PLC Purification User's Guide (EN) Gilson
PLC Purification User's Guide (EN) Gilson
Chemical Hazards | 9
Electrical Hazards | 9
Mechanical Hazards | 9
Flammable Solvents | 10
Lifting | 10
Replacement Parts | 10
Signs of Damage | 10
Site Requirements | 10
Spacing | 10
CHAPTER 1 | INTRODUCTION | 11
Description | 12
Unpacking | 14
Technical Specifications | 19
Detailed Description | 22
Customer Service | 32
CHAPTER 2 | INSTALLATION | 33
Place the Solvent Bottle Tray | 34
Plumbing Connections | 34
Column Setup | 40
Power Down | 54
CHAPTER 4 | MAINTENANCE | 55
Helpful Hints | 56
Maintenance Schedule | 58
CHAPTER 5 | TROUBLESHOOTING | 67
Troubleshooting Table | 68
APPENDIX A | WARRANTY | 75
General | 75
Limited Warranty | 76
Detector Parts | 78
Fittings | 79
TABLE OF CONTENTS
Fraction Collector | 79
Miscellaneous | 80
Pump Parts | 81
Tubing | 83
Valves | 84
APPENDIX C | MATERIALS | 85
Liquid Contact Materials | 85
System Setup | 90
Control Software | 90
Installation | 91
Plumbing Connections | 92
Start Up | 96
Start Up | 104
Description | 107
Operation | 114
Maintenance | 118
Troubleshooting | 119
IN THIS CHAPTER:
Performing any task not listed in this user’s guide will result in loss of warranty and may be
WARNING
dangerous.
SYMBOL EXPLANATION
Direct Current
Alternating Current
| Electrical Power ON
Caution
WARNING Indicates a potentially hazardous situation which, if not avoided, may result in personal injury.
SAFETY
Indicates a potentially hazardous situation which, if not avoided, may result in minor or
CAUTION
moderate injury.
Indicates a potentially hazardous situation which, if not avoided, may result in equipment
NOTICE
damage.
Chemical Hazards
Any chemicals used for analysis should be handled according to good laboratory practice
(GLP). They should also be stored, used, and disposed of in accordance with the manufacturer’s
specifications, as well as local and national regulations. Potentially hazardous chemicals can be
used with the instrument. Use care when handling chemicals and wear appropriate personal
protective equipment (PPE), such as safety glasses, gloves, etc.
The responsible individual must ensure that personnel are not exposed to hazardous levels of toxic
substances as outlined in the Safety Data Sheets (SDSs), or any documentation provided by local
governing bodies such as The Health Protection Agency (United Kingdom) or The Occupational
Safety and Health Administration (United States).
Electrical Hazards
Unless specifically instructed, do not remove any protective covers. Detach all sources of voltage
from the instrument before the service, repair, or exchange of parts.
Use only the grounded AC cord provided. Ungrounded power cords can result in electrical shock
and serious personal injury. Faulty or frayed power cords must be immediately replaced with one
of the same type and rating. When it is necessary to use a non-original power cord, make sure the
replacement cord adheres to following specifications and local building safety codes: 1) European
Union Model; Connector A: Male, Type E or F (Schuko), 16A; Connector B: Female, IEC320/C13,
10A; 250 V~, H05VV-F 3G1.0 mm² and 2) United States and Canadian Model; Connector A: Male
Type, NEMA 15-5, 15A; Connector B: Female, IEC320/C13, 10A; 125 V~, SVT 3x18 AWG.
Use only approved fuses with the specified current rating. The PLC Purification System must be
NOTICE
operated within the voltage specified on the marking plate of the instrument.
Mechanical Hazards
The housing of the PLC Purification System must be attached permanently with all screws
WARNING tightened at 3 N m. Unless specifically instructed, do not remove any protective covers.
All protections must be in place before starting the PLC Purification System.
Lifting
The PLC Purification System exceeds the weight one person can lift safely. Two or more people
SAFETY
are required to lift the PLC Purification System safely. Always lift the PLC Purification System
from the base and follow any unpacking instructions provided with the instrument. For more
information, see Unpacking on page 14.
Replacement Parts
Be sure to use only replacement parts mentioned in this user’s guide.
See Replacement Parts and Accessories on page 77.
Signs of Damage
Do not attempt to use the instrument if there are visible signs of damage.
Site Requirements
Do not operate the instrument if site conditions are not within specifications. Refer to
Technical Specifications on page 19 for more details.
Spacing
Allow sufficient spacing around the system for proper cooling and for the connection of power
cords, plumbing, injection pump, liquid handler, external detectors, or any other peripherals.
Chapter 1 | INTRODUCTION
IN THIS CHAPTER:
● Description | 12
● Unpacking | 14
● Technical Specifications | 19
● Detailed Description | 22
● Terms and Abbreviations | 32
● Customer Service | 32
Standard Components
The following diagrams provide a general
description of the PLC Purification System with
standard components.
Figure 1
PLC 2050 Purification System
Touchscreen PC
Refer to page 30 for a
detailed description
Electronic Connections:
Analog, USB, and Collection Probe and
External Controls Collection Valve
Refer to page 31 for Refer to page 30 for
detailed descriptions detailed descriptions
Racks
Two-Way Gradient Valve
Refer to page 42 for
Refer to page 26 for a installation details
detailed description
Dynamic Mixing Chamber Prime and Purge Valve and Pump Heads
Refer to page 25 for a Refer to page 22 for detailed descriptions
detailed description
Figure 2
Standard Components for the PLC Purification System (Left and Front View)
USB Connections
Description
Refer to page 31 for a
detailed description
Figure 3
Standard Components for the PLC Purification System (Right and Front View)
Optional Components
The following diagrams provide a general description of the PLC Purification System with optional
components.
Acquisition Module for
Backflush Valve Additional Detectors
Refer to page 28 for a (Not Shown)
detailed description Refer to page 45 for
installation details
Figure 4
Optional Components for the PLC Purification System (Left and Front View)
GX-241
Unpacking
The optional GX-241 is a compact X/Y/Z instrument designed for automating general liquid
handling procedures. For more information, refer to the GX-241 Liquid Handler User’s Guide.
Other Configurations
PLC Purification Systems can be configured with an external injection pump for high volume
injections and additional detectors, such as an evaporative light scattering detector (ELSD). All
these optional devices can be controlled with GLIDER.
Unpacking
Upon receipt of the instrument, inspect the exterior of all shipping boxes. All boxes should arrive
unopened and undamaged. If examination reveals that damage has occurred in shipment, notify
the carrier and Gilson immediately. Refer to Customer Service on page 32.
WARNING
Do not plug in the PLC Purification System if any damage is detected. Powering the
PLC Purification System in a damaged state may result in serious injury and may damage
internal components of the PLC Purification System.
The PLC Purification Systems are delivered with major components already assembled. Keep the
original container and packing assembly so the unit may be shipped safely, if necessary. Carefully
unpack the PLC Purification System and its accessories. Verify that all parts are included and
undamaged. Do this immediately, even if the system will not be used right away. Many carriers
must receive concealed damage claims within seven days of delivery.
Unpacking
4. Lift the unit out of the box and place it at suitable location, such as a lab bench or cart, and
always on a flat, level surface.
It is recommended to operate the PLC Purification System inside a fume hood to ensure
proper ventilation.
Allow sufficient spacing around the system for proper cooling and for the different
NOTICE connections.
Free spaces of at least 20 cm must be kept around the instrument at all times to allow
for proper tubing and cable connections and to provide adequate ventilation during
operation.
5. Verify that all components from the Unpacking List on page 15 are included.
Unpacking List
The following items are considered standard equipment and are provided with
the PLC Purification Systems. For detailed descriptions and part numbers, refer to
Replacement Parts and Accessories on page 77.
STANDARD EQUIPMENT
Unpacking
OPTIONAL ACCESSORIES
Factory Installed
● Backflush valve, automatic (qty. 1)
○ 4-way, 2-position electric valve for 1/8” tubing, special screws, and cable
○ Control module, power supply, bracket, and cables
○ Tubing kit, 1 mm (ID) (PLC 2050) or 2.1 mm (ID) (PLC 2250 and PLC 2500)
● Column switching valve, automatic (qty. 1)
○ 6-way, 2-position electric valve for 1/8” tubing, special screws, and cable
○ Control module, power supply, bracket, and cables
○ Tubing kit 1 mm (ID) (PLC 2050) or 2.1 mm (ID) (PLC 2250 and PLC 2500)
● Integrated ELS Detector
○ SEDEX FP™ nebulizer
○ Glassware nebulization chamber
○ Chamber shield
○ Accessory kit:
○ Splitter tubing (PEEK), 1/16” (OD) with PEEK fittings, and PEEK union 10-32
► Gas tube (PA), 6mm (OD), 3 m
► Gas tube (PU), 4 mm (OD), 23 cm
► Drain tube assembly (including SS fitting and PTFE seal)
► 2 adjustable releasable clips for black exhaust tube
○ Gas regulator with 5 µm filter (ordered separately):
► Manometer
► Fittings
► Set of two side mounting brackets
● Loop injection valve, automatic (qty. 1)
○ 6-way, 2-position electric valve for 1/8” tubing, special screws, and cable
○ Control module, power supply, bracket, and cables
○ Tubing kit 1 mm (ID) (PLC 2050) or 2.1 mm (ID) (PLC 2250 and PLC 2500)
● Quaternary gradient system 1/8” (PLC 2050 and PLC 2250) or 3/16” (PLC 2500) (qty. 1)
○ 4-way gradient valve
○ Inlet/suction tubing with fittings (qty. 4)
Changes or modifications to the instrument not expressly approved by Gilson could void the
NOTICE
factory-authorized warranty.
Technical Specifications
This instrument complies with part 15 of the Federal Communications Commission (FCC) rules.
Operation is subject to the following two conditions: (1) this instrument may not cause harmful
interference, and (2) this instrument must accept any interference received, including interference
that may cause undesired operation.
Shielded cables must be used with the instrument to ensure compliance with the FCC Class A
limits.
Collection Options Several rack options available. Refer to Fraction Collector on page 79 in the
Replacement Parts and Accessories appendix.
Rack overlays for easy tube identification (optional). Refer to
Install the Rack Overlays on page 46.
Funnel rack with 16 outlets for large and variable collection volume in external high-
capacity containers (optional). Refer to Install the Funnel Rack on page 47.
Fraction collector cover kit (optional).
Install the Fraction Collector Cover Kit on page 46.
Column Holder Built-in column clamp (standard). 40 mm maximum (OD) for columns. Refer to
Column Setup on page 40.
Additional standalone holder for flash columns. 400 mm maximum height. 115 mm
maximum diameter.
Contact Control Analog signal input: Two contacts +/- (-1 V 𝌂 ; and +1V 𝌂, use of
10‑bit A/D converter)
Two inputs (contact closure, CMOS open drain): Start In and Stop In
(5 V 𝌂 maximum)
Two outputs (static relay contacts): Start Out and Stop Out (5 V 𝌂 maximum)
Electrical Protection General: Delayed action fuses 6.3A H 250 V~, T-type (qty. 2)
24 V 𝌂 and 5 V 𝌂: Delayed action fuses with different ratings, L 250 V~, T-type
Altitude Up to 2000 m
*Refer to Materials on page 85 for Gradient valve (gradient former) and PEEK and Kalrez®
Technical Specifications
more information. collection valve
Maximum pressure
PLC 2250 250 mL/min (5 to 250 mL/min)
230 bar (3336 psi)
Safety and Compliance The instrument has been certified to safety standards specified for Canada, Europe,
and the United States. Refer to the instrument rear panel label and the Declaration
of Conformity document for the current standards to which the instrument has
been found compliant.
Detailed Description
Two-Head Pump and Purge Valve
The two-head pump can deliver mobile phases at flow rates up to 500 mL/min (depending
on model) with low residual pulsation. PLC Purification Systems utilize high performance, dual
piston, reciprocal pumps. The pump generates suction through a system of pulleys, belt, and a
stepper motor, which drives the cams and pistons. For plumbing connections, refer to the table
on page 24.
Cleaning Purge
Belt Cam Discs Valve
Detailed Description
Pump
Heads
Pistons
Pulley
Motor
Figure 5 Figure 6
Interior View of Pump Pump Heads and Purge Valve
H J
P Q
Q
N
G
M M
C2
L L
C2
K
F
F
Z
Z
Z
D A
B
A
C1
A
Figure 7
Component and Connection Diagram for the Pump Heads and Prime/Purge Valve
A PTFE tubing, 1/8" (OD), 2.4 mm (ID) J SS, preformed tubing, 1/8” (OD),
with springs (PN 21040137) for 1 mm (ID) (PN 21040121) for PLC 2050
Detailed Description
PLC 2050 and PLC 2250 SS, preformed tubing, 1/8” (OD),
PFA tubing, 3/16" (OD), 1/8" (ID) 2.1 mm (ID) (PN 21040117) for PLC 2250
(PN 21040139) for PLC 2500 and PLC 2500
B PEEK tee union, 1/4"-28, 2.4 mm K ETFE junction tubing for cleaning discs
bore (PN 21040201) for PLC 2050
and PLC 2250
PEEK Y assembly, 5/16"-24 to 10‑32
(PN 4957515) for PLC 2500
C2 PEEK short nuts for 1/8" tubing M SS nut for 1/8" tubing (PN 21040199)
(PN 21040194) SS ferrule for 1/8" tubing
(PN 21040200)
Detailed Description
Mixing volume depends on pump model installed:
● 4 mL for a 50 mL/min pump (PLC 2050)
Figure 8
● 12 mL for a 250 mL/min pump (PLC 2250) Dynamic Mixing Chamber
● 16 mL for a 500 mL/min pump (PLC 2500)
Refer to the diagram below for components and connections to the dynamic mixing chamber.
B A
D
Figure 9
Component and Connection Diagram for the Mixing Chamber
21040143 1/8" with cable, PLC 2050/2250 21040145 1/8" with cable, PLC 2050/2250
21040144 3/16" with cable, PLC 2500 21040175 3/16" with cable, PLC 2500
INLET TUBING
Inlet tubing connects the gradient valve to the solvent reservoirs. Use the two (standard) or
four (optional) inlet lines supplied. Each inlet line has a fitting and spring on one end and an
SS ballast on the other end.
Low
Pressure
Plug
Valve
Body
Spring
ETFE
Ferrule
PEEK
Nut
Figure 11
Inlet Tubing Connections
Detailed Description
Status
Injection E
LEDs
Port D
F
J
H
C J
Sample
Loop
G 1
6 2
5 3
4
A
G G
Figure 12 Figure 13
Automatic Loop Injection Valve Component and Connection Diagram
for the Automatic Loop Injection Valve
D Shut-off valve outlet tubing, 1/8" (OD), J PEEK short nuts for 1/8" tubing
1.6 mm (ID) (PN 21040169) (PN 21040194)
ETFE ferrules for 1/8” tubing
(PN 21040193)
Sample
Loop
Injection
Port
Handle
Figure 14
Manual Loop Injection Valve
BACKFLUSH VALVE
The four-way automatic backflush valve allows users to change the pumping direction (elution
mode) into the connected column at any time while the system is running:
● Direct 'Dir' / Reverse 'Rev', for a HPLC column
● Ascending 'Asc' / Descending 'Dsc', for a CPC column
Outlet
To
Column
To
Inlet Column
Figure 15
Backflush Valve
Detailed Description
spectrum, which is displayed on the main screen.
Fraction Collector
After detection of the signal (chromatogram), the results obtained at the outlet of the column
(peaks) can be directed to tubes in racks. A three-way collection valve located on the dispense
arm allows users to direct fractions to collection tubes or to waste.
Dispense
Arm
Collection
Probe
Tubes
Figure 16 Figure 17
Fraction Collector with Standard Racks 3-Way Collection Valve
Additional
Probe
Standard (ETFE
Dispense Luer
Probe Male-
to-1/4”-28
Male)
Figure 18
Collector Probes
For the PLC 2500, the back pressure regulator is replaced by a simple union because it
creates too much back pressure at high flow rates. The back pressure regulator is supplied as
NOTICE an accessory.
You can install and use the back pressure regulator 20 psi on PLC 2500 if you are working
with flow rates less than 300 mL/min.
Racks
The integrated collector is provided with three
SS racks (standard, PN 21040019) designed for USB USB
18 mm diameter tubes, accommodating up to 192 Touchscreen Ports Ports
tubes and a maximum volume of 6.1 L. Refer to
Fraction Collector on page 79 for a complete list of
rack options.
Integrated Touchscreen PC
The PLC Purification System is controlled by
GLIDER Software that runs on an integrated
touchscreen PC, which allows the user to create
methods and run the PLC Purification System.
Seven USB ports are available, two in the front of
the PLC Purification System, below the screen, four
on the right side, and one more on the left side Power
Indicator Stylus
of the PLC Purification System. USB cables and all
Light
other external connections should not exceed two
meters (6.5 ft.). Figure 19
Touchscreen PC
Local
Detailed Description
Area
Analog RS-232 Serial Network
Signal USB External Communications (LAN)
Inputs Port Controls Port (COM4) Port
Figure 20
Analog Inputs, USB Port, and I/O Ports Figure 21
(Left Side) RS-232 and LAN Ports (Right Side)
USB PORTS
The USB ports can be used to connect any USB peripherals (keyboard, mouse, USB drive,
printer, etc.) or optional acquisition module for additional detectors, or to control the
VERITY 4020 Syringe Pump.
LAN PORTS
The LAN port can used to make direct data transfer to a private server.
EXTERNAL CONTROLS
The inputs 'Start In' and 'Stop In' can be used to start or stop the pump via a pulse (with 10 mA
current limitation) or a closure of a NO/NC contact (O>C) coming from an external device (max.
5 V𝌂). The outputs are static relay contacts (max. 5 V𝌂). 'Start Out' induces a short closure of a
contact NO/NC (Open>Close) at each starting of the pump (Run/Resume). 'Stop Out' induces a
short closure of a contact NO/NC at each stop of the pump (Stop/Pause).
Customer Service
Gilson, Inc. and its worldwide network of authorized representatives provide customers with
the following types of assistance: sales, technical support, applications, and instrument repair.
If you need assistance, please contact your local Gilson representative. Specific contact
information can be found at www.gilson.com. To help us serve you quickly and efficiently, please
refer to Before Calling Us on page 73.
Chapter 2 | INSTALLATION
IN THIS CHAPTER:
A PLC Purification System and its components should be set up and installed in order per the
instructions in this chapter.
Plumbing Connections
Wear appropriate personal protection
CAUTION equipment (PPE) according to the solvent
pumped.
Figure 22
The pump is shipped with high-pressure stainless steel
Solvent Tray
tubing connected, and with plugs on the inlets and
outlets. Remove the plugs before connecting any
tubing.
Inlet Tubing
Inlet tubing connects the gradient valve to the solvent
reservoirs. Use the two (standard) or four (optional)
inlet lines supplied. Take care not to kink the inlet tubing.
Each inlet line has a fitting and spring on one end and
an SS ballast on the other end.
For each inlet line, place the ETFE ferrule into the PEEK
body of the gradient valve and screw on the PEEK nut,
finger tight, and slide the spring to place its end in
the knurled part of the PEEK nut. Springs are used to
prevent the tubes from bending. Colored cable ties are
provided to easily identify the tubing (A=red, B=green,
C=blue, and D=yellow).
Put the ends of the tubing with ballasts into the
appropriate solvent reservoirs.
An adjustable, releasable clip on the left side of the
system allows users to properly guide the tubing into
the bottles positioned on the solvent tray atop the Figure 23
Springs and Colored Cable Ties
system.
Ballast
PEEK
Ballast
Nut
Tubing
Figure 24 Figure 25
Inlet Tubing Ballasts Close Up of Ballast and Fitting
Plumbing Connections
L
Gradient
(Solvent) F
Reservoirs
K
Left
Side
A B C D H
G N
F
E
F
A11 M
B11
C1 K
D1
Waste
Figure 26
Component and Connection Diagram for the Gradient Valve and Waste Tubing
• Quaternary (optional)
Plumbing Connections
1. Screw two PEEK nuts into the check valve housings beneath the pump heads without fully
tightening them.
2. Screw the PEEK nut into the gradient valve outlet on the left side of the PLC Purification System.
3. Finger-tighten the three fittings, and then ensure that the three fittings on the tee union are
fully tightened.
Gradient
Valve Pump
Outlet Heads
Check
Valve
Housings
Tee
Union
Figure 27
Gradient Valve to Pump Heads Connections
Nut
Ferrule
Tube
Fitting
Body
Figure 28
Outlet Tubing Assembly
Waste
Outlet
Figure 29
Waste Outlet
The waste tubing (PN 21040115 for PLC 2050 and PLC 2250; PN 21040116 for PLC 2500) connects
the waste outlet to a waste reservoir. There is an SS fitting at one end and an SS ballast on the
other end.
Connect the tubing to the SS bulkhead union labeled 'Waste', located on the left side of the
system. Tighten the SS nut with a 3/8” spanner, and then place the other end in suitable waste
receptacle.
Plumbing Connections
Purge
Purge Tubing
Valve
Outlet
Figure 30
Purge Valve Outlet
Sample Loop
Most PLC Purification Systems come with the sample loop pre-installed at the factory; however, for
NOTE
custom orders, please follow the instructions below.
The injection valve waste outlet (PN 21040169) tubing connects the waste outlet of the injection
valve to a waste reservoir. There is a fitting at one end and an SS ballast on the other end. Connect
Column Setup
the tubing to the shut-off valve outlet and finger-tighten the PEEK nut. The shut-off valve is
connected to port 2 of the automatic loop injection valve.
Waste
Outlet
Shut-
Off
Valve
Figure 31
Injection Valve Waste Outlet
Column Setup
A wide range of columns can be used with PLC Purification Systems, including normal- or
reverse-phase columns and those with an overall external diameter up to 40 mm. The device used
to install the column is a clamp with vinyl coating. The clamp is mounted on the black holder on
the left side of the PLC Purification System.
Clamp
Holder
Screws
Wing
Nut
Figure 32
Column Positioning
Column Setup
G
slide the column inside the clamp. Securely tighten the
clamp. H
Arm
holds the collector arm during
NOTE
transport by manually loosening it.
Keep this cable tie, it is reusable.
When using short racks, use the additional positioning guide supplied and place it in appropriate
NOTE
holes.
Back
Guide
Back Guide
for
Short Racks
Rack
Rack
Slot
Front
Guide
Figure 35 Figure 36
Rack Installation Short Racks Guide
Never use any cabling not supplied or recommended by Gilson. Use of unspecified cabling may
NOTICE
lead to improper operation or failure to comply with safety or EMC regulations.
When operating the PLC Purification System, it must be possible to disconnect it from mains
supply at any time. In the event of an emergency, the power connector of the instrument must
be easily accessible and removable.
CAUTION The PLC Purification System must never be operated from a power outlet that has no ground
connection. The absence of a ground connection can lead to electric shock or short circuit.
The PLC Purification System is designed for use with liquids; however, liquid contact with
external equipment may lead to the risk of electric shock or short circuit. Ensure that fluid
connections are not close to ancillary equipment and are checked for leaks prior to use. In the
event of a leak, any ancillary equipment not designed for use with liquids must be turned off
until the liquid is removed.
Power
To make the power connection, plug Fuse Receptacle
in the AC power cord into the power Drawer
receptacle located on the right side Voltage
Power Selector
of the PLC Purification System. Then
Switch
make the connection between the
PLC Purification System and the
AC power source.
● 110–120/220–240 V~ (single-phase)
● 50/60 Hz
● 450 W max.
Figure 38
● General fuses: T 6.3A H 250 V~ (qty. 2) Power Switch and Voltage Selector
| indicates that the electrical power is switched on; O indicates that the electrical power is
NOTE
switched off.
on the standard column clamp and an additional column. Secure the additional column with a
second column clamp or use a standalone column holder (PN 21040015).
Manual Automatic
Column Column
Switching Backflush
Valve Valve
Figure 39
Column Switching Valve with Flow Diagrams
1. Secure the manual column switching valve (already mounted on its bracket) to the two
dedicated holes located on the left side of the PLC Purification System with the two
domed-head screws supplied with a 3 mm Allen wrench.
2. Connect the manual column switching valve inlet (Port 1) to the loop injection valve outlet.
3. Connect the column switching valve outlet (Port 4) to the bulkhead union detector inlet.
4. Connect the columns inlets to the column switching valve:
● To Port 2 for the Column 1 inlet
● To Port 6 for the Column 2 inlet
5. Connect the columns outlets to the column switching valve:
● To Port 3 for the Column 1 outlet
● To Port 5 for the Column 2 outlet
6. Use a 3/8” spanner to tighten the nuts and properly crimp the ferrules.
7. Switch the manual switching valve to the right to direct flow to the Column 1. Switch the
valve to the left to direct flow to Column 2.
Do not switch the manual column switching valve while the pump is running as internal
NOTICE
damage may occur due to a sudden overpressure.
Figure 40
Standalone Column Holder
Figure 41 Figure 42
Acquisition Module Acquisition Module Connections
Collector
Cover
with
Magnetic Metallic
Lock Clamp
Transparent
Door
Adapter for
Pipe Connection
and
Air Extraction
Figure 43
Fraction Collector Cover
Figure 45
Overlays for Racks 18 mm Tubes (Front Side Numbering)
Figure 46
Overlays for Racks 18 mm Tubes (Back Side Numbering)
1. Use a 2.5 mm Allen wrench to remove the two cylindrical-head screws from the front
positioning guide.
2. Remove the front positioning guide from the tray.
3. Place the funnel rack on the tray and fix it with the two cylindrical-head screws previously
removed. Tighten the two screws with the 2.5 mm Allen wrench.
4. Connect the 16-outlet tubing and securely tighten the black nuts.
Front Front
Guide Guide
Screws Screws
Figure 48
Funnel Rack Installation
Chapter 3 | OPERATION
IN THIS CHAPTER:
● Start Up | 50
● Prime the Pump | 51
● Loop Injection with Automatic Valve | 52
● Loop Injection with Manual Valve | 53
● Power Down | 54
1. Turn the power on using the switch on the side panel (Refer to External Connections on page 31).
● The power indicator light below the touchscreen on the front panel illuminates.
Start Up
NOTICE Do not use sharp tools to touch the screen. Use the stylus or your fingers.
Figure 50
Purge Valve Knurled Knob
Figure 51
Close Purge Valve Message
For best results, elevate the solvent containers for priming the pump. Or, connect a large
NOTICE
volume syringe to create a slight vacuum at the outlet.
equilibration phase. In both cases, the red LED Load light will illuminate.
Figure 52
Automatic Load and Injection Diagrams
Figure 53
Injection Mode Setting
Mobile phase spray from the needle port may occur when the valve is turned from the
Rinse the valve with water after using buffer solutions to prevent crystals from forming,
NOTICE
which can cause scratches on the rotor seal of the valve.
The sample loop should be loaded when the system is stopped or during an equilibration phase.
To inject the sample into the column:
Locking
Pin
Needle
Port
Plug
Needle
Port
Figure 54
Manual Injection Valve - Needle Port
Figure 55
Manual Injection - Load and Inject Diagrams
5. Switch the valve from the LOAD to INJECT position with a 60° rotation of the handle.
● The sample is pumped into the fluid path and pushed into the chromatography column
via the mobile phase used.
● When the injection mode Wait for Injection is selected in the method edition (refer to
the GLIDER Software User’s Guide), the elution phase will start upon user confirmation
that the valve has been manually switched.
Figure 56
Injection Mode Setting - Wait for Injection Message
Power Down
1. Stop the pump and the running method.
2. Close GLIDER.
3. Turn off the PC and wait for the touchscreen to turn off.
● The UV-VIS detector lamp will remain on.
4. Switch the main power switch to ‘0’. Refer to External Connections on page 31.
● All indicator lights turn off.
● The UV-VIS detector lamp switches off.
● The fans stop.
Chapter 4 | MAINTENANCE
IN THIS CHAPTER:
● Helpful Hints | 56
● Cleaning and Decontamination | 56
● Maintenance Schedule | 58
● Replace a Pump Seal | 59
● Clean or Replace a Pump Check Valve | 62
● Clean the Pump Pistons | 64
● Replace the Mixing Chamber Piston Seal | 65
● Test and Repair the UV-VIS Detector | 66
This chapter describes actions that should be performed on a routine basis to ensure the long
term safe and trouble-free operation of the PLC Purification System. The frequency of the
maintenance activities is dependent on the nature of the application, such as the solvents used,
the volume of the mobile phases delivered by the pump, the level of cleanliness of the facility,
etc. The PLC Purification System has been designed for reliability and needs very little routine
maintenance when operated correctly.
When performing the maintenance described in this chapter, use good laboratory practice (GLP),
including, but not limited to, wearing protective clothing and preparing the maintenance space for
service. Only qualified personnel can perform the maintenance. After completing the maintenance
operation, verify the safe and good working order of the part and PLC Purification System.
Three models of pump are available in the range (50 mL/min, PLC 2050; 250 mL/min, PLC 2250;
and 500 mL/min, PLC 2500). Parts of the heads kit are different according to the pump model
(seals, pistons, cleaning discs, heads, etc.).
Reduced accuracy of the pump may be caused by seal wear or by a fault in the check valves. This
chapter contains information and procedures for replacing consumable parts and maintaining
the PLC Purification System.
Do not open the PLC Purification System or attempt to perform any maintenance other than
the ones described in this chapter, failure to comply with this directive will void the warranty
WARNING and may be potentially dangerous.
Before performing any maintenance, make sure that the PLC Purification System is unplugged
and that the fluid paths are flushed and decontaminated according to the cleaning procedure.
NOTICE Follow local regulations when discarding any used fluidic parts.
● Do not leave buffer in the system. Doing this may cause blockages and wear the seals.
● Run a clean and appropriate solvent through all fluid paths before storing the system.
● Clean the valves if the system has not been used for a while.
● Wipe up all spills immediately.
● Allow liquids to equilibrate to room temperature before running them through the system;
cold liquids may cause leakage.
Exterior
The instrument should be cleaned occasionally, using a dry, clean cloth. If necessary, use a cloth
dipped in water. If liquid is accidentally spilled on the instrument, wipe it up immediately. If a
contamination is suspected, the external surfaces of the instrument should be wiped down with
a laboratory detergent (alkaline) and rinsed with water and then with a 10% ethanolic (or IPA)
aqueous solution (v/v) to remove any residue.
WARNING Beware of splashes when cleaning the fluidic panel. Wear appropriate PPE.
CAUTION Do not to use a detergent that will damage the paint or pollute the work environment.
Before running an application, prime and flush the fluid path with the solvents of this
application.
After having run an application, flush the fluid path with the suitable mixture and let those
NOTICE
clean solvents in the fluid path. Alcohol prevents bacterial development.
Only use water and ethanol to clean the system so not to damage the seals, cables, tubing
and all the plastic parts.
WARNING
Potentially hazardous chemicals can be used with the system. Use care when handling
chemicals and wear appropriate PPE. Handling toxic, flammable, and hazardous chemicals
can lead to health and safety risks.
The pump pistons can also get dirty and interfere with the fluid path, refer to
Clean the Pump Pistons on page 64.
injected.
NOTICE Follow all guidelines listed in this table to avoid damaging the PLC Purification System.
Maintenance Schedule
OPERATION FREQUENCY
• Clean and rinse the gradient valve and all other switching valves with suitable solution. As needed
• Perform maintenance any time an aqueous solution with strong buffers is used.
• Clean all parts in contact with solvents or samples with suitable solutions. Rinse thoroughly using a As needed
mix of water and alcohol.
• Perform maintenance any time an aqueous solution with strong buffers is used.
Replace the detector deuterium lamp when efficiency is unreliable. Refer to the As needed
UV-VIS Detector Service Guide.
• Check the suitable tightening of fittings and the proper state of tubing (not kinked, damaged, etc.). Weekly
• Perform maintenance any time solvent leaks or air bubbles are observed.
Clean the pump check valves. Refer to Clean or Replace a Pump Check Valve on page 62. Quarterly
Clean the pump piston rods (manually). Refer to Clean the Pump Pistons on page 64. Perform Quarterly
maintenance any time the pump head is disassembled.
• Check the efficiency of the detector deuterium lamp. Refer to the UV-VIS Detector Service Guide. Quarterly
• Perform maintenance any time higher noise levels or decreased sensitivity is observed in the
chromatogram.
Clean the detector flow cell. Refer to the UV-VIS Detector Service Guide. Quarterly
• Replace the tubing and associated fittings (except the preformed, SS tubing). Yearly
• Replace damaged tubing and fittings any time damage is observed.
Replace the pump seals. Refer to Replace a Pump Seal on page 59 Yearly
• Replace the pump check valves. Refer to Clean or Replace a Pump Check Valve on page 62. Yearly
• Perform maintenance any time abnormal pressure, flow rate fluctuation, or noisy baselines occur.
When disassembling or reassembling the pump, make sure that each component is clean and
take care that the system is assembled in a clean environment.
There are four seals: one for each pump head, and one for each cleaning disc.
To disassemble a pump head with cleaning disc, and to replace a seal, follow the instructions
below:
SS
Outlet
Tubing
Check
Valve
Housings
Figure 57 Figure 58
Removing the Outlet and Inlet Tubing Removing the Head Screws
Cleaning Exposed
Disc Piston
Cleaning Exposed
Discs Pistons
Figure 59 Figure 60
Removing the Pump Heads Removing the Cleaning Discs
If needed, disconnect the ETFE junction tubing for cleaning discs by hand (PEEK nuts under
NOTE
pump head).
9. Once the part is removed, carefully take out the old seal.
Do not score the seat of the seal. If the seat is scored, it is necessary to replace the pump
NOTICE
head or the cleaning disc (not under warranty).
Figure 61
Back Pump Head and Back Cleaning Disc
B
A
A
Seal
A B
Figure 62
Pump Seal Seating Instructions
Figure 65
Removing the Check Valve Housing and the Check Valve
Ensure that the arrow on the check valve is directed downwards when it is cleaned in
NOTICE
an ultrasonic bath, or damage to the surface quality, the ball and the seat will occur.
Figure 66
Inlet Valve Housing (Bottom)
Inlet Check Valve Placement
Figure 67
Outlet Valve Housing (Top)
Outlet Check Valve Placement
7. Tighten slowly the valve housing in the pump head by hand until finger tight and tighten
the housing with a 17 mm spanner until hearing a little crunch, which means that the new
check valve is properly seated.
Do not overtighten the valve housing as this could degrade the check valve more quickly.
NOTICE
If a liquid leakage or air bubble is detected, tighten slightly to improve the sealing.
8. Reconnect the tubing. When tightening the SS nut on the outlet check valve housing, use a
17 mm spanner to hold the housing and keep the suitable tightening.
1. Follow steps 1 through 8 in Remove the Pump Head and the Cleaning Disc on page 59.
Clean the Pump Pistons
2. Thoroughly clean the piston rods with a lint‑free cloth and alcohol.
3. Follow the steps 8 through 12 in Replace the Seal on page 61 to reinstall the cleaning discs
and pump heads.
Connect the Inlet and Outlet Tubing for the Cleaning Discs
The high pressure heads incorporate a piston washing device called cleaning discs. These are
typically used when aqueous buffers are pumped and function to manually wash the buffers
from the back of the seals and to prevent early wear of these seals.
1. Fit the 1/8" PEEK nuts and ETFE ferrules (supplied) onto suitable lengths of wide bore 1/8”
PTFE tubing.
2. Screw the fittings into the cleaning discs inlet and outlet at the back of the high pressure
heads. Both connections are interchangeable, choose one for the inlet, and the other for
the outlet.
3. Connect the cleaning discs inlet tubing to a dedicated device (a combination syringe with
Luer or another external pumping system).
4. Put the cleaning discs outlet tubing into a waste container.
Inlet
and
Outlet
for
Cleaning
Discs
Figure 68
Pistons Cleaning
To replace the seal of the mixing chamber, follow the instructions below:
Chamber M5
Outlet Screws
Figure 69
Mixing Chamber Piston Removal
Figure 70
Mixing Chamber Piston
lamp is used. The efficiency of the lamp and the cleanliness of the flow cell must be checked
periodically.
If efficiency of detection is unreliable, detector lamp wear could be to blame. Deuterium lamp
life can be checked with GLIDER Software. From the GLIDER main menu, select the ? icon, and
then Information.
When disassembling or reassembling the detector, make sure that each component is clean and
take care that the system is in a clean environment.
Refer to the UV-VIS Detector Service Guide for more information on detector testing, lamp
replacement, and flow cell cleaning.
Chapter 5 | TROUBLESHOOTING
IN THIS CHAPTER:
● Troubleshooting Table | 68
● Repair and Return Policies | 73
Leakage or air bubbles. The fittings are not Tighten the fittings.
sufficiently tightened.
The pump makes • The motor or motor • Stop the pump. Refer to
abnormal noises. driver is damaged or • Switch off the Power Down on page 54.
not properly installed. PLC Purification System.
• The system belt or
pulleys are damaged.
• Inaccurate or irregular • The fittings are not • Tighten the fittings and Refer to page 56.
flow rates. sufficiently tightened. check the state of the
• Lack of solvent suction. • Improper state of tubing.
• Abnormal pressure The pump check valves Check and/or clean Refer to page 62.
fluctuation. are dirty or damaged. and/or replace the pump
• Noise on the data check valves.
system.
Leakage near the pump Worn pump piston seals. Replace the pump piston Refer to page 59.
heads or cleaning discs. seals.
Troubleshooting Table
Leakage near the purge Worn purge valve seal. • Check the purge valve
valve. seal.
• Replace the purge
valve seal, if damaged.
Leakage near the mixing Worn mixing chamber • Check the mixing Refer to page 65.
chamber. piston seal. chamber piston seal.
• Replace the mixing
chamber piston seal if
damaged.
Leakage near a • The fittings are not • Tighten the fittings and
high-pressure valve. sufficiently tightened. check the state of the
• Improper state of tubing.
tubing or fitting. • Replace the fittings
• The fluid path is and/or tubing, if
partially or totally damaged.
blocked. • Clean the valve and the
fluid path.
Leakage near the • The fittings are not Check the tubing and
UV detector flow cell sufficiently tightened. fittings connection to
(liquid flow under the • Improper state of and from the UV cell by
system through the tubing or fitting. removing the plastic
detector drain pan panel on the left side of
tubing). the system.
Cracked quartz window Replace the UV flow cell. Refer to the UV-VIS Detector
inside the UV flow cell. Service Guide.
• Higher noise levels or • Low UV lamp intensity. • Check the UV lamp for Refer to the UV-VIS
decreased sensitivity on • Unreliable efficiency of efficiency. Detector Service Guide.
UV signals. UV lamp. • Check for UV noise and
• Improper autozero. drift.
• Check the light intensity
curve.
• No signal during UV Worn UV deuterium lamp. Replace the UV deuterium Refer to the UV-VIS
acquisition. lamp and perform an Detector Service Guide.
• UV detector lamp does intensity recalibration.
not switch on.
Low signal during UV Worn UV halogen lamp. Replace the UV halogen Refer to the UV-VIS
acquisition in visible lamp and perform a scan Detector Service Guide.
range. intensity.
Lowered UV light Dirty or contaminated UV Check the UV cell for Refer to the UV-VIS
transmission, persistent flow cell. cleanliness. Detector Service Guide.
higher noise levels or
decreased sensitivity on
UV signals with efficient
lamp.
Peaks detected by UV UV flow cell optical path Replace the UV cell with Refer to the UV-VIS
have a too low intensity length is too small. another cell with a higher Detector Service Guide.
whatever the application. path length.
Peaks detected by UV UV flow cell optical path Replace the UV cell with Refer to the UV-VIS
have a too high intensity length is too large. another cell with a lower Detector Service Guide.
(signal saturation) path length.
whatever the application.
Troubleshooting Table
commands. firmware failure. the Traces log files from User’s Guide.
GLIDER.
Fraction collector does • Wrong racks used. • Check rack code in the Refer to the
not collect properly into • Wrong racks software matches the GLIDER Software
tubes. positioning. racks you are using. User’s Guide.
• Wrong collection • Check positioning of
parameters. racks on the tray.
• Check the collection
parameters in GLIDER
and then adjust if
needed.
• Check that the
collection probe is
sufficiently tightened.
The PLC Purification The PLC Purification • Check the AC power source.
System does not System is not supplied with • Try a different AC outlet.
Troubleshooting Table
GLIDER does not An abnormal malfunction • Export the Traces log files Refer to the
work correctly. not detected beforehand. from GLIDER. GLIDER Software
• Upgrade GLIDER to the newest User’s Guide.
version.
The stylus points to An unwanted modification Recalibrate the touchscreen with Refer to the
the wrong place on of touchscreen calibration. the calibration tool installed on GLIDER Software
the touchscreen. the PC. User’s Guide.
The automatic, Defective valve or control • Use GLIDER to manually switch Refer to the
high-pressure module. the valve several times. GLIDER Software
valve (injection, • Check the valve’s status LEDs User’s Guide.
backflush, column on the front panel.
switching, etc.) • Check the control module and
does not respond to its LEDs.
commands and does
not switch properly.
Before Calling Us
Warranty Repair
Units covered under warranty will be repaired and returned to you at no charge. If you have any
questions about applicability, contact your local Gilson representative.
Non-Warranty Repair
For out-of-warranty repairs, contact your local Gilson representative who will discuss service
options with you and can assist in making arrangements to return the equipment, if necessary.
Return Procedure
Contact your local Gilson representative to obtain authorization before returning any Gilson
equipment. To return a piece of equipment:
● Carefully pack the unit to prevent damage in transit. Check with your local Gilson representative
regarding proper method of shipment. No responsibility is assumed by Gilson or your local
Gilson representative for damage caused by improperly packaged instruments. Indicate the
authorization reference on the carton and on the packing slip.
● Always insure for the replacement value of the unit.
● Include a description of symptoms, your name, address, phone number, and purchase order
to cover repair costs, return and shipping charges, if your institution requires it.
Unit End-of-Life
When a unit reaches the end of its useful life, refer to www.gilson.com for
directives and information on the end-of life policy. This is in accordance with
the European Union Directive 2012/19/UE on Waste Electrical and Electronic
Equipment (WEEE).
Appendix A | WARRANTY
IN THIS CHAPTER:
● General | 75
● Limited Warranty | 76
General
A PLC Purification System is under warranty to be free from defects in material or workmanship
under normal installation, use and maintenance, as described in this user’s guide and under
conditions given below.
The warranty period is defined in the order/contract and starts from the date of initial shipment
of the manufacturer. Note that it does not apply to the instrument if modified by the user or
resold without permission from the manufacturer, nor to consumable parts, nor to any failure of
lifetime-expired parts. Refer to Repair and Return Policies on page 73.
The warranty is void in the following cases:
● Failure due to incorrect installation
● Failure due to using the incorrect AC power supply
● Failure due to mechanical force applied to the unit
● Failure due to improper handling by the user or an individual not authorized to operate the
equipment
● Failure due to the use of improper spare parts or hardware
● Damage to the software, data, or hard disk due to a PC virus infection
● Corrosion caused by leakage of solvent or samples
● Corrosion of electronic parts caused by highly corrosive atmospheric gas
● Failure due to the disassembly, modification, relocation, or transport after initial installation
● Failure due to the disconnection of main power without taking the specified normal shutdown
procedure
● Failure due to the disregard of safety regulations
● Failure due to not following the maintenance schedule
● Failure due to acts of nature
Gilson is not responsible for any damage caused by improper use of the system, improper
NOTICE maintenance, unauthorized modifications or failure to comply with the procedures
detailed in Gilson documentation.
● Use of clean solvents and containers (Filter the solvents to avoid presence of fibers and
particles in the pump.)
● Proper positioning, according to the instructions in this guide.
Valves
TERMS OF USE
● Normal tightening of fittings.
● Use of clean solvents and containers. (Filter the solvents to avoid presence of fibers and
particles in the pump.)
● Preventative maintenance schedule followed.
IN THIS CHAPTER:
All items listed must only be supplied by Gilson or an agent thereof. Use of alternate parts
CAUTION may lead to improper operation of the system or failure to comply with safety or EMC
regulations.
The manufacturer’s declaration of conformity becomes invalid if the user modifies the
NOTICE
original product or installs additional components.
Detector Parts
PART NUMBER DESCRIPTION
Fraction Collector
PART NUMBER DESCRIPTION
Miscellaneous
PART NUMBER DESCRIPTION
Pump Parts
PART NUMBER DESCRIPTION
Sample Loops
PART NUMBER DESCRIPTION
Appendix C | MATERIALS
Al2O3 Ceramic Alumina ceramic (Aluminum Oxide or Al2O3) is an excellent electrical insulator and one of
the most widely used advanced ceramic materials. Additionally, it is extremely resistant
to wear and corrosion. Alumina components are used in a wide range of applications such
as electronics, pump components and automotive sensors.
ETFE Ethylene tetrafluoroethylene is the generic name for the material such as Tefzel®. A
fluoropolymer used for sealing surfaces, it is resistant to most chemical attack; however,
some chlorinated chemicals will cause a physical swelling of ETFE tubing.
Fused Silica, SiO2 Glass High purity sand deposits provide the raw material for bulk refractory grade, which is
electric arc melted at extremely high temperatures. Optical and general purpose fused
silica rods and tubing are drawn from a melt made from high purity chemicals. Fiber
optic purity is made by thermal decomposition of high purity gaseous silica containing
chemicals. The glass may be clear or translucent, in which case it is often referred to
as fused quartz. The glass has very high viscosity, and this property allows the glass to
be formed, cooled and annealed without crystallizing. Fused silica glass is a very low
thermal expansion material, and so is extremely thermal shock resistant. The material
is also chemically inert up to moderate temperatures except to hydrofluoric acid, which
dissolves silica. It will devitrify above about 1100°C in the presence of contaminants
such as sodium, phosphorus and vanadium, with the formation of cristobalite crystals
which destroy the properties of the glass. The dielectric properties are stable up through
gigahertz frequencies.
Graphite GFP is a reinforced graphite fiber PTFE material used in moderate to extreme service
Fiber-Reinforced conditions. This material has excellent wear resistance in a wide range of speeds, in
Polytetrafluoroethylene medium pressures and in high temperatures. GFP performs well in water and other liquid
(GFP) solutions and has limited use in vacuum or inert gas applications. Dynamic surfaces in
contact with a seal made from GFP should have a hardness of Rc 40 or higher. GFP is
recommended for applications that require good wear in liquids and humid conditions
at temperatures from -320°F to +500°F (-196°C to +260°C), such as down-hole logging
tools, adhesive, and epoxy dispensing equipment, and chemical and laboratory
equipment. GFP has excellent chemical compatibility. This material is compatible with
most fluids and gases, except some acids, such as sulfuric, nitric and hydrofluoric acids.
LIQUID CONTACT MATERIALS CONTINUED ON PAGE 86
Hastelloy C This is the material most often recommended for corrosion resistance. This versatile
nickel-chromium molybdenum alloy has excellent resistance to most acids, including
Materials
strong oxidizers such as ferric and cupric chlorides; nitric, formic and acetic acids; wet
chlorine; sea water and brine solutions; and mixtures containing nitric acid or oxidizing
acids with chloride ions.
The best choice for most special applications where HPLC grade stainless cannot be
used, Hastelloy C has excellent resistance to pitting, stress corrosion cracking, and
oxidizing atmospheres up to temperatures well beyond any other standard components
of the chromatographic system.
Kalrez® Kalrez® perfluoroelastomer parts (FFKM) resist over 1,800 different chemicals, while
offering the high temperature stability of PTFE (327°C).
They are used in highly aggressive chemical processing, semiconductor wafer
fabrication, pharmaceutical, oil and gas recovery, and aerospace applications. The
long-term, proven performance of Kalrez® parts can mean less frequent seal changes,
repairs and inspections, increasing process and equipment uptime for greater
productivity and yield.
Ketron® CA30 PEEK Ketron® CA30 PEEK is a polyetheretherketone material with 30% carbon fiber reinforced.
The addition of carbon fibers enhances the compressive strength and stiffness of PEEK,
and dramatically lowers its expansion rate. It offers designers optimum wear resistance
and load carrying capability in a PEEK-based product. This grade provides more thermal
conductivity than unreinforced PEEK, increasing heat dissipation from bearing surfaces
improving bearing life and capability.
Materials
member of the fluorocarbon family with similar chemical properties. It is generally
more rigid than PTFE, with somewhat increased tensile strength. It is typically more
transparent than PTFE, slightly less porous, and less permeable to oxygen. FEP is not as
subject to compressive creep at room temperature as PTFE, and because of its slightly
higher coefficient of friction is easier to retain in a compression fitting.
PTFE Polytetrafluoroethylene is the generic name for the class of materials such as Teflon®.
It offers superior chemical resistance but is limited in pressure and temperature
capabilities. Because it’s so easy to handle, it is often used in low pressure situations
where stainless steel might cause adsorption. PTFE tubing is relatively porous, and
compounds of low molecular weight can diffuse through the tubing wall.
Ruby / Sapphire Synthetic rubies and sapphires are single-crystal aluminum oxides, practically pure for
the sapphire (+99,99% Al2O3). The color of the ruby is produced by adding a few ppm
(parts per million) of chromium oxide (CrO3). Synthetic rubies and sapphires have a
hexagonal-rhombic crystal structure, density of 3.99 g/cm³ and a water absorption
coefficient of 0%.
The principal properties of synthetic rubies and sapphires include a hardness and high
mechanical strength, excellent resistance to wear, very low friction coefficient, chemically
inert, good thermal conductivity, ideal electrical insulation.
LIQUID CONTACT MATERIALS CONTINUED ON PAGE 88
Stainless Steel, Stainless Steel 316 is a chromium-nickel based steel that possesses increased levels
Type 316/316L of resistance against several substances, due to the addition of molybdenum in its
Materials
Zirconium Oxide The zirconium oxide or zirconia (ZrO2) has a tetragonal crystal structure with a grain size
not exceeding 0.50 microns, density greater than 6,00 g/cm³, and a Vickers hardness
value of about 1200.
The principal properties of zirconia ZrO2 include a high mechanical strength, fracture
toughness, extreme hardness, compressive strength, low thermal conductivity,
excellent resistance to corrosion and wear, excellent tribological properties and good
biocompatibility.
IN THIS CHAPTER:
as possible after receipt. Keep packing materials in case the instrument needs to be shipped in
the future.
System Setup
Follow the instructions to set up the GX-241 Liquid Handler and VERITY® 4020 Single Syringe Pump
as described in each instrument’s user’s guide.
Do not connect the transfer tubing as instructed. This is a system connection that will be described
NOTE
later in this section.
Control Software
Control of the PLC Purification System with the GX-241 as autosampler with
VERITY 4020 Syringe Pump is via GLIDER Software accessed via the touchscreen of the
PLC Purification System.
90 PLC with GX-241 Liquid Handler Configuration | PLC 2050/2250/2500 Purification Systems
Installation
Install the GX-241 and the VERITY 4020 Syringe Pump as described in each instrument’s
user’s guide. Complete instructions for each step are included in the ‘Installation’ chapter. The
instrument and its components should be set up in the order shown below.
25025344 Syringe 5 mL
25025345 Syringe 10 mL
25025346 Syringe 25 mL
Embedded
D
Detector
K 1
2
C F 6
3
5
E 4
B
A
G
Embedded Pump
System
L
H
J
Figure 73
Plumbing Diagram for the PLC Purification System with GX-241 Liquid Handler
and VERITY® 4020 Single Syringe Pump
Transfer tubing
FROM Syringe pump outlet • 5.5 mL for 5 mL syringe
(PN 499671112)
B • 10.5 mL for 10 mL syringe PEEK
Port 2 (PN 499474103)
TO Automated sample loop • 30 mL for 25 mL syringe
injection valve (PN 499483602)
PLC PURIFICATION SYSTEM WITH GX-241 LIQUID HANDLER PLUMBING CONNECTIONS CONTINUED ON PAGE 93
92 PLC with GX-241 Liquid Handler Configuration | PLC 2050/2250/2500 Purification Systems
PLC Purification System with GX-241 Liquid Handler Plumbing Connections
Connection Tubing Fittings
Port 5
FROM Automated sample loop Varies, based on
Varies, based on PLC model and
E injection valve PLC model and
connected column
connected column
TO Column
Transfer Tubing
Part Number Syringe Size Description
1. Connect one end with 1/4”-28 fitting to the transfer tubing side of the valve of the
VERITY 4020 Syringe Pump (right-hand port), and finger-tighten the nut.
2. Connect the other end to the outlet of the shut-off valve, connected to the waste outlet
(port 2) of the automatic loop injection valve of the PLC Purification System and finger-
tighten the nut.
Right Left
Side Side
OR
C
E
3
2
+1 0 9
4 5 6
4 5 6
8
7
A B 3
2
7 F
8
109
G
VERITY® 4020 GX-241 PLC
Single Syringe Liquid Handler Purification
Pump System
Figure 74
Electrical Diagram for the PLC Purification System with GX-241 Liquid Handler
and VERITY® 4020 Single Syringe Pump
94 PLC with GX-241 Liquid Handler Configuration | PLC 2050/2250/2500 Purification Systems
GX-241, VERITY 4020 Syringe Pump, and
PLC Purification System Electrical Connections
Connection Description
A FROM VERITY 4020 Syringe Pump External power supply and power cord (appropriate
C FROM VERITY 4020 Syringe Pump USB cable, B-type end (square)
TO Power outlet
G Unit ID
GX-241
The GX-241 is shipped configured for RS-232 communication with a unit ID of 8, which is the
supported configuration in the PLC Purification System.
To make the RS-232 connection between the GX-241 Liquid handler and the PLC Purification
System, use the RS-232 cable supplied.
1. Attach the male end of the RS-232 cable to the RS-232 port located on the rear panel of the
liquid handler. Tighten the retaining screws.
2. Attach the female end of the RS-232 cable to the RS-232 serial communication port (COM4)
located on the right panel of the PLC system. Tighten the retaining screws.
UNIT ID
At the factory, the unit ID on the liquid handler is set to 30. The selector on the rear panel has
the arrow aligned with 0. The unit ID is 30 plus the selected number. There is no need to change
this number.
Z-ARM CONNECTION
Connect the cable from the Z-arm to the Z-ARM port on the rear panel and then use a small,
flat-blade screwdriver to tighten the screws to secure it
with the B-type (square) connector to connect to the VERITY 4020 Syringe Pump.
Power Connections
Use the power cord on the external power supply to make the connection between the power
receptacle on the GX-241 Liquid Handler and the external power supply.
The connection from the external power supply to the GX-241 Liquid Handler uses a connector
with a locking collar. Check the alignment of the pins and then push in until it clicks and locks in
place. To disconnect, pull back on the locking collar and then disconnect the cable from the rear
panel of the GX-241 Liquid Handler.
Locate the appropriate power cord for the line voltage and then make the connection between
the external power supply and the AC power source.
Repeat the previous operation for the power connection of the VERITY 4020 Syringe Pump and
its external power supply.
Start Up
Follow all prior instructions to make all rear panel and plumbing connections.
To start the GX-241 Liquid Handler and the VERITY 4020 Syringe Pump:
1. Power on the external supply, and then power on the GX-241 Liquid Handler, using the
power switch on the rear panel.
● The green power indicator light on the front panel illuminates.
● The two-digit front panel display indicates 0 (no error).
2. Power on the VERITY 4020 Syringe Pump, using the power switch on the side panel.
● The green power indicator light on the front panel illuminates.
● The syringe pump initializes. It stops with the valve set to the outlet (transfer tubing)
position.
3. Power on the PLC Purification System and wait for GLIDER to start.
96 PLC with GX-241 Liquid Handler Configuration | PLC 2050/2250/2500 Purification Systems
Operational Description of the VERITY 4020 Syringe Pump
The VERITY 4020 Syringe Pump has a single syringe and valve. The instrument only dispenses
to the GX-241 probe via the automatic loop injection valve of the PLC Purification System.
IN THIS CHAPTER:
Control Software
Control of the PLC Purification System with VERITY 1920 MS, VERITY 3011 Pump and MRA Splitter
is via GLIDER Software accessed via the touchscreen of the PLC Purification System.
System Setup
Install each system component per the instructions provided in each instrument’s user’s guide
and in the site requirements document.
Install the rotary vane pump underneath the bench or fume hood when possible. If a
diaphragm pump is used, place it on top of the rotary vane pump, at the back. The MRA Splitter
can be stacked on top of the VERITY 3011 Pump.
Position the chromatographic column (HPLC or CPC) between the PLC Purification System and
VERITY 1920 MS system.
Figure 78
Stacking Diagram
Electrical Connections
The rear panel of the VERITY 1920 MS includes connections for power and nitrogen gas. The front
panel includes vacuum exhaust, house exhaust, waste outlet, and a USB communication port.
The rear panel of the VERITY 3011 Pump includes connections for power and communication.
The USB port is used for the communication with the PLC Purification System.
The rear panel of the MRA Splitter includes connections for power and communication. The
RS-232 port is used for the communication with the PLC Purification System.
100 PLC with VERITY 1920 MS Configuration | PLC 2050/2250/2500 Purification Systems
Communication Connections
USB
VERITY 1920 MS
The USB port on the VERITY 1920 MS is used for remote control with GLIDER.
RS-232
MRA Splitter
The RS-232 port on the MRA Splitter is used for remote control with GLIDER.
To make the RS-232 connection between the MRA Splitter and the PLC Purification System,
use the RS-232 gender changer and the USB/RS-232 converter cable supplied. Connect the
gender changer to the RS-232 port located on the rear panel of the MRA Splitter and tighten the
retaining screws. Connect the male RS-232 end of the converter cable to the gender changer
and tighten the retaining screws. Connect the USB A-type end of the converter cable to the
PLC Purification System.
Power Connections
Do not turn on the instruments until instructed.
Exhaust Connections
Vacuum Exhaust
The vacuum exhaust tubing is already connected inside the VERITY 1920 MS and goes out
from the front left of the instrument. This large, spiral-supported 1.3" OD- 1" ID tubing must be
connected to the rotary vane pump (clamp connection).
House Exhaust
The house exhaust tubing is already connected inside the VERITY 1920 MS and goes out from
the front right of the instrument. This spiral-supported 1" OD–3/4" ID tubing must be connected
to the Ion Source Exhaust Trap, the 4L bottle with cap and adapters (barbed connection).
carried out using a splitter valve set up with the correct split ratio.
The outlet from the column is directed to the HPLC stream inlet of the MRA Splitter, where
the fluid stream is split between the integrated UV detector and the VERITY 1920 MS. The
HPLC stream outlet of the MRA Splitter is redirected to the UV detector bulkhead union of the
PLC Purification System.
The VERITY 3011 Pump is used as the make-up pump and its outlet is connected to the MS stream
inlet of the MRA Splitter. The VERITY 1920 ACTIVE SPLITTER KIT (PN 18005005) includes all
necessary tubing to connect the MS stream outlet of the MRA Splitter to the VERITY 1920 MS.
Use 1 m of the red PEEK tubing (PN 499500591) to connect the MRA Splitter to the flow injection
analysis (FIA) injection valve and use 0.5 m of the same tubing to connect the FIA injection valve
to the ion source.
The waste outlet tubing of the VERITY 1920 MS is already connected inside and goes out from
the front right of the instrument. Place the end of this Tygon® 3/8”OD-1/4” ID tubing into suitable
waste receptacle.
The following diagram shows and the table explains the plumbing connections between system
components.
3
2 5
A B
1 F E
4
D
Make Up MRA
Solvent Splitter
C 6
G
7
H
CPC
PLC Purification System
VERITY 1920 MS
VERITY 3011 Pump
Figure 79
System Plumbing Connections
102 PLC with VERITY 1920 MS Configuration | PLC 2050/2250/2500 Purification Systems
System Plumbing Connections
ID DESCRIPTION
3 MRA Splitter to integrated UV/VIS detector or DAD in the PLC Purification System
Cut the tubing to the appropriate length at installation:
PEEK tubing, green, 1/16" (OD), 0.030"/0.75mm (ID), 50 ft/15 m (included in the semi-prep tubing kit
PN 18005100)
or
PEEK tubing, natural, 1/16" (OD), 0.040"/1.0 mm (ID), 50 ft/15 m (included in the prep tubing kit
PN 18005101)
This tubing corresponds to the “Tubing Kit” in GLIDER.
communication established, there is no communication error and the system is ready to operate.
For more information about how to control a VERITY 1920 MS and PLC Purification System using
GLIDER, refer to the instructions document on the documentation USB drive shipped with the
system.
104 PLC with VERITY 1920 MS Configuration | PLC 2050/2250/2500 Purification Systems
Appendix F
IN THIS CHAPTER:
Data rate Up to 30 Hz
Gain setting Dynamic gain management: SAGA (SEDEX Automated Gain Adjustment)
Light source Selected high efficiency blue LED (470 mm) with elapsed-time counter
Liquid
Nebulizer
Inlet Large
Black
Nut
Shield
Holder
Small Nebulization
Gas
Black Chamber
Inlet
Nut
Shield
Holder
Nebulization
Chamber Nebulizing
Gas
Inclined
Tubing
Support
Figure 80 Figure 81
ELSD Components ELSD Installed on PLC
Shield
Figure 82
Shield Installed over ELSD Components
Figure 83
Rear Panel Components Diagram for PLC with ELSD
A Gas
B Exhaust
Highly purified gas is not required to operate the ELS detector; however, a good quality of gas
NOTICE is mandatory for high performance detection. The gas should be free from particles (dust) and
The gas supply must include a pressure gauge that is stable and regulated by an external
manometer. The ELS detector is typically operated with a gas pressure of 2 bar (29 psi) and
gas consumption of 3 L/min. A gas regulator (sold separately, PN 21040205) with a 5 µm filter,
manometer, and fittings is available for purchase.
To connect the PLC Purification System to the gas supply:
1. Connect a suitable length of clean, 6 mm (OD) tubing from the gas regulator outlet to the
gas inlet (Port A).
2. Connect a suitable length of clean, 6 mm (OD) tubing from the gas source to the gas
regulator inlet.
3. Insert the tube into the fitting until it reaches the back, and then pull the tube to check
engagement of the grab.
Ensure that the gas pressure supplied to the detector is less than 4.5 bar (67 psi). If the
NOTICE pressure increases above 4.5 bar, the pressure sensor may be damaged (not covered by
the warranty).
Ensure that the fume hood draws gas from the detector, i.e., there is positive pressure between
the detector and the fume hood.
Use a moderate vacuum to avoid turbulence in the nebulization chamber siphon or to prevent
liquid from spilling into the evaporation tube.
Avoid loops or bends in the black exhaust tubing to avoid condensation traps that can result
NOTICE
in poor measurement results.
Negative pressure between the fume hood and ELS detector can introduce contaminants
from the fume hood into the ELS detector.
Connect the exhaust tubing to the fume hood. If necessary, use the two adjustable, self-adhesive
clips supplied to guide the vent tube.
Figure 84
Exhaust Tubing and Self-Adhesive Clips
Evaporation Evaporation
Tube Cap Tube
Small
Black
Nut
Inclined
Support
Figure 85
Removing the Evaporation Tube Cap Figure 86
Positioning the Nebulization Chamber
6. Fill the siphon of the nebulization chamber with the mobile phase that will be used for the
separation. The liquid should fill the bend in the siphon, but not pool in the evaporation
tube. Ensure that no liquid leak could affect the detector performance or create laboratory
pollution.
Figure 87
Fill Area for Siphon Overflow
Figure 88
Siphon Overflow Connections
Ensure that the waste container is directly below the siphon outlet and that there are no
loops or bends in the drain tubing.
Affix the drain tubing to the inlet of the waste container, so that the end of the tubing
never submerges into the liquid.
NOTICE
If the solvent you are using is not compatible with the drain tube, such as
tetrahydrofuran (THF), use a piece of PTFE tubing or any material compatible with your
solvent in its place. When using this type of tubing, which is generally more rigid, ensure
that it is safely installed, so that it will not damage the nebulization chamber (glassware).
Gas
Tubing
Gas
Outlet
Fitting
Figure 89
Nebulization Gas Supply Connections
To install the chamber shield slide the chamber shield into the exposed notches of the two shield
holders.
Shield
Holders
Figure 91
Installing the Chamber Shield
Operation
Start Up
Make all rear and front panel connections before start up.
To use the ELS detector:
7. Power on the PLC Purification System and wait for GLIDER to load.
● If the PLC Purification System and the ELS detector are properly configured and
communication is established, a second detector cell with two green dots appears on
the software screen.
● The ELS detector warms to reach the default temperature selected in the
‘System Configuration’ window of GLIDER.
5 0.08 1.62
10 0.29 2.87
20 0.70 3.51
50 1.58 3.15
75 2.04 2.72
Split Flow Rates and Ratios with 90% Water - 10% Acetonitrile (ACN)
Elution Flow Rate (mL/min) Split Flow Rate in ELSD (mL/min) Split Ratio (%)
5 0.12 2.32
10 0.24 2.44
20 0.47 2.33
50 0.80 1.60
75 1.01 1.34
Closing the gas supply while the pump is still running may result in serious nebulizer
damage.
NOTICE
If the ELS detector will not be used for one hour, close the shut-off valve on the splitter to
avoid clogging the nebulizer or depositing foreign matter into the detector.
After each run and before shutting down the PLC Purification System, the ELS detector should
be cleaned to ensure good performance. Follow the preventative maintenance steps below to
clean the ELS detector after the final analysis:
1. Allow the mobile phase to continue flowing after analysis to flush any remaining particles
from the detector. The mobile phase should not contain any additive or buffer.
2. Incrementally increase the temperature to dissolve any remaining deposits.
3. Stop the mobile phase flow, but allow the gas to flow for at least 30 minutes to prevent
particle deposit.
● If needed, stop the gas flow at the source.
● If needed, shut down the system.
The time required for each step depends on the application, so solvent type and sample
concentration should be determined accordingly. Access the inside of the instrument is not
required for routine operation. If the suggestions provided in this chapter do not remedy the
NOTE problem, contact your local Gilson representative. Refer to Customer Service on page 32.
The LED has a long but finite lifetime (5000 hours) and should be replaced periodically by
trained personnel. Decreasing light intensity from a failing LED will result in decreasing
signal heights. Contact your local Gilson representative to replace the LED.
Never disassemble the nebulizer. Disassembling the rear part of the nebulizer (at the level of
1. Check to ensure that the liquid level in the siphon is appropriate, and there is no liquid
pooling close to the evaporation tube inlet.
2. Check that the gas pressure is sufficient and stable. The selected pressure for most
applications is 2 bar (29 psi) and gas consumption is 3 L/min. Pressure above 4.5 bar
(67 psi) can damage the pressure sensor. The gas filter should be clean and in place. Only
use gas free of particle and oil residue.
3. Ensure that the flow rate of the pump is constant and check that there are no leaks in the
PLC system.
4. The mist from the nebulizer should be homogeneous. If it is not homogeneous, the nebulizer,
the needle or the PTFE tube may be partially obstructed. To remove the obstruction, pump
a solvent that can dissolve the foreign material. As an alternative, the nebulizer can be
placed in an ultrasonic bath to dissolve the foreign material. Instructions about cleaning of
the nebulizer are given on next page.
5. If the sensitivity of the detector is low, ensure that there are no leaks in the system. Make
sure you are using a fresh sample and consider running the test using a backpressure loop
instead of a column. Alternatively, the light source may need to be replaced or the nebulizer
could be obstructed.
6. A decrease in the sensitivity is often caused by the nebulizer (main cause). Clean the
nebulizer. If the sensitivity does not return to normal, the nebulizer might need to be
replaced. Please note that the root cause might also be in different module, i.e. volumes
injected by the autosampler might be too low or dead volumes in capillary connections
may cause peak broadening.
7. If the detector signal is saturated or if there is a decrease in the dynamic range of the
system, it is possible that a residue is passing through the detector cell. This will lead to an
intense signal due to a significant amount of light-scattering. This residue may be a result of
the elution of strongly retained materials from the column, or may come from the solvent.
To determine the cause of the problem, bypass the column and observe the signal intensity.
● If the signal returns to normal, strongly retained materials are eluting from the column.
Flush the column with a strong solvent to elute all material.
● If the signal does not return to normal, the solvent contains a too high residue material,
after evaporation and is not suitable for use with the detector.
8. If the noise of the detector without solvent is high or if ghost peaks occur, it is possible that
foreign material is present in the drift tube. In this situation, increase the temperature to 95°C
and pump appropriate solvent at the rate of 2 mL/min, using a gas pressure of 2.0 bar (29 psi).
The solvent will be determined by the nature of the samples that were previously analyzed with
the detector. If you do not know the nature of the sample, ethanol is a good choice. Do not
use solvents that can potentially corrode the instrument. Maintain the flow and temperature
during three hours at least.
of use and care. This section provides general directions for nebulizer maintenance, but if cleaning
procedures are ineffective, consider a nebulizer replacement.
Handle the nebulizer carefully, and do not disassemble the rear component protected by a
white, heat-shrinking tubing. Improper handling of the nebulizer will permanently damage it.
NOTICE The nebulizer rear part results from a very tricky setting which must not be disassembled
for any reason. If it has been removed or unscrewed, the only solution is to proceed to a
nebulizer replacement.
Replace the nebulizer if the cleaning and preventative maintenance steps provided are
NOTICE
ineffective.