12th Experiment 4p
12th Experiment 4p
EXPE RIMENT
A9
AIM
To convert the given galvanometer (of known resistance and figure of merit) into a
voltmeter of desired range (say 0 to 3 V) and to verify the same.
WORKING FORMULAE
1. In nis the total number of divisions on either side of the zero mark of the galvanometer and k its
figure of merit, then current required for full scale deflection in the galvanometer is , = nk
2. To convert the galvanometer of resistance Ginto a voltmeter of range 0-V volt, a high resistanceR
is connected in series with it so that only current I, flows through it, as shown in Fig. 6.12. From
Ohm's law,
OI R G
R+G
CIRCUIT DIAGRAMS
Battery K
Rh
ww
Voltmeter
R
w
volts
Lo - -
Converted voltmeter
Fig. 6.12 Conversion of galvanometer into a voltmeter. Fig. 6.13 Circuit to verify conversion of galvanometer into a voltmeter.
PROCEDURE
1. Record the values of the resistance G of the galvanometer and its figure of merit k.
2. Count the total number of divisions n on either side of the zero mark of the galvanometer scale.
3. Calculate the current I, required to produce full scale deflection in the galvanometer by using the
relation, I, = nk.
4. For conversion of galvanometer into voltmeter of range (0-V) volt, calculate the value of series
resistance R using the formula,
R G
74
Chopter 6
GALVANOMETER, AMMETER
:
AND VOLTMETE
5. nect
Connect aa
high resistance box of
h
a
resistance equal to the range 0- 10 K2 in series eter and
introduce
calculated
voltmeter of range (0--V) volt. value of R. The with the
gaia its USe
a
a
ga
ne galvanometer is now ready for
-V2)
Deflection V, =k'0 (V)
(divisions) (V)
(V)
75
LAB MANUAL PHYSICS-XII
RESULT
1. Series resistance R required for conversion of given galvanometer into v o l m e t e r or range 0 to
PRECAUTIONS
I h e standard voltmeter used should have nearly the same range as that required tor converted
voltmeter.
2. The resistance box used should be used of high resistance
3. Low resistance rheostat should be used as potential divider.
4.The exact high resistance (correct to the nearest ohm) equal to the calculated value should be
introduced from the resistance box connected in series with the galvanometer. Only after
introducing the high series resistance, the battery key should be closed to avoid any damage to the
galvanometer.
5. Both the standard and the converted voltmeters should be connected in parallel with the rheostat.
See that the positive marked terminal of the standard voltmeter is connected to the positive
terminal of the battery.
6. Zero error, if any, of the galvanometer should be eliminated or accounted for in the calculations,
SOURCES OF ERROR
1. The divisions marked on the galvanometer scale may not be of equal size.
2. Plugs in the resistance box may be loose or they may not be clean.
3. The emf of the battery may not be constarnt.