37648IM
37648IM
Model 35080B
kVp Divider
Instruction Manual
CONTENTS
APPENDICES
Appendix A: MAMMOGRAPHIC MEASUREMENTS
WITH THE MODEL 35080B KVP DIVIDER A-1
A.1 K-edge Measurement ................................................. A-2
A.2 Linear Measurement ................................................... A-4
LIST OF FIGURES
3-1 Simplified functional block diagram of the
Model 35080B kVp Divider ............................................ 3-2
3-2 Output ratio plotted as a function of kV ............................ 3-3
3-3 Output of the Model 35080B kVp Divider
compared to voltage applied to x-ray tube ..................... 3-3
4-1ab Suggested oscilloscope settings for use with the
Wide Range Filter Pack (37617C) ................................. 4-3
4-2 Geometry of setup for above-table x-ray tube .................. 4-5
4-3 Geometry of setup for below-table x-ray tube .................. 4-5
4-4 The Wide Range Filter Pack (37617C) Filtration
Effect Correction Curve ................................................. 4-6
4-5 The Wide Range Filter Pack (37617C) Linearity
Correction Curve ........................................................... 4-7
5-1 The Model 35080B kVp Divider single-phase and
three-phase output with the Test Filter Pack (32863C) .. 5-2
6-1 Response curve for the Cadmium K-edge
Mammo Filter Pack (37355C) ........................................ 6-1
6-2 The kVp correction curves for the Cadmium
K-edge Mammo Filter Pack (37355C) ........................... 6-3
6-3 K-edge measurement setup ............................................ 6-4
6-4 DC mammographic generator input waveforms .............. 6-5
6-5 The Model 35080B kVp Divider output waveforms with
the Cadmium K-edge Mammo Filter Pack (37355C) .... 6-6
6-6 Single-phase mammographic generator input waveform . 6-7
6-7 The Model 35080B kVp Divider output waveform with
the Cadmium K-edge Mammo Filter Pack (37355C) .... 6-8
7-1 Correction curves for 12, 24, and 35 microns
molybdenum filtrations ................................................. 7-2
7-2 The correction curve that shows the effect of adding
0.25 mm Al (0.01 in. Al) to the 24 microns
molybdenum filter ......................................................... 7-3
7-3 Linear measurement setup .............................................. 7-4
8-1 CT Filter Pack (33551C) Filtration Effect Correction
Curve ........................................................................... 8-3
8-2 CT Filter Pack (33551C) Linearity Correction Curve ........ 8-4
9-1 Mobile Filter Pack Plus (37946C) Filtration Effect
Correction Curve .......................................................... 9-3
9-2 Mobile Filter Pack Plus (37946C) Linearity Correction
Curve ........................................................................... 9-4
LIST OF TABLES
1-1 Permitted Tube Current Settings (less than ± 1 kV effect) 1-1
1-2 The Wide Range Filter Pack (37617C) Specifications ..... 1-2
11-1 Replaceable Parts Information ..................................... 11-10
Section 1
SPECIFICATIONS
Range: 50-150 kVp with the Wide Range Filter Pack (37617C)
Accuracy: ± 2% of reading, exclusive of linearity, filtration, and gain effects
Calibration: Internally-generated signal provides a calibration check.
Permitted Tube Current Range: 1 mA to 3000 mA, dependent on waveform
and distance
Response Time: 150 microseconds 10-90%
Environment: operating temperature 15° to 35°C; storage temperature
-35° to 50°C; relative humidity to 90%
Orientation: The long axis of the Model 35080B kVp Divider is oriented
perpendicular to the axis of the x-ray tube to eliminate heel effect.
Single-phase Single-phase
Three-phase
half - wave full - wave
Minimum Tube
Current (12" Focal 1 mA 2 mA 4 mA
Spot to kVp divider)
Maximum Tube
Current (22" Focal 750 mA 1500 mA 3000 mA
Spot to kVp divider)
Factory calibration performed at 200 mA, 2.5 mm Al inherent filtration,
22 inches focal spot to kVp divider. Six-point calibration check.
Standard Filter Pack: One Wide Range Filter Pack* (37617C) is provided for
measurement from 50 to 150 kVp.
Optional Filter Packs: A Test Filter Pack (32863C) that provides verification
of instrument accuracy, a K-edge Mammo Filter Pack (37355C)* and a
Linear Mammo Filter Pack (37351C) used for precision measurements from
22 to 40 kVp, a CT Filter Pack (33551C) used for CT system kVp
measurements from 70 to 140 kVp, a Mobile Filter Pack Plus (37946C) used
for measurements on x-ray systems with kV waveforms above 1 kHz with
substantial ripple, and a Low Range Filter Pack (38237C) used for
measurements from 30 kVp to 90 kVp where the kVp range or sensitivity of
other filter packs is not adequate.
Section 2
INTRODUCTION
AND APPLICATIONS
The user should note, however, that as with an invasive divider, the accuracy of
readings will depend on the accuracy of the oscilloscope or readout device
used in the measurements.
The symbol on the instrument indicates that the user should refer
to the operating instructions located in the manual.
Section 3
THEORY OF
OPERATION
3.1 INTRODUCTION
When used with the Wide Range Filter Pack (37617C) the Model
35080B kVp Divider produces an output voltage which closely follows
the voltage applied to the x-ray tube. It does this with a 10-90%
response time of 150 microseconds, so the waveform can be used not
only for determining kVp, but for making precise time determinations
and for diagnosing generator malfunctions.
1.1
0.9
Model 35080B Output in Volts
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
40 50 60 70 80 90 100 110 120 130 140 150
kV
Figure 3-3. Output of the Model 35080B kVp Divider compared to voltage
applied to x-ray tube
3.4 OUTPUT
In the Model 35080B kVp Divider, the intensity ratio is taken with a
high speed logarithmic ratio module. Thus, the output of the Model
35080B kVp Divider follows the kVp waveform applied to the x-ray
tube within the linear range of the filter pack. The waveform differs
from the waveform output of a resistive divider in that information
below the kV baseline of the filter is not displayed (see Figure 3-3).
NOTE
Section 4
OPERATING
INSTRUCTIONS
NOTE
1. Insert the Wide Range Filter Pack (37617C) into the Model 35080B
kVp Divider and connect the kVp divider to the oscilloscope.
2. Set the oscilloscope gain to 0.1 V / div, DC Coupled. This setting
corresponds to a 10 kV / div sensitivity.
3. If making kV measurements from 50 to 120 kV, set the FUNCTION
switch of the kVp divider to the OFF position and adjust the
oscilloscope vertical position control so that the oscilloscope trace
coincides with the bottom horizontal line on the screen, which
corresponds to 40 kV. Then set the FUNCTION switch of the kVp
Divider to the ‘‘CAL’’ position and verify that after a delay of
several seconds the trace on the oscilloscope coincides with middle
horizontal line on the screen, which corresponds to 80 kV. When
switching to the ‘‘RUN’’ position, a slight offset may be noted on the
oscilloscope. Do NOT readjust the oscilloscope trace to compensate
for this offset. Refer to Figure 4-1a.
NOTE
a) b)
50 kV ________ 90 kV ________
Figure 4-1a and b. Suggested oscilloscope settings for use with the Wide
Range Filter Pack (37617C)
kV = 0.300 x 100 + 40
= 30 + 40
= 70 kV
The kVp divider must be inverted for under-table x-ray head evaluation
(refer to Figure 4-3). Again, the filter pack should be on beam center,
perpendicular to the axis of the x-ray tube and the beam normal to the
table. The beam area must be large enough to cover the active area of
the filter pack.
In positioning the kVp divider, the long axis of the instrument should be
perpendicular to the x-ray tube axis to eliminate “heel effect.” The
beam must strike the filter pack side of the kVp divider, the unit is
designed to be turned over for under-table x-ray heads. In setting the
x-ray generator, 200 mA at 22 inches is recommended, but the limits of
operation allow higher or lower settings and distances. Refer to
Section 1, Specifications, for low current limits.
Figure 4-2. Geometry of setup for Figure 4-3. Geometry of setup for
above-table x-ray tube below-table x-ray tube
Example: Suppose the kVp Divider reads 90 kV, using the Wide
Range Filter Pack (37617C). The x-ray machine has 5.0 mm Al total
filtration. The filtration effect curve for this filter pack indicates a
filtration effect of 0.5 kV per mm Al at 90 kVp. To determine the actual
reading, the filtration in the x-ray machine is subtracted from the
standard calibration filtration (2.5 mm Al) and the resulting number is
multiplied by the filtration effect: (2.5-5.0) x 0.5 = -1.25 kV. The
resulting number, -1.25 kV, is the correction which is added to the
apparent reading. Therefore, the actual reading is
90 kV + (-1.25 kV) = 88.75 kV.
Filtration Effect
kV Per mm AL
kVp Per mm AL
Figure 4-4. The Wide Range Filter Pack (37617C) Filtration Effect
Correction Curve
Figure 4-5. The Wide Range Filter Pack (37617C) Linearity Correction
Curve. The kVp Correction Value read from the vertical scale
should be added to the Model 35080B kVp Divider reading to
produce the corrected kVp result.
Section 5
TEST FILTER PACK
OPTION (32863C)
5.1 SPECIFICATIONS
Output: 3.00 V ± 1%
Cal: 0.750 V ± 1%
X-ray Generator kV setting: 100 kV is recommended
X-ray Generator mA setting: 200 mA at 22 inches focal spot to kVp
divider is recommended
5.2 INTRODUCTION
The Test Filter Pack (32863C) provides a means for a complete system
check of the Model 35080B kVp Divider’s calibration. The CAL
position on the kVp divider provides a check of the electronics only.
The Test Filter Pack (32863C) verifies the calibration of the kVp
divider from radiation input through to the output signal under actual
operating conditions.
Figure 5-1. The Model 35080B kVp Divider single-phase and three-phase
output with the Test Filter Pack (32863C)
Section 6
CADMIUM K-EDGE MAMMO
FILTER PACK OPTION (37355C)
6.1 SPECIFICATIONS
Intended Use: beryllium window, molybdenum anode mammographic
generators
Range: 27.5 kVp to 29.5 kVp
Accuracy: ±0.5 kV
Sensitivity: 0.1 volt per kV
Zero: 27.0 kV
Cal: 0.75 V
Minimum Anode Current: 50 mA
Target to kVp Divider Distance: 12 in (30.5 cm)
Effect of Molybdenum Filtration: negligible, 12 to 35 microns
Cadmium K-edge Mammo Filter Pack (37355C)
Response Curve (24 microns Mo Filter)
Output Voltage
kVp
6.2 INTRODUCTION
The Cadmium K-edge Mammo Filter Pack (37355C) provides a non-
invasive way to make kVp measurements of mammographic x-ray
generators. This filter pack uses the inherent stability of the Cadmium
K-edge as an absolute reference for precision measurements between
27.5 and 29.5 kVp (for beryllium window, molybdenum anode
mammographic generators with molybdenum filtration). When used in
conjunction with the Linear Mammo Filter Pack (37351C), precision
calibrations of ± 1.0 kV accuracy can be performed from 22 to 40 kV.
The Model 35080B kVp Divider takes the ratio of the x-ray signals
passing through the two filters. As long as the applied voltage to the
x-ray generator is less than 26.7 kV, the output of the kVp Divider is
constant (approximately 3 volts). When the applied voltage exceeds
26.7 kV, the ratio and thus the output voltage increases rapidly. The
filters are carefully manufactured so that the voltage increases at
0.1 V / kV above 27.0 kV. For molybdenum anode, beryllium window
generators, there is very little effect on the calibration for changes in
molybdenum filtration from 12 to 35 microns. As shown by Figure 6-2,
the addition of aluminum filtration does influence the sensitivity.
Adding 0.25 mm of aluminum can increase the kVp divider's reading by
up to 0.5 kVp.
Indicated kVp
Figure 6-2. The kVp correction curves for the Cadmium K-edge Mammo
Filter Pack (37355C)
Minimum
Model
35080B
kVp Divider
Generator Input
DC Generator 24 kV, 26 kv, 28 kV
Input Kilovolts
Milliseconds
Milliseconds
Figure 6-5. The Model 35080B kVp Divider output waveforms with the
Cadmium K-edge Mammo Filter Pack (37355C) (DC
mammographic generator)
Generator Input
Filtered Single-Phase Generator
Input Kilovoltage
Seconds
Seconds
Figure 6-7. The Model 35080B kVp Divider output waveform with the
Cadmium K-edge Mammo Filter Pack (37355C) (single-phase
mammographic generator)
NOTE
Section 7
LINEAR MAMMO FILTER
PACK OPTION (37351C)
7.1 SPECIFICATIONS
Intended Use: beryllium window, molybdenum anode mammographic
generators
Range: 22.0 kVp to 40.0 kVp
Accuracy: ± 1.0 kV exclusive of filtration effects
Sensitivity: 0.01 volt per kV
Cal: 0.200 V (20.0 kV)
Minimum Anode Current: 50 mA
Target to kVp Divider Distance: 12 in to 24 in (30.5 to 61 cm)
7.2 INTRODUCTION
The Linear Mammo Filter Pack (37351C) provides a non-invasive way
to make kVp measurements of mammographic x-ray generators. This
filter pack features nearly perfect linearity for precision measurements
between 22 and 40 kVp (for beryllium window, molybdenum anode
mammographic generators with molybdenum filtration).
7.4 CORRECTIONS
If the amount of molybdenum filtration were known, the correction
could be made with the correction curves in Figures 7-1 and 7-2.
Figure 7-1 shows the correction required for total molybdenum filtration
of 12 microns, 24 microns, and 35 microns (no correction is required for
24 microns). Figure 7-2 shows the effect of adding 0.25 mm Al to the
24 microns of molybdenum.
Linear Mammo Filter Pack (37351C)
Correction for Molybdenum Filtration
kV Correction
Indicated kVp
Indicated kVp
Figure 7-2. The correction curve that shows the effect of adding
0.25mm Al (0.01in Al) to the 24 microns
molybdenum filter. The aluminum makes the Linear
Mammo Filter Pack (37351C) read high. To make
corrections, subtract the kV correction from the
indicated reading.
NOTE
Model
35080B
kVp Divider
NOTE
NOTE
NOTE
5. Subtract this kVp voltage from the true K-edge kVp value (see
Section 6.4). (This will be the CORRECTION VALUE applied to
all kVp measurements made with the Linear Mammo Filter Pack--
37351C.)
Section 8
CT FILTER PACK
OPTION (33551C)
8.1 SPECIFICATIONS
Range: 70 to 140 kV
Sensitivity: 0.1 V / 10 kV
Cal: 0.644 V
Zero: 60 kV
Cal: 124.4 kV
Linearity: less than 1 kV correction from 70 to 140 kV
Calibration Filtration: 2.5 mm Al
Filtration Effect: less than 1 kV per mm Al difference in filtration
(from the 2.5 mm Al calibration standard) over the 70 to 140 kV
range
Active Area: 40 mm x 6.5 mm
Minimum Anode Current: 25 mA at 25 inches
8.2 INTRODUCTION
The CT Filter Pack (33551C) provides a non-invasive way to make
independent tests of the internal voltage divider of a CT system -- and a
convenient and safe way to check tube voltage.
These applications are made possible with the Inovision CT Filter Pack
(33551C) in conjunction with the Model 35080B kVp Divider.
The filters are carefully selected to give the best response curve possible
over the kV range of CT interest (see Specifications).
Effect of Filtration on kV
Correction
kV
kV
Section 9
MOBILE FILTER PACK
PLUS OPTION (37946C)
9.1 SPECIFICATIONS
Range: 50 to 135 kV
Sensitivity: 0.1 V / 10 kV
Cal: 0.600 V
Zero: 40 kV
Cal: 100 kV
Calibration Filtration: 2.5 mm Al
9.2 INTRODUCTION
The Mobile Filter Pack Plus (37946C) is a specialized filter pack
designed primarily for use on mid- to high-frequency generators
(≥ 1 kHz) with substantial kV ripple (≥ 10%). For single-phase and
3-phase generators, or other low-frequency generators (< 1 kHz), use of
the Mobile Filter Pack Plus (37946C) is not recommended. The Wide
Range Filter Pack (37617C) should be used for these applications.
Measurements below the 50kV operating range of either the Mobile
Filter Pack Plus (37946C) or the Wide Range Filter Pack (37617C)
should be made with the Low Range Filter Pack (38237C).
The formula used to calculate the kV is the same as for the Wide Range
pack:
where the ZERO kV is 40 for both the Mobile and Wide Range Filter
Packs (37946C and 37167C). However, when the kVp divider is placed
in the “CAL” mode with a Mobile Filter Pack Plus (37946C) inserted,
an output voltage of 0.600 V ± 1% should be obtained, corresponding
to a CAL kV of 100.
The filtration and linearity correction curves for the Mobile Filter Pack
Plus (37946C) are given in the next two sections. Two linearity curves
are shown. The solid line on the linearity graph is the standard linearity
curve for an x-ray generator with a 2 kHz waveform with 20% observed
ripple. The dashed line is normally not used, but shows the linearity
curve for generators with DC to 720 Hz ripple components.
Example: The Model 35080B kVp Divider with a Mobile Filter Pack
Plus (37946C) inserted reads 80kVp. The x-ray system has 3.3 mm Al
total filtration. The filtration effect curve of Figure 9-1 indicates a
filtration effect of 1.05 kVp per mm Al at 80 kVp. The difference in
filtration, 2.5 - 3.3 = -0.8 mm Al, is multiplied by the correction value to
yield -0.8 x 1.05 = -0.84 kVp. This number is then added to the
apparent reading to obtain the actual kVp, which is
80 kVp + (-0.84 kVp) = 79.16 kVp.
Filtration Effect
kVp Per mm Al
1.6
1.5
1.4
1.3
1.2
1.1
kVp Per mm Al
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
40 50 60 70 80 90 100 110 120 130 140
Figure 9-1. Mobile Filter Pack Plus (37946C) Filtration Effect Correction
Curve
Example: The Model 35080B kVp Divider with a Mobile Filter Pack
Plus (37946C) has a reading of 83 kVp. From the chart, a kVp
Correction value of -0.9 kVp is obtained. Assuming filtration
corrections are not required for this x-ray system, the actual kVp is
83 + (-0.9) = 82.1 kVp.
0.5
0
kVp Correction
-0.5
-1
-1.5
-2
-2.5
-3
-3.5
40 50 60 70 80 90 100 110 120 130 140
Model 35080B kVp Reading
Figure 9-2. Mobile Filter Pack Plus (37946C) Linearity Correction Curve.
The kVp Correction Value read from the vertical scale should be
added to the Model 35080B kVp Divider reading to obtain the
corrected kVp result.
Section 10
LOW RANGE FILTER
PACK OPTION (38237C)
10.1 SPECIFICATIONS
Range: 30 kVp to 90 kVp
Sensitivity: 0.1 V / 10 kV
Cal: 0.150 V
Zero: 26 kV
Cal: 41 kV
Accuracy: ± 2%, exclusive of linearity and filtration effects
Minimum Anode Current for less than ± 1 kV effect: Typical at
22 inches with a 3 phase x-ray generator, 3 mA from 90 kVp down to
50 kVp, increasing to 10 mA @ 40 kVp and 30 mA @ 30 kVp.
10.2 INTRODUCTION
The Low Range Filter Pack (38237C) is optimized for measurements in
the 30 kVp to 90 kVp range. It can be used to make measurements
below the 50 kVp lower limit of the Wide Range Filter Pack (37617C)
and Mobile Filter Pack Plus (37946C) on x-ray generators with settings
below 50 kVp. The Low Range Filter Pack (38237C) will typically
operate at lower mA settings, making it useful in taking measurements
difficult to obtain with either the Wide Range Filter Pack (37617C) or
Mobile Filter Pack Plus (38237C) at the lower end of their kVp
operating range.
Oscilloscope Setup
Two possible methods for reading the output of the Model 35080B
kVp Divider on a storage oscilloscope are to read the kVp divider
output voltage and then convert the voltage into the equivalent kV
using the conversion formula, or to set up the oscilloscope to allow
direct readout of kV. Both methods are described and assume that the
operator is familiar with the use of his storage oscilloscope.
Geometry of Setup
The Model 35080B kVp Divider with the Low Range Filter Pack
(38237C) is set up for use in the same manner as described in
Section 4.5 for the Wide Range Filter Pack (37617C).
The normal setup described in Section 4.5 calls for 200 mA and
22 inches. Lower current settings may be used in many cases;
however, the detectors in the kVp divider require a minimum intensity
below which the output voltage of the kVp Divider develops noise
and errors. The minimum mA at a given distance varies from
generator to generator. The typical specification is given for guidance
only. If a particular test setup seems to be giving noisy or erroneous
readings, increase the mA setting of the generator, reduce the distance
from the source to the kVp divider, or both. When reducing the
distance from the focal spot to the kVp divider, make sure that the
x-ray beam is wide enough to cover the active area of the filter pack at
the distance used.
2.5
2
kVp Per mm Al
1.5
0.5
-0.5
30 40 50 60 70 80 90 100
Model 35080B kVp Reading
Figure 10-1. Low Range Filter Pack (38237C) Filtration Effect Correction
Curve
Example: The Model 35080B kVp Divider with a Low Range Filter
Pack (38237C) has a reading of 45 kVp. From the chart, a kVp
Correction value of -0.2 kVp is obtained. Assuming filtration
corrections are not required for this x-ray system, the actual kVp is
45 + (-0.2) = 44.8 kVp.
1.5
1.0
kVp Correction
0.5
0.0
-0.5
-1.0
20 30 40 50 60 70 80 90
-1.5
Model 35080B kVp Reading
Figure 10-2. Low Range Filter Pack (38237C) Linearity Correction Curve.
The kVp Correction Value read from the vertical scale should be
added to the Model 35080B Model 35080B kVp Divider reading
to obtain the corrected kVp result.
Section 11
MAINTENANCE
11.2 TROUBLESHOOTING
Difficulties experienced by the Model 35080B kVp Divider user may be
caused by problems from one or more of the three sources described
below: problems with the storage oscilloscope used to obtain the kV
reading; problems with the x-ray machine; and either improper
technique or an electronic fault with the kVp divider. Typically, the
existence of a problem would be signalled by one of the following:
NOTE
If the battery voltage drops below 7.0 V, the battery should be replaced,
as the remaining life of the battery is quite limited. If the battery voltage
is already below 6.5 volts when checked, the battery must be replaced
for proper operation of the kVp divider.
To replace the battery, remove the screw from the cover on the bottom
of the kVp divider (refer to Figure 11-1). Lift the cover from the
instrument case. Place your hand over the battery compartment and roll
the instrument upright, letting the battery drop into your hand.
11.4 CALIBRATION
WARNING
NOTE
NOTE
Calibration Adjustment
NOTE
1. Insert the Test Pack (32863C). Place the Model 35080B kVp
Divider on the x-ray table, such that the distance from the focal
spot to the kVp divider is 22 inches, and the instrument is normal
to the beam--with the long axis of the kVp divider perpendicular to
the axis of the x-ray tube; this configuration will avoid “heel
effect.” Settings on the generator for kVp should be 100 kV and
200 mA.
NOTE
LEGEND
kV WAVEFORM
Inovision Part
Description
Number
NOTE
Appendix A
MAMMOGRAPHIC
MEASUREMENTS WITH
THE MODEL 35080B
Part of the correction factor for the Linear Mammo Filter Pack (37351C) is due
to variations in tube filtration and part is due to positional dependence.
Therefore, if the Model 35080B non-invasive kVp Divider is moved when
using the Linear Mammo Filter Pack (37351C), a new measurement at the
K-edge comparison kVp point must be made with the Linear Mammo Filter
Pack (37351C) to compute a new correction factor.
The following procedure enables the first-time user to achieve good results.
An experienced user can take many short cuts. The kVp divider with the
mammo kVp option is for use on molybdenum anode, beryllium window tubes
only.
Oscilloscope Setup
1. Connect the Model 35080B kVp Divider to the scope using a
shielded cable.
2. Set the vertical sensitivity to 0.5 volts per division.
3. With the input in the ‘‘GROUND’’ position, adjust the vertical
position control to place the trace on the bottom graticule line.
4. Set the scope in single trace, storage mode.
5. Place the input selector switch in the ‘‘DC’’ position.
Part C
1. Set the generator for 29 kVp and make an exposure.
2. Measure the peak voltage. (If this voltage is greater than the
BASE VOLTAGE by 0.1V but not greater than 0.25 V above the
BASE VOLTAGE, record in box C (Figure A-2). If the voltage is
greater than 0.25 volts above the BASE VOLTAGE, repeat this
step at one kVp lower. If the voltage is not greater than the BASE
VOLTAGE by 0.1 volts, repeat this step at one kVp higher.)
Part D
Repeat the last measurement to verify that the generator is repeatable.
The voltage reading should repeat to within 0.020 V. If the reading is
not repeatable, the generator needs service.
Part E
1. Subtract the BASELINE voltage from the voltage measured in
Part C, step 2 (box C).
2. Multiply this voltage by 10, add to 27.0, and record in box D
(Figure A-2). This is the true kVp for this station.
3. Set the scope for single sweep, storage mode, and in the ‘‘DC’’
input mode.
Oscilloscope Setup
1. Set the vertical sensitivity to 0.05 volts per division.
2. With the input switch in the ‘‘GROUND’’ position use the vertical
position control to place the trace on the bottom graticule line.
3. Set scope for single sweep, storage mode, and in the ‘‘DC’’ input
mode.
Part B
1. Subtract the kVp in box F from the kVp in box D (Figure A-2).
(This will be the CORRECTION VALUE applied to all kVp
measurements made with the Linear Mammo Filter Pack--
37351C.)
2. Insert this value in all spaces in the column labeled
‘‘CORRECTION VALUE.’’
NOTE
Part C
1. Measure other kVp values as desired.
2. Enter the voltage in the ‘‘LINEAR VOLTAGE’’ column on the
data sheet; multiply by 100.
3. Enter the value in the ‘‘LINEAR kVp’’ column.
4. Apply the CORRECTION VALUE to all kVp's by either adding to
or subtracting from the ‘‘LINEAR kVp’’ column, and record the
results in the ‘‘CORRECTED kVp’’ column. (This column now
has the true kVp value for all kVp's measured.)
Part D
If the position of the Model 35080B kVp Divider changes during the
Linear kVp measurement, you must repeat the measurement starting
with Part C: some of the correction value is positional dependent.
Appendix B
CORRECTING kVp
MEASUREMENTS
FOR GAIN EFFECTS
NOTE
B.1 INTRODUCTION
The calibration mode feature of the Model 35080B kVp Divider
provides a method to monitor the gain of the internal circuitry. When
the kVp divider is placed in CAL mode, the specific voltage and kVp
stated on each individual filter pack should be obtained. If the output
voltage in CAL mode is not within ± 1% of the value stated on the pack,
kVp measurements should be corrected in one of the following ways,
depending on the device used for measuring the output of the kVp
divider. A deviation of less than ± 1% will result in a maximum kVp
error of 0.75% of reading.
1. Insert the filter pack to be used into the kVp divider and connect to
the oscilloscope. Set the oscilloscope gain to an appropriate value
depending on the sensitivity of the filter pack.
2. With the function switch of the kVp divider set to OFF, adjust the
oscilloscope vertical position such that the trace coincides with the
bottom horizontal line on the screen. This line now corresponds to
the Zero offset kVp given on the pack.
3. Switch the kVp divider to the CAL position. The oscilloscope
trace should now be at a level corresponding to the CAL kVp
given on the filter pack. If this is not the case, the gain effects
should be corrected before making kVp measurements in the RUN
position.
4. The gain effects can be corrected by adjusting the fine tuning knob
on the Volts / div switch of the oscilloscope. While still in the
CAL position, adjust this knob until the trace on the screen
corresponds to the kVp value expected for the filter pack in use.
Measurements made in RUN mode will now be corrected for gain
effects.
Voltages obtained in RUN mode will deviate from correct values by the
same ratio as the deviation in CAL voltages. That is,
0.400V
Corrected Voltage = 0.515V = 0.500V
0.412
Using the sensitivity of 0.1 V / 10 kV and 40 kVp Zero offset given for
this filter pack, the Corrected kVp is 90. A kVp of 91.5 would have
been measured if the above gain correction had not been performed.
( Actual CAL kVp − Zero offset kVp) ( Measured kVp in RUN mode − Zero offset kVp)
=
( Expected CAL kVp − Zero offset kVp) ( Corrected kVp − Zero offset kVp)
where the Expected CAL kVp and Zero offset kVp correspond to the
values given on the actual filter pack. Note that the sensitivity need not
be considered since it cancels out of the above equation. The Corrected
kVp can be obtained directly from this equation by rearranging.
Example: For the Wide Range Filter Pack (37617C) (50-150kVp), the
Expected CAL kVp is given as 80, and the Zero offset value is 40 kVp.
However, a Model 35080B kVp Divider in CAL mode with this filter
pack inserted is connected to a Model 35050A Dosimeter, and a kVp of
81.2 is measured. When placed in an x-ray beam in RUN mode, the
Model 35050A Dosimeter reads 91.5 kVp. The Corrected kVp for this
exposure is calculated from the previous equation.
80 kVp - 40 kVp
Corrected kVp = (91.5 kVp - 40 kVp ) + 40 kVp = 90 kVp
81.2 kVp - 40 kVp
Appendix C
PROCEDURE FOR MAKING kVp
MEASUREMENTS ON RHODIUM
ANODE, RHODIUM FILTRATION
SELECTION OF GE DMR
MAMMOGRAPHY X-RAY
SYSTEMS
C.1 INTRODUCTION
The GE DMR mammography x-ray system has both molybdenum and
rhodium anode selections. Measurements on the Mo anode track can be
made in the same manner as on any other Mo anode mammography
generator, provided Mo is also selected as the additional filter material.
This involves using the Model 35080B Dosimeter and its two associated
mammo filter packs, the Cadmium K-edge Mammo Filter Pack
(37355C) and the Linear Mammo Filter Pack (37351C), as instructed in
the Model 10100A TRIADTM Field Service Kit Instruction Manual
(37580IM) or the Model 35080B kVp Divider Instruction Manual
(37648IM). This addendum is intended to provide a method to measure
kVp on the rhodium anode track, with rhodium also selected as the
additional filter material.
C.2 PROCEDURE
NOTE
2. Insert the Linear Mammo Filter Pack (37351C) into the Model
35080B kVp Divider and acquire a kVp measurement using the
same generator settings as used for the k-edge measurement in
step 1. Use the look-up-table in Table C-2 to determine the actual
kVp of the exposure. For example, if the measured value was
34.3 kVp, the actual kVp would be 28.7.
3. Determine the k-edge correction value as usual, by subtracting the
Linear Mammo Filter Pack (37351C) reading (28.2 in the
example) from the measurement obtained with the Cadmium
K-edge Mammo Filter Pack (37355C) in step 1 (28.7 in the
example). Note that the k-edge correction value can be positive or
negative (the value is (28.2 - 28.7) = -0.5 kVp in this example).