Astm D5453 19

this is fgor it specially

Uploaded by

freeguys2010
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what

changes have been made to the previous version. Because


it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.

Designation: D5453 − 16´1 D5453 − 19

Standard Test Method for


Determination of Total Sulfur in Light Hydrocarbons, Spark
Ignition Engine Fuel, Diesel Engine Fuel, and Engine Oil by
Ultraviolet Fluorescence1
This standard is issued under the fixed designation D5453; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

ε1 NOTE—Subsection 1.1 was corrected editorially in June 2016.

1. Scope*
1.1 This test method covers the determination of total sulfur in liquid hydrocarbons, boiling in the range from approximately
25 °C to 400 °C, with viscosities between approximately 0.2 cSt and 20 cSt (mm2/s) at room temperature.
1.2 Three separate interlaboratory studies (ILS) on precision, and three other investigations that resulted in an ASTM research
report, have determined that this test method is applicable to naphthas, distillates, engine oil, ethanol, Fatty Acid Methyl Ester
(FAME), and engine fuel such as gasoline, oxygen enriched gasoline (ethanol blends, E-85, M-85, RFG), diesel, biodiesel,
diesel/biodiesel blends, and jet fuel. Samples containing 1.0 mg ⁄kg to 8000 mg ⁄kg total sulfur can be analyzed (Note 1).
NOTE 1—Estimates of the pooled limit of quantification (PLOQ) for the precision studies were calculated. Values ranged between less than 1.0 mg ⁄kg
and less than 5.0 mg/kg (see Section 8 and 15.1).

iTeh Standards
1.3 This test method is applicable for total sulfur determination in liquid hydrocarbons containing less than 0.35 % (m ⁄m)
halogen(s).

(https://standards.iteh.ai)
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
Document
applicability of regulatory limitations prior Preview
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
to use. For warning statements, see 3.1, 6.3, 6.4, Section 7, and 8.1.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
ASTM D5453-19
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
https://standards.iteh.ai/catalog/standards/sist/9b015069-d212-41c2-afa3-0e67d7cbc8b9/astm-d5453-19
2. Referenced Documents
2.1 ASTM Standards:2
D1250 Guide for Use of the Petroleum Measurement Tables
D1298 Test Method for Density, Relative Density, or API Gravity of Crude Petroleum and Liquid Petroleum Products by
Hydrometer Method
D4052 Test Method for Density, Relative Density, and API Gravity of Liquids by Digital Density Meter
D4057 Practice for Manual Sampling of Petroleum and Petroleum Products
D4177 Practice for Automatic Sampling of Petroleum and Petroleum Products
D6299 Practice for Applying Statistical Quality Assurance and Control Charting Techniques to Evaluate Analytical Measure-
ment System Performance
3. Summary of Test Method
3.1 A hydrocarbon sample is either directly injected or placed in a sample boat. The sample or boat, or both, is inserted into
a high temperature combustion tube where the sulfur is oxidized to sulfur dioxide (SO2) in an oxygen rich atmosphere. Water

1
This test method is under the jurisdiction of ASTM Committee D02 on Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of Subcommittee
D02.03 on Elemental Analysis.
Current edition approved April 15, 2016May 1, 2019. Published April 2016June 2019. Originally approved in 1993. Last previous edition approved in 20122016 as
D5453 – 12.D5453 – 16ɛ1. DOI: 10.1520/D5453-16E01.10.1520/D5453-19.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at [email protected]. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.

*A Summary of Changes section appears at the end of this standard


Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

1
D5453 − 19

iTeh Standards
(https://standards.iteh.ai)
Document Preview
FIG. 1 Conventional Combustion Tubes

produced during the sample combustion is removed and ASTM D5453-19


the sample combustion gases are next exposed to ultraviolet (UV) light.
The SO2 absorbs the energy from the UV light and is converted to excited sulfur dioxide (SO2*). The fluorescence emitted from
https://standards.iteh.ai/catalog/standards/sist/9b015069-d212-41c2-afa3-0e67d7cbc8b9/astm-d5453-19
the excited SO2* as it returns to a stable state, SO2, is detected by a photomultiplier tube and the resulting signal is a measure of
the sulfur contained in the sample. (Warning—Exposure to excessive quantities of ultraviolet (UV) light is injurious to health. The
operator must avoid exposing any part of their person, especially their eyes, not only to direct UV light but also to secondary or
scattered radiation that is present.)
4. Significance and Use
4.1 Some process catalysts used in petroleum and chemical refining can be poisoned when trace amounts of sulfur bearing
materials are contained in the feedstocks. This test method can be used to determine sulfur in process feeds sulfur in finished
products, and can also be used for purposes of regulatory control.
5. Apparatus
5.1 Furnace—An electric furnace held at a temperature (1050 °C min to 1150 °C max) sufficient to pyrolyze all of the sample
and oxidize sulfur to SO2. The actual temperature to be recommended by the specific apparatus manufacturer.
5.2 Combustion Tube—A quartz combustion tube constructed to allow the direct injection of the sample into the heated
oxidation zone of the furnace or constructed so that the inlet end of the tube is large enough to accommodate a quartz sample boat.
The combustion tube shall have one or more side arms for the introduction of oxygen and carrier gas or air. The oxidation section
shall be large enough (see Fig. 1) to ensure complete combustion of the sample. Fig. 1 depicts conventional combustion tubes.
Other configurations are acceptable if precision is not degraded.
5.3 Flow Control—The apparatus shall be equipped with a means of flow control that is capable of maintaining a constant
supply of oxygen and carrier gas or air.
5.4 Drier Tube—The apparatus must be equipped with a mechanism for the removal of water vapor. The oxidation reaction
produces water vapor which must be eliminated prior to measurement by the detector. This can be accomplished with a membrane
drying tube, or a permeation dryer, that utilizes a selective capillary action for water removal.

2
D5453 − 19

iTeh Standards
(https://standards.iteh.ai)
FIG. 2 Direct Inject Syringe Drive

5.5 UV Fluorescence Detector—A qualitative and quantitative detector capable of measuring light emitted from the fluorescence
of sulfur dioxide by UV light.
Document Preview
5.6 Microlitre Syringe—Capable of accurately delivering from 5 µL to 90 µL of test specimen. Check with the instrument
manufacturer for specific test specimen volume and needle size requirements.
5.7 Sample Inlet System—Either of two types of sampleASTM D5453-19
inlet systems can be used.
https://standards.iteh.ai/catalog/standards/sist/9b015069-d212-41c2-afa3-0e67d7cbc8b9/astm-d5453-19
5.7.1 Direct Injection—A direct injection inlet system must be capable of allowing the quantitative delivery of the material to
be analyzed into an inlet carrier stream which directs the sample into the oxidation zone at a controlled and repeatable rate. A
syringe drive mechanism which discharges the sample from the microlitre syringe at a rate of approximately 1 µL/s is typical.
Check with the instrument manufacturer for specific test specimen injection rates. For example, see Fig. 2.
5.7.2 Boat Inlet System—An extended combustion tube provides a seal to the inlet of the oxidation area and is swept by a carrier
gas. The system provides an area to position the sample carrying mechanism (boat) at a retracted position removed from the
furnace. The boat drive mechanism will fully insert the boat into the hottest section of the furnace inlet. The sample boats and
combustion tube are constructed of quartz. The combustion tube provides a cooling jacket for the area in which the retracted boat
rests awaiting sample introduction from a microlitre syringe. A drive mechanism which advances and withdraws the sample boat
into and out of the furnace at a controlled and repeatable rate is required. For example, see Fig. 3.
5.8 Refrigerated Circulator—An adjustable apparatus capable of delivering a coolant material at a constant temperature as low
as 4 °C could be required when using the boat inlet injection method (optional).
5.9 Strip Chart Recorder, (optional).
5.10 Balance, with a precision of 60.01 mg (optional).
6. Reagents
6.1 Purity of Reagents—Reagent grade chemicals shall be used in tests. Unless otherwise indicated, it is intended that all
reagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, where
such specifications are available.3 Other grades may be used, provided it is first ascertained that the reagent is of sufficiently high
purity to permit its use without lessening the accuracy of the determination.

3
Reagent Chemicals, American Chemical Society Specifications, American Chemical Society, Washington, DC. For Suggestions on the testing of reagents not listed by
the American Chemical Society, see Annual Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia and National
Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, MD.

3
D5453 − 19

iTeh Standards
(https://standards.iteh.ai)
FIG. 3 Boat Inlet System

Document Preview
6.2 Carrier Gas—Inert gas or air. The actual gas to be recommended by the specific apparatus manufacturer.
6.2.1 Inert Gas—Argon or helium only, high purityASTM D5453-19
grade (that is, chromatography or zero grade), 99.998 % minimum purity,
moisture 5 mg ⁄kg maximum.
https://standards.iteh.ai/catalog/standards/sist/9b015069-d212-41c2-afa3-0e67d7cbc8b9/astm-d5453-19
6.2.2 Air—High purity grade (that is, chromatography or zero grade), 99.99 % minimum purity, moisture 5 mg ⁄kg maximum.
6.3 Oxygen (as required)—High purity (that is, chromatography or zero grade), 99.75 % minimum purity, moisture 5 ppm w/w
maximum, dried over molecular sieves. (Warning—Vigorously accelerates combustion.)
6.4 Toluene, Xylenes, Isooctane, reagent grade (other solvents similar to those occurring in samples to be analyzed are also
acceptable). Correction for sulfur contribution from solvents (solvent blank) used in standard preparation and sample specimen
dilution is required. Alternatively, use of a solvent with nondetectable level of sulfur contamination relative to the sulphur content
in the sample unknown makes the blank correction unnecessary. (Warning—Flammable solvents.)
6.5 Dibenzothiophene, FW184.26, 17.399 % (m/m) S (Note 2).
6.6 Butyl Sulfide, FW146.29, 21.92 % (m/m) S (Note 2).
6.7 Thionaphthene (Benzothiophene), FW134.20, 23.90 % (m/m) S (Note 2).
NOTE 2—A correction for chemical impurity can be required.
6.8 Quartz Wool, or other suitable absorbent material that is stable and capable of withstanding temperatures inside the furnace
(see Note 3).
NOTE 3—Materials meeting the requirements in 6.8 provide a more uniform injection of the sample into the boat by wicking any remaining drops of
the sample from the tip of the syringe needle prior to introduction of the sample into the furnace. Consult instrument manufacturer recommendations for
further guidance.
6.9 Sulfur Stock Solution, 1000 µg S/mL—Prepare a stock solution by accurately weighing approximately 0.5748 g of
dibenzothiophene or 0.4562 g of butyl sulfide or 0.4184 g of thionaphthene into a tared 100 mL volumetric flask. Dilute to volume
with selected solvent. This stock can be further diluted to desired sulfur concentration (Notes 4-7).
NOTE 4—Working standards that simulate or match the composition or matrix of the samples analyzed can reduce test result bias between direct inject
and boat sample inlet systems.

4
D5453 − 19
TABLE 1 Typical Operating Conditions
Syringe drive (direct inject) drive rate (700–750) 1 µL/s
Boat drive (boat inlet) drive rate (700–750) 140 mm/min – 160 mm/min
Furnace temperature 1075 °C ± 25 °C
Furnace oxygen flowmeter setting (3.8–4.1) 450 mL ⁄min –500 mL/min
Inlet oxygen flowmeter setting (0.4–0.8) 10 mL ⁄min –30 mL/min
Inlet carrier flowmeter setting (3.4–3.6) 130 mL ⁄min –160 mL/min

NOTE 5—Working standards should be remixed on a regular basis depending upon frequency of use and age. Typically, stock solutions have a useful
life of about three months.
NOTE 6—Calibration standards can be prepared and diluted on a mass/mass basis when result calculations are adjusted to accommodate them.
NOTE 7—Calibration standards from commercial sources can be used if checked for accuracy and if precision is not degraded.
6.10 Quality Control (QC) Samples, preferably are portions of one or more liquid petroleum materials that are stable and
representative of the samples of interest. These QC samples can be used to check the validity of the testing process as described
in Section 14.
6.11 Oxidation Reagent (as required)—Tungsten trioxide (WO3), granular (typical particle size >2.0 mm), high purity, 99.75 %
minimum.
7. Hazards
7.1 High temperature is employed in this test method. Extra care must be exercised when using flammable materials near the
oxidative pyrolysis furnace.
8. Sampling
8.1 Obtain a test unit in accordance with Practice D4057 or Practice D4177. To preserve volatile components which are in some
iTeh Standards
samples, do not uncover samples any longer than necessary. Samples shall be analyzed as soon as possible after taking from bulk
supplies to prevent loss of sulfur or contamination due to exposure or contact with sample container. (Warning—Samples that are

(https://standards.iteh.ai)
collected at temperatures below room temperature can undergo expansion and rupture the container. For such samples, do not fill
the container to the top; leave sufficient air space above the sample to allow room for expansion.)

Document Preview
8.2 If the test unit is not used immediately, then thoroughly mix in its container prior to taking a test specimen.
9. Preparation of Apparatus
9.1 Assemble and leak check apparatus according to manufacturer’s instructions.
ASTM D5453-19
9.2 Adjust the apparatus, depending upon the method of sample introduction, to meet conditions described in Table 1.
https://standards.iteh.ai/catalog/standards/sist/9b015069-d212-41c2-afa3-0e67d7cbc8b9/astm-d5453-19
9.3 Adjust the instrument sensitivity and baseline stability and perform instrument blanking procedures following manufac-
turer’s guidelines.
10. Calibration and Standardization
10.1 Based on anticipated sulfur concentration, select one of the suggested curves outlined in Table 2. Narrower ranges than
those indicated may be used, if desired. However, the test method precision using narrower ranges than those indicated have not
been determined. Ensure the standards used for calibration bracket the concentrations of the samples being analyzed. Carefully
prepare a series of calibration standards accordingly. Make other volumetric dilutions of the stock solution to cover the various
ranges of operation within these calibration curve guidelines. The number of standards used per curve can vary, if equivalent results
are obtained.
10.2 Flush the microlitre syringe several times with the sample prior to analysis. If bubbles are present in the liquid column,
flush the syringe and withdraw a new sample.
10.3 A sample injection size recommended for the curve selected from Table 2 shall be quantitatively measured prior to
injection into the combustion tube or delivery into the sample boat for analysis (Notes 8-10). There are two alternative techniques
available.
NOTE 8—Injection of a constant or similar sample size for all materials analyzed in a selected operating range promotes consistent combustion
conditions.
NOTE 9—Injection of 10 µL of the 100 ng ⁄µL standard would establish a calibration point equal to 1000 ng or 1.0 µg.
NOTE 10—Other injection sizes can be used when complete sample combustion is not compromised and accuracy/precision are not degraded.
10.3.1 The volumetric measurement of the injected material can be obtained by filling the syringe to the selected level. Retract
the plunger so that air is aspirated and the lower liquid meniscus falls on the 10 % scale mark and record the volume of liquid in
the syringe. After injection, again retract the plunger so that the lower liquid meniscus falls on the 10 % scale mark and record
the volume of liquid in the syringe. The difference between the two volume readings is the volume of sample injected (Note 11).

5
D5453 − 19
TABLE 2 Typical Sulfur Calibration Ranges and Standard
Concentrations
Curve I Curve II Curve III
Sulfur, ng/µL Sulfur, ng/µL Sulfur, ng/µL
0.50 5.00 100.00
1.00 25.00 500.00
2.50 50.00 1000.00
5.00 100.00
10.00
Injection Size Injection Size Injection Size
10 µL – 20 µL 5 µL – 10 µL 5 µL

NOTE 11—An automatic sampling and injection device can be used in place of the described manual injection procedure.
10.3.2 Fill the syringe as described in 10.3.1. Weigh the device before and after injection to determine the amount of sample
injected. This procedure can provide greater accuracy than the volume delivery method, provided a balance with a precision of
60.01 mg is used.
10.4 Once the appropriate sample size has been measured into the microlitre syringe, promptly and quantitatively deliver the
sample into the apparatus. Again, there are two alternative techniques available.
10.4.1 For direct injection, carefully insert the syringe into the inlet of the combustion tube and the syringe drive. Allow time
for sample residues to be burned from the needle (Needle Blank). Once a stable baseline has reestablished, promptly start the
analysis. Remove syringe once the apparatus has returned to a stable baseline.
10.4.2 For the boat inlet, quantitatively discharge the contents of the syringe into the boat containing quartz wool or suitable
equivalent (see 6.8) at a slow rate being careful to displace the last drop from the syringe needle. Remove the syringe and promptly
start the analysis. The instrument baseline shall remain stable until the boat approaches the furnace and vaporization of the sample

iTeh Standards
begins. Instrument baseline is to be reestablished before the boat has been completely withdrawn from the furnace (Note 12). Once
the boat has reached its fully retracted position, allow at least 1 min for cooling before the next sample injection (Note 12).

(https://standards.iteh.ai)
NOTE 12—Slowing boat speed or briefly pausing the boat in the furnace can be necessary to ensure complete sample combustion. Direct injection can
ease sample handling and improve sample combustion characteristics for materials containing very volatile sulfur compounds.

Document Preview
10.4.3 The level of boat cooling required and the onset of sulfur detection following sample injection are directly related to the
volatility of the materials analyzed. For volatile materials, effective cooling of the sample boat prior to sample injection is essential.
The use of a refrigerated circulator to minimize the vaporization of the sample until the boat begins approaching the furnace or
an increased time for boat cooling can be required.
ASTMtwo
10.5 Calibrate the instrument using one of the following D5453-19
techniques.
https://standards.iteh.ai/catalog/standards/sist/9b015069-d212-41c2-afa3-0e67d7cbc8b9/astm-d5453-19
10.5.1 Perform measurements for the calibration standards and blank using one of the procedures described in 10.2 – 10.4.
Measure the calibration standards and blank three times. Subtract the average response of the blank injections from each calibration
standard response. Then determine the average integrated response of each concentration (see 6.4). Construct a curve plotting of
the average integrated detector response ( y-axis) versus micrograms of sulfur injected (x-axis) (Note 13). This curve shall be linear
and system performance must be checked each day of use. See Section 14.
NOTE 13—Other calibration curve techniques can be used when accuracy and precision are not degraded.
10.5.2 If the apparatus features self calibration routine, measure the calibration standards and blank three times using one of the
procedures described in 10.2 – 10.4. If blank correction is required and is not an available instrument option (see 6.4 or 10.5.1),
calibrate the analyzer in accordance with manufacturer’s instructions to yield results expressed as nanograms of sulfur (Note 13).
This curve shall be linear and system performance must be checked with each day of use (see Section 14).
10.6 If analyzer calibration is performed using a different calibration curve than listed in Table 2, select an injection size based
on the curve closest in concentration to the measured solution(s). Construct the calibration curve to yield values that can be used
to report sulfur content on a mass/mass basis.

11. Procedure
11.1 Obtain a test specimen using the procedure described in Section 8. The sulfur concentration in the test specimen must be
less than the concentration of the highest standard and greater than the concentration of the lowest standard used in the calibration.
If required, a dilution can be performed on either a weight or volume basis.
11.1.1 Gravimetric Dilution (mass/mass)—Record the mass of the test specimen and the total mass of the test specimen and
solvent.
11.1.2 Volumetric Dilution (mass/volume)—Record the mass of the test specimen and the total volume of the test specimen and
solvent.
11.2 Measure the response for the test specimen solution using one of the procedures described in 10.2 – 10.4.