1 Abutment Cap. > Inner cutting surface for anti-rotation on abutment.
> For Cement Retained Prosthesis. > Insert into the Impression Cap.
> Cutting surface for anti-rotation of prosthesis.
> Integrated with screw and abutment.
> Tightened with the Shoulder Driver.
> Tightening torque force : 30 [Link].
> Abutment level impression :
Impression cap in platform Ø4.1 fixture and direct impression in platform Ø5.8 fixture.

4IPVMEFSⰝ KRR19L

Solid Protection Cap Solid Abutment


Diameter
Ø3.5 Solid Lab Analog
Solid Abutment
Diameter D Diameter Ø3.5
Height Ø5.4
D L Diameter Ø4.8
5.2 IASR130
Length 3 4 5.5 7
H 6.2 IASR140
ILSR030 ILSR040 ILSR055 ILSR070
7.7 IASR155
9.2 IASR170 > Packing unit : 1 Solid Lab Analog.
> Replacement of abutment shape in working cast.
> Packing unit : 1 Solid Protection Cap. > Choose by abutment length.
> Protection from cheek and tongue for gingival healing period.
> Alternative usage for sub-structure of temporary prosthesis.

Solid Impression Cap Solid Abutment Ø3.5


Solid Plastic Coping
Diameter
D Diameter
Height 8 Type Crown Bridge
8 IICR001 Solid Abutment
H H Diameter Ø3.5 Ø3.5
Diameter
> Packing unit : 1 Solid Impression Cap. Height Ø5 Ø5
> Connected with the Solid Positioning Cylinder. 10 IPCB001
D IPCC001
> Confirm locking with abutment by rotation of clockwise and anti-clockwise direction.
> Packing unit : 1 Solid Plastic Coping.
> Connect with the Lab Analog.
Crown Bridge > Burn out and casting for metal framework.

080 / 081 COWELL® Implant System


Shoulder Abutment Type Octa N-Octa
D
Platform [Fixture Dia.] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.9 [Ø5.0 / Ø6.0] Ø4.8 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.9 [Ø5.0 / Ø6.0]

Prosthetic Procedure III 2 Pieces


L Platform Length
Diameter

4
Ø3.5
ISAC404
Ø4.5
ISAC504
Ø3.5
ISAB404
Ø4.5
ISAB504
Component Selection Guide for Shoulder Abutment Fixture 5.5 ISAC405 ISAC505 ISAB405 ISAB505
Diameter
7 ISAC407 ISAC507 ISAB407 ISAB507

> Packing unit : 1 Shoulder Abutment + 1 Abutment Cap.


> For Cement Retained Prosthesis.
> Dual anti-rotation grip with single crown for prevention of screw loosening.
Octa N-Octa > Integrated with the Screw and Abutment.
> Tightened with the Shoulder Driver.
> Tightening torque force : 30 [Link].
> Abutment level impression.

(Octa) (N-Octa)
4IPVMEFSⰝ KRR19L 4IPVMEFSⰝ KRW19L
Shoulder Lab Analog

Shoulder Abutment Cap Shoulder Abutment


Diameter Ø3.5 Ø4.5
D Diameter
Height Ø5.4 Ø5.4
6.2 IASR140 IASW140
H
Shoulder Impression Cap Shoulder Positioning Cylinder 7.7 IASR155 IASW155
9.2 IASR170 IASW170
> Packing unit : 1 Shoulder Abutment Cap.
> Protection from cheek and tongue for gingival healing period.
> Alternative usage for sub-structure of temporary prosthesis.

Shoulder Impression Cap


Shoulder Abutment Cap Shoulder Abutment Ø3.5 Ø4.5
D Diameter
Diameter
Height 8 9
H 8 IICR001 IICW001

> Packing unit : 1 Shoulder Impression Cap.


> Connected with the Shoulder Positioning Cylinder.
> Confirm locking with abutment by rotation of clockwise and anti-clockwise direction.

Shoulder Positioning Cylinder Shoulder Abutment


Ø3.5 Ø4.5
(Octa) (N-Octa) D Diameter
Diameter
Shoulder Abutment Height 5.7 6.8
10.7 SAPR001 SAPW001
H
> Packing unit : 1 Shoulder Positioning Cylinder.
> Inner cutting surface for anti-rotation on abutment.
> Insert into the Impression Cap.

Shoulder Lab Analog


D Type Octa N-Octa
Shoulder Abutment Ø3.5 Ø4.5 Ø3.5 Ø4.5
L Diameter
Healing Diameter
Fixture Cover Screw Headless Screw Abutment Length Ø4.8 Ø5.9 Ø4.8 Ø5.9
4 SLCR040 SLCW040 SLBR040 SLBW040
5.5 SLCR055 SLCW055 SLBR055 SLBW055
7 SLCR070 SLCW070 SLBR070 SLBW070
> Packing unit : 1 Shoulder Lab Analog.
> Replacement of abutment shape in working cast.
Octa N-Octa > Choose according to width and length of abutment.

082 / 083 COWELL® Implant System


Prosthetic Procedure IV Sonator S Abutment
Black
Block-out Processing
Component Selection Guide for Sonator Abutment Outer Cap Spacer Male H-Type Inner Caps

2.25mm

Ø5.4
Black Blue : 10N Pink : 15N White : 22N
Block-out Processing
Outer Cap Spacer Male
Carrier
Diameter Ø4.0
Cuff 0.6 2 3 4
Length
1.5 IONS401 IONS402 IONS403 IONS404
> Packing unit : 1 Sonator S Abutment + 1 Carrier + 3 H-Type Inner Caps + 1 Outer Cap + 1 Block-out Spacer
+ 1 Black Processing Male.
> For Implant Supported Over-denture Prosthesis.
> Stable with low vertical height.
> 6 kinds of the Inner Caps give various holding force (Both, H and NH-Type Inner Caps are used for the Sonator S Abutment).
H-Type Inner Caps NH-Type Inner Caps > Path compensation up to 20° based on 2 implants.
Ratchet Driver
D > Carrier : Used for delivery of the abutment.
L > Tightened with the Ratchet Driver and Torque wrench.
C
> Tightening torque force : 30 [Link].
> Abutment level impression.

Sonator Lab Analog


Outer Cap
Diameter
D
Black Processing Male Height Ø5.4
2.25 SONOC01
H
> Packing unit : 2 Outer Caps and 2 Black Processing Males.
> Black Processing Male : Inserted and Removed with the I&R Driver.

Sonator Impression Coping

H-Type Inner Cap Code SONIC01


> Packing unit : 3 Block-out Spacers + 3 Inner Caps (1 Blue, 1 Pink and 1 White).
> Path compensation up to 20° based on 2 implants.
Sonator S Abutment Block-out
Spacer Inner Cap > Mainly used for the Sonator S Abutment.
> Inner Caps : Inserted and Removed with the I&R Driver.

Blue : Pink : White :


Retention Force Retention Force Retention Force
About 10N About 15N About 22N

Healing
Fixture Cover Screw Headless Screw Abutment

084 / 085 COWELL® Implant System


NH-Type Inner Cap Code SONIC02 Sonator S Ratchet Driver
H Type
> Packing unit : 3 Block-out Spacers + 3 Inner Caps (1 Red, 1 Orange and 1 Green). Height Ratchet
Block-out > Path compensation up to 40° based on 2 implants. 18 SONRD19L
Spacer Inner Cap > Mainly used for the Sonator A Abutment.
> Inner Caps : Inserted and Removed with the I&R Driver. > Used to tighten and untighten the Sonator S Abutment.

Red : Orange : Green :


Retention Force Retention Force Retention Force
About 10N About 15N About 22N

Sonator Impression Coping I & R Driver


Diameter Height 95.4
Length Ø4.8
D
Black Processing Male SONIR002
3 SONIP04
H
L
> Packing unit : 4 Impression Copings and 4 Black Processing Males.
> Abutment level pick-up impression.
> Connected over the Sonator Abutment after placing the Block-out Spacer. For Removal For Insertion
> For close tray impression.
> Used to insert and remove the Inner Caps and Block Processing Male.

Sonator Lab Analog


Diameter
D Length Ø4
L
1.4 SONLA04

> Packing unit : 4 Sonator Lab Analogs.


> Replacement of abutment shape in working cast.

086 / 087 COWELL® Implant System


Ball Abutment
Outer Cap Block-out Spacer

Prosthetic Procedure V Ø3.4


2.4mm Length
4
Diameter Ø3.5
IBAT404R
Component Selection Guide for Ball Abutment
> Packing unit : 1 Ball Abutment + 3 Inner Caps (1 per each color) + 1 Block-out Spacer + 1 Outer Cap.
> For Implant Supported Over-denture Prosthesis.
Red : 10N Orange : 15N Green : 20N
> Tightened with the Ball Driver and Torque Wrench.
> Tightening torque force : 30 [Link].
Ball Inner Cap > Direct impression.
D

Ball Outer Cap Block-out Spacer

Ball Abutment

Ball Outer Cap Diameter


D Height Ø3.4
Red : 10N Orange : 15N Green : 20N 2.4 BATC003C
Ball Inner Cap H
> Packing unit : 2 Outer Caps.

Ball Inner Cap


Block-out Ball Inner Cap
Spacer
BATC003I
> Packing unit : 2 Block-out Spacers + 6 Inner Caps (2 per each color).
Red : 10N Orange : 15N Green : 20N
> Retention force : Red 10N, Orange 15N & Green 20N.
Ball Analog

Ball Analog
D
Diameter
Length Ø4.0
L
4 SBAL400

> Packing unit : 4 Lab Analogs.


> Replacement of abutment shape in working cast.

Ball Abutment

Type
Ball Driver Height Ratchet
H 19 KRB19L
> Packing unit : 1 Ball Driver. *Extra Product
> Used with the Torque Wrench to tighten and untighten the Ball Abutment.

Healing
Fixture Cover Screw Headless Screw Abutment

I&R Driver Height 100


KBIR01

For Removal For Insertion

> Packing unit : 1 I&R Driver.


> Used to insert and remove the Inner Caps into and out of the Outer Cap.

088 / 089 COWELL® Implant System


INNO EXTERNAL IMPLANT (Ext.)
System Flow

Fixture Abutment Impression

096p 096p

098p 098p 098p


Cemented Angulated
Procedure I

Fixture Level
Prosthetic

Impression
096p 097p 097p Replica Pick-up Squared Transfer Post
Impression Coping

Temporary Meta G UCLA Plastic UCLA

Hex : 2.7 / 3.4

Platform
Ø4.1 / 5.1

100p 100p 101p 101p 101p


Length
Procedure II
Prosthetic

7 / 8 / 10 / 12 / 14mm

Shoulder Shoulder Shoulder Positioning Shoulder


Shoulder Abutment Cap Impression Cap Cylinder Lab Analog
Diameter
Ø3.5 / 4.0 / 4.5 / 5.0 / 6.0

Abutment Level
Impression
103p 103p
Procedure III
Prosthetic

Ball Ball Analog

090 / 091 COWELL® Implant System


Pre-Mount
INNO External Implant (Ext.)
> Packing unit :1 Fixture + 1 Mount + 1 Mount Screw.

* Diameter
Ø3.5 * Hex 2.7
Length * Platform : Ø4.1

External Fixture
EXT.
Surface Treatment : SLA-SH® 7 -
HEXAGON
SYSTEM > Interchangeable with external hexagonal fixture. 8 ET3508S
>External hex connection (Hex 2.7 / 3.4). 10 ET3510S
12 ET3512S
14 ET3514S

Hex 2.7 / 3.4


* Diameter
Ø4.0 * Hex 2.7
Length * Platform : Ø4.1

Platform 7
8
ET4007S
ET4008S
10 ET4010S
12 ET4012S
14 ET4014S

* Diameter
Ø4.5 * Hex 2.7
Length * Platform : Ø4.1

L 7 ET4507S
8 ET4508S
10 ET4510S
12 ET4512S
14 ET4514S

* Diameter
Ø5.0 * Hex 3.4
Length * Platform : Ø5.1
D
7 ET5007S
8 ET5008S
INNO Fixture Code 10 ET5010S
12 ET5012S
E T 40 10 S Ex.) 14 ET5014S
Type body Diameter Length Surface Treatment Mount SLA Pre-Mount ET4010S
External Taper Ø4.0 10mm SLA Pre-Mount
* Hex 3.4
* Diameter * Platform : Ø5.1
Ø6.0
Length

7 ET6007S
8 ET6008S
10 ET6010S
12 ET6012S
14 -

092 / 093 COWELL® Implant System


Fixture Mount
D

Hex Hex2.7 Hex3.4 Prosthetic Procedure I


L Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0]
Diameter
Component Selection Guide for Cemented & UCLA Abutment
Length Ø4.9 Ø5.5
7.2 MER001 MEW002

> Packing unit : 1 Mount + 1 Mount Screw.


> Tightened with the 1.2 Hex Driver.
> Tightening torque force : 5~10 [Link].

Abutment Screw

(15°) (25°)

Cover Screw (Hex)


(H
(Hex
(He
Hex
ex)

(Hex) (N-Hex) (Hex) (Hex) (N-Hex) (Hex) (N-Hex) (Hex) (N-Hex)

D Hex Hex2.7 Hex3.4 Cemented Angulated Abutment Temporary Abutment Meta G UCLA Abutment Plastic UCLA Abutment
1.1mm Abutment
Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0]
H
Diameter
Height Ø4.3 Ø5.4
5.8 VNR001 VNW001

> Packing unit : 1 Cover Screw.


> To seal the conical interface of fixture.
> Tightened with the 1.2 Hex Driver.
> Tightening torque force : 5~10 [Link].

Replica

Healing Abutment Hex Hex2.7 Hex3.4


(Hex) (N-Hex) (Hex) (N-Hex)
Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0]
Pick-up Squared
D Diameter Impression Coping Transfer Post
Length Ø5.0 Ø6.0
L
2.8 HNR502 HNW602
3.8 HNR503 HNW603
4.8 HNR504 HNW604
5.8 HNR505 HNW605
6.8 HNR506 HNW606
7.8 HNR507 HNW607

> Packing unit : 1 Healing Abutment.


> For remodeling gingival contour during soft tissue healing. Healing
> Select according to gingival height and abutment type. Fixture Cover Screw Abutment
> Tightened with the 1.2 Hex Driver.
> Tightening torque force : 5~10 [Link].

094 / 095 COWELL® Implant System


Cemented Abutment Type[Hex] Hex[2.7] Hex[3.4]
Meta G UCLA Abutment
Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0] 1.2 Hex.
1.2 Hex. Type[Hex] Hex[2.7] Hex[3.4] N-Hex N-Hex
D Diameter Ø5.0 Ø6.0
Length
Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0]
Cuff 6 8 6 8 L
Diameter Ø4.5 Ø5.5 Ø4.5 Ø5.5
L
1 CHR516 CHR518 CHW616 CHW618 Length
Cuff 13 13 13 13
C
2 CHR526 CHR528 CHW626 CHW628 C
D 1.2 GHR001N GHW001N GNR001N GNW001N
3 CHR536 CHR538 CHW636 CHW638
> Packing unit : 1 Meta G UCLA Abutment + 1 Abutment Screw.
4 CHR546 CHR548 CHW646 CHW648 > For Screw-Cement or Screw Retained Prosthesis.
Hex N-Hex Hex N-Hex > Modification to angulated abutment, customized abutment and telescopic abutment.
Type[Hex] N-Hex > CCM alloy core for precise connection.
Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0] > Cast with non-precious metal or gold alloy.
> Connected with the Abutment Screw.
Diameter Ø5.0 Ø6.0 > Tightened with the 1.2 Hex Driver and Torque Wrench.
Hex Length > Tightening torque force : 30 [Link].
Cuff 6 8 6 8
Platform > Fixture level impression.
1 CNR516 CNR518 CNW616 CNW618
2 CNR526 CNR528 CNW626 CNW628
3 CNR536 CNR538 CNW636 CNW638
4 CNR546 CNR548 CNW646 CNW648

> Packing unit : 1 Cemented Abutment + 1 Abutment Screw. > Tightened with the 1.2 Hex Driver and Torque Wrench.
> For Cement Retained and Screw-Cement Retained Prosthesis. > Tightening torque force : 30 [Link]
> Cutting surface for anti-rotation of prosthesis. > Fixture level impression.
> Connected with the Abutment Screw.

Angulated Abutment Plastic UCLA Abutment


Type[Hex] Hex[2.7] Hex[3.4] Hex[2.7] Hex[3.4] Type[Hex] Hex[2.7] Hex[3.4] N-Hex N-Hex
A 1.2 Hex. Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0] 1.2 Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0]
Diameter (Angle) Ø5 (15°) Ø6 (15°) Ø5 (25°) Ø6 (25°) Diameter Ø4.5 Ø5.5 Ø4.5 Ø5.5
Length L Length
L Cuff 8 8 8 8 Cuff 11.8 11.8 11.8 11.8
2 AAR152 AAW152 AAR252 AAW252 1.2 PHR001 PHW001 PNR001 PNW001
C
3 AAR153 AAW153 AAR253 AAW253 C > Packing unit : 1 Plastic UCLA Abutment + 1 Abutment Screw.
D D > Same purpose of use as Meta G UCLA Abutment but low accuracy of connection.
4 AAR154 AAW154 AAR254 AAW254
> PMMA material.
> Packing unit : 1 Angulated Abutment + 1 Abutment Screw. > Connected with the Abutment Screw.
> For Screw-Cement or Cement Retained Prosthesis. > Tightened with the 1.2 Hex Driver and Torque Wrench.
Hex N-Hex
Hex > Solution for anterior esthetic zone. > Tightening torque force : Finger light force during wax Pattern fabrication, 30 [Link] after casting.
> Connected with the Abutment Screw.
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 30 [Link].
> Fixture level impression.

Temporary Abutment
1.2 Hex.
Type[Hex] Hex[2.7] Hex[3.4] N-Hex N-Hex Abutment Screw
Type[Hex] Hex[2.7] Hex[3.4]
L Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0] 1.2 Hex. 1.2 Hex.
Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0]
Diameter Ø5.4 Ø5.95 Ø5.4 Ø5.95 Diameter
Length D D Height Ø2.5 Ø3.0
Cuff 12 12 12 12
C 8 SHR100 SHW100
D
1.5 THR001 THW001 TNR001 TNW001 H
> Packing unit : 1 Abutment Screw.
> Packing unit : 1 Temporary Abutment + 1 Abutment Screw. > Tightened with the 1.2 Hex Driver and Torque Wrench.
> For Screw-Cement Retained Prosthesis.
Hex N-Hex > For provisional restoration.
> Connected with the Abutment Screw.
> Tightened with the 1.2 Hex Driver and Torque Wrench.
> Tightening torque force : 20 [Link].

096 / 097 COWELL® Implant System


Replica
D Type[Hex]
Platform [Fixture Dia.]
Hex[2.7]
Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5]
Hex[3.4]
Ø5.1 [Ø5.0 / Ø6.0] Prosthetic Procedure II
Diameter
H Height Ø4.1 Ø5.1 Component Selection Guide for Shoulder Abutment
12 LHR001 LHW001
> Packing unit : 1 Replica.
> Mimicking of conical interface of fixture.
> Analog of fixture for working cast.

(Crown) (Bridge)

Shoulder Lab Analog

Pick-up Squared Impression Coping


D
Type[Hex] Hex[2.7] Hex[3.4] N-Hex N-Hex
Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0]
L Diameter
Length Ø5 Ø5.8 Ø5 Ø5.8
Shoulder Impression Cap Shoulder Positioning Cylinder
17 IHR500 IHW600 INR500 INW600
2 Pieces
> Packing unit : 1 Pick-up Squared Impression Coping + 1 Guide Pin.
> Connected with the Guide Pin (Regular : UHR115 / Wide : UHW115).
> For Open Tray Impression.
> Tightened with the 1.2 Hex Driver and Torque Wrench.
Hex N-Hex > Tightening torque force: 12~15 [Link].

Shoulder Abutment Cap

Transfer Post (Crown) (Bridge)


D
Type[Hex] Hex[2.7] Hex[3.4] N-Hex N-Hex Shoulder Abutment
Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0]
L
Diameter
Length Ø4.8 Ø5.8 Ø4.8 Ø5.8
2 Pieces
13.1 IHR510 IHW610 INR510 INW610
> Packing unit : 1 Transfer Post + 1 Guide Pin.
> Connected with the Guide Pin (Regular : IHR510S, IHR610S / Wide : IHW610S).
> For Closed Tray Impression.
> Tightened with the 1.2 Hex Driver and Torque Wrench.
Hex N-Hex
> Tightening torque force: 12~15 [Link].
Healing
Fixture Cover Screw Abutment

098 / 099 COWELL® Implant System


Shoulder Abutment Shoulder Impression Cap
Type[Hex] Hex[2.7] Hex[3.4] Shoulder Abutment Ø4.8 Ø5.9
D Diameter
Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0] Diameter 8 9
Height
2 Pieces L Diameter Ø4.8 Ø5.9 H 8 IICR001 IICW001
Length
Cuff 4 5.5 7 4 5.5 7
C
D > Packing unit : 1 Shoulder Impression Cap.
1 SAC414 SAC415 SAC417 SAC514 SAC515 SAC517 > Connected with the Shoulder Positioning Cylinder.
2 SAC424 SAC425 SAC427 SAC524 SAC525 SAC527 > Confirm locking with abutment by rotation of clockwise and anti-clockwise direction.
3 SAC434 SAC435 SAC437 SAC534 SAC535 SAC537
4 SAC444 SAC445 SAC447 SAC544 SAC545 SAC547
Hex Hex
Platform
Type[Hex] N-Hex N-Hex
Platform [Fixture Dia.] Ø4.1 [Ø3.5 / Ø4.0 / Ø4.5] Ø5.1 [Ø5.0 / Ø6.0]
Diameter Ø4.8 Ø5.9
D Length
Cuff 4 5.5 7 4 5.5 7
L 1 SAB414 SAB415 SAB417 SAB514 SAB515 SAB517
2 Pieces
2 SAB424 SAB425 SAB427 SAB524 SAB525 SAB527
C
3 SAB434 SAB435 SAB437 SAB534 SAB535 SAB537
4 SAB444 SAB445 SAB447 SAB544 SAB545 SAB547
Shoulder Positioning Cylinder Shoulder Abutment
Ø4.8 Ø5.9
> Packing unit : 1 Shoulder Abutment. Diameter
> For Cement Retained Prosthesis. Diameter Ø4.4 Ø5.5
Height
N-Hex
> Dual anti-rotation grip for prevention of screw loosening.
> Integrated with screw and abutment. 10.7 SAPR001 SAPW001
H
> Tightened with the Shoulder Driver. > Packing unit : 1 Shoulder Positioning Cylinder.
> Tightening torque force : 30 [Link]. > Inner cutting surface for anti-rotation on abutment.
> Abutment level impression : Impression cap in platform Ø4.1 fixture and direct impression in platform Ø5.8 fixture. D > Insert into the Impression Cap.

4IPVMEFSⰝ KRR19L 4IPVMEFSⰝ KRW19L

Shoulder Lab Analog


Shoulder Abutment Cap D Type[Hex] Hex[2.7&3.4] N-Hex
Shoulder Abutment Ø4.8 Ø5.9
Diameter Shoulder Abutment
Diameter Ø4.8 Ø5.9 Ø4.8 Ø5.9
L
Diameter Ø5.4 Ø6.5 Diameter
D Height Length Ø4.8 Ø5.9 Ø4.8 Ø5.9
6.2 IASR140 IASW140 4 SLCR040 SLCW040 SLBR040 SLBW040
H
7.7 IASR155 IASW155 5.5 SLCR055 SLCW055 SLBR055 SLBW055
9.2 IASR170 IASW170 7 SLCR070 SLCW070 SLBR070 SLBW070
> Packing unit : 1 Shoulder Abutment Cap. > Packing unit : 1 Shoulder Lab Analog.
> Protection from cheek and tongue for gingival healing period. > Replacement of abutment shape in working cast.
> Alternative usage for sub-structure of temporary prosthesis. Hex N-Hex > Choose by abutment length.

100 / 101 COWELL® Implant System


Ball Abutment
Outer Cap Block-out Spacer

Prosthetic Procedure III Ø3.4


2.4mm Diameter

Cuff
Length
Ø5.0
4
Ø6.0
4
Component Selection Guide for Ball Abutment 1 EBAT411R EBAT511R
2 EBAT412R EBAT512R
Red : 10N Orange : 15N Green : 20N 3 EBAT413R EBAT513R
Ball Inner Cap 4 EBAT414R EBAT514R
D > Packing unit : 1 Ball Abutment + 3 Inner Caps (1 per each color) + 1 Block-out Spacer + 1 Outer Cap.
> For Implant Supported Over-denture Prosthesis.
L > Tightened with the Ball Driver and Torque Wrench.
C
> Tightening torque force : 30 [Link].
> Direct impression.
Ball Outer Cap Block-out Spacer

Ball Abutment

Ball Outer Cap Diameter


D Height Ø3.4
Red : 10N Orange : 15N Green : 20N
2.4 BATC003C
Ball Inner Cap H
> Packing unit : 2 Outer Caps.

Ball Inner Cap


Block-out Ball Inner Cap
Spacer BATC003I
> Packing unit : 2 Block-out Spacers + 6 Inner Caps (2 per each color).
> Retention force : Red 10N, Orange 15N & Green 20N.
Red : 10N Orange : 15N Green : 20N

Ball Analog

Ball Analog
D
Diameter
Length Ø4.0
L
4 SBAL400

> Packing unit : 4 Lab Analogs.


> Replacement of abutment shape in working cast.

Ball Abutment

Ball Driver
H
Type
Height Ratchet
19 KRB19L
> Packing unit : 1 Ball Driver. *Extra Product
> Used with the Torque Wrench to tighten and untighten the Ball Abutment.

Healing
Fixture Cover Screw Abutment I&R Driver Height 100
KBIR01

For Removal For Insertion

> Packing unit : 1 I&R Driver.


> Used to insert and remove the Inner Caps into and out of the Outer Cap.

102 / 103 COWELL® Implant System


INNO SUB. FULL SURGICAL KIT [KCA010F] INNO INT. FULL SURGICAL KIT [KCA010FI]
SUB. INT. > For the selection and use guide of the components for the Sub. Int. & Ext. Full Surgical Kit, refer to pages 107 to 113.
> For INNO Submerged Implant System (Sub.). OCTAGON
HEXAGON > All components are for Sub. / Int. / Ext. except for the Fixture Drivers and the Depth Gauge used for Sub. exclusively.
> For the INNO Internal Implant System (Int.).
SYSTEM SYSTEM > All components are for Sub. / Int. / Ext. except for the Fixture Drivers used for Int. exclusively.

Ⱍ'JYUVSF Ⱍ'JYUVSF Ⱍ'JYUVSF Ⱍ'JYUVSF Ⱍ'JYUVSF Ⱍ'JYUVSF Ⱍ'JYUVSF Ⱍ'JYUVSF Ⱍ'JYUVSF Ⱍ'JYUVSF
'JOBM%SJMM 'JOBM%SJMM 'JOBM%SJMM 'JOBM%SJMM 'JOBM%SJMM 'JOBM%SJMM 'JOBM%SJMM 'JOBM%SJMM 'JOBM%SJMM 'JOBM%SJMM

Path Drill Point Drill Ø2.2 X 7 Ø3.5 X 7 Ø4.0 X 7 Ø4.5 X 7 Ø5.0 X 7 Ø6.0 X 7 Path Drill Point Drill Ø2.2 X 7 Ø3.5 X 7 Ø4.0 X 7 Ø4.5 X 7 Ø5.0 X 7 Ø6.0 X 7
2KTD18 KPD01S KPSD2207 2KTD3707 2KTD4007 2KTD4507 2KTD5007 2KTD6007 2KTD18 KPD01S KPSD2207 2KTD3707 2KTD4007 2KTD4507 2KTD5007 2KTD6007

Parallel Pin Parallel Pin Ø2.2 X 8 Ø3.5 X 8 Ø4.0 X 8 Ø4.5 X 8 Ø5.0 X 8 Ø6.0 X 8 Parallel Pin Parallel Pin Ø2.2 X 8 Ø3.5 X 8 Ø4.0 X 8 Ø4.5 X 8 Ø5.0 X 8 Ø6.0 X 8
KPP002 KPP002 KPSD2208 2KTD3708 2KTD4008 2KTD4508 2KTD5008 2KTD6008 KPP002 KPP002 KPSD2208 2KTD3708 2KTD4008 2KTD4508 2KTD5008 2KTD6008

1.2 Hex Driver L 1.2 Hex Driver XL Ø2.2 X 10 Ø3.5 X 10 Ø4.0 X 10 Ø4.5 X 10 Ø5.0 X 10 Ø6.0 X 10 1.2 Hex Driver L 1.2 Hex Driver XL Ø2.2 X 10 Ø3.5 X 10 Ø4.0 X 10 Ø4.5 X 10 Ø5.0 X 10 Ø6.0 X 10
KHD1221 KHD1227 KPSD2210 2KTD3710 2KTD4010 2KTD4510 2KTD5010 2KTD6010 KHD1221 KHD1227 KPSD2210 2KTD3710 2KTD4010 2KTD4510 2KTD5010 2KTD6010

M. Mount Driver. L R. Mount Driver. L Ø2.2 X 12 Ø3.5 X 12 Ø4.0 X 12 Ø4.5 X 12 Ø5.0 X 12 Ø6.0 X 12 M. Mount Driver. L R. Mount Driver. L Ø2.2 X 12 Ø3.5 X 12 Ø4.0 X 12 Ø4.5 X 12 Ø5.0 X 12 Ø6.0 X 12
KMMD06L KRMD19L KPSD2212 2KTD3712 2KTD4012 2KTD4512 2KTD5012 2KTD6012 KMMD06L KRMD19L KPSD2212 2KTD3712 2KTD4012 2KTD4512 2KTD5012 2KTD6012
4VC'JYUVSF%SJWFS *OU'JYUVSF%SJWFS

M. Fixture Driver. S R. Fixture Driver. L Ø2.2 X 14 Ø3.5 X 14 Ø4.0 X 14 Ø4.5 X 14 Ø5.0 X 14 M. Fixture Driver. S R. Fixture Driver. S Ø2.2 X 14 Ø3.5 X 14 Ø4.0 X 14 Ø4.5 X 14 Ø5.0 X 14
2KMMS01S 2KHDS01L KPSD2214 2KTD3714 2KTD4014 2KTD4514 2KTD5014 KMMI01S KHDI01S KPSD2214 2KTD3714 2KTD4014 2KTD4514 2KTD5014

M. Fixture Driver. L R. Fixture Driver. XL Drill Extension Ø3.5 Countersink Ø4.0 Countersink Ø4.5 Countersink Ø5.0 Countersink Ø6.0 Countersink M. Fixture Driver. L R. Fixture Driver. L Drill Extension Ø3.5 Countersink Ø4.0 Countersink Ø4.5 Countersink Ø5.0 Countersink Ø6.0 Countersink
2KMMS01L 2KHDS01X KDE002 4KCS35 4KCS40 4KCS45 4KCS50 4KCS60 KMMI01L KHDI01L KDE002 4KCS35 4KCS40 4KCS45 4KCS50 4KCS60

Torque Wrench Depth Gauge Torque Wrench Depth Gauge


A common tool for A common tool for
KTW001 KDG004 Sub. / Int. / Ext. An exclusive tool by type KTW001 KDG001 Sub. / Int. / Ext. An exclusive tool by type

104 / 105 COWELL® Implant System


INNO EXT. FULL SURGICAL KIT [KCA010FE]

01
EXT. > For the selection and use guide of the components for the Sub. Int. & Ext. Full Surgical Kit, refer to pages 107 to 113. SUB. INT. EXT.
HEXAGON OCTAGON HEXAGON
HEXAGON > For the INNO External Implant System (Ext.). SYSTEM SYSTEM SYSTEM
SYSTEM > All components are for Sub. / Int. / Ext. except for the Fixture Drivers and the Multi Countersink used for Ext. exclusively.
Drill / Surgical Tool

Point Drill
> Primarily used for marking the implant recipient site and to determine spacing.
> The point drill has a unique pointed tip, making this an excellent drill for starting the osteotomy
L
7 through the hard cortical plate.

31.8 3.5
Length 15
KPD01S

Path Drill
L > Used for the case that path modification is required.
> Excellent ablation force that does not slip in slanted bone.
35.2 14
> Easy to drill even in extraction socket without slipping.
12
10
8
7 Length 15
Ⱍ'JYUVSF Ⱍ'JYUVSF Ⱍ'JYUVSF Ⱍ'JYUVSF Ⱍ'JYUVSF 3
'JOBM%SJMM 'JOBM%SJMM 'JOBM%SJMM 'JOBM%SJMM 'JOBM%SJMM 2KTD18
Ø2

Path Drill Point Drill Ø2.2 X 7 Ø3.5 X 7 Ø4.0 X 7 Ø4.5 X 7 Ø5.0 X 7 Ø6.0 X 7
2KTD18 KPD01S KPSD2207 2KTD3707 2KTD4007 2KTD4507 2KTD5007 2KTD6007
Length band 7mm Fixture
Initial Drill > Initial stepped Drill - Ø2.2 / Ø2.8 / Ø3.3 mm 8mm Fixture
10mm Fixture
stepped osteotomy at Ø1.8 Drill site. 12mm Fixture
Ø2.2 X 8 18.5 4 L
Parallel Pin Parallel Pin Ø3.5 X 8 Ø4.0 X 8 Ø4.5 X 8 Ø5.0 X 8 Ø6.0 X 8 14mm Fixture
KPP002 KPP002 KPSD2208 2KTD3708 2KTD4008 2KTD4508 2KTD5008 2KTD6008 16&18mm Fixture

Length 8 9 11 13 15 17&19
1.2 Hex Driver L 1.2 Hex Driver XL Ø2.2 X 10 Ø3.5 X 10 Ø4.0 X 10 Ø4.5 X 10 Ø5.0 X 10 Ø6.0 X 10 KPSD2207 KPSD2208 KPSD2210 KPSD2212 KPSD2214 KPSD2218
KHD1221 KHD1227 KPSD2210 2KTD3710 2KTD4010 2KTD4510 2KTD5010 2KTD6010 *Extra product

18.5 4 2 17

M. Mount Driver. L R. Mount Driver. L Ø2.2 X 12 Ø3.5 X 12 Ø4.0 X 12 Ø4.5 X 12 Ø5.0 X 12 Ø6.0 X 12
KMMD06L KRMD19L KPSD2212 2KTD3712 2KTD4012 2KTD4512 2KTD5012 2KTD6012
&YU'JYUVSF%SJWFS
%SJMM Stopper with irrigation groove.
%SJMM
M. Fixture Driver. S R. Fixture Driver. L Ø2.2 X 14 Ø3.5 X 14 Ø4.0 X 14 Ø4.5 X 14 Ø5.0 X 14
4.9
KMME01S KHDE01L KPSD2214 2KTD3714 2KTD4014 2KTD4514 2KTD5014 Ø4.9 Crestal cutter – Flattening the slopped and sharp crestal ridge which
2.5 makes crestal bone surround around the whole fixture platform.
3.3

1.5 2.8
Ø3.3/2.8 Pilot cutter – Stepped hole guiding the pathway of Ø3.5 / Ø4.0 /Ø4.5
fixture's final drill.
M. Fixture Driver. S R. Fixture Driver. L Drill Extension Ø3.5 Countersink Ø4.0 Countersink Ø4.5 Countersink Ø5.0 Countersink Ø6.0 Countersink
KMME02S KHDE02L KDE002 4KCS35 4KCS40 4KCS45 4KCS50 4KCS60
Ø2.2 Path cutter – Determining the orientation of fixture installation.

The Initial Drill guides the pathway of


Torque Wrench Depth Gauge
Multi Countersink the Final Drills. The Final Drill is inserted a
A common tool for half into the Initial Drill's hole without drilling.
4KCS01 KTW001 KDG001 Sub. / Int. / Ext. An exclusive tool by type

106 / 107 COWELL® Implant System


Final Drill > Ø3.5 / 4.0 / 4.5 / 5.0 / 6.0 fixture’s Final Drill.
Tap Drill
> Used in preparation of osteotomy in dense bone only.
18.5 4 L > 7 / 8 / 10 / 12 / 14 / 16 / 18mm fixture’s Final Drill.
Ø3.5
Fixture Dia. Ø3.5 Ø4.0 Ø4.5 Ø5.0 Ø6.0
Ø4.0
Fixture Dia.
Length Ø3.5 Ø4.0 Ø4.5 Ø5.0 Ø6.0 3KMTD35A 3KMTD40A 3KMTD45A 3KMTD50A 3KMTD60A
Ø4.5
8 2KTD3707 2KTD4007 2KTD4507 2KTD5007 2KTD6007 *Extra product
Ø5.0
9 2KTD3708 2KTD4008 2KTD4508 2KTD5008 2KTD6008
11 2KTD3710 2KTD4010 2KTD4510 2KTD5010 2KTD6010 Ø6.0

13 2KTD3712 2KTD4012 2KTD4512 2KTD5012 2KTD6012


15 2KTD3714 2KTD4014 2KTD4514 2KTD5014
18.5 4 2 17
17&19 2KTD3718 2KTD4018 2KTD4518
*Extra product

Parallel Pin > Insert the Parallel Pin after Ⱍ2.2 or 3.5 Drill to check the osteotomy path.
Countersink > Bone quality 1 : high compressive placement of fixtures induces the failure of osseointegration > In order to prevent losing Parallel Pin in patient’s mouth, be sure to tie floss through the hole
and bone loss. in the Parallel Pin.
> Countersink decreases torque force (Ø4.0 Fixture : 80 [Link] -> 45 [Link] / Ø5.0 Fixture : 150 [Link] -> 45 [Link]). H

D
> Preventing compressive necrosis of dense cortical bone. Height 21 After Ø2.2 After Ø3.5
initial drilling. final drilling.
28.6 KPP002
Fixture Dia. Ø3.5 Ø4.0 Ø4.5 Ø5.0 Ø6.0
Diameter Ø3.7 Ø4.2 Ø4.6 Ø5.1 Ø6.0
4KCS35 4KCS40 4KCS45 4KCS50 4KCS60
1mm
1mm

Lower margin of depth Upper margin of depth


marking line indicating marking line indicating
fixture platform level. 1 mm over fixture platform level.

Drill Extension > Used for lengthening the Drill when using a Hand-piece.
Multi Countersink > Do not go over recommended torque when using the Drill Extension.
H
Diameter Ø6.5
D Ø6.0 Fixture line. Height 27.5
4KCS01 Ø5.0 Fixture line.
34 Ø4.5 Fixture line. KDE002
> Only for Ext. Ø3.5 / Ø4.0 Fixture line.

Triangle mark indicating the cutting surface of the drill shaft.

108 / 109 COWELL® Implant System


Mount Driver > Used to install Pre-Mount type fixtures. Fixture Driver > Used to install No-Mount type fixtures.

Type Machine
Type Sub.
H Height Machine System
Length Sub. Int. Ext.(Hex 2.7) Ext.(Hex 3.4)
20.5(Short) KMMD06S
Int.
28.1 / 26.3 / 26.4 (Short) 2KMMS01S KMMI01S KMME01S KMME02S
26.3(Long) KMMD06L
33.3 / 30.5 / 31.4 (Long) 2KMMS01L KMMI01L KMME01L
32.3(X-Long) KMMD12X
Ext. 40.3 / 35.5 / 36.4 (X-Long) 2KMMS01X KMMI01X KMME01X
*Extra product
*Extra product

Type Ratchet
Sub. System
Length Sub. Int. Ext.(Hex 2.7) Ext.(Hex 3.4)
H Type
Height Ratchet 20.7 / 19.5 / 19.9 (Short) 2KHDS01S KHDI01S KHDE01S
Int.
12(Short) KRMD12S 25.7 / 24.5 / 24.9 (Long) 2KHDS01L KHDI01L KHDE01L KHDE02L
19(Long) KRMD19L Ext. 30.7 / 29.5 / 29.9 (X-Long) 2KHDS01X KHDI01X KHDE01X
*Extra product *Extra product
4
3
2 4
1 3
2
1

Hex Driver > Used to insert or remove the Cover Screw, Healing Abutment and Abutment Screw etc.
> The Machine Drivers are used with contra angle, while the Ratchet Drivers are used with
the Torque Wrench.
H
Type Machine Torque Wrench > Torque control in implant placement and abutment connection.
Height
Hex
Hex 0.9 Hex 1.2 > Torque force 10, 25, 30 & 35 [Link] are able to be controlled by pulling the elastic bar.
> Maximal torque force 120 [Link] with pulling the rigid main bar.
22(Short) KMD09S KMD12S
28(Long) KMD09L KMD12L
Code KTW001
*Extra product

H Type Ratchet
Hex
Height Hex 0.9 Hex 1.2
12(X-Short) - KHD1212
17(Short) KHD0915 KHD1215
23(Long) KHD0921 KHD1221
29(X-Long) KHD0927 KHD1227 Depth Gauge > Measuring the drill depth with scaled rod.
> Measuring the 5mm space between adjacent fixtures with flat end of the other side.
*Extra product


Code KDG001 






> One side of the Depth Gauge measures the drilling depth and the other side
measures the gingival height from the top of the implant.

Code KDG004 ⲼExclusive for Sub.

110 / 111 COWELL® Implant System


02
୔ Sloped edentulous ridge adjacent of tooth > Crestal cutter of the Initial Drill and Final Drill.
> Longer drills than fixture’s length.

Drilling Sequence E.g. 12mm Fixture

> Minimal drill frequency with the Point Drill, Initial Drill and Final Drills > Length Marking
(Ø3.5, Ø4.0 and Ø4.5 Fixtures).
Initial Drill Final Drill 12mm Drill 10mm Fixture
Crestal flatting Longer Final Drill




୔ Wide extraction socket

 > Absence of the cortical bone & spongy bone.
 > Narrower diameter than the fixture’s diameter.


Initial Drill Ø4.0 Drill Ø4.5 Fixture

Actual length of Drill : Fixture + 1mm ୔ Torque force control


> The 0.5mm deeper placement increases the initial torque force of fixture.
0.5mm deeper level.
Fixture Placement Level
Level Crestal Level 0.5mm Deeper Level
Density D1 D2 D3 D1 D2 D3
Torque 34.1 29 15.5 44.4 38.4 19.1

> The pumping action of drill removes the bone chip in the hole.
> In dense bone, the debriment action decreases the high torque force.
Point Ø2.2 Ø3.5 Ø4.0 Ø4.5 Ø5.0 Ø6.0

Ø3.5 Pumping action of final drill for debriment


Density D1 D2 D3
Non- Debriment 34.1 29 19.6
Ø4.0
Debriment 30 25 15.5

Ø4.5

Ø5.0
୔ In weak bone quality 4 of maxillary tuberosity
> No pumping action.
Ø6.0
> 0.5mm deeper placement of fixture.
> Wider fixture than Final Drill.

> Ø5.0 fixture : a series of the Point Drill, Initial Drill, Ø4.0 Final Drill and Ø5.0 Final Drill.
> Ø6.0 fixture : a series of the Point Drill, Initial Drill, Ø4.0 Final Drill, Ø5.0 Final Drill and Ø6.0 Final Drill.
0.5mm deeper level.
Level Crestal level 0.5mm Deeper Level
Debriment with without with without
Ø4.5 Fixture 4.4 10.2 - 12.9
Ø5.0 Fixture 11.6 19.9 14.1 24.5
Ø4.5 Drill Ø5.0 Fixture
ⲼThe Countersink and Tap Drill are used in hard bone quality.
*Extra product

112 / 113 COWELL® Implant System


INNO SUB. SMART SURGICAL KIT [KSA002]
SUB. Length Marking & Stopper Actual Length of Drill : Fixture length + 1mm
> For the INNO Submerged Implant System (Sub. Ø3.5, 4.0, 4.5 & 5.0).
HEXAGON
SYSTEM > A simpler kit for implant surgery, applied to fixtures with 8~14mm in length using the Drills and Stoppers.

Diameter band Ø2.2


Ø3.5
Ø4.0
Ø4.5
Ø5.0





8mm 10mm 12mm 14mm


Fixture Line Fixture Line Fixture Line Fixture Line

* While the 2.2 Initial Drill is used at 1,000 rpm with irrigation,
the Final Drills from the 3.5 to 5.0 should be used at 50 rpm without irrigation.

Drilling Sequence

Drill
Ø3.5
Point Ø2.2 Ø3.5 Countersink

Normal Bone Hard Bone

Point Drill Initial Drill Ø3.5 Final Drill Ø4.0 Final Drill Ø4.5 Final Drill Ø5.0 Final Drill
KPD01S 2KTD22 2KTD35 2KTD40 2KTD45 2KTD50 Ø4.0
Point Ø2.2 Ø4.0 Countersink

Stopper Countersink
Normal Bone Hard Bone

Ø4.5
Point Ø2.2 Ø3.5 Ø4.5 Countersink
8 Drill Stopper 10 Drill Stopper 12 Drill Stopper Ø3.5 Countersink Ø4.0 Countersink Ø4.5 Countersink Ø5.0 Countersink
KSDS08S KSDS10S KSDS12S 4KCS35 4KCS40 4KCS45 4KCS50 Normal Bone Hard Bone

Extension & Driver

Ø5.0

Point Ø2.2 Ø4.0 Ø5.0 Countersink

Normal Bone Hard Bone

Torque Wrench
Drill Extension M. Fixture Driver L R. Fixture Driver L 1.2 Hex Driver L 1.2 Hex Driver XL
KDE002 2KMMS01L 2KHDS01L KHD1221 KHD1227 KTW001

114 / 115 COWELL® Implant System


INNO SUB-SHORT SURGICAL KIT [KSI001] Length Marking & Stopper Actual length of drill : Fixture + 0.5mm. Diameter band Ø2.0
Ø4.0
Ø4.5
Ø5.0
Ø5.5
SUB. 4 5.5 7 Ø6.0
HEXAGON > For the INNO Submerged Short Implant System (Sub.).
SYSTEM




4mm 5.5mm 7mm


Fixture Line Fixture Line Fixture Line

Drilling Sequence
Normal Bone Hard Bone

Ø4.0
Point Ø2.0 Ø4.0 Countersink

A common tool for An exclusive tool for


the INNO Surgical Kits. the Short Implants.

Normal Bone Hard Bone

Drill

Ø4.5
Point Ø2.0 Ø4.0 Ø4.5 Countersink

A common tool for An exclusive tool for


the INNO Surgical Kits. the Short Implants.

Point Drill Ø2.0 Step Drill Ø4.0 Step Drill Ø4.5 Step Drill Ø5.0 Step Drill Ø5.5 Step Drill Ø6.0 Step Drill
Normal Bone Hard Bone
KPD01S KSSD2004 KSSD4004 KSSD4504 KSSD5004 KSSD5504 KSSD6004

Stopper Countersink

Ø5.0
Point Ø2.0 Ø4.0 Ø4.5 Ø5.0 Countersink

A common tool for An exclusive tool for


the INNO Surgical Kits. the Short Implants.

Normal Bone Hard Bone


4mm Drill Stopper 5.5mm Drill Stopper 7mm Drill Stopper Ø4.0 Countersink Ø4.5 Countersink Ø5.0 Countersink Ø5.5 Countersink Ø6.0 Countersink
SIDS04 SIDS05 SIDS07 4KCS40S 4KCS45S 4KCS50S 4KCS55S 4KCS60S

Extension & Driver


Ø5.5 Countersink
Point Ø2.0 Ø4.0 Ø4.5 Ø5.0 Ø5.5

A common tool for An exclusive tool for


the INNO Surgical Kits. the Short Implants.

Normal Bone Hard Bone

Torque Wrench
M. Mount Driver. L R. Mount Driver. L 1.2 Hex Driver L
KMMD06L KRMD19L KHD1221 KTW001
Ø6.0 Point Ø2.0 Ø4.0 Ø4.5 Ø5.0 Ø5.5 Ø6.0 Countersink

A common tool for An exclusive tool for


the INNO Surgical Kits. the Short Implants.

116 / 117 COWELL® Implant System


INNO SUB-NARROW SURGICAL KIT [KNA001] Fixture Driver
30.6
18

2.8 Type Ratchet Machine


SUB-N.
HEXAGON > For the INNO Submerged Narrow Implant System (Sub-N). Ratchet KHDS01XN KMMS01XN
SYSTEM
38.1

2.8

Machine
arrow K I T Ratchet X-Long Machine X-Long
12

10
Drill Ø 3.3 Fixture Driver
Stopper Fixture
Mount Driver
8
Ø 3.1
Fixture

Ø 2.6 Type Ratchet Machine


Parallel Pin
Countersink KRMD19L KMMD12X
Ratchet Long
H
Ø 2.2
H > Used to install fixture. *Extra product
Spare
Point 1.2 Hex Driver
Depth Gauge
Ratchet Machine

Torque Wrench

Drill

Parallel Pin
21
Code KPP003

After Ø2.2 After Ø3.1 / Ø3.3


Drilling. Drilling.
Point Drill Ø2.2 Twist Drill Ø2.6 Twist Drill Ø3.1 Twist Drill Ø3.3 Twist Drill
Cuff Height
KNPD20 KNSD22L KNSD26L KNSD31L KNSD33L

Stopper Countersink
Normal Bone
Drilling Sequence Hard Bone

8 Drill Stopper 10 Drill Stopper 12 Drill Stopper Ø3.1 Countersink Ø3.3 Countersink
KNDS08 KNDS10 KNDS12 4KCS31N 4KCS33N

Ø3.1
Extension & Driver Point Ø2.2 Ø2.6 Ø3.1 Ø3.1
Countersink Normal Bone

Hard Bone
Depth Gauge

KDG001

Torque Wrench
Fixture Driver(Ratchet) Fixture Driver(Machine) Parallel Pin 1.2 Hex Driver L
KHDS01XN KMMS01XN KPP003 KHD1221 KTW001
Ø3.3
Point Ø2.2 Ø2.6 Ø3.3 Ø3.3
Countersink

118 / 119 COWELL® Implant System


COWELL® PROSTHETIC KIT [KPA003]
SUB.
> For the INNO Submerged Implant System.
HEXAGON
SYSTEM > Try-in Kit for determining abutment specifications.

Mini Plus® Implant


Cement Type
SUB. HEXAGON
SYSTEM Prosthetic Kit Ball Type

Surgical kit

Abutment Gauge > Measuring of the diameter, cuff length and angulation of abutment.

7
5.5
4
4
2 3
1

Cuff 2 Cuff 4 15° / Cuff 2 15° / Cuff 4 25° / Cuff 2 25° / Cuff 4
2SAG002 2SAG004 2SAG152 2SAG154 2SAG252 2SAG254

Straight Type Angulated Type

Abutment Driver > Used in the Shoulder, Solid, Absolute, and Straight Abutment placement.

Shoulder Ø4.0 Shoulder Ø5.0 Hex 1.2 Ratchet Hex 1.2 Machine Tissue Punch
KRR19L KRW19L KHD1221 KMD12L KTP001 2 Pieces
Trephine soft tissue over
the Cover Screw of fixture.

120 / 121 COWELL® Implant System


System Flow
DESIGN OF MINI PLUS® FIXTURE (1P-C.)
Cement Type
>G For mandible anterior spaces and edentulous arch.
> For semi-permanent or temporary solution.

Fixture Protection Cap Impression Coping Lab Analog

2
Implant
Top Anti-Rotation Square
> Promotes firm attachment to prosthetics.
> Direct impression is recommended when
2
modified/altered.
Diameter Diameter
Ø2.5 Ø3.0
Cuff Cuff
Length 2.0mm 4.0mm Length 2.0mm 4.0mm

10mm AMC2210S AMC2410S 10mm AMC3210S AMC3410S


5 12mm AMC2212S AMC2412S 12mm AMC3212S AMC3412S
14mm AMC2214S AMC2414S 14mm AMC3214S AMC3414S

Cutting Groove > Packing unit : 1 Fixture. > Packing unit : 1 Fixture.
> Two cutting grooves at 5mm, 7mm facilitates > Abutment level impression. > Abutment level impression.
alteration of height.

Cuff

Impression Coping / Lab Analog

Impression Coping
> Packing unit : 1 Impression Coping.
Surface Treatment : SLA-SH®
> Fastened on the shoulder of the cement post and connected with
> Uniform micro-surface geometry, increased BIC
the Lab Analog for working cast.
and improved surface wetness.
> Direct impression is recommended when modified/altered.

Length
Lab Analog
Impression Coping Lab Analog
> Packing unit : 1 Lab Analog.
AMCI001 AMCL001
> The same adjustment must be made for the Lab Analog when the abutment
portion of the fixture is modified/altered.
> Replacement of the cement post shape in working mode.
Cutting edge allows Self-Tapping function

Protection Cap > Provides temporary protection of mucogingiva


and tongue after implantation.

Diameter Diameter
Height Ø4.0
Height
7mm AMCC001

9mm AMCC002
Diameter 11mm AMCC003

> Packing unit : 1 Protection Cap.

122 / 123 COWELL® Implant System


System Flow
DESIGN OF MINI PLUS® FIXTURE (1P-B.) Outer Cap Block-out Spacer
Ball Type
>G For semi-permanent or temporary solution for overdenture prosthesis. 2.4mm

Red : 10N Orange : 15N Green : 20N

Fixture Impression Coping Lab Analog Ball Inner Cap

Implant

Diameter Diameter
Ø2.5 Ø3.0
Cuff Cuff
4.2 Easy delivery using the ball type drivers Length
2.0mm 4.0mm Length
2.0mm 4.0mm

10mm AMB2210S AMB2410S 10mm AMB3210S AMB3410S

12mm AMB2212S AMB2412S 12mm AMB3212S AMB3412S

14mm AMB2214S AMB2414S 14mm AMB3214S AMB3414S


Cuff
> Packing unit : 1 Fixture. > Packing unit : 1 Fixture.

Ball Outer Cap Diameter


Surface Treatment : SLA-SH® D Height Ø3.4

> Uniform micro-surface geometry, increased BIC 2.4 BATC003C


H
and improved surface wetness. > Packing unit : 2 Outer Caps.

Length
Ball Inner Cap
Block-out Ball Inner Cap
Spacer BATC003I
> Packing unit : 2 Block-out Spacers + 6 Inner Caps (2 per each color).
> Retention force : Red 10N, Orange 15N & Green 20N.
Red : 10N Orange : 15N Green : 20N

Cutting edge allows Self-Tapping function

Impression Coping / Lab Analog


Diameter

Impression Coping
> Packing unit : 1 Impression Coping.
> Used for impression taking of the ball post.

Impression Coping Lab Analog


Lab Analog
AMBI001 AMBL001 > Packing unit : 1 Lab Analog.
> Replacement of the ball post shape in working model.

124 / 125 COWELL® Implant System


Tissue Punch
SURGICAL KIT [KMA003] KTP16
*Extra product

Ø2.35(Ø2.0 Implant)

KTP21 Ø2.85(Ø2.5 Implant)


6mm 4mm 2mm
KTP26 Ø3.35(Ø3.0 Implant)
> Easy removal of soft tissue for flapless
surgery.
KTP28 Ø3.55(Ø3.3 Implant) > 0.3mm wider than implant diameter
allows more predictable results.

Point Drill
Multi Gauge *Extra product > Allows precise measurement of drilling depth and path.
36.8mm
KPD03L Length
30
20
10mm 8mm 5mm
(Ø3.0, Ø3.3 Implant) (Ø2.5 Implant) (Ø2.0 Implant) KDGP13 Ø1.3

Ø1.3 Twist Drill KDGP19 Ø1.9 18mm 16mm 14mm 12mm 10mm

KTWD13S Ø1.3 Ø2.0 Implant Final Drill

16mm 14mm 12mm 10mm


Drilling Sequence
Ø1.8 Twist Drill Ø2.3 Twist Drill
Hard Bone
D1 / D2
KTWD18S Ø1.8 Ø2.5 Implant Final Drill
Ø2.5

Point Drill Ø1.8 Twist Drill Ø2.5 Implant (12mm)


16mm 14mm 12mm 10mm
Soft Bone
D3 / D4

Ø2.3 / Ø2.6 Twist Drill


KTWD23S Ø2.3 Ø3.0 Implant Final Drill

KTWD26S Ø2.6 Ø3.3 Implant Final Drill

Ø1.8 Twist Drill Ø2.6 Twist Drill


Hard Bone Hard Bone
16mm 14mm 12mm 10mm
D1 / D2 D1 / D2

Driver Ø3.0
Cement Type Ball Type
Point Drill Ø2.3 Twist Drill Ø3.0 Implant (12mm)
Machine KMDC31L Machine KMDB27L
Soft Bone Soft Bone
D3 / D4 D3 / D4

Ratchet KRDC19L Ratchet KRDB19L


ⲼIn case of D4 bone quality, make proper adjustment in drilling. Due to narrow diameter,
the Mini Plus® Implants have self-tapping characteristic.

126 / 127 COWELL® Implant System


®   
More than 20 years of clinical outcomes

Cowellmedi Co.,Ltd Cowell R&D Institute


Floor 6, Blue Fin Tower, 42, Seochojungang-ro,Seocho-gu, Seoul, Korea 48, Hakgam-daero 221beon-gil, Sasang-gu, Busan, 46986, Korea
Tel. +82-2-3453-5085 Fax. +82-2-3453-5086 E-mail. cib@[Link] Tel. +82-51-314-2028 Fax. +82-51-314-2026

Cowellmedi USA INC


218 Trianon LN Villanova PA 19085-1442 USA
Tel. 1-623-939-1344 Fax. 1-623-939-1472

You might also like