0% found this document useful (0 votes)
428 views4 pages

CL-162 Calibration Accreditation Notice

This document provides accreditation for AL KAABI FOR TECHNOLOGY SUPPORT to calibrate various measurement devices. It lists the measurement capabilities and uncertainties for calibrating things like scales, pressure gauges, viscometers, and nuclear density gauges.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
428 views4 pages

CL-162 Calibration Accreditation Notice

This document provides accreditation for AL KAABI FOR TECHNOLOGY SUPPORT to calibrate various measurement devices. It lists the measurement capabilities and uncertainties for calibrating things like scales, pressure gauges, viscometers, and nuclear density gauges.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

This is to attest that

AL KAABI FOR TECHNOLOGY SUPPORT


P.O. BOX 4875
EASTERN PROVINCE 31412, KINGDOM OF SAUDI ARABIA

Calibration Laboratory CL-162

has met the requirements of AC204, IAS Accreditation Criteria for Calibration Laboratories, and has demonstrated
compliance with ISO/IEC Standard 17025:2017, General requirements for the competence of testing and calibration
laboratories. This organization is accredited to provide the services specified in the scope of accreditation.
Effective Date October 22, 2023
Expiration Date February 1, 2025

President
AL KAABI FOR TECHNOLOGY SUPPORT
[Link]

Contact Name Shadi Nofal Contact Phone + 966-138584005


Accredited to ISO/IEC 17025:2017 Effective Date October 22, 2023

CALIBRATION AND MEASUREMENT CAPABILITY (CMC)*


MEASURED RANGE UNCERTAINTY1,2 CALIBRATION METHOD OR
QUANTITY or DEVICE (±) PROCEDURE, STANDARD
TYPE CALIBRATED EQUIPMENT (OPTIONAL)
Dimensional
Dial Gauge / 0 mm to 25 mm 6 µm Dial Gauge Calibrator
Dial Indicators (JIS B7503 & BS 907)
Calipers Up to 300 mm 18 µm Gauge Blocks Grade 0
(ASME B89.1.14)
Fine Sieves 20 µm to 1.7 mm 20 µm Microscope
(ASTM E11)
Coarse Sieves 2.0 mm to 125 mm 0.5 mm Digital Caliper
(ASTM E11)
Mechanical
Scales and Balances Up to 220 g 0.2 mg Class E2 Weights
Up to 720 g 8.0 mg (ASTM E898)
Up to 6 kg 25 mg
Up to 16 kg 52 mg
Up to 36 kg 90 mg Class F1 Weights
(ASTM E898)
Up to 1000 kg 0.8 kg Class M1 Weights
(ASTM E898)
Pressure Up to 200 bar 0.11 bar Dead Weight Tester
201 bar to 400 bar 0.19 bar (BS/EN 837-1)
401 bar to 800 bar 0.29 bar
801 bar to 1000 bar 0.44 bar
Vacuum -0.95 bar to 0 bar 14 mbar Test Pump/ Digital Master
Gauge
(BS/EN 837-1)
Mass (Standard Weights) 20 kg 130 g Precision balance / Class E2
Class M1, M2 & M3 10 kg 65 g standard weights
5 kg 25 g
2 kg 27 g (OIML R111)
1 kg 27 g

* If information in this CMC is presented in non-SI units, the conversion factors stated in NIST Special Publication
811 “Guide for the Use of the International System of Units (SI)” apply.

CL-162 Effective Date October 22, 2023


Al Kaabi for Technology Page 2 of 4
IAS/CL/100-3
Support
MEASURED RANGE UNCERTAINTY1,2 CALIBRATION METHOD OR
QUANTITY or DEVICE (±) PROCEDURE, STANDARD
TYPE CALIBRATED EQUIPMENT (OPTIONAL)
Batch Plants 0 kg to 1000 kg 0.8 kg Class M1 Weights
1001 kg to 2000 kg 1.8 kg as per the requirements of
2001 kg to 3000 kg 2.8 kg ASTM Standard C94/94M and
3001 kg to 4000 kg 4.0 kg Handbook 44 – 2019, Section
4001 kg to 5000 kg 5.3 kg 2.22. Automatic Bulk
5001 kg to 6000 kg 6.7 kg Weighing Systems Clause
6001 kg to 7000 kg 8.2 kg N.1.1.2
7001 kg to 8000 kg 9.9 kg
8001 kg to 9000 kg 12 kg
9001 kg to 10000 kg 14 kg

Parameters of Gyratory Digital Dynamometer


Compactor Height Standard
Dynamic Angle Validator
Force 0.25 kN to 25 kN 0.23 % (DAV) Model: II HMS

Height 114.36 mm 41 µm (ASHTO T312)


(ASTM D6925)
Angle 1.36º 0.005º
Rotational Speed 10 rpm to 28800 rpm 0.3 % Photo Tachometer
(KTS/WI/033)
(KTS/WI/025)
(KTS/WI/031)

Force - Compression 0.1 kN to 10 kN 0.36 % 10 kN Load Cell


(BS/EN/ISO 7500)
1 kN to 100 kN 0.13 % 100 kN Load Cell
(BS/EN/ISO 7500)
7.5 kN to 750 kN 0.13 % 750 kN Load Cell
(BS/EN/ISO 7500)
50 kN to 5000 kN 0.13 % 5000 kN Load Cell
(BS/EN/ISO 7500)

Force - Tension 0.1 kN to 10 kN 0.4 % 10 kN Load Cell


(BS/EN/ISO 7500)
1 kN to 100 kN 0.25 % 100 kN Load Cell
(BS/EN/ISO 7500)
7.5 kN to 750 kN 0.23 % 750 kN Load Cell
(BS/EN/ISO 7500)
10 kN to 1000 kN 0.1 % 1000 kN Load Cell
(BS/EN/ISO 7500)
Load Cell (Compression) 0 kN to 500 kN 0.15 % Master Load Cell and Master
501 kN to 1500 kN 0.14 % Digital Indicator
1501 kN to 5000 kN 0.13 % (ISO 376:2011)

CL-162 Effective Date October 22, 2023


Al Kaabi for Technology Page 3 of 4
IAS/CL/100-3
Support
MEASURED RANGE UNCERTAINTY1,2 CALIBRATION METHOD OR
QUANTITY or DEVICE (±) PROCEDURE, STANDARD
TYPE CALIBRATED EQUIPMENT (OPTIONAL)
Viscometer Up to 12,500 cP 96 cP Certified Viscosity Standard
(ASTM D2196 & AASHTO
T316)
Nuclear Density Gage 569.34 kg/m3 7.1 % Mg/Polyethylene Standard
Moisture Block
(ASTM D6938 & ASTM
D7759)
1784.82 kg/m3 0.44 % Magnesium Standard Block
(ASTM D6938 & ASTM
D7759)
2191.75 kg/m3 0.46 % Mg/Al Standard Block
(ASTM D6938 & ASTM
D7759)
2732.53 kg/m3 0.16 % Aluminum Standard Block
(ASTM D6938 & ASTM
D7759)
Dynamic Shear Rheometer 133,800 mPa·s to 1400 mPa·s Reference Certified Silicon
(DSR) 621,800 mPa·s Fluid
(ASTM D7175)
Thermal
Thermometers – SIKA TPM 165S-U
Digital 0 ºC to 160 ºC 0.16 ºC Multifunction Calibrator

Dial Type 0 ºC to 160 ºC 1.2 ºC ASTM E2877-12

Infrared 0 ºC to 160 ºC 2 ºC ASTM E2847


Ovens and Furnaces Up to 160 ºC 0.6 ºC Thermocouple Type K and
Up to 800 ºC 1.1 ºC Multifunction Simulator
(ASTM E145)
Chemical/Gas
pH Meter 0 pH to 14 pH 0.08 pH Certified Buffer Solutions
(KTS/WI/009)
TDS / Conductivity Meter Up to 1417 µS/cm 2.3 µS/cm Conductivity Solutions
(KTS/WI/009)

1The uncertainty covered by the Calibration and Measurement Capability (CMC) is expressed as the expanded
uncertainty having a coverage probability of approximately 95 %. It is the smallest measurement uncertainty that a
laboratory can achieve within its scope of accreditation when performing calibrations of a best existing device. The
measurement uncertainty reported on a calibration certificate may be greater than that provided in the CMC due to the
behavior of the calibration item and other factors that may contribute to the uncertainty of a specific calibration.
2When uncertainty is stated in relative terms (such as percent, a multiplier expressed as a decimal fraction or in scientific
notation), it is in relation to instrument reading or instrument output, as appropriate, unless otherwise indicated.

CL-162 Effective Date October 22, 2023


Al Kaabi for Technology Page 4 of 4